Abstract
Neurofilament light chain (NfL) has emerged as a biomarker of neuroaxonal damage in several neurologic conditions. With increasing availability of fourth-generation immunoassays detecting NfL in blood, aspects of pre-analytical stability of this biomarker remain unanswered. This study investigated NfL concentrations in serum and plasma samples of 32 patients with neurological diagnoses using state of the art Simoa technology. We tested the effect of delayed freezing of up to 7 days and statistically determined stability and validity of measured concentrations. We found concentrations of NfL in serum and plasma to remain stable at room temperature when processing of samples is delayed up to 7 days (serum: mean absolute difference 0.9 pg/mL, intraindividual variation 1.2%; plasma: mean absolute difference 0.5 pg/mL, intraindividual variation 1.3%). Consistency of these results was nearly perfect for serum and excellent for plasma (intraclass correlation coefficients 0.99 and 0.94, respectively). In conclusion, the soluble serum and plasma NfL concentration remains stable when unprocessed blood samples are stored up to 7 days at room temperature. This information is essential for ensuring reliable study protocols, for example, when shipment of fresh samples is needed.
Subject terms: Neurology, Neuroimmunology
Introduction
Neurofilaments are part of the axon’s cytoskeleton and are released into the blood and cerebrospinal fluid following axonal damage1,2. Over the past decade, neurofilament light chain (NfL) has emerged as a multifaceted biomarker in blood allowing for the detection of damage to the central and peripheral nervous system3–6. With its ability to predict treatment response and neurological outcome in some cases, NfL has since been coined the c-reactive protein of neurology7. State of the art single molecule array (Simoa) technology allows for reliable detection of this low-concentrated biomarker in the range of a few picograms per milliliter8,9.
With an ever-increasing amount of NfL research and study protocols employing the inclusion of NfL from blood, there is an important pre-analytical consideration for sample storage and handling. Currently, there is some evidence from a small cohort suggesting that serum NfL are stable if unprocessed blood is stored up to 7 days at room temperature10. Furthermore, a more recent study shows stability for NfL concentrations in serum if samples are left up to 24 h at room temperature11. However, there is no investigation with the primary hypothesis of confirming stability of NfL up to 7 days in unprocessed samples of both serum and plasma.
This study explores the pre-analytical stability of NfL in serum and plasma with conditions in mind, for which standard processing might not be possible. These circumstances may apply to, for instance, an out of hospital collection requiring shipment to other facilities or sample collection outside of regular working hours. Therefore, the core investigation involved measurement of NfL concentration in serum and plasma from 32 volunteers whose sample processing (i.e., centrifuging and freezing) was delayed up to 7 days. We also report on the analytical performance characteristics of this assay with respect to precision and repeatability.
Methods
Ethics statement
The ethics committee at the Medical University of Vienna (protocol number EK2100/2019, amended in July 2020) approved this study. Inclusion of participants into the biobanking protocol followed written informed consent. This study complies with the World Medical Association Declaration of Helsinki regarding ethical conduct of research involving human subjects.
Study population
For this study, we used serum and plasma samples of participants from our outpatient clinic at the Department of Neurology at the Medical University of Vienna, Austria. To avoid interference with our comparing measurements in serum and plasma, only patients not taking any medication affecting thrombocyte function of blood coagulation were included. Based on a previous study we performed an a priori power calculation for detecting a systemic measurement error of 2 pg/mL (at 80% power and an α-level of 0.05), we included groups consisting of (i) 32 patients for the main study of delayed processing of serum and plasma samples, and (ii) a validity cohort of 4 patients whose NfL concentrations in paired plasma and serum aliquots were determined repeatedly on four assay plates to determine inter-assay reliability11.
Blood sampling
From all participants, peripheral venous blood was collected in Greiner Bio-One Vacuette serum and ethylenediaminetetraacetic acid (EDTA) tubes (GBO, Kremsmuenster, Austria). Blood vials were sent unaltered and at room temperature to the local biobank (Department of Laboratory Medicine, Medical University of Vienna) at different points in time depending on the group they were assigned to. Once in the biobank laboratory, samples were processed following standard operating procedures12. In brief, serum tubes were centrifuged after clotting had been completed at 1.884×g at room temperature. Serum was transferred to 500 µL polypropylene tubes and stored as 400 µL aliquots at − 70 °C until analysis. Plasma tubes were equally processed and stored without waiting for the material to coagulate before centrifugation.
