Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2022 Jun 1.
Published in final edited form as: Magn Reson Imaging. 2021 Mar 20;79:103–111. doi: 10.1016/j.mri.2021.03.013

Improved Nerve Conspicuity with Water-Weighting and Denoising in Two-Point Dixon Magnetic Resonance Neurography

Ek T Tan 1,*, Sophie C Queler 1, Bin Lin 1, Yoshimi Endo 1, Alissa J Burge 1, Julia Sternberg 1, Hollis G Potter 1, Darryl B Sneag 1
PMCID: PMC8107136  NIHMSID: NIHMS1686922  PMID: 33753136

Abstract

Background:

T2-weighted, two-point Dixon fast-spin-echo (FSE) is an effective technique for magnetic resonance neurography (MRN) that can provide quantitative assessment of muscle denervation. Low signal-to-noise ratio and inadequate fat suppression, however, can impede accurate interpretation.

Purpose:

To quantify effects of principal component analysis (PCA) denoising on tissue signal intensities and fat fraction (FF) and to determine qualitative image quality improvements from both denoising and water-weighting (WW) algorithms to improve nerve conspicuity and fat suppression.

Study Type:

Prospective.

Subjects:

Twenty-one subjects undergoing MR neurography evaluation (11/10 male/female, mean age=46.3+/−13.7 years) with 60 image volumes. Twelve subjects (23 image volumes) were determined to have muscle denervation based on diffusely elevated T2 signal intensity.

Field Strength/Sequence:

3T, 2D, two-point Dixon FSE.

Assessment:

Qualitative assessment included overall image quality, nerve conspicuity, fat suppression, pulsation and ringing artifacts by 3 radiologists separately on a three-point scale (1=poor, 2=average, 3=excellent). Quantitative measurements for FF and signal intensity relative to normal muscle were made for nerve, abnormal muscle and subcutaneous fat.

Statistical Tests:

Linear and ordinal regression models were used for quantitative and qualitative comparisons, respectively; 95% confidence intervals (CIs) and p-values for pairwise comparisons were adjusted using the Holm-Bonferroni method. Inter-rater agreement was assessed using Gwet’s agreement coefficient (AC2).

Results:

Simulations showed PCA-denoising reduced FF error from 2.0% to 1.0%, and from 7.6% to 3.1% at noise levels of 10% and 30%, respectively. In human subjects, PCA-denoising did not change signal levels and FF quantitatively. WW decreased fat signal significantly (−83.6%, p<0.001). Nerve conspicuity was improved by WW (odds ratio, OR=5.8, p<0.001). Fat suppression was improved by both PCA (OR=3.6, p<0.001) and WW (OR=2.2, p<0.001). Overall image quality was improved by PCA+WW (OR=1.7, p=0.04).

Conclusions:

WW and PCA-denoising improved nerve conspicuity and fat suppression in MR neurography. Denoising can potentially provide improved accuracy of FF maps for assessing fat-infiltrated muscle.

Keywords: magnetic resonance neurography, Dixon, principal component analysis, quantitative magnetic resonance imaging, denoising, fat quantification

1. Introduction

Peripheral nerve MRI (MR neurography, MRN) is effectively performed with T2-weighted imaging[1] using a fast spin echo (FSE) readout to exploit the contrast between the higher T2 of nerves and the lower T2 of surrounding soft tissue. T2-weighted imaging also highlights muscle edema-like patterns in actively denervated muscle[2], which may be secondary to extracellular fluid accumulation from increased muscle perfusion and capillary permeability [3]. T2-weighted FSE requires effective fat suppression, as fat has a high T2 value and therefore appears hyperintense relative to other soft tissues. Fat suppression is especially important in MRN, where hyperintense perineural fat obscures adjacent, small nerves. Although inversion recovery[4] and chemical saturation[5] are used routinely for fat suppression in musculoskeletal imaging[6], Dixon-based FSE provides high signal-to-noise ratio (SNR) relative to inversion recovery-based techniques and superior robustness to B0-inhomogeneity relative to chemical fat saturation [7]. Dixon-based FSE[8,9] is therefore particularly effective for fat suppression in MRN [10], including of the brachial plexus[11].

In Dixon-based fat suppression, images are acquired at different echo times (TE) that have different phases between spins at “water” resonance and “fat” resonance. “Water” and “fat” images are resolved by either subtracting echoes with opposite phase shifts or by solving a system of equations from three or more echoes acquired at varying phase shifts[12]. Prior to solving these equations, Dixon-processing requires removal of the confounding phase due to B0-inhomogeneity for which several effective methods, including phase-fitting[13] and region-growing[14], have become routinely used. In Dixon-based FSE, each additional echo increases scan time proportionately to maintain the same echo train length (ETL) [9]. MRN requires high spatial resolution (~0.3–0.5 mm in-plane) to visualize small nerves and their fascicular architecture, and to delineate these from the surrounding soft tissues; consequently, scan times can be long (4–7 minutes). Since two-point or two-echo Dixon is effective at providing spatially uniform fat suppression, Dixon T2-weighted FSE is often acquired with just two echoes to minimize scan time.

