Regional Differences in Diffusion Tensor Imaging Measurements: Assessment of Intrarater and Interrater Variability

A Ozturk a, AD Sasson a, JAD Farrell a,b, BA Landman c, ACBS da Motta a, A Aralasmak d, DM Yousem a
PMCID: PMC8118842  PMID: 18356471

Abstract

BACKGROUND AND PURPOSE: Diffusion tensor imaging (DTI) has become a valuable tool in both the research and clinical evaluation of subjects. We sought to quantify interobserver and intraobserver variability of diffusivity and diffusion anisotropy measurements with regard to specific regions of interest (ROIs).

MATERIALS AND METHODS: The subject group consisted of 5 healthy control subjects and 7 study subjects (all males; 16–19 years old; mean age = 17.5 years), as part of a protocol for closed head injury. Two whole-brain DTI scans were acquired on a 3T scanner for each subject. Analysis was performed using a ROI approach. Two independent observers analyzed the apparent diffusion coefficient (ADC) and fractional anisotropy (FA) indices in the corpus callosum, cortical spinal tract, internal capsules (ICs), basal ganglia, and centrum semiovale (CSO). Intraobserver and interobserver variability were calculated for the mean ADC, FA, and ordered eigenvalues of the diffusion tensor (λ1, λ2, and λ3).

RESULTS: The overall κ statistic for intraobserver variability for both observers showed slight-to-substantial agreement (κ = 0.02–0.69), however FA values in the CSO showed only slight agreement. Interobserver agreement was also slight to substantial for these DTI measurements with high variability in FA values in the IC and CSO.

CONCLUSIONS: When one is comparing 2 DTI measurements, it is important to assess intraobserver and interobserver variability. We recommend caution in the analysis of DTI contrasts in the IC and CSO, because we have found the widest range of variability in measurements within these structures.


Diffusion tensor imaging (DTI) has become well established as a research tool to investigate water diffusion properties in the central nervous system and is making inroads into clinical imaging. It is a safe, noninvasive in vivo method that allows a superior assessment (compared with conventional MR imaging) of white matter tracts by reconstructing their 3D shape and connectivity. Parameters derived from DTI can provide information about tissue organization, degree of myelination, and water mobility, enabling the study of white matter tract direction, integrity, and damage in the brain.1 Although DTI was initially used for anatomic purposes to understand human brain anatomy and to topographically depict the white matter tracts of the brain, the technique has increasingly been used to study changes in pathology by comparing quantifiable metrics of diffusion.

Two important parameters derived from DTI are the apparent diffusion coefficient (ADC) and fractional anisotropy (FA).1 The ADC and FA parameters characterize the average amount of diffusion and the diffusion anisotropy, respectively.2 These parameters are derived from the eigenvalues (λ1, λ2, and λ3) computed from the diffusion tensor. The diffusivity measurements from the eigenvalues themselves can also be used to study tissue properties.3,4 Recent reviews5,6 outline the methodology and clinical applications of DTI.

In clinical practice, these parameters can be used for comparison between individual patients, for serial examinations in the same patient, and for the evaluation of maturation during childhood. The quantification of diffusion can be especially helpful, because it may allow earlier diagnosis of the presence and extent of pathology. Pathologic changes due to damage in the central nervous system start at the microstructural level. Previous studies have shown that changes in diffusion characteristics (ie, diffusivity and diffusion anisotropy) can be detected in stroke7 and multiple sclerosis8 before abnormalities can be detected with conventional MR imaging.7,9,10

The assessment of the variability of DTI measurements, both when the data are analyzed repeatedly by one reviewer or by different reviewers, is an essential and important step in evaluating the clinical utility of DTI and its strength as a quantitative measurement. We sought to provide measurements of the intrarater and interrater variability for DTI measurements and to assess the degree of difference that can be detected comparing 2 DTI studies.

Materials and Methods

Subjects

The subjects were recruited as part of an ongoing study looking at closed-head injury in athletes performing contact sports. This study was approved by the institutional review board, and written informed consent was obtained from all of the adult subjects and from the parents of the minors before data acquisition. The subjects were all male and ranged in age from 16 to 19 years (mean age = 17.5 years). Five healthy control subjects with no history of neurologic disease and 7 research subjects participated in this study. All of the subjects were also evaluated for trauma-related changes in MR imaging by one observer (D.M.Y). There were no abnormal findings in MR images of subjects.