Sample handling for the delayed freezing-group
We drew a total of 4 serum and 4 plasma samples from each of the 32 participants in this group. Serum and plasma pairs from the same patient were stored at four different conditions before being sent to the local biobank for processing. The first serum and plasma tubes were sent to the local biobank immediately and at room temperature. The second serum and plasma tubes were placed in a refrigerator (temperature controlled at 4–8 °C) for 3 days before being sent to the local biobank. The third serum und plasma tubes were kept at room temperature for 3 days before being sent to the local biobank. The fourth serum and plasma tubes were kept at room temperature for 7 days before being sent to the local biobank.
Sample handling for the control group measuring assay precision
In order to determine inter-assay reproducibility, we repeatedly measured NfL concentrations in 8 paired serum and plasma samples from 4 volunteers on four consecutive assay plates.
Analysis of neurofilament light chain concentrations in serum and plasma
Serum and plasma aliquots were thawed for an hour at room temperature and analyzed by a blinded investigator. NfL concentrations in serum (sNfL) and plasma (pNfL) samples were determined using the Simoa Nf-light kits on a Simoa SR-X Analyzer following the manufacturer’s instructions (Quanterix, Billerica, MA, USA)13. In brief, thawed samples, calibrators and internal controls were equilibrated to room temperature, diluted (1:4) in sample diluent and transferred to the provided 96-well plates as duplicates. Well and plate assignment was designed so that all aliquots from the same participant were analyzed in the same run to avoid within-subject run-to-run variability. All assay plates (a total of 8) were run on a daily basis. Plates and samples were consecutively incubated with 20 µL of detector reagent, 25 µL of paramagnetic beads, and 100 μL streptavidin β-galactosidase. Over the course of these steps, plates were processed on the Simoa microplate incubator and Simoa microplate washer as per instructions. After final washing, pellets were allowed to dry for 10 min before transfer to the Quanterix SR-X analyzer for readout. As per usual convention, only samples yielding a coefficient of variance (CV) between paired replicates below 0.20 were included in this study.
Statistical analyses
Statistical analysis was performed using SPSS 25.0 (SPSS Inc, Chicago, IL, USA). Categorical variables were expressed in frequencies and percentages, continuous variables were tested for Gaussian distribution by the Shapiro–Wilk test and displayed as mean and 95% CI or median and range as appropriate. Differences between sNfL and pNfL values after different freezing intervals were analyzed using repeated measurement ANOVA. A two-sided p-value of 0.05 was considered the level of significance. Reproducibility of sNfL measurement across different freezing intervals was assessed by calculating intraindividual standard deviations (iSD) and intraindividual covariances (iCOV). We also created Bland–Altman plots to visualize the agreement between different freezing intervals (Y-axis shows the difference between immediate freezing and different freezing intervals and the X-axis represents the average of these measures)14. Finally, intraclass correlation coefficients (ICC) were calculated. ICC calculation is based on a repeated measure ANOVA model using the variance among subjects, the variance among measurements, and the residual error variance14. An ICC of 1 represents perfect reproducibility, while an ICC above 0.9 is rated excellent for laboratory tests.
Results
Patient characteristics
Table 1 lists characteristics of participants in the main study setup (delayed freezing) and their mean NfL concentrations in serum (sNfL) and plasma (pNfL) at standard conditions (immediate processing). The mean age in this cohort was 40.2 years (95% CI 34.5–45.9), 38% of participants were female. The overall mean sNfL concentration was 14.5 pg/mL (95% CI 9.6–19.4), the mean pNfL concentration was 12.5 pg/mL (95% CI 8.6–16.5). Neurological diagnoses of patients included in this study were multiple sclerosis (n = 20), Parkinson’s disease (n = 6), Alzheimer disease (n = 4), and primary headache (n = 2).
Table 1.
