Abstract
Continuous glucose monitors (CGMs) have suddenly become part of routine care in many hospitals. The coronavirus disease 2019 (COVID-19) pandemic has necessitated the use of new technologies and new processes to care for hospitalized patients, including diabetes patients. The use of CGMs to automatically and remotely supplement or replace assisted monitoring of blood glucose by bedside nurses can decrease: the amount of necessary nursing exposure to COVID-19 patients with diabetes; the amount of time required for obtaining blood glucose measurements, and the amount of personal protective equipment necessary for interacting with patients during the blood glucose testing. The United States Food and Drug Administration (FDA) is now exercising enforcement discretion and not objecting to certain factory-calibrated CGMs being used in a hospital setting, both to facilitate patient care and to obtain performance data that can be used for future regulatory submissions. CGMs can be used in the hospital to decrease the frequency of fingerstick point of care capillary blood glucose testing, decrease hyperglycemic episodes, and decrease hypoglycemic episodes. Most of the research on CGMs in the hospital has focused on their accuracy and only recently outcomes data has been reported. A hospital CGM program requires cooperation of physicians, bedside nurses, diabetes educators, and hospital administrators to appropriately select and manage patients. Processes for collecting, reviewing, storing, and responding to CGM data must be established for such a program to be successful. CGM technology is advancing and we expect that CGMs will be increasingly used in the hospital for patients with diabetes.
Keywords: Blood glucose, COVID-19, Diabetes mellitus, Glucose, Hospitals, Intensive care units, Technology
INTRODUCTION
Recent escalating interest in the use of continuous glucose monitor (CGM) technology for hospitalized patients has been fueled by (1) improvements in the sensing and data management technology; (2) increasing popularity of these devices among outpatients with diabetes as well as others from the athlete community and the quantified self movement; (3) a recent surge of articles in leading diabetes journals describing both good accuracy and improved clinical outcomes in hospitalized patients using these devices; (4) a recent decision by the United States Food and Drug Administration (FDA) related to the coronavirus disease 2019 (COVID-19) pandemic to exercise enforcement discretion and not block hospitals from providing devices and technical support to hospitals wishing to use these systems; and (5) dissemination of guidelines and review articles describing policies for using CGMs in a hospital setting.
DEFINITION OF A CONTINUOUS GLUCOSE MONITOR
A CGM is a device that measures the glucose concentration automatically around the clock [1]. The device must be attached to the body in some way and it can be either a wearable device or an implanted device. A CGM sensor can be situated in the subcutaneous space to measure the glucose concentration in interstitial fluid (ISF) or in a blood vessel and measure the glucose concentration in blood. Intravascular sensors are rarely used and confer a risk of bleeding and thrombosis and will not be covered in this article. Subcutaneous glucose sensors can require calibration anywhere from not at all up to four times daily [2]. One of the advantages of automatic glucose measurements is the time saved for the nurses who do not have to check blood glucoses as often, but a disadvantage is that compared to blood glucose monitors, these sensors track ISF glucose rather than reference blood glucose concentrations (which are used for registration studies) less closely. Therefore, if a particular CGM requires regular calibration with blood glucose testing and provides less accurate reading as well, then there will be little interest in using such a product for hospitalized patients.
A CGM sensor with two attractive features: (1) factory calibration and (2) accuracy close to that of most blood glucose monitors (which many clinicians and mathematicians consider to be a mean absolute relative difference (MARD) from reference of below 10%) [3] would be well received in a hospital setting, providing there are no special accuracy concerns about physiological states altering ISF composition in critically ill patients. Two currently available types of sensors, the FreeStyle Libre series (FreeStyle Libre 14 day and FreeStyle Libre 2) from Abbott Diabetes Care (Alameda, CA, USA) [4,5], and the Dexcom G6 from Dexcom (San Diego, CA, USA) [6] provide these two features for most patients. The FreeStyle Libre 3 has a Conformité Européenne (CE) mark, but is not cleared by the FDA. However, none of these products are cleared by the FDA for use by critically ill hospitalized patients. Table 1 presents a list of currently available (in the United States) factory-calibrated subcutaneous CGMs that have product labels available and their known interferences from chemical substances [7–14].
Table 1.
CGM system | Warm up time, hr | No. of wear, day | Technical features [7] | Can the device be connected to an AID system? | Associated mobile app | Known interferences from chemical substances |
---|---|---|---|---|---|---|
Abbott Diabetes Care FreeStyle Libre 14 day system [8] | 1 | 14 | Range 40–500 mg/dL; no predictive alerts; glucose measured every minute and real-time data can be viewed by scanning sensor which holds 8 hours of data | No | Freestyle LibreLink, LibreLinkUp [9] | Ascorbic acid Salicylic acid |
Abbott Diabetes Care FreeStyle Libre 2 [10,11] | 1 | 14 | Range 40–400 mg/dL; optional real-time alarms for hypoglycemia, hyperglycemia, and signal loss; no predictive alerts, since the sensor monitors glucose every minute; real-time data can be viewed by scanning sensor which holds 8 hours of data | No | FreeStyle Libre 2 app (currently under FDA review), LibreLinkUp [9] | Ascorbic acid |
Dexcom G6 [12,13] | 2 | 10 | Range 40–400 mg/dL; sensor monitors glucose every 5 minutes; urgent low alarm (55 mg/dL) and optional hypoglycemia predictive alert, hypoglycemia and hyperglycemia threshold alerts, and rate of change alerts | Yes | Dexcom G6 app, Follow app [14] | Hydroxyurea |
CGM, continuous glucose monitor; AID, automated insulin delivery; FDA, U.S. Food and Drug Administration.
For a given glucose concentration, CGMs from different manufacturers might report different percentages of time spent below range and above range, and might generate different individual times spent below or above range that could lead to different therapy recommendations [15]. Some hospitalized patients are volume depleted or they can experience peripheral vasoconstriction from pressors, and it has not been clearly established whether CGMs that are currently intended for outpatient use can deliver adequately accurate results for these people [16]. More data will be needed about the analytical accuracy of CGMs in hospitalized patients before these devices will receive regulatory clearance for use in this setting.
With the recent COVID-19 pandemic straining hospital resources in 2020, the FDA notified Abbott Diabetes Care and Dexcom that they would exercise enforcement discretion and not object if these companies provided devices and technical support to hospitals who used CGMs for off-label use [17–19]. This plan has allowed many hospitals to assign nurses to use these devices for remote monitoring of glucose concentrations in COVID-19 patients with diabetes without needing to spend time performing fingerstick testing, having as much contact with contagious patients, and using up as much (in some cases limited) personal protective equipment (PPE) during fingerstick testing. Furthermore these programs have benefitted both these two manufacturers and the FDA by facilitating collection of real world data on large numbers of patients, whereas in clinical trials of these products in a hospital setting testing protocols would have been expensive and time consuming.
ESTABLISHING A CGM PROGRAM
The implementation of a CGM program in the hospital demands the interaction of multiple stakeholders including hospital leadership, physicians (many times including endocrinologists), nursing leadership, information technologists, quality implementation officers, laboratories representatives, pharmacists, and risk management officers. Ideally, information technology services can facilitate the documentation of CGM in the electronic health records (EHRs) to facilitate the comparison of point of care (POC) tests with CGM values and confirm periodically that CGM readings used for patient care are within an acceptable range [20]. Steps for a successful implementation in the hospital were recently described by Galindo et al. [21].
Current platforms for CGM implementation with factory-calibrated devices (that do not need a confirmatory test for decisions in the outpatient setting) include apps for the patient, apps for remote followers of glucose values, and platforms for population-based management. The Dexcom G6 device from Dexcom can transmit glucose data via Bluetooth to a receiver or smartphone (Android or iOS) within 20 feet. If using a compatible smartphone, then information can then be shared via the “Follow” app to up to 10 selected people, such as clinicians at a nursing station (who can use a dedicated tablet), the primary care team, bedside nurses, pharmacists, and/or endocrinologists that follow remotely. The data can also be transferred to online platforms, such as Dexcom CLARITY, that generates CGM-based glucose reports and summarizes data on a daily basis or over any specified period of time [14]. With the FreeStyle Libre 14 day and FreeStyle Libre 2 systems, a similar approach can allow the use of this technology in the hospital. With the FreeStyle Libre 14 day system the information can be transferred to a receiver or a smartphone via the FreeStyle Librelink app, which then sends data to FreeStyle LibreView. Data on FreeStyle LibreView can be viewed by the patient and clinicians. Data can be viewed by patients via the LibreLink app, then to followers via the LibreLinkUp app, and then to Libreview.com to monitor multiple patients [9,22]. It should be noted that the both of the FreeStyle Libre systems measure glucose every minute; real-time glucose data including the trend and retrospective information is available by scanning the sensor, which holds 8 hours of data. The FreeStyle Libre 2 system has real-time optional alarms, which alert the user to high and low glucose levels without scanning.
