Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2021 Apr 29;16(4):e0250942. doi: 10.1371/journal.pone.0250942

Development of multiplex real-time RT-PCR assay for the detection of SARS-CoV-2

Huseyin Tombuloglu 1,*, Hussein Sabit 1, Ebtesam Al-Suhaimi 2, Reem Al Jindan 3, Khaled R Alkharsah 3
Editor: Turgay Unver4
PMCID: PMC8084238  PMID: 33914804

Abstract

The outbreak of the new human coronavirus SARS-CoV-2 (also known as 2019-nCoV) continues to increase globally. The real-time reverse transcription polymerase chain reaction (rRT-PCR) is the most used technique in virus detection. However, possible false-negative and false-positive results produce misleading consequences, making it necessary to improve existing methods. Here, we developed a multiplex rRT-PCR diagnostic method, which targets two viral genes (RdRP and E) and one human gene (RP) simultaneously. The reaction was tested by using pseudoviral RNA and human target mRNA sequences as a template. Also, the protocol was validated by using 14 clinical SARS-CoV-2 positive samples. The results are in good agreement with the CDC authorized Cepheid`s Xpert® Xpress SARS-CoV-2 diagnostic system (100%). Unlike single gene targeting strategies, the current method provides the amplification of two viral regions in the same PCR reaction. Therefore, an accurate SARS-CoV-2 diagnostic assay was provided, which allows testing of 91 samples in 96-well plates in per run. Thanks to this strategy, fast, reliable, and easy-to-use rRT-PCR method is obtained to diagnose SARS-CoV-2.

Introduction

The outbreak of novel Betacoronavirus, SARS-CoV-2, which began in Wuhan, China in December 2019, has spread rapidly to multiple countries as a global pandemic. As of March 10, 2021, about 135 million people were confirmed with SARS-CoV-2 infection, and three million were died (https://www.worldometers.info/coronavirus/#countries). The increasing number of infections worldwide necessitates the need for a less-invasive, reliable, and fast diagnostic tool [1]. To achieve this goal, several studies have tackled this challenge and these efforts have yielded several diagnostic kits. Many approaches have been proposed to detect SARS-CoV-2 virus in nasopharyngeal fluids such as multiplex RT-PCR [2, 3], CRISPR/Cas12 [4, 5], and CRISPR-Cas3 [6], lateral flow immunoassay [7], paper-based biomolecular sensors [8], SHERLOCK testing in one pot [9], DNA aptamer [10], loop-mediated isothermal amplification (LAMP) [11], etc. Each of these methods has its own strong and weak points in terms of sensitivity and specificity. Among these methods, nucleic acid amplification-based tests are the most common for the diagnosis of SARS-CoV-2. To date, the US Food and Drug Administration (FDA) has approved 196 molecular diagnostic tests for the detection of SARS-CoV-2 nucleic acids as Emergency Use Authorization (EUA) (https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/vitro-diagnostics-euas). In addition, US CDC (The Centers for Disease Control and Prevention) suggest a protocol for the detection of SARS-CoV-2, based on the amplification of two regions of the nucleocapsid gene, namely N1 and N2, and a human internal control gene RNase P (RP) [12]. In addition to these strategies, efforts to develop SARS-CoV-2 detection methods with high efficiency and accuracy, with less reaction time and effort, are still ongoing.

SARS-CoV-2 is a positive-sense single-stranded RNA ((+) ssRNA) virus. Its genome consists of 29,900 nucleotides (nt) enclosing five open reading frames (ORFs) (5′–3′); ORF1ab polyprotein (P, 7,096 amino acids), spike glycoprotein (S, 1,273 amino acids), nucleocapsid protein (N, 419 amino acids), envelope protein (E, 75 amino acids), and membrane protein (M, 222 amino acids) [1, 13]. Therefore, several viral genes could be targeted for the detection of SARS-CoV-2 by RT-PCR methods. These genes include such as RdRP and S genes (Yan et al. 2020) [14], N and S genes [2], and E gene [15]. For the best RT-PCR performance, the combination of these potential gene targets should be optimized.

