Abstract
Phase-contrast synchrotron-based X-ray imaging using an X-ray interferometer provides high sensitivity and high spatial resolution, and it has the ability to depict the fine morphological structures of biological soft tissues, including tumors. In this study, we quantitatively compared phase-contrast synchrotron-based X-ray computed tomography images and images of histopathological hematoxylin-eosin-stained sections of spontaneously occurring rat testicular tumors that contained different types of cells. The absolute densities measured on the phase-contrast synchrotron-based X-ray computed tomography images correlated well with the densities of the nuclear chromatin in the histological images, thereby demonstrating the ability of phase-contrast synchrotron-based X-ray imaging using an X-ray interferometer to reliably identify the characteristics of cancer cells within solid soft tissue tumors. In addition, 3-dimensional synchrotron-based phase-contrast X-ray computed tomography enables screening for different structures within tumors, such as solid, cystic, and fibrous tissues, and blood clots, from any direction and with a spatial resolution down to 26 μm. Thus, phase-contrast synchrotron-based X-ray imaging using an X-ray interferometer shows potential for being useful in preclinical cancer research by providing the ability to depict the characteristics of tumor cells and by offering 3-dimensional information capabilities.
Keywords: phase-contrast synchrotron-based X-ray imaging, X-ray interferometer, histopathological images, rats, testicular tumors
Introduction
Ex-vivo micro-magnetic resonance imaging has been used to generate 2- and 3-dimensional high-resolution images of tumors, but even with 9.4-T magnetic resonance scanners spatial resolution is limited to around 60 μm.1 Micro X-ray computed tomography (CT) provides excellent anatomic imaging of bones and teeth with a spatial resolution on the sub-micrometer order,2 but since biological soft tissues, including tumors, are composed of light elements, they cannot be imaged well because of their significantly weaker X-ray absorption. However, the phase-contrast X-ray CT imaging technique, which is based on phase shifts of X-rays, has the potential to overcome this problem. When an X-ray passes through an object, its amplitude decreases and its phase shifts. In the hard X-ray region, the cross section of the phase shift is around 1000 times greater than that of absorption for light elements, such as biological soft tissues,3 and as a result phase-contrast X-ray CT imaging is capable of providing better contrast for biological soft tissue than conventional absorption based X-ray imaging does.4-8
Several imaging methods, including interferometry performed with an X-ray interferometer, diffraction-enhanced imaging (Analyzer-based imaging), propagation-based method with a Fresnel pattern and Talbot interferometry with a Talbot grating interferometer, are capable of detecting object-induced phase distributions.9-13
A high coherence synchrotron source is an ideal means of improving the high image resolution and sensitivity of phase-contrast X-ray CT imaging for biological soft tissues, and because of its simple system set-up, the Talbot interferometry method7,14 has begun to be used with conventional polychromatic X-ray sources. However, the synchrotron-based crystal interferometry imaging method directly detects the phase shift by the superposition of the waves, while other methods detect the first or second spatial derivation of the phase shift. Thus, crystal interferometry has greater sensitivity for observation of biological soft tissue than other methods, because spatial distribution of phase shift (phase maps) can be obtained without integral calculation, which may amplify noise.
In previous comparative studies among absorption, diffraction-enhanced method, Talbot interferometry and crystal X-ray interferometry, the sensitivity was higher, approximately 230 times in crystal X-ray interferometer, approximately 40 times in diffraction-enhanced method, and approximately 13 times in Talbot interferometry compared to conventional absorption imaging method with monochromatic X-ray at 17.8 keV.15,16 Therefore, crystal interferometry is the most suitable method for making observations that require high-density resolution, such as observations of tumor microstructure and amyloid plaques. Diffraction-enhanced methods and Talbot interferometry have a wide dynamic range of density and are suitable for observation of samples containing large differences in density, such as bone and soft tissue.15,16
Actually, phase-contrast synchrotron-based X-ray CT imaging using an X-ray interferometer makes it possible to visualize the fine morphological structures of biological soft tissues without using contrast agents.4–5,17 Histopathological imaging is a well-established approach to visualizing tissue architecture and properties down to the sub-micrometer resolution level and is widely used for tumor diagnosis, and phase-contrast synchrotron-based X-ray CT imaging has been found to provide images that close resemble conventional histopathological images.5,18,19
In this article, based on a study of spontaneously occurring rat testicular tumors, we report finding that phase-contrast synchrotron-based X-ray CT imaging yielded images of soft tissue tumors whose information content was quantitatively comparable to that of histopathological images. We also demonstrated easy screening of an entire tumor from any direction by a 3-dimensional (3D) phase-contrast synchrotron-based X-ray CT method without the need for a contrast agent.
