Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2022 May 4.
Published in final edited form as: Biosens Bioelectron. 2020 Nov 4;173:112764. doi: 10.1016/j.bios.2020.112764

Wireless, continuous monitoring of daily stress and management practice via soft bioelectronics

Hojoong Kim a,b, Yun-Soung Kim a, Musa Mahmood a, Shinjae Kwon a, Fayron Epps c, You Seung Rim b,**, Woon-Hong Yeo a,d,e,*
PMCID: PMC8093317  NIHMSID: NIHMS1660849  PMID: 33190046

Abstract

Stress has become a significant factor, directly affecting human health. Due to the numerous sources of stress that are inevitable in daily life, effective management of stress is essential to maintain a healthy life. Recent advancements in wearable devices allow monitoring stress levels via the detection of galvanic skin response on the skin. Some of these devices show the capability of assessing stress relief methods. However, prior works have been limited in a controlled laboratory setting with a short period assessment (<1 h) of stress intervention. The existing systems’ main issues include motion artifacts and discomfort caused by rigid and bulky electronics and mandatory device connection on active fingers. Here, we introduce soft, wireless, skin-like electronics (SKINTRONICS) that offers continuous, portable daily stress and management practice monitoring. The ultrathin, lightweight, all-in-one device captures the change of a subject’s stress over six continuous hours during everyday activities, including desk work, cleaning, and resting. At the same time, the SKINTRONICS proves that typical stress alleviation methods (mindfulness and meditation) can reduce stress levels, even in the middle of the day, which is supported by statistical analysis. The low-profile, wireless, gel-free device shows enhanced breathability and minimized motion artifacts compared to a commercial stress monitor. Collectively, this study shows the first demonstration of soft, nanomembrane bioelectronics for long-term, continuous assessment of stress and intervention effectiveness throughout daily life.

Keywords: Wireless soft electronics, Continuous stress monitoring, Stress management practice, Galvanic skin response (GSR), Skin temperature

1. Introduction

Work-related stress and its adverse outcomes have resulted in severe risks to individuals and public health (Alberdi et al., 2016; Ebert et al., 2018). For example, excessive workplace stress caused 120,000 deaths, with $190 billion in healthcare costs per year in the United States (Goh et al., 2016). Individuals with high levels of stress suffer from negative physical and psychological consequences, such as sleeping problems (Akerstedt, 2006), depression (Heber et al., 2016; Kalia 2002), and cardiovascular disease (Can et al., 2019; McEwen 2008). Therefore, early detection and management of excessive stress are critical to maintaining a healthy life. Stress levels have been estimated by a self-graded response or behavior data, which is subjective and hardly quantifiable (Can et al., 2019; Sharma and Gedeon 2012). Recently, to evaluate the effectiveness of stress management, wearable systems have been developed by measuring physiological signals from the heart, muscle, and brain (Can et al., 2019; Lim et al., 2020; Sharma and Gedeon 2012).

Electrodermal activity, also known as galvanic skin response (GSR), is one of the key indicators that directly evaluate stress arousal and cognitive states by sensing the sympathetic nervous system (Cabibihan and Chauhan, 2017; Dehzangi et al., 2018; Hernando-Gallego et al., 2018). The cognitive and emotional stressor activates the sympathetic nervous system (SNS), promoting the eccrine glands’ secretion to generate sweat on the skin (Boucsein, 2012). The GSR sensors can monitor the sympathetic activity by detecting the variation of the ionic permeability of sweat gland membranes (Fig. S1). The phasic signal of GSR, rapid time-varying response, is correlated with the arousals by SNS (Dehzangi et al., 2018; Hernando-Gallego et al., 2018). Thus, identification of the phasic component of GSR allows the quantification of the stress. GSR is typically measured by attaching wired gel-electrodes on fingers or hands where the sweat glands exist densely (Kappeler-Setz et al., 2011; Subramanian et al., 2018; Yin et al., 2019). However, these locations cause significant issues of motion artifacts and data loss from frequently disconnected wires. Recent advances in miniaturized wearable devices allow GSR’s wireless detection on other areas, such as wrists and arms (Gautam et al., 2018; Olbrich et al., 2019). These devices, however, still suffer from side effects, including substantial motion artifacts caused by rigid sensor dissociation from the skin, skin irritation from electrolyte gels, and discomfort due to aggressive fixtures and straps (Herbert et al., 2018; Kwon et al., 2020; Lim et al., 2020). Recent reports show that wearable devices can measure stress intervention and the effectiveness of the management practice (Svetlov et al., 2019) (Yin et al., 2019) (Liu et al., 2019) (Fallon et al., 2020). Nevertheless, these devices still rely on bulky, rigid sensors and electronics, limiting the continuous and long-term use in daily life. In addition, all of those studies have been conducted in controlled laboratory environments with limited monitoring time, less than an hour.

This paper introduces a new class of technologies that develop soft, nanomembrane biosensors, and stretchable bioelectronics. The all-in-one wearable device, named SKINTRONICS, offers wireless, portable, continuous, and long-term (>6 h) recording of GSR and temperature to assess stress on the skin during daily life. Also, the ultrathin and lightweight system can monitor stress management practice and the intervention efficacy via statistical analysis. The soft wearable device that makes conformal, intimate contact to the skin without electrolyte gels allows the recording of high-quality physiological data with minimized motion artifacts. A set of experimental studies proves the low-profile device’s enhanced breathability along with no side effects to the skin after a 6-h long device wearing. Real-time monitoring of stress alleviation methods demonstrates the effectiveness of mindfulness of meditation to relieve stress in the middle of daily activities.

2. Materials and Methods

2.1. Fabrication of SKINTRONICS

Nanomembrane electrodes and stretchable circuits were fabricated using the combination of standard microfabrication, material transfer printing, and soft material packaging (Kim et al., 2019; Mahmood et al., 2019; Mishra et al., 2020). For the electrodes, Cr and Au were sputtered on a Si wafer and patterned into an open-mesh, meander structure, followed by etching steps. For the stretchable circuit fabrication, polyimide and polydimethylsiloxane (PI/PDMS) layers were spin-coated on a Si wafer. Then, the 1st Cu layer was deposited by sputtering and patterned into a serpentine mesh network. Additional layers (PI/Cu/PI) were deposited, and the 1st and the 2nd Cu layers were connected. Then, the circuit surface was etched by reactive ion etcher, leaving PI-insulated Cu traces and exposed Cu pads for subsequent soldering of electronic chip components. Additional details of the electrodes and circuits’ fabrication methods are summarized in Supplementary Note S1, Fig. S2, and Table S1.

2.2. Data recording, signal processing, and analysis

For stress monitoring, a fabricated SKINTRONICS was mounted on a subject’s non-dominant hand’s inner wrist. The SKINTRONCS measures GSR as the Wheatstone bridge’s potential variation. A digital potentiometer actively moves a baseline potential to the central position in the detection range (0–1 V) to increase signal sensitivity. The sampling rate was varied from 1 to 5 Hz, depending on the target recording time. Recorded GSR data was stored directly onto the microcontroller’s flash memory. For validation of the device performance, a commercial physiological monitor (BioRadio, Great Lakes NeuroTechnologies) was utilized with three gel-covered snap electrodes. These sensors were mounted on the middle phalanx of the index finger, the middle finger, and the hand’s back for grounding. An armband strap secured these sensors on the skin, connected to the BioRadio. The sampling rate was 250 Hz and averaged to 10 Hz for further analysis. The phasic components of GSR signals were extracted from raw data using the band-pass filter (0.2–1 Hz). The root-mean-square (RMS) value of the phasic components was calculated as a threshold level to detect GSR peaks. The peaks above the threshold were defined as stress arousals and those numbers were counted every minute. Signal-to-noise ratio (SNR) was calculated based on the phasic GSR and noise signals, including motion artifacts identified by the high-pass filter (1 Hz). After RMS conversion of both signals, SNR values (unit: dB) are calculated by the following equation:

SNR=10Log10(RMS_signalRMS_noise) (1)

2.3. Procedures for monitoring of stress in daily life

The all-in-one, low-profile SKINTRONICS offers portable, continuous monitoring of stress. Subjects were asked to wear the device on the inner wrist with multiple activities, such as deskwork (reading articles and data analysis) and vacuum cleaning. In addition, each subject attempted stress alleviation activities, such as mindfulness or meditation. For mindfulness, each subject was asked to maintain any restful activity for 10 min. For meditation, each subject was asked to turn on calm music, close the eyes, and keep their mind peaceful while concentrating on breathing for 10 min. We used a small lithium polymer battery (110 mAh) for the continuous data recording, integrated with the wearable device (Fig. S3). This battery could record both GSR and temperature up to 7 h. Afterward, a magnet-assisted connection was required to recharge the battery. Overall, the human pilot study involved two healthy volunteers; the study followed the approved IRB (#H20204). All participants agreed and signed to the consent form to allow the experiment procedure.

3. Results and discussion

3.1. System architecture of SKINTRONICS and applications for daily stress management

In this study, we develop a wearable, wireless, all-in-one system (SKINTRONICS), which enables continuous, long-term monitoring of daily stress as well as a management practice for multiple days. Fig. 1A captures the overview of the use of SKINTRONICS on the skin for monitoring stress factors and the effectiveness of stress alleviation practice throughout daily activities. The SKINTRONICS is composed of a pair of nanomembrane electrodes and a stretchable circuit enclosed by a soft elastomeric membrane. Overall, the device is ultrathin (<5 mm in thickness), lightweight (<6 g even with a battery), and low modulus (<100 kPa), providing an exceptionally small form factor for comfortable and long-term wearability on the skin. A flow chart in Fig. 1B illustrates the overall procedure to quantify stress levels based on GSR and temperature recording on the skin. The measured signals are stored in the flash memory of the on-board microcontroller chip. Temperature signals, measured on the skin, are used to compensate for undesirable GSR fluctuations due to the temperature change. Details of the device structure appear in Fig. 1C, which shows a multi-layered arrangement, including electrodes, conductive films, wireless circuit, rechargeable battery, soft encapsulant, and medical film. Flexible film cables connect the nanomembrane electrodes to the stretchable circuit on top of the soft elastomer. Photos in Fig. 1D captures the form factor of the fabricated SKINTRONICS, integrated with a clinical-grade medical tape for skin mounting.

Fig. 1. Overview of wireless, continuous monitoring of daily stress and management practice via soft nanomembrane bioelectronics.

Fig. 1.

A) SKINTRONICS for real-time, continuous monitoring of daily stress factors and management practice. B) Flowchart showing the use of SKINTRONICS for monitoring of GSR and temperature on the skin. C) Schematic illustration of the fully integrated device composed of a multi-layered circuit, a pair of nanomembrane electrodes, and a rechargeable battery. D) Photos showing an ultrathin and flexible device in an angled view (left) and side view (right).

3.2. Manufacturing, mechanical reliability, and breathability of SKINTRONICS

A series of optical microscope images and photos in Fig. 2A summarizes the device fabrication steps, including circuit preparation, transfer printing, electrodes integration, and skin mounting process. The fabrication of a microstructured circuit on a Si wafer follows the standard micromachining techniques with photolithography, metallization, and etching. We used an open-mesh, meander design to construct the circuit interconnects for mechanical flexibility and stretchability when mounted on the skin. A water-soluble tape facilitates the retrieval of the fabricated circuit patterns for transfer printing onto an elastomeric membrane. A follow-up integration of functional chips (microcontroller, analog front end, thermistor, digital potentiometer, and charging controller) finishes the circuit fabrication. The next step is to assemble the stretchable Au electrodes and circuits with a clinical-grade medical film (Tegaderm, 3 M, 7 × 6 cm2). The electrodes are attached to the film’s adhesive side while the circuit is placed on top of the film, facilitated by a gel elastomer (1:2 mixture of Ecoflex 00–30 and Gels, Smooth-On). A pair of GSR electrodes have 200 nm in thickness, with the pattern size 2 × 2 cm2 and inter-distance of 0.3 cm between the pair. The circuit size is 3.1 × 2.2 cm2’ with 2 mm in thickness. The total thickness of the SKINTRONICS, including a rechargeable battery, is less than 5 mm. For continuous monitoring of stress, we mounted the SKINTRONICS on the inner wrist. Details of the circuit design and electronic wiring appear in Figs. S4 and S5, respectively.

Fig. 2. Fabrication, mechanical flexibility, and breathability of SKINTRONICS.

Fig. 2.

A) A multi-step fabrication process of SKINTRONICS: (step 1) preparation of circuit interconnects on a Si wafer via microfabrication, (step 2) transfer-printing of the interconnects on a soft elastomer, followed by integration of microchips, (step 3) integration with electrodes (red square) on a Tegaderm patch, along with a rechargeable battery (blue square), and (step 4) mounting of the device on the wrist for stress monitoring and management. B) Photos showing the device’s stretchable, bendable, and compressible characteristics. C) Photo of a device with three different locations: case 1 – green, Tegaderm film only, case 2 – red, electrode only, and case 3 – blue, electrode and circuit. D) Results of water weight loss according to the membrane type, compared with a medical-grade, breathable silicone tape. E) Measured water-vapor transmission rates (WVTR) for four membrane types in D.

Collective photos in Fig. 2B capture the mechanical characteristics of the compliant circuit and electrodes, which can endure stretching, bending, and compression without mechanical failure. The open-mesh structures, used in both electrodes and circuits, accommodate applied strains from bending and stretching. Considering the continuous use of the wearable device, it is critical to validate the mechanical robustness. Our prior work proves the electrode’s functionality with 30% uniaxial stretching up to 400 cycles (Kim et al., 2019). Also, a mechanical stretching test in Fig. S6 shows the circuit’s maximum stretchability over 20% and reliability upon repeated tensile strain (10%) over 500 cycles. The measured electrical signals show negligible changes.

Another essential feature of a wearable device is the breathability for long-term wearing on the skin (Mishra et al., 2017; Sun et al., 2018). The ability to allow air permeation through the membrane improves prolonged adhesion of the device as well as minimized skin irritation. The experimental results in Fig. 2C-E shows the characterized breathability of materials via measured changes of water loss. There are four membrane types, including Tegaderm film only, electrodes with the film, electrodes and circuit with the film, and another clinical-grade silicone film (Kind Removal, 3 M). In this study, we measured the change of water vapor transmission rate (WVTR) over time, details of the experimental method in Fig. S7. WVTR for each membrane was measured at the constant temperature (24 °C) and humidity (50%). The graph in Fig. 2D shows water loss per each case over 48 h. WVTR is calculated by the following equation (Hu et al., 2019):

WVTR=ΔgA×t (2)

where Δg, A, and t are the weight changes of water, membrane area, and elapsed time, respectively. As summarized in Fig. 2E, the laminated circuit and electrode slightly decrease the values. However, the fully integrated membrane with additional components still shows a similar WVTR level as the clinical-grade, breathable silicone tape, demonstrating the device feasibility for multi-hour use on the skin. Collectively, the presented experimental results in Fig. 2 show the acceptable ranges of mechanical reliability and breathability of the SKINTRONICS for a pilot study with human subjects.

3.3. Validation of the device performance on stress monitoring

In this study, we validate the performance of SKINTRONICS on stress monitoring with human subjects. Specifically, Fig. 3 summarizes the side-to-side comparison of recorded data between SKINTRONICS and a clinical-grade commercial device (BioRadio, Great Lakes Neuro-Technologies). Fig. 3A captures the exceptionally small form factor of the SKINTRONICS on the inner wrist compared to the commercial wearable system that wirelessly transfers GSR signals via Bluetooth. The commercial system utilizes a pair of gel-covered rigid electrodes mounted on two fingers and a data acquisition module, worn on the arm. The wireless, all-in-one SKINTRONICS (5.3 g) is about 25 times lighter than the BioRadio (134.1 g) with multiple electronic components connected by wires. The skin-electrode contact impedance in Fig. 3B shows the difference between two devices where the lowered value from SKINTRONICS is caused by the highly conformal contact of large-area electrodes. In this study, we increased the electrode size of SKINTRONICS up to 4.0 cm2 compared to the gel (1.4 cm2) to enhance the areal coverage. As a result, the dry, skin-like electrode shows a lower skin-electrode contact impedance than the gel electrode on the finger.

Fig. 3. Validation of the device performance compared to a commercial device on stress monitoring.

Fig. 3.

A) Photo that compares the form factor and weight of two devices. B) Comparison of skin-electrode contact impedance. Error bars show standard deviation (n = 5). C) Three representative daily activities: deskwork (DW), housework (HW), and meditation (MD). D) Flowchart showing the data processing sequence. E) Raw GSR and F) their phasic signals measured by BioRadio (top) and SKINTRONICS (bottom). G) Number of peaks for three activities (n = 3). H) Noise data caused by motion artifacts. I) Comparison SNR values for three activities (n = 3).

Fig. 3C captures three representative daily activities that a subject follows while measuring GSR signals from both devices simultaneously. Recorded data are analyzed by going through the multi-step processes, as described in Fig. 3D, including phasic element extraction, threshold determination, and peak identification. A set of graphs in Fig. 3E presents raw GSR data, measured by BioRadio on fingers and SKINTRONICS on the wrist from three activities: deskwork, housework, and meditation (details in Materials and Methods). Fig. 3F summarizes the collected data after the extraction of phasic elements with RMS values. The average numbers of peaks counted from different activities in Fig. 3G show a significant reduction in peak numbers during meditation (1.61 and 1.33 for SKINTRONICS and BioRadio, respectively), compared to during other work-related activities (6.53–8.90 and 8.83–11.90). The result validates that both deskwork and housework cause stress to the user, while the stationary meditation effectively reduces the stress values significantly. Note that the counted number of peaks from BioRadio is higher than that from SKINTRONICS because fingers have more sweat glands than the wrist (Payne et al., 2016). Three times of the GSR measurements by SKINTRONICS with replacing electrodes show reliable data recording.

Considering real-time, continuous monitoring of GSR in daily life, it is critical to minimize noise, mostly from motion artifacts (Kwon et al., 2020; Lim et al., 2020; Mahmood et al., 2019). The lightweight, soft, skin-conformal SKINTRONICS demonstrates minimal effects from motion artifacts compared to BioRadio, as summarized in Fig. 3H and I. The conductive gel on a rigid sensor has drift motions at the interface during daily activities, while wires and other electronic components are susceptible to movements during deskwork and housework. Overall, the set of experimental studies in Fig. 3 validates the enhanced performance of the SKINTRONICS in wireless, continuous, and ambulatory monitoring of GSR with negligible motion artifacts.

3.4. Long-term recording of stress and the effectiveness of management practice

The human body has the function of thermoregulation that actively controls body temperature by sweating through the skin. In other words, GSR signals, measuring the change of skin conductance on sweat glands, are affected by the variation of body temperature (Doberenz et al., 2011; Gerrett et al., 2019). The uncontrollable ambient temperature must be carefully considered when measuring ambulatory GSR during a subject’s daily life. As shown in Fig. 1, SKINTRONICS includes two sensors that measure GSR and skin temperature simultaneously to compensate for undesirable thermal effects during everyday activities.

Photos in Fig. 4A show thermographic images of the inner wrist without (top) and with (bottom) SKINTRONICS, captured by an infrared camera (E8, FLIR). Compared to the camera, the temperature sensor in SKINTRONICS shows lower values (Fig. 4B), caused by the device encapsulants that absorb and distribute external thermal energy. Moving activity results in temperature reduction (Fig. 4C), which is also confirmed by the thermal camera (Fig. S8). Temperature change, departed from normal lukewarm level, causes the GSR signal boost in the peak counting (Fig. 4D). These results indicate that constant moving and other daily life activities can overestimate the stress level, showing the importance of simultaneous temperature recording for necessary compensation. Fig. 4E shows temperature data (left) measured during daily activities over 6 h and averaged variation per minute (right). Fig. 4F summarizes a set of stress levels by counting GSR peaks per minute during multiple tasks, including desk work, housework, and meditation. Note that the number of GSR peaks include the compensation of temperature variation, while the colorized scale shows the level of stress intensity from high (red color) to low (green color). The stress level increases gradually with continuous deskwork. At the elapsed time between 1.5 and 3 h, the stress level becomes extremely high as the various work-related events are experienced. After meditation at 3 h, the stress level is decreased and lower than the initial work period. Another meditation after 5 h causes the lowest stress level for the subject. The SKINTRONICS, enclosed by a medical film (Tegaderm), is designed for a single-use since it loses the adhesion when detached from the skin. Considering possible reuse of the device, we developed another type of SKINTRONICS encapsulated by a reusable silicone tape (Kind Removal, 3 M), which offered multiple uses more than three times (Fig. S9). The reusable device measures continuous, high-quality GSR signals during multiple activities and meditation.

Fig. 4. Long-term recording of stress and effectiveness of management practice.

Fig. 4.

A) Thermographic images of the inner wrist without (top) and with SKINTRONICS (bottom). The average temperature value is calculated from the indicated area in the white box. B) Comparison of measured temperature by an IR camera and SKINTRONICS. Error bars show standard deviation (n = 3). C) Temperature variation with moving activity (n = 3). D) Temperature effects on the number of stress peaks (n = 3). E) Continuously measured skin temperature by SKINTRONICS for over 6 h (left) and averaged variation per minute (right). F) Temperature-calibrated stress levels during multiple daily activities with deskwork, housework, and meditation. The upper bar displays a visualized stress intensity according to the measured stress peaks. G) Photo showing no side effects of the wearable device on the skin after 6 h. H) Comparison of two types of stress alleviation: mindfulness (left) and meditation (right), measured for 6 days with two subjects. Each stress level is averaged within 0.5 h prior and post to the alleviation. I) Calculated effectiveness of stress management practice. Control data shows no alleviation activity.

The breathable SKINTRONICS with good air permeability (studied in Fig. 2) shows no skin irritation and allergic reaction after the continuous use over 6 h (Fig 4G). The novelty of real-time, continuous recording with SKINTRONICS is summarized in Fig. 4H and I, which can quantify the effectiveness of stress management practice such as mindfulness and meditation. Fig. 4H presents various stress levels from two subjects who use two types of stress alleviation methods for six days. Video S1 captures real-time, continuous monitoring of GSR and temperature during meditation and desk work. Details of the measured data appear in Figs. S10 and S11. Each work-related stress level is averaged within 0.5 h prior and post to the alleviation activities. Fig. 4I compares each alleviation method’s efficacy via the mean difference of stress indicators, showing more effective stress reduction from meditation (60%) than mindfulness (38%). To confirm the impact of stress management, we used statistical analysis by using a two-way analysis of variance (Christensen 2020; Liu et al., 2019). The univariate assessment finds a significant effect on the alleviation activities as F = 7.72, p = 0.003 (α = 0.05) where F value is the ratio of the variance of group means over the average values in group variances, p is the evidence for a hypothesis, and α is the confidence level. Table 1 compares the performance of SKINTRONICS with other wearable stress monitors and intervention methods. High F ratio in this work clearly shows significant alleviation of stress with reliability (smaller p-value than α value). This study collectively shows the first demonstration of a soft, wearable biosystem for long-term (>6 h), continuous assessment of stress, and intervention effectiveness throughout daily life.

Table 1.

Comparison of wearable stress detection sensors and intervention methods.

Reference Device type
(electrode)
Device
location
Stress inducing
method
Stress alleviating
method
Intervention
effectiveness
(Δmean)
Statistical values Intervention
monitoring time
Data
recording
place
This work Soft, stretchable patch (dry electrode) Wrist Working (desk work/housework) Meditation/mindfulness 60%/38% F = 7.72, p = 0.003 (two-way ANOVA) >6 h Home/Laboratory
Svetlov et al. (2019) Rigid, wristband (dry electrode) Wrist Word memorization Mindfulness relaxation 49% Not applicable 40 min Laboratory
Yin et al. (2019) Rigid, wristband (gel electrode) Palm VR VR 4% Not applicable 52 min Laboratory
Liu et al. (2019) Rigid, bulky electronics (gel electrode) Finger TSST Watching a video 43% F = 5.27, p = 0.001 (multivariate ANOVA) <60 min Laboratory
Fallon et al. (2020) Rigid, bulky electronics (gel electrode) Finger TSST Music therapy 8–23% F = 5.10, p < 0.01 (two-way ANOVA) 20 min Laboratory
Akmandor and Jha (2017) Rigid, bulky electronics (gel electrode) Finger Memory game, fly sound, ice test, and picture system Music, warm stone, meditation, and news 49% Not applicable 64 min Laboratory

VR: virtual reality, TSST: trier social stress test, and ANOVA: analysis of variance.

4. Conclusion

This paper reports an all-in-one, wireless, soft bioelectronic system for portable, continuous monitoring of stress and management practice in daily life. The fully integrated stretchable system incorporates skin-conformal nanomembrane electrodes and wireless circuits. The wearable device on the wrist measures high-fidelity GSR and temperature with minimized motion artifacts and enhanced breathability, compared to a commercial device. Simultaneous skin temperature recording provides accurate detection of stress by removing unwanted contributions of temperature changes from sweating. In vivo demonstration with human subjects captures the device performance of continuous stress detection over 6 h during multiple daily activities, including desk work, cleaning, and stress alleviation. Collectively, the bioelectronic stress monitor, presented in this work, would provide a novel wearable platform for users to monitor daily stress factors actively and control them via management practice.

Supplementary Material

Supplementary Information
Supplementary Video
Download video file (7.8MB, mp4)

Acknowledgments

W.-H.Y. acknowledges the support by the Goizueta Alzheimer’s Disease Research Center (P50 AG025688). This work was partially supported by the National Science Foundation (grant NRI-2024742) and the Georgia Tech IEN Center Grant. This work was performed in part at the Institute for Electronics and Nanotechnology, a member of the National Nanotechnology Coordinated Infrastructure, which is supported by the National Science Foundation (Grant ECCS-2025462). H.K. acknowledges the support by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2019R1A6A3A01096242).

Footnotes

Declaration of competing interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Georgia Tech has a pending US patent application.

Appendix A. Supplementary data

Supplementary data to this article can be found online at https://doi.org/10.1016/j.bios.2020.112764.

References

  1. Åkerstedt T, 2006. Psychosocial stress and impaired sleep. Scand. J. Work. Environ. Hea 32 (6), 493–501. [PubMed] [Google Scholar]
  2. Akmandor AO, Jha NK, 2017. Keep the stress away with SoDA: stress detection and alleviation system. IEEE Trans. Multi-Scale Comput. Syst 3 (4), 269–282. [Google Scholar]
  3. Alberdi A, Aztiria A, Basarab A, 2016. Towards an automatic early stress recognition system for office environments based on multimodal measurements: a review. J. Biomed. Inf 59, 49–75. [DOI] [PubMed] [Google Scholar]
  4. Boucsein W, 2012. Electrodermal Activity. Springer Science & Business Media. [Google Scholar]
  5. Cabibihan J-J, Chauhan SS, 2017. Physiological responses to affective tele-touch during induced emotional stimuli. IEEE T. Affect. Comput 8 (1), 108–118. [Google Scholar]
  6. Can YS, Arnrich B, Ersoy C, 2019. Stress detection in daily life scenarios using smart phones and wearable sensors: a survey. J. Biomed. Inf 92, 103139. [DOI] [PubMed] [Google Scholar]
  7. Christensen R, 2020. One-way ANOVA. Plane Answers to Complex Questions. Springer, pp. 107–121. [Google Scholar]
  8. Dehzangi O, Rajendra V, Taherisadr M, 2018. Wearable driver distraction identification on-the-road via continuous decomposition of galvanic skin responses. Sensors 18 (2). [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Doberenz S, Roth WT, Wollburg E, Maslowski NI, Kim S, 2011. Methodological considerations in ambulatory skin conductance monitoring. Int. J. Psychophysiol 80 (2), 87–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Ebert DD, Kahlke F, Buntrock C, Berking M, Smit F, Heber E, Baumeister H, Funk B, Riper H, Lehr D, 2018. A health economic outcome evaluation of an internet-based mobile-supported stress management intervention for employees. Scand. J. Work. Environ. Health 44 (2), 171–182. [DOI] [PubMed] [Google Scholar]
  11. Fallon VT, Rubenstein S, Warfield R, Ennerfelt H, Hearn B, Leaver E, 2020. Stress reduction from a musical intervention. Psycho. 30 (1), 20–27. [Google Scholar]
  12. Gautam A, Simoes-Capela N, Schiavone G, Acharyya A, de Raedt W, Van Hoof C, 2018. A data driven empirical iterative algorithm for GSR signal pre-processing. In: 26th European Signal Processing Conference (EUSIPCO), pp. 1162–1166. [Google Scholar]
  13. Gerrett N, Amano T, Havenith G, Inoue Y, Kondo N, 2019. The influence of local skin temperature on the sweat glands maximum ion reabsorption rate. Eur. J. Appl. Physiol 119 (3), 685–695. [DOI] [PubMed] [Google Scholar]
  14. Goh J, Pfeffer J, Zenios SA, 2016. The relationship between workplace stressors and mortality and health costs in the United States. Manag. Sci 62 (2), 608–628. [Google Scholar]
  15. Heber E, Lehr D, Ebert DD, Berking M, Riper H, 2016. Web-based and mobile stress management intervention for employees: a randomized controlled trial. J. Med. Internet Res 18 (1), e21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Herbert R, Kim J-H, Kim YS, Lee HM, Yeo W-H, 2018. Soft material-enabled, flexible hybrid electronics for medicine, healthcare, and human-machine interfaces. Materials 11 (2), 187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Hernando-Gallego F, Luengo D, Artes-Rodriguez A, 2018. Feature extraction of galvanic skin responses by nonnegative sparse deconvolution. IEEE J. Biomed. Hea. Inf 22 (5), 1385–1394. [DOI] [PubMed] [Google Scholar]
  18. Hu X, Jia X, Zhi C, Jin Z, Miao M, 2019. Improving the properties of starch-based antimicrobial composite films using ZnO-chitosan nanoparticles. Carbohydr. Polym 210, 204–209. [DOI] [PubMed] [Google Scholar]
  19. Kalia M, 2002. Assessing the economic impact of stress [mdash] the modern day hidden epidemic. Metabol. 51 (6), 49–53. [DOI] [PubMed] [Google Scholar]
  20. Kappeler-Setz C, Gravenhorst F, Schumm J, Arnrich B, Tröster G, 2011. Towards long term monitoring of electrodermal activity in daily life. Pers. Ubiquit. Comput 17 (2), 261–271. [Google Scholar]
  21. Kim YS, Mahmood M, Lee Y, Kim NK, Kwon S, Herbert R, Kim D, Cho HC, Yeo WH, 2019. All-in-One, wireless, stretchable hybrid electronics for smart, connected, and ambulatory physiological monitoring. Adv. Sci 6 (17), 1900939. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Kwon S, Kwon Y-T, Kim Y-S, Lim H-R, Mahmood M, Yeo W-H, 2020. Skin-conformal, soft material-enabled bioelectronic system with minimized motion artifacts for reliable health and performance monitoring of athletes. Biosens. Bioelectron 151, 111981. [DOI] [PubMed] [Google Scholar]
  23. Lim HR, Kim HS, Qazi R, Kwon YT, Jeong JW, Yeo WH, 2020. Advanced soft materials, sensor integrations, and applications of wearable flexible hybrid electronics in healthcare, energy, and environment. Adv. Mater 32 (15), e1901924. [DOI] [PubMed] [Google Scholar]
  24. Liu JJW, Reed M, Vickers K, 2019. Reframing the individual stress response: balancing our knowledge of stress to improve responsivity to stressors. Stress Health 35 (5), 607–616. [DOI] [PubMed] [Google Scholar]
  25. Mahmood M, Mzurikwao D, Kim Y-S, Lee Y, Mishra S, Herbert R, Duarte A, Ang CS, Yeo W-H, 2019. Fully portable and wireless universal brain–machine interfaces enabled by flexible scalp electronics and deep learning algorithm. Nat. Mac. Intel 1 (9), 412–422. [Google Scholar]
  26. McEwen BS, 2008. Central effects of stress hormones in health and disease: understanding the protective and damaging effects of stress and stress mediators. Eur. J. Pharmacol 583 (2–3), 174–185. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Mishra S, Kim Y-S, Intarasirisawat J, Kwon Y-T, Lee Y, Mahmood M, Lim H-R, Herbert R, Yu KJ, Ang CS, 2020. Soft, wireless periocular wearable electronics for real-time detection of eye vergence in a virtual reality toward mobile eye therapies. Sci. Adv 6 (11) eaay1729. [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Mishra S, Norton JJS, Lee Y, Lee DS, Agee N, Chen Y, Chun Y, Yeo WH, 2017. Soft, conformal bioelectronics for a wireless human-wheelchair interface. Biosens. Bioelectron 91, 796–803. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Olbrich S, Jahn I, Stengler K, 2019. Exposure and response prevention therapy augmented with naltrexone in kleptomania: a controlled case study using galvanic skin response for monitoring. Behav. Cognit. Psychother 47 (5), 622–627. [DOI] [PubMed] [Google Scholar]
  30. Payne AF, Schell AM, Dawson ME, 2016. Lapses in skin conductance responding across anatomical sites: comparison of fingers, feet, forehead, and wrist. Psychophysiology 53 (7), 1084–1092. [DOI] [PubMed] [Google Scholar]
  31. Sharma N, Gedeon T, 2012. Objective measures, sensors and computational techniques for stress recognition and classification: a survey. Comput. Meth. Progr. Biomed 108 (3), 1287–1301. [DOI] [PubMed] [Google Scholar]
  32. Subramanian R, Wache J, Abadi MK, Vieriu RL, Winkler S, Sebe N, 2018. ASCERTAIN: emotion and personality recognition using commercial sensors. IEEE T. Affect. Comput 9 (2), 147–160. [Google Scholar]
  33. Sun B, McCay RN, Goswami S, Xu Y, Zhang C, Ling Y, Lin J, Yan Z, 2018. Gaspermeable, multifunctional on-skin electronics based on laser-induced porous graphene and sugar-templated elastomer sponges. Adv. Mater 30 (50), 1804327. [DOI] [PubMed] [Google Scholar]
  34. Svetlov AS, Nelson MM, Antonenko PD, McNamara JPH, Bussing R, 2019. Commercial mindfulness aid does not aid short-term stress reduction compared to unassisted relaxation. Heli. 5 (3), e01351. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Yin J, Arfaei N, MacNaughton P, Catalano PJ, Allen JG, Spengler JD, 2019. Effects of biophilic interventions in office on stress reaction and cognitive function: a randomized crossover study in virtual reality. Indoor Air 29 (6), 1028–1039. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Information
Supplementary Video
Download video file (7.8MB, mp4)

RESOURCES