In the present study, it is aimed to develop multiplex real-time reverse transcription polymerase chain reaction (rRT-PCR) assay for the detection of SARS-COV-2. The assay simultaneously targets two viral genes (RdRP and E) and a human gene (RP) as an internal control by using the Applied Biosystems 7500 Fast Real-Time PCR instrument (ABI, Thermo Fisher Sci). In addition, the clinical performance of the assay was evaluated on SARS-CoV-2 samples collected from COVID-19 positive patients and compared by using the GeneXpert Dx instrument (Cepheid, Sunnyvale, CA, USA).

Materials and methods

Alignment of SARS-nCoV-19 genome sequences

Genomic sequences of all SARS-nCoV-19 types that have been sequenced worldwide were downloaded from the database of GISAID (Global Initiative on Sharing All Influenza Data, https://www.gisaid.org). The comparative analyses by aligning the sequences at a base level were made with bioinformatics programs such as NCBI BLAST [16], MUSCLE [17], and Clustal Omega [18]. More than 100 annotated genomes, whose genome sequence information have been determined and which include samples from Europe and Asia, were selected. In this way, virus gene targets are adjusted as sensitive, specific, and accurate as possible.

Multiplex primer / probe design

Multiplex PCR compatible primer and probe arrays that are specific to viral and human gene targets were designed using programs such as PrimerPooler [19], PrimerPlex (http://www.premierbiosoft.com/primerplex/index.html), and Primer3 [20]. 5’ Fluorescein amidites (FAM) and 3’ black hole quencher-1 (BHQ-1)-labeled probe for the viral RdRp gene, 5’ hexachloro-fluorescein (HEX) and 3’ BHQ-1-labeled probe for the viral E gene, and a 5’ carboxyrhodamine (ROX) and 3’ BHQ-2-laballed probe for the human RP gene were designed and synthesized (Fig 1A). The concentration and sequence of each primer or probe are stated in Table 1.

Fig 1. Experimental design and the cycle threshold (Ct) value of rRT-PCR repeats.

Fig 1

(a) Genome organization of SARS-CoV-2 and selected gene/probe sets. (b) Simplex assay targeting RdRP, E, and RP genes in separate reaction tubes. Multiplex or duplex assay amplifies two genes simultaneously: (c) RdRP and RP, (d) E and RP. (e) Multiplex or triplex assay that targets two viral (RdRP and E) and one human internal control (RP) gene simultaneously.

Table 1. The sequence and concentration of primer and probe sets used in PCR reactions.

Primer/probe Sequence (5`- 3`) Concentration (μM) Reference
RdRp—F GTCATGTGTGGCGGTTCACT 20 This study
RdRp—R CAACACTATTAGCATAAGCAGTTGT 20
RdRp- P FAM-CAGGTGGAACCTCATCAGGAGATGC- BHQ1 5 [33]
E- F GGAAGAGACAGGTACGTTAATA 20 This study
E- R AGCAGTACGCACACAATCGAA 20
E- P HEX-ACACTAGCCATCCTTACTGCGCTTCG-BHQ1 5 [32]
RP—F AGATTTGGACCTGCGAGCG 20 Universal
RP—R GATAGCAACAACTGAATAGCCAAGGT 20
RP—P ROX-TTCTGACCTGAAGGCTCTGCGCG-BHQ2 5

RNA isolation

The study is approved by the Institutional Review Board (IRB) at Imam Abdulrahman bin Faisal University (IAU) with an IRB number of IRB-2020-13-406. Also, it is approved by the institutional research committee and the de-identified samples were left over after completion of diagnostic tests; hence this study requires no consenting as per institutional ethics committee regulations.

Viral RNA was extracted from nasopharyngeal swabs in virus transport medium (VTM) which were sent to the microbiology laboratory at King Fahd Hospital of the University (KFHU), Al Khobar for SARS-CoV-2 detection. RNA extraction was performed from 280 μL of the VTM using the QIAamp Viral RNA Mini kit (Qiagen, Hilden, Germany) according to the manufacturer’s instructions.

rRT-PCR reaction

The reaction mixture (20 μL) includes the following reagents: 2 μL of 10x Buffer, 0.25 μL of dNTPs (10 mM each), 0.2 μL of uracil-DNA glycosylase (UDG) (1 U/μL), 0.4 μL of VitaTaq® HS polymerase (2 U/μL), 0.05 μL VitaScript® Enzyme mix including M-MLV (Procomcure, Austria), 0.05 μL of Triton™ X-100 (molecular biology grade, Merck), the primer and probe mixture, and RNase/DNase-free ddH2O up to 20 μL. The mixture for the primer and probe is varied according to the kit design. For the multiplex kit that simultaneously targeting the three genes, the final concentration of the primers/probes were adjusted as follow:

  1. 10 pM for RdRP-F, 13 pM for RdRP-R, and 4 pM for RdRP-P

  2. 4 pM for E-F, 4 pM for E-R, and 2 pM for E-P

  3. 10 pM for RP-F, 3.75 pM for RP-R, and 4 pM for RP-P

The mixture was dispensed in 96-well plates (MicroAmp™ Fast Optical 96-well reaction Plate 0.1 mL, Applied Biosystems) and sealed with optical film (MicroAmp™ Optical Adhesive Film, Applied Biosystems). Pseudoviral RNAs including viral RdRP and E gene and human RNaseP (RP) mRNA sequences were used as the positive template. Meanwhile, RNase/DNase-free ddH2O was added to the negative control tubes to check any contamination or primer dimer.

Quantitation experiments were performed in a real-time PCR instrument (Applied Biosystems™, 7500 Fast Real-Time PCR System). Before the operation, the instrument was calibrated by using Applied Biosystems™ 7500 Fast Real-Time PCR Systems Spectral Calibration Kit. Then, the qPCR reaction conditions were adjusted as follow: 1) Reverse transcription at 45°C for 5 min, 2) Pre-denaturation at 95°C for 30 sec, 3) 40 cycles of denaturation at 95°C for 5 sec and amplification at 60°C for 30 sec. The reporter dye channel sets as FAM for viral RdRp gene; and VIC for E gene; and ROX for human RNAseP (RP) gene. For the Applied Biosystems™ real-time PCR instrument (7500 and StepOne models), set to “passive reference” dye as “none”.

Amplification efficiency

To find out the amplification efficiency (E) of the genes, a standard curve from the dilution series of templates was prepared. Ct values versus the logarithmic amount of the template were plotted. The amplification efficiency was obtained by using the following equation:

E=100x(101/slope) (1)

Validation of the assay by Xpert Xpress SARS-CoV-2

The nasopharyngeal swabs used for the validation of the current assay were also tested for SARS-CoV-2 using the Xpert Xpress SARS-CoV-2 kit (n = 14) (Cepheid, Sunnyvale, CA, USA). About 300 μL of the VTM were transferred to the Xpert Xpress SARS-CoV-2 kit cartridge. The kit includes direct RNA extraction and rRT-PCR targeting the E and N2 gene fragments of the SARS-CoV-2. The assay was run on the GeneXpert Dx instrument (Cepheid, Sunnyvale, CA, USA).

Data analysis

The results were evaluated by determining the amplification curve of the target gene and the internal control gene. For the ABI 7500 device, the cycle threshold (Ct or Cq) line was automatically adjusted to ensure that the curves are all straight position. For this purpose, ABI 7500 software (v2.3) was used. The cycle threshold number ≤ 38 with a sigmoidal curve is accepted as `positive`.

Results

Simplex rRT-PCR standardization

Before multiplexing, all targeted gene primer/probe set, and RT-PCR reagents were tested in simplex rRT-PCR. The quantity of SARS-CoV-2 specific E and RdRP gene primer and probe set were optimized by using 105 copy/μL of synthetic viral RNA as a template. Before the reactions, the rRT-PCR instrument (Applied Biosystems™, 7500 Fast Real-Time PCR System) was calibrated to get the best fluorescent performance. The simplex reactions were repeated at least six times for two viral E and RdRP genes and one internal control (IC) gene RP. The average cycle threshold (Ct) value with standard deviations (SD) were 24.1 ± 1.05, 27.1 ± 1.5, and 31.5 ± 1.0 for RP, E, and RdRP, respectively (Fig 1B).

Duplex rRT-PCR standardization

The primer and probe sets were designed to detect one viral (E or RdRP) and one human IC gene (RP) at the same rRT-PCR reaction. Two different reaction mixtures were designed including the primer and probe sets either for RdRP with RP, or E with RP. Triplicate reactions yielded the Ct value as 23.6 ± 0.77 and 32.3 ± 1.8 for RP and RdRP genes, respectively (Fig 1C). In the second reaction mixture, the Ct value was detected as 25.7 ± 0.84 and 25.5± 0.73 for E and RP genes, respectively (Fig 1D). The Ct values of both reactions are less than 38, which is the recommended limit of CDC (Center of Disease Control and Prevention).

Multiplex (triplex) rRT-PCR standardization

Multiplex rRT-PCR protocol was set up for the amplification of three genes (E, RdRP, and RP). Three primer and probe set for each gene were combined in the same reaction tube. Sigmoidal amplification curves were obtained with an average Ct value ± SD as 28.8 ± 0.51, 27.3± 0.7, and 23.6± 1.04 for RdRP, E, and RP genes, respectively (Fig 1E). This shows that the assay is capable to detect three genes in the same reaction tube.

Limit-of-detection (LOD) and rRT-PCR efficiency

A serial dilution of synthetic RNA (105, 104, 103, 102 and 101 copies/μL) was prepared to find the limit-of-detection (LOD) for RdRP and E genes. The amplification plots, the amplification efficiencies (E), and R2 score are represented in Fig 2. The LOD of RdRP gene was ≥10 copy/μL. Nevertheless, the LOD of E gene was ≥103 copy/μL. The E value of RdRP and E genes were determined as 99.9. The R2 scores were determined as 0.977 for the RdRP and 0.995 for the E gene.

Fig 2. The limit-of-detection (LOD) of RdRP and E genes.

Fig 2

A serial dilution of synthetic RNA (105, 104, 103, 102 and 101 copies/ μL) was prepared. The amplification plots (a, c) and the amplification efficiencies (E) (b, d) were represented.

Validation of the assay using SARS-CoV-2-positive samples detected by Cepheid`s system

To validate the outcome of multiplex rRT-PCR assay (named as COV2-kit) in the clinical SARS-CoV-2-positive samples, we compared our results by using Cepheid’s GeneXpert® System. For this purpose, firstly the nasopharyngeal swabs were collected from COVID-19 infected patients between the 1st of November and the 5th of December 2020 at King Fahd Hospital of University (KFHU), Al Khobar. Then, the samples were kept in VTM medium and 300 μL of the solution was directly transferred to the Cepheid’s GeneXpert® cartridge. The samples were run by using Xpert® Xpress SARS-CoV-2 detection kit. The SARS-CoV-2 positive samples were selected for viral RNA extraction (QIAamp Viral RNA Mini Kit, Qiagen, Germany). Then, the extracted RNA was used as template to test our assay. The comparative Ct performances of each assay were shown in Fig 3. The COV2-kit detected all confirmed SARS-CoV-2 positive samples. All Ct values of the E gene are below the threshold accepted by the CDC (≤37) (Fig 3A). For the RdRP gene, only one clinical sample was out of the threshold, which is consistent with Cepheid’s assay (Fig 3B). According to the comparison of Ct values between Cepheid’s E and N2 (Fig 3C), and the RdRP and E genes of COV2-kit, there is agreement between the obtained data (Fig 3D).

Fig 3. Comparison of the cycle threshold (Ct) value of COVID-19 positive samples using Cepheid`s and the current (COV2-kit) assays.

Fig 3

(a) The data shows the Ct value of E gene, which is the common gene in both assays. (b) The comparison of Cepheid`s N2 and COV2-kit`s RdRP genes. The plots show the comparison of Ct values of the target genes for (c) Cepheid`s E and N2; and (d) COV2-kit`s RdRP and E. The red dashed line shows threshold value of 37, which is accepted as the upper limit for SARS-CoV-2 detection by CDC.

Discussion

The global increase in the COVID-19 pandemic makes the development of faster, reliable, and sensitive virus detection methods a priority [21]. Extensive studies to find an efficient, reliable, and sensitive detection method for SARS-CoV-2 are still ongoing. This study aims to develop a multiplex rRT-PCR test that targets two viral (RdRP and E) and one human (RP) gene simultaneously in a single reaction tube. The assay is named as COV2-kit. Thanks to the specific probes that were labelled with different fluorescence dyes (HEX, ROX, and FAM), the gene amplifications can be identified in the same reaction tube by using different filters of the Applied Biosystems™, 7500 Fast Real-Time PCR System. Three different approaches were performed to optimize the protocol: simplex (targets single gene), duplex (simultaneously targets two genes) and triplex (simultaneously targets three genes). For this purpose, a synthetic viral template together with mRNA of RP gene was used. The Ct values of the reactions are ranged in between 24 to 34. According to the CDC (Center of Disease Control and Prevention) and WHO recommendations, the samples with a Ct value of 37.01 or greater are considered as negative (CDC, https://www.cdc.gov/dengue/healthcare-providers/testing/molecular-tests/faq_rt-pcr.html; WHO, https://extranet.who.int/pqweb/sites/default/files/documents/201005_final_pqpr_eul_0517_204_00_sars_cov2_nucleic_acid_detection%20%281%29.pdf). In other words, the Ct value should not exceed 37 to accept the sample as positive. The Ct values in all tested approaches (simplex, duplex, or triplex) were under this threshold. The LOD (limit-of-detection) for RdRP and E genes were at least 101 and 103 copy/μL, respectively. The primers and probe for the RdRP gene were found to be more sensitive than these for the E gene. The viral load of the COVID-19 patients is the critical factor for the test efficiency [22]. It can be concluded that RdRP gene provides the maximum detection capability on the patients having low viral load (≥101 copy/μL). In addition, the patients with a viral load of ≥103 copy/μL can be detected by using the E gene as a target. Thus, the samples with a low viral load can be detected by using these two sensitive gene targeting approaches. In addition to these viral genes, the reaction includes a human gene target, RP, as an internal control (IC). In general, the IC gene is tested in a separate tube for each reaction which decreases the sample size to be tested and increase the expenses. By using the current strategy, the number of reactions per sample is reduced. For instance, the assay allows testing 91 patients in 96-well plates per run, thus provides less time and save expensive RT-PCR reagents. Fast, reliable, high-sensitivity and low-cost SARS-CoV-2 detection is achieved by designing and using effective primer / probe sets.

The COV2-kit assay was tested to verify its performance on clinically SARS-CoV-2 positive samples. In addition, the results were compared by using Cepheid’s GeneXpert® System that uses Xpert® Xpress SARS-CoV-2 detection kit (Fig 3). Rather than our strategy that targets the RdRP and E genes, the Xpert® Xpress SARS-CoV-2 detects N2 and E genes. The Ct value of the E gene was in between 18 and 41 using the Cepheid`s system. COV2-kit revealed the Ct scores of the same gene as 27–35.5. CDC recommends the upper limit of the Ct as 37. Accordingly, all tested samples were found below this limit by using the COV2-kit approach. In addition, the performance of COV2-kit`s RdRP gene versus Cepheid`s N2 gene was compared. Accordingly, the Ct value of the RdRP gene (COV2-kit) was found to be lower than N2 gene (Cepheid`s), which points out the sensitivity of COV2-kit than Cepheid`s system. Nevertheless, the Ct score of three COVID19 patients of N2 gene and one of RdRP gene were out of the upper threshold (Fig 3B). Additionally, multiplex (triplex) rRT-PCR standardization test showed that this assay is efficient to detect three genes in the same reaction tube.

Reverse transcription polymerase chain reaction (RT-PCR) is a sensitive assay for the detection of specified gene sequences encoding the proteins of the virus, such as RNA-dependent RNA polymerase (RdRP), nucleocapsid (N), envelope (E), and spike (S). Although RT-PCR tests are widely applied, and substitution tests are being developed, the current testing capability must meet new universal demands for fast, credible, and possible molecular detection [23]. In this method, we tried to bridge many challenges and weakness resulted in former assays and take benefits from their results and applications to improve novel assay. Compared to simplex or duplex analyses, the triplex analysis provides more accuracy and avoids possible negative false results that may be caused by a mutation in one of the viral genes [24]. E gene is an important gene codes for the envelope (E) protein of SARS-CoV-2. It is an integral membrane protein functional in several biological processes of the virus such as assembly, budding, envelope formation, and pathogenesis [25]. Additionally, RNA-dependent RNA polymerase (RdRP) gene encodes an enzyme responsible for viral replication. These indispensable genes are the main target for the molecular detection of SARS-CoV-2 [12, 2629]. Kakhki et al. [30] reported that ORF8 gene is a possible target for detecting COVID-19 and it is different from reported genes (RdRP, E and N genes), and ORF8 is fully specific to COVID-19 and has no cross-reactivity with other types of coronaviruses. They reported that ORF8 gene can be considered as a supplemental confirmatory test.

