Abstract
Even though ultrahigh frequency ultrasonic transducers over 60 MHz have been used for single-cell-level manipulation such as intracellular delivery, acoustic tweezers, and stimulation to investigate cell phenotype and cell mechanics, no techniques have been available to measure the actual acoustic radiation force (ARF) applied to target cells. Therefore, we have developed an approach to measure the ARF of ultrahigh frequency ultrasonic transducers using a theoretical model of the dynamics of a solid sphere in a gelatin phantom. To estimate ARF at the focus of a 130 MHz transducer, we matched measured maximum displacements of a solid sphere with theoretical calculations. We selected appropriate ranges of input voltages and pulse durations for single-cell applications, and the estimated ARF was in the range of tens of μN. To gauge the influence of pulse duration, an impulse of different pulse durations was estimated. Fluorescence resonance energy transfer live cell imaging was demonstrated to visualize calcium transport between cells after a target single cell was stimulated by the developed ultrasonic transducer.
Single-cell engineering has great potential to visualize active molecular events during cell cycle while studying cell–cell interactions and sort intact cells without labeling.1–7 While lasers have been used for single-cell-level manipulation due to their short wavelengths, photobleaching and thermal damage to cells may affect cell viability and functional and phenotypical changes after the manipulation.8,9 As the fabrication technique of ultrahigh-frequency ultrasonic transducers ranging from 60 MHz to 300 MHz has emerged in the last two decades,10–12 the wavelength of ultrasound is comparable to the size of single cells, which allows us to manipulate single cells and micrometer-sized particles with minimal disturbance in deep tissues. Ultrahigh-frequency ultrasound-based imaging improves the spatial resolution to better identify abnormalities in tissue samples. We have developed intravascular ultrasound (IVUS) imaging transducers to improve the resolution for detecting lesion on the lumen wall12,13 and single-cell intracellular delivery of various types of macromolecules into cell cytoplasm5,6 using ultrahigh-frequency transducers up to 150 MHz. Single-cell acoustic trapping14 and cell signaling monitoring using transducers with the center frequency of up to 200 MHz were demonstrated.3,4,14
The main and immediate questions using ultrahigh-frequency ultrasonic transducers are the measurements of acoustic radiation force (ARF) at the focus of transducers. Current technology is not mature enough to measure ARF of ultrahigh-frequency ultrasound with the center frequency of over 60 MHz.15,16 Many applications using ultrahigh-frequency ultrasound only refer to the input voltages and duty factor without measuring actual ARF at focus. Numerical simulation using commercial packages was used to estimate ARF as an alternative.
Since we engineered immune cells and developed the next generation intracellular delivery technique using ultrasonic transducers, our lab standardized transducer's specification such as aperture diameter, center frequency (fc), fnumber, and housing sizes to systemically optimize the performance of transducers for these applications and to investigate biological effects of high-frequency ultrasonic transducers to cells.5,6,17–20
Here, we quantitatively measured ARF of ultrahigh frequency transducers, which provides a better understanding of the mechanism of high frequency ultrasound-based single-cell manipulation. We estimated ARF of a pushing ultrasonic transducer [1, PUT in Figs. 1(a) and 1(b)] using a model-based approach by mapping measured displacements of a solid sphere in gelatin phantom with theoretical predictions. The theoretical model of the motion of a solid sphere in a viscoelastic medium was developed and experimentally validated in our lab.21–24 In our previous study, we developed a theoretical model of a gas bubble and a solid sphere under ARF and measured the mechanical properties of the crystalline lens and the vitreous humor of bovine and porcine eyes using a 3-MHz transducer for pushing and a 25-MHz transducer for tracking.25–28 In this study, we used a 130-MHz PUT to displace a solid sphere in a gelatin phantom and a 45-MHz tracking ultrasonic transducer (TUT) to track the displacement of the sphere. If this approach is established to measure the ARF of ultrahigh frequency ultrasonic transducers, this approach will become a gold standard to be widely used in ultrasonic community.
To develop a system, we fabricated a PUT and TUT using lithium niobate plates (LNB, Boston Piezo-Optics) by following our lab's protocol.12 SEM was used to measure the thickness of LNB of the PUT, and it was approximately 10 μm, which matched our designed specification. The measured center frequency (fc) of the PUT and TUT was 130 MHz and 45 MHz shown in Fig. 1(b) and Table I.
TABLE I.
fc (MHz) | LNB thickness (μm) | Aperture (mm) | fnumber | Bandwidth (BW) (%) | |
---|---|---|---|---|---|
TUT | 45 | 50 | 3 | 1.16 | 80 |
PUT | 150 | 10 | 1 | 1 | 58 |
We first placed a carbon steel sphere (CS) or a ruby sphere (Ruby) in a 2% gelatin phantom. We measured shear elastic modulus (μ) using a Discovery Hybrid HR-20 rheometer (TA instruments). The measured shear elastic modulus of 2% gelatin phantom was 2300 ± 900 Pa from ten specimens. The PUT and TUT were carefully aligned using a goniometer and translation stages with a CS or a Ruby sphere as shown in Fig. 1(a). PUT was positioned in the gelatin phantom block with a sphere to minimize ARF loss due to reflection at water and gelatin phantom boundary. After filling the water cuvette with water, the PUT was connected to a power amplifier [8 in Fig. 1(a)] and the TUT was connected to a pulser/receiver [7 in Fig. 1(a)] and an oscilloscope. A function generator [6 in Fig. 1(a)] controlled tracking and pushing triggering for radio frequency (RF) data acquisition for post-processing to reconstruct the displacement of a sphere under ARF [Fig. 1(c)]. Tracking pulses were emitted from the TUT with a pulse repetition frequency of 100 kHz (PRF), and a pushing pulse was emitted from the PUT with a pulse duration (tp) and peak-to-peak voltage (Vpp) as shown in Fig. 2(c). Five hundred pulse/echo was saved in a storage for post-processing to visualize the displacement of a sphere. The sampling frequency (fs) was 10 Gsamples/s, and 50 × 106 samples were saved to track the dynamics of the sphere for 5 ms. The cross correlation speckle tracking method was used to define the displacement of the sphere.29
The previously developed and validated theoretical model formulates the dynamics of a solid sphere in a viscoelastic medium under impulsive ARF. The densities of CS (7850 kg/m3) and Ruby (4000 kg/m3) were used. A brief summary of a theory is the following. We used transient and impulse-like acoustic pulse to displace a solid sphere in nearly incompressible and elastic medium. Similar to the Stoke's formula for a solid sphere moving in liquid, the displacement of a solid sphere in an elastic medium can be found using the balance equation for stress tensor and strain tensor.23,30 The governing equation of motion for a solid sphere in an elastic medium in a frequency domain is22,23,26
(1) |
where ρ is the medium density, is the pressure in the medium in the frequency domain, is the oscillation frequency under excitation (EX) frequency, is the displacement vector, and μ and η are the shear elastic modulus and shear viscosity, respectively.
The general solution of Eq. (1) is
(2) |
where and are complex constants to be determined by the boundary conditions. and , is the radial direction in the polar coordinate system.
Displacement components of are
(3) |
(4) |
Boundary conditions at the surface of a solid sphere (r = R),
(5) |
(6) |
were used to find constants and as
(7) |
(8) |
where is an angle between the radial direction and displacement of a sphere, is the magnitude of , and is the spectral component of a displacement, which will be inverse Fourier transformed at the end to find the actual displacement of a solid sphere in the time domain.
Displaced solid spheres are experiencing forces similar to the drag force in response to the medium on the surface of a solid sphere,
(9) |
The pressure gradient in Eq. (1) and Eqs. (7) and (8) are used to find an expression for pressure:23
(10) |
Stress components are
(11) |
(12) |
Equations (10)–(12) are substituted in Eq. (9) to find the spectral component of force acting on a solid sphere from elastic medium
(13) |
To consider viscoelastic medium, and are changed to and , where is a shear viscosity coefficient.
To couple the displacement of a solid sphere and externally applied force to a sphere in frequency domain () using Newton's second law
(14) |
(15) |
where is a mass of a solid sphere. The spectral component of the displacement of a solid sphere is obtained from Eq. (15)
(16) |
can be determined under impulsive and transient external force defined as follows [Fig. 1(c)]:
(17) |
where is ARF to be measured and is the duration of acoustic radiation pulse [Fig. 1(c)]. In the frequency domain, Eq. (17) becomes
(18) |
Once is found using Eqs. (16) and (18), an inverse Fourier transform { is performed to determine and in Eqs. (7) and (8). Then, using Eqs. (3) and (4), and can be found to determine the displacement of a solid sphere under ARF. Since we consider the displacement of a sphere along the ultrasound propagation, θ is set to be zero. Therefore, represents the displacement of a sphere under consideration.
We first used a 150-μm radius CS sphere to investigate the feasibility of mapping between measurements and theoretical calculations of the dynamics of a solid sphere under ARF generated by a 130-MHz transducer. To estimate ARF (), was fixed to 20 μs and Vpp was increased from 30 to 90 mV. These values were chosen from other publications of our lab's3,5,6 for the intracellular delivery of macromolecules and acoustic single-cell stimulation. Because the gain of the power amplifier is 50 dB, 30 and 90 mV represent 9 to 28 V. Measured displacements of a CS sphere are plotted in Fig. 3(a), and the theoretical calculation to match with the measured displacements is shown in Fig. 3(b). After matching between measurements and theory, was estimated to be 60 , 40 , 24 , and 9 μN for 90 , 70 , 50 , and 30 mV, respectively. To confirm estimated , we fixed 70 mV and changed as 1, 10, 20, and 30 μs. If of 40 μN was correctly estimated, theory and measurement using different values should be matched. As shown in Figs. 3(c) and 3(d), measured displacements and theoretical calculations match for varying values with of 40 μN. Therefore, the measured ARF of Vpp of 90 , 70 , 50 , and 30 mV with fixed 20 μs was confirmed as 60 , 40 , 24 , and 9 μN, respectively.
As a second confirmation, we changed the material of spheres. We used a 150-μm-radius Ruby sphere and compared the displacements of a Ruby sphere with a CS sphere, when Vpp is 70 mV and is 20 μs. Considering the density of Ruby (4000 kg/m3) and CS (7850 kg/m3), time to reach maximum displacement (tmax) was decreased and maximum displacement (Umax) was increased as shown in Figs. 3(e) and 3(f). Under the same ARF, the lighter Ruby sphere travels longer distance within a short time.
Because we confirmed this model-based ARF measurements by varying Vpp, , and the materials of a sphere, we measured ARF and impulse by comparing Umax of a CS and Ruby spheres within a design space of Vpp of 30 , 50 , 70 , and 90 mV and of 5, 10, 15, 20, and 30 μs. We selected Vpp and , which have been frequently used for intracellular delivery of macromolecules, acoustic tweezers, and single cell stimulations using ultrahigh frequency ultrasound. The PUT, TUT, and a sphere were co-aligned in gelatin phantom not to cause any reflection and refraction, which may occur when ultrasound waves are traveling across water and gelatin boundaries. Measured are 9.9 ± 1.4 , 26.5 ± 4.2 , 42.4 ± 3.0 , and 65.1 ± 5.8 μN for 30 , 50 , 70 , and 90 mV, respectively, as shown in Fig. 4(a). We can conclude here that the applied ARF to cells during intracellular delivery and single cell stimulation was in the range of several tens of μN.
To include as a means of a biophysical effect of high frequency ultrasound to cells, impulse was estimated as shown in Fig. 4(b). Briefly, the area under ARF () and time curve represent momentum changes in time. Momentum changes during are impulse, which can be used to measure the biophysical influence on cells of high frequency ultrasound.
We used the PUT to demonstrate single cell stimulation and investigate calcium transport between the cells. An integrated system was used to stimulate a target cell, while capturing fluorescence resonance energy transfer (FRET) changes of the target cell and a neighboring cell to visualize calcium transport between two cells [Fig. 5(a)]. The 3D stage was computer-controlled with a sub-micrometer precision, where the transducer was attached for focusing and stimulation of the target single cell. Foci of the microscope and the transducer were coaligned for FRET-based live cell imaging. To prevent standing waves between a PUT aperture and the bottom of a cell culture dish, PUT was positioned at 45° with respect to cell culture dish [Fig. 5(a)]. Particularly, standing waves induce pressure doubling under the rigid boundary condition.
Ultrasound pulses with Vpp of 70 mV and of 5 μs were applied to the region of interest 1 (ROI 1) shown in Fig. 5(c) for 1 min between 90 s and 150 s with a PRF of 1 kHz. From ARF measurements in Fig. 4(a), 42.4 ± 3.0 μN ARF was applied every 1 ms for 1 min. FRET images were collected with the Ti2 Nikon fluorescence microscope and a cooled charge-coupled device (CCD) camera using the Nikon NIS-Elements AR software. Nikon NIS-Elements AR was used to calculate the pixel-by-pixel ratio images of FRET-YPet over Enhanced cyan fluorescent protein (ECFP) after subtracting the basal fluorescence level. In other words, FRET is the ratio between the intensity of the YPet signal [Y in Fig. 5(a)] and ECFP signal [C in Fig. 5(a)], represented by a simple equation (FRET = Y/C) in Fig. 5(a). A high FRET value of 0.6, in this case, represents an increased intracellular calcium concentration, and a low FRET value of 0.4 indicates a decreased and basal calcium level in cytoplasm of cells. Microscope filter setting for ECFP images was excitation (EX) 440/20 nm and emission (EM) 480/40 nm. FRET-YPet image filter setting was EX 440/20 nm and EM 535/30 nm. Calcium transport between two cells was clearly visualized due to the time shift between maximum FRET in ROIs 2–3 [black dots in Fig. 5(b) and red arrows in Figs. 5(b) and 5(c)]. Figure 5(c)–5(g) represents FRET images at 0, 130, 155, 160, and 220 s, respectively. First FRET increases only within the stimulated cell as shown in Fig. 5(d). Then, FRET increases in both the cells as shown in Fig. 5(e). Quantitatively, the first FRET peak of ROI 1 is at 155 s and the first FRET peak of ROI 5 is at 182 s. Peak FRET of ROIs 2–4 shifts from 159, 181, and 184 s, respectively. These results confirm calcium transport between two cells starting from the stimulated cell.
We developed an approach to measure ARF of ultrahigh-frequency ultrasonic transducers using the theoretical model of the dynamics of a sphere in a gelatin phantom. The impulse due to ARF was estimated to include the influence of . To estimate ARF at the focus of a 130 MHz transducer, we matched measured Umax with theoretical calculations. We selected appropriate ranges of Vpp and for single-cell applications, and estimated ARFs were in the range of tens of μN. To increase the accuracy of the proposed approach in this paper, the absorption and boundary conditions of different targets should be considered. In this paper, we used a solid sphere to measure ARF of the ultrahigh-frequency ultrasound. ARF for cell applications needs some corrections. The boundary condition between the gelatin medium and solid sphere can be approximated as a rigid boundary condition, where pressure doubling, a phase change, and nearly perfect reflection occur. On the other hand, the boundary condition for cells may be considered as a pressure release boundary condition, where no pressure doubling, no phase change, and less reflection occur. Due to these reasons, the measured displacements may be smaller than what we observed in this paper. To further fine-tune the proposed method, point targets with impedance similar to cells may be used for future study. FRET live cell imaging was performed to visualize calcium transport between cells after a target single cell was stimulated by the developed ultrasonic transducer. We conclude that the calcium ion may be intracellularly delivered by acoustic radiation force in the range of μN. Because the focal diameter of 130 MHz is approximately 10 μm, the estimated pressure was 0.54 MPa using ARF of 42.4 μN and used to intracellularly deliver calcium ions. This study may establish a gold standard to estimate ARF of ultrahigh-frequency transducers for single-cell applications.
Acknowledgments
This work was supported in part by the National Institutes of Health (NIH) under Grant No. GM120493, the National Science Foundation (NSF) under Grant No. CBET 1943852, 2019 AD&T Discovery Fund, and the Harper Cancer Research Institute, CCV grant.
DATA AVAILABILITY
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- 1. Kim M. G., Park J., Lim H. G., Yoon S., Lee C., Chang J. H., and Shung K. K., “ Label-free analysis of the characteristics of a single cell trapped by acoustic tweezers,” Sci. Rep. 7(1), 14092 (2017). 10.1038/s41598-017-14572-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Lee C., Lee J., Kim H. H., Teh S. Y., Lee A., Chung I. Y., Park J. Y., and Shung K. K., “ Microfluidic droplet sorting with a high frequency ultrasound beam,” Lab Chip 12(15), 2736–2742 (2012). 10.1039/c2lc21123h [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Yoon S., Pan Y., Shung K., and Wang Y., “ FRET-Based Ca(2+) biosensor single cell imaging interrogated by high-frequency ultrasound,” Sensors 20(17), 4998 (2020). 10.3390/s20174998 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Hwang J. Y., Yoon C. W., Lim H. G., Park J. M., Yoon S., Lee J., and Shung K. K., “ Acoustic tweezers for studying intracellular calcium signaling in SKBR-3 human breast cancer cells,” Ultrasonics 63, 94–101 (2015). 10.1016/j.ultras.2015.06.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Yoon S., Wang P., Peng Q., Wang Y., and Shung K. K., “ Acoustic-transfection for genomic manipulation of single-cells using high frequency ultrasound,” Sci. Rep. 7(1), 5275 (2017). 10.1038/s41598-017-05722-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Yoon S., Kim M. G., Chiu C. T., Hwang J. Y., Kim H. H., Wang Y., and Shung K. K., “ Direct and sustained intracellular delivery of exogenous molecules using acoustic-transfection with high frequency ultrasound,” Sci. Rep. 6, 20477 (2016). 10.1038/srep20477 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Hwang J. Y., Lim H. G., Yoon C. W., Lam K. H., Yoon S., Lee C., Chiu C. T., Kang B. J., Kim H. H., and Shung K. K., “ Non-contact high-frequency ultrasound microbeam stimulation for studying mechanotransduction in human umbilical vein endothelial cells,” Ultrasound Med. Biol. 40(9), 2172–2182 (2014). 10.1016/j.ultrasmedbio.2014.03.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Kalies S., Kuetemeyer K., and Heisterkamp A., “ Mechanisms of high-order photobleaching and its relationship to intracellular ablation,” Biomed. Opt. Express 2(4), 805–816 (2011). 10.1364/BOE.2.000805 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Le Harzic R., Riemann I., König K., Wüllner C., and Donitzky C., “ Influence of femtosecond laser pulse irradiation on the viability of cells at 1035, 517, and 345 nm,” J. Appl. Phys. 102(11), 114701 (2007). 10.1063/1.2818107 [DOI] [Google Scholar]
- 10. Cannata J. M., Ritter T. A., Chen W. H., Silverman R. H., and Shung K. K., “ Design of efficient, broadband single-element (20–80 MHz) ultrasonic transducers for medical imaging applications,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 50(11), 1548–1557 (2003). 10.1109/TUFFC.2003.1251138 [DOI] [PubMed] [Google Scholar]
- 11. Lam K. H., Ji H. F., Zheng F., Ren W., Zhou Q., and Shung K. K., “ Development of lead-free single-element ultrahigh frequency (170–320 MHz) ultrasonic transducers,” Ultrasonics 53(5), 1033–1038 (2013). 10.1016/j.ultras.2013.01.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Yoon S., Kim M. G., Williams J. A., Yoon C., Kang B. J., Cabrera-Munoz N., Shung K. K., and Kim H. H., “ Dual-element needle transducer for intravascular ultrasound imaging,” J. Med. Imaging 2(2), 027001 (2015). 10.1117/1.JMI.2.2.027001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Yoon S., Williams J., Kang B. J., Yoon C., Cabrera-Munoz N., Jeong J. S., Lee S. G., Shung K. K., and Kim H. H., “ Angled-focused 45 MHz PMN-PT single element transducer for intravascular ultrasound imaging,” Sens. Actuators, A 228, 16–22 (2015). 10.1016/j.sna.2015.02.037 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Lee J. and Shung K. K., “ Radiation forces exerted on arbitrarily located sphere by acoustic tweezer,” J. Acoust. Soc. Am. 120(2), 1084–1094 (2006). 10.1121/1.2216899 [DOI] [PubMed] [Google Scholar]
- 15. Nagle S. M., Sundar G., Schafer M. E., Harris G. R., Vaezy S., Gessert J. M., Howard S. M., Moore M. K., and Eaton R. M., “ Challenges and regulatory considerations in the acoustic measurement of high-frequency (>20 MHz) ultrasound,” J. Ultrasound Med. 32(11), 1897–1911 (2013). 10.7863/ultra.32.11.1897 [DOI] [PubMed] [Google Scholar]
- 16. Umchid S., Gopinath R., Srinivasan K., Lewin P. A., Daryoush A. S., Bansal L., and El-Sherif M., “ Development of calibration techniques for ultrasonic hydrophone probes in the frequency range from 1 to 100 MHz,” Ultrasonics 49(3), 306–311 (2009). 10.1016/j.ultras.2008.09.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Pan Y., Yoon S., Zhu L., and Wang Y., “ Acoustic mechanogenetics,” Curr. Opin. Biomed. Eng. 7, 64–70 (2018). 10.1016/j.cobme.2018.10.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Kim M. G., Yoon S., Kim H. H., and Shung K. K., “ Impedance matching network for high frequency ultrasonic transducer for cellular applications,” Ultrasonics 65, 258–267 (2016). 10.1016/j.ultras.2015.09.016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Kim M. G., Yoon S., Chiu C. T., and Shung K. K., “ Investigation of optimized treatment conditions for acoustic-transfection technique for intracellular delivery of macromolecules,” Ultrasound Med. Biol. 44(3), 622–634 (2018). 10.1016/j.ultrasmedbio.2017.11.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Pan Y., Yoon S., Sun J., Huang Z., Lee C., Allen M., Wu Y., Chang Y. J., Sadelain M., Shung K. K., Chien S., and Wang Y., “ Mechanogenetics for the remote and noninvasive control of cancer immunotherapy,” Proc. Natl. Acad. Sci. U. S. A. 115(5), 992–997 (2018). 10.1073/pnas.1714900115 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Karpiouk A. B., Aglyamov S. R., Ilinskii Y. A., Zabolotskaya E. A., and Emelianov S. Y., “ Assessment of shear modulus of tissue using ultrasound radiation force acting on a spherical acoustic inhomogeneity,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 56(11), 2380–2387 (2009). 10.1109/TUFFC.2009.1326 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Aglyamov S. R., Karpiouk A. B., Ilinskii Y. A., Zabolotskaya E. A., and Emelianov S. Y., “ Motion of a solid sphere in a viscoelastic medium in response to applied acoustic radiation force: Theoretical analysis and experimental verification,” J. Acoust. Soc. Am. 122(4), 1927–1936 (2007). 10.1121/1.2774754 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Ilinskii Y. A., Meegan G. D., Zabolotskaya E. A., and Emelianov S. Y., “ Gas bubble and solid sphere motion in elastic media in response to acoustic radiation force,” J. Acoust. Soc. Am. 117(4 Pt 1), 2338–2346 (2005). 10.1121/1.1863672 [DOI] [PubMed] [Google Scholar]
- 24. Aglyamov S., Karpiouk A., Mehrmohammadi M., Yoon S., Kim S., Ilinskii Y., Zabolotskaya E., and Emelianov S., “ Elasticity imaging and sensing using targeted motion: From macro to nano,” Curr. Med. Imaging 8(1), 3–15 (2012). 10.2174/157340512799220599 [DOI] [Google Scholar]
- 25. Yoon S., Aglyamov S., Karpiouk A., and Emelianov S., “ Correspondence: Spatial variations of viscoelastic properties of porcine vitreous humors,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 60(11), 2453–2460 (2013). 10.1109/TUFFC.2013.6644747 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Yoon S., Aglyamov S. R., Karpiouk A. B., Kim S., and Emelianov S. Y., “ Estimation of mechanical properties of a viscoelastic medium using a laser-induced microbubble interrogated by an acoustic radiation force,” J. Acoust. Soc. Am. 130(4), 2241–2248 (2011). 10.1121/1.3628344 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Yoon S., Aglyamov S., Karpiouk A., and Emelianov S., “ A high pulse repetition frequency ultrasound system for the ex vivo measurement of mechanical properties of crystalline lenses with laser-induced microbubbles interrogated by acoustic radiation force,” Phys. Med. Biol. 57(15), 4871–4884 (2012). 10.1088/0031-9155/57/15/4871 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Yoon S., Aglyamov S., Karpiouk A., and Emelianov S., “ The mechanical properties of ex vivo bovine and porcine crystalline lenses: Age-related changes and location-dependent variations,” Ultrasound. Med. Biol. 39(6), 1120–1127 (2013). 10.1016/j.ultrasmedbio.2012.12.010 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Lubinski M. A., Emelianov S. Y., and O'Donnell M., “ Speckle tracking methods for ultrasonic elasticity imaging using short-time correlation,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 46(1), 82–96 (1999). 10.1109/58.741427 [DOI] [PubMed] [Google Scholar]
- 30. Landau L. D. and Lifshitz E. M., Fluid Mechanics, 3rd ed. ( Pergamon Press, 1987). [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.