Patient ID | Age | Sex | sNfL (pg/mL) | pNfL (pg/mL) |
---|---|---|---|---|
1 | 42 | m | 58.3 | 36.6 |
2 | 44 | m | 5.2 | 4.3 |
3 | 30 | f | 16.8 | 14.7 |
4 | 42 | m | 12.6 | 8.3 |
5 | 60 | f | 12.8 | 12.6 |
6 | 41 | f | 17.4 | 15.1 |
7 | 28 | m | 6.2 | 6.4 |
8 | 47 | f | 7.4 | 6.5 |
9 | 42 | f | 11.1 | 9.2 |
10 | 26 | m | 4.9 | 3.6 |
11 | 39 | f | 8.0 | 7.1 |
12 | 32 | m | 5.9 | 4.3 |
13 | 31 | f | 5.8 | 4.6 |
14 | 70 | m | 36.5 | 32.8 |
15 | 47 | f | 14.1 | 10.9 |
16 | 55 | m | 13.6 | 10.7 |
17 | 70 | m | 35.4 | 29.3 |
18 | 32 | m | 44.1 | 26.9 |
19 | 23 | m | 8.1 | 6.5 |
20 | 29 | m | 6.1 | 6.9 |
21 | 27 | m | 16.7 | 15.0 |
22 | 35 | m | 7.7 | 7.6 |
23 | 37 | m | 6.8 | 6.2 |
24 | 33 | f | 5.3 | 5.9 |
25 | 81 | m | 38.1 | 45.6 |
26 | 33 | m | 6.7 | 6.0 |
27 | 26 | f | 4.5 | 4.7 |
28 | 26 | f | 5.0 | 5.2 |
29 | 23 | m | 4.7 | 5.8 |
30 | 34 | m | 7.0 | 6.2 |
31 | 76 | f | 26.7 | 29.2 |
32 | 26 | m | 5.8 | 6.3 |
f female, m male, pNfL plasma neurofilament light chain, sNfL serum neurofilament light chain.
NfL concentrations in serum and plasma withstand delayed processing up to 7 days
Mean concentrations of sNfL and pNfL across all four pre-freezing intervals are shown in Table 2a (serum) and Table 2b (plasma). After storing of serum and blood tubes for 3 days at 4–8 °C, mean absolute differences compared to immediate freezing were 0.9 pg/mL for sNfL and 0.5 pg/mL for pNfL. Keeping samples stored for 3 days at room temperature yielded a mean absolute difference in concentration compared to immediate freezing of 1.4 pg/mL for sNfL and 0.8 pg/mL for pNfL. Ultimately, storing serum and blood vials for 7 days at room temperature resulted in a mean absolute difference in NfL concentrations compared to immediate freezing of 0.9 pg/mL for sNfL and 0.5 pg/mL for pNfL. Bland–Altman plots for these differences are shown in Fig. 1a–c (serum) and Fig. 2a–c (plasma). The reliability of sNfL and pNfL concentrations at the different pre-freezing intervals in comparison to immediate freezing are shown in Table 3a (serum) and Table 3b (plasma). For all three altered pre-freezing intervals, intraindividual variation (iCOV) was minimal (for samples stored for 7 days at room temperature; serum: 9.0%, plasma: 6.5%) and reproducibility close to perfect (for samples stored for 7 days at room temperature; serum: ICC = 0.99, plasma: ICC = 0.94). Separate analysis for the lowest and highest quartiles revealed still acceptable ICCs for sNfL and pNfL though intraclass correlation was better for serum.
Table 2.
[sNfL] freezing interval: immediate | [sNfL] freezing interval: 3 days RT | [sNfL] freezing interval: 3 days 4–8 °C | [sNfL] freezing interval: 7 days RT | p-valuea | |||||
---|---|---|---|---|---|---|---|---|---|
Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | ||
(a) | |||||||||
Whole cohort | 14.6 | 9.7–19.5 | 15.5 | 9.8–21.2 | 16.0 | 10.3–21.8 | 15.5 | 10.1–20.9 | 0.192 |
Lowest quartile | 5.2 | 4.8–5.6 | 5.3 | 4.4–6.1 | 5.4 | 4.9–5.4 | 5.6 | 4.9–6.4 | 0.633 |
Highest quartile | 34.2 | 22.6–45.7 | 36.8 | 20.8–52.7 | 38.5 | 23.7–53.3 | 36.8 | 23.5–50.1 | 0.422 |
[pNfL] freezing interval: immediate | [pNfL] freezing interval: 3 days RT | [pNfL] freezing interval: 3 days 4–8 °C | [pNfL] freezing interval: 7 days RT | p-valuea | |||||
---|---|---|---|---|---|---|---|---|---|
Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | Mean (pg/mL) | 95% CI | ||
(b) | |||||||||
Whole cohort | 12.0 | 8.1–15.9 | 12.5 | 8.3–16.8 | 12.8 | 8.1–17.5 | 12.5 | 8.4–16.6 | 0.385 |
Lowest quartile | 4.8 | 4.1–5.5 | 4.8 | 4.4–5.3 | 4.3 | 3.7–4.8 | 5.0 | 4.4–5.5 | 0.071 |
Highest quartile | 28.8 | 18.5–39.1 | 30.3 | 19.6–41.1 | 31.6 | 18.1–45.2 | 30.0 | 19.5–40.5 | 0.541 |
Mean values and 95% confidence intervals (95% CI) for (a) serum neurofilament light chain (sNfL) and (b) plasma neurofilament light chain (pNfL).
CI confidence interval, pNfL plasma neurofilament light chain concentration (pg/mL), RT room temperature, sNfL serum neurofilament light chain concentration (pg/mL).
aCalculated by repeated measurement ANOVA.
Table 3.
Freezing interval: 3 days RT | Freezing interval: 3 days 4–8 °C | Freezing interval: 7 days RT | |||||||
---|---|---|---|---|---|---|---|---|---|
iSD [x10e3pg/L] | iCOV (%) | ICC | iSD [x10e3pg/L] | iCOV (%) | ICC | iSD [x10e3pg/L] | iCOV (%) | ICC | |
(a) sNfL | |||||||||
Whole cohort | 1.5 | 10.2 | 0.97 | 1.7 | 9.7 | 0.98 | 1.3 | 9.0 | 0.99 |
0.3–2.6 | 8.2–12.2 | 0.94–0.99 | 0.9–2.6 | 6.5–12.8 | 0.96–0.99 | 0.5–2.0 | 6.5–11.5 | 0.97—0.99 | |
Lowest quartile | 0.5 | 7.0 | 0.90 | 0.4 | 9.5 | 0.91 | 0.5 | 8.1 | 0.90 |
0.1–0.9 | 3.5–11.0 | 0.65–0.94 | 0.2–0.6 | 2.5–16.4 | 0.66–0.94 | 0.1–0.8 | 2.0–14.2 | 0.64–0.93 | |
Highest quartile | 3.4 | 11.1 | 0.92 | 4.3 | 10.2 | 0.94 | 2.7 | 8.5 | 0.95 |
1.5–8.4 | 3.6–18.5 | 0.60–0.98 | 1.2–7.3 | 1.1–21.4 | 0.69–0.99 | 0.5–5.9 | 0.5–16.6 | 0.78–0.99 | |
(b) pNfL | |||||||||
Whole cohort | 1.3 | 11.6 | 0.98 | 1.7 | 8.9 | 0.96 | 0.7 | 6.5 | 0.94 |
0.6–2.0 | 8.1–15.1 | 0.96–0.99 | 0.6–2.7 | 6.5–11.2 | 0.92–0.98 | 0.4–1.1 | 4.8–8.2 | 0.90–0.99 | |
Lowest quartile | 0.4 | 16.3 | 0.90 | 0.6 | 7.8 | 0.90 | 0.5 | 7.6 | 0.88 |
0.2–0.6 | 7.2–25.4 | 0.61–0.92 | 0.2–1.0 | 4.1–11.5 | 0.60–0.93 | 0.3–0.7 | 3.4–11.8 | 0.59–0.91 | |
Highest quartile | 3.4 | 14.2 | 0.91 | 4.5 | 11.9 | 0.86 | 1.7 | 5.6 | 0.84 |
0.6–6.2 | 2.8–25.6 | 0.53–0.98 | 0.2–9.0 | 6.2–17.6 | 0.27–0.98 | 0.1–3.3 | 0.4–10.6 | 0.59–0.90 |
Mean values and 95% confidence intervals for (a) serum neurofilament light chain (sNfL) and (b) plasma neurofilament light chain (pNfL).
