. 2020 Dec 31;27(6):897–898. doi: 10.5603/CJ.2020.0173

Successful optical coherence tomography-guided stent ablation with rotational atherectomy for an underexpanded stent

Yongcheol Kim 1, Deok-Kyu Cho 1,, Ji Woong Roh 1, Oh-Hyun Lee 1, Eui Im 1, Donghoon Choi 1
PMCID: PMC8079103  PMID: 33432568

A 62-year-old man with a history of ischemic heart disease and dyslipidemia presented with aggravating effort angina. He underwent stent implantation with a 3.0 × 40 mm sirolimus-eluting stent at the mid portion of left anterior descending artery 2 years prior, but a heavily calcified lesion led to severe stent underexpansion, which did not resolve despite postdilation with a non-compliant (NC) balloon (Suppl. Video 1). Angiography demonstrated aggravated luminal narrowing in the underexpanded stent site (Fig. 1A, Suppl. Video 2); therefore, stent ablation with rotational atherectomy was planned. Pre-interventional optical coherence tomography (OCT) demonstrated that the diameter of the underexpanded stent was between 1.24 and 1.66 mm (Fig. 1B, C, Suppl. Video 3), which was also confirmed on three-dimensional OCT (Fig. 1D). OCT assessment led to stent ablation using a stepwise increase in burr size, from 1.25 mm to 1.50 mm and finally 1.75 mm. Following stent ablation, OCT showed no visible strut area in the site of the previously underexpanded stent (Fig. 1E, F). However, a minimal lumen area of 2.27 mm2 led to the performance of drug-coated balloon angioplasty with a 3.0 × 20 mm Pantera Lux (Biotronik, Bülach, Switzerland) after balloon dilation with a 3.0 × 15 mm NC balloon at 24 atm. Final angiography showed good distal flow without residual stenosis (Fig. 1G, Suppl. Video 4).

Figure 1.

Figure 1

A. Initial angiography demonstrating underexpanded stent (white box); B, C. Cross-sectional optical coherence tomography (OCT) imaging of underexpanded stent site; D. Three-dimensional OCT reconstructed strut image of underexpanded stent site; E, F. OCT demonstrating no visible struts after stent ablation with rotational atherectomy; G. Final angiography demonstrating no residual stenosis after treatment with 3.0 × 20 mm drug-coated balloon.

This report highlights the superior resolution of OCT, which can aid in choosing the optimal burr size, and presents OCT images of successful stent ablation.

Supplementary Information

Download video file (1.9MB, wmv)
Download video file (463.2KB, wmv)
Download video file (2.7MB, wmv)
Download video file (285.1KB, wmv)

Acknowledgements

The authors would like to thank all the staff, especially Dae Seok Jang, working in the cardiac catheterization laboratories at Yongin Severance Hospital for their commitment to this study.

Footnotes

Conflict of interest: None declared

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Download video file (1.9MB, wmv)
Download video file (463.2KB, wmv)
Download video file (2.7MB, wmv)
Download video file (285.1KB, wmv)

Articles from Cardiology Journal are provided here courtesy of Via Medica sp. z o.o. sp. k.

RESOURCES