Even though CGMs often appear to be reliable in the hospital, we believe regular routine POC blood glucose testing is still needed plus additional confirmatory POC tests are needed when: (1) glucose values are <85 mg/dL or >300 mg/dL; (2) hypoglycemic symptoms occur; (3) glucose readings and/or glucose trend arrows are not present on the monitor; (4) a blood drop symbol appears on the monitor; (5) hemodynamic instability occurs; and (6) a patient is in the immediate postoperative period.
WHO IS A CANDIDATE FOR USING A CGM IN THE INTENSIVE CARE UNIT
Reports are emerging on the use of wearable CGMs in the intensive care unit (ICU) setting during the COVID-19 pandemic. CGMs may be practical in the ICU for patients that require continuous intravenous insulin infusion, where hourly POC blood glucose testing is not practical. CGMs can be attractive alternatives to hourly POC capillary blood glucose testing because of the exposure risk of healthcare workers performing assisted monitoring of blood glucose [23], the amount of time needed to don and doff PPE in order to perform a POC capillary blood glucose test, and the depletion of scarce PPE used up just to perform a single POC capillary blood glucose test. Also, at some hospitals there can be a shortage of POC blood glucose monitors. Careful monitoring of POC blood glucose along with CGM readings are required to identify potential mechanical interference [24], such as pressure induced sensor attenuation [25]. Table 2 presents a review of the literature of clinical trials of CGM use in ICU settings for adult patients [26–68]. Table 3 presents a review of the literature of clinical trials of CGM use in ICU settings for pediatric patients [68–74]. Every table in this article presenting studies of CGM trials in hospital settings refers to protocols where these devices are intended to determine insulin doses administered manually rather than by way of automated delivery. In these intended settings, patients are already being closely monitored and there is no convincing data currently that CGM technology will be useful in this setting. Furthermore, patients in an intensive care setting frequently require pressors, which can cause peripheral vasoconstriction. It is possible that in such patients the circulation to the skin where CGMs are placed might be decreased and the CGM readings might or might not be accurate. Recently, Perez-Guzman et al. [26] performed a prospective study with the Dexcom G6 in the ICU, for 15 patients with stress hyperglycemia treated with continuous insulin infusion and vasopressors. The reported accuracy was a MARD of 12.9%. Agarwal et al. [27] conducted a retrospective analysis of 11 diabetes patients who were using Dexcom G6 CGMs. Their series included patients who had anasarca and/or were receiving renal replacement therapy, vasopressors, mechanical ventilation support, or high-dose glucocorticoids. The MARD of these CGMs was 12.58% [27]. In a different cohort of 30 critically ill patients, Chow et al. [28] found a decrease in mean glucose in 77% of the patients after Dexcom G6 monitoring was started. For the full cohort, there was a 14% decrease (235.7 to 202.7 mg/dL, P=0.0003) in mean sensor glucose. Sadhu et al. [29] reported similar accuracy between the Medtronic Guardian Connect (Medtronic Minimed, Northridge, CA, USA) and Dexcom G6 devices in a cohort of 11 ICU patients. This ICU data is promising for eventual routine use of CGMs in an ICU setting. However, these studies have limitations, including small sample sizes, inconsistent reference matrices (either capillary or arterial blood), and little information on glycemic or clinical outcomes.
Table 2.
Study | Year | First author country | Population | CGM type | CGM manufacturer | Performance measurement | Comparator |
---|---|---|---|---|---|---|---|
Goldberg et al. [30] | 2004 | USA | ICU (n=22) | CGMs | Medtronic MiniMed | Accuracy | Capillary by POC |
Vriesendorp et al. [31] | 2005 | Netherlands | OR, SICU (n=8) | CGMs and GlucoDay | Medtronic MiniMed and A. Menarini Diagnostics (A. Menarini Diagnostics Ltd., Florence, Italy) | Accuracy and feasibility | Arterial by blood gas analyzer |
Corstjens et al. [32] | 2006 | Netherlands | MICU (n=45) | System Gold | Medtronic MiniMed | Accuracy | Arterial by blood gas analyzer, YSI (YSI 2300 STAT Plus glucose and lactate analyzer, YSI Life Science, Yellow Springs, OH, USA) and POC |
De Block et al. [33] | 2006 | Belgium | MICU (n=50) | Glucoday | A. Menarini Diagnostics | Reliability | Arterial |
Price et al. [34] | 2008 | UK | Mixed ICU (n=17) | Guardian | Medtronic MiniMed | Accuracy | Arterial by blood gas analyzer and POC |
Logtenberg et al. [35] | 2009 | Netherlands | Cardiac surgery ICU (n=30) | Paradigm | Medtronic MiniMed | Accuracy and glycemic control | Capillary, arterial, and venous by POC |
Yamashita et al. [36] | 2009 | Japan | ICU (n=50) | STG 22 | Nikkiso Co. Ltd. (Tokyo, Japan) | Accuracy | Arterial by blood gas analyzer |
Rabiee et al. [37] | 2009 | USA | SICU/Burn (n=19) | Dexcom STS | Dexcom | Accuracy and reliability | Capillary by POC and serum by lab |
Holzinger et al. [38] | 2009 | Austria | MICU (n=50) | System Gold | Medtronic MiniMed | Accuracy and reliability | Arterial by blood gas analyzer |
Holzinger et al. [39] | 2010 | Austria | ICU, mechanical ventilation (n=24) | Guardian | Medtronic MiniMed | Glycemic control (% time at glucose <110 mg/dL), LOS, mortality | Arterial by blood gas analyzer and blinded Medtronic MiniMed System Gold CGM |
Jacobs et al. [40] | 2010 | USA | ICU (n=29) | Guardian RT | Medtronic MiniMed | Accuracy and feasibility | Capillary by POC |
Brunner et al. [41] | 2011 | Austria | MICU (n=174) | Guardian & System Gold | Medtronic MiniMed | Accuracy and reliability | Arterial by blood gas analyzer |
Kalmovich et al. [42] | 2012 | Israel | Perioperative cardiac surgery (n=32) | System Gold Blinded | Medtronic MiniMed | Accuracy and feasibility | Venous by blood gas analyzer |
Lorencio et al. [43] | 2012 | Spain | ICU (n: 41) | Guardian | Medtronic MiniMed | Accuracy | Arterial by blood gas analyzer |
Kopecky et al. [44] | 2013 | Czech Republic | Cardiac ICU (n=24) | Guardian RT | Medtronic MiniMed | Accuracy and glycemic control | Arterial by blood gas analyzer and computer (enhanced model predictive control) algorithm alone |
Leelarathna et al. [45] | 2013 | UK | Neurosurgical ICU (n=24) | FreeStyle Navigator | Abbott Diabetes Care | Glycemic control | Arterial by blood gas analyzer |
Rodriguez-Quintanilla et al. [46] | 2013 | Mexico | ICU (n=16) | Guardian RT | Medtronic MiniMed | Time to normoglycemia | Venous and capillary by POC |
Schuster et al. [47] | 2014 | USA | SICU (n=24) | Guardian | Medtronic MiniMed | Accuracy | Capillary by POC |
Kosiborod et al. [48] | 2014 | USA | Cardiac ICU (n=21) | Sentrino | Medtronic MiniMed | Accuracy and reliability | Central venous by POC or lab |
Boom et al. [49] | 2014 | Netherlands | MICU/SICU (n=156) | FreeStyle Navigator | Abbott Diabetes Care | Accuracy and glycemic control | Arterial by blood gas analyzer, and POC |
Umbrello et al. [50] | 2014 | Italy | MICU (n=6) | OptiScanner 5000 | OptiScan Biomedical | Glycemic control | Central venous by blood gas analyzer or lab (reported elsewhere) |