ICC intraclass correlation coefficient, iCOV intraindividual coefficient of variance, iSD intraindividual standard deviation, pNfL plasma neurofilament light chain, RT room temperature, sNfL serum neurofilament light chain.
Inter-assay variability
Inter-assay variability within four serum and four plasma samples measured on four consecutive runs was perfect. For sNfL, the iSD was 0.6 pg/mL (range 0.5–1.2), the iCOV 4.8% (range 2.0–6.0) and the ICC 1.0 (range 0.99–1.00). For pNFL, the iSD was 1.6 pg/mL (range 0.8–2.7), the iCOV 12.7% (range 6.0–18.0) and the ICC 0.99 (range 0.90–1.00).
Discussion
Neurofilament light chain is making a run at influencing how we monitor disease course and treatment in neurology. This study investigated a potential limitation for NfL research being constrained to immediate availability and accessibility of standardized sample processing, i.e., a biobanking procedure. We explored whether sNfL and pNfL concentrations remain stable when current standards in sample storage and handling cannot be met. Powered for detecting a measurement error of 2 pg/mL, we found that samples collected and stored at room temperature up to 7 days showed near-perfect stability of NfL concentrations in serum and a little less but still excellent stability in plasma (indicated by ICCs of 0.99 for serum and 0.94 for plasma). Our results are in line with two other studies reporting pre-analytical stability of serum NfL in smaller cohorts making this the largest study confirming this observation for both serum and plasma10,11. With this evidence at hand, we can recommend future practice in NfL research to allow delayed processing of serum or plasma samples. Native tubes can be stored or shipped for up to 7 days at room temperature and still yield reliable results for NfL concentrations.
Emphasis should be placed on the strengths and limitations of this study. First, we produced results for NfL measured in both serum and plasma allowing for a broad application of this knowledge. Second, particular focus was put on the analytical performance characteristics of the NfL assay itself, including precision and repeatability in serum and plasma samples. We provide data for inter-assay validity and found those results comparable to another large study investigating NfL and reporting assay precision15. Third, the results obtained from our study align well within the manufacturer statements regarding the limits of detection13. The mean lower limit of quantification is given as 0.316 pg/mL and the mean lower limit of detection as 0.0552 pg/mL. As for CVs reported by the manufacturer, the within run CVs ranged from 3.3–10.7 and the between run CVs from 5.3–11.7. Variation in our runs were comparably low as indicated by ICCs of 1.0 for serum and 0.99 for plasma. Fourth, the tested pre-freezing interval of 7 days at room temperature is of high practical relevance for the future. Once NfL enters clinical routine as a biomarker, we could expect peripheral hospitals to ship their native samples to a core facility.
There may be some limitations concerning the generalizability of this study. We used Greiner Bio-One Vacuette serum tubes as they are the most widely used product in our country. Other collection systems may produce different results. That being said, this issue per se applies to every biomarker study. Furthermore, the validity of our results depends on meticulous quality control in a specialized biomarker laboratory. A real-world setting in a busy lab is expectedly more prone to accidental protocol deviation, human error, and other preanalytical errors such as hemolysis or the fact that various analyses are performed on a Quanterix reader for different analytes at the same time. Moreover, our results relate to a population showing sNfL concentrations ranging from 4.5 to 58.3 pg/mL and may not be applied to higher concentrations.
A few remarks should be made on comparing the analytical performance of serum and plasma samples in this pre-analytical study. First, the mean concentration of NfL in plasma was 10% lower (95% CI 5–16) than in serum which compares well to another investigation10. Second, our results suggest that reliability for prolonged storage was slightly better in serum than plasma (ICC 0.99 vs. 0.94). Possible explanations for this observation may constitute an interference of persisting coagulation factors through EDTA. Therefore, one may conclude that native serum samples may be slightly more suitable for shipment or any case of delayed processing for up to 7 days.
In summary, we demonstrated how sNfL and pNfL concentrations remain stable if sample are left at room temperature up to 7 days before being processed at standardized laboratory conditions. Our results may help NfL to further evolve as a biomarker in neurological disease as future protocols can account for pre-analytical stability in terms of delayed sample storage.
Acknowledgements
The authors would like to thank Ms. Sylvia Baeumel, Mr. Martin Krutzler, and Ms. Elke Hohenbichler (outpatient clinic, Department of Neurology, Medical University of Vienna, Austria) for helping with this study.
Abbreviations
- sNfL
Serum neurofilament light chain
- pNfL
Plasma neurofilament light chain
- NfL
Neurofilament light chain
- EDTA
Ethylenediaminetetraacetic acid
- CI
Confidence interval
- iSD
Intraindividual standard deviation
- iCOV
Intraindividual coefficient of variation
- ICC
Intraclass correlation coefficient
Author contributions
Study conception and design: P.A., M.P., G.B.. Acquisition of data and biobanking: P.A., F.L., P.S.R., H.H., P.M., C.W. Analysis and interpretation of data: M.P., R.L., A.P., H.Z. Statistical analyses: G.B., P.A. Writing of the manuscript: P.A., M.P., G.B. Review and revision of the manuscript: T.B., H.Z., A.P., G.B. All authors have accepted responsibility for the entire content of this submitted manuscript and approved submission. They have read the journal's policy on disclosure of potential conflicts of interest.
Funding
This study was supported by the Medical University of Vienna.
Competing interests
There are no direct conflicts of interest with respect to this study. This study was not influenced by a third party. PA has participated in meetings sponsored by, received speaker honoraria or travel funding from Biogen, Merck, Roche, Sanofi-Genzyme and Teva, and received honoraria for consulting from Biogen. He received a research grant from Quanterix International and received funding for the development of a smartphone application for a clinical study from Biogen, Merck, Roche, Sanofi-Genzyme and Teva. MP has nothing to disclose. PSR has received honoraria for consultancy/speaking from AbbVie, Alexion, Almirall, Biogen, Merck, Novartis, Roche, Sandoz, Sanofi-Genzyme, has received research grants from Amicus, Biogen, Merck and Roche. HH received an infrastructural grant from the Austrian Federal Ministry for Science, Research and Education (BBMRI.at) and compensations for Biobanking services in the framework of clinical trials from Glock Health and Blue-Skye Vaccines. PM has nothing to disclose. FL has participated in meetings sponsored by or received honoraria for acting as an advisor/speaker for Bayer, Biogen, Celgene, MedDay, Merck, Novartis, Roche, Sanofi-Genzyme and Teva. AP is a member of the steering committee for the OCTiMS study (Novartis) and ARI network (Zeiss), for which he did not receive any consulting fees. He performed quality control on optical coherence tomography for the Passos study (Novartis), for which he received consulting fees. He received speaker fees from Heidelberg Engineering. CW has nothing to disclose. RL received conference speaker honoraria within the last three years from Bruker BioSpin MR, and support from Siemens Healthcare regarding clinical research using PET/MR. He is a shareholder of the start-up company BM Health GmbH since 2019. TB has participated in meetings sponsored by and received honoraria (lectures, advisory boards, consultations) from pharmaceutical companies marketing treatments for MS: Allergan, Bayer, Biogen, Bionorica, Celgene, MedDay, Merck, Novartis, Octapharma, Roche, Sanofi-Genzyme, Teva. His institution has received financial support in the past 12 months by unrestricted research grants (Biogen, Bayer, Merck, Novartis, Sanofi Aventis, Teva and for participation in clinical trials in multiple sclerosis sponsored by Alexion, Bayer, Biogen, Merck, Novartis, Octapharma, Roche, Sanofi-Genzyme, Teva. HZ has served at scientific advisory boards for Denali, Roche Diagnostics, Wave, Samumed and CogRx, has given lectures in symposia sponsored by Fujirebio, Alzecure and Biogen, and is a co-founder of Brain Biomarker Solutions in Gothenburg AB, a GU Ventures-based platform company at the University of Gothenburg, all unrelated to the work presented in this paper. GB has participated in meetings sponsored by, received speaker honoraria or travel funding from Biogen, Celgene, Merck, Novartis, Sanofi-Genzyme and Teva, and received honoraria for consulting Biogen, Roche and Teva. The other authors report no competing interests or any financial or personal relationships with organizations that could potentially be perceived as influencing the described research.
Footnotes
The original online version of this Article was revised: The original version of this Article contained errors, where the intraindividual coefficient of variation (iCOV) was calculated incorrectly. Modifications have been made to the Results and Table 3. Full information regarding the corrections made can be found in the correction for this Article.
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Change history
4/20/2022
A Correction to this paper has been published: 10.1038/s41598-022-10916-3
References
- 1.Gaetani L, Blennow K, Calabresi P, Di Filippo M, Parnetti L, Zetterberg H. Neurofilament light chain as a biomarker in neurological disorders. J. Neurol. Neurosurg. Psychiatry. 2019;90:870–881. doi: 10.1136/jnnp-2018-320106. [DOI] [PubMed] [Google Scholar]
- 2.Khalil M, Teunissen CE, Otto M, Piehl F, Sormani MP, Gattringer T, et al. Neurofilaments as biomarkers in neurological disorders. Nat. Rev. Neurol. 2018;14:577–589. doi: 10.1038/s41582-018-0058-z. [DOI] [PubMed] [Google Scholar]
- 3.Altmann P, De Simoni D, Kaider A, Ludwig B, Rath J, Leutmezer F, et al. Increased serum neurofilament light chain concentration indicates poor outcome in Guillain–Barré syndrome. J. Neuroinflammation. 2020;17:86. doi: 10.1186/s12974-020-01737-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Bsteh G, Berek K, Hegen H, Teuchner B, Buchmann A, Voortman MM, et al. Serum neurofilament levels correlate with retinal nerve fiber layer thinning in multiple sclerosis. Mult. Scler. J. 2016;22:1550–1559. doi: 10.1177/1352458515623365. [DOI] [PubMed] [Google Scholar]
- 5.Disanto G, Barro C, Benkert P, Naegelin Y, Schädelin S, Giardiello A, et al. Serum Neurofilament light: A biomarker of neuronal damage in multiple sclerosis. Ann. Neurol. 2017;81:857–870. doi: 10.1002/ana.24954. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Zetterberg H. Neurofilament light: A dynamic cross-disease fluid biomarker for neurodegeneration. Neuron. 2016;91:1–3. doi: 10.1016/j.neuron.2016.06.030. [DOI] [PubMed] [Google Scholar]
- 7.Lambertsen KL, Soares CB, Gaist D, Nielsen HH. Neurofilaments: The C-reactive protein of neurology. Brain Sci. 2020;10:56. doi: 10.3390/brainsci10010056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Jens K, Christian B, Ulf A, Tobias D, Raija L, Åsa S, et al. Comparison of three analytical platforms for quantification of the neurofilament light chain in blood samples: ELISA, electrochemiluminescence immunoassay and Simoa. Clin. Chem. Lab. Med. (CCLM) 2016;54:1655–1661. doi: 10.1515/cclm-2015-1195. [DOI] [PubMed] [Google Scholar]
- 9.Rissin DM, Kan CW, Campbell TG, Howes SC, Fournier DR, Song L, et al. Single-molecule enzyme-linked immunosorbent assay detects serum proteins at subfemtomolar concentrations. Nat. Biotechnol. 2010;28:595–599. doi: 10.1038/nbt.1641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hviid CVB, Knudsen CS, Parkner T. Reference interval and preanalytical properties of serum neurofilament light chain in Scandinavian adults. Scand. J. Clin. Lab. Investig. 2020 doi: 10.1080/00365513.2020.1730434. [DOI] [PubMed] [Google Scholar]
- 11.Altmann P, Leutmezer F, Zach H, Wurm R, Stattmann M, Ponