Table 2.

Positive and negative values obtained by qRT-PCR and sensitivity, specificity, and accuracy of RAPID 1.0 using NP/OP and saliva samples

RAPID (NP/OP) qRT-PCR
Sensitivity Specificity Accuracy
Positive (n = 109a) Negative (n = 30) Total (n = 139)
Positive 91 0 91 91/109 (83.5%) 121/139 (87.1%)
Negative 18 30 48 30/30 (100%)
RAPID (saliva) qRT-PCR
Sensitivity Specificity Accuracy
Positive (n = 13) Negative (n = 37) Total (n = 50)
Positive 13 5 18 13/13 (100.0%) 45/50 (90.0%)
Negative 0 32 32 32/37 (86.5%)
a

Clinical sample set includes a highly contagious SARS-CoV-2 UK variant B.1.1.7 from a patient.