Abstract
Lung disease is the main cause of morbidity and mortality in cystic fibrosis (CF). CF patients inhale antibiotics regularly as treatment against persistent bacterial infections. The goal of this study was to investigate the effect of clinical intervention on aerosol therapy during the escalation of care using a bench model of adult CF. Droplet size analysis of selected antibiotics was completed in tandem with the delivered aerosol dose (% of total dose) assessments in simulations of various interventions providing oxygen supplementation or ventilatory support. Results highlight the variability of aerosolised dose delivery. In the homecare setting, the vibrating mesh nebuliser (VMN) delivered significantly more than the jet nebuliser (JN) (16.15 ± 0.86% versus 6.51 ± 2.15%). In the hospital setting, using VMN only, significant variability was seen across clinical interventions. In the emergency department, VMN plus mouthpiece (no supplemental oxygen) was seen to deliver (29.02 ± 1.41%) versus low flow nasal therapy (10 L per minute (LPM) oxygen) (1.81 ± 0.47%) and high flow nasal therapy (50 LPM oxygen) (3.36 ± 0.34%). In the ward/intensive care unit, non-invasive ventilation recorded 19.02 ± 0.28%, versus 22.64 ± 1.88% of the dose delivered during invasive mechanical ventilation. These results will have application in the design of intervention-appropriate aerosol therapy strategies and will be of use to researchers developing new therapeutics for application in cystic fibrosis and beyond.
Keywords: cystic fibrosis, lung disease, inhalation, nebuliser, aerosol therapy, antibiotic, escalation of care, home, hospital
1. Introduction
Cystic fibrosis (CF) is one of the most commonly diagnosed genetic diseases worldwide. It is caused by mutations in the cystic fibrosis transmembrane conductance regulator (CFTR) gene [1]. This encodes a cyclic AMP dependent chloride channel expressed in the epithelia of many exocrine tissues including the airways, lung, pancreas, liver, intestine, vas deferens and sweat gland/duct [2]. Mutations in CFTR cause abnormal ion transport in the epithelium of several tissues, which results in the production of abnormally thick and sticky mucus that blocks the organ and is responsible for CF pathology [3]. However, the main cause of morbidity and mortality in CF is lung disease with lung malfunction and pulmonary failure [4]. The production of sticky mucus in the lumen of the lungs impedes mucociliary clearance [5]. Clinically, this manifests as chronic inflammation and recurrent bacterial infections by pathogens such as Pseudomonas aeruginosa and Staphylococus aureus [6]. Mucus and inflammatory cells cause bronchiectasis, and the continuous cycle of infection and inflammation lead to the progressive destruction of the lung tissue [7,8]. Survival of patients with CF has continued to improve to a median age of 46.2 years. This has resulted in a rapid increase in the adult CF population and complications arising from the disease are becoming more common [9]. Therefore, pulmonary disease, one of the major factors linked to mortality [10], is a crucial point in the optimal management of the patients. Consequently, CF patients inhale antibiotics and other therapeutics regularly as treatment against those persistent bacterial infections. Daily treatment with mucolytics, DNase, antibiotics, bronchodilators (considered the “core treatments”), as well as corticosteroids, hypertonic saline and amilorides have been used in the treatment of CF lung disease [11]. As an example, adults with CF who are colonised with Pseudomonas will typically be prescribed at least twice daily nebulised antibiotics in addition to a once daily nebulised mucolytic [12]. The inhaled route offers the potential for high drug concentrations delivered directly to the target organ (airways and lung), eliminating the side effects of the systemic delivery route [13]. Thus, nebuliser performance is a key factor when optimising the delivery of inhaled medications.
The treatment burden is so high in these CF patients that the optimisation of the delivery device may assist individuals in completing the prescribed treatment regimes in different care settings, such as at home or when requiring further treatment during escalation of care. For example, in the case of an acute pulmonary exacerbation, when the patient is admitted in the hospital [14]. The escalation of respiratory support with concurrent aerosol therapy for CF patients likely plays a role in the success of clinical outcomes. These respiratory supports act to deliver supplemental oxygen to the patient, for example, low flow nasal therapy (LFNT) or high flow nasal therapy (HFNT), or provide ventilatory support, for example, non-invasive mechanical ventilation (NIV) or invasive mechanical ventilation (IMV). Figure 1 illustrates the potential escalation of care of CF patients, from home to the hospital setting, and their respiratory support upon the site of care. During that escalation of care, concurrent aerosol therapy may be performed, and the nebulisers used by CF patients may differ within the different scenarios.
Of the two main types of nebulisers, compressed-air-driven jet nebulisers (JN) are the most common in the home. Vibrating mesh nebuliser-type (VMN) is also used in the home, but to a lesser degree. Whilst JN is available in the hospital setting, the VMN is the predominant type. VMN is the only one considered suitable for concurrent use across the various patient interventions encountered throughout the escalation of care from home to intensive care [15,16]. Additionally, current global COVID−19 guidance on the selection of nebuliser recommends the use of devices that maintain a closed system for the prevention of release of both medical and patient-derived bioaerosols into the local environment. VMN is the only nebuliser type that allow this [17,18,19,20,21].
The aim of this study was to investigate the effect of the escalation of care related to the aerosol therapy delivered to the patient under a simulated adult CF breathing profile, and to provide insight into the variability of nebulised dose delivery across clinical interventions either in the homecare or hospital setting.
2. Results
2.1. Respirable Fraction Aerosol Droplet Size Characterisation
The result of droplet size characterisation for a selection of commonly nebulised antibiotic formulations using the Aerogen Solo vibrating mesh nebuliser are presented in Table 1 below. All formulations characterised were within the respirable range with relatively high fine respirable fractions.
Table 1.
Formulation | Concentration | Manufacturer | VMD (μm) |
Flow Rate (mL/min) |
FPF (%) <5 μm |
FPF (%) <3 μm |
FPF (%) 1–5 μm |
---|---|---|---|---|---|---|---|
Saline | 0.9% | BBraun, Ireland | 4.43 ± 0.01 | 0.45 ± 0.01 | 57.45 ± 0.08 | 28.59 ± 0.07 | 56.05 ± 0.03 |
Salbutamol | 2.5 mg/2.5 mL | GlaxoSmithKline | 4.07 ± 0.43 | 0.46 ± 0.01 | 61.38 ± 0.38 | 34.66 ± 0.25 | 56.11 ± 0.59 |
Genticin | 80 mg/2 mL | Amdipharm | 4.21 ± 0.01 | 0.48 ± 0.01 | 59.77 ± 0.07 | 32.49 ± 0.21 | 55.87 ± 0.34 |
Tobramycin | 80 mg/2 mL | Hospira | 4.25 ± 0.01 | 0.47 ± 0.01 | 59.28 ± 0.10 | 31.85 ± 0.11 | 56.15 ± 0.27 |
Likacin | 500 mg/2 mL | TitoLare | 4.78 ± 0.01 | 0.51 ± 0.01 | 52.08 ± 0.08 | 31.25 ± 0.04 | 43.11 ± 0.13 |
Colistin | 0.25 million IU/4 mL | Teva | 4.05 ± 0.03 | 0.21 ± 0.01 | 57.97 ± 0.43 | 39.45 ± 0.51 | 44.97 ± 0.53 |
2.2. Simulated Aerosol Delivery during Spontaneous Breathing in the Homecare Setting
Following tests, the JN was seen to deliver 6.51 ± 2.15% of the nominal dose to the end of the trachea, i.e., the expected lung dose. However, the portable VMN was seen to deliver significantly more, with 16.15 ± 0.86% of the dose delivered to the end of the trachea (p value < 0.001). See Figure 2. The consequence of device selection is apparent here with choice of nebuliser having a significant bearing on the amount of drug delivered to the patient.
2.3. Simulated Aerosol Delivery in the Hospital Setting
2.3.1. Emergency Department
In simulated aerosol drug delivery interventions and oxygen therapy interventions that facilitate concurrent aerosol drug delivery commonly prescribed for patients presenting to the hospital emergency department, mouthpiece-mediated VMN aerosol delivery with no supplemental gas flow rate was seen to deliver the highest dose, with 29.02 ± 1.41% of the nominal dose delivered to the level of the trachea (p value < 0.001 versus low flow and high flow systems, respectively). When the same VMN device was included in a low flow nasal therapy system at 10 L per minute, 1.81 ± 0.47% of the dose was delivered, compared with 3.36 ± 0.34% delivered during high flow nasal therapy at 50 L per minute oxygen flow rate. See Figure 3. In this instance, whilst the same nebuliser type was used, it is obvious that clinical intervention can have a significant bearing on the amount of drug delivered to the patient.
2.3.2. Ward/Intensive Care Unit
During both non-invasive and invasive mechanical ventilation, typical of the intensive care and ward settings, similar levels of drug were seen delivered under ventilation strategies appropriate for the adult cystic fibrosis patient. NIV in combination with VMN recorded 19.02 ± 0.28%, whereas IMV recorded 22.64 ± 1.88% of the dose delivered (p value = 0.012). See Figure 3.
3. Discussion
Aerosol therapy is a mainstay in the routine treatment of CF wherein many therapeutics, including antibiotics, are delivered via nebuliser. Here, we have assessed aerosol drug delivery across a range of CF-appropriate interventions in both the homecare and hospital settings. Using a tracer drug, salbutamol, in combination with adult CF breath patterns, we describe the variability in dose delivery that could be expected to be delivered to the CF patient. This study, for the first time, investigates the effect of device and intervention on the amount of drug delivered to the patient throughout the escalation of care of a simulated exacerbated adult CF patient.
Droplet size is known to affect aerosol drug delivery in the patient, and here we characterised the respirable fraction and droplet size of a selection of commonly nebulised antibiotics [22]. Using a single nebuliser type as the control, indicative droplet size was seen to be relatively consistent with all formulations under test, recording a consistent respirable fraction and droplet size. This would indicate that whilst some differences are noted between formulations, they all remain within the respirable range. Factors affecting droplet size include the physicochemical characteristics, drug concentration, ionic content and temperature of the formulation being aerosolised [23,24,25,26]. Further, significant variation in aerosol characteristics can be expected between devices [27,28]. As such, there will never be a single droplet size distribution that can be relied on, and so these results should be considered as only a feasibility for generation of inhalable antibiotic.
In the home, the predominant nebuliser type used is the JN [29]. In more recent times, however, the vibrating mesh type devices have seen increasing adoption [30,31,32]. These VMN devices claim higher drug delivery rates, with less drug wastage. Here, the results demonstrate that, in this case at least, the portable VMN delivered approximately 2.5 times more to the simulated patient than the JN. The drug/mass amount delivered, whilst not described in the literature previously using the breathing pattern used herein, are in line with results published in imaging studies assessing performance of the JN in a spontaneously breathing adult patient (for example, 6.51% versus 5.2%) [31]. Whilst only one JN type was assessed here, variability in JN performance can be significant, and so these results should be considered representative of the nebuliser type [32]. The dramatic difference in tracheal dose delivered is due to a combination of factors including the large residual volume recorded in JN of all types [33,34], as well as the fact that the drug mass tends to concentrate in the residual volume remaining, as buffer is preferentially aerosolised [23,35].