Materials and Methods
Animal Preparation
Two spontaneously occurring testicular tumors obtained from aging rats (2-year-old male Wistar rats) were used as the specimens in this study. Under anesthesia, the apex of the left ventricle of the heart of each rat was surgically cannulated, and the rat’s blood was replaced with physiological saline solution containing heparin in order to eliminate blood coagulation artifacts within the blood vessels. The tumor-bearing testes were quickly excised and fixed in 10% formalin for imaging.
One testis measured 14 mm in width and 20 mm in length, and the other testis measured 10 mm in width and 15 mm in length. The testes were irregular in shape and had a tense capsule, and few contusions were seen on their surface.
Phase-Contrast Synchrotron-Based X-ray CT Imaging
The phase-contrast synchrotron-based X-ray CT imaging system (Figure 1)20,21 consisted of an asymmetric Si (220) crystal that formed a 2-dimensional beam, a 2-crystal X-ray interferometer, a phase shifter, a target sample cell, and a lens-coupling X-ray charged-coupled device (CCD) camera with a pixel size of 18 × 18 μm2. The field of view was set at 16 × 13 mm2 to detect interference patterns.
Figure 1.

Phase-contrast synchrotron-based X-ray CT system with two-crystal X-ray interferometer.
The interferometer consisted of 2 crystal blocks containing 2 crystal wafers each. The incident X-ray beam was divided into 2 beams by the first wafer (Splitter) of the first crystal block by Laue-case X-ray diffraction, and the 2 beams were similarly reflected by the second wafer (Mirror) of the first crystal block and the first wafer (Mirror) of the second crystal block, respectively. Reflected beams were then recombined at the second wafer (Analyzer) of the second crystal block to create 2 interference beams. Placing the sample on one of the interference beam (object beam) paths produced interference fringes corresponding to the phase shift caused by the sample.
During the imaging, the specimen was placed in a sample cell filled with 10% formalin to prevent the sample from drying shrinkage. The total image acquisition time for each sample was 2.5 hr, which included the exposure time, sample rotation time, and data transmission time. The X-ray exposure time was 9 sec per projection, and since the total number of projections was 250 over 180°, the actual X-ray exposure time was less than 40 min.
A phase map for each projection was obtained by the fringe scanning method,22 with subtraction of the background phase. The number of fringe scans was set to 3, with a 3-sec exposure time to obtain each interference pattern. The total exposure time for a phase map was 9 sec. Phase-contrast X-ray CT images were reconstructed by using a filtered back-projection algorithm with Shepp-Logan filter.20,21
Experiments were carried out at the vertical wiggler beamline BL14C of the Photon Factory, High Energy Accelerator Research Organization, Tsukuba, Japan. The X-ray energy was set at 35 keV by the monochromator, and the X-ray flux in front of the sample cell was approximately 4.0 × 107 counts/sec/mm2 at a ring current of 450 mA with 2.5 GeV storage energy. The density and spatial resolution of the phase-contrast synchrotron-based X-ray CT system were approximately 0.57 mg/cm3 and 26 μm, respectively.15
Histopathological Staining
After performing the phase-contrast synchrotron-based X-ray CT imaging, the specimens were sliced into 3-μm-thick sections, and the sections were stained with hematoxylin-eosin (HE) to examine their histological structure. The sections were examined with an optical microscope (Olympus FSX100; Olympus, Tokyo, Japan), and the optical microscopic images (×20) were used for comparison with the phase-contrast synchrotron-based X-ray CT image. The original color images were converted into 8-bit gray-scale images for comparison with the phase-contrast synchrotron-based X-ray CT images (PSCIs).23
Image Analysis
The PSCIs were analyzed with real-time 3D volume-rendering software (Real INTAGE; KGT Inc., Tokyo, Japan), which contains several integrated tools for 3D image data analysis, including density measurement, data processing, and 3D display. Interactive selection of processing parameters made it easy to depict the microstructure of the tumors.
Image-manipulating software (NIH Image version 1.41, https://imagej.nih.gov/nih-image) was used to calculate mass density in regions of interest (ROI) placed on different areas of the tumors: solid areas, cystic areas, and fibrous areas.
The X-ray complex refractive index, n, is described as n = 1-δ-iβ, where the real refractive index δ and imaginary refractive index β are related to the X-ray phase shift and X-ray absorption. δ is given by24:
where λ, re, Ni, Zi, and f’i are X-ray wavelength, classical electron radius, atomic density, atomic number, and the real part of the anomalous atomic scattering factor of element i, respectively.
By using these constituents of the refractive index, the phase shift, Δp, is given by Δp = 2πδt/λ, where t is the thickness of the sample. The real part of the refractive index δ was then calculated directly from the phase-shift data by using the following formula24,25:
Since the electron density of the sample (ρ) is almost proportional to δ,24 approximate mass density (ρ) was calculated by using the formula24,25:
where δ is the real part of the refractive index, λ is X-ray wavelength, re is the classical electron radius, and NA is Avogadro’s number.
The area of each testicular tumor that yielded the best morphological match between the PSCIs and HE images was selected for the analysis and divided into 3 regions according to the gray-scale gradation. The mean relative gray value of each region was calculated to compare the absolute densities of the corresponding regions on the PSCIs. The gray level distribution was also measured in detail in the 3 different magnified regions.
Results
The PSCIs of the tumors showed heterogeneous dense lobulations and multiple low-density cysts in each tumor, and a few compressed seminiferous tubules were seen at the periphery of the testis as a result of almost total replacement of the testis by the tumor (Figure 2A). The lobulations appeared to be separated by high-density capsule-like structures, and they contained slightly low-density heterogeneous structures. The HE images revealed that most of the testis was occupied by the tumor, and a few compressed atrophic tubules were seen at the periphery of the testis. The tumor was composed of multiple solid nodules and cysts with dense fibrotic septa (Figure 2B and C). Thus, the HE images yielded findings very similar to the findings revealed by the PSCIs. In addition, blood clots in the vessels were visualized as multiple white spots on the PSCIs, and the HE images showed artifacts due to the cutting procedure, but the PSCIs did not.
Figure 2.
Tumor-bearing rat testis: (A) Phase-contrast synchrotron-based X-ray CT image; (B) corresponding histopathological (HE-stained) gray-scale image; (C) color image; periphery of the testis (brown arrow); cystic areas (red arrow); solid areas (green arrow); fibrous septa (blue arrow); artificially torn region in the HE image (black arrow).
Good linear correlations were observed between the absolute densities and the relative gray values in different areas of the tumor: solid areas, cystic areas, and fibrous areas (r = 0.87) in Figure 3.
Figure 3.

Correlation between the absolute densities and relative gray values of different areas of the tumor, i.e. solid portions, cystic areas, and fibrous septa, shown in Figure 2. ①: solid tumor area 1, ②: solid tumor area 2, ③: solid tumor area 3, ④: fibrous septa, ⑤: cyst.
We also measured absolute densities in the PSCIs (Figure 4A) and mean relative gray values in the HE images (Figure 4B) in 3 different regions of the solid soft tissue tumor (Table 1). The absolute density values for region 1, region 2, and region 3 were 1.040 ± 0.001 g/cm3, 1.078 ± 0.003 g/cm3, and 1.065 ± 0.002 g/cm3, respectively, and the corresponding mean relative gray values were 74.9 ± 1 5.4, 87.6 ± 18.4, and 76.9 ± 11.7, respectively. The pixel numbers corresponding to the gray value distribution in different regions on the HE images are shown in Figure 4C.
Figure 4.
Magnified image of the area of the testicular tumor that yielded the closest morphological match between the phase-contrast synchrotron-based X-ray CT images and HE-stained images. (A) Phase-contrast synchrotron-based X-ray CT image, (B) gray-scale HE image, (C) gray value distribution curve of the boxed region in (B), and (D) magnified image of the boxed region in (B).
Table 1.
Absolute Densities Obtained From the Phase-Contrast Synchrotron-Based X-Ray CT Image Shown in Figure 4(A) and the Relative Gray Values Obtained From the Gray-Scale HE-Stained Histopathological Images Shown in Figure 4(B).
| Region 1 | Region 2 | Region 3 | |
|---|---|---|---|
| Density (g/cm3) | 1.040 ± 0.001 | 1.078 ± 0.003 | 1.065 ± 0.002 |
| Relative gray value | 74.9 ± 15.4 | 87.6 ± 18.4 | 76.9 ± 11.7 |
The magnified HE images showed different types of cells within the solid soft tissue tumors, including cells containing large cytoplasmic vacuoles and small hyperchromatic nuclei (region 1), dense lymphocytes containing hyperchromatic nuclei and scant cytoplasm (region 2), and cells containing moderately enhancing abundant cytoplasm and relatively large hypochromatic nuclei (region 3) (Figure 4D). The density of the nuclear chromatin in the HE images appeared to be closely correlated with the absolute density in the PSCIs.
The excellent contrast and high resolution of the PSCIs enabled clear separation of the tumor and surrounding seminiferous tubules from any direction (Figure 5A and B). The morphology of the seminiferous tubules, blood vessels, and tunica albuginea could also be identified. The enlarged images showed the significant tumor heterogeneity information caused by solid, cystic, blood clots and fibrous septa (Figure 5C).
Figure 5.
Three-dimensional phase-contrast synchrotron-based X-ray CT image of a tumor-bearing rat’s testis. Serial cut sections of the tumor; (A) surface area, (B) internal area, (C) enlargement of the boxed region in (B); a: blood vessel, b: seminiferous tubules, c: tunica albuginea, d: tumor, e: different densities apparent of the solid area of the tumor, f: cystic area of the tumor, g: blood clot, h: fibrous septa.
Discussion
Without requiring the use of any contrast agents, phase-contrast synchrotron-based X-ray imaging provides morphological information down to a spatial resolution of approximately 26 μm, which is comparable to the spatial resolution obtained by low power microscopic image. The PSCIs proved capable of depicting the various morphological structures in the soft tissue tumors, including their solid, cystic, and fibrous areas, and their hemorrhagic areas as well. In particular, minutely differing densities within the solid tumors on the PSCIs corresponded well to the different types of cells in the soft tissue tumors observed on the HE images. Quantitative comparisons between the absolute densities in the PSCIs and the relative contrast values in the HE images yielded excellent correlations, and the high-density sites on the PSCIs corresponded to the high nuclear chromatin locations on the HE images. A detailed quantitative analysis focused on the solid tumor components revealed that the density differences in the PSCIs were mainly generated by differences in the density of the nuclear chromatin on the HE images. The nuclear-cytoplasmic ratio appears to be more important to making a diagnosis of cancer than the density of the cytoplasm in tumor cells. Because cancer cells are characterized by large nuclei and a high nuclear-cytoplasmic ratio, and the latter is well-known to be a strong predictor of malignancy.26 The spatial resolution provided by the high power microscopic images of 3-μm-thick HE-stained tissue sections can yield detailed cellular morphology information at less than a 1-µm resolution, which enables observation of tumor cell nuclei measuring 2- to 3-µm in diameter. Although 26-µm resolution provided by the PSCIs was considerably lower than the resolution of the histological images, the PSCIs clearly depicted the morphological structures of testis (10-14 mm in width, 15-20 mm in length) with tumor and density information that roughly corresponded to the density of the nuclear chromatin of the tumor cells.
Thus, phase-contrast synchrotron-based X-ray imaging appears to depict the characteristics of different type of cells within a solid soft tissue tumor.
Moreover, the 3D information capabilities of phase-contrast synchrotron-based X-ray imaging enable imaging of entire organs, and, in addition to revealing tumor morphology, they make it possible to detect precise density differences within tumors themselves in the form of heterogeneity that is probably attributable to differences in cell types. While histopathological staining methods are generally used in biomedical research and routine clinical practice, histological analysis of specific areas requires serial sectioning and staining, which can be labor-intensive and time-consuming and sometimes lead to inaccurate results. Phase-contrast synchrotron-based X-ray imaging can be performed repeatedly and can be used to access multiple tumor areas. Since phase-contrast synchrotron-based X-ray imaging does not require the use of contrast agents or staining, tissue morphology is well preserved when the tissue is subsequently processed histopathologically. Thus, the 3D visualization can provide advance information that is useful in deciding on histopathological staining methods to precisely identify places to section in order to view specific areas of interest.
There are some limitations in our study. First, our dedicated imaging system can only be used with highly brilliant X-ray sources like synchrotrons and it cannot yet be used with the clinically more readily available X-ray sources such as conventional X-ray tubes. However, synchrotron-based imaging can be used as a reference standard for pathologic tissue, especially for spatial resolution on a micrometer order.27 Second, the absorption by the wafers of the interferometer is not negligible when the monochromatic synchrotron 35 keV X-ray source is used. Since each wafer is 1 mm thick and has a transmission ratio of about 85%, the intensity of the outgoing interfering X-ray is reduced to about 60% by passing through 3 wafers. Thus, even with synchrotron radiation, an exposure time of several seconds is required. Moreover, current X-ray CCD cameras (lens-coupling X-ray charged couple device cameras) have limited sensitivity and low transfer speed. In order to obtain images within a shorter time, as our next step we are planning to use the interferometer containing crystal wafers as thin as 0.2 mm to reduce the absorption by the wafer, and to use a high-speed/high-sensitivity X-ray CCD camera.
Conclusions
Phase-contrast synchrotron-based X-ray imaging using an X-ray interferometer makes it possible to depict the characteristics of different type of cells in solid tissue tumor tissues, and it offers 3D information capabilities and can be used as an ancillary approach to histopathological analyses in preclinical cancer research.
Abbreviations
- CT
computed tomography
- 3D
three-dimensional
- HE
hematoxylin-eosin
- CCD
charged-coupled device
- PSCIs
phase-contrast synchrotron-based X-ray CT images
Footnotes
Authors’ Note: All animal experiments were performed in accordance with the experimental protocol approved by the President of Kitasato University through the judgment of the Animal Care and Use Committee of Kitasato University (reference number:14-02). This research was carried out with the approval (proposal nos. 2013G584, 2016G034) of the photon factory program advisory committee of the High Energy Accelerator Research Organization.
Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.
ORCID iD: Thet-Thet -Lwin, MD, PhD
https://orcid.org/0000-0002-7225-9228
References
- 1. Kim E, Zhang J, Hong K, Benoit NE, Pathak AP. Vascular phenotyping of brain tumors using magnetic resonance microscopy (μMRI). J Cereb Blood Flow Metab. 2011;31(7):1623–1636. doi:10.1038/jcbfm.2011.17 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Jud C, Schaff F, Zanette I, Wolf J, Fehringer A, Pfeiffer F. Dentinal tubules revealed with x-ray tensor tomography. Dent Mater. 2016;32(9):1189–1195.doi:10.1016/j.dental.2016.06.021 [DOI] [PubMed] [Google Scholar]
- 3. Momose A, Fukuda J. Phase-contrast radiographs of nonstained rat cerebellar specimen. Med. Phys. 1995;22(4):375–379. doi:10.1118/1.597472 [DOI] [PubMed] [Google Scholar]
- 4. Momose A, Takeda T, Itai Y, Hirano K. Phase–contrast X–ray computed tomography for observing biological soft tissues. Nat Med. 1996;2(4):473–475. doi:10.1038/nm0496-473 [DOI] [PubMed] [Google Scholar]
- 5. Takeda T, Momose A, Hirano K, et al. Hunan carcinoma: early experiment X-ray CT with synchrotron radiation comparative specimen study with optical microscopy. Radiology. 2000;214(1):298–301. doi:10.1148/radiology.214.1.r00ja08298 [DOI] [PubMed] [Google Scholar]
- 6. Stampanoni M, Groso A, Isenegger A, et al. Trends in synchrotron-based tomographic imaging: the SLS experience. Developments in X-Ray Tomography V, Proceedings of the Society of Photo-Optical Instrumentation Engineers (Spie). 2006;6318:U199–U212. [Google Scholar]
- 7. Pfeiffer F, Weitkamp T, Bunk O, David C. Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources. Nat Phys. 2006;2(4):2258–2261. [Google Scholar]
- 8. Davis TJ, Gao D, Gureyev TE, Stevenson AW, Wilkins SW. Phase-contrast imaging of weakly absorbing materials using hard X-rays. Nature. 1995;373(6515):595–598. [Google Scholar]
- 9. Fitzgerald R. Phase-sensitive X-ray imaging; new approaches that can detect X-ray phase shift within soft tissues show promise for clinical and biological applications. Phys today. 2000;53(7):23–28. [Google Scholar]
- 10. Momose A. Recent advanced in X-ray phase imaging. Jpn J Appl Phys. 2005;44(9R):6355–6367. [Google Scholar]
- 11. Endrizzi M. X-ray phase-contrast imaging. Nucl Inst Meth Phys Res A. 2018;878:88–98. [Google Scholar]
- 12. Wilkins SW, Gureyev TE, Gao D, Pogany A, Stevenson AW. Phase-contrast imaging using polychromatic hard X-rays. Nature. 1996:384(6607):335–338. [Google Scholar]
- 13. Nugent KA, Gureyev TE, Cookson DF, Paganin D, Barnea Z. Quantitative phase imaging using hard X-rays. Phys Rev Lett. 1996;77(14):2961–2964. [DOI] [PubMed] [Google Scholar]
- 14. Momose A. X-ray phase imaging reaching clinical uses. Phys Med. 2020;79:93–102. doi:10.1016/j.ejmp.2020.11.003 [DOI] [PubMed] [Google Scholar]
- 15. Yoneyama A, Wu J, Hyodo K, Takeda T. Quantitative comparison of imaging performance of x-ray interferometric imaging and diffraction enhanced imaging. Med Phys. 2008;35(10):4724–4734. doi:10.1118/1.2977543 [DOI] [PubMed] [Google Scholar]
- 16. Yoneyama A, Baba R, Hyodo K, Takeda T. Quantitative comparison of performance of absorption, Talbot interferometric, and crystal x-ray interferometric imaging. Electronic Presentation Online System. In: Proceedings of the ECR 2015 Conference of the European Society of Radiology; 2015:4–8. doi:10.1594/ecr2015/C–0531 [Google Scholar]
- 17. Noda-Saita K, Yoneyama A, Shitaka Y, et al. Quantitative analysis of amyloid plaques in a mouse model of Alzheimer’s disease by phase-contrast X-ray computed tomography. Neuroscience. 2006;138(4):1205–1213. doi:10.1016/j.neuroscience.2005.12.036 [DOI] [PubMed] [Google Scholar]
- 18. Huang S, Kou B, Chi Y, et al. In-line phase-contrast and grating-based phase-contrast synchrotron imaging study of brain micrometastasis of breast cancer. Sci Rep. 2015;5:9418.doi:10.1038/srep09418 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Baran P, Mayo S, McCormack M, et al. High-resolution X-ray phase-contrast 3-D imaging of breast tissue specimens as a possible adjunct to histopathology. IEEE Trans Med Imag. 2018;37(12):2642–2650. [DOI] [PubMed] [Google Scholar]
- 20. Yoneyama A, Takeda T, Tsuchiya Y, et al. A phase-contrast x-ray imaging system-with a 60 x 30 mm field of view based on a skew-symmetric two-crystal x-ray interferometer. Nucl Instrum Methods Phys Res. 2004;A523(1-2):217–222. doi:10.1016/j.nima.2003.12.008 [Google Scholar]
- 21. Yoneyama A, Takeda T, Tsuchiya Y, et al. High-energy phase-contrast X-ray imaging using a two-crystal X-ray interferometer. J Synch Rad. 2005;12(Pt 4):534–536. doi:10.1107/S0909049505008356 [DOI] [PubMed] [Google Scholar]
- 22. Momose A. Demonstration of phase-contrast X-ray computed tomography using an X-ray interferometer. Nucl Instrum Methods Phys Res. 1995;A352:622–628. doi:10.1016/0168-9002(95)90017-9 [Google Scholar]
- 23. Lwin T-T, Yoneyama A, Imai M, Maruyama H, Hyodo K, Takeda T. Testicular seminoma in the aged rat visualized by phase-contrast X-ray computed tomography. Acta Radiologica Open. 2018;7(10):2058460118806657. doi:10.1177/2058460118806657 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Yoneyama A, Yamada S, Takeda T. Fine biomedical imaging using X-ray phase-sensitive technique. In: Gargiulo DG, Mcewan A, eds. Advanced Biomedical Engineering. InTech; 2011;2:107–128.doi:10.5772/20456 [Google Scholar]
- 25. Shirai R, Kunii T, Yoneyama A, et al. Enhanced renal image contrast by ethanol fixation in phase-contrast X-ray computed tomography. J Synch Rad. 2014;21(4):795–800. doi:10.1107/S1600577514010558 [DOI] [PubMed] [Google Scholar]
- 26. Baba AI, Câto C. Tumor cell morphology. In Comparative Oncology. The Publishing House of the Romanian Academy [Internet]; 2007. ISBN-10:973-27-1457-3ISBN-13:978-073-27- 1457-7. [PubMed] [Google Scholar]
- 27. Wu J, Takeda T, Lwin TT, et al. Imaging renal structures by X-ray phase-contrast microtomography. Kid Int. 2009;75(9):945–948. doi:10.1038/ki.2009.42 [DOI] [PubMed] [Google Scholar]