Dixon-based T2-weighted FSE currently has several limitations. First, the contrast between fat and nerves can be sub-optimal due to the wide spectrum of frequency shift and T2 in fatty tissue[15]. Incomplete fat suppression can result in reduced conspicuity of small nerves in regions directly adjacent to perineural fat. Neither adding echoes nor changing TE improves fat suppression due to the need for contrast against both low-T2 muscle and high-T2 fat. Second, the need for high spatial resolution in MRN often results in sub-optimal SNR, which motivates performing image denoising. A promising denoising approach utilizes principal components analysis (PCA) across multiple images by reorganizing multi-echo data as matrices and attributing noise to the smaller principal components[16]. While PCA-denoising may be readily applied to T2-mapping by reorganizing matrices in the echo- and spatial-dimensions[17], having only two echoes in MRN makes it practically infeasible to perform PCA with a matrix size of 2. However, if data from multi-channel phased-array coils are considered, the matrix dimensionality of Dixon-FSE can be increased to make it suitable for PCA-denoising.

An additional motivation for denoising is fat fraction (FF) quantification, which requires high SNR[18]. During the chronic phase of muscle denervation, the presence and extent of fat infiltration may be assessed qualitatively from the “water” image without relying on the “fat” image. However, quantitative estimation of the MRI-based fat content[18], while different from actual fat content[19], may help characterize the extent and stage of muscle denervation or muscle function loss[20]. While FF quantification is more efficiently performed using gradient recalled echo readout[12], the ability to quantify FF using an FSE readout[21] that can be performed in combination with T2-mapping[22,23] would add to the utility of this MRN pulse sequence.

This study’s objective was to improve key image quality metrics in two-point Dixon FSE MRN by exploiting the “fat” image. We hypothesized that the proposed water-weighting and PCA-denoising would improve overall image quality, nerve conspicuity and fat suppression in two-point Dixon FSE MRN.

2. Material and Methods

Fig. 1a shows the overall processing steps for the PCA-denoising and water-weighting methods introduced in this work. These steps are described in more detail below.

Figure 1.

Figure 1.

Schematics of methodology, including (a) a flowchart of processing steps in this work, including principal components analysis (PCA)-denoising and water-weighting processing, for generating water (W), fat (F) and fat fraction (FF) images; (b) organization of image kernel data into matrix S for PCA.

2.1. Denoising with Principal Components Analysis

In PCA-denoising[16], the image data is first organized into an M-by-N matrix S and undergoes eigenvalue decomposition into its eigenvector matrices U and V, and eigenvalue matrix Λ, which is a diagonal matrix of values with scalar values Λi,i sorted in decreasing order where i ∈ [1,N]. When M=N:

S=VT,Λ=[Λ1,100ΛN,N]. (1)

The denoised eigenvalue matrix Λ′ is obtained by nulling all Λi,i except the P largest eigenvalues, where 1 ≤ P ≤ {M, N}. When MN, singular value decomposition may instead be used without loss of generality. The denoised image data S′ is obtained with S′ = UΛ′VT. PCA-denoising may be applied ascribing the number of echoes to one dimension (M), and pixels to the other (N) as suggested previously[17]. However, the choice of these dimensions could be sub-optimal for denoising two-point Dixon data as the rank of Λ is low at 2 (M=2).

To increase dimensions and rank of S (and Λ), we propose utilizing the channel dimension from multi-channel phased-array coils, similar to that proposed elsewhere [24]. This requires denoising to be performed just prior to complex coil combination, which means the data would be complex, containing real and imaginary components. In addition, the image can be separated into real/imaginary channels to double the dimensionality. In this work, the first dimension (M) consisted of echoes, the image-kernel (of 3×3 to 7×7 pixels), concatenated by real/imaginary channels; the second dimension (N) was for coil channels. The matrix organization is illustrated in Fig. 1b. For a 3×3 kernel, two real/imaginary channels and two echoes for two-point Dixon, and 32-channel phased-array coil, one obtains MxN=36×32, which would have sufficiently high matrix rank.

2.2. Water Weighting (WW)

To exploit the fat image for further suppression of fat signal in the water image, we propose water-weighting (WW) processing following PCA-denoising. WW exploits the weighting from FF, which is defined by the ratio between fat (F) and water (W) signals, FF=F/(W+F) and is sensitive to noise due to normalization. With PCA-denoising, the noise in FF will be reduced, which incidentally also benefits WW processing. To apply WW, we propose a simple weighting of:

WWW=W(1FF)=W2(W+F). (2)

The rationale for this weighting is to further suppress fat signal while maintaining the contrast of most tissue in the WW in a similar range as the original water image. Signal intensities in image regions with high FF will be significantly attenuated, while those with low FF will be minimally attenuated, creating increased contrast of water signal. It is expected that nerve conspicuity will be improved, as fat suppression will be improved. The contrast of non-denervated and denervated but non-fat-infiltrated muscle (low FF) would also be expected to increase compared to fat-infiltrated denervated muscle, but this was not evaluated in this work.

2.4. Simulations - PCA

To determine the effects from PCA-denoising, signal simulations were performed at different levels of noise. The signals were processed using mostly the same steps as the in vivo MRI data (Fig. 1a). The simulated signals were generated using coil-sensitivities ρc obtained from 400 representative pixels acquired from a 16-channel phased-array coil on human subject data. The impact of additive Gaussian noise, η was tested at 7 different noise levels ranging from 0.1% to 100%. To exclude possible effects on the phase from PCA-denoising that could impact Dixon-processing, the B0 differences between the in-phase (sin) and out-of-phase signals (sout) were simulated with a small amount of B0 inhomogeneity, leading to phase accrual θ (−30° to +30 °) between the in-phase (sin) and out-of-phase (sout) signals from the c-th coil:

sin,c=ρc(W+F)+η,and (3)
sout,c=ρceiθ(W+F)+η. (4)

The signals were combined in the coil-dimension using optimal reconstruction[25] to obtain the coil-combined signals sin and sout. To perform B0-correction, the measured phase accrual, ϕ=(s¯outsin), was used to correct the out-of-phase signal under the premise that ϕ ≈ −θ if there were no deleterious effects from PCA:

sout,c=eiϕsout,c. (5)

The simulated water and fat signals were then obtained using the standard Dixon-based addition and subtraction of sin and s’out and sum-of-squares coil combination. These steps were repeated without and with PCA-denoising and at different PCA thresholds, P to determine the optimum threshold.

2.5. MR Imaging

For in vivo imaging, 21 subjects (11 male/10 female; age 46.3+/−13.7 years old) undergoing MRN of varying anatomies were prospectively included in this IRB-approved study. The clinical indications for MRI included suspected brachial plexopathy (n= 5), Parsonage-Turner syndrome (n=4), pain and weakness following motor vehicle or bicycle accident (n=3), lower lumbosacral plexopathy or proximal sciatic neuropathy (n=2), concern for mass (n=2), finger pain, paresthesia or weakness (n=2), foot drop (n= 1), and pudendal neuralgia (n=1). 2D, two-point-Dixon, T2-weighted FSE images, acquired as part of standard-of-care, used parameters that varied depending on the imaged anatomy (TR/TE=3077–7520/78.7–88.5 ms, FOV=10–36 cm, matrix=320×224, 30–100 slices, slice thickness=1.2–4.0 mm (no slice gap), ETL=11–18, bandwidth-per-pixel=244–391 Hz). Acquisitions were performed at 3T MRI (Signa Premier, and MR750, GE Healthcare, Waukesha, WI, USA) using either one or two 16-channel flexible coils or a 32-channel torso coil, depending on the anatomy imaged. Raw data was reconstructed offline using in-house code developed on Matlab (Mathworks Inc., Natick, USA); image reconstruction was performed without and with PCA. A 3×3 kernel was used in PCA (Fig. 1b), with real and imaginary components separated for PCA matrix sizes of either MxN=36×16 or 36×32 depending on the coils used. The optimum number of eigenvalues being retained, P, was determined from the simulations. Phase-fitting was applied for B0-correction. The same P was used for the entire image, so as not to introduce any spatially varying artifacts. WW-processing was applied to the PCA images. The processing times for PCA and water-weighting for each image series were 26–58 minutes and <1 minute, respectively (on a 2.8GHz Pentium IV 16GB RAM workstation).

2.6. Hypotheses and Statistical Analysis

We hypothesized that PCA-denoising would reduce FF in normal muscle and increase FF in subcutaneous fat, without significantly changing water and fat signals in nerve, abnormal muscle, or subcutaneous fat ROIs. To test this hypothesis, the mean values of water signal, fat signal, and FF from the standard vs. PCA images for each ROI were measured. In order to facilitate averaging of signal intensities (water and fat) across subjects, the signals for nerve, subcutaneous fat, and abnormal muscle ROIs were normalized to that of the normal muscle ROI. As FF is a normalized quantity, normalization of FF was not performed. The FF was adjusted for relaxation by assuming a normal muscle T2* of 21.6 ms. We hypothesized that WW would improve fat suppression by reducing fat signal without reducing water signal. The water and fat signals from the PCA and WW+PCA images were also measured and normalized to normal muscle ROIs. For quantitative outcomes, marginal linear regression models estimated with small sample bias-corrected empirical standard errors were specified to evaluate for mean differences between image type by ROI [26]. Subjects were treated as the repeated factor to account for within-patient correlations between image type and ROIs.

We also hypothesized that, qualitatively, PCA-denoising and WW would improve overall image quality, nerve conspicuity, and fat suppression without increasing pulsation and ringing artifacts. With reduced noise and a new intensity-weighting scheme, it was conceivable that such artifacts common to MRN could become more pronounced. Three radiologists with 7 to 10 years of dedicated musculoskeletal MR experience (YE, AJB, DBS) graded the water images from the Dixon-FSE acquisition. These were assessed for overall image quality, nerve conspicuity (i.e. ability to confidently visualize the nerve, separate from the surrounding soft tissue), fat suppression (1=poor, 2=average, 3=excellent), and pulsation and ringing artifacts (1=none, 2=mild, 3=severe). For each qualitative outcome, an ordinal logistic regression model with random subject effects was specified to compare standard, PCA, and WW+PCA images, accounting for the three readers. Odds ratios (OR) were estimated from the three pairwise image comparisons (PCA vs standard, WW+PCA vs PCA, WW+PCA vs Standard) with 95% confidence intervals (CIs) and p-values subsequently adjusted using the Holm-Bonferroni method. Statistical significance was set at p<0.05. To assess interrater agreement of the readers’ assessments, ordinal weighted Gwet’s agreement coefficients (AC2) were estimated, with AC2 ≥ 0.7 considered substantial. All statistical analyses were performed using SAS v9.4 (SAS Institute, Cary, NC, USA).

The ROIs were manually drawn by one radiologist with 7 years of musculoskeletal subspecialty experience (DBS) on the uncorrected water image – one nerve, subcutaneous fat, normal muscle (expected signal intensity), and abnormal muscle (diffusely increased T2 signal intensity). The qualitative diagnosis of ‘normal’ versus ‘abnormal’ muscle was agreed upon independently by all three readers.

2.7. Muscle and nerve ROIs

Table 1 provides a summary of the locations of normal and abnormal muscle ROIs. Muscle abnormality was determined qualitatively by the presence of diffuse signal hyperintensity of the muscle on the T2-weighted Dixon water image (without denoising). In total, 60 image volumes covering a broad range of anatomy (27 brachial plexus, 15 upper-arm/elbow, 18 pelvis/lower extremities) were evaluated in 21 subjects. All subjects had normal muscle as determined on MRI, with normal muscle ROIs drawn in 59 image volumes (1 image volume was excluded due to excessive motion). Twelve of the 21 subjects had abnormal muscle determined on MRI, with abnormal muscle ROIs drawn in 23 image volumes. There were no cases of fatty infiltrated muscle. EMGs were available in 14 of the 21 subjects. EMG demonstrated abnormal motor unit recruitment in 16 of the 23 abnormal muscles; the remaining 7 abnormal muscles were not tested on EMG.

Table 1.

Patient demographics and body parts imaged and analyzed

Age (Years) Mean SD Minimum Maximum
46.3 13.7 19 65
Sex Normal Muscle Abnormal Muscle
Subjects Image
Series
Subjects Image
Series
Male 3 30 8 17
Female 6 29 4 6
Total: 9 59 12 23
Body Part by MRI Study Normal Muscle Image Abnormal Muscle Image
Description Series Series
Brachial Plexus (Total): 15 -
Nerves: C5 nerve root 1 -
C6 nerve root 3 -
C7 nerve root 2 -
Anterior division of upper trunk 1 -
Posterior division of upper trunk 2 -
Medial cord 1 -
Posterior cord 5 -
Trunk/shoulder girdle/neck (Total): 26 16
Muscles: Supraspinatus 2 8
Infraspinatus 2 2
Subscapularis 0 1
Teres minor 1 0
Trapezius 10 1
Deltoid 5 3
Anterior Scalene 1 0
Middle Scalene 4 0
Posterior paraspinals 0 1
Pectoralis major 1 0
Upper Extremities Nerves (Total): 12 -
Nerves: Axillary 1 -
Ulnar 7 -
Median 3 -
Radial 1 -
Upper Extremities Muscles (Total): 15 6
Muscles: Biceps 4 0
Brachialis 2 1
Pronator Teres 2 0
Flexor Carpi Ulnaris 2 2
Flexor Carpi Radialis 1 0
Flexor Digitorum Profundus 2 1
Pronator Quadratus 1 0
Flexor Pollicis Longus 0 2
1st Dorsal Interosseous 1 0
Lumbar Plexus (Total): 18 -
Nerves: L5 nerve root 3 -
S1 nerve root 4 -
S2 nerve root 1 -
Sciatic nerve 10 -
Lower Extremities Muscles (Total): 18 1
Muscles: Gluteus maximus 10 0
Gluteus medius 2 0
Gluteus minimus 3 0
Piriformis 1 0
Obturator internus 1 0
Vastus medialis 1 0
Long head bicep femoris 0 1

In total, 45 nerve ROIs were drawn, undifferentiated by the presence of a nerve abnormality. Fifteen nerve ROIs could not be drawn reliably because the nerves were either too small or the acquired images were not orthogonal to the longitudinal course of the nerve.

3. Results

3.1. Simulations

Fig. 2ab show that the error in FF (measured FF minus true FF) was low at low noise and was not improved significantly with PCA-denoising with P=1. Increased noise levels (Fig. 2c) increased FF error, which showed positive bias at low FF and negative bias at high FF. The FF error was reduced significantly with PCA-denoising (Fig. 2d). Also, there were no increased errors due to B0 inhomogeneity, either with or without PCA-denoising.

Figure 2.

Figure 2.

Simulation results of normalized mean error in computed fat fraction (FF) as a function of the simulated FF and off-resonance. At low noise of (a) 3% (b) PCA (P=1) did not result in significant improvement to FF error. When noise was increased to (c) 30%, bias in FF increased, with positive bias at low FF and negative bias at high FF. Applying (d) PCA-denoising to the 30% noise case reduced the bias to levels similar to that in 3%. The level of bias was invariant to the degree of off-resonance.

Fig. 3 summarizes the simulation results for different noise levels and values of P. At noise levels of 1% and below, the root-mean-square error (RMSE) was less than 0.1%. The error increased with increased noise levels. PCA-denoising with P=1 resulted in the lowest error at all noise levels compared to higher P and with no PCA-denoising. The largest improvement in error was seen with the highest noise level of 100% (RMSE reduced from 20.2% to 12.6%). However, when normalized to its non-PCA-denoised RMSE, the RMSE reduction was more substantial at noise levels of 30% (RMSE reduced from 7.6% to 3.1%) and 10% (RMSE reduced from 2.0% to 1.0%).

Figure 3.

Figure 3.

Summary of root-mean-square error (RMSE) of fat fraction results simulated at various noise levels and without (none) and with PCA-denoising at different threshold P levels.

3.2. Quantitative Analysis

Using the previous simulation results, the lowest P=1 was used to process the in vivo images. PCA-denoising did not reduce the relative nerve, subcutaneous fat, and abnormal muscle signal (Table 2). PCA-denoising also did not significantly alter FF, but for subcutaneous fat the average change was positive (+0.1%) and for other ROIs the average change was negative (−0.9 to −0.3%). WW decreased the relative signal intensity of subcutaneous fat significantly (−83.6%, p<0.001), without significantly altering the relative signal intensities of nerve (−4.0%, p=0.66) and abnormal muscle (+6.0%, p=0.62).

Table 2.

Summary of quantitative results. The results indicate mean percentage changes in water image signal (normalized to normal muscle signal) and fat fraction (FF) from application of principal components analysis (PCA)-denoising and water-weighting (WW) in nerve, normal muscle, subcutaneous fat, and abnormal muscle. WW was applied on PCA images.

Change in Water Image Signal Intensity (% Relative to Normal Muscle Signal) Change in FF (%)
Quantitative PCA vs Standard WW+PCA vs PCA PCA vs Standard
Mean (95% CI) p-value −2.2 (−19.0, 14.5) −4.0 (−21.8, 13.8) −0.3 (−2.2, 1.6)
Nerve p-value 0.796 0.657 0.751
Mean (95% CI) - - −0.9 ( −2.7, 0.8)
Normal Muscle p-value - - 0.301
Mean (95% CI) −2.1 (−16.6, 12.3) −83.6 (−100.3, −66.8) 0.1 (−1.5, 1.8)
Subcutaneous Fat p-value 0.770 <.001* 0.864
Mean (95% CI) −0.3 (−23.7, 23.0) 6.0 (−17.9, 30) −0.3 (−2.9, 2.4)
Abnormal Muscle p-value 0.977 0.621 0.845

3.3. Qualitative Analysis

Fig. 4 shows the distribution of qualitative results averaged across all readers. The fraction of scans rated as excellent for overall image quality was increased by PCA (from 19.0% to 21.0%), and further by WW (to 30.3%). For nerve conspicuity, scans rated as excellent were increased by PCA (from 25.7% to 27.0%) and further by WW (to 67.0%). For fat suppression, scans rated as excellent were increased by PCA (from 22.0% to 53.7%) and further by WW (to 69.7%). Artifact scores were mostly unchanged.

Figure 4.

Figure 4.

Distribution of qualitative scoring, showing averaged fraction of responses from all readers with ratings (1=poor, 2=average, 3=excellent) for (a) overall image quality, (b) nerve conspicuity, and (c) fat suppression, and artifact ratings (1=none, 2=mild, 3=severe) for (d) pulsatility and (e) ringing.

Table 3 summarizes results of the paired comparisons. PCA resulted in significantly improved fat suppression by an OR=3.6 (p<0.001); the improvements in overall image quality (OR=1.08) and nerve conspicuity (OR=1.05) were, however, not statistically significant (p=0.72 to 0.81). WW resulted in significantly improved nerve conspicuity by OR=5.7 (p<0.001), and fat suppression by OR=2.2 (p<0.001). WW with PCA resulted in improved overall image quality by OR=1.7 (p=0.04), along with improved nerve conspicuity by OR=6.1 (p<0.001) and fat suppression by OR=8.1 (p<0.001). Pulsatility and ringing artifacts were unchanged by PCA with or without WW.

Table 3.

Summary of qualitative results comparing standard images vs. principal components analysis (PCA) denoising, vs. water-weighted (WW) images. WW was applied on PCA-denoised images. Results of the odds ratio (OR) were based on a 3-point assessment of image quality (IQ) or artifacts.

Qualitative PCA vs Standard WW vs PCA WW vs Standard
Odds Ratio (95% CI) 1.08 (0.65,1.78) 1.56 (0.94,2.57) 1.68 (1.02, 2.76)
Overall IQ p-value 0.724 0.067 0.040*
Odds Ratio (95% CI) 1.05 (0.64, 1.73) 5.77 (3.41, 9.76) 6.08 (3.59, 10.28)
Nerve Conspicuity p-value 0.805 <.001* <.001*
Odds Ratio (95% CI) 3.62 (2.17, 6.02) 2.24 (1.28, 3.9) 8.09 (4.65, 14.08)
Fat Suppression p-value <.001* <.001* <.001*
Odds Ratio (95% CI) p-value 1 (0.54, 1.86) 1.04 (0.56, 1.94) 1.04 (0.56, 1.94)
Pulsatility Artifact p-value 0.999 0.999 0.999
Odds Ratio (95% CI) 1 (0.61, 1.64) 1.18 (0.72, 1.95) 1.18 (0.72, 1.94)
Ringing Artifact p-value 0.999 0.999 0.999

Interpretation:

OR > 1 for A vs B implies superior scores in A compared to B

OR = 1 for A vs B implies no difference in scores between A and B

OR < 1 for A vs B implies superior scores in B compared to A

Table 4 summarizes the inter-rater agreement between the three readers. There was substantial agreement (≥0.7) in 10 out of 15 of the assessed image metrics. The range of the agreement coefficient was between 0.59 and 0.89.

Table 4.

Inter-rater agreement of qualitative measures assessed using ordinal weighted Gwet’s agreement coefficient (standard error in parentheses), where a coefficient greater than or equal to 0.7 was considered substantial agreement.

Standard PCA WW+PCA
Overall Image Quality 0.77 ( 0.04) 0.74 ( 0.04) 0.69 ( 0.05)
Nerve Conspicuity 0.62 ( 0.06) 0.59 ( 0.06) 0.76 ( 0.04)
Fat Suppression 0.73 ( 0.06) 0.62 ( 0.06) 0.77 ( 0.05)
Pulsatility Artifact 0.89 ( 0.03) 0.89 ( 0.03) 0.63 ( 0.08)
Ringing Artifact 0.76 ( 0.04) 0.76 ( 0.04) 0.75 ( 0.04)

Fig. 5 shows water images and FF maps in a patient with polyneuropathy of unknown etiology and denervated muscle. PCA improved fat suppression and conspicuously decreased FF in the musculature, especially in regions that were more central. These regions corresponded to reduced noise in the PCA+WW images. The nerve conspicuity was unaltered by PCA. The image contrasts for denervated vs. normal muscle were similar, but the subcutaneous fat signal was significantly suppressed.

Figure 5.

Figure 5.

Images without PCA (Std, top row) and with PCA-denoising (bottom row) from a 65-year-old man with denervation edema pattern of multiple forearm flexor and extensor muscles, showing (from left to right) water, fat, fat fraction (FF) and water-weighted (WW) images. PCA provided similar levels of proximal forearm ulnar nerve conspicuity (yellow arrows) and abnormal muscle edema (red dashed arrows). PCA improved homogeneity of subcutaneous fat suppression (green arrows) relative to the adjacent muscle. PCA also reduced the FF levels and noisy appearance of FF in the normal muscle (blue dashed arrows). In the WW images, the regions where FF were reduced in normal muscle corresponded to improved muscle signal homogeneity as well.

Fig. 6 compares standard water images, PCA water images, and WW images from two different subjects and anatomies. The WW images show significantly improved subcutaneous fat signal in both subjects (Fig. 6d, f), as well as suppression of perineural fat (Fig. 6d, f, insets), which improved the conspicuity of the sciatic (Fig. 6d) and radial nerves (Fig. 6f). The contrast of the denervated muscle vs. normal muscle was altered neither by PCA nor WW.

Figure 6.

Figure 6.

Comparisons in (a-c) pelvic MR neurography in a 41-year-old man with right-sided sciatica but normal appearance of the sciatic nerve. The right sciatic nerve was not conspicuous with (a) standard and (b) PCA (yellow inset and yellow arrows), but was conspicuous on the (c) water-weighted (WW) PCA image. WW markedly improved suppression of subcutaneous fat signal (white arrow). Aliasing and subcutaneous fat (green arrow) were less conspicuous with PCA and WW. Similar comparison of (d-f) elbow MR neurography in a 65-year-old man with mild denervation edema pattern of multiple forearm flexor and extensor muscles (red arrows). The branches of the radial nerve were more conspicuous on (f) WW than on (d) standard and (e) PCA only (yellow inset and yellow arrows).

4. Discussion

WW qualitatively and quantitatively improved fat suppression and qualitatively improved nerve conspicuity in T2-weighted, two-point Dixon FSE MRN. PCA-denoising demonstrated quantitative reduction in FF error in simulations, with significantly improved fat suppression qualitatively. PCA-denoising did not alter image signal significantly, suggesting no strong bias was introduced to any tissue contrast; the improvement in qualitative fat suppression might in part be due to the denoised muscle signal. Together, WW and PCA-denoising significantly improved the overall image quality of MRN.

The in vivo results from PCA-denoising were consistent with simulations between 3–10% noise level, i.e. resulting in about −1% FF in normal muscle. While not statistically significant, PCA-denoising increased FF in subcutaneous fat and decreased FF in other tissues, consistent with the expected physiology; variation from the wide sampling parameters and range of anatomical regions may have contributed to a non-statistically significant result. There was also variation in effects from motion, especially respiratory-induced motion about the brachial plexus, despite the use of respiratory-gating[27].

The denoising and WW processing techniques may also be applicable to other Dixon-based acquisitions. For example, Dixon-based fat suppression has been applied in reversed fast imaging steady-state free precession (PSIF) in MRN [28] and dual-echo steady-state (DESS) in musculoskeletal imaging[29]. Denoising may also be applicable to quantitative T2-mapping sequences that use the Dixon method for separating water and fat components [22].

4.1. Limitations

In this study, the choice of two-point Dixon imaging was based on protocols optimized for high in-plane spatial resolution and scan times commonly used in clinical practice. This necessarily limited the potential size of the PCA matrix and capability for denoising that generally increases with the matrix size. For example, matrix size can be increased by 50% with a three-point Dixon technique and doubled with four-point Dixon, albeit at the cost of scan time. Furthermore, more echoes could be useful for solving for T2* simultaneously[12] and could provide a more accurate B0-map for reducing FF calculation bias[18]. Also, we did not utilize the Marchenko-Pastur theory previously proposed for determining the optimum threshold[16]; instead, a uniform threshold was chosen to provide spatial uniformity in the extent of denoising. The lowest possible threshold of P=1was found to be optimal, which suggests a high level of informational redundancy in coil-sensitivity and is likely a reasonable assumption for small kernel sizes. In this work, we did not evaluate WW without denoising, which did not allow evaluation of interaction between PCA-denoising and WW. We speculate that WW would also provide improved nerve conspicuity and fat suppression in non-denoised images, given it works complementarily to PCA-denoising. Given the significant contrast improvement from WW processing observed in this study, we anticipate that WW effects would dominate those of PCA-denoising. Another limitation was that this study did not include normal subject controls; muscle was deemed to be ‘normal’ by qualitative assessment by the radiologists rather than by electrophysiological or physical exam testing.