MR Imaging

All of the subjects were scanned using a 3T MR imaging scanner (Philips Medical Systems, Best, the Netherlands) capable of 60-mT/m magnetic field gradients, using the body coil for excitation and an 8-channel phased-array sensitivity-encoding coil for reception. Each DTI dataset was acquired with the following protocol. A multisection, single-shot echo-planar imaging, spin-echo sequence (TR/TE = 7897/84 ms; FOV = 212 × 212 mm) was used to acquire 60 transverse sections with no section gap and 2.2-mm nominal isotropic resolution (acquired matrix = 96 × 96, reconstructed to 256 × 256). Diffusion weighting was applied in 32 noncollinear directions with a b- value of 700 s/mm2. A volume minimal diffusion weighting (b = 0 s/mm2) was also acquired. Two DTI datasets were acquired for each subject, and the total scan time for DTI was approximately 10 minutes.

Image Processing

DTI data were processed off-line using the Coregistration, Adjustment and Tensor-solving, a Nicely Automated Program (CATNAP; http://iacl.ece.jhu.edu/∼bennett/catnap/, Johns Hopkins University School of Medicine), an in-house–designed data processing pipeline.4 CATNAP performs motion-correction with the Oxford Center for Functional Magnetic Resonance Imaging Linear Image Registration Tool (Oxford, United Kingdom), computes diffusion-weighted gradient tables adjusted for section angulation, and calculates the diffusion tensor, as well as the diffusivity (ADC, eigenvalues) and diffusion anisotropy (FA) metrics. Motion correction was performed by using a 6-degree of freedom rigid body model, and the 2 DTI datasets were concatenated and processed simultaneously in 1 diffusion tensor calculation. Image visualization and region of interest (ROI) placement were then performed using DTIStudio (Johns Hopkins University, Baltimore, Md).11

ROI Placement

The processed DTI contrasts were analyzed separately and independently by 2 observers (A.O. and A.D.S) who had been trained on the DTIStudio software package. To estimate the intraobserver and interobserver reliability of ROI-based DTI parameters, ROI placement was carried out by both observers on 2 separate occasions, 4–12 weeks apart, without the use of a template. The information about the first ROI placement was not available during the second assessment, and the investigator was blinded to the other observer's evaluation. We selected 9 different ROIs in locations with different FA. The locations, which are commonly used in many clinical DTI studies, were easily visualized and delineated on DTI color maps. Predetermined circular single ROIs were manually placed on color maps at the following anatomic locations: corticospinal tract at the level of the pons, middle cerebellar peduncles (MCP), anterior limb of the internal capsule, posterior limb of the internal capsule, genu of the internal capsule, centrum semiovale (CSO), thalamus (Th), putamen (Pt), splenium, and genu of the corpus callosum (CCG). With the exception of the corpus callosum, all of the ROIs were positioned bilaterally. The ROIs were then propagated onto the DTI contrast images (ie, FA, ADC, etc), and the mean and SD over all voxels in the ROI were calculated. Statistical analysis (intrarater and interrater comparisons) were performed for the mean ADC, FA, λ1, λ2, and λ3.

Data Analysis

To compare first and second measurements of 1 observer (intraobserver reliability) and to compare the first and the second measurements of 2 independent investigators (interobserver reliability), κ statistics was used. (The κ value of 0.11–0.20 is considered as “slight,” 0.21–0.40 as “fair,” 0.41–0.60 as “moderate,” 0.61–0.80 as “substantial,” and 0.81–1.00 as “almost perfect” agreement.)12 Finally, to determine the degree of differences between measurements as a percentage, we calculated the difference between 2 measurements divided by the mean of these 2 measurements for both evaluators for all of the ROIs.

Results

Intrarater and Interrater Agreement

The overall κ statistic for intraobserver variability for both observers showed slight to substantial agreement (κ = 0.02–0.69) for FA, ADC, and eigenvalues (On-line Table 1); however, FA values in the CSO showed only slight agreement (κ < 0.20). The interobserver agreement was also slight to substantial for these DTI measurements (κ = 0.02–0.69; On-line Table 1) with high variability in FA values in the internal capsule and CSO. Eigenvalues also showed slight agreement in internal capsules, CSO, and MCPs for interobserver agreement. λ1 was more reliable than the λ2 and λ3 (On-line Table 2).

Regional Distribution of Measurements

When the individual ROIs were analyzed to determine whether the reliability was lower in some location than others, the overall span of variability of interobserver readings showed the lowest agreement within the internal capsule and CSO, whereas the splenium and CCG showed the highest agreement (On-line Tables 1 and 2). FA measurements are least reliable in the CSO and gray matter ROIs (On-line Table 2). The percentage of variability (as measured by the difference between the 2 readings divided by the mean of the 2 readings) was obtained for all of the DTI measures for each reader (On-line Table 2).

Discussion

The Potential Clinical Use of DTI Measurements

The potential use of DTI assessments in clinical practice is encouraging. ADC and FA measurements are frequently used as potential biomarkers of the degree of tissue injury in brain diseases. Some authors have used DTI in the assessment of brain disorders and have shown abnormal hemispheric fiber connections in acquired disease or congenital abnormalities.13

In a similar vein, the degree of white matter disruption due to MS or diffuse axonal injury may be quantitatively assessed with DTI metrics, as opposed to merely depicting the tracts as smaller in size or finer in quality. Although the interpretation of diffusion changes measured by DTI is not straightforward, measures of the severity of demyelination and axonal damage would be of great clinical relevance.8,1419 Ptak et al20 have used DTI indices to propose a cerebral FA score to serve as an index of white matter injury due to trauma that successfully correlates with outcome and predictor variables.

DTI is a valuable tool to assess the impact of neoplasms on the white matter tracts.21 Another widespread white matter disorder, such as adrenoleukodystrophy, could be assessed quantitatively with DTI as the dietary therapy is implemented. In this way, an improvement from baseline can be demonstrated eloquently in a quantitative matter.22,23 Other clinical applications include the use of DTI to investigate the affects of small vessel ischemic disease on white matter integrity, determining thresholds by which the patient becomes symptomatic. More severe white matter disease, such as cerebral autosomal dominant arteriopathy with subcortical infarcts and leukoencephalopathy or Binswanger disease, which both tend to affect the white matter more so than the gray matter,24 could also be investigated. Neurophysiologic disorders have also been investigated with DTI, and some researchers propose that errant white matter connectivity may be the intrinsic problem in schizophrenia, autism, or dyslexia.25 From the clinical perspective, it is hoped that the sensitivity of DTI contrasts to changes in microstructure may permit the detection of recovery (as a function of time or medical treatment), which could be correlated with standard clinical outcome measures.

Variability in DTI Measurements

The acceptable range of variability of a quantifiable index in medicine cannot be arbitrarily set. Nonetheless, it is valuable to measure this variability if one hopes to be able to make a claim of abnormality when a value deviates from “the norm.” DTI will increasingly be used to access pathology in the brain, and, therefore, knowing the interobserver and intraobserver variability is critical. Calculating DTI indices requires an intimate knowledge of the anatomy of the brain so that the ROIs can be placed appropriately. The variability from one observer to the other largely lies in the placement of the ROIs and, when tractography is performed, is also influenced by the subjective determination of the fiber tracking termination criteria (lower bound FA and turning angle thresholds). The variability produced by ROI placement is due, in part, to the neuroanatomic training of the observer, as well as the selection of the proper location for analysis as visualized on DTI contrasts.

According to our results, measurements of FA were most reliable in the CCG and least reliable in the Pt, Th, and CSO. For ADC, the percentage of variability was highest in Th and MCP and lowest in CSO and CCG. Reliability can be affected by a number of factors, including scanner performance, initial signal intensity-to-noise and acquisition resolution, positioning, segmentation, alignment, warping, and resectioning.26 Pfefferbaum et al26 were the first to address reproducibility of FA and ADC images in detail. Many studies have compared measured ADC and FA values in healthy children and adults on 1.5T MR imaging units.4,10,2729 Characterizing regional variation in measurement error for this method is important for the understanding of the results of group comparisons and longitudinal studies. This is particularly true in the case of some diseases where differences from controls may be quite subtle.30 The possible explanation of higher variability in the internal capsule and CSO may be due to section shifts of the ROI location leading to addition or subtraction of a group of pixels within the section which would have a large effect on the κ value.