van Hooijdonk et al. [51] | 2015 | Netherlands | ICU (n=50) | Sentrino | Medtronic MiniMed | Accuracy and reliability | Arterial by blood gas analyzer |
Sechterberger et al. [52] | 2015 | Netherlands | Cardiac ICU (n=8) | FreeStyle Navigator | Abbott Diabetes Care | Accuracy | Arterial by blood gas analyzer |
Punke et al. [53] | 2015 | Germany | SICU (n=14) | Sentrino | Medtronic MiniMed | Accuracy | Arterial by blood gas analyzer |
Ballesteros et al. [54] | 2015 | Spain | MICU (n=18) | Soft Sensor | Medtronic MiniMed | Accuracy | Capillary by POC |
De Block et al. [55] | 2015 | Belgium | MICU (n=35) | GlucoDay S | A. Menarini Diagnostics | Accuracy and glycemic control | Arterial by blood gas analyzer and blinded microdialysis-based CGM |
Gottschalk et al. [56] | 2016 | Germany | Extracorporeal cardiac life support (n=25) | Sentrino | Medtronic MiniMed | Accuracy | Arterial by blood gas analyzer |
Nohra et al. [57] | 2016 | USA | SICU (n=23) | Optiscanner 5000 | Optiscan Biomedical | Accuracy | Central venous by YSI |
Righy Shinotsuka et al. [58] | 2016 | Belgium | ICU (n=88) | OptiScanner 5000 | Optiscan Biomedical | Accuracy | Arterial by YSI |
Wollersheim et al. [59] | 2016 | Germany | MICU (n=20) | Sentrino | Medtronic MiniMed | Accuracy and feasibility | Arterial, central venous, or venous by blood gas analyzer |
Schierenbeck et al. [60] | 2017 | Sweden | Cardiac ICU (n=26) | FreeStyle Libre Subcutaneous-CGM vs. Eirus Intravascular | Abbott Diabetes Care and Maquet Getinge Group | Accuracy | Arterial by blood gas analyzer and capillary by POC |
Song et al. [61] | 2017 | Republic of Korea | OR, ICU (n=22) | Guardian | Medtronic MiniMed | Accuracy and reliability | Arterial by blood gas analyzer |
Ancona et al. [62] | 2017 | Australia | ICU (n=8) | FreeStyle Libre CGM | Abbott Diabetes Care | Accuracy and feasibility | Arterial by blood gas analyzer or capillary by POC |
Bochicchio et al. [63] | 2017 | USA | ICU (n=243) | OptiScanner 5000 | OptiScan Biomedical | Accuracy | Arterial, central venous, or venous by YSI |
Rijkenberg et al. [64] | 2018 | Netherlands | Mixed ICU (n=155) | FreeStyle Navigator | Abbott Diabetes Care | Accuracy and reliability | Arterial by blood gas analyzer |
Nukui et al. [65] | 2019 | Japan | Acute stroke (n=39) | FreeStyle Pro CGM | Abbott Diabetes Care | Accuracy and efficacy | Capillary by POC |
Furushima et al. [66] | 2020 | Japan | ICU (n=40) | FreeStyle Libre CGM | Abbott Diabetes Care | Determine the mean amplitude of glycemic excursions | Arterial by blood gas analyzer |
Chow et al. [67] | 2020 | ||||||
USA | ICU (n=1) | Dexcom G6 | Dexcom | Accuracy | Capillary by POC and venous (methods not specified) | ||
Sadhu et al. [29] | 2020 | USA | ICU (n=11) | Guardian Connect Dexcom G6 |
Medtronic MiniMed Dexcom |
Accuracy and feasibility | Capillary, venous, and arterial by POC |
Garelli et al. [68] | 2020 | Argentina | ICU patients with COVID-19 (n=3) (2 other pediatric patients were also studied) | Dexcom G6 for the ICU patient_s | Dexcom | Glycemic control using a new multisensor platform | None |
Agarwal et al. [27] | 2020 | ||||||
USA | ICU (n=47) | Dexcom G6 | Dexcom | Accuracy | Capillary by POC | ||
Chow et al. [28] | 2021 | USA | ICU patients with COVID-19 (n=30) | Dexcom G6 | Dexcom | Clinical utility and accuracy | Arterial by POC |
Perez-Guzman et al. [26] | 2021 | USA | OR and cardiac ICU patients without diabetes undergoing scheduled or urgent coronary artery bypass surgery | Dexcom G6 | Dexcom | Accuracy | Capillary by POC |
CGM, continuous glucose monitor; ICU, intensive care unit; POC, point of care; OR, operating room; SICU, surgical ICU; MICU, medical ICU; COVID-19, coronavirus disease 2019.
Table 3.
Study | Year | First author country | Population | Type of CGM | CGM manufacturer | Performance measurement | Comparator |
---|---|---|---|---|---|---|---|
Bridges et al. [69] | 2010 | USA | ICU (n=47) | Guardian | Medtronic MiniMed | Accuracy | Arterial, venous, and capillary by iSTAT POC and lab |
Steil et al. [70] | 2011 | USA | Cardiac ICU (n=311) | Guardian | Medtronic MiniMed | Accuracy and hypoglycemia prevention | Arterial by POC and lab |
Prabhudesai et al. [71] | 2015 | UK | ICU (n=19) | Guardian | Medtronic MiniMed | Accuracy | Arterial by lab |
Kotzapanagiotou et al. [72] | 2020 | Greece | ICU (n=16) | FreeStyle Libre | Abbott Diabetes Care | Accuracy | Arterial by blood gas analyzer capillary by POC, biochemical serum by lab |
Sopfe et al. [73] | 2020 | USA | Stem cell transplantation (n=29) | FreeStyle Libre Pro | Abbott Diabetes Care | Accuracy | Central venous by lab |
Garelli et al. [68] | 2020 | Argentina | ICU patients with COVID-19 (n=2) (3 other adult patients were also studied) | Dexcom G6 for the ICU patients | Dexcom | Glycemic control using a new multisensor platform | None |
Chesser et al. [74] | 2021 | USA | Children with postprandial hypoglycemia due to late dumping syndrome following gastric surgeries (n=3) | Dexcom G4 Dexcom G5 Dexcom G6 |
Dexcom | Diagnose hypoglycemia due to late dumping syndrome. Also to determine best treatment strategies and feeding regimens. | None |
AID, automated insulin delivery; CGM, continuous glucose monitoring; ICU, intensive care unit; POC, point of care; COVID-19, coronavirus disease 2019.
As the information about CGMs in the ICU is evolving, we currently recommend using CGM in the ICU for selected candidates, such as patients with COVID-19, who (1) are treated with continuous intravenous insulin infusion; (2) develop steroid-induced hyperglycemia; or (3) have medical nutrition therapy-induced hyperglycemia or high glycemic variability. We also recommend using a hybrid approach (combining CGM with periodical POC blood glucose testing) in the ICU to ensure consistent accuracy. CGM is currently not widely used for non-COVID patients outside of research settings.
WHO IS A CANDIDATE FOR USING A CGM IN THE WARDS
An emerging use for CGM in the hospital will be detection and prevention of hypoglycemia on the wards among insulin-treated patients. This is a new application of CGM and the early evidence from trials to detect this complication of hospitalizations are promising. Table 4 presents a review of the literature of clinical trials of CGM use in non-ICU settings for adult patients [75–89]. Ward patients sometimes need to have their ongoing insulin dose significantly decreased suddenly if they miss a meal or if they are receiving a regimen of corticosteroids or parenteral nutrition, which is suddenly curtailed or discontinued, and the communication between the physicians, nurses, and pharmacists is not optimal. Hypoglycemia detection is particularly useful in a setting where ward nurses are not always with the patient and where a patient might have hypoglycemia unawareness or might fail to complain of hypoglycemic symptoms and then slowly become confused or even unconscious. Table 5 presents a review of the literature of clinical trials of CGM use to detect hypoglycemia in hospitalized patients [70,74,79,84,86, 87]. It is likely that CGM technology will be used frequently for this purpose in the future.
Table 4.
Study | Year | First author country | Patient population | CGM type | CGM manufacturer | Performance measurement | Comparator |
---|---|---|---|---|---|---|---|
Dungan et al. [77] | 2012 | USA | T1DM and T2DM (n=58), on intravenous or subcutaneous insulin | iPro system | Medtronic MiniMed | Accuracy | Capillary by POC |
Burt et al. [78] | 2013 | Australia | T1DM and T2DM, on basal bolus insulin (n=26) | System Gold | Medtronic MiniMed | Accuracy and glycemic control | Capillary by POC |
Gomez et al. [79] | 2015 | Colombia | T2DM, on basal bolus insulin (n=38) | iPro2 system | Medtronic MiniMed | Glycemic control and hypoglycemia detection | Capillary by POC |
Schaupp et al. [80] | 2015 | Austria | T2DM, on basal bolus insulin (n=84) | iPro2 system | Medtronic MiniMed | Accuracy | Capillary by POC |
Spanakis et al. [81] | 2018 | USA | T2DM, on insulin therapy (n=5) | Dexcom G4 CGM with Share2 application | Dexcom | Glucose telemetry system feasibility | None |
Migdal et al. [82] | 2020 | USA | Adult medicine and surgery patients with T1DM and T2DM (n=49) | Dexcom (CGM type unspecified) | Dexcom | Precision and accuracy | Capillary by POC |
Shehav-Zaltzman et al. [83] | 2020 | Israel | T1DM on CSII (n: 1) and T2DM on basal bolus (n=3), COVID-19 wards (n=5) | Guardian | Medtronic MiniMed | Feasibility | None |
Singh et al. [84] | 2020 | USA | T2DM, on basal-bolus insulin (n=13) | Dexcom G4 Platinum CGM | Dexcom | Feasibility and prevention of hypoglycemia | Blinded CGM |
Tripyla et al. [75] | 2020 | Switzerland | Prediabetes patients undergoing elective abdominal surgery (n=20) | Dexcom G6 | Dexcom | Accuracy | Capillary by POC |
Reutrakul et al. [85] | 2020 | USA | Diabetes (unspecified type) on subcutaneous insulin injection with COVID-19 (n=1) | Dexcom G6 | Dexcom | Feasibility | Capillary by POC |
Galindo et al. [86] | 2020 | USA | T2DM, on basal-bolus insulin (n=97) | FreeStyle Libre Pro CGM | Abbott Diabetes Care | Accuracy and hypoglycemia detection | Capillary by POC |
Nair et al. [76] | 2020 | USA | Surgical ward (n=10) | Dexcom G6 Blinded | Dexcom | Accuracy | Capillary by POC |
Singh et al. [87] | 2020 | USA | T2DM, on basal-bolus insulin (n=72) | Dexcom G6 | Dexcom | Prevention of hypoglycemia | Blinded CGM |
Fortmann et al. [88] | 2020 | USA | T2DM on subcutaneous insulin (n=110) | Dexcom G6 | Dexcom | Effectiveness | Capillary by POC |
Ushigome et al.[89] | 2021 | Japan | Diabetes (unspecified type) with COVID-19 (n=1) | Dexcom G4 Platinum | Dexcom | Safety and effectiveness | Lab |
CGM, continuous glucose monitor; ICU, intensive care unit; T1DM, type 1 diabetes mellitus; T2DM, type 2 diabetes mellitus; POC, point of care; CSII, continuous subcutaneous insulin infusion; COVID-19, coronavirus disease 2019.
Table 5.
Study | Year | First author country | Patient population | CGM type | CGM manufacturer | Definition of hypoglycemia | Outcome |
---|---|---|---|---|---|---|---|
Steil et al. [70] | 2011 | USA | Cardiac ICU (n=311) | Guardian | Medtronic MiniMed | Blood glucose <60 mg/dL (3.3 mmol/L) | No reduction if CGM alarm was set at 60 mg/dL. 18 out of 40 episodes of hypoglycemia detected when the alarm threshold set to 70 mg/dL. One to two false hypoglycemia alarms in each patient |
Gomez et al. [79] | 2015 | Colombia | T2DM, on basal bolus insulin (n=38) | iPro2 system | Medtronic MiniMed | Hypoglycemia was defined as blood glucose <70 mg/dL (3.9 mmol/L) or <60 mg/dL (3.3 mmol/L). | CGM is more effective than POC testing for detecting hypoglycemic episodes and asymptomatic hypoglycemia using either definition of hypoglycemia. |
Singh et al. [84] | 2020 | USA | T2DM, on basal-bolus insulin (n=13) | Dexcom G4 Platinum CGM | Dexcom | Blood glucose <70 mg/dL (3.9 mmol/L) | A hypoglycemia prevention protocol using a specific Glucose Telemetry System can reduce incidence of inpatient hypoglycemia |
Galindo et al. [86] | 2020 | USA | T2DM, on basal-bolus insulin (n=97) | FreeStyle Libre Pro CGM | Abbott Diabetes Care | Hypoglycemia was defined as <70 mg/dL (3.9 mmol/L) or <54 mg/dL (3.0 mmol/L). | Hypoglycemic events were detected more often by CGM use than POC testing. |
Singh et al. [87] | 2020 | USA | T2DM, on basal-bolus insulin (n=72) | Dexcom G6 | Dexcom | Hypoglycemia was defined as blood glucose <70 mg/dL (3.9 mmol/L) for over 15 minutes. Clinically significant hypoglycemia was defined as blood glucose <54 mg/dL (3.0 mmol/L). |
In patients with type 2 diabetes who have been treated with insulin, hypoglycemia can be decreased by a combination of RT-CGM use with a protocol for hypoglycemia prevention. |
Chesser et al. [74] | 2021 | USA | Children with postprandial hypoglycemia due to late dumping syndrome following gastric surgeries (n=3) | Dexcom G4 Dexcom G5 Dexcom G6 |
Dexcom | Hypoglycemia was not explicitly defined, but one patient whose glucose level was up to 65 mg/dL was considered hypoglycemic. | CGM can be used for early diagnosis of dumping syndrome by revealing glycemic dysregulation. It can also be used to evaluate the effectiveness of treatments and feeding regimens for postprandial hypoglycemia due to late dumping syndrome. |
CGM, continuous glucose monitor; ICU, intensive care unit; T2DM, type 2 diabetes mellitus; POC, point of care; RT-CGM, real-time continuous glucose monitor.
Based on our experience and the results of recent studies of currently available CGM systems, we believe that selected types of patients could benefit from wearing a CGM in a medical or surgical ward. These include patients with (1) a high risk of hypoglycemia (e.g., with a fragile habitus, end stage renal disease, advanced age, or poor nutrition); (2) type 1 diabetes; (3) a requirement for multiple daily insulin injections; (4) high glycemic variability; (5) steroid-induced hyperglycemia; and (6) enteral or parenteral feeding-induced hyperglycemia.
WHO IS A CANDIDATE FOR USING A CGM IN SURGERY
The opportunity to see a full glycemic profile during the operative time will be ideal for an anesthesiologist (who will receive glucose readings every 5 to 15 minutes) managing patients with or without diabetes (including those with stress hyperglycemia). However, until recently, accuracy and performance data of CGMs during surgery has been unreliable. One recent study by Tripyla et al. [75] reported a MARD of 12.7% with the Dexcom G6 during elective abdominal surgery. Nair et al. [76] performed a prospective pilot study, in which they included 10 adult patients with a diagnosis of diabetes who were undergoing elective general surgery. They found that postoperatively, Dexcom G6 had a MARD of 9.4%. In 15 patients undergoing coronary artery bypass graft surgery, Perez-Guzman et al. [26] noted that CGM technology was not consistently reliable in the operating room (OR), which they attributed to electrocautery interference. They observed signal loss and negative bias during surgery in 60% of their patients. However, some sensors recovered immediately after surgery and had sustained accuracy postoperatively, even during exposure to vasopressors in the ICU. Based on these results they suggested avoiding the use of a CGM to make clinical decisions during surgery unless adequate accuracy of the device is confirmed and the possibility of needing a new sensor is excluded [26]. We do not recommend that any patient in the OR should be managed with a CGM system. For patients undergoing surgery who are already using a CGM, we recommend confirming accuracy in the immediate postoperative period, because interference due to electrocautery may occur in the OR causing temporary or permanent device dysfunction [26].
RESPONSIBILITIES OF PHYSICIANS, HOSPITALS, AND INDUSTRY
In September 2020, Diabetes Technology Society published the ‘Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline’ to address how and when to best use both subcutaneous CGMs and automated insulin delivery (AID) systems, as well as to promote clinical research utilizing these devices [21]. The consensus panel of 24 international experts in the use of CGM developed recommendations for physicians (which the guideline referred to as healthcare professionals), nurses, and hospitals, as well as for industry. The panel covered five topics: (1) continuation of home CGMs after hospitalization; (2) initiation of CGMs in the hospital; (3) continuation of automated insulin dosing systems in the hospital; (4) logistics and hands-on care of hospitalized patients using CGMs and automated insulin dosing systems; and (5) data management of CGMs. The panelists voted on 78 proposed recommendations and 77 recommendations were endorsed and classified as either strong (80% to 100% agreement) or mild (60% to 79% agreement). One recommendation failed to reach consensus.
IMPLICATIONS OF CGM USE FOR BEDSIDE NURSES
As inpatient CGM use increases in people with both type 1 and type 2 diabetes, the bedside nurse needs to become familiar with this technology and its advantages and limitations. Understanding that ISF glucose lags behind capillary glucose is important when recognizing and treating hypoglycemia. For example, a patient may feel fine after receiving an appropriate hypoglycemia treatment, and the fingerstick value may be above the recommended threshold of 80 to 100 mg/dL, but the sensor glucose may still reflect a value under 70 mg/dL. Any discrepancies between patients’ symptoms and sensor numbers require a confirmatory finger stick POC capillary blood glucose measurement.
Most medical centers still require finger stick values for dosing insulin as a safety precaution, as if the devices are intended for adjunctive use, even though for outpatients the FDA-cleared factory-calibrated CGMs are approved for non-adjunctive use. Hospitals need a written policy and/or care guideline to provide medical and nursing staff with a system for charting decisions based on CGM data because these devices are not cleared for hospital use and it may be difficult to enter sensor values into the EHR.
Another care concern is documentation of the sensor placement location and inspection of the site any time from every shift to daily. For all radiologic studies, the sensor will need to be removed from the skin and the transmitter and/or receiver should be safely stored. As most medical centers do not carry sensor supplies, patients being admitted should be advised to bring additional sensor supplies.
HOSPITAL CGM DATA
CGM data for a hospitalized patient cannot be confirmed through a test using a reference method because it is almost impossible to assay ISF. Many hospital clinical chemists consider CGM data analogous to vital sign information, which is being monitored frequently but which cannot be individually confirmed for accuracy. A CGM can be calibrated in the factory for outpatient use, but it is not known whether this type of calibration will be adequate for a hospitalized patient.
CGM data is currently not compatible with the vast majority of EHRs. A screen shot of a tracing cannot be searched and therefore, this type of record will have only limited value after the day it is collected. The selection of which types of CGM data to store will likely be various 24-hour or 14-day metrics, such as time in range, glycemic variability, or percentage of time spent in a slightly low, extremely low, slightly high, and extremely high range [90], because the amount of storage space needed to save every data point (up to 288 glucose readings per day or more) could overwhelm many hospital EHRs. Hospitals, CGM manufacturers, and EHR vendors must work together to integrate the most useful CGM data into the EHR. This data acquisition process must be compliant with regulatory privacy rules and sound cybersecurity policies. Successful integration of real-time CGM data into the EHR has been reported in a limited number of cases, so there is hope for the future [91,92]. If CGM data of a hospitalized patient must reside on the website of a CGM manufacturer or a software integrator of data from multiple CGM manufacturers, then the hospital might have concerns about how the data is protected from cyber breaches and who is liable if such data from a hospitalization is lost.
FUTURE TECHNOLOGY
It is likely that future CGM sensors will be more accurate with continued improvements in sensor sensitivity and algorithm fidelity for converting a signal to a glucose concentration. These sensors will likely have a smaller form factor, a longer life, and louder alarms for out of range glucose concentrations. With longer-duration sensors, it will be necessary to develop better adhesives that are less likely to cause a rash [93]. In order to prolong viable implantation time, it will be necessary to develop better coatings that will minimize the foreign body response that limits integration of the sensor into the subcutaneous space [94]. Future CGM data will become interconnected with multiple other data streams to provide a more nuanced pattern of how behavior affects glycemia and to better predict glucose patterns [95]. Data fusion technology will be combined with decision support and behavioral modification software for real-time management and in the hospital this management will be in the form of warnings and treatment recommendations for the hospital staff [96].
Just as in the past, many did not expect that a subcutaneous glucose sensor would ever provide accurate glucose readings comparable to blood glucose testing, many now believe that a wearable optical noninvasive glucose monitor [97] or a monitor that is based on noninvasive collection and measurement of a body fluid (such as sweat, saliva, or tears) [98] cannot be developed to measure glucose almost as accurately as an invasive blood test. These doubters of optical noninvasive technology might prove to be just as wrong as the doubters of subcutaneous minimally invasive technology were 20 years ago. Whether these new technologies can become established will depend significantly on whether human physiology will allow for body compartments other than blood or skin to be measured without significant lag during periods of dynamic fluctuations—not on the expected accuracy of future sensors to measure these matrices. Eventually it will become clear which individual or composite metrics for CGMs (if any) and which definitions of a hypoglycemic episode will best correlate with specific hospital outcomes [99].
CONCLUSIONS
CGMs are gradually migrating from the outpatient setting into the inpatient setting because these devices are becoming more compatible with the needs of the healthcare team that cares for these inpatients. The trend accelerated because of the COVID-19 pandemic, which necessitated collecting data remotely from patients when possible. It is too soon to know whether the accuracy and clinical benefits of CGMs will result in widespread adoption or regulatory clearance of this technology for hospital use. CGMs work extremely well on outpatients and there is reason to believe that they could prove to be effective to achieve well-defined endpoints for hospitalized patients.
Acknowledgments
We thank Francisco Pasquel, MD for his helpful advice. We also thank Annamarie Sucher-Jones for her expert editorial assistance.
Footnotes
CONFLICTS OF INTEREST
M. Citlalli Perez-Guzman, Trisha Shang, and Jennifer Y. Zhang have nothing to disclose. Donna Jornsay was a speaker for Abbott Diabetes Care and BD, and holds Medtronic stock. David C. Klonoff is a consultant for EOFlow, Fractyl, Lifecare, Novo, Roche, Samsung, and Thirdwayv.
REFERENCES
- 1.Galindo RJ, Aleppo G. Continuous glucose monitoring: the achievement of 100 years of innovation in diabetes technology. Diabetes Res Clin Pract. 2020;170:108502. doi: 10.1016/j.diabres.2020.108502. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Acciaroli G, Vettoretti M, Facchinetti A, Sparacino G. Calibration of minimally invasive continuous glucose monitoring sensors: state-of-the-art and current perspectives. Biosensors (Basel) 2018;8:24. doi: 10.3390/bios8010024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring. Diabetes Technol Ther. 2015;17:177–86. doi: 10.1089/dia.2014.0272. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Abbott. The FreeStyle Libre 14 day system [Internet] Chicago: Abbott; 2021. [cited 2021 Feb 24]. Available from: https://www.freestyle.abbott/us-en/products/freestyle-14-day.html. [Google Scholar]
- 5.Abbott. FreeStyle Libre 2 [Internet] Chicago: Abbott; 2021. [cited 2021 Feb 24]. Available from: https://www.freestyle.abbott/us-en/products/freestyle-libre-2.html. [Google Scholar]
- 6.Dexcom. Dexcom G6 CGM System [Internet] San Diego: Dexcom; 2021. [cited 2021 Feb 24]. Available from: https://www.dexcom.com/g6-cgm-system. [Google Scholar]
- 7.Galindo RJ, Aleppo G, Klonoff DC, Spanakis EK, Agarwal S, Vellanki P, et al. Implementation of continuous glucose monitoring in the hospital: emergent considerations for remote glucose monitoring during the COVID-19 pandemic. J Diabetes Sci Technol. 2020;14:822–32. doi: 10.1177/1932296820932903. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Abbott. FreeStyle Libre flash glucose monitoring system: user’s manual 2016 [Internet] Chicago: Abbott; 2016. [cited 2021 Feb 24]. Available from: https://freestyleserver.com/Payloads/IFU/2017_dec/ART34745-107_rev-A-WEB.pdf. [Google Scholar]
- 9.Abbott. Chicago: Abbott; 2021. FreeStyle Libre: Apps [Internet] [cited 2021 Feb 24]. Available from: https://www.freestyle.abbott/us-en/products/continous-glucose-monitor-app.html. [Google Scholar]
- 10.Abbott. FreeStyle Libre 2 flash glucose monitoring system: user’s manual 2020 [Internet] Chicago: Abbott; 2020. [cited 2021 Feb 24]. Available from: https://freestyleserver.com/Payloads/IFU/2020/q2/ART40703-001_rev-D-Web.pdf. [Google Scholar]
- 11.Alva S, Bailey T, Brazg R, Budiman ES, Castorino K, Christiansen MP, et al. Accuracy of a 14-day factory-calibrated continuous glucose monitoring system with advanced algorithm in pediatric and adult population with diabetes. J Diabetes Sci Technol. 2020 Sep 19; doi: 10.1177/1932296820958754. [Epub]. [DOI] [PMC free article] [PubMed]
- 12.Calhoun P, Johnson TK, Hughes J, Price D, Balo AK. Resistance to acetaminophen interference in a novel continuous glucose monitoring system. J Diabetes Sci Technol. 2018;12:393–6. doi: 10.1177/1932296818755797. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Dexcom. Dexcom G6 continuous glucose monitoring system: user guide 2020 [Internet] San Diego: Dexcom; 2020. [cited 2021 Feb 24]. Available from: https://s3-us-west-2.amazonaws.com/dexcompdf/G6-CGM-Users-Guide.pdf. [Google Scholar]
- 14.Dexcom. Dexcom: Apps [Internet] San Diego: Dexcom; 2021. [cited 2021 Feb 24]. Available from: https://www.dexcom.com/apps. [Google Scholar]
- 15.Freckmann G, Pleus S, Schauer S, Link M, Jendrike N, Waldenmaier D, et al. Choice of continuous glucose monitoring systems may affect metrics: clinically relevant differences in times in ranges. Exp Clin Endocrinol Diabetes. 2021 Jan 28; doi: 10.1055/a-1347-2550. [Epub]. [DOI] [PubMed]
- 16.Klonoff DC, Ahn D, Drincic A. Continuous glucose monitoring: a review of the technology and clinical use. Diabetes Res Clin Pract. 2017;133:178–92. doi: 10.1016/j.diabres.2017.08.005. [DOI] [PubMed] [Google Scholar]
- 17.Dexcom. Information request for healthcare providers: use of Dexcom continuous glucose monitoring systems during the COVID-19 pandemic [Internet] San Diego: Dexcom; 2021. [cited 2021 Feb 24]. Available from: https://www.dexcom.com/hospitalcovid-19. [Google Scholar]
- 18.Dexcom. Fact sheet for healthcare providers: use of Dexcom continuous glucose monitoring systems during the COVID-19 pandemic [Internet] San Diego: Dexcom; 2021. [cited 2021 Feb 24]. Available from: https://www.dexcom.com/hospitalfacts. [Google Scholar]
- 19.Abbott. Abbott’s FreeStyle® Libre 14 day system now available in U.S. for hospitalized patients with diabetes during COVID-19 pandemic [Internet] Chicago: Abbott; 2021. [cited 2021 Feb 24]. Available from: https://abbott.mediaroom.com/2020-04-08-Abbotts-FreeStyle-R-Libre-14-Day-System-Now-Available-in-U-S-for-Hospitalized-Patients-with-Diabetes-During-COVID-19-Pandemic. [Google Scholar]
- 20.Yeh T, Yeung M, Mendelsohn Curanaj FA. Managing patients with insulin pumps and continuous glucose monitors in the hospital: to wear or not to wear. Curr Diab Rep. 2021;21:7. doi: 10.1007/s11892-021-01375-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Galindo RJ, Umpierrez GE, Rushakoff RJ, Basu A, Lohnes S, Nichols JH, et al. Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline. J Diabetes Sci Technol. 2020;14:1035–64. doi: 10.1177/1932296820954163. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Pasquel FJ, Lansang MC, Dhatariya K, Umpierrez GE. Management of diabetes and hyperglycaemia in the hospital. Lancet Diabetes Endocrinol. 2021;9:174–88. doi: 10.1016/S2213-8587(20)30381-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Klonoff DC, Perz JF. Assisted monitoring of blood glucose: special safety needs for a new paradigm in testing glucose. J Diabetes Sci Technol. 2010;4:1027–31. doi: 10.1177/193229681000400501. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Davis GM, Faulds E, Walker T, Vigliotti D, Rabinovich M, Hester J, et al. Remote continuous glucose monitoring with a computerized insulin infusion protocol for critically Ill patients in a COVID-19 medical ICU: proof of concept. Diabetes Care. 2021;44:1055–8. doi: 10.2337/dc20-2085. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Baysal N, Cameron F, Buckingham BA, Wilson DM, Chase HP, Maahs DM, et al. A novel method to detect pressure-induced sensor attenuations (PISA) in an artificial pancreas. J Diabetes Sci Technol. 2014;8:1091–6. doi: 10.1177/1932296814553267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Perez-Guzman MC, Duggan E, Gibanica S, Cardona S, Corujo-Rodriguez A, Faloye A, et al. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care. 2021;44:e50–2. doi: 10.2337/dc20-2386. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Agarwal S, Mathew J, Davis GM, Shephardson A, Levine A, Louard R, et al. Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic. Diabetes Care. 2021;44:847–9. doi: 10.2337/dc20-2219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Chow KW, Kelly DJ, Rieff MC, Skala PA, Kravets I, Charitou MM, et al. Outcomes and healthcare provider perceptions of real-time continuous glucose monitoring (rtCGM) in patients with diabetes and COVID-19 admitted to the ICU. J Diabetes Sci Technol. 2021 Jan 12; doi: 10.1177/1932296820985263. [Epub]. [DOI] [PMC free article] [PubMed]
- 29.Sadhu AR, Serrano IA, Xu J, Nisar T, Lucier J, Pandya AR, et al. Continuous glucose monitoring in critically Ill patients with COVID-19: results of an emergent pilot study. J Diabetes Sci Technol. 2020;14:1065–73. doi: 10.1177/1932296820964264. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Goldberg PA, Siegel MD, Russell RR, Sherwin RS, Halickman JI, Cooper DA, et al. Experience with the continuous glucose monitoring system in a medical intensive care unit. Diabetes Technol Ther. 2004;6:339–47. doi: 10.1089/152091504774198034. [DOI] [PubMed] [Google Scholar]
- 31.Vriesendorp TM, DeVries JH, Holleman F, Dzoljic M, Hoekstra JB. The use of two continuous glucose sensors during and after surgery. Diabetes Technol Ther. 2005;7:315–22. doi: 10.1089/dia.2005.7.315. [DOI] [PubMed] [Google Scholar]
- 32.Corstjens AM, Ligtenberg JJ, van der Horst IC, Spanjersberg R, Lind JS, Tulleken JE, et al. Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients. Crit Care. 2006;10:R135. doi: 10.1186/cc5048. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.De Block C, Manuel-Y-Keenoy B, Van Gaal L, Rogiers P. Intensive insulin therapy in the intensive care unit: assessment by continuous glucose monitoring. Diabetes Care. 2006;29:1750–6. doi: 10.2337/dc05-2353. [DOI] [PubMed] [Google Scholar]
- 34.Price GC, Stevenson K, Walsh TS. Evaluation of a continuous glucose monitor in an unselected general intensive care population. Crit Care Resusc. 2008;10:209–16. [PubMed] [Google Scholar]
- 35.Logtenberg SJ, Kleefstra N, Snellen FT, Groenier KH, Slingerland RJ, Nierich AP, et al. Pre- and postoperative accuracy and safety of a real-time continuous glucose monitoring system in cardiac surgical patients: a randomized pilot study. Diabetes Technol Ther. 2009;11:31–7. doi: 10.1089/dia.2008.0028. [DOI] [PubMed] [Google Scholar]
- 36.Yamashita K, Okabayashi T, Yokoyama T, Yatabe T, Maeda H, Manabe M, et al. Accuracy and reliability of continuous blood glucose monitor in post-surgical patients. Acta Anaesthesiol Scand. 2009;53:66–71. doi: 10.1111/j.1399-6576.2008.01799.x. [DOI] [PubMed] [Google Scholar]
- 37.Rabiee A, Andreasik V, Abu-Hamdah R, Galiatsatos P, Khouri Z, Gibson BR, et al. Numerical and clinical accuracy of a continuous glucose monitoring system during intravenous insulin therapy in the surgical and burn intensive care units. J Diabetes Sci Technol. 2009;3:951–9. doi: 10.1177/193229680900300443. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Holzinger U, Warszawska J, Kitzberger R, Herkner H, Metnitz PG, Madl C. Impact of shock requiring norepinephrine on the accuracy and reliability of subcutaneous continuous glucose monitoring. Intensive Care Med. 2009;35:1383–9. doi: 10.1007/s00134-009-1471-y. [DOI] [PubMed] [Google Scholar]