Should the patient require medical attention and present at the emergency department (ED) in the hospital, a variety of patient interventions may be prescribed in an effort to have the patient return to stable-state lung function. On confirmation of infection, or infection-mediated bronchospasm, inhaled antibiotics, mucolytics or bronchodilators may be prescribed [36,37]. Whereas JN are found and used in the ED, VMN is also, with a mounting body of published clinical literature supporting its adoption [38,39,40]. As mentioned, VMN is also considered the only nebuliser type suitable for concurrent use with a variety of oxygenation supplementation or ventilatory support strategies. In fact, in some cases, JN is contra-indicated against, as the compressed air required for its normal function interferes with the finely tittered oxygen flow prescribed for the patient [41]. In the ED setting, and depending on how severe the state of the patient, and their supplemental oxygen requirement, e.g., if they become hypoxic, they may be prescribed further aerosol therapy. This may be concurrent with low flow nasal therapy (LFNT) or high flow nasal therapy (HFNT). Across these frontline interventions, our studies indicate that aerosol delivery performance can vary significantly. As might be expected, an interface such as the mouthpiece-mediated delivery using the VMN and chamber was seen to deliver the greatest amount of tracer aerosol during simulated breathing, with LFNT and HFNT delivering significantly less. The results recorded here for VMN-mediated aerosol delivery are in line with previous reports for the VMN in combination with the chamber (for example, 29.02% versus 34.1%) [31,42], and HFNT (for example, 3.36% versus 3.46%) [42,43]. No experimentally comparable published data for LFNT could be found.
The larger dose delivered by the combination of VMN and chamber, via mouthpiece, may be explained through a combination of bolus aerosol inhalation, whereby aerosol builds up in the chamber during the exhalation phase of the breath, and the relatively large bore oropharynx, which minimises losses before the lung. The HFNT result is explained through significant aerosol losses throughout the tubing between the nebuliser and the patient but, additionally, the impaction losses seen in the nasal passages after the aerosol exits the nasal cannula [44]. One would expect that, at the lower oxygen flow rates used during LFNT, the ballistic fraction would be smaller and, thus, impaction losses would be minimised; however, a significant fraction of the aerosol is lost in the nasal passages [45]. Further, a major contributor to the reduced aerosol dose here is the very narrow bore oxygen cannula that are typically used [44]. In this instance, the inner diameter of the cannula was 1.8 mm. This is low, compared with the typical internal diameter of adult HFNT cannula, measuring up to as wide as 10 mm [44]. The ratio of nasal cannula outer diameter to nostril inner diameter determines the leak of oxygen, and consequently aerosol from the system also. This ratio also plays a part in the delivered aerosol dose, as well as the levels of fugitive emissions to the local environment, which is a key, but often forgotten consideration for healthcare professionals or bystanders during aerosol therapy [46,47,48,49]. A recent review of aerosol delivery during HFNT by Li and colleagues provides additional insight into the factors at play [50].
Clinically, should these oxygenation supplementation strategies not be successful, and the patient continues to deteriorate, non-invasive ventilation (NIV), or as a last resort, invasive mechanical ventilation (IMV), may be prescribed. These ventilatory support strategies provide pre-set minimum pressures and/or volumes of air and aid the patient in returning to normal pulmonary arterial oxygen levels. NIV is often deployed as pressure support in a conscious spontaneously breathing patient, with IMV having a selection of ventilatory modes available for ventilation of the anaesthetised, unconscious patient via an endotracheal tube. In this study, there were only small differences between NIV and IMV with respect to the delivered aerosol dose (19.02% versus 22.64% respectively). Again, these figures are in line with previous published reports [51,52].
Whilst little difference was noted between NIV and IMV, these interventions deliver significantly more aerosol compared to LFNT and HFNT. The likely reasons for the poor delivery during those interventions has already been explained; however, NIV benefits from large bore tubing (22 mm inner diameter) and a tight-fitting facemask (no leak). Additionally, with the addition of a fixed minimum positive end expiratory pressure (PEEP), aerosol distribution and deposition is likely to be increased over the lower PEEP levels delivered by LFNT and HFNT [53]. IMV also benefits from larger bore tubing, and depending on the ventilator mode, the bias flow-mediated build-up of an aerosol bolus between inhalations [54]. Furthermore, and to ensure safe volumes and pressures are delivered to the patient’s lung on each breath, air leak is minimal, with cuffed endotracheal tubes creating a complete seal within the trachea. This again not only facilitates increased aerosol delivery, but also reduces fugitive medical aerosol emissions to the local environment, and, cognizant of COVID−19, the risk of patient derived transmission of infectious disease [55,56].
The implications of these findings may be far-reaching. For example, patients need to be aware of the potential differences between nebulisers when purchasing from a community pharmacy; however, in some instances this choice may be limited if a particular nebuliser is indicated for use on the drug label. Further, clinicians now have insight into potential variability in aerosol delivery across common clinical interventions, and may, where appropriate, titrate the dose accordingly. Finally, developers of novel therapeutics for the treatment of CF are now provided insight into the significant variance in delivery between devices and interventions, and should consider the intended patient and treatment carefully in the selection of devices and interventions for delivery of their therapeutic.
4. Materials and Methods
4.1. Nebulisers
Experiments were conducted using a selection of appropriate and commonly used nebulisers in the treatment of cystic fibrosis, and the escalating treatment of lung infection. In the homecare setting, both a compressed air-driven jet nebuliser (Up-draft II jet nebuliser, Hudson, Teleflex Medical, Wake County, NC, USA) as well as a portable vibrating mesh nebuliser (InnoSpire Go, Philips, Farnborough, UK) were evaluated. In the hospital setting, a second VMN was used (Aerogen Solo, Aerogen Ltd., Galway, Ireland).
4.2. Respirable Fraction Aerosol Droplet Size Characterisation
The respirable fraction and aerosol droplet size for a selection of antibiotic formulations were characterised using laser diffraction as previously described [57]. Briefly, the devices were loaded with sample and connected to the inlet of the droplet sizer (Spraytec, Malvern Instruments, Worcestershire, UK). The nebuliser was turned on and run until the entire dose was delivered. Testing was carried out in triplicate. The aerosol characteristics of the nebulisers, with respect to volumetric median diameter (VMD) and fine particle fraction (FPF), were measured using laser diffraction (Spraytec, Malvern Instruments, Worcestershire, UK).