The pattern of high variability in the internal capsule and in other white matter structures with different FA might indicate a combination of effects due to noise, partial volume effects, and complex fiber architectures within a pixel, which could easily vary from one section to another section. Image noise produces errors in the calculated tensor and, hence, in its eigenvalues and eigenvectors. Our results showed that λ1 was more reliable than λ2 and λ3 in several regions, such as the corpus callosum and CSO and corticospinal tract at the pons level. Overall percentage of variability was higher for λ3. Random variations in these quantities complicate the analysis and interpretation of DTI experiments. It is known that, in anisotropic systems, the expectation value of the largest eigenvalue is overestimated, and the lowest eigenvalue is underestimated.31,32

ADC values of the healthy brain are very similar in gray matter and white matter; however, FA shows distinct mean values between these structures. This may explain inaccuracies of FA values due to partial volume effects and ROI outlines of fiber tracts close to their borders.33 Overall, a lower percentage of variability of our ADC measurements is consistent with other studies reported in the literature (On-line Table 2).34,35

Rater performance is also an important source of variability and is one of the limiting factors in the detection of variance in both cross-sectional and longitudinal studies. Standardized rater training is increasingly used to improve the quality of the investigated outcome parameters.35 It is possible that, while 2 observers may have equivalent training for anatomic localization and use of software packages for DTI analysis, delineation of ROIs without the assistance of a template or reliance on single ROIs that may be subject to section shift errors may introduce sources of error and contribute to data variability. In a similar fashion to methods used in fiber tracking, the confidence level of validity could be improved by the use of anatomic limitations using multiple ROIs.36

One of the purposes of this article was to look at the degree of variability and to determine at what level a difference in DTI values would be significant. If one uses 2 SDs of variability as a marker for 95% accuracy corresponding to a typical p value of .05, it would suggest that identifying a difference more than twice that of the variability would be reasonable to compare a single subject versus a population norm. For example, given our variability of 2.6% for FA of the CCG, a difference of more than twice that, or 5.2% from either baseline or from a matched control would be a reasonable standard to use. For each DTI parameter, for each location, there are different thresholds. Overall, if one reviews on-line Table 2, these threshold values would range from 6.6% for the ADC of the CSO to as high as 25.0% for the ADC of the MCP. For comparative group studies, the empirical variability for each region may guide appropriate power analyses.

The limitations of this study include the relatively small number of subjects evaluated. One would expect nonetheless that the degree of variability would not change significantly with increasing numbers of subjects. The results herein reflect that of a 3T protocol, which may not apply to the more widely used 1.5T magnets. It would also be of benefit to have more than 2 observers for every location and to have evaluated additional locations with additional ROIs, which may be used for various analyses.

DTI measurements are sensitive to differences in hardware, acquisition parameters, analysis software, and data processing strategies. We suggest that standardizing and using schemes of ROIs should allow reduction of interobserver and intraobserver variability. The variability is greatest when slight differences in ROI placement can have a large effect on measured values. It would be of value for all researchers reporting DTI-based measurements of diffusivity and diffusion anisotropy to provide their interobserver and intraobserver variability so that the validity of their measurements can be better assessed.

Supplementary Material

A0998.pdf
A0998.pdf (90.6KB, pdf)

Footnotes

Indicates articles with supplemental on-line tables.

References

  • 1.Basser PJ. Inferring microstructural features and the physiological state of tissues from diffusion-weighted images. NMR Biomed 1995;8:333–44 [DOI] [PubMed] [Google Scholar]
  • 2.Sigmund EE, Song YQ. Multiple echo diffusion tensor acquisition technique. Magn Reson Imaging 2006;24:7–18 [DOI] [PubMed] [Google Scholar]
  • 3.Ni H, Kavcic V, Zhu T, et al. Effects of number of diffusion gradient directions on derived diffusion tensor imaging indices in human brain. AJNR Am J Neuroradiol 2006;27:1776–81 [PMC free article] [PubMed] [Google Scholar]
  • 4.Landman BA, Farrell JA, Jones CK, et al. Effects of diffusion weighting schemes on the reproducibility of diffusion tensor imaging-derived fractional anisotropy, mean diffusivity, and principal eigenvector measurements at 1.5T. J Magn Reson Imaging 2007;26:756–67 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Horsfield MA, Jones DK. Applications of diffusion-weighted and diffusion tensor MRI to white matter diseases—a review. NMR Biomed 2002;15:570–77 [DOI] [PubMed] [Google Scholar]
  • 6.Sundgren PC, Dong Q, Gomez-Hassan D, et al. Diffusion tensor imaging of the brain: review of clinical applications. Neuroradiology 2004;46:339–50 [DOI] [PubMed] [Google Scholar]
  • 7.Sorensen AG, Wu O, Copen WA, et al. Human acute cerebral ischemia: detection of changes in water diffusion anisotropy by using MR imaging. Radiology 1999;212:785–92 [DOI] [PubMed] [Google Scholar]
  • 8.Vrenken H, Pouwels PJ, Geurts JJ, et al. Altered diffusion tensor in multiple sclerosis normal-appearing brain tissue: cortical diffusion changes seem related to clinical deterioration. J Magn Reson Imaging 2006;23:628–36 [DOI] [PubMed] [Google Scholar]
  • 9.Pierpaoli C, Jezzard P, Basser PJ, et al. Diffusion tensor MR imaging of the human brain. Radiology 1996;201:637–48 [DOI] [PubMed] [Google Scholar]
  • 10.Schneider JF, Il'yasov KA, Boltshauser E, et al. Diffusion tensor imaging in cases of adrenoleukodystrophy: preliminary experience as a marker for early demyelination? AJNR Am J Neuroradiol 2003;24:819–24 [PMC free article] [PubMed] [Google Scholar]
  • 11.Jiang H, van Zijl PC, Kim J, et al. DTIStudio: resource program for diffusion tensor computation and fiber bundle tracking. Comput Methods Programs Biomed 2006;81:106–16 [DOI] [PubMed] [Google Scholar]
  • 12.Landis JR, Koch GG. An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 1977;33:363–74 [PubMed] [Google Scholar]
  • 13.Lee SK, Kim DI, Kim J, et al. Diffusion-tensor MR imaging and fiber tractography: a new method of describing aberrant fiber connections in developmental CNS anomalies. Radiographics 2005;25:53–65, discussion 66–68 [DOI] [PubMed] [Google Scholar]
  • 14.Filippi M, Cercignani M, Inglese M, et al. Diffusion tensor magnetic resonance imaging in multiple sclerosis. Neurology 2001;56:304–11 [DOI] [PubMed] [Google Scholar]
  • 15.Tievsky AL, Ptak T, Farkas J. Investigation of apparent diffusion coefficient and diffusion tensor anisotrophy in acute and chronic multiple sclerosis lesions. AJNR Am J Neuroradiol 1999;20:1491–99 [PMC free article] [PubMed] [Google Scholar]
  • 16.Werring DJ, Clark CA, Barker GJ, et al. Diffusion tensor imaging of lesions and normal-appearing white matter in multiple sclerosis. Neurology 1999;52:1626–32 [DOI] [PubMed] [Google Scholar]
  • 17.Ciccarelli O, Werring DJ, Wheeler-Kingshott CA, et al. Investigation of MS normal-appearing brain using diffusion tensor MRI with clinical correlations. Neurology 2001;56:926–33 [DOI] [PubMed] [Google Scholar]
  • 18.Ge Y, Law M, Grossman RI. Applications of diffusion tensor MR imaging in multiple sclerosis. Ann N Y Acad Sci 2005;1064:202–19 [DOI] [PubMed] [Google Scholar]
  • 19.Alexopoulos GS, Kiosses DN, Choi SJ, et al. Frontal white matter microstructure and treatment response of late-life depression: a preliminary study. Am J Psychiatry 2002;159:1929–32 [DOI] [PubMed] [Google Scholar]
  • 20.Ptak T, Sheridan RL, Rhea JT, et al. Cerebral fractional anisotropy score in trauma patients: a new indicator of white matter injury after trauma. AJR Am J Roentgenol 2003;181:1401–07 [DOI] [PubMed] [Google Scholar]
  • 21.Price SJ, Burnet NG, Donovan T, et al. Diffusion tensor imaging of brain tumours at 3T: a potential tool for assessing white matter tract invasion? Clin Radiol 2003;58:455–62 [DOI] [PubMed] [Google Scholar]
  • 22.Ito R, Melhem ER, Mori S, et al. Diffusion tensor brain MR imaging in X-linked cerebral adrenoleukodystrophy. Neurology 2001;56:544–47 [DOI] [PubMed] [Google Scholar]
  • 23.Schneider JF, Il'yasov KA, Hennig J, et al. Fast quantitative diffusion-tensor imaging of cerebral white matter from the neonatal period to adolescence. Neuroradiology 2004;46:258–66 [DOI] [PubMed] [Google Scholar]
  • 24.Holtmannspötter M, Peters N, Opherk C, et al. Diffusion magnetic resonance histograms as a surrogate marker and predictor of disease progression in CADASIL: a two-year follow-up study. Stroke 2005;36:2559–65 [DOI] [PubMed] [Google Scholar]
  • 25.Kubicki M, Westin CF, Maier SE, et al. Diffusion tensor imaging and its application to neuropsychiatric disorders. Harv Rev Psychiatry 2002;10:324–36 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Pfefferbaum A, Adalsteinsson E, Sullivan EV. Replicability of diffusion tensor imaging measurements of fractional anisotropy and trace in brain. J Magn Reson Imaging 2003;18:427–33 [DOI] [PubMed] [Google Scholar]
  • 27.Neil JJ, Shiran SI, McKinstry RC, et al. Normal brain in human newborns: apparent diffusion coefficient and diffusion anisotropy measured by using diffusion tensor MR imaging. Radiology 1998;209:57–66 [DOI] [PubMed] [Google Scholar]
  • 28.Nusbaum AO, Tang CY, Buchsbaum MS, et al. Regional and global changes in cerebral diffusion with normal aging. AJNR Am J Neuroradiol 2001;22:136–42 [PMC free article] [PubMed] [Google Scholar]
  • 29.Schmithorst VJ, Wilke M, Dardzinski BJ, et al. Correlation of white matter diffusivity and anisotropy with age during childhood and adolescence: a cross-sectional diffusion-tensor MR imaging study. Radiology 2002;222:212–28 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Marenco S, Rawlings R, Rohde GK, et al. Regional distribution of measurement error in diffusion tensor imaging. Psychiatry Res 2006;147:69–78 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Anderson AW. Theoretical analysis of the effects of noise on diffusion tensor imaging. Magn Reson Med 2001;46:1174–88 [DOI] [PubMed] [Google Scholar]
  • 32.Bastin ME, Armigate PA, Marshall I. A theoretical study of the effect of experimental noise on the measurement of anisotrophy in diffusion imaging. Magn Reson Med 1998;16:7:773–85 [DOI] [PubMed] [Google Scholar]
  • 33.Basser PJ, Pajevic S. Statistical artifacts in diffusion tensor MRI (DT-MRI) caused by background noise. Magn Reson Med 2000;44:41–50 [DOI] [PubMed] [Google Scholar]
  • 34.Heim S, Hahn K, Samann PG, et al. Assessing DTI data quality using bootstrap analysis. Magn Reson Med 2004;42:582–89 [DOI] [PubMed] [Google Scholar]
  • 35.Bonekamp D, Nagae LM, Degaonkar M, et al. Diffusion tensor imaging in children and adolescents: reproducibility, hemispheric, and age-related differences. Neuroimage 2007;34:733–42 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Huang H, Zhang J, van Zijl PC, et al. Analysis of noise effects on DTI-based tractography using the brute-force and multi-ROI approach. Magn Reson Med 2004;52:559–65 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

A0998.pdf
A0998.pdf (90.6KB, pdf)

Articles from AJNR: American Journal of Neuroradiology are provided here courtesy of American Society of Neuroradiology

RESOURCES