Our study partly agrees with a protocol of a study published by Park et al. [31], based on a strategy for styling dynamic primer kits, and nominating reaction status leads to the top utilization of these primers. They reported three-steps guidelines for designing and optimization specific primer sets: 1) the selection of primer sets for target genes (RdRP, N, E, and S) in SARS-COV-2 genome, 2) the in-silico effectiveness of primer and sequences of the amplicon, and 3) the optimization of PCR conditions. The real-time PCR-dependent protocol was developed to rather than traditional PCR-based protocols and used a multiplex PCR-based protocol permitting the simultaneous amplification of multiple gene targets such as RdRP, N, E, and S. This allows to run the reaction in a single tube. The described protocol which is like ours is wide-scale, high-integrity and accepted as gold standard. It will be applied to a larger number of viral extracted clinical RNA samples in subsequent tests to verify the specificity and sensitivity of the study. Although this strategy focused on SARS-CoV-2, but it is also applicable to other viruses.

Conclusions

This study demonstrates a multiplex rRT-PCR method for the diagnosis of SARS-CoV-2. Simultaneous targeting of two viral genes (RdRP and E) and one human gene (RP) in the same PCR reaction provides an accurate, reliable, and easy-to-use SARS-CoV-2 diagnostic test. Additionally, the assay allows working with large numbers of patients using less PCR reagent. Therefore, the cost and reliability of the assay are improved. Although the primer/probe sequences are designed to minimize the possibility of cross-reactivity with other coronavirus strains, they should be tested with emerging mutant variants. The test must be validated using different RT-PCR instruments and can be used in infection prevention and control organizations, hospitals, medical centers, and diagnostic laboratories.

Acknowledgments

We would like to thank Mr. Geoffrey James T. Moro for technical assistance in conducting RT-PCR experiments and Assist Prof Mehmet Ozdemir for the language editing and proofreading.

Data Availability

All relevant data are available within the paper.

Funding Statement

The authors extend their appreciation to the Deputyship for Research & Innovation, Ministry of Education in Saudi Arabia for funding this research work through the project number COVID19-2020-026-IRMC at lmam Abdulrahman bin Faisal University / Institute for Research and Medical Consultations (IRMC) (https://seu.edu.sa/dammam/en/news/deputy-minister-of-education-for-universities-research-and-innovation; https://www.iau.edu.sa/en). Also, this study is funded by Institute for Research and Medical Consultations (IRMC) (https://www.iau.edu.sa/en/administration/centers/institute-for-research-and-medical-consultations-irmc) under the project number of 2020-IRMC-S-3. Huseyin Tombuloglu received the awards.

References

  • 1.Liu R, Fu A, Deng Z, Li Y, Liu T. Promising methods for detection of novel coronavirus SARS‐CoV‐2. View. 2020;1(1):e4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ulloa S, Bravo C, Parra B, Ramirez E, Acevedo A, Fasce R, et al. A simple method for SARS-CoV-2 detection by rRT-PCR without the use of a commercial RNA extraction kit. Journal of virological methods. 2020;285:113960. 10.1016/j.jviromet.2020.113960 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Kudo E, Israelow B, Vogels CB, Lu P, Wyllie AL, Tokuyama M, et al. Detection of SARS-CoV-2 RNA by multiplex RT-qPCR. PLoS Biology. 2020;18(10):e3000867. 10.1371/journal.pbio.3000867 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Broughton JP, Deng X, Yu G, Fasching CL, Servellita V, Singh J, et al. CRISPR–Cas12-based detection of SARS-CoV-2. Nature biotechnology. 2020;38(7):870–4. 10.1038/s41587-020-0513-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Ali Z, Aman R, Mahas A, Rao GS, Tehseen M, Marsic T, et al. iSCAN: An RT-LAMP-coupled CRISPR-Cas12 module for rapid, sensitive detection of SARS-CoV-2. Virus research. 2020;288:198129. 10.1016/j.virusres.2020.198129 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Yoshimi K, Takeshita K, Yamayoshi S, Shibumura S, Yamauchi Y, Yamamoto M, et al. Rapid and accurate detection of novel coronavirus SARS-CoV-2 using CRISPR-Cas3. MedRxiv. 10.2139/ssrn.3640844 [DOI] [Google Scholar]
  • 7.Chen Z, Zhang Z, Zhai X, Li Y, Lin L, Zhao H, et al. Rapid and sensitive detection of anti-SARS-CoV-2 IgG, using lanthanide-doped nanoparticles-based lateral flow immunoassay. Analytical chemistry. 2020;92(10):7226–31. 10.1021/acs.analchem.0c00784 [DOI] [PubMed] [Google Scholar]
  • 8.Kasetsirikul S, Umer M, Soda N, Sreejith KR, Shiddiky MJ, Nguyen NT. Detection of the SARS-CoV-2 humanized antibody with paper-based ELISA. Analyst. 2020;145(23):7680–6. 10.1039/d0an01609h [DOI] [PubMed] [Google Scholar]
  • 9.Joung J, Ladha A, Saito M, Kim NG, Woolley AE, Segel M, et al. Detection of SARS-CoV-2 with SHERLOCK one-pot testing. New England Journal of Medicine. 2020;383(15):1492–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Chen Z, Wu Q, Chen J, Ni X, Dai J. A DNA aptamer based method for detection of SARS-CoV-2 nucleocapsid protein. Virologica Sinica. 2020;35:351–4. 10.1007/s12250-020-00236-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Jang WS, Lim DH, Yoon J, Kim A, Lim M, Nam J, et al. Development of a multiplex Loop-Mediated Isothermal Amplification (LAMP) assay for on-site diagnosis of SARS CoV-2. PloS one. 2021;16(3):e0248042. 10.1371/journal.pone.0248042 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Zhen W, Berry GJ. Development of a New Multiplex Real-Time RT-PCR Assay for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Detection. The Journal of Molecular Diagnostics. 2020;22(12):1367–72. 10.1016/j.jmoldx.2020.09.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Khan S, Tombuloglu H, Hassanein SE, Rehman S, Bozkurt A, Cevik E, et al. Coronavirus diseases 2019: current biological situation and potential therapeutic perspective. European Journal of Pharmacology. 2020;886:173447. 10.1016/j.ejphar.2020.173447 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Yan C, Cui J, Huang L, Du B, Chen L, Xue G, et al. Rapid and visual detection of 2019 novel coronavirus (SARS-CoV-2) by a reverse transcription loop-mediated isothermal amplification assay. Clinical Microbiology and Infection. 2020;26(6):773–9. 10.1016/j.cmi.2020.04.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Moran A, Beavis KG, Matushek SM, Ciaglia C, Francois N, Tesic V, et al. The detection of SARS-CoV-2 using the cepheid xpert xpress SARS-CoV-2 and Roche cobas SARS-CoV-2 assays. Journal of clinical microbiology. 2020. 10.1128/JCM.00772-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Altschul SF, Gish W, Miller W, Myers EW, Lipman DJ. Basic local alignment search tool. Journal of molecular biology. 1990;215(3):403–10. 10.1016/S0022-2836(05)80360-2 [DOI] [PubMed] [Google Scholar]
  • 17.Edgar RC. MUSCLE: a multiple sequence alignment method with reduced time and space complexity. BMC bioinformatics. 2004;5(1):1–9. 10.1186/1471-2105-5-113 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Sievers F, Wilm A, Dineen D, Gibson TJ, Karplus K, Li W, et al. Fast, scalable generation of high‐quality protein multiple sequence alignments using Clustal Omega. Molecular systems biology. 2011;7(1):539. 10.1038/msb.2011.75 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Brown SS, Chen YW, Wang M, Clipson A, Ochoa E, Du MQ. PrimerPooler: automated primer pooling to prepare library for targeted sequencing. Biology Methods and Protocols. 2017;2(1):bpx006. 10.1093/biomethods/bpx006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Untergasser A, Cutcutache I, Koressaar T, Ye J, Faircloth BC, Remm M, et al. Primer3—new capabilities and interfaces. Nucleic acids research. 2012;40(15):e115–. 10.1093/nar/gks596 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Orooji Y, Sohrabi H, Hemmat N, Oroojalian F, Baradaran B, Mokhtarzadeh A, et al. An overview on SARS-CoV-2 (COVID-19) and other human coronaviruses and their detection capability via amplification assay, chemical sensing, biosensing, immunosensing, and clinical assays. Nano-Micro Letters. 2021;13(1):1–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wacharapluesadee S, Kaewpom T, Ampoot W, Ghai S, Khamhang W, Worachotsueptrakun K, et al. Evaluating the efficiency of specimen pooling for PCR‐based detection of COVID‐19. Journal of medical virology. 2020;92(10):2193–9. 10.1002/jmv.26005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Feng W, Newbigging AM, Le C, Pang B, Peng H, Cao Y, et al. Molecular diagnosis of COVID-19: challenges and research needs. Analytical chemistry. 2020;92(15):10196–209. 10.1021/acs.analchem.0c02060 [DOI] [PubMed] [Google Scholar]
  • 24.Petrillo S, Carrà G, Bottino P, Zanotto E, De Santis MC, Margaria JP, et al. A novel multiplex qRT-PCR assay to detect SARS-CoV-2 infection: high sensitivity and increased testing capacity. Microorganisms. 2020;8(7):1064. 10.3390/microorganisms8071064 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Schoeman D, Fielding BC. Coronavirus envelope protein: current knowledge. Virology journal. 2019;16(1):1–22. 10.1186/s12985-018-1108-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Al-Saud H, Al-Romaih K, Bakheet R, Mahmoud L, Al-Harbi N, Alshareef I, et al. Automated SARS-COV-2 RNA extraction from patient nasopharyngeal samples using a modified DNA extraction kit for high throughput testing. Annals of Saudi Medicine. 2020;40(5):373–81. 10.5144/0256-4947.2020.373 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Abasiyanik MF, Flood B, Lin J, Ozcan S, Rouhani S, Pyzer A, et al. Sensitive detection and quantification of SARS-CoV-2 in saliva. medRxiv. 2020. 10.1101/2020.12.04.20241059 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Reijns MA, Thompson L, Acosta JC, Black HA, Sanchez-Luque FJ, Diamond A, et al. A sensitive and affordable multiplex RT-qPCR assay for SARS-CoV-2 detection. PLoS biology. 2020;18(12):e3001030. 10.1371/journal.pbio.3001030 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Chung HY, Jian MJ, Chang CK, Lin JC, Yeh KM, Chen CW, et al. Novel Dual Multiplex Real-time RT-PCR Assays for the Rapid Detection of SARS-CoV-2, Influenza A/B, and Respiratory Syncytial Virus using the BD Max Open System. Emerging microbes & infections. 2021;8:1–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Kakhki RK, Kakhki MK, Neshani A. COVID-19 target: A specific target for novel coronavirus detection. Gene Reports. 2020;20:100740. 10.1016/j.genrep.2020.100740 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Park M, Won J, Choi BY, Lee CJ. Optimization of primer sets and detection protocols for SARS-CoV-2 of coronavirus disease 2019 (COVID-19) using PCR and real-time PCR. Experimental & molecular medicine. 2020;52(6):963–77. 10.1038/s12276-020-0452-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Giri B, Pandey S, Shrestha R, Pokharel K, Ligler FS, Neupane BB. Review of analytical performance of COVID-19 detection methods. Analytical and bioanalytical chemistry. 2020;18:1–4. 10.1007/s00216-020-02889-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.WHO (World Health Organization). Diagnostic detection of 2019-nCoV by real-time RT-PCR. 2020. https://www.who.int/docs/default-source/coronaviruse/protocol-v2-1.pdf?sfvrsn=a9ef618c_2 Accessed 10 December 2020.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All relevant data are available within the paper.


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES