Abstract
Background
The PSP (predilatation, sizing, post-dilatation)-technique was developed to improve the prognosis of patients after bioresorbable vascular scaffold (BVS) implantation. In acute coronary syndrome (ACS) the use of BVS is particularly demanding and carries some potential risk regarding aggressive lesion preparation, proper vessel sizing due to spasm and thrombus inside the artery. The aim herein, was to determine the long-term results of BVS stenting in ACS patients depending on the scaffold implantation technique.
Methods
The present study is a prospective, two-center study, which consisted of 182 patients who underwent percutaneous coronary intervention (PCI) with BVS (Absorb, Abbott Vascular, Santa Clara, California, USA) implantation for the ACS. All patients were divided into two groups. The first consisted of 52 patients treated with the PSP-technique (PSP group). The second group enrolled 130 patients treated with a non-PSP procedure (non-PSP group).
Results
The procedure was successful in all patients. The mean observation time was 28.8 ± 16.5 months (median 28.3 months, interquartile range 24.0 [17.0–41.0] months). It was found that target vessel failure (TVF) was consistently reduced in patients using the PSP-technique as compared with the non-PSP group (5.8% vs. 17.7%, p = 0.03). Moreover, PSP-technique was superior to non-PSP-technique concerning major adverse cardiac events (MACE) (3.7% vs. 22.3%, p = 0.02). Logistic regression analysis revealed that the use of PSP technique significantly decreased the risk of target vessel revascularization (odds ratio [OR] 0.11, p = 0.01), TVF (OR 0.28, p = 0.03) and MACE (OR 0.29, p = 0.02).
Conclusions
The PSP-technique for BVS implantation improves long-term results and should also be recommended for newer generations of the bioresorbable scaffold.
Keywords: acute coronary syndrome, acute myocardial infarction, STEMI, NSTEMI, angiography, coronary, bioresorbable devices/polymers
Introduction
Bioresorbable vascular scaffolds (BVSs) are a first-generation technology introduced to overcome the limitations of metallic stents [1, 2]. Unfortunately, recent reports of randomized trials revealed several negative results compared with drug eluting stents (DESs) [3–5], especially a higher rate of target-vessel myocardial infarction (MI) and scaffold thrombosis [6]. Thick struts of BVS delay endothelialization, correlate with flow disturbance and, in consequence, increase the risk of scaffold thrombosis [7]. Different constructions and mechanical properties make the proper choice of scaffold diameter and its implantation crucial to the results of the procedure. The recent studies have focused on optimal pre-dilatation, sizing of the vessel and post-dilatation to improve treatment results. Ortega-Paz et al. [8] presented the predictive value of PSP (predilatation, sizing, post-dilatation) scores on clinical outcomes. It was an independent predictor of a 1-year device-oriented composite endpoint composed of cardiac death, target vessel MI, and clinically driven target lesion revascularization (TLR). However, the use of BVS and its implantation using PSP-technique in acute coronary syndrome (ACS), the most prothrombotic form of atherosclerosis, is demanding and carries some potential risk regarding aggressive lesion preparation, proper vessel sizing due to spasm and thrombus inside the artery. Moreover, BVS has raised concerns regarding over-expansion, disruption, and the effect of post-dilatation following implantation [9, 10].
Evidence regarding optimal BVS implantation technique in ACS remains limited. These data would be useful in subsequent generations of bioresorbable scaffolds. The aim of the study is to determine results of BVS stenting in ACS depending on scaffold implantation technique.
Methods
Study design
In this prospective, two-center study, a total of 182 patients were consecutively selected who underwent percutaneous coronary intervention (PCI) with BVS (Absorb, Abbott Vascular, Santa Clara, California, USA) implantation for ACS between December 2012 and October 2015. Eligible patients were hemodynamically stable with left ventricular ejection fraction > 30% and had a life expectancy of at least 5 years. In angiography, they had at least one significant coronary artery stenosis, with no restrictions as to the number, severity or lesion location. Patients were divided into two groups, depending on implantation technique. The first consisted of 52 patients treated with the PSP-technique (PSP group). The second group enrolled 130 patients treated with a non-PSP procedure (non-PSP group). In this group, predilatation was performed in 120 (92.3%) and 17 (13.1%) in post-dilatation patients, respectively.
Patients excluded from the study were with: cardiogenic shock, the life expectancy of less than 1 year. The use of metallic stents during the index procedure and the target vessel reference diameter were < 2.3 mm and > 3.7 mm by visual estimate. Detailed exclusion criteria are presented in Table 1.
Table 1.
Exclusion criteria.
| Known intolerance to acetylsalicylic acid, heparin, Poly L-lactic acid, everolimus, contrast material Active bleeding or coagulopathy or patients on chronic anticoagulation therapy Poor compliance Severe tortuous, calcified or angulated coronary anatomy of the study vessel Fibrinolysis prior to percutaneous coronary intervention |
Ethics approval was obtained from the Institutional Review Committees in each institution. The study was performed following ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments.
Implantation technique
The PCI procedure was performed according to current PCI guidelines. PSP technique is the recommended optimal implantation method of ABSORB BVS. The definition was derived from the GHOST registry and included three steps: predilatation, proper vessel sizing, and post-dilatation. In PSP group, these implantation criteria were met in all patients. Predilatation was performed using non-compliant (NC) balloon 1:1 ratio with reference vessel diameter (RVD) to obtain optimal lesion preparation. The alternative balloons (scoring or cutting) were considered if NC balloon was not completely expanded. Proper scaffold sizing was based on angiography guidance and online quantitative coronary angiography (QCA) according to RVD. During implantation, the balloon was inflated slowly with 2 atmospheres every 5 s, maintaining the final pressure for 20 s in the scaffold. Post-dilatation was carried out with an NC balloon > 1:1 ratio with RVD up to 0.5 mm at ≥ 16 atmospheres to confirm the full expansion of the scaffold and optimize overlap zone. In non-PSP group pre- and post-dilatation were at the discretion of the operator, however, were performed according to the principles of PSP technique.
Peri- and post-procedural pharmacotherapy
Each patient naive to antiplatelet therapy, received a loading dose of 300 mg acetylsalicylic acid followed by the maintenance daily dose of 75 mg and one of the following: clopidogrel 600 mg (n = 97; 53.3%), prasugrel 60 mg (n = 1; 0.5%), or ticagrelor 180 mg (n = 84; 46.2%) in loading doses before or immediately after PCI, followed by a maintenance dose of clopidogrel (75 mg o.d.), prasugrel (10 mg o.d.), or ticagrelor (90 mg twice daily) for a minimum of 12 months. The decision about the continuation of dual antiplatelet therapy (DAPT) after 12 months was made individually for the patient depending on risk of thrombosis. A bolus of unfractionated heparin, 100 U/kg was administered intravenously during the procedure. The remaining pharmacotherapy was applied according to contemporary guidelines.
Data collection
All data were collected in an electronic database. Clinical follow-up was obtained 30 days, 6 months, 1 year and every following year after the procedure by direct contact with patients or telephone interview, and additionally a review of medical reports if the patient had been hospitalized.
Patients were monitored the for the following endpoints: death, MI, scaffold thrombosis, TLR, target vessel revascularization (TVR) and target vessel failure (TVF), defined as cardiac death, target vessel MI, and TVR. Additionally, cumulative major adverse cardiac events (MACE) rate, composed of cardiac death, non-fatal infarction or reintervention were analyzed.
Definitions
ST-segment elevation myocardial infarction (STEMI) was defined as an electrocardiographic ST-segment elevation concomitant with characteristic symptoms of myocardial ischemia and subsequent release of biomarkers of myocardial necrosis [11]. New or presumed new left bundle branch block has been considered a STEMI equivalent. Non-ST-segment elevation myocardial infarction (NSTEMI) definition involved the presence of angina chest pain with a marked elevation of myocardial necrosis biomarkers and no evidence of ST-segment elevation in the electrocardiogram (ECG). Unstable angina was diagnosed in patients with symptoms of myocardial ischemia and no troponin elevation, with or without ECG changes indicative of ischemia (e.g., ST-segment depression or transient elevation or new T wave inversion) [12]. Death was defined as all-cause mortality during the follow-up. Scaffold thrombosis was determined according to the Academic Research Consortium definition [13, 14]. TLR was set as a target segment reintervention including 5 mm proximal and distal to the scaffold.
Revascularization was indicated if symptoms of myocardial ischemia occurred, and positive stress test, electrocardiographic evidence of ischemia at rest, and/or > 70% diameter in-lesion stenosis on angiography were observed. A procedure was angiographically successful with residual diameter stenosis of less than 30% after scaffold implantation in combination with Thrombolysis in Myocardial Infarction (TIMI) III coronary flow. Procedure success was defined as angiographic success in the absence of in-hospital MACE.
Results
The baseline characteristics of the study groups are presented in Table 2. According to these data, clinical presentation and prevalence of cardiovascular risk factors did not differ between groups (for all, p > 0.05). In both, middle aged men with hypertension predominated. About one-third of patients suffered from diabetes mellitus. It was noticed that previous MI was significantly more often in the non-PSP group. In turn, the PSP group had more complex lesions, such as higher rate of left main disease, target bifurcation lesion, and significant calcification. There was a significantly higher rate of lesions of type B2/C in the non-PSP group (Table 3).
Table 2.
Baseline clinical characteristics.
| Characteristics | PSP-technique | Non-PSP-technique |
|---|---|---|
| N | 52 | 130 |
| Male | 34 (65.4%) | 94 (72.3%) |
| Age [years] | 60 ± 11 | 58 ± 11 |
| STEMI | 7 (13.5%) | 22 (16.9%) |
| NSTEMI | 11 (21.2%) | 37 (28.5%) |
| Unstable angina | 24 (46.2%) | 24 (53.8%) |
| Cardiovascular risk factors: | ||
| Hypertension | 34 (65.4%) | 114 (87.7%) |
| Diabetes mellitus | 9 (17.3%) | 32 (24.6%) |
| IDDM | 4 (7.7%) | 12 (9.2%) |
| Cardiovascular history: | ||
| Prior MI | 14 (26.9%) | 24 (18.5%)* |
| Prior CABG | 2 (3.8%) | 7 (5.4%) |
| Prior PCI | 13 (25.0%) | 35 (26.9%) |
| Chronic kidney disease | 4 (7.7%) | 12 (9.2%) |
p < 0.05, Mann-Whitney test or t-student test, as appropriate;
CABG — coronary artery bypass grafting; IDDM - insulin-dependent diabetes mellitus; MI — myocardial infarction; NSTEMI — non-ST-segment elevation myocardial infarction; PCI — percutaneous coronary intervention; STEMI — ST-segment elevation myocardial infarction
Table 3.
Baseline angiography characteristics.
| Characteristics | PSP-technique | Non-PSP-technique |
|---|---|---|
| Multivessel disease | 16 (30.8%) | 71 (55.0%) |
| Target vessel location: | ||
| LM | 4 (7.7%) | 0 (0.0%)* |
| LAD | 26 (50.0%) | 62 (47.7%) |
| RCA | 10 (19.2%) | 30 (23.1%) |
| LCX | 8 (15.4%) | 22 (16.9%) |
| Other | 4 (7.7%) | 16 (12.3%) |
| Lesion type B2/C | 32 (61.5%) | 128 (98.5%)* |
| Calcification | 5 (9.6%) | 1 (0.8%)* |
| Bifurcation lesion | 14 (26.9%) | 14 (10.8%)* |
| Thrombus | 3 (5.8%) | 5 (3.8%) |
| RVD [mm] | 3.1 ± 0.4 | 2.91 ± 0.4 |
| MLD [mm] | 0.4 ± 0.2 | 0.31 ± 0.2 |
| Diameter stenosis [%] | 87.3 ± 8.2 | 88.01 ± 6.7 |
| Quantitative coronary analysis | 34 (65.4%) | 9 (6.8%) |
| Visual estimate | 18 (34.6%) | 121 (93.2%) |
| Total number of scaffolds | 61 | 149 |
| Mean scaffolds per lesion | 1.17 | 1.15 |
| Mean scaffold length per lesion | 27.2 ± 10.7 | 22.5 ± 11.1 |
| Mean scaffold diameter per lesion | 3.0 ± 0.4 | 3.0 ± 0.4 |
| Radial approach | 50 (96.1%) | 125 (96.1%) |
| Pre-dilatation | 52 (100%) | 120 (92.3%) |
| Mean pre-dilatation balloon diameter [mm] | 2.9 ± 0.5 | 2.7 ± 0.5 |
| Maximum pre-dilatation pressure [atm] | 13.1 ± 2.3 | 13.5 ± 1.2 |
| Post-dilatation | 52 (100%) | 2 (0.5%)* |
| Mean post-dilatation balloon diameter [mm] | 3.1 ± 0.8 | 3.5 ± 0.0 |
| Max post-dilatation pressure [atm] | 17.8 ± 2.4 | 16.0 ± 2.3 |
| Complications occurring any time during the procedure: | ||
| MACE | 0 (0.0%) | 0 (0.0%) |
| Dissection | 2 (1.5%) | 1 (1.9%) |
| Distal embolism | 0 (0.0%) | 0 (0.0%) |
| No-reflow | 0 (0.0%) | 0 (0.0%) |
| Angiographic success | 52 (100%) | 130 (100%) |
| Procedure success | 52 (100%) | 130 (100%) |
p < 0.05, Mann-Whitney test or t-student test, as appropriate;
LAD — left artery descending; LCx — left circumflex artery; LM — left main; MACE — major adverse cardiac events; MLD — minimal lumen diameter; RCA — right coronary artery; RVD — reference vessel diameter
Total length of the implanted scaffold was significantly higher in the PSP group compared with the non-PSP group (26.8 ± 12.5 mm vs. 22.5 ± 10.3 mm, p = 0.02). Quantitative coronary analysis demonstrated a significant more upper reference vessel diameter and lower post-procedural diameter stenosis in patients treated with PSP technique.
Procedural success was obtained in all patients. In 3 cases coronary dissection occurred and was successfully covered with an additional scaffold. No peri-procedural MACE were reported. Detailed angiographic characteristics are presented in Table 3.
Complete follow-up was available in 88.5% after 12 months, 83.5% after 24 months, and 63.2% after 36 months. The mean observation time was 28.8 ± 16.5 months (median 28.3 months, inter-quartile range 24.0 [17.0–41.0] months). The rate of all-cause death and cardiac death was similar in both groups. There was a trend to a higher incidence of MI and TLR in the non-PSP-technique group, however, it was not statistically significant. Scaffold thrombosis occurred only in 1 patient during hospitalization (definite sub-acute thrombosis). No further scaffold thrombosis occurred at follow-up. TVF was consistently reduced in patients using the PSP technique as compared with non-PSP-technique group (5.8% vs. 17.7%, p = 0.03). Moreover, PSP-technique was superior to non-PSP-technique concerning MACE (3.7% vs. 22.3%, p = 0.02) (Table 4).
Table 4.
Results.
| Characteristics | PSP-technique | Non-PSP-technique |
|---|---|---|
| All cause death | 3 (5.8%) | 4 (3.1%) |
| Cardiac death | 2 (3.8%) | 2 (1.5%) |
| Any MI | 1 (1.9%) | 10 (7.7%) |
| Target vessel MI | 1 (1.9%) | 6 (4.6%) |
| Scaffold thrombosis | 0 (0%) | 1 (0.8%) |
| Target lesion revascularization | 0 (0%) | 7 (5.4%) |
| Target vessel revascularization | 1 (1.9%) | 19 (14.6%)* |
| Target vessel failure | 3 (5.8%) | 23 (17.7%)* |
| MACE | 4 (3.7%) | 29 (22.3%)* |
p < 0.05, Mann-Whitney test, as appropriate;
MACE — major adverse cardiac events; MI — myocardial infarction
The logistic regression analysis revealed that use of the PSP technique significantly decreased the risk of TVR (odds ratio [OR] = 0.11, p = 0.01), TVF (OR = 0.28, p = 0.03) and MACE (OR = 0.29, p = 0.02).
Discussion
In this study, it was found that pre-dilatation, proper sizing, and post-dilatation, cold PSP-technique, improve long-term clinical results of bioresorbable absorb scaffolds in patients with ACS. PSP-technique reduces the risk of TVR, TVF, and MACEs by almost 8-fold.
Recently, preliminary results for BVS are not very encouraging. The main concerns regard thrombosis and restoration of vessel functionality at long-term follow-up. Data from the randomized ABSORB Japan (2 years), ABSORB III (2 years), ABSORB II (3 years), and AIDA (2-year mean follow-up) trials demonstrated a higher rate of very late scaffold thrombosis with BVS compared to CoCr-EES [3, 4, 15]. Increased strut thickness delays endothelialization and correlates with flow disturbance, increased risk of strut fracture and disruption because of overexpansion [16]. Additionally, BVSs are not as stretchable as metallic stents and cannot be expanded beyond specified limits. Due to these specific properties, implantation of BVS should be performed particularly carefully. PSP technique (precise pre-dilatation and vessel sizing before BVS implantation and post-dilation following implantation) was associated with lower risk of thrombotic events in context with a non-PSP technique [17]. Predilatation and proper vessel sizing increases the rate of successful device delivery and correct expansion. The data showed that a correct size of the vessel is the most critical determinant of event-free rate during the year subsequent to implantation [18, 19]. The MICAT authors suggested that very-late events could also be associated with a suboptimal sizing of the vessel [20]. In turn, an optimal post-dilation prevents adverse events by maximizing scaffold dimensions, embed struts into plaque, avoid acute malapposition, and reduce shear stress [19].
The PSP technique has been investigated for ABSORB BVS technology [21]. Firstly, it was considered as the five golden “P”s: prepare the lesion, properly size, pay attention to expansion limits, post-dilate with non-compliant balloon as well as pay attention to DAPT [22]. This concept was supported by a group of European experts in a consensus document regarding optimal implantation technique [16] and by results from the MICAT registry (The Coronary Slow-flow and Microvascular Diseases Registry) [20]. In this study, optimal implantation technique significantly reduced the rate of scaffold thrombosis. The post-hoc analysis of the GHOST-EU registry showed a reduction of device-oriented composite endpoint at 1-year follow-up when all three steps of the PSP technique were performed correctly [18].
Moreover, a pooled analysis of the ABSORB trials (ABSORB II, III, CHINA, JAPAN, and EXTEND) revealed that an optimal PSP-technique was strongly associated with clinical outcomes during long-term follow-up [19]. The rationale for the use of BVS in the setting of the ACS are data suggesting that implantation of a temporary scaffold is associated with stabilization of atherosclerotic plaque without a permanent metallic cage. According to recent data, the safety and clinical outcomes of BVS in ACS patients are comparable to that of modern DESs [23]. From a retrospective study, it is also known that scaffold thrombosis can be reduced, when appropriate BVS size, pre- and post-dilatation were employed [24]. In the present study, the rate of scaffold thrombosis was not negligible and occurred in only 1 patient in the non-PSP group who did not have post-dilatation.
Although aggressive lesion preparation improves the rate of successful device delivery, predilatation potentially increases the risk of plaque disruption, thrombus mobilization, and distal embolism [25]. Usually, it is recommended to use semi-or non-compliant balloons with a diameter 0.5 mm smaller or equal to the size of the planned device and characteristics [24]. In the present study, all lesions were pre-dilated in PSP group and 92.3% lesions in non-PSP groups without complications. In all these cases, manual thrombus aspiration was applied before pre-dilatation. The overall procedural success rate was 100%, including all cases with evident thrombus.
Due to the limited expansion and BVS sizes available, vessel sizing is crucial in performing accurate scaffold implantation, especially in patients with ACS. In this group, proper vessel sizing can be limited due to spasm and thrombus inside the artery [26, 27]. Scaffold diameter should be selected according to the reference vessel diameter. The gold standard of correct RVD estimation after proper pre-dilation, and excluded under expansion or malapposition is intravascular imaging [21]. Tanaka et al. [28] reported that patients treated with intravascular imaging guidance, post-dilation balloon/scaffold ratio was higher and final residual percentage stenosis was lower compared with those treated with an angio-guided approach. However, despite the angio- and QCA guided PSP technique has several limitations, such as limited information of the atherosclerotic plaque composition, limited visibility of the scaffold in the angiography, difficulties in the estimation of RVD, and uncertainty of possible scaffold under expansion or malapposition, these techniques are used in the 3 steps of implantation in most patients. This results from the still limited availability of optical coherence tomography and intravascular ultrasound due to high cost. In the current study the QCA guided approach dominated in the PSP group (65.4%), and angiography-guided approach in non-PSP group (93.2%).
According to the recommended PSP technique, all scaffolds should be post-dilated with NC balloon. However, ACS patients have a potentially increased risk of over-expansion, disruption, and the effect of post-dilatation following its implantation [9, 10]. The ASSURE Registry (21.3% unstable angina) showed that a slight systematic oversizing of BVS, followed by high pressure post-dilatation, is safe and effective [29]. In turn, short-term results of the RAI registry (1,505 patients, 59% ACS) confirmed that high post-dilation rate (96.8%) might mitigate BVS-related events [30]. In a pooled analysis of the BVS Expand and BVS STEMI registries (351 patients, 72.6% ACS), post-dilation in ACS group was only 41.3% [31].
A comparison of BVS vs. everolimus eluting stent (EES) in STEMI patients with a high rate of post-dilation showed favorable mid-term results [32]. In the BVS STEMI first propensity score matching comparisons between 151 BVS patients and 151 EES patients, the MACE rate was higher in the BVS group (9.8% vs. 3.6%, p = 0.02, and TLR was 5.7% vs. 1.3%, p = 0.05) [33]. Interestingly, the 30-day MACE rate in BVS patients without post-dilatation was 6.8% and 3.6% in patients with post-dilatation. Of note, all BVS cases with acute scaffold thrombosis had no post-dilatation at the index procedure suggesting that optimization of the implantation technique is of paramount importance even in the acute setting. Imori et al. [17] also confirmed the importance of BVS post-dilation in the ACS setting. At 24 month follow-up, a higher rate of MACE was observed in BVS compared to EES in consecutive ACS patients before and after propensity score matching. However, after sensitivity analysis, MACE rates in BRS patients with post-dilation were significantly lower than in those without post-dilation and were comparable to EES patients (6.0% vs. 12.6% vs. 4.7%, p < 0.001). scaffold thrombosis rates were only slightly lower in the BVS group with post-dilatation, but were higher in both BVS groups than in EES patients (2.0% vs. 2.6% vs. 1.2%, p = 0.09).
Contrarily to the ABSORB III 2-year results, the investigators did not find any relation between clinical outcomes with either the implantation technique (74% BVS post-dilation rate) or the diameter of the treated vessels or the presenting symptoms. However, among the patients in the scaffold group who had definite or probable device thrombosis, 19% had a residual diameter stenosis of 30% or greater; among the patients who did not have device thrombosis, 9% had a residual percent diameter stenosis of 30% or greater (p = 0.05) highlighting the importance to obtain maximal BVS expansion at the end of the procedure. In the present study, post-dilatation was applied in all patients in the PSP group without complications and only in 2 (1.5%) patients in the non-PSP group.
Conclusions
The implantation of BVSs according to the PSP-technique reduced rates of TVR, TVF, as well as MACE, compared with non-PSP-technique implantation during long-term observation. The PSP-technique for BVS implantation improves long-term results and should also be recommended for newer generations of bioresorbable scaffold.
Footnotes
Conflict of interest: Maciej Lesiak has received payments as an individual for the advisory board and speaker honoraria from Abbott Vascular, Astra-Zeneca, Biotronik, Boston Scientific and Tryton Medical. Stefan Grajek has received payments as an individual on the advisory board and speaker honoraria from Astra-Zeneca, Servier, Pfizer, Sandoz, Adamed, Polpharma. Aleksander Araszkiewicz has received payments as an individual for the advisory board and speaker honoraria from Abbott Vascular. None of the other authors have relationships with Industry to declare.
References
- 1.Giacchi G, Ortega-Paz L, Brugaletta S, et al. Bioresorbable vascular scaffold implantation in acute coronary syndromes: clinical evidence, tips and tricks. Post Kardiol Interw. 2015;11(3):161–169. doi: 10.5114/pwki.2015.54006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Gomez-Lara J, Brugaletta S, Jacobi F, et al. Five-Year optical coherence tomography in patients with ST-segment-elevation myocardial infarction treated with bare-metal versus everolimus-eluting stents. Circ Cardiovasc Interv. 2016;9(10) doi: 10.1161/CIRCINTERVENTIONS.116.003670. [DOI] [PubMed] [Google Scholar]
- 3.Serruys PW, Chevalier B, Sotomi Y, et al. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet. 2016;388(10059)(16):2479–2491. 32050–5. doi: 10.1016/S0140-6736. [DOI] [PubMed] [Google Scholar]
- 4.Wykrzykowska JJ, Kraak RP, Hofma SH, et al. Bioresorbable Scaffolds versus Metallic Stents in Routine PCI. N Engl J Med. 2017;376(24):2319–2328. doi: 10.1056/NEJMoa1614954. [DOI] [PubMed] [Google Scholar]
- 5.Kereiakes DJ, Ellis SG, Metzger C, et al. 3-year clinical outcomes with everolimus-eluting bioresorbable coronary scaffolds: the absorb III trial. J Am Coll Cardiol. 2017;70(23):2852–2862. doi: 10.1016/j.jacc.2017.10.010. [DOI] [PubMed] [Google Scholar]
- 6.Ali ZA, Serruys PW, Kimura T, et al. 2-year outcomes with the Absorb bioresorbable scaffold for treatment of coronary artery disease: a systematic review and meta-analysis of seven randomised trials with an individual patient data substudy. Lancet. 2017;390(10096)(17):760–772. 31470–8. doi: 10.1016/S0140-6736. [DOI] [PubMed] [Google Scholar]
- 7.Kolandaivelu K, Swaminathan R, Gibson WJ, et al. Stent thrombogenicity early in high-risk interventional settings is driven by stent design and deployment and protected by polymer-drug coatings. Circulation. 2011;123(13):1400–1409. doi: 10.1161/CIRCULATIONAHA.110.003210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Ortega-Paz L, Brugaletta S, Sabaté M. Impact of PSP technique on clinical outcomes following bioresorbable scaffolds implantation. J Clin Med. 2018;7(2) doi: 10.3390/jcm7020027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.De Ribamar Costa J, Abizaid A, Bartorelli AL, et al. Impact of post-dilation on the acute and one-year clinical outcomes of a large cohort of patients treated solely with the Absorb Bioresorbable Vascular Scaffold. EuroIntervention. 2015;11(2):141–148. doi: 10.4244/EIJY15M05_06. [DOI] [PubMed] [Google Scholar]
- 10.Ormiston JA, Serruys PW, Regar E, et al. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet. 2008;371(9616)(08):899–907. 60415–8. doi: 10.1016/S0140-6736. [DOI] [PubMed] [Google Scholar]
- 11.Onuma Y, Serruys PW, Perkins LEL, et al. Intracoronary optical coherence tomography and histology at 1 month and 2, 3, and 4 years after implantation of everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: an attempt to decipher the human optical coherence tomography images in the ABSORB trial. Circulation. 2010;122(22):2288–2300. doi: 10.1161/CIRCULATIONAHA.109.921528. [DOI] [PubMed] [Google Scholar]
- 12.Thygesen K, Alpert JS, Jaffe AS, et al. Third universal definition of myocardial infarction. Circulation. 2012;126(16):2020–2035. doi: 10.1161/CIR.0b013e31826e1058. [DOI] [PubMed] [Google Scholar]
- 13.Laskey WK, Yancy CW, Maisel WH. Thrombosis in coronary drug-eluting stents: report from the meeting of the Circulatory System Medical Devices Advisory Panel of the Food and Drug Administration Center for Devices and Radiologic Health, December 7 – 8, 2006. Circulation. 2007;115(17):2352–2357. doi: 10.1161/CIRCULATIONAHA.107.688416. [DOI] [PubMed] [Google Scholar]
- 14.Cutlip DE, Windecker S, Mehran R, et al. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007;115(17):2344–2351. doi: 10.1161/CIRCULATIONAHA.106.685313. [DOI] [PubMed] [Google Scholar]
- 15.Onuma Y, Sotomi Y, Shiomi H, et al. Two-year clinical, angiographic, and serial optical coherence tomographic follow-up after implantation of an everolimus-eluting bioresorbable scaffold and an everolimus-eluting metallic stent: insights from the randomised ABSORB Japan trial. EuroIntervention. 2016;12(9):1090–1101. doi: 10.4244/EIJY16M09_01. [DOI] [PubMed] [Google Scholar]
- 16.Tamburino C, Latib A, van Geuns RJ, et al. Contemporary practice and technical aspects in coronary intervention with bioresorbable scaffolds: a European perspective. EuroIntervention. 2015;11(1):45–52. doi: 10.4244/EIJY15M01_05. [DOI] [PubMed] [Google Scholar]
- 17.Imori Y, D’Ascenzo F, Gori T, et al. Impact of postdilatation on performance of bioresorbable vascular scaffolds in patients with acute coronary syndrome compared with everolimus-eluting stents: A propensity score-matched analysis from a multicenter “real-world” registry. Cardiol J. 2016;23(4):374–383. doi: 10.5603/CJ.a2016.0052. [DOI] [PubMed] [Google Scholar]
- 18.Ortega-Paz L, Capodanno D, Gori T, et al. Predilation, sizing and post-dilation scoring in patients undergoing everolimus-eluting bioresorbable scaffold implantation for prediction of cardiac adverse events: development and internal validation of the PSP score. EuroIntervention. 2017;12(17):2110–2117. doi: 10.4244/EIJ-D-16-00974. [DOI] [PubMed] [Google Scholar]
- 19.Stone GW, Abizaid A, Onuma Y, et al. Effect of technique on outcomes following bioresorbable vascular scaffold implantation: analysis from the ABSORB trials. J Am Coll Cardiol. 2017;70(23):2863–2874. doi: 10.1016/j.jacc.2017.09.1106. [DOI] [PubMed] [Google Scholar]
- 20.Puricel S, Cuculi F, Weissner M, et al. Bioresorbable Coronary Scaffold Thrombosis: Multicenter Comprehensive Analysis of Clinical Presentation, Mechanisms, and Predictors. J Am Coll Cardiol. 2016;67(8):921–931. doi: 10.1016/j.jacc.2015.12.019. [DOI] [PubMed] [Google Scholar]
- 21.Absorb Bioresorbable Vascular Scaffold System [Internet] https://www.vascular.abbott:80/us/products/coronary-intervention/absorb-bioresorbable-scaffold-dissolving-stent.html (cited 2018 Jun 12 ).
- 22.Everaert B, Felix C, Koolen J, et al. Appropriate use of bioresorbable vascular scaffolds in percutaneous coronary interventions: a recommendation from experienced users : A position statement on the use of bioresorbable vascular scaffolds in the Netherlands. Neth Heart J. 2015;23(3):161–165. doi: 10.1007/s12471-015-0651-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Iwańczyk S, Hiczkiewicz J, Araszkiewicz A, et al. Evaluation of bioresorbable vascular scaffolds in acute coronary syndrome: A two-center, one-year follow-up analysis. Cardiol J. 2018;25(4):479–486. doi: 10.5603/CJ.a2017.0131. [DOI] [PubMed] [Google Scholar]
- 24.Ellis SG, Steffenino G, Kereiakes DJ, et al. Clinical, angiographic, and procedural correlates of acute, subacute, and late absorb scaffold thrombosis. JACC Cardiovasc Interv. 2017;10(18):1809–1815. doi: 10.1016/j.jcin.2017.06.067. [DOI] [PubMed] [Google Scholar]
- 25.Pan M, Romero M, Ojeda S, et al. Fracture of bioresorbable vascular scaffold after side-branch balloon dilation in bifurcation coronary narrowings. Am J Cardiol. 2015;116(7):1045–1049. doi: 10.1016/j.amjcard.2015.07.015. [DOI] [PubMed] [Google Scholar]
- 26.Spaulding C, Henry P, Teiger E, et al. Sirolimus-eluting versus uncoated stents in acute myocardial infarction. N Engl J Med. 2006;355(11):1093–1104. doi: 10.1056/NEJMoa062006. [DOI] [PubMed] [Google Scholar]
- 27.Stone GW, Lansky AJ, Pocock SJ, et al. Paclitaxel-eluting stents versus bare-metal stents in acute myocardial infarction. N Engl J Med. 2009;360(19):1946–1959. doi: 10.1056/NEJMoa0810116. [DOI] [PubMed] [Google Scholar]
- 28.Tanaka A, Latib A, Kawamoto H, et al. Clinical outcomes of a real-world cohort following bioresorbable vascular scaffold implantation utilising an optimised implantation strategy. EuroIntervention. 2017;12(14):1730–1737. doi: 10.4244/EIJ-D-16-00247. [DOI] [PubMed] [Google Scholar]
- 29.Wöhrle J, Naber C, Schmitz T, et al. Beyond the early stages: insights from the ASSURE registry on bioresorbable vascular scaffolds. EuroIntervention. 2015;11(2):149–156. doi: 10.4244/EIJY14M12_10. [DOI] [PubMed] [Google Scholar]
- 30.Cortese B, Ielasi A, Moscarella E, et al. Thirty-Day outcomes after unrestricted implantation of bioresorbable vascular scaffold (from the prospective RAI registry) Am J Cardiol. 2017;119(12):1924–1930. doi: 10.1016/j.amjcard.2017.03.017. [DOI] [PubMed] [Google Scholar]
- 31.Felix CM, Onuma Y, Fam JM, et al. Are BVS suitable for ACS patients? Support from a large single center real live registry. Int J Cardiol. 2016;218:89–97. doi: 10.1016/j.ijcard.2016.05.037. [DOI] [PubMed] [Google Scholar]
- 32.Farag M, Spinthakis N, Gorog DA, et al. Use of bioresorbable vascular scaffold: a meta-analysis of patients with coronary artery disease. Open Heart. 2016;3(2):e000462. doi: 10.1136/openhrt-2016-000462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Fam JM, Felix C, van Geuns RJ, et al. Initial experience with everolimus-eluting bioresorbable vascular scaffolds for treatment of patients presenting with acute myocardial infarction: a propensity-matched comparison to metallic drug eluting stents 18-month follow-up of the BVS STEMI first study. Eurointervention. 2016;12(1):30–37. doi: 10.4244/EIJV12I1A6.. [DOI] [PubMed] [Google Scholar]