4.3. Determination of Tracheal Dose
For all test combinations, 2.5 mL of 2.5 mg/2.5 mL of Albuterol Sulphate (Ventolin, GlaxoSmithKline, Brentford, UK) was used as a tracer aerosol drug to determine drug dose delivered. This is in line with previous studies [44,54,58,59] and international nebuliser test standard ISO27427. A 3D printed adult head model [60,61] was attached to a breathing simulator (BRS2100, Copley Scientific Ltd., Nottingham, UK) via a capture filter, at the level of the trachea bifurcation (Respirgard 303, Vyaire, Basingstoke, UK) and breathing pattern applied. See Table 2 for summary of simulated aerosol delivery test scenarios with associated nebuliser types, interfaces and breath patterns utilised for each test. At the end of each dose, the drug captured on the filter was eluted using 10 mL of deionised water. The mass of drug was quantified by means of UV spectrophotometry at a wavelength of 276 nm and interpolation on a standard curve of albuterol sulphate concentrations (100–3.125 µg/mL). Results for tracheal dose were expressed as the percentage of the nominal dose placed in the nebuliser’s medication cup. All testing was carried out in independent quintuplicate.
Table 2.
Simulated Aerosol Delivery Test Scenario |
Nebuliser Type | Interface | Simulated Breath Parameters |
---|---|---|---|
Spontaneous breathing (SB) | VMN | Mouthpiece | Vt 410 mL RR 22 BPM I:E Ratio 1:2 [62,63] |
JN | Open facemask | ||
Non-invasive patient intervention nasal oxygen | VMN | Nasal cannula with LFNT at 10 LPM | |
VMN | Nasal cannula with HFNT at 60 LPM | ||
Non-invasive ventilation (NIV) | VMN | Non-vented NIV mask | |
Invasive mechanical ventilation (IMV) | VMN | Endotracheal tube | Vt 550 mL RR 13 BPM I:E Ratio 1:2 [64] |
4.4. Simulated Aerosol Delivery during Spontaneous Breathing
To replicate aerosol therapy in a spontaneously breathing adult CF patient undergoing standard care in the homecare setting, both the JN plus open facemask (Up-draft II jet nebuliser, Hudson, Teleflex Medical, Wake County, NC, USA) and portable VMN in combination with a mouthpiece (InnoSpire Go, Philips, Farnborough, UK) were used. The JN was operated at 7 L per minute compressed air flow rate, in line with the accompanying instruction manual. To replicate aerosol therapy in a spontaneously breathing adult CF patient post-exacerbation in the hospital setting a VMN in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Baxter, Dublin, Ireland) with no supplemental oxygen was used to deliver drug to the simulated adult CF patient. See Figure 4A,B.
4.5. Simulated Aerosol Delivery during A Non-Invasive Patient Intervention–Nasal Oxygen Support
Low flow oxygen therapy (LFNT): A nasal oxygen cannula (Hudson RCI comfort flo® nasal cannula, Teleflex Medical, Delaware County PA, USA) placed in the nostrils of the head model was used to simultaneously deliver low flow oxygen therapy at 10 LPM while aerosol from the VMN (Aerogen Solo, Aerogen Limited, Galway, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke, UK), with no supplemental oxygen. See Figure 5.
High flow oxygen therapy (HFNT): A nasal cannula (OPT944, Optiflow + Medium, Fisher & Paykel, New Zealand) placed on the head model was used to deliver oxygen at 50 LPM with the turbine driven HFNT device (AIRVO2, Fisher & Paykel, Auckland New Zealand). Aerosol from the VMN (Aerogen Solo, Aerogen Limited, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke UK), with no supplemental oxygen. See Figure 6.
4.6. Simulated Aerosol Delivery during Mask-Mediated Non-Invasive Ventilation (NIV)
A critical care mechanical ventilator with non-invasive functionality (Bellavista, IMT Medical, Buchs Switzerland), was set to deliver continuous positive airway pressure (CPAP) of 10 cmH20. It was attached to the breathing simulator via a single limb, vented circuit with a non-vented NIV mask (RT045, Nivairo, Fisher and Paykel, Auckland New Zealand). The breathing simulator was set to deliver the same breathing pattern as the spontaneously breathing patient (see Table 2). In line with published in vivo imaging studies of the optimal nebuliser placement, the VMN was placed between the exhalation port of the single limb circuit and the mask [65]. The mask was attached to the breathing simulator via a capture filter. See Figure 7.
4.7. Simulated Aerosol Delivery during Endotracheal Tube-Mediated Invasive Mechanical Ventilation (IMV)
A critical care mechanical ventilator (Servo-U, Maquet, Baden-Württemberg, Germany) incorporating a dual limb circuit (RT200, Fisher & Paykel, Auckland New Zealand) was used in combination with a humidifier (MR850, Fisher & Paykel, Auckland New Zealand). The VMN was placed at the dry side of the humidifier. A lung protective, low tidal volume ventilation strategy of 8 mL/Kg for a 69 Kg adult was adopted (Vt 550 mL, RR 13 BPM, and I:E Ratio 1:2) [64]. The capture filter was placed between the endotracheal tube (ETT) (Flexicare, 8.00 mm, UK) and a test lung (IMT Medical, Bachs Switzerland). See Figure 8.
4.8. Statistical Analysis
Data are expressed as the arithmetic mean ± standard deviation (SD) tracheal dose (percentage). Statistical analysis was carried out using Minitab Software, version 18 (Coventry, UK). Student’s t-tests were conducted to establish if the tracheal dose varied significantly across two different aerosol generators in the homecare setting. A One-Way ANOVA was performed to assess different drug delivery interfaces, across various patient interventions using a VMN only, in the hospital setting. Differences were considered statistically significant when p ≤ 0.05 (*). Aerosol droplet size characterisation work was carried out in triplicate. Tracheal dose experiments were conducted in quintuplicate and with at least five technical replicates per experiment.
5. Study Limitations
The experimental approach taken here to describe the gross or total aerosol inhaled and delivered to the level of the trachea does not consider the disease specific features of CF such as mucus secretions, inflammation, infection etcetera and the influence they may have on distribution and deposition throughout the airways.
6. Conclusions
This study highlights the potential for sometimes significant variance in aerosol drug delivery between both nebuliser type, and clinical intervention. These results will have application in the design of intervention-appropriate aerosol therapy strategies and will be of use to researchers developing new therapeutics for application in cystic fibrosis and beyond, as well as clinicians in the selection of nebuliser/intervention combinations in the treatment of patients.
Author Contributions
Conceptualization, E.F.F. and R.M.; methodology, M.J. and A.O.; formal analysis, E.F.F., M.J. and A.O., writing—original draft preparation, E.F.F., M.J., A.O. and R.M.; writing—review and editing, E.F.F., M.J., A.O., and R.M.; All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by Aerogen Limited.
Data Availability Statement
All relevant data is contained within the article.
Conflicts of Interest
This research was funded by Aerogen Limited. All authors are employees of Aerogen Limited.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Kerem B., Rommens J.M., A Buchanan J., Markiewicz D., Cox T.K., Chakravarti A., Buchwald M., Tsui L.C. Identification of the cystic fibrosis gene: Genetic analysis. Science. 1989;245:1073–1080. doi: 10.1126/science.2570460. [DOI] [PubMed] [Google Scholar]
- 2.Welsh M.J., Smith A.E. Molecular mechanisms of CFTR chloride channel dysfunction in cystic fibrosis. Cell. 1993;73:1251–1254. doi: 10.1016/0092-8674(93)90353-R. [DOI] [PubMed] [Google Scholar]
- 3.Fernandez E., De Santi C., De Rose V., Greene C.M. CFTR dysfunction in cystic fibrosis and chronic obstructive pulmonary disease. Expert Rev. Respir. Med. 2018;12:483–492. doi: 10.1080/17476348.2018.1475235. [DOI] [PubMed] [Google Scholar]
- 4.Gibson R.L., Burns J.L., Ramsey B.W. Pathophysiology and Management of Pulmonary Infections in Cystic Fibrosis. Am. J. Respir. Crit. Care Med. 2003;168:918–951. doi: 10.1164/rccm.200304-505SO. [DOI] [PubMed] [Google Scholar]
- 5.Boucher R.C. Cystic fibrosis: A disease of vulnerability to airway surface dehydration. Trends Mol. Med. 2007;13:231–240. doi: 10.1016/j.molmed.2007.05.001. [DOI] [PubMed] [Google Scholar]
- 6.Davis P.B. Cystic Fibrosis Since 1938. Am. J. Respir. Crit. Care Med. 2006;173:475–482. doi: 10.1164/rccm.200505-840OE. [DOI] [PubMed] [Google Scholar]
- 7.Hoenderdos K., Condliffe A. The neutrophil in chronic obstructive pulmonary disease. Am. J. Respir. Cell. Mol. Biol. 2013;48:531–539. doi: 10.1165/rcmb.2012-0492TR. [DOI] [PubMed] [Google Scholar]
- 8.King P., Holdsworth S., Freezer N., Holmes P. Bronchiectasis. Intern. Med. J. 2006;36:729–737. doi: 10.1111/j.1445-5994.2006.01219.x. [DOI] [PubMed] [Google Scholar]
- 9.Dickinson K.M., Collaco J.M. Cystic fibrosis. [(accessed on 11 March 2021)];Pediatr Rev. Am. Acad. Pediatrics. 2021 42:55–65. doi: 10.1542/pir.2019-0212. Available online: https://pedsinreview.aappublications.org/content/42/2/55. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Bhatt J.M. Treatment of pulmonary exacerbations in cystic fibrosis. [(accessed on 12 March 2021)];Eur. Respir. Rev. Eur. Respir. Soc. 2013 :205–216. doi: 10.1183/09059180.00006512. Available online: http://ow.ly/mK5df. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Borsje P., De Jongste J.C., Tiddens H.A. Aerosol therapy in cystic fibrosis: A survey of 54 CF centers. Pediatr. Pulmonol. 2000;30:368–376. doi: 10.1002/1099-0496(200011)30:5<368::AID-PPUL2>3.0.CO;2-8. [DOI] [PubMed] [Google Scholar]
- 12.Mogayzel P.J., Naureckas E.T., Robinson K.A., Mueller G., Hadjiliadis D., Hoag J.B., Lubsch L., Hazle L., Sabadosa K., Marshall B., et al. Cystic Fibrosis Pulmonary Guidelines. Am. J. Respir. Crit. Care Med. 2013;187:680–689. doi: 10.1164/rccm.201207-1160OE. [DOI] [PubMed] [Google Scholar]
- 13.Labiris N.R., Dolovich M.B. Pulmonary drug delivery. Part I: Physiological factors affecting therapeutic effectiveness of aerosolized medications. Br. J. Clin. Pharmacol. 2003:588–599. doi: 10.1046/j.1365-2125.2003.01892.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Goss C.H. Acute Pulmonary Exacerbations in Cystic Fibrosis. Semin. Respir. Crit. Care Med. 2019;40:792–803. doi: 10.1055/s-0039-1697975. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.O’Callaghan C., White J., Jackson J., Crosby D., Dougill B., Bland H. The Effects of Heliox on the Output and Particle-Size Distribution of Salbutamol Using Jet and Vibrating Mesh Nebulizers. J. Aerosol Med. 2007;20:434–444. doi: 10.1089/jam.2007.0614. [DOI] [PubMed] [Google Scholar]
- 16.Ehrmann S. Vibrating Mesh Nebulisers—Can Greater Drug Delivery to the Airways and Lungs Improve Respiratory Outcomes? Eur. Respir. Pulm. Dis. Touch Med. Media. 2018;4:33. doi: 10.17925/ERPD.2018.4.1.33. [DOI] [Google Scholar]
- 17.Halpin D.M.G., Criner G.J., Papi A., Singh D., Anzueto A., Martinez F.J., Agusti A.A., Vogelmeier C.F. Global Initiative for the Diagnosis, Management, and Prevention of Chronic Obstructive Lung Disease. The 2020 GOLD Science Committee Report on COVID-19 and Chronic Obstructive Pulmonary Disease. Am. J. Respir. Crit. Care Med. 2021;203:24–36. doi: 10.1164/rccm.202009-3533SO. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Respiratory care committee of Chinese Thoracic Society Expert consensus on protective measures related to respiratory therapy in patients with severe coronavirus infection. [(accessed on 11 March 2021)];Zhonghua Jie He He Hu Xi Za Zhi. 2020 17:E020. doi: 10.3760/cma.j.issn.1001-0939.2020.0020. Available online: http://www.ncbi.nlm.nih.gov/pubmed/32077661. [DOI] [PubMed] [Google Scholar]
- 19.AARC American Association for Respiratory Care (AARC) 2020. [(accessed on 11 March 2021)]; Available online: https://www.aarc.org/wp-content/uploads/2020/03/guidance-document-SARS-COVID19.pdf.
- 20.Kulkarni A.P., Kumar G.P., Gopal P.B. Airway Management and Related Procedures in Critically Ill COVID-19 Patients: Position Statement of the Indian Society of Critical Care Medicine. Indian J. Crit. Care Med. 2020;24:630–642. doi: 10.5005/jp-journals-10071-23471. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Gómez C.C., Rodríguez Ó.P., Torné M.L., Santaolalla C.E., Jiménez J.F.M., Fernández J.G. Clinical Consensus Recommendations Regarding Non-Invasive Respiratory Support in the Adult Patient with Acute Respiratory Failure Secondary to SARS-CoV-2 infection. [(accessed on 11 March 2021)];Arch. Bronconeumol. 2020 56:11–18. doi: 10.1016/j.arbres.2020.03.005. Available online: https://pubmed.ncbi.nlm.nih.gov/32336563/ [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Usmani O.S., Biddiscombe M.F., Barnes P.J. Regional Lung Deposition and Bronchodilator Response as a Function of β2-Agonist Particle Size. Am. J. Respir. Crit. Care Med. 2005;172:1497–1504. doi: 10.1164/rccm.200410-1414OC. [DOI] [PubMed] [Google Scholar]
- 23.Steckel H., Eskandar F. Factors affecting aerosol performance during nebulization with jet and ultrasonic nebulizers. Eur. J. Pharm. Sci. 2003;19:443–455. doi: 10.1016/S0928-0987(03)00148-9. [DOI] [PubMed] [Google Scholar]
- 24.Ghazanfari T., Elhissi A.M., Ding Z., Taylor K.M. The influence of fluid physicochemical properties on vibrating-mesh nebulization. Int. J. Pharm. 2007;339:103–111. doi: 10.1016/j.ijpharm.2007.02.035. [DOI] [PubMed] [Google Scholar]
- 25.Mc Callion O.N.M., Patel M.J. Viscosity effects on nebulisation of aqueous solutions. Int. J. Pharm. 1996;130:245–249. doi: 10.1016/0378-5173(95)04291-1. [DOI] [Google Scholar]
- 26.Najlah M., Vali A., Taylor M., Arafat B.T., Ahmed W., Phoenix D.A., Taylor K.M., Elhissi A. A study of the effects of sodium halides on the performance of air-jet and vibrating-mesh nebulizers. Int. J. Pharm. 2013;456:520–527. doi: 10.1016/j.ijpharm.2013.08.023. [DOI] [PubMed] [Google Scholar]
- 27.Pritchard J.N., Hatley R.H., Denyer J., Von Hollen D. Mesh nebulizers have become the first choice for new nebulized pharmaceutical drug developments. Ther. Deliv. 2018;9:121–136. doi: 10.4155/tde-2017-0102. [DOI] [PubMed] [Google Scholar]
- 28.Pritchard J.N. Nebulized drug delivery in respiratory medicine: What does the future hold? Ther. Deliv. 2017;8:391–399. doi: 10.4155/tde-2017-0015. [DOI] [PubMed] [Google Scholar]
- 29.Rosenfeld M., Emerson J., Astley S., Joy P., Williams-Warren J., Standaert T.A., Yim D.L., Crist D., Thykkuttathil M., Torrence M., et al. Home nebulizer use among patients with cystic fibrosis. J. Pediatr. 1998;132:125–131. doi: 10.1016/S0022-3476(98)70497-4. [DOI] [PubMed] [Google Scholar]
- 30.Nickerson C., Von Hollen D., Garbin S., Doty K., Jasko J., Cain C. Preference and quality of life of adult chronic obstructive lung disease (COPD) patients when using a novel mesh nebulizer compared to traditional jet nebulizer (TJN) Airway Pharmacol. Treat. 2020;56:640. doi: 10.1183/13993003.congress-2020.640. [DOI] [Google Scholar]
- 31.Dugernier J., Hesse M., Vanbever R., Depoortere V., Roeseler J., Michotte J.-B., Laterre P.-F., Jamar F., Reychler G. SPECT-CT Comparison of Lung Deposition using a System combining a Vibrating-mesh Nebulizer with a Valved Holding Chamber and a Conventional Jet Nebulizer: A Randomized Cross-over Study. Pharm. Res. 2016;34:290–300. doi: 10.1007/s11095-016-2061-7. [DOI] [PubMed] [Google Scholar]
- 32.Adorni G., Seifert G., Buttini F., Colombo G., Stecanella L.A., Krämer I., Rossi A. Aerosolization Performance of Jet Nebulizers and Biopharmaceutical Aspects. Pharmaceutics. 2019;11:406. doi: 10.3390/pharmaceutics11080406. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.McGrath J.A., O’Sullivan A., Bennett G., O’Toole C., Joyce M., Byrne M.A., MacLoughlin R. Investigation of the Quantity of Exhaled Aerosols Released into the Environment during Nebulisation. Pharmaceutics. 2019;11:75. doi: 10.3390/pharmaceutics11020075. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Hurley P., Smye S., Cunliffe H. Assessment of Antibiotic Aerosol Generation Using Commercial Jet Nebulizers. J. Aerosol Med. 1994;7:217–228. doi: 10.1089/jam.1994.7.217. [DOI] [PubMed] [Google Scholar]
- 35.Newman S.P., Pellow P.G., Clay M.M., Clarke S.W. Evaluation of jet nebulisers for use with gentamicin solution. Thorax. 1985;40:671–676. doi: 10.1136/thx.40.9.671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Bell S.C., Mall M.A., Gutierrez H., Macek M., Madge S., Davies J.C. The future of cystic fibrosis care: A global perspective. [(accessed on 13 January 2021)];Lancet Respir. Med. 2020 :65–124. doi: 10.1016/S2213-2600(19)30337-6. Available online: https://pubmed.ncbi.nlm.nih.gov/31570318/ [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Smyth A.R., Bell S.C., Bojcin S., Bryon M., Duff A., Flume P., Kashirskaya N., Munck A., Ratjen F., Schwarzenberg S.J., et al. European Cystic Fibrosis Society Standards of Care: Best Practice guidelines. J. Cyst. Fibros. 2014;13:S23–S42. doi: 10.1016/j.jcf.2014.03.010. [DOI] [PubMed] [Google Scholar]
- 38.Dunne R.B., Shortt S. Comparison of bronchodilator administration with vibrating mesh nebulizer and standard jet nebulizer in the emergency department. Am. J. Emerg. Med. 2018;36:641–646. doi: 10.1016/j.ajem.2017.10.067. [DOI] [PubMed] [Google Scholar]
- 39.Moody G. Comparison of Vibrating Mesh Nebulizer Versus Jet Nebulizer in the Pediatric Asthma Patient: A Randomized Controlled Trial. Respir. Care. 2019;64:3217971. [Google Scholar]
- 40.Cantu T., Jenkins L. Quality Improvement Project to Compare Vibrating Mesh Nebulizer Therapy with Hour Long Jet Nebulizer Therapy for Albuterol Delivery in Asthma and Reactive Airway Disorder Patients in a Pediatric Emergency Department. Respir. Care. 2019;64:3238888. [Google Scholar]
- 41.Caille V., Ehrmann S., Boissinot E., Perrotin D., Diot P., Dequin P.-F. Influence of Jet Nebulization and Oxygen Delivery on the Fraction of Inspired Oxygen: An Experimental Model. J. Aerosol Med. Pulm. Drug Deliv. 2009;22:255–261. doi: 10.1089/jamp.2008.0718. [DOI] [PubMed] [Google Scholar]
- 42.Forde É., Kelly G., Sweeney L., Fitzgerald-Hughes D., MacLoughlin R., Devocelle M. Vibrating Mesh Nebulisation of Pro-Antimicrobial Peptides for Use in Cystic Fibrosis. Pharmaceutics. 2019;11:239. doi: 10.3390/pharmaceutics11050239. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Alcoforado L., Ari A., Barcelar J.D.M., Brandão S.C.S., Fink J.B., De Andrade A.D. Impact of Gas Flow and Humidity on Trans-Nasal Aerosol Deposition via Nasal Cannula in Adults: A Randomized Cross-Over Study. Pharmaceutics. 2019;11:320. doi: 10.3390/pharmaceutics11070320. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Bennett G., Joyce M., Sweeney L., MacLoughlin R. In Vitro Determination of the Main Effects in the Design of High-Flow Nasal Therapy Systems with Respect to Aerosol Performance. Pulm. Ther. 2018;4:73–86. doi: 10.1007/s41030-018-0054-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Réminiac F., Vecellio L., Mac Loughlin R., Le Pennec D., Cabrera M., Vourc’H N.H., Fink J.B., Ehrmann S. Nasal high flow nebulization in infants and toddlers: An in vitro and in vivo scintigraphic study. Pediatr. Pulmonol. 2016;52:337–344. doi: 10.1002/ppul.23509. [DOI] [PubMed] [Google Scholar]
- 46.McGrath J.A., O’Toole C., Bennett G., Joyce M., Byrne M.A., MacLoughlin R. Investigation of Fugitive Aerosols Released into the Environment during High-Flow Therapy. Pharmaceutics. 2019;11:254. doi: 10.3390/pharmaceutics11060254. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Mirabelli M.C., Zock J.-P., Plana E., Antó J.M., Benke G., Blanc P.D., Dahlman-Höglund A., Jarvis D.L., Kromhout H., Lillienberg L., et al. Occupational risk factors for asthma among nurses and related healthcare professionals in an international study. Occup. Environ. Med. 2007;64:474–479. doi: 10.1136/oem.2006.031203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Beach J.R., Campbell M., Andrews D.J. Exposure of health care workers to pentamidine isethionate. Occup. Med. 1999;49:243–245. doi: 10.1093/occmed/49.4.243. [DOI] [PubMed] [Google Scholar]
- 49.Tsai R.J., Boiano J.M., Steege A.L., Sweeney M.H. Precautionary Practices of Respiratory Therapists and Other Health-Care Practitioners Who Administer Aerosolized Medications. Respir. Care. 2015;60:1409–1417. doi: 10.4187/respcare.03817. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Li J., Fink J.B., MacLoughlin R., Dhand R. A narrative review on trans-nasal pulmonary aerosol delivery. Crit. Care. 2020;24:1–13. doi: 10.1186/s13054-020-03206-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Galindo-Filho V.C., Alcoforado L., Rattes C., Paiva D.N., Brandão S.C.S., Fink J.B., de Andrade A.D. A mesh nebulizer is more effective than jet nebulizer to nebulize bronchodilators during non-invasive ventilation of subjects with COPD: A randomized controlled trial with radiolabeled aerosols. Respir. Med. 2019;153:60–67. doi: 10.1016/j.rmed.2019.05.016. [DOI] [PubMed] [Google Scholar]
- 52.Dugernier J., Reychler G., Wittebole X., Roeseler J., Depoortere V., Sottiaux T., Michotte J.-B., Vanbever R., Dugernier T., Goffette P., et al. Aerosol delivery with two ventilation modes during mechanical ventilation: A randomized study. Ann. Intensiv. Care. 2016;6:1–10. doi: 10.1186/s13613-016-0169-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Fauroux B., Itti E., Pigeot J., Isabey D., Meignan M., Ferry G. Optimization of aerosol deposition pressure support in children with cystic fibrosis: An experimental and clinical study. [(accessed on 14 March 2021)];Am. J. Respir. Crit. Care Med. 2000 162:2265–2271. doi: 10.1164/ajrccm.162.6.2003069. Available online: https://pubmed.ncbi.nlm.nih.gov/11112150/ [DOI] [PubMed] [Google Scholar]
- 54.Ari A., Atalay O.T., Harwood R., Sheard M.M., A Aljamhan E., Fink J.B. Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation. Respir. Care. 2010;55:281–289. [PubMed] [Google Scholar]
- 55.O’Toole C., McGrath J.A., Joyce M., Bennett G., Byrne M.A., MacLoughlin R. Fugitive Aerosol Therapy Emissions during Mechanical Ventilation: In Vitro Assessment of the Effect of Tidal Volume and Use of Protective Filters. Aerosol Air Qual. Res. 2020;20:2604–2613. doi: 10.4209/aaqr.2020.04.0176. [DOI] [Google Scholar]
- 56.Joyce M., McGrath J.A., Mac Giolla Eain M., O’Sullivan A., Byrne M., MacLoughlin R. Nebuliser Type Influences Both Patient-Derived Bioaerosol Emissions and Ventilation Parameters during Mechanical Ventilation. [(accessed on 11 March 2021)];Pharmaceutics. 2021 13:199. doi: 10.3390/pharmaceutics13020199. Available online: https://www.mdpi.com/1999-4923/13/2/199. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Hibbitts A., O’Mahony A., Forde E., Nolan L., Ogier J., Desgranges S., Darcy R., MacLoughlin R., O’Driscoll C., Cryan S.-A., et al. Early-Stage Development of Novel Cyclodextrin-siRNA Nanocomplexes Allows for Successful Postnebulization Transfection of Bronchial Epithelial Cells. J. Aerosol Med. Pulm. Drug Deliv. 2014;27:466–477. doi: 10.1089/jamp.2013.1045. [DOI] [PubMed] [Google Scholar]
- 58.Sweeney L., McCloskey A.P., Higgins G., Ramsey J.M., Cryan S.-A., MacLoughlin R. Effective nebulization of interferon-γ using a novel vibrating mesh. Respir. Res. 2019;20:66. doi: 10.1186/s12931-019-1030-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Moody G.B., Ari A. Quantifying continuous nebulization via high flow nasal cannula and large volume nebulizer in a pediatric model. Pediatr. Pulmonol. 2020;55:2596–2602. doi: 10.1002/ppul.24967. [DOI] [PubMed] [Google Scholar]
- 60.Rygg A., Longest P.W. Absorption and Clearance of Pharmaceutical Aerosols in the Human Nose: Development of a CFD Model. J. Aerosol Med. Pulm. Drug Deliv. 2016;29:416–431. doi: 10.1089/jamp.2015.1252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Golshahi L., Longest P.W., Azimi M., Syed A., Hindle M. Intermittent Aerosol Delivery to the Lungs During High-Flow Nasal Cannula Therapy. Respir. Care. 2014;59:1476–1486. doi: 10.4187/respcare.02903. [DOI] [PubMed] [Google Scholar]
- 62.Hart N., Polkey M.I., Clément A., Boulé M., Moxham J., Lofaso F., Fauroux B. Changes in Pulmonary Mechanics with Increasing Disease Severity in Children and Young Adults with Cystic Fibrosis. Am. J. Respir. Crit. Care Med. 2002;166:61–66. doi: 10.1164/rccm.2112059. [DOI] [PubMed] [Google Scholar]
- 63.Tiddens H., Koopman L., Lambert R., Elliott W.M., Hop W., Van Der Mark T.W., De Boer W.J., De Jongste J. Cartilaginous airway wall dimensions and airway resistance in cystic fibrosis lungs. Eur. Respir. J. 2000;15:735–742. doi: 10.1034/j.1399-3003.2000.15d18.x. [DOI] [PubMed] [Google Scholar]
- 64.Rouby J.J., Sole-Lleonart C., Rello J., Monsel A., Constantin J.M., Bouglé A. Ventilator-associated pneumonia caused by multidrug-resistant Gram-negative bacteria: Understanding nebulization of aminoglycosides and colistin. [(accessed on 10 March 2021)];Intensive Care Med. 2020 46:766–770. doi: 10.1007/s00134-019-05890-w. Available online: https://pubmed.ncbi.nlm.nih.gov/31915838/ [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Galindo-Filho V.C., Ramos M.E., Rattes C.S.F., Barbosa A.K., Brandão D.C., Brandão S.C.S., Fink J.B., De Andrade A.D. Radioaerosol Pulmonary Deposition Using Mesh and Jet Nebulizers During Noninvasive Ventilation in Healthy Subjects. Respir. Care. 2015;60:1238–1246. doi: 10.4187/respcare.03667. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data is contained within the article.