Abstract
Respiratory pressure responses to cervical magnetic stimulation are important measurements in monitoring the mechanical function of the respiratory muscles. Pressures can be measured using balloon catheters or a catheter containing integrated micro‐transducers. However, no research has provided a comprehensive analysis of their pressure measurement characteristics. Accordingly, the aim of this study was to provide a comparative analysis of these characteristics in two separate experiments: (1) in vitro with a reference pressure transducer following a controlled pressurization; and (2) in vivo following cervical magnetic stimulations. In vitro the micro‐transducer catheter recorded pressure amplitudes and areas which were in closer agreement to the reference pressure transducer than the balloon catheter. In vivo there was a main effect for stimulation power and catheter for esophageal (Pes), gastric (Pga), and transdiaphragmatic (Pdi) pressure amplitudes (p < 0.001) with the micro‐transducer catheter recording larger pressure amplitudes. There was a main effect of stimulation power (p < 0.001) and no main effect of catheter for esophageal (p = 0.481), gastric (p = 0.923), and transdiaphragmatic (p = 0.964) pressure areas. At 100% stimulator power agreement between catheters for Pdi amplitude (bias =6.9 cmH2O and LOA −0.61 to 14.27 cmH2O) and pressure areas (bias = −0.05 cmH2O·s and LOA −1.22 to 1.11 cmH2O·s) were assessed. At 100% stimulator power, and compared to the balloon catheters, the micro‐transducer catheter displayed a shorter 10–90% rise time, contraction time, latency, and half‐relaxation time, alongside greater maximal rates of change in pressure for esophageal, gastric, and transdiaphragmatic pressure amplitudes (p < 0.05). These results suggest that caution is warranted if comparing pressure amplitude results utilizing different catheter systems, or if micro‐transducers are used in clinical settings while applying balloon catheter‐derived normative values. However, pressure areas could be used as an alternative point of comparison between catheter systems.
Keywords: balloon catheter, Esophageal catheter, micro‐transducer catheter, respiratory pressures
The aim of this study was to provide a comparative analysis of micro‐transducer and balloon catheter pressure characteristics: (1) in vitro with a reference pressure transducer; and (2) in vivo following cervical magnetic stimulations. In vivo there was a main effect for catheter for esophageal, gastric, and transdiaphragmatic pressure amplitudes and no main effect for pressure areas, indicating that pressure areas could be used as a point of comparison between catheter systems. Caution is warranted if comparing pressure amplitude results from different catheter systems or if micro‐transducers are used in clinical settings with balloon catheter‐derived normative values.

1. INTRODUCTION
Respiratory pressure responses to nerve stimulation are important measurements in monitoring the mechanical function of the respiratory muscles (American Thoracic Society, 2003, Laveneziana et al., 2019, Macklem, 2004, Romer & Polkey, 2008). As measurements of pleural and abdominal pressures are invasive, they are typically estimated using surrogate measures of esophageal (Pes) and gastric (Pga) pressures, respectively (Benditt, 2005, Laveneziana et al., 2019). Traditionally, these measurements are collected with balloon catheters (Baydur et al., 1982; Milic‐Emili et al., 1964), but variations in catheter design, manual inflation of the balloon with either air or fluid, and catheter placement can lead to under or overestimation of pressure (Mead et al., 1955; Milic‐Emili et al., 1964; Mojoli et al., 2015; Petit & Milic‐Emili, 1958; Walterspacher et al., 2014).
There are a variety of commercially available balloon catheter designs and each requires a different quantity of air for optimum performance, and under and over inflation of balloons can produce invalid estimations of pressure (Milic‐Emili et al., 1964; Mojoli et al., 2015; Walterspacher et al., 2014). The perimeter and length of a balloon, along with its elastance, can also affect measurement accuracy (Mead et al., 1955; Mojoli et al., 2015; Petit & Milic‐Emili, 1958). Pressures are also affected by the location of the balloon within the body and are therefore dependent on placement technique (Mead & Whittenberger, 1953; Petit & Milic‐Emili, 1958). The proximal end of a balloon catheter is attached via plastic tubing to a pressure transducer located outside the body. Increasing the tubing length between the balloon and the transducer leads to reduced flow within the tubing (i.e., Poiseuille's Law), which may compromise dynamic response characteristics in balloon catheter systems (Cross et al., 2016; Mead et al., 1955; Mojoli et al., 2015; Walterspacher et al., 2014). Furthermore, balloon elasticity may change over time due to repeated sterilization and re‐use. These issues may explain the limited uptake of balloon catheters in clinical settings (Mauri et al., 2016; Mojoli et al., 2015) despite their many medical applications (Akoumianaki et al., 2014; Mauri et al., 2016).
The primary alternative to a balloon catheter is a catheter containing one or two integrated micro‐transducers (Beardsmore et al., 1982; Evans et al., 1993; Gilbert et al., 1979). Since micro‐transducer catheters do not utilize a balloon or require tubing to connect to an external transducer, they may overcome some of the limitations associated with traditional balloon catheters. However, despite these benefits, micro‐transducer measurements of Pes are more susceptible to mucus adhesion and contact with the esophageal wall, which reduces the surface area and therefore the spread of Van der Waals forces (Peters et al., 1998). Unpredictable shifts in baseline Pes have also been reported and are partly attributed to the micro‐transducers susceptibility to differences in pressures across the esophagus (Beardsmore et al., 1982), to regional artifacts (Panizza & Finucane, 1992) and baseline pressure drift in the device over time (1999). Recently, Augusto et al. reported no clinically relevant drift following 1 h of submersion with a Gaeltech micro‐transducer catheter. Micro‐transducer measurements of Pga may be also affected by immersion in gastric fluids (Stell et al., 1999).
Despite the potential benefits of the micro‐transducer catheter, only a limited number of studies have compared their pressure responses with those of a balloon catheter, and the results remain controversial. Poor agreement has been reported for absolute Pes and Pga (Augusto et al., 2017; Beardsmore et al., 1982; Peters et al., 1998; Stell et al., 1999), whereas both good (Peters et al., 1998; Stell et al., 1999) and poor (Augusto et al., 2017; Beardsmore et al., 1982) agreement has been reported for relative Pes and Pga (i.e., amplitude relative to baseline). Moreover, ambiguous evidence is provided by other studies that describe micro‐transducer and balloon catheters as “measuring pressures similarly” (Evans et al., 1993) and as “providing comparable measurements of absolute Pes and Pga” (Gilbert et al., 1979). As such, it is not clear how comparable the two devices are and which device measures pressure more accurately.
Analysis of magnetic or electrical cervical stimulation is important for the comprehensive assessment of the mechanical and neural properties of the respiratory muscles (Laghi et al., 1996; Man et al., 2004; Similowski et al., 1989, 1991, 1996, 1998; Taylor et al., 2006). Thus, understanding the accuracy and comparability of the two devices in measuring these responses is important for the correct interpretation of these measurements. While previous studies have evaluated the differences in pressures between balloon and micro‐transducer catheters (Augusto et al., 2017; Beardsmore et al., 1982; Panizza & Finucane, 1992; Stell et al., 1999), none have provided a comprehensive analysis of their pressure measurement characteristics following electric or magnetic stimulations. Accordingly, this study provides a thorough assessment of a range of characteristics for Pes, Pga, and transdiaphragmatic pressure (Pdi) in response to controlled pressurizations in vitro and to cervical magnetic stimulation in vivo.
2. METHODS
2.1. Experimental overview
This study comprised two separate experiments to evaluate the pressure measurement characteristics of a micro‐transducer catheter and balloon catheters. Experiment 1 evaluated, in vitro, the pressure amplitudes and areas of both catheter types following a controlled pressurization, with their responses compared to a reference pressure. Experiment 1 was also used to identify whether differences in catheter responses are present after removal of physiological factors such as mucus adhesion and immersion in gastric fluids. Experiment 2 evaluated, in vivo, the characteristics of both catheter types in human participants following cervical magnetic stimulation. The study was approved by the University of Southern Queensland's Ethics Committee and all procedures conformed to the standards set by the Declaration of Helsinki.
2.1.1. Experiment 1—in vitro
Protocols
The micro‐transducer catheter and a single balloon catheter were positioned in a sealed pressurized polyvinylchloride chamber (length =25 cm; radius =1 cm) alongside a reference pressure transducer (piezo‐resistive pressure transmitter MRB20; Bestech, Brisbane, Australia). The reference pressure was the standard against which pressures recorded by the micro‐transducer and balloon catheters were compared (measurement range =500 cmH2O; frequency response =1 kHz). The reference pressure transducer was calibrated at room temperature using a water manometer with a 1 m water column. The balloon catheter was inflated with 1 mL of air from a glass syringe, and both catheter types were then calibrated within the chamber at 100 cmH2O as measured by the reference pressure transducer. The catheters were then exposed to chamber pressures of 25, 50, 75, and 100 cmH2O (n = 100 for each) with a constant pressurization time of 0.2 s. For experiment 1, the same micro‐transducer catheter and a single balloon catheter were used, and all measurements were taken on the same day.
The micro‐transducer catheter and balloon catheter were secured on a mounting board with the micro‐transducers aligned to the centers of the balloons. This assembly and the reference pressure transducer were placed inside the airtight chamber which was pressurized using a gas supply (79% N2, 16% O2, and 5% CO2; BOC, North Ryde, Australia). The cylinder was fitted with a Type 10 valve (flow coefficient =0.4; BOC, North Ryde, Australia) leading to a regulator (6000 Argon Gas Regulator; BOC) with an upstream pressure of 2900 PSI. Maximum chamber pressures were adjusted via the regulator to obtain maximum pressure at the end of a 0.2 s pressurization time. Pressurization was automated using the Powerlab 16/35 to control a two‐way normally open isolation valve (NR3‐2–12; VFV, Mitcham, Australia). When the gas flow was switched off by the isolation valve, depressurization was complete within 150 – 250 ms.
2.1.2. Experiment 2—in vivo
Participants
Healthy young male (n = 4) and female (n = 4) participants (age =29 ± 3 years; height =173 ± 11 cm; body mass =84.7 ± 9.6 kg) with normal pulmonary function (forced vital capacity =98 ± 9% predicted; forced expiratory volume in 1 s = 95 ± 9% predicted) provided written informed consent to participate in this study. Exclusion criteria included current cigarette smokers, a history or current symptoms of cardiopulmonary disease, and a body mass index of <18.5 or >30 kg/m2.
Experimental design
Each participant visited the laboratory on two occasions, at a similar time of day, separated by a minimum of 24 h and a maximum of 7 days. Before each visit, participants abstained from food for 4 h, caffeine for 12 h, and exercise for 48 h. During visit 1, anthropometric measures and pulmonary function were assessed using a spirometer (Vmax® Encore PFT system; Vyaire Medical, Chicago, USA) according to published guidelines (Miller et al., 2005). Participants were instrumented with a micro‐transducer catheter to evaluate Pes, Pga, and Pdi responses to cervical magnetic stimulation. The micro‐transducer catheter was then removed, and participants were instrumented with esophageal and gastric balloon catheters and pressure responses to cervical magnetic stimulation were re‐evaluated. During visit 2, the order of catheter placement was reversed. The duration between removal of catheter(s) and instrumentation of the next catheter(s) was ~10 min.
Respiratory pressure catheters
The micro‐transducer catheter (Gaeltech) housed two pressure transducers (~5 × 2 mm), separated by 22.8 cm, which were constructed using half‐bridge thin‐film resistive strain gauge sensors coated with a silicone elastomer with frequency responses of 10–20 kHz. The catheter comprised a 100 cm silicon shaft (2.7 mm diameter) that also contained nine silver electrodes spaced 1 mm apart (electromyography data not reported here) and the pressure transducers were positioned proximally and distally to the electrodes. Prior to instrumentation in vivo the catheter was soaked for 1 h as per manufacturer's instructions to reduce baseline drift. The micro‐transducer catheter was then placed inside a small section of airtight plastic tubing and calibrated by injecting or withdrawing air, via a three‐way open valve connected to a glass syringe and a handheld respiratory pressure meter (Micro RPM; Vyaire Medical, Chicago, USA). Pes was calibrated to −100 cmH2O and Pga to +100 cmH2O. The external transducers of the balloon catheters were connected, via a 3‐way open valve, directly to the respiratory pressure meter and glass syringe. These transducers were calibrated between −27 cmH2O and +100 cmH2O by injecting and withdrawing air. The two balloon catheters consisted of a thin‐walled (~0.6 mm) polytetrafluoroethylene balloon (9.5 cm in length) sealed over an 86‐cm‐long polyethylene catheter (Adult esophageal balloon catheter; Cooper Surgical). These were connected to external pressure transducers with maximum frequency responses of 300 Hz and a pressure range of −27 to 407 cmH2O (SP844 Pressure Transducer; MEMSCAP). Pdi was calculated automatically using LabChart Pro software (AD Instruments, Bella Vista, Australia) by subtracting Pes from Pga.
Catheter placement
Catheter placement was preceded by intranasal administration of 1 mL of anesthetic lidocaine hydrochloride gel (Instillagel; MD Solutions Australasia). The positioning of the micro‐transducer catheter was achieved as previously described (Luo et al., 2001). The catheter was passed peri‐nasally into the stomach until a positive deflection in Pga and a negative deflection in Pes were observed during repeated sniffs. The catheter was then repositioned based on the strength of the crural diaphragm EMG simultaneously from different pairs of electrodes and was then secured in place. An occlusion test was then performed to confirm the catheter's location in the esophagus (Baydur et al., 1982). As esophageal diaphragm EMG is sensitive to differences in positioning (Luo et al., 2000), the micro‐transducer was positioned first to ensure the collection of quality EMG data. Subsequently, the deflated balloon catheters were inserted through the same nostril used for the micro‐transducer catheter. The centers of the respective balloons were positioned at the same distance from the nares as the micro‐transducers. The esophageal and gastric balloons were inflated with 1 and 2 mL of air, respectively. Pes and Pga deflections were then observed during repeated sniffs to check positioning, before being further assessed by an occlusion test. If required, the location of the balloon catheters was then altered to ensure accurate Pes and Pga measurements. The position of the catheters, relative to the nares, was identical during visits 1 and 2. This process allowed for the optimization of Pes, Pga, and EMG signals.
Cervical magnetic stimulation
After an initial 20‐min seated rest period to minimize post‐activation potentiation (Wragg et al., 1994), cervical magnetic stimulation was performed using a 90‐mm circular coil attached to a magnetic stimulator (2002; Magstim, Whitland, United Kingdom). Participants wore a nose‐clip and were seated in a chair with their neck flexed. Stimulations were performed with the glottis closed at functional residual capacity, which was inferred from visual feedback of Pes (i.e., an elevated plateau at the end of a tidal breath). The optimal stimulation site was determined by performing multiple stimulations at submaximal intensity (50% stimulator power) along C5‐C7 until the maximal Pdi, and thus the optimal stimulation site, was determined. This site was marked with indelible ink and used for all subsequent stimulations. Pes, Pga, and Pdi amplitudes were not different between visits, indicating that all stimulations were delivered with the same thoracoabdominal configuration. Pressure systems were compared at intensities of 50, 60, 70, 80, 85, 90, 95, and 100% of stimulator power output, with a minimum of three stimulations recorded at each intensity. Additional stimulations were performed when Pes or Pga values at end‐expiration were not at a stable baseline value. A 30 s pause was maintained between stimulations to prevent twitch‐on‐twitch potentiation (Guenette et al., 2010; Polkey et al., 1995; Taylor & Romer, 2009; Welch et al., ,2017, 2018).
2.2. Pressure capture and response analyses
Pressures were amplified with a Quad Bridge Amplifier (FE224; ADInstruments, Bella Vista, Australia) and all data were sampled continuously at 10 kHz using a Powerlab 16/35 and recorded using LabChart v8.1.2 software (ADInstruments). Low pass filters were set at 10 Hz for the balloon catheter pressure transducers and 1 kHz for the micro‐transducer catheter and the reference pressure transducer. In experiment 1, pressure amplitudes and areas were analyzed. In experiment 2, pressure amplitude, percentage of maximum amplitude, latency, contraction time, pressure area, 10–90% rise time, half‐relaxation time, time constant, maximal rate of pressure development (MRPD), maximal relaxation rate (MRR), and time to peak pressure using customized macroinstructions (LabChart v8.1.2 software; ADInstruments) (Figure 1).
FIGURE 1.

Pressure response analysis. (A) stimulation event; (B) pressure 5% above baseline; (A‐B) latency; (C‐E) 10–90% rise time; (D) point of the maximal rate of pressure development calculated as derivative at D divided by pressure amplitude at (F; G) point of the maximal relaxation rate calculated as derivative at G divided by pressure amplitude at F; (F) peak pressure; (A‐F) time to peak pressure; (B‐F) contraction time; (F‐I) half‐relaxation time; (H‐J) time constant calculated from 60–5% pressure amplitude
Pressure amplitude was calculated as the difference between baseline and peak pressure. Response onset was defined as the point at which pressure deviated 5% from baseline. Offset was defined as the point at which pressure returned to ±5% of baseline. Latency was defined as the time difference between magnetic stimulation and response onset (Experiment 2) or the time difference between valve opening and response onset (Experiment 1). Contraction time was defined as the duration between response onset and 100% of peak pressure. Pressure area was calculated using the integration between response onset and offset. The 10–90% rise time was defined as the elapsed time between 10% and 90% of peak pressure. Half‐relaxation time was defined as the elapsed time between 100% and 50% of peak pressure. The time constant was calculated between 60% and 10% of pressure amplitude. Time to peak pressure was defined as latency plus contraction time. MRPD and MRR were calculated based on equations [1] and [2] from previous work (Similowski et al., 1991).
| (1) |
| (2) |
Where dP/dt is the rate of change in pressure and A is the amplitude of the pressure response.
2.2.1. Statistical analyses
Statistical analyses were performed using SPSS for Windows (IBM). An initial power calculation was performed on the basis of the Pdi amplitudes for the balloon catheters and micro‐transducer catheter following cervical magnetic stimulation at 100% of stimulator power output. Power analysis indicated that a sample size of 8 would be required to detect differences in Pdi amplitudes between catheters (alpha =0.05 and power =0.8). Normality was assessed using a Shapiro–Wilk test. Supramaximality was determined by identifying a plateau in mean twitch Pdi at increasing stimulation power using a one‐way repeated measures ANOVA followed by pairwise comparisons (Guenette et al., 2010).
Between‐visit and between‐catheter pressure measurement characteristics at 100% of maximum stimulator output in response to cervical magnetic stimulation were analyzed using a paired sample t‐tests or Wilcoxon signed ranks test for parametric and nonparametric data, respectively. Between‐catheter differences for pressure amplitudes and areas at increasing stimulation intensities were analyzed using a two‐way repeated measures ANOVA to determine the effects of stimulation “intensity” (50, 60, 70, 80, 85, 90, 95, and 100% of maximum stimulation output) and “catheter” (micro‐transducer vs. balloon catheter). Significant intensity ×catheter interaction effects were followed by planned pairwise comparisons between catheters using the Bonferroni method.
The agreement, relationship, and reliability characteristics for pressure amplitudes and areas between the micro‐transducer catheter and balloon catheters were determined from data collected from all chamber pressures (Experiment 1—in vitro) or stimulation intensities (Experiment 2—in vivo). Bland–Altman analysis was used to evaluate the agreement between balloon and micro‐transducer catheter pressure measurements (Giavarina, 2015). Bias was defined as the micro‐transducer catheter measurement minus the balloon catheter measurement (experiment 1, in vivo), or as the reference transducer measurement minus the catheter measurement (experiment 2, in vitro). Limits of agreement (LOA) were calculated as the mean difference (bias) ±1.96 SD. Pearson's product moment correlation coefficient was used to examine the relationship between catheters. Within‐day reliability was assessed using coefficients of variation (CV) with the method error of the measurement (i.e., standard deviation divided by the mean). Between‐day reliability was assessed by using CV and the intraclass correlation coefficient (ICC(2,k)). Statistical significance was set at p < 0.05. Results are presented as means ±SD unless stated otherwise.
3. RESULTS
3.1. Experiment 1—in vitro
Ensemble averaged pressure responses to increasing chamber pressurizations for the micro‐transducer catheter, balloon catheter, and reference transducer are shown in Figure 2. Table 1 shows the measurement characteristics and agreement for pressure amplitudes and areas between the micro‐transducer catheter and balloon catheter and at increasing chamber pressures of 25, 50, 75, and 100 cmH2O with a constant pressurization time of 0.2 s. Pressure amplitudes were higher for the micro‐transducer catheter compared to the balloon catheter at all chamber pressures. Pressure areas for the micro‐transducer catheter were slightly higher than for the balloon catheter, with some exceeding that of the reference pressure at chamber pressures of 25 and 50 cmH2O, respectively (Table 1). Despite this, micro‐transducer catheter pressure amplitudes and areas were closer to reference values than the balloon catheters with the largest differences between the catheters occurring at the lowest chamber pressure (25 cmH2O; Table 1).
FIGURE 2.

Experiment 2—in vivo: Bland–Altman plots of esophageal, gastric, and transdiaphragmatic pressure amplitudes (top panels) and areas (bottom panels) between balloon catheters (BC) and micro‐transducer catheter (MC) following cervical magnetic stimulation at increasing stimulation intensities. Bias is represented by the solid line and the limits of agreement by the dotted lines (± 1.96 SD). Each participant has one datapoint per stimulation power and each datapoint was calculated as the mean value from visits 1 and 2
TABLE 1.
Experiment 1—in vitro: Measurement characteristics and agreement for pressure amplitudes and areas between the balloon catheter (BC) and micro‐transducer catheter (MC) at increasing chamber pressures of 25, 50, 75, and 100 cmH2O with a constant pressurization time of 0.2 s. Bias values were calculated as catheter pressure subtracted from reference pressure. Values are mean ±SD calculated from 100 responses to each chamber pressure
| 25 cmH2O | 50 cmH2O | 75 cmH2O | 100 cmH2O | |||||
|---|---|---|---|---|---|---|---|---|
| BC | MC | BC | MC | BC | MC | BC | MC | |
| Amplitude (cmH2O) | 22.8 ± 0.1 | 24.7 ± 0.1 | 44.9 ± 0.1 | 47.7 ± 0.1 | 66.2 ± 0.1 | 69.3 ± 0.1 | 84.8 ± 0.1 | 89.7 ± 0.1 |
| Amplitude (%RP) | 91 ± 0 | 99 ± 0 | 90 ± 0 | 95 ± 0 | 89 ± 0 | 93 ± 0 | 86 ± 0 | 90 ± 0 |
| Amplitude Bias (cmH2O) | 2.2 | 0.4 | 5.0 | 2.3 | 8.6 | 5.5 | 14.4 | 9.6 |
| Amplitude LOA (cmH2O) | 2.2 to 2.3 | 0.3 to 0.5 | 5.0 to 5.1 | 2.2 to 2.4 | 8.5 to 8.7 | 5.4 to 5.6 | 13.8 to 15.0 | 8.9 to 10.2 |
| Area (cmH2O·s) | 4.17 ± 0.02 | 4.39 ± 0.02 | 8.59 ± 0.02 | 8.83 ± 0.02 | 13.2 ± 0.03 | 13.4 ± 0.03 | 17.8 ± 0.03 | 18.0 ± 0.04 |
| Area (%RP) | 97 ± 0 | 102 ± 0 | 98 ± 0 | 101 ± 0 | 98 ± 0 | 99 ± 0 | 97 ± 0 | 98 ± 0 |
| Area Bias (cmH2O·s) | 0.12 | −0.10 | 0.17 | −0.08 | 0.24 | 0.08 | −0.51 | −0.36 |
| Area LOA (cmH2O·s) | 0.11 to 0.13 | −0.12 to −0.08 | 0.16 to 0.18 | −0.11 to −0.05 | 0.22 to 0.26 | 0.04 to 0.11 | −0.52 to −0.49 | −0.39 to −0.33 |
Abbreviations: RP, reference pressure; LOA, limits of agreement (bias ±1.96 SD).
For pressure amplitudes and areas, agreement with the reference pressure transducer was closer (reflected by a lower bias) for the micro‐transducer catheter than the balloon catheter (Table 1). Significant correlations between the catheters for pressure amplitude were present at chamber pressures of 25 (r = 0.84), 50 (r = 0.78), 75 (r = 0.91), and 100 (r = 0.91) cmH2O (p < 0.001). Similarly, correlations between the catheters for pressure area were also present at chamber pressures of 25 (r = 0.77), 50 (r = 0.79), 75 (r = 0.84), and 100 (r = 0.90) cmH2O (p < 0.001). Within‐day reliability was high for both micro‐transducer and balloon catheters for pressure amplitudes (micro‐transducer vs. balloon catheters): 0.25 (CI 0.22 to 0.27) vs. 0.22 (CI 0.20 to 0.24%) and areas 0.29 (CI 0.27 to 0.31) vs. 0.25 (CI 0.24 to 0.27) %.
3.2. Experiment 2—in vivo
Representative pressure responses to cervical magnetic stimulation at 100% of stimulator power output for the balloon catheters and micro‐transducer catheter are shown in Figure 3. There were no between‐visit differences for all pressure measurement characteristics for the micro‐transducer (p = 0.055) and balloon catheters (p = 0.314). Therefore, data from visits 1 and 2 were pooled. Supramaximality was achieved from 80% (p > 0.055) and 90% (p > 0.105) stimulator power output for the balloon and micro‐transducer catheters.
FIGURE 3.

Experiment 1—in vitro: Ensemble average waveforms (each from 100 waves) from the micro‐transducer catheter (MC), balloon catheter (BC), and reference (RP) pressures in response to chamber pressures of 25, 50, 75, and 100 cmH2O with a constant pressurization time of 0.2 s
Table 2 shows the Pes, Pga, and Pdi pressure measurement characteristics for the balloon catheters and micro‐transducer catheter following cervical magnetic stimulation at 100% of stimulator power output. Compared to the balloon catheters, the micro‐transducer catheter displayed shorter 10–90% rise times, contraction times, latencies and half‐relaxation times, and greater maximal rates of changes in pressure (MRPD and MRR) and pressure amplitudes (p < 0.05). When pressure amplitudes were normalized to the percentage of maximum, there was no difference between catheters, nor were there any differences between catheters for pressure area. Pga and, subsequently, Pdi were higher (p < 0.05) at end‐expiration for the micro‐transducer catheter than the balloon catheters.
TABLE 2.
Experiment 2—in vivo: Esophageal pressure (Pes), gastric pressure (Pga), and transdiaphragmatic pressure (Pdi) measurement characteristics for balloon catheters (BC) and micro‐transducer catheter (MC) following cervical magnetic stimulation at 100% of stimulator power output. Data are mean ±SD and pooled from visits 1 and 2
| Pes | Pga | Pdi | ||||
|---|---|---|---|---|---|---|
| BC | MC | BC | MC | BC | MC | |
| Amplitude (cmH2O) | 15.8 ± 4.1* | 20.5 ± 6.4 | 9.0 ± 3.1* | 13.1 ± 4.2 | 24.2 ± 5.0* | 32.1 ± 8.3 |
| Amplitude (%max) | 89 ± 9 | 87 ± 12 | 78 ± 16 | 74 ± 19 | 94 ± 4 | 92 ± 5 |
| Area (cmH2O·s) | 2.4 ± 0.7 | 2.3 ± 0.7 | 2.9 ± 1.3 | 2.5 ± 1.6 | 4.5 ± 0.9 | 4.3 ± 1.2 |
| 10–90% Rise time (ms) | 66 ± 9* | 43 ± 8 | 78 ± 21* | 38 ± 18 | 69 ± 8* | 47 ± 8 |
| Time to peak pressure (ms) | 97 ± 13* | 66 ± 12 | 121 ± 36* | 58 ± 28 | 146 ± 13* | 95 ± 12 |
| Latency (ms) | 49 ± 5* | 33 ± 6 | 39 ± 3* | 27 ± 7 | 42 ± 3* | 27 ± 3 |
| Half‐relaxation (ms) | 89 ± 12* | 60 ± 12 | 132 ± 67* | 82 ± 58 | 108 ± 14* | 70 ± 7 |
| Time constant (ms) | 70 ± 30 | 54 ± 24 | 197 ± 182 | 125 ± 135 | 106 ± 13 | 98 ± 39 |
| MRPD (%gain/10 ms) | 12.8 ± 2.1* | 18.4 ± 1.7 | 13.6 ± 2.3* | 18.7 ± 2.9 | 12.6 ± 1.4* | 17.3 ± 1.8 |
| MRR (%loss/10 ms) | 8.1 ± 2.3* | 10.4 ± 2.5 | 5.9 ± 3.2* | 8.2 ± 3.2 | 5.6 ± 0.7* | 8.9 ± 2.0 |
| Pressure at end‐expiration (cmH2O) | −1.4 ± 2.1 | 0.8 ± 2.5 | 13.5 ± 5.1* | 10.6 ± 2.2 | 16.0 ± 3.5* | 9.7 ± 3.0 |
Significantly different from micro‐transducer catheter (*p < 0.05).
Abbreviations: MRPD, maximum rate of pressure development; MRR, maximum rate of relaxation.
Pes, Pga, and Pdi amplitudes and areas from the micro‐transducer and balloon catheters in response to increasing stimulation intensities are shown in Figure 4. Both catheters responded linearly to increasing stimulation intensities. For Pes, Pga, and Pdi amplitude, there were main effects of stimulation intensity (p < 0.001) and catheter (p < 0.001). That is, pressure amplitudes increase with stimulation intensity and are higher for the micro‐transducer catheter. No intensity ×catheter interaction effects (p > 0.935) were observed. For Pes, Pga, and Pdi pressure areas, there was a main effect of stimulation intensity (p < 0.001) with pressure area increasing with stimulation intensity. There were no main effects of catheter (p = 0.481) or stimulation intensity ×catheter interaction effects (p > 0.995).
FIGURE 4.

Experiment 2—in vivo: Representative esophageal, gastric, and transdiaphragmatic pressure characteristics for the balloon catheters and micro‐transducer catheter following cervical magnetic stimulation at 100% of stimulator power output. Three repeated twitches from one participant are shown superimposed. Stimulation artifacts are marked with an arrow (↑)
Bland–Altman plots for the agreement between the micro‐transducer and balloon catheters for Pes, Pga, and Pdi amplitudes and areas in response to cervical magnetic stimulation are shown in Figure 5. Pes, Pga, and Pdi amplitudes had biases of 3.8 (LOA −0.55 to 8.26), 4.2 (LOA −6.64 to 15.09), and 6.9 (LOA −0.61 to 14.27) cmH2O, respectively. Significant correlations between the catheters for Pes (r = 0.96), Pga (r = 0.77), and Pdi (r = 0.94) amplitudes were moderate to strong (p < 0.001). Pes, Pga, and Pdi pressure areas had biases of −0.08 (LOA −0.70 to 0.54), −0.03 (LOA −3.75 to 3.68), and −0.05 (LOA −1.22 to 1.11) cmH2O∙s, respectively. Significant correlations between the catheters for Pes (r = 0.94), Pga (r = 0.84), and Pdi (r = 0.91) were moderate to strong (p < 0.001).
FIGURE 5.

Experiment 2—in vivo: Esophageal, gastric, and transdiaphragmatic pressure amplitudes (top panels) and areas (bottom panels) for balloon catheters and micro‐transducer catheter following cervical magnetic stimulation at increasing stimulation intensities. Data are mean ±SD and pooled from visits 1 and 2. Significant difference between catheters (*p < 0.05; **p < 0.01)
Within‐ and between‐day reliability coefficients for Pes, Pga, and Pdi amplitudes and areas in response to cervical magnetic stimulation at 100% of stimulator output for the micro‐transducer and balloon catheters are shown in Table 3. Within‐ and between‐day reliability for Pes and Pdi amplitudes and areas were similar between the catheters. For the micro‐transducer compared to the balloon catheters, Pga amplitudes, and areas had lower within‐day reliability and higher between‐day reliability.
TABLE 3.
Experiment 2—in vivo: Within‐ and between day reliability of esophageal pressure (Pes), gastric pressure (Pga), and transdiaphragmatic pressure (Pdi) amplitudes and areas for balloon catheters (BC) and micro‐transducer catheter (MC) following cervical magnetic stimulation at 100% of stimulator power output. Data are presented as means with 95% confidence intervals in parentheses
| Pes | Pga | Pdi | ||||
|---|---|---|---|---|---|---|
| BC | MC | BC | MC | BC | MC | |
| Within‐day (CV) | ||||||
| Amplitude (%) | 8.7 (5.2 to 12.3) | 10.7 (6.4 to 14.9) | 6.7 (4.1 to 9.2) | 12.9 (8.3 to 17.5) | 6.2 (3.0 to 9.4) | 6.1 (4.0 to 8.3) |
| Area (%) | 14.5 (9.3 to 19.6) | 12.8 (8.2 to 17.4) | 14.9 (7.1 to 22.8) | 23.4 (12.1 to 34.6) | 9.6 (5.3 to 14.0) | 8.6 (4.6 to 12.6) |
| Between‐day (CV) | ||||||
| Amplitude (%) | 10.7 (8.1 to 13.3) | 10.9 (7.8 to 14.0) | 20.7 (17.5 to 23.9) | 17.8 (11.1 to 24.4) | 9.8 (6.0 to 13.6) | 11.3 (5.3 to 17.2) |
| Area (%) | 15.0 (12.1 to 18.0) | 16.0 (12.4 to 19.7) | 30.6 (17.9 to 43.3) | 26.4 (21.1 to 31.8) | 13.0 (9.0 to 17.0) | 18.5 (7.8 to 29.2) |
| Between‐day (ICC) | ||||||
| Amplitude | 0. 93 (0.69 to 0.99) | 0.934 (0.70 to 0.99) | 0.72 (−0.58 to 0.95) | 0.60 (−1.54 to 0.92) | 0.81 (−0.05 to 0.96) | 0.82 (0.08 to 0.96) |
| Area | 0. 94 (0.71 to 0.99) | 0.903 (0.56 to 0.98) | 0.68 (−0.92 to 0.93) | 0.60 (−0.87 to 0.92) | 0.79 (−0.12 to 0.96) | 0.58 (−1.37 to 0.92) |
Abbreviations: CV, coefficient of variation; ICC, intraclass correlation coefficient.
4. DISCUSSION
4.1. Main findings
This study is the first to provide a comprehensive analysis of a range of balloon and micro‐transducer catheter pressure measurement characteristics in vitro with a reference pressure following controlled pressurizations (Experiment 1) and in vivo following cervical magnetic stimulation (Experiment 2). The main findings were: (1) in vitro the micro‐transducer catheter showed closer agreement to the reference pressure amplitudes and areas than the balloon catheter; (2) in vivo the micro‐transducer catheter recorded higher pressure amplitudes and similar pressure areas than the balloon catheters; and (3) in vivo the micro‐transducer catheter displayed shorter pressure response times and half‐relaxation times, and greater maximal rates of changes in pressure than the balloon catheters.
4.2. Pressure amplitudes
In vivo the micro‐transducer catheter had higher pressure amplitudes compared to the balloon catheters. While no Pes agreement data following cervical magnetic stimulation have previously been reported, the values here are similar to those reported during quiet breathing (bias = −3.6 cmH2O, LOA −14.3 to 7 cmH2O) and demonstrate better agreement than those reported during sniff maneuvers (bias = −50.6 cmH2O, LOA −60.6 to −40.6 cmH2O) (Augusto et al., 2017). The presence of differences in pressure measurement is also consistent with previous work (Augusto et al., 2017; Beardsmore et al., 1982; Peters et al., 1998; Stell et al., 1999). The in vivo Pdi results presented here with a bias of 6.9 (LOA −0.61 to 14.27) cmH2O are higher than those previously reported by Stell et al. with a bias of 2.1 (LOA −10.5 to 6.3) cmH2O. This difference is likely due to methodological and technical differences between the studies. For instance, Stell et al. placed micro‐transducer and balloon catheters simultaneously into their participants, thus exposing them to identical physiological conditions (i.e., excluding some of the within‐day variability potentially experienced during sequential catheter placements). The balloon catheters utilized by Stell et al. were from a different manufacturer, with a longer catheter (+24 cm) and balloons (+0.5 cm) and a different filling volumes for Pes (0.5 mL). These differences, respectively, may affect the dynamic compliance, whereas differences in balloon filling volumes affect pressure measurements (Cross et al., 2016; Milic‐Emili et al., 1964; Mojoli et al., 2015; Walterspacher et al., 2014). There are no published values of Pga available against which to compare our results.
The in vitro results also demonstrated that the micro‐transducer catheter recorded higher pressure amplitudes than the balloon catheter and the pressures obtained were closer to the reference pressure. The differences in pressure amplitude between the catheters are likely due to the faster dynamic responses of the micro‐transducer catheter, allowing it to reach higher pressures more quickly than the balloon catheter, and thus more closely tracking rapid pressurization. In vivo, the within‐ and between‐day reliability coefficients for Pes and Pdi amplitudes were similar between the catheters and to those reported previously for balloon catheters (Taylor & Romer, 2009; Tiller et al., 2017; Wüthrich et al., 2015). However, the within‐ and between‐day reliability coefficients for Pga from the micro‐transducer catheter were higher than those of the balloon catheter and slightly higher than those reported previously for balloon catheters (Tiller et al., 2017). The differences may be explained by the greater sensitivity of the micro‐transducer catheter to pressure changes that occur readily within the stomach. The within‐day repeatability of pressure amplitudes and areas in vitro was high for both catheters, which suggests that when physiological factors are excluded, there are no inherent differences in the reliability of balloon and micro‐transducer catheters.
4.3. Pressure areas
The most common measurement of respiratory muscle strength is pressure amplitude (i.e., twitch pressures), however, the pressure area is also indicative of muscular work output (Bazzucchi et al., 2011; Carámbula et al., 2019; Celichowski et al., 2000). Areas have been reported for twitch tension (Lepers et al., 2000; Lewis et al., 2017) and twitch peak torque (Lepers et al., 2002) following electrical quadricep stimulations, but to the best of our knowledge have not been reported for the diaphragm following cervical magnetic stimulation. The pressure area envelope is “triangular” and pressure amplitude determines the perpendicular height of the triangle from base to apex, whereas the pressure response and relaxation rates control the slopes up and down from the apex. Thus, changes in pressure area are reflective predominantly of pressure amplitude, while also being influenced by differences in response and relaxation rates.
The micro‐transducer catheter demonstrated higher pressure amplitudes and sharper waveforms. Conversely, the balloon catheter displayed lower pressure amplitudes and blunter waveforms. Thus, despite the shape of the waveform recorded by the catheters being visibly different, the pressure areas are similar. This is evidenced in vitro by agreement values closer to zero and relative pressure area values that were closer to 100% for the micro‐transducer catheter. In vivo this is shown by the lack of main effect of catheter on pressure area results. However, the CV values for the within‐ and between‐day reliability indicate that pressure area measurements are less reliable than pressure amplitudes. Assessment of between‐day reliability using ICC indicates a higher degree of variability in Pga and Pdi amplitudes and areas as these values had wide CI, with some incorporating negative lower limits. While this indicates that the measures are unreliable, there is no significant evidence of differences in reliability between devices, or between pressure amplitudes and areas. Hence, these data indicate that pressure area could provide a measurement suitable for direct comparisons between micro‐transducer and balloon catheters.
4.4. Pressure responses, half‐relaxation times, and rates of pressure change
This is the first study to provide a comparative analysis of the pressure measurement characteristics of a micro‐transducer and balloon catheters following cervical magnetic stimulations. In vivo, the Pes, Pga, and Pdi responses of the micro‐transducer catheter had shorter latencies, 10–90% rise times, time to peak pressure and a greater MRPD than the balloon catheter in response to cervical magnetic stimulation. Furthermore, as pressures returned to baseline, the micro‐transducer catheter had shorter half‐relaxation times and greater maximal relaxation rates. No differences were observed in the time constant for Pes, Pga, or Pdi. The larger variability of time constant values observed in Pga (and thus Pdi) are due to the secondary peaks occurring in some gastric response curves. These alter the decay waveform from the standard exponential form, causing variability in the calculation of the time constant. Hence, caution is advised when collecting and analyzing time constant data. These response characteristic data show that the micro‐transducer catheter demonstrated “faster” responses to changes in pressures than balloon catheters. This does not imply that it performs better than the balloon catheter in measuring pressures in vivo. However, their faster responses do produce different waveforms in response to cervical magnetic stimulation, with the micro‐transducer catheter providing sharper and shorter response curves than the balloon catheters. The differences in catheter responses can be attributed to their unique designs, with the micro‐transducer having a greater inherent capacity for fast responses.
4.5. Methodological considerations
Experiment 1. Ideally any reference waveform used in in vitro respiratory testing should include waveforms with spectral content greater than 20 Hz. However, those presented in Experiment 1 were approximately 5 Hz and thus a deeper comparison of these data to assess the dynamic response characteristics of the catheters was not possible.
4.6. Clinical implications
Low Pdi amplitudes (i.e., twitch pressures) in response to un‐potentiated cervical magnetic stimulation have been utilized for the identification of diaphragm weakness. Pressures below 20 cmH2O for bilateral phrenic nerve stimulation (such as that performed in this study) are potentially indicative of bilateral diaphragm weakness (ATS/ERS Taskforce, 2002). Pressures below 18 cmH2O correlate with observations of muscle weakness in some diseases (Steier et al., 2007), whereas those below 10 cmH2O in critically ill patients indicate acquired diaphragm weakness (Supinski & Callahan, 2013). Recently, Dubé and Dres (2016) produced algorithms for the suspicion and treatment of diaphragm dysfunction and proposed a twitch Pdi <20 cmH2O (or <10 cmH2O for unilateral phrenic nerve stimulation) is indicative of bilateral diaphragm weakness. However, as these cut‐off values are based on respiratory pressures measured using balloon catheters, which based on our findings record lower Pdi. For example, the mean Pdi twitch pressure for patients with severe stable COPD, measured using balloon catheters by Polkey et al., is 18.5 cmH2O (1996). If a micro‐transducer catheter was used, and the twitch Pdi bias from our Experiment 2 (~6.9 cmH2O higher) factored in, the recorded value would have been closer to ~25.4 cmH2O indicating that diaphragm weakness is instead unlikely. Thus, applying the aforementioned cut‐off values measured using balloon catheters to those measured using a micro‐transducer catheter may lead to incorrect clinical assessments and diagnoses. This should therefore be considered if micro‐transducer catheters are used in the evaluation of diaphragm weakness, and it may be necessary to establish new normative and cut‐off values.
Alternatively, our results have demonstrated that a surrogate measurement for direct comparisons between micro‐transducer and balloon catheters may be pressure area, which corrects for differences in the pressure response shape between the catheters. If normative values and cut‐off values for pressure areas were ascertained, then these measurements would allow for comparisons between the catheters to be made. Given the presence of a main effect of catheter on Pdi, and the significant differences observed between catheters at 100% stimulation power, we would also expect significant differences between catheters when measuring potentiated twitch Pdi (e.g., twitches delivered after a maximal volitional inspiratory maneuver). Thus, between catheter comparisons of diaphragm contractility test results should be interpreted with care. Response and relaxation rates (e.g., muscle shortening and relaxation rates) following cervical magnetic stimulation also provide valuable information pertaining to the mechanical properties of the diaphragm (ATS/ERS Taskforce, 2002, Laveneziana et al., 2019, Wilcox et al., 1988). The present study shows, however, that response and relaxation rates differ between the micro‐transducer and balloon catheters. Therefore, caution is warranted when comparing studies that have used different catheter systems to obtain these measurements.
5. CONCLUSION
This is the first study to provide a comparative analysis of the pressure measurement characteristics of micro‐transducer and balloon catheters in response to controlled pressurizations in vitro (Experiment 1) and cervical magnetic stimulations in vivo (Experiment 2). Under in vivo and in vitro conditions, the micro‐transducer catheter recorded higher pressure amplitudes, and under in vivo conditions, shorter response and relaxation rates and greater rates of changes in pressure compared to the balloon catheters. Accordingly, caution is warranted when comparing the results of studies that used different catheter systems to obtain these measurements. Furthermore, in a clinical setting caution is warranted if pressure amplitude measurements made with micro‐transducer catheters are compared to normative values derived from balloon catheters. However, this limitation may be mitigated if comparisons are made based on pressure area, which does not differ between micro‐transducer and balloon catheters.
COMPETING INTERESTS
The authors declare no conflict of interest.
AUTHOR CONTRIBUTIONS
W.M and D.E.M conceptualized and designed the experiments. W.M collected and analyzed the data. D.E.M, B.H contributed to data interpretation and statistical analysis. M.A.J and G.R.S contributed to revisions of intellectual content. All authors approved the final manuscript.
ACKNOWLEDGMENT
The authors acknowledge the participants for their time and dedication to this study.
Funding information
University of Southern Queensland.
REFERENCES
- Akoumianaki, E. , Maggiore, S. M. , Valenza, F. , Bellani, G. , Jubran, A. , Loring, S. H. , Pelosi, P. , Talmor, D. , Grasso, S. , & Chiumello, D. (2014). The application of esophageal pressure measurement in patients with respiratory failure. American Journal of Respiratory and Critical Care Medicine, 189, 520–531. [DOI] [PubMed] [Google Scholar]
- American thoracic society (2003). ATS/ACCP statement on cardiopulmonary exercise testing. American Journal of Respiratory and Critical Care Medicine, 167, 211. [DOI] [PubMed] [Google Scholar]
- ATS, ERS TASKFORCE , (2002). ATS/ERS Statement on respiratory muscle testing. American journal of respiratory and critical care medicine, 166, 518. [DOI] [PubMed] [Google Scholar]
- Augusto, R. M. , Albuquerque, A. L. P. , Jaeger, T. , de Carvalho, C. R. R. , & Caruso, P. (2017). Stability and agreement of a microtransducer and an air‐filled balloon esophageal catheter in the monitoring of esophageal pressure. Respiratory Care, 62, 215–221. [DOI] [PubMed] [Google Scholar]
- Baydur, A. , Behrakis, P. K. , Zin, W. A. , Jaeger, M. , & Milic‐Emili, J. (1982). A Simple method for assessing the validity of the esophageal balloon technique 1–2. American Review of Respiratory Disease, 126, 788–791. [DOI] [PubMed] [Google Scholar]
- Bazzucchi, I. , Felici, F. , Montini, M. , Figura, F. , & Sacchetti, M. (2011). Caffeine improves neuromuscular function during maximal dynamic exercise. Muscle & Nerve, 43, 839–844. [DOI] [PubMed] [Google Scholar]
- Beardsmore, C. , Wong, Y.‐C. , Stocks, J. , & Silverman, M. (1982). Assessment of the catheter tip pressure transducer for use in infant lung function tests. Medical & biological engineering & computing, 20, 657–660. [DOI] [PubMed] [Google Scholar]
- BENDITT, J. O. (2005). Esophageal and Gastric Pressure Measurements. Respiratory Care, 50, 68–77. [PubMed] [Google Scholar]
- Carámbula, A. , Visca, A. , D’Amico, S. , & Angulo, M. (2019). Respiratory and peripheral muscle assessment in the intensive care unit. Archivos de bronconeumología, 55, 258–265. [DOI] [PubMed] [Google Scholar]
- Celichowski, J. , Grottel, K. , & Bichler, E. (2000). Relationship between the tension‐time area and the frequency of stimulation in motor units of the rat medial gastrocnemius muscle. Journal of Physiology and Pharmacology, 51, 291–302. [PubMed] [Google Scholar]
- Cross, T. J. , Beck, K. C. , & Johnson, B. D. (2016). Correcting the dynamic response of a commercial esophageal balloon‐catheter. Journal of Applied Physiology, 121, 503–511. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dubé, B.‐P. , & Dres, M. (2016). Diaphragm dysfunction: diagnostic approaches and management strategies. Journal of clinical medicine, 5, 113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Evans, S. A. , Watson, L. , Cowley, A. J. , Johnston, I. D. , & Kinnear, W. J. (1993). Normal range for transdiaphragmatic pressures during sniffs with catheter mounted transducers. Thorax, 48, 750–753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Giavarina, D. (2015). Understanding bland altman analysis. Biochemia Medica, 25, 141–151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gilbert, R. , Peppi, D. , & Auchincloss, J. (1979). Measurement of transdiaphragmatic pressure with a single gastric‐esophageal probe. Journal of Applied Physiology, 47, 628–630. [DOI] [PubMed] [Google Scholar]
- Guenette, J. A. , Romer, L. M. , Querido, J. S. , Chua, R. , Eves, N. D. , McKenzie, D. C. , & Sheel, A. W. (2010). Sex differences in exercise‐induced diaphragmatic fatigue in endurance‐trained athletes. Journal of Applied Physiology, 109, 35–46. [DOI] [PubMed] [Google Scholar]
- Laghi, F. , Harrison, M. J. , & Tobin, M. J. (1996). Comparison of magnetic and electrical phrenic nerve stimulation in assessment of diaphragmatic contractility. Journal of Applied Physiology, 80, 1731–1742. [DOI] [PubMed] [Google Scholar]
- Laveneziana, P. , Albuquerque, A. , Aliverti, A. , Babb, T. , Barreiro, E. , Dres, M. , Dube, B. P. , Fauroux, B. , Gea, J. , Guenette, J. A. , Hudson, A. L. , Kabitz, H. J. , Laghi, F. , Langer, D. , Luo, Y. M. , Neder, J. A. , O'donnell, D. , Polkey, M. I. , Rabinovich, R. A. , & Verges, S. , 2019. ERS statement on respiratory muscle testing at rest and during exercise. The European respiratory journal, 53(6). [DOI] [PubMed] [Google Scholar]
- Lepers, R. , Hausswirth, C. , Maffiuletti, N. , Brisswalter, J. , & van Hoecke, J. (2000). Evidence of neuromuscular fatigue after prolonged cycling exercise. Medicine and science in sports and exercise, 32, 1880–1886. [DOI] [PubMed] [Google Scholar]
- Lepers, R. , Maffiuletti, N. A. , Rochette, L. , Brugniaux, J. , & Millet, G. Y. (2002). Neuromuscular fatigue during a long‐duration cycling exercise. Journal of applied physiology, 92, 1487–1493. [DOI] [PubMed] [Google Scholar]
- Lewis, E. J. , Stucky, F. , Radonic, P. W. , Metherel, A. H. , Wolever, T. M. , & Wells, G. D. (2017). Neuromuscular adaptations to sprint interval training and the effect of mammalian omega‐3 fatty acid supplementation. European Journal of Applied Physiology, 117, 469–482. [DOI] [PubMed] [Google Scholar]
- Luo, Y. , Hart, N. , Mustfa, N. , Lyall, R. , Polkey, M. , & Moxham, J. (2001). Effect of diaphragm fatigue on neural respiratory drive. Journal of Applied Physiology, 90, 1691–1699. [DOI] [PubMed] [Google Scholar]
- Luo, Y. , Lyall, R. , Harris, M. , Hawkins, P. , Hart, N. , Polkey, M. , & Moxham, J. (2000). Effect of lung volume on the oesophageal diaphragm EMG assessed by magnetic phrenic nerve stimulation. European Respiratory Journal, 15, 1033–1038. [DOI] [PubMed] [Google Scholar]
- Macklem, P. T. (2004). A Century of the mechanics of breathing. American journal of respiratory and critical care medicine, 170, 10–15. [DOI] [PubMed] [Google Scholar]
- Man, W.‐ D.‐C. , Moxham, J. , & Polkey, M. I. (2004). Magnetic stimulation for the measurement of respiratory and skeletal muscle function. European Respiratory Journal, 24, 846–860. [DOI] [PubMed] [Google Scholar]
- Mauri, T. , Yoshida, T. , Bellani, G. , Goligher, E. C. , Carteaux, G. , Rittayamai, N. , Mojoli, F. , Chiumello, D. , Piquilloud, L. , & Grasso, S. (2016). Esophageal and transpulmonary pressure in the clinical setting: meaning, usefulness and perspectives. Intensive Care Medicine, 42, 1360–1373. [DOI] [PubMed] [Google Scholar]
- Mead, J. , McIlroy, M. , Selverstone, N. , & Kriete, B. (1955). Measurement of intraesophageal pressure. Journal of Applied Physiology, 7, 491–495. [DOI] [PubMed] [Google Scholar]
- Mead, J. , & Whittenberger, J. L. (1953). Physical properties of human lungs measured during spontaneous respiration. Journal of Applied Physiology, 5, 779–796. [Google Scholar]
- Milic‐Emili, J. , Mead, J. , Turner, J. , & Glauser, E. (1964). Improved technique for estimating pleural pressure from esophageal balloons. Journal of Applied Physiology, 19, 207–211. [DOI] [PubMed] [Google Scholar]
- Miller, M. R. , Hankinson, J. , Brusasco, V. , Burgos, F. , Casaburi, R. , Coates, A. , Crapo, R. , Enright, P. , van der Grinten, C. , & Gustafsson, P. (2005). Standardisation of spirometry. European Respiratory Journal, 26, 319–338. [DOI] [PubMed] [Google Scholar]
- Mojoli, F. , Chiumello, D. , Pozzi, M. , Algieri, I. , Bianzina, S. , Luoni, S. , Volta, C. , Braschi, A. , & Brochard, L. (2015). Esophageal pressure measurements under different conditions of intrathoracic pressure. an in vitro study of second generation balloon catheters. Minerva Anestesiologica, 81, 855–864. [PubMed] [Google Scholar]
- Panizza, J. A. , & Finucane, K. (1992). Comparison of balloon and transducer catheters for estimating lung elasticity. Journal of Applied Physiology, 72, 231–235. [DOI] [PubMed] [Google Scholar]
- Peters, R. , Meijer, J. , Kingma, R. , Scheffer, G. , & Heethaar, R. (1998). Evaluation of catheter‐mounted transducers for intra‐oesophageal pressure recording in respiratory function tests. Medical & biological engineering & computing, 36, 562–567. [DOI] [PubMed] [Google Scholar]
- Petit, J. , & Milic‐Emili, G. (1958). Measurement of endoesophageal pressure. Journal of Applied Physiology, 13, 481–485. [DOI] [PubMed] [Google Scholar]
- Polkey, M. I. , Kyroussis, D. , Hamnegard, C.‐H. , Mills, G. H. , Green, M. , & Moxham, J. (1996). Diaphragm strength in chronic obstructive pulmonary disease. American Journal of Respiratory and Critical Care Medicine, 154, 1310–1317. [DOI] [PubMed] [Google Scholar]
- Polkey, M. I. , Kyroussis, D. , Keilty, S. , Hamnegard, C. H. , Mills, G. H. , Green, M. , & Moxham, J. (1995). Exhaustive treadmill exercise does not reduce twitch transdiaphragmatic pressure in patients with COPD. American Journal of Respiratory and Critical Care Medicine, 152, 959–964. [DOI] [PubMed] [Google Scholar]
- Romer, L. M. , & Polkey, M. I. (2008). Exercise‐induced respiratory muscle fatigue: implications for performance. Journal of Applied Physiology, 104, 879–888. [DOI] [PubMed] [Google Scholar]
- Similowski, T. , Duguet, A. , Straus, C. , Attali, V. , Boisteanu, D. , & Derenne, J.‐P. (1996). Assessment of the voluntary activation of the diaphragm using cervical and cortical magnetic stimulation. European Respiratory Journal, 9, 1224–1231. [DOI] [PubMed] [Google Scholar]
- Similowski, T. , Fleury, B. , Launois, S. , Cathala, H. , Bouche, P. , & Derenne, J. (1989). Cervical magnetic stimulation: a new painless method for bilateral phrenic nerve stimulation in conscious humans. Journal of Applied Physiology, 67, 1311–1318. [DOI] [PubMed] [Google Scholar]
- Similowski, T. , Straus, C. , Attali, V. , Duguet, A. , & Derenne, J.‐P. (1998). Cervical magnetic stimulation as a method to discriminate between diaphragm and rib cage muscle fatigue. Journal of Applied Physiology, 84, 1692–1700. [DOI] [PubMed] [Google Scholar]
- Similowski, T. , Yan, S. , Gauthier, A. P. , Macklem, P. T. , & Bellemare, F. (1991). Contractile properties of the human diaphragm during chronic hyperinflation. New England Journal of Medicine, 325, 917–923. [DOI] [PubMed] [Google Scholar]
- Steier, J. , Kaul, S. , Seymour, J. , Jolley, C. , Rafferty, G. , Man, W. , Luo, Y. M. , Roughton, M. , Polkey, M. I. , & Moxham, J. (2007). The value of multiple tests of respiratory muscle strength. Thorax, 62, 975–980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stell, I. , Tompkins, S. , Lovell, A. , Goldstone, J. , & Moxham, J. (1999). An in vivo comparison of a catheter mounted pressure transducer system with conventional balloon catheters. European Respiratory Journal, 13, 1158–1163. [DOI] [PubMed] [Google Scholar]
- Supinski, G. S. , & Callahan, L. A. (2013). Diaphragm weakness in mechanically ventilated critically ill patients. Critical Care, 17, R120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Taylor, B. J. , How, S. C. , & Romer, L. M. (2006). Exercise‐induced abdominal muscle fatigue in healthy humans. Journal of Applied Physiology, 1985(100), 1554–1562. [DOI] [PubMed] [Google Scholar]
- Taylor, B. J. , & Romer, L. M. (2009). Effect of expiratory resistive loading on inspiratory and expiratory muscle fatigue. Respiratory Physiology & Neurobiology, 166, 164–174. [DOI] [PubMed] [Google Scholar]
- Tiller, N. B. , Campbell, I. G. , & Romer, L. M. (2017). Influence of upper‐body exercise on the fatigability of human respiratory muscles. Medicine and Science in Sports and Exercise, 49, 1461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Walterspacher, S. , Isaak, L. , Guttmann, J. , Kabitz, H.‐J. , & Schumann, S. (2014). Assessing respiratory function depends on mechanical characteristics of balloon catheters. Respiratory Care, 59, 1345–1352. [DOI] [PubMed] [Google Scholar]
- Welch, J. F. , Archiza, B. , Guenette, J. A. , West, C. R. , & Sheel, A. W. (2018). Sex differences in diaphragmatic fatigue: the cardiovascular response to inspiratory resistance. Journal of Physiology, 596, 4017–4032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Welch, J. F. , Mildren, R. L. , Zaback, M. , Archiza, B. , Allen, G. P. , & Sheel, A. W. (2017). Reliability of the diaphragmatic compound muscle action potential evoked by cervical magnetic stimulation and recorded via chest wall surface EMG. Respiratory Physiology & Neurobiology, 243, 101–106. [DOI] [PubMed] [Google Scholar]
- Wilcox, P. , Eisen, A. , Wiggs, B. , & Pardy, R. (1988). Diaphragmatic relaxation rate after voluntary contractions and uni‐and bilateral phrenic stimulation. Journal of Applied Physiology, 65, 675–682. [DOI] [PubMed] [Google Scholar]
- Wragg, S. , Hamnegard, C. , Kyroussis, D. , Moran, J. , Green, M. , & Moxham, J. (1994). Potentiation of diaphragmatic twitch after voluntary contraction in normal subjects. Thorax, 49, 1234–1237. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wüthrich, T. U. , Marty, J. , Benaglia, P. , Eichenberger, P. A. , & Spengler, C. M. (2015). Acute effects of a respiratory sprint‐interval session on muscle contractility. Medicine and Science in Sports and Exercise, 47, 1979. [DOI] [PubMed] [Google Scholar]