- 39.Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-time continuous glucose monitoring in critically ill patients: a prospective randomized trial. Diabetes Care. 2010;33:467–72. doi: 10.2337/dc09-1352. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Jacobs B, Phan K, Bertheau L, Dogbey G, Schwartz F, Shubrook J. Continuous glucose monitoring system in a rural intensive care unit: a pilot study evaluating accuracy and acceptance. J Diabetes Sci Technol. 2010;4:636–44. doi: 10.1177/193229681000400318. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Brunner R, Kitzberger R, Miehsler W, Herkner H, Madl C, Holzinger U. Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients. Crit Care Med. 2011;39:659–64. doi: 10.1097/CCM.0b013e318206bf2e. [DOI] [PubMed] [Google Scholar]
- 42.Kalmovich B, Bar-Dayan Y, Boaz M, Wainstein J. Continuous glucose monitoring in patients undergoing cardiac surgery. Diabetes Technol Ther. 2012;14:232–8. doi: 10.1089/dia.2011.0154. [DOI] [PubMed] [Google Scholar]
- 43.Lorencio C, Leal Y, Bonet A, Bondia J, Palerm CC, Tache A, et al. Real-time continuous glucose monitoring in an intensive care unit: better accuracy in patients with septic shock. Diabetes Technol Ther. 2012;14:568–75. doi: 10.1089/dia.2012.0008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Kopecky P, Mraz M, Blaha J, Lindner J, Svacina S, Hovorka R, et al. The use of continuous glucose monitoring combined with computer-based eMPC algorithm for tight glucose control in cardiosurgical ICU. Biomed Res Int. 2013;2013 doi: 10.1155/2013/186439. 186439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Leelarathna L, English SW, Thabit H, Caldwell K, Allen JM, Kumareswaran K, et al. Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trial. Crit Care. 2013;17:R159. doi: 10.1186/cc12838. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Rodriguez-Quintanilla KA, Lavalle-Gonzalez FJ, Mancillas-Adame LG, Zapata-Garrido AJ, Villarreal-Perez JZ, Tamez-Perez HE. Continuous glucose monitoring in acute coronary syndrome. Arch Cardiol Mex. 2013;83:237–43. doi: 10.1016/j.acmx.2013.08.001. [DOI] [PubMed] [Google Scholar]
- 47.Schuster KM, Barre K, Inzucchi SE, Udelsman R, Davis KA. Continuous glucose monitoring in the surgical intensive care unit: concordance with capillary glucose. J Trauma Acute Care Surg. 2014;76:798–803. doi: 10.1097/TA.0000000000000127. [DOI] [PubMed] [Google Scholar]
- 48.Kosiborod M, Gottlieb RK, Sekella JA, Peterman D, Grodzinsky A, Kennedy P, et al. Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit. BMJ Open Diabetes Res Care. 2014;2:e000037. doi: 10.1136/bmjdrc-2014-000037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. Crit Care. 2014;18:453. doi: 10.1186/s13054-014-0453-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Umbrello M, Salice V, Spanu P, Formenti P, Barassi A, Melzi d’Eril GV, et al. Performance assessment of a glucose control protocol in septic patients with an automated intermittent plasma glucose monitoring device. Clin Nutr. 2014;33:867–71. doi: 10.1016/j.clnu.2013.10.007. [DOI] [PubMed] [Google Scholar]
- 51.van Hooijdonk RT, Leopold JH, Winters T, Binnekade JM, Juffermans NP, Horn J, et al. Point accuracy and reliability of an interstitial continuous glucose-monitoring device in critically ill patients: a prospective study. Crit Care. 2015;19:34. doi: 10.1186/s13054-015-0757-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Sechterberger MK, van der Voort PH, Strasma PJ, DeVries JH. Accuracy of intra-arterial and subcutaneous continuous glucose monitoring in postoperative cardiac surgery patients in the ICU. J Diabetes Sci Technol. 2015;9:663–7. doi: 10.1177/1932296814564993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Punke MA, Decker C, Wodack K, Reuter DA, Kluge S. Continuous glucose monitoring on the ICU using a subcutaneous sensor. Med Klin Intensivmed Notfmed. 2015;110:360–3. doi: 10.1007/s00063-014-0453-1. [DOI] [PubMed] [Google Scholar]
- 54.Ballesteros D, Martinez O, Blancas Gomez-Casero R, Martin Parra C, Lopez Matamala B, Estebanez B, et al. Continuous tissue glucose monitoring correlates with measurement of intermittent capillary glucose in patients with distributive shock. Med Intensiva. 2015;39:405–11. doi: 10.1016/j.medin.2014.09.004. [DOI] [PubMed] [Google Scholar]
- 55.De Block CE, Gios J, Verheyen N, Manuel-y-Keenoy B, Rogiers P, Jorens PG, et al. Randomized evaluation of glycemic control in the medical intensive care unit using real-time continuous glucose monitoring (REGIMEN Trial) Diabetes Technol Ther. 2015;17:889–98. doi: 10.1089/dia.2015.0151. [DOI] [PubMed] [Google Scholar]
- 56.Gottschalk A, Welp HA, Leser L, Lanckohr C, Wempe C, Ellger B. Continuous glucose monitoring in patients undergoing extracorporeal ventricular assist therapy. PLoS One. 2016;11:e0148778. doi: 10.1371/journal.pone.0148778. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Nohra E, Buckman S, Bochicchio K, Chamieh J, Reese S, Merrill C, et al. Results of a near continuous glucose monitoring technology in surgical intensive care and trauma. Contemp Clin Trials. 2016;50:1–4. doi: 10.1016/j.cct.2016.07.007. [DOI] [PubMed] [Google Scholar]
- 58.Righy Shinotsuka C, Brasseur A, Fagnoul D, So T, Vincent JL, Preiser JC. Manual versus automated moNitoring accuracy of glucosE II (MANAGE II) Crit Care. 2016;20:380. doi: 10.1186/s13054-016-1547-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Wollersheim T, Engelhardt LJ, Pachulla J, Moergeli R, Koch S, Spies C, et al. Accuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: a prospective clinical trial. Ann Intensive Care. 2016;6:70. doi: 10.1186/s13613-016-0167-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 different continuous glucose monitoring systems in patients undergoing cardiac surgery. J Diabetes Sci Technol. 2017;11:108–16. doi: 10.1177/1932296816651632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Song IK, Lee JH, Kang JE, Park YH, Kim HS, Kim JT. Continuous glucose monitoring system in the operating room and intensive care unit: any difference according to measurement sites? J Clin Monit Comput. 2017;31:187–94. doi: 10.1007/s10877-015-9804-6. [DOI] [PubMed] [Google Scholar]
- 62.Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Martensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19:167–74. [PubMed] [Google Scholar]
- 63.Bochicchio GV, Nasraway S, Moore L, Furnary A, Nohra E, Bochicchio K. Results of a multicenter prospective pivotal trial of the first inline continuous glucose monitor in critically ill patients. J Trauma Acute Care Surg. 2017;82:1049–54. doi: 10.1097/TA.0000000000001444. [DOI] [PubMed] [Google Scholar]
- 64.Rijkenberg S, van Steen SC, DeVries JH, van der Voort PHJ. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J Clin Monit Comput. 2018;32:953–64. doi: 10.1007/s10877-017-0086-z. [DOI] [PubMed] [Google Scholar]
- 65.Nukui S, Akiyama H, Soga K, Takao N, Tsuchihashi Y, Iijima N, et al. Risk of hyperglycemia and hypoglycemia in patients with acute ischemic stroke based on continuous glucose monitoring. J Stroke Cerebrovasc Dis. 2019;28:104346. doi: 10.1016/j.jstrokecerebrovasdis.2019.104346. [DOI] [PubMed] [Google Scholar]
- 66.Furushima N, Egi M, Obata N, Sato H, Mizobuchi S. Mean amplitude of glycemic excursions in septic patients and its association with outcomes: a prospective observational study using continuous glucose monitoring. J Crit Care. 2021;63:218–22. doi: 10.1016/j.jcrc.2020.08.021. [DOI] [PubMed] [Google Scholar]
- 67.Chow KW, Kelly DJ, Gupta R, Miller JD. Use of continuous glucose monitoring to assess parenteral nutrition-induced hyperglycemia in an adult patient with severe COVID-19. JPEN J Parenter Enteral Nutr. 2021;45:208–11. doi: 10.1002/jpen.2032. [DOI] [PubMed] [Google Scholar]
- 68.Garelli F, Rosales N, Fushimi E, Arambarri D, Mendoza L, De Battista H, et al. Remote glucose monitoring platform for multiple simultaneous patients at coronavirus disease 2019 intensive care units: case report including adults and children. Diabetes Technol Ther. 2020 Dec 18; doi: 10.1089/dia.2020.0556. [Epub]. [DOI] [PubMed]
- 69.Bridges BC, Preissig CM, Maher KO, Rigby MR. Continuous glucose monitors prove highly accurate in critically ill children. Crit Care. 2010;14:R176. doi: 10.1186/cc9280. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Steil GM, Langer M, Jaeger K, Alexander J, Gaies M, Agus MS. Value of continuous glucose monitoring for minimizing severe hypoglycemia during tight glycemic control. Pediatr Crit Care Med. 2011;12:643–8. doi: 10.1097/PCC.0b013e31821926a5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Prabhudesai S, Kanjani A, Bhagat I, Ravikumar KG, Ramachandran B. Accuracy of a real-time continuous glucose monitoring system in children with septic shock: a pilot study. Indian J Crit Care Med. 2015;19:642–7. doi: 10.4103/0972-5229.169331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Kotzapanagiotou E, Tsotridou E, Volakli E, Dimitriadou M, Chochliourou E, Kalamitsou S, et al. Evaluation of continuous flash glucose monitoring in a pediatric ICU setting. J Clin Monit Comput. 2020;34:843–52. doi: 10.1007/s10877-019-00384-y. [DOI] [PubMed] [Google Scholar]
- 73.Sopfe J, Vigers T, Pyle L, Giller RH, Forlenza GP. Safety and accuracy of factory-calibrated continuous glucose monitoring in pediatric patients undergoing hematopoietic stem cell transplantation. Diabetes Technol Ther. 2020;22:727–33. doi: 10.1089/dia.2019.0521. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Chesser H, Abdulhussein F, Huang A, Lee JY, Gitelman SE. Continuous glucose monitoring to diagnose hypoglycemia due to late dumping syndrome in children after gastric surgeries. J Endocr Soc. 2021;5:bvaa197. doi: 10.1210/jendso/bvaa197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Tripyla A, Herzig D, Joachim D, Nakas CT, Amiet F, Andreou A, et al. Performance of a factory-calibrated, real-time continuous glucose monitoring system during elective abdominal surgery. Diabetes Obes Metab. 2020;22:1678–82. doi: 10.1111/dom.14073. [DOI] [PubMed] [Google Scholar]
- 76.Nair BG, Dellinger EP, Flum DR, Rooke GA, Hirsch IB. A pilot study of the feasibility and accuracy of inpatient continuous glucose monitoring. Diabetes Care. 2020;43:e168–9. doi: 10.2337/dc20-0670. [DOI] [PubMed] [Google Scholar]
- 77.Dungan KM, Han W, Miele A, Zeidan T, Weiland K. Determinants of the accuracy of continuous glucose monitoring in non-critically ill patients with heart failure or severe hyperglycemia. J Diabetes Sci Technol. 2012;6:884–91. doi: 10.1177/193229681200600420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Burt MG, Roberts GW, Aguilar-Loza NR, Stranks SN. Brief report: comparison of continuous glucose monitoring and finger-prick blood glucose levels in hospitalized patients administered basal-bolus insulin. Diabetes Technol Ther. 2013;15:241–5. doi: 10.1089/dia.2012.0282. [DOI] [PubMed] [Google Scholar]
- 79.Gomez AM, Umpierrez GE, Munoz OM, Herrera F, Rubio C, Aschner P, et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J Diabetes Sci Technol. 2015;10:325–9. doi: 10.1177/1932296815602905. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Schaupp L, Donsa K, Neubauer KM, Mader JK, Aberer F, Holl B, et al. Taking a closer look: continuous glucose monitoring in non-critically Ill hospitalized patients with type 2 diabetes mellitus under basal-bolus insulin therapy. Diabetes Technol Ther. 2015;17:611–8. doi: 10.1089/dia.2014.0343. [DOI] [PubMed] [Google Scholar]
- 81.Spanakis EK, Levitt DL, Siddiqui T, Singh LG, Pinault L, Sorkin J, et al. The effect of continuous glucose monitoring in preventing inpatient hypoglycemia in general wards: the glucose telemetry system. J Diabetes Sci Technol. 2018;12:20–5. doi: 10.1177/1932296817748964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Migdal AL, Spanakis EK, Galindo RJ, Davis G, Singh LG, Satyarengga M, et al. Accuracy and precision of continuous glucose monitoring in hospitalized patients undergoing radiology procedures. J Diabetes Sci Technol. 2020;14:1135–6. doi: 10.1177/1932296820930038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Shehav-Zaltzman G, Segal G, Konvalina N, Tirosh A. Remote glucose monitoring of hospitalized, quarantined patients with diabetes and COVID-19. Diabetes Care. 2020;43:e75–6. doi: 10.2337/dc20-0696. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Singh LG, Levitt DL, Satyarengga M, Pinault L, Zhan M, Sorkin JD, et al. Continuous glucose monitoring in general wards for prevention of hypoglycemia: results from the glucose telemetry system pilot study. J Diabetes Sci Technol. 2020;14:783–90. doi: 10.1177/1932296819889640. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Reutrakul S, Genco M, Salinas H, Sargis RM, Paul C, Eisenberg Y, et al. Feasibility of inpatient continuous glucose monitoring during the COVID-19 pandemic: early experience. Diabetes Care. 2020;43:e137–8. doi: 10.2337/dc20-1503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Galindo RJ, Migdal AL, Davis GM, Urrutia MA, Albury B, Zambrano C, et al. Comparison of the FreeStyle Libre pro flash continuous glucose monitoring (CGM) system and point-of-care capillary glucose testing in hospitalized patients with type 2 diabetes treated with Basal-Bolus insulin regimen. Diabetes Care. 2020;43:2730–5. doi: 10.2337/dc19-2073. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Singh LG, Satyarengga M, Marcano I, Scott WH, Pinault LF, Feng Z, et al. Reducing inpatient hypoglycemia in the general wards using real-time continuous glucose monitoring: the glucose telemetry system, a randomized clinical trial. Diabetes Care. 2020;43:2736–43. doi: 10.2337/dc20-0840. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Fortmann AL, Spierling Bagsic SR, Talavera L, Garcia IM, Sandoval H, Hottinger A, et al. Glucose as the fifth vital sign: a randomized controlled trial of continuous glucose monitoring in a non-ICU hospital setting. Diabetes Care. 2020;43:2873–7. doi: 10.2337/dc20-1016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Ushigome E, Yamazaki M, Hamaguchi M, Ito T, Matsubara S, Tsuchido Y, et al. Usefulness and safety of remote continuous glucose monitoring for a severe COVID-19 patient with diabetes. Diabetes Technol Ther. 2021;23:78–80. doi: 10.1089/dia.2020.0237. [DOI] [PubMed] [Google Scholar]
- 90.Carlson AL, Criego AB, Martens TW, Bergenstal RM. HbA1c: the glucose management indicator, time in range, and standardization of continuous glucose monitoring reports in clinical practice. Endocrinol Metab Clin North Am. 2020;49:95–107. doi: 10.1016/j.ecl.2019.10.010. [DOI] [PubMed] [Google Scholar]
- 91.Kumar RB, Goren ND, Stark DE, Wall DP, Longhurst CA. Automated integration of continuous glucose monitor data in the electronic health record using consumer technology. J Am Med Inform Assoc. 2016;23:532–7. doi: 10.1093/jamia/ocv206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Espinoza J, Shah P, Raymond J. Integrating continuous glucose monitor data directly into the electronic health record: proof of concept. Diabetes Technol Ther. 2020;22:570–6. doi: 10.1089/dia.2019.0377. [DOI] [PubMed] [Google Scholar]
- 93.Khatsenko K, Khin Y, Maibach H. Allergic contact dermatitis to components of wearable adhesive health devices. Dermatitis. 2020;31:283–6. doi: 10.1097/DER.0000000000000575. [DOI] [PubMed] [Google Scholar]
- 94.Didyuk O, Econom N, Guardia A, Livingston K, Klueh U. Continuous glucose monitoring devices: past, present, and future focus on the history and evolution of technological innovation. J Diabetes Sci Technol. 2020 Jan 13; doi: 10.1177/1932296819899394. [Epub]. [DOI] [PMC free article] [PubMed]
- 95.Kenda K, Kazic B, Novak E, Mladenic D. Streaming data fusion for the internet of things. Sensors (Basel) 2019;19:1955. doi: 10.3390/s19081955. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Faintuch J, Faintuch S. Glucose control in the intensive care unit. Chapter 42. Cham: Springer International Publishing; 2020. Obesity and diabetes: scientific advances and best practice; pp. 579–89. [Google Scholar]
- 97.Tang L, Chang SJ, Chen CJ, Liu JT. Non-invasive blood glucose monitoring technology: a review. Sensors (Basel) 2020;20:6925. doi: 10.3390/s20236925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Min J, Sempionatto JR, Teymourian H, Wang J, Gao W. Wearable electrochemical biosensors in North America. Biosens Bioelectron. 2021;172:112750. doi: 10.1016/j.bios.2020.112750. [DOI] [PubMed] [Google Scholar]
- 99.Ehrhardt N, Hirsch IB. The impact of COVID-19 on CGM Use in the hospital. Diabetes Care. 2020;43:2628–30. doi: 10.2337/dci20-0046. [DOI] [PubMed] [Google Scholar]