In this work, we focused on improvements in visualization of nerve conspicuity and potential improvements in FF accuracy. However, we did not evaluate potential improvements in visualizing edema patterns in denervated muscle due to the smaller subset of denervated muscle. In addition, we did not evaluate the potential for using FF to quantify fat infiltration in denervated muscle. In this cohort, only 11 subjects were diagnosed with abnormal muscle and none of these muscles demonstrated fat infiltration qualitatively, as seen in states of chronic denervation. We speculate that PCA-denoising and WW together would provide improved depiction of denervated muscle, especially in discerning patterns of fat infiltration due to the increased fat suppression. Given the simplicity of WW as compared to PCA, WW would be more feasibly implemented in real-time. Future work could include comparing Dixon FSE-based FF measurements with gradient-echo-based methods, as well as correlating Dixon FSE-based FF measurements with electrophysiology results in a larger cohort with varying stages and severities of muscle denervation.

5. Conclusions

This study demonstrated that PCA-denoising and water-weighting improved nerve conspicuity and fat suppression on T2-weighted, Dixon FSE MRN sequences. PCA-denoising also improved quantitative accuracy of FF maps, which might be useful in assessing chronic muscle denervation.

Acknowledgements

The authors will like to acknowledge technical assistance from Maggie Fung, Jaemin Shin, and Shiv Kaushik.

Funding

The project described was supported by Grant Number 1R21TR003033-01A1 from the National Institutes of Health (NIH). Its contents are solely the responsibility of authors and do not necessarily represent the official views of the NIH. This project also received research support from GE Healthcare.

Footnotes

Declarations of Interest

None

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • [1].Filler AG, Maravilla KR, Tsuruda JS. MR neurography and muscle MR imaging for image diagnosis of disorders affecting the peripheral nerves and musculature. Neurol Clin 2004;22:643–682. doi: 10.1016/j.ncl.2004.03.005. [DOI] [PubMed] [Google Scholar]
  • [2].Kim S-J, Hong SH, Jun WS, Choi J-Y, Myung JS, Jacobson JA, et al. MR Imaging Mapping of Skeletal Muscle Denervation in Entrapment and Compressive Neuropathies. Radiographics 2011;31:319–32. doi: 10.1148/rg.312105122. [DOI] [PubMed] [Google Scholar]
  • [3].Polak JF, Jolesz FA, Adams DF. Magnetic Resonance Imaging of Skeletal Muscle Prolongation of T1 and T2 Subsequent to Denervation. Invest Radiol 1988;23:365–9. doi: 10.1097/00004424-198805000-00007. [DOI] [PubMed] [Google Scholar]
  • [4].Bydder GM, Young IR. MR Imaging: Clinical Use of the Inversion Recovery Sequence. J Comput Assist Tomo 1985;9:659–75. doi: 10.1097/00004728-198507010-00002. [DOI] [PubMed] [Google Scholar]
  • [5].Haase A, Frahm J, Hanicke W, Matthaei D. 1 H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985;30:341–4. doi: 10.1088/0031-9155/30/4/008. [DOI] [PubMed] [Google Scholar]
  • [6].Grande FD, Santini F, Herzka DA, Aro MR, Dean CW, Gold GE, et al. Fat-suppression techniques for 3-T MR imaging of the musculoskeletal system. Radiographics Rev Publ Radiological Soc North Am Inc 2014;34:217–33. doi: 10.1148/rg.341135130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Rybicki FJ, Chung T, Reid J, Jaramillo D, Mulkern RV, Ma J. Fast Three-Point Dixon MR Imaging Using Low-Resolution Images for Phase Correction: A Comparison with Chemical Shift Selective Fat Suppression for Pediatric Musculoskeletal Imaging. Am J Roentgenol 2001;177:1019–23. doi: 10.2214/ajr.177.5.1771019. [DOI] [PubMed] [Google Scholar]
  • [8].Ma J, Son JB, Bankson JA, Stafford RJ, Choi H, Ragan D. A fast spin echo two-point Dixon technique and its combination with sensitivity encoding for efficient T2-weighted imaging. Magn Reson Imaging 2005;23:977–82. doi: 10.1016/j.mri.2005.10.005. [DOI] [PubMed] [Google Scholar]
  • [9].Hardy PA, Hinks RS, Tkach JA. Separation of fat and water in fast spin-echo MR imaging with the three-point dixon technique. J Magn Reson Imaging 1995;5:181–5. doi: 10.1002/jmri.1880050213. [DOI] [PubMed] [Google Scholar]
  • [10].Kalia V, Leung D, Sneag D, Grande F, Carrino J. Advanced MRI Techniques for Muscle Imaging. Semin Musculoskel R 2017;21:459–69. doi: 10.1055/s-0037-1604007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Tomura N, Saginoya T, Kokubun M, Tsuchiya M, Yamamoto T. T2-weighted IDEAL fast spin echo imaging of the brachial plexus: comparison with STIR. Acta Radiol 2015;56:1242–7. doi: 10.1177/0284185114552292. [DOI] [PubMed] [Google Scholar]
  • [12].Reeder SB, Wen Z, Yu H, Pineda AR, Gold GE, Markl M, et al. Multicoil Dixon chemical species separation with an iterative least-squares estimation method. Magnet Reson Med 2004;51:35–45. doi: 10.1002/mrm.10675. [DOI] [PubMed] [Google Scholar]
  • [13].Chen Q, Schneider E, Aghazadeh B, Weinhous MS, Humm J, Ballon D. An automated iterative algorithm for water and fat decomposition in three-point Dixon magnetic resonance imaging. Med Phys 1999;26:2341–7. doi: 10.1118/1.598748. [DOI] [PubMed] [Google Scholar]
  • [14].Szumowski J, Coshow WR, Li F, Quinn SF. Phase unwrapping in the three-point Dixon method for fat suppression MR imaging. Radiology 1994;192:555–61. doi: 10.1148/radiology.192.2.8029431. [DOI] [PubMed] [Google Scholar]
  • [15].Azzabou N, Sousa PL de, Caldas E, Carlier PG. Validation of a generic approach to muscle water T2 determination at 3T in fat-infiltrated skeletal muscle. J Magn Reson Imaging 2015;41:645–53. doi: 10.1002/jmri.24613. [DOI] [PubMed] [Google Scholar]
  • [16].Veraart J, Novikov DS, Christiaens D, Ades-aron B, Sijbers J, Fieremans E. Denoising of diffusion MRI using random matrix theory. Neuroimage 2016;142:394–406. doi: 10.1016/j.neuroimage.2016.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Bydder M, Du J. Noise reduction in multiple-echo data sets using singular value decomposition. Magn Reson Imaging 2006;24:849–56. doi: 10.1016/j.mri.2006.03.006. [DOI] [PubMed] [Google Scholar]
  • [18].Liu C, McKenzie CA, Yu H, Brittain JH, Reeder SB. Fat quantification with IDEAL gradient echo imaging: Correction of bias from T1 and noise. Magnet Reson Med 2007;58:354–64. doi: 10.1002/mrm.21301. [DOI] [PubMed] [Google Scholar]
  • [19].Reeder SB, Hu HH, Sirlin CB. Proton density fat-fraction: A standardized mr-based biomarker of tissue fat concentration. J Magn Reson Imaging 2012;36:1011–4. doi: 10.1002/jmri.23741. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Karampinos DC, Holwein C, Buchmann S, Baum T, Ruschke S, Gersing AS, et al. Proton Density Fat-Fraction of Rotator Cuff Muscles Is Associated With Isometric Strength 10 Years After Rotator Cuff Repair: A Quantitative Magnetic Resonance Imaging Study of the Shoulder. Am J Sports Medicine 2017;45:1990–9. doi: 10.1177/0363546517703086. [DOI] [PubMed] [Google Scholar]
  • [21].Reeder SB, Pineda AR, Wen Z, Shimakawa A, Yu H, Brittain JH, et al. Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): Application with fast spin-echo imaging. Magnet Reson Med 2005;54:636–44. doi: 10.1002/mrm.20624. [DOI] [PubMed] [Google Scholar]
  • [22].Janiczek RL, Gambarota G, Sinclair CDJ, Yousry TA, Thornton JS, Golay X, et al. Simultaneous T2 and lipid quantitation using IDEAL-CPMG. Magnet Reson Med 2011;66:1293–302. doi: 10.1002/mrm.22916. [DOI] [PubMed] [Google Scholar]
  • [23].Mankodi A, Bishop CA, Auh S, Newbould RD, Fischbeck KH, Janiczek RL. Quantifying disease activity in fatty-infiltrated skeletal muscle by IDEAL-CPMG in Duchenne muscular dystrophy. Neuromuscular Disord 2016;26:650–8. doi: 10.1016/j.nmd.2016.07.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Lemberskiy G, Baete S, Veraart J, Shepherd T, Fieremans E, Novikov DS. Achieving sub-mm clinical diffusion MRI resolution by removing noise during reconstruction using random matrix theory. Proc ISMRM 2019:770. [Google Scholar]
  • [25].Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magnet Reson Med 1990;16:192–225. doi: 10.1002/mrm.1910160203. [DOI] [PubMed] [Google Scholar]
  • [26].Morel JG, Bokossa MC, Neerchal NK. Small Sample Correction for the Variance of GEE Estimators. Biometrical J 2003;45:395–409. doi: 10.1002/bimj.200390021. [DOI] [Google Scholar]
  • [27].Sneag DB, Mendapara P, Zhu JC, Lee SC, Lin B, Curlin J, et al. Prospective respiratory triggering improves high-resolution brachial plexus MRI quality. J Magnetic Reson Imaging Jmri 2018;49:1723–9. doi: 10.1002/jmri.26559. [DOI] [PubMed] [Google Scholar]
  • [28].Yoon D, Biswal S, Rutt B, Lutz A, Hargreaves B. Feasibility of 7T MRI for imaging fascicular structures of peripheral nerves. Muscle Nerve 2018;57:494–8. doi: 10.1002/mus.26035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Chaudhari AS, Sveinsson B, Moran CJ, McWalter EJ, Johnson EM, Zhang T, et al. Imaging and T2 relaxometry of short-T2 connective tissues in the knee using ultrashort echo-time double-echo steady-state (UTEDESS). Magnet Reson Med 2017;78:2136–48. doi: 10.1002/mrm.26577. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES