Skip to main content
Journal of Medical Internet Research logoLink to Journal of Medical Internet Research
. 2021 Apr 23;23(4):e19187. doi: 10.2196/19187

Web-Based Smartphone Algorithm for Calculating Blood Pressure From Photoplethysmography Remotely in a General Adult Population: Validation Study

Paul Holyoke 1,, Karthika Yogaratnam 1, Elizabeth Kalles 1
Editor: Gunther Eysenbach
Reviewed by: Cesar Calvo-Lobo, Keith Brunt, Tim McGough
PMCID: PMC8105754  PMID: 33890856

Abstract

Background

Outside of a clinical setting, oscillometric devices make remote monitoring of blood pressure and virtual care more convenient and feasible. HeartBeat Technologies Ltd developed a novel approach to measuring blood pressure remotely after an initial blood pressure reading by a nurse using the conventional measurement method. Using a finger pulse oximeter, a photoplethysmogram wave is transmitted by Bluetooth to a smartphone or tablet. A smartphone app (MediBeat) transmits the photoplethysmogram to a server for analysis by a proprietary algorithm—the person’s current blood pressure is sent back to the smartphone and to the individual’s health care provider.

Objective

This study sought to determine whether the HeartBeat algorithm calculates blood pressure as accurately as required by the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) for validation of blood pressure measuring devices.

Methods

ESH-IP2 requirements, modified to conform to a more recent international consensus statement, were followed. The ESH-IP2 establishes strict guidelines for the conduct and reporting of any validation of any device to measure blood pressure, including using the standard manual blood pressure instrument as a comparator and specific required accuracy levels for low, medium, and high ranges of blood pressure readings. The consensus statement requires a greater number of study participants for each of the blood pressure ranges. The validation of the accuracy of the algorithm was conducted with a Contec CMS50EW pulse oximeter and a Samsung Galaxy XCover 4 smartphone.

Results

The differences between the HeartBeat-calculated and the manually measured blood pressures of 62 study participants did not meet the ESH-IP2 standards for accuracy for either systolic or diastolic blood pressure measurements. There was no discernible pattern in the inaccuracies of the HeartBeat-calculated measurements.

Conclusions

The October 4, 2019 version of the HeartBeat algorithm, implemented in combination with the MediBeat app, a pulse oximeter, and an Android smartphone, was not sufficiently accurate for use in a general adult population.

Trial Registration

ClinicalTrials.gov NCT04082819; http://clinicaltrials.gov/ct2/show/NCT04082819

Keywords: blood pressure measurement, remote monitoring, hypertension

Introduction

Accurately assessing blood pressure is necessary for the proper diagnosis of hypertension, implementing treatments and monitoring whether those treatments are working [1,2]. To measure blood pressure accurately, standardized measurement techniques, calibrated equipment, and valid interpretation of readings are necessary [2-4]. Aneroid sphygmomanometers that are properly calibrated and maintained are considered to be at least as accurate as mercury-filled sphygmomanometers [5]. However, such instruments are not convenient for remote monitoring or virtual (remote) care because they can be difficult for a person to self-administer.

A novel approach to measuring blood pressure remotely has been developed (HeartBeat, HeartBeat Technologies Ltd). The measurement is taken after an initial blood pressure reading using the conventional measurement method, supplemented by specific characteristics of a person (age, gender, height, weight, and heart rate) to establish a baseline for the person. Unlike other devices [6-9], this approach uses a finger pulse oximeter that detects the changes in blood volume directly below the person’s skin and indirectly measures oxygen saturation in the blood. The resulting photoplethysmogram (PPG) wave is transmitted by Bluetooth technology to a smartphone or tablet. The MediBeat (HeartBeat Technologies Ltd) smartphone or tablet app then transmits the PPG wave to a server where a proprietary algorithm analyzes the baseline measurement for the person and the PPG waveform to calculate the person’s current blood pressure. The blood pressure measurement is then transmitted to the person’s smartphone, and if applicable, to the person’s health care provider’s devices.

The purpose of this study was to test whether the accuracy of the HeartBeat algorithm, available as of October 4, 2019 and for a general adult population (ages 25 years and older), was sufficient to seek regulatory approval when compared with a standard aneroid sphygmomanometer with stethoscope for measuring blood pressure.

Methods

Protocol

The European Society of Hypertension International Protocol revision (ESH-IP2) for the validation of blood pressure measuring devices in adults [10] was followed, with modifications to account for additional guidance from an international consensus statement [11]. ESH-IP2 [10] requires strict adherence to its method and reporting requirements, which are reflected in this report: to “ensure a uniform distribution of test pressures across a representative range,” 33 participants must be enrolled. Among the 33 participants, at least 10 must be male and 10 must be female; all should be at least 25 years of age; and between 10 and 12 participants must have blood pressure readings within each of the following recruitment ranges—systolic blood pressure: below 130 mmHg, between 130 and 180 mmHg, above 180 mmHg; diastolic blood pressure: below 80mmHg, between 80 and 130 mmHg, above 130 mmHg. All potential participants must be screened against these criteria and can be excluded if the gender criteria or age criteria are not met, or if including them in the study would exceed the maximum number of participants in any of the recruitment ranges.

An international consensus statement group [11] reviewed the adequacy of the number of participants in a variety of established protocols, including that of the ESH-IP2, and recommended that, to increase the power and accuracy of validation studies, at least 85 participants, rather than 33, should be included. To align with the consensus statement, the research team recalculated the ESH-IP2 requirement numbers to reflect the increased participant pool of 85. The recalculations are reflected in the tables and discussion in this report.

Devices

The pulse oximeter (Contec CMS50EW, Contec Medical Systems Co Ltd [12]) used to capture and transmit PPG is shown in Figure 1. The 8-bit Contec pulse oximeter is approved for use by Health Canada [13] and the US Food and Drug Administration [14]. The smartphone used in this study (upon which HeartBeat’s MediBeat app was installed and used to transmit the data to HeartBeat servers and to receive and display the results) was a Samsung Galaxy XCover 4 (Figure 2).

Figure 1.

Figure 1

Contec CMS50EW.

Figure 2.

Figure 2

Samsung Galaxy XCover 4.

Recruitment

Potential participants were recruited by mass email and offline posters at a single-site multistory office building in Markham, Ontario, Canada, where they worked; participants provided informed consent (Multimedia Appendix 1) and were scheduled for a measurement appointment during their working day by a member of the research team.

Procedure

Three registered nurses who had training and expertise in blood pressure measurement carried out the measurements required for this study between October 4, 2019 to November 22, 2019. One nurse was the supervisor, and two nurses were the observers throughout the duration of the study. The nurses were trained as per ESH-IP2 directives and were instructed on how to follow the protocol guidelines for data collection. Blood pressure measurements were recorded to the nearest 2 mmHg.

Two new sphygmomanometers (Prosphyg 775 Model, American Diagnostics Corporation [15]), whose components were checked against one another and calibrated before the study, were used as the reference instruments by the observers. With respect to reliability and validity, the ESH-IP2 embeds an ongoing reliability mechanism requiring that any individual participants’ blood pressure readings be excluded from the analysis if the difference in readings from the two observer nurses differ by 4 mmHg twice. The instruments were placed within one meter of the observers while they interacted with a participant, which allowed the observers to follow the instruments’ dials at eye level from 40 mmHg to 180 mmHg. Multiple sizes of bladders (cuffs) were available for the sphygmomanometers to ensure that 80% to 100% of each participant’s arm circumference could be encircled. The observers were also supplied with good quality nonelectronic stethoscopes with well-fitted earpieces.

The observers were instructed that, in the event that an abnormal blood pressure reading was detected (for example, if the blood pressure was outside clinically normal range for a specific person), they were to remeasure the blood pressure using conventional means to determine if the abnormal reading was accurate. If the abnormal blood pressure measurement was accurate, the nurse supervisor was to act according professional standards. The abnormal blood pressure measurement should also be recorded on ESH-IP2 Form 2.

Participants’ blood pressure measurements were taken in a room that was a comfortable temperature, without any noises or other influences that would have caused disturbances, such as telephones. The temperature in the room was the same as that throughout the building—20 °C to 24 °C. All participants participated in the study between the working hours of 9 AM and 4:30 PM. The pulse oximeter and the supervisor’s smartphone were charged and tested each day prior to interactions with participants. The MediBeat app was calibrated for each participant as follows: (1) Reference data for the individual (year of birth, height, weight, baseline manual blood pressure reading, and heart rate) were entered. (2) The participant rested for 10 minutes while seated comfortably with their legs uncrossed, back supported, and arm resting on the table at heart level. (3) The supervisor attached the finger pulse oximeter to the participant’s index finger (on the opposite arm to the manual blood pressure cuff) and waited for a few seconds until a stable signal was received and the device displayed both the participant’s pulse and SpO2 (oxygen saturation) values. (4) Once the values were displayed, the supervisor initiated a new session on the smartphone to connect the phone to the Contec device via Bluetooth. This connection started the 1-minute calibration process, at which point each participant was asked to remain still until the calibration had been completed successfully.

Overseen by an independent supervisor, measurements were recorded by the two observers who were blinded from both the other observer’s readings and from the reading calculated by the HeartBeat algorithm. Each participant’s blood pressure was measured as illustrated in Figure 3. Each systolic blood pressure measurement and each diastolic blood pressure measurement calculated by the HeartBeat algorithm was compared with the nearest of the previous and next observer systolic and diastolic measurements, respectively, and the difference between each HeartBeat-calculated measurement and the nearest observer measurement was calculated. Every such difference was then classified in one of these error groups for each systolic and diastolic measurements: 0-5 mmHg, 6-10 mmHg, 11-15 mmHg, and >15 mmHg. The results were then compared with the number of tolerable differences allowed by the protocol at both the level of individual measurements (part 1) and at the level of the participants (part 2). The device gets an overall pass (an acceptable level of accuracy) if it meets 4 pass criteria.

Figure 3.

Figure 3

Timeline of measurement of blood pressure for each participant.

Ethics

The study was reviewed and approved by the Southlake Regional Health Centre Research Ethics Board, Newmarket, Ontario, Canada (007-1920).

Results

A total of 105 individuals volunteered and were screened for the study, of whom 62 participants met the ESH-IP2 criteria, and 43 were excluded (Table 1; adapted version of the CONSORT flow diagram [16] in Figure 4). As required by the protocol, 5 participants out of the 105 were excluded because the differences in the measurements by the observer nurses exceeded 4 mmHg twice for the same participant, 34 were excluded participants were excluded because their blood pressure measurements fell within ranges for which the protocol quota had been fulfilled, and 4 participants were excluded because they were younger than 25 years of age. There was difficulty in recruiting hypertensive participants, that is, individuals with both systolic and diastolic blood pressures in the high ranges. This is reflected in the overall distribution (Figure 5 and Figure 6), in which most of the points fall below 125 mmHg and 90 mmHg, for systolic and diastolic blood pressures, respectively. By the time 105 participants had been recruited, only 2 met the criterion for high systolic blood pressure, and none met the criterion for diastolic blood pressure. The modified criteria would have required between 26 and 31 participants in the high blood pressure group to meet the protocol requirements. A calculation was performed to determine if it would be possible for the algorithm to achieve a pass if additional participants with high blood pressure were recruited, and there were a perfect match between their HeartBeat-calculated measurements and standard measurements. It was determined that such a result was not possible, and recruitment was discontinued. Therefore, the study concluded before the required number of participants was reached. The characteristics of participants are described in Tables 2 and 3.

Table 1.

Participants screened, excluded, and recruited for the study.

Reason for exclusion Excluded, n Total, n
Participants screened
105
Participants excluded
43

Ranges completea 34

Range adjustment 0

Arrhythmias 0

Device failure 0

Poor quality sounds 0

Cuff size unavailable 0

Observer disagreementb 5

Distributionc 0

Other reasonsc 4
Participants recruited
62

aIntake assessment placed them in ranges that had already been filled. The ESH-IP2 requires that the last participants to be recruited are the ones who are excluded.

bDiscrepancy >4 mmHg twice for the same participant; therefore, these participants were, in accordance with the ESH-IP2, excluded from the study.

cAge <25 years or older, as required by the ESH-IP2.

Figure 4.

Figure 4

Adapted version of the CONSORT flow diagram [16].

Figure 5.

Figure 5

Bland-Altman plot of the differences between the Heartbeat-calculated measurements and the manual measurements of systolic blood pressure.

Figure 6.

Figure 6

Bland-Altman plot of the differences between the Heartbeat-calculated measurements and the manual measurements of diastolic blood pressure.

Table 2.

Blood pressure and medication details for initially screened participants.

Category Blood pressure (mmHg), range Medicationa No medication Total
Systolic blood pressure




Low-low <90 0 0 0

Low 90-129 5 40 45

Medium 130-160 7 9 15

High 161-180 1 1 2

High-high >180 0 0 0
Diastolic blood pressure




Low-low <40 0 0 0

Low 40-79 3 30 33

Medium 80-100 10 19 29

High 101-130 0 0 0

High-high >130 0 0 0

aParticipants taking medication for hypertension.

Table 3.

Participant details.

Category Value
Sex, n

Male 13

Female 49
Age (years)

Mean (SD) 43.5 (11.2)

Range 25-66
Arm circumference (cm)

Mean (SD) 29.5 (4.2)

Range 23-40
Cuff size used, n

Small 0

Standard 61

Large 0

Missing data 1
Recruitment blood pressure (mmHg)

Systolic blood pressure


Mean (SD) 119.5 (18.9)


Range 90-180

Diastolic blood pressure


Mean (SD) 77.6 (10.6)


Range 57-110

Table 4 and Table 5 present the blood pressure readings by the international consensus statement’s predetermined levels. Table 4 shows the distribution of systolic and diastolic blood pressure readings across the recruited participants in the study. Table 5 shows the range of differences by observers 1 and 2.

Table 4.

Blood pressure readings (3 per participant) after the initial screening measurement.


Overall range (mmHg) (low-high) Lowa, n Mediumb, n Highc, n Maximum difference among highest and lowest totals in low, medium, and high ranges
Systolic blood pressure 90-180 150 30 6 144
Diastolic blood pressure 60-101 90 75 21 69

aSystolic ≤130 mmHg; diastolic ≤80 mmHg.

bSystolic 130 mmHg-160mmHg, diastolic 80 mmHg-100 mmHg.

cSystolic ≥160 mmHg, diastolic ≥100 mmHg.

Table 5.

Observer differences.


Range (low to high) (mmHg) Mean (SD) Repeated measurements
Systolic blood pressurea –3 to 3 0.0081 (0.4) b
Diastolic blood pressurea –4 to 3 0.0000 (0.6) b

aMeasurements are calculated by subtracting observer 1 measurements from observer 2 measurements.

bThere were no documented repeated measurements to calculate.

Table 6 shows the results in terms of each of the protocol requirements. According to the ESH-IP2 standards, for an acceptable level of accuracy, a device must pass all the protocol requirements in both parts 1 and 2. Note that in this study, the algorithm achieved a pass in only 1 category and a fail in 3 categories, for an overall fail.

Table 6.

Validation results.

Protocol requirements Range, n Pass or fail

<5 mmHg <10 mmHg <15 mmHg
Part 1 (N=186a)




Required (2 of 3) At least 137 At least 163 At least 180

Achieved differences





Systolic blood pressurea 120 147 169 Fail


Diastolic blood pressurea 159 180 186 Pass

Required (all) At least 122 At least 152 At least 175

Achieved differences





Systolic blood pressure 120 147 169


Diastolic blood pressure 159 180 186
Part 2 (N=62a)




Required At least 45

Fail

Achieved





Systolic blood pressure 43 b


Diastolic blood pressure 10

Allowed At most 6

Fail

Achieved





Systolic blood pressure 54


Diastolic blood pressure 5
Grade 3 (final result)—must pass all of parts 1 and 2


Fail

aThese reflect the number of readings and participants included in the study when recruitment was stopped.

bNot applicable.

Figure 5 shows the Bland-Altman plots [17,18] of the differences in systolic blood pressure measurements between the MediBeat device and the manual observer measurements (y-axis) and the average of the 2 measurements (x-axis). Figure 6 shows the plots for the differences in diastolic measurements. In both cases, the averages of the 3 MediBeat measurements were plotted against their absolute differences from the manual measurements for both systolic blood pressure (bias 3.89, LOA –11.98 to 19.77) and diastolic blood pressure (bias –0.28, LOA –9.66 to 9.09). Although values fall within the 95% confidence interval for both Bland-Altman plots, there is no statistically significant correlation between the HeartBeat-calculated blood pressure measurements and the manual measurements (systolic blood pressure: r=0.04, P=.53; diastolic blood pressure: r=.04, P=.54). This indicates that the HeartBeat-calculated blood pressure measurements are not sufficiently accurate. There was no discernible pattern in the inaccuracies in the HeartBeat-calculated measures; therefore, the source of the inaccuracies is not known.

Discussion

The Bland-Altman plots show that the October 4, 2019 version of the HeartBeat algorithm, used in conjunction with a Contec CMS50EW pulse oximeter, MediBeat app, and a Samsung Galaxy XCover 4 smartphone, was not sufficiently accurate to meet ESH-IP2 standards for measuring the blood pressure of adults in the general population; therefore, as per the protocol, it is not recommended for personal or clinical use in this configuration.

Acknowledgments

We would like to acknowledge the nurses who conducted the measurements for this study: Favour Onyido, RN, Laura Agborchi, RN, and Valerie Mgbatogu, RN. We would also like to acknowledge the contributions of Brianna Croft, MMI, and Mary Lou Ackerman, MBA, who helped with the logistics of this study, and of Grace Lui, MSc, who helped with the early phase of this study.

Abbreviations

ESH-IP2

European Society of Hypertension International Protocol revision 2010

LOA

limits of agreement

PPG

photoplethysmogram

SpO2

peripheral capillary oxygen saturation

Appendix

Multimedia Appendix 1

Information given during recruitment.

jmir_v23i4e19187_app1.pdf (406.2KB, pdf)

Footnotes

Authors' Contributions: PH led the design of the study. KY and EK collected and analyzed the data. All authors contributed to the interpretation of the results and reviewed and edited the manuscript. All authors read and approved the final version of the manuscript.

Conflicts of Interest: Two Contec pulse oximeters were supplied for the purposes of the study by HeartBeat Technologies Ltd. None of the authors have any association with, or have received any personal benefit from, HeartBeat Technologies Ltd.

References

  • 1.Lindsay P, Connor Gorber S, Joffres M, Birtwhistle R, McKay D, Cloutier L, Canadian Task Force on Preventive Health Care (CTFPHC) Recommendations on screening for high blood pressure in Canadian adults. Can Fam Physician. 2013 Sep;59(9):927–33, e393. http://www.cfp.ca/cgi/pmidlookup?view=long&pmid=24029505. [PMC free article] [PubMed] [Google Scholar]
  • 2.Kaczorowski J, Chambers LW, Dolovich L, Paterson JM, Karwalajtys T, Gierman T, Farrell B, McDonough B, Thabane L, Tu K, Zagorski B, Goeree R, Levitt CA, Hogg W, Laryea S, Carter MA, Cross D, Sabaldt RJ. Improving cardiovascular health at population level: 39 community cluster randomised trial of Cardiovascular Health Awareness Program (CHAP) BMJ. 2011 Feb 07;342:d442. doi: 10.1136/bmj.d442. http://europepmc.org/abstract/MED/21300712. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Nerenberg KA, Zarnke KB, Leung AA, Dasgupta K, Butalia S, McBrien K, Harris KC, Nakhla M, Cloutier L, Gelfer M, Lamarre-Cliche M, Milot A, Bolli P, Tremblay G, McLean D, Padwal RS, Tran KC, Grover S, Rabkin SW, Moe GW, Howlett JG, Lindsay P, Hill MD, Sharma M, Field T, Wein TH, Shoamanesh A, Dresser GK, Hamet P, Herman RJ, Burgess E, Gryn SE, Grégoire Jean C, Lewanczuk R, Poirier L, Campbell TS, Feldman RD, Lavoie KL, Tsuyuki RT, Honos G, Prebtani APH, Kline G, Schiffrin EL, Don-Wauchope A, Tobe SW, Gilbert RE, Leiter LA, Jones C, Woo V, Hegele RA, Selby P, Pipe A, McFarlane PA, Oh P, Gupta M, Bacon SL, Kaczorowski J, Trudeau L, Campbell NRC, Hiremath S, Roerecke M, Arcand J, Ruzicka M, Prasad GVR, Vallée Michel, Edwards C, Sivapalan P, Penner SB, Fournier A, Benoit G, Feber J, Dionne J, Magee LA, Logan AG, Côté Anne-Marie, Rey E, Firoz T, Kuyper LM, Gabor JY, Townsend RR, Rabi DM, Daskalopoulou SS, Hypertension Canada Hypertension Canada's 2018 guidelines for diagnosis, risk assessment, prevention, and treatment of hypertension in adults and children. Can J Cardiol. 2018 May;34(5):506–525. doi: 10.1016/j.cjca.2018.02.022. [DOI] [PubMed] [Google Scholar]
  • 4.McAlister F A, Straus S E. Evidence based treatment of hypertension. Measurement of blood pressure: an evidence based review. BMJ. 2001 Apr 14;322(7291):908–11. doi: 10.1136/bmj.322.7291.908. http://europepmc.org/abstract/MED/11302909. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Mercury versus nonmercury sphygmomanometers: a review of the comparative clinical effectiveness and guidelines. Canadian Agency for Drugs and Technologies in Health. 2012. [2018-06-25]. https://www.cadth.ca/mercury-versus-non-mercury-sphygmomanometers-review-comparative-clinical-effectiveness-and.
  • 6.Mena LJ, Félix Vanessa G, Ostos R, González Armando J, Martínez-Peláez Rafael, Melgarejo JD, Maestre GE. Mobile personal health care system for noninvasive, pervasive, and continuous blood pressure monitoring: development and usability study. JMIR Mhealth Uhealth. 2020 Jul 20;8(7):e18012. doi: 10.2196/18012. https://mhealth.jmir.org/2020/7/e18012/ [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Plante TB, Urrea B, MacFarlane ZT, Blumenthal RS, Miller ER, Appel LJ, Martin SS. Validation of the instant blood pressure smartphone app. JAMA Intern Med. 2016 May 01;176(5):700–2. doi: 10.1001/jamainternmed.2016.0157. http://europepmc.org/abstract/MED/26938174. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Plante TB, O'Kelly AC, Urrea B, Macfarlane ZT, Appel LJ, Miller III ER, Blumenthal RS, Martin SS. Auralife instant blood pressure app in measuring resting heart rate: validation study. JMIR Biomed Eng. 2018 Nov 21;3(1):e11057. doi: 10.2196/11057. [DOI] [Google Scholar]
  • 9.Mazoteras-Pardo V, Becerro-De-Bengoa-Vallejo R, Losa-Iglesias ME, López-López Daniel, Palomo-López Patricia, Rodríguez-Sanz David, Calvo-Lobo C. Validation in the general population of the iHealth track blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol Revision 2010: descriptive investigation. JMIR Mhealth Uhealth. 2019 Mar 19;7(3):e13137. doi: 10.2196/13137. https://mhealth.jmir.org/2019/3/e13137/ [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y, Wang J, Mengden T, Shennan A, Working Group on Blood Pressure Monitoring of the European Society of Hypertension European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010 Feb;15(1):23–38. doi: 10.1097/MBP.0b013e3283360e98. [DOI] [PubMed] [Google Scholar]
  • 11.Stergiou G, Alpert B, Mieke S, Asmar R, Atkins N, Eckert Siegfried, Frick Gerhard, Friedman Bruce, Graßl Thomas, Ichikawa Tsutomu, Ioannidis John P, Lacy Peter, McManus Richard, Murray Alan, Myers Martin, Palatini Paolo, Parati Gianfranco, Quinn David, Sarkis Josh, Shennan Andrew, Usuda Takashi, Wang Jiguang, Wu Colin O, O'Brien Eoin. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) collaboration statement. J Hypertens. 2018 Mar;36(3):472–478. doi: 10.1097/HJH.0000000000001634. http://europepmc.org/abstract/MED/29384983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.CMS50EW Pulse Oximeter. Contec Medical Systems Co Ltd. 2020. [2021-04-15]. https://www.contecmed.com/productinfo/602619.html.
  • 13.Licence no.: 90629. Health Canada. [2021-03-22]. https://health-products.canada.ca/mdall-limh/information.do?companyId_idCompanie=130273&lang=eng.
  • 14.K141362. U.S. Food and Drug Administration. [2021-03-22]. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K141362.
  • 15.K962655. U.S. Food and Drug Administration. [2021-03-22]. https://www.accessdata.fda.gov/SCRIPTS/cdrh/devicesatfda/index.cfm?db=pmn&id=K962655.
  • 16.Schulz Kenneth F, Altman Douglas G, Moher David, CONSORT Group CONSORT 2010 statement: updated guidelines for reporting parallel group randomised trials. PLoS Med. 2010 Mar 24;7(3):e1000251. doi: 10.1371/journal.pmed.1000251. https://dx.plos.org/10.1371/journal.pmed.1000251. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Mar 08;1(8476):307–10. [PubMed] [Google Scholar]
  • 18.Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999 Jun;8(2):135–60. doi: 10.1177/096228029900800204. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Multimedia Appendix 1

Information given during recruitment.

jmir_v23i4e19187_app1.pdf (406.2KB, pdf)

Articles from Journal of Medical Internet Research are provided here courtesy of JMIR Publications Inc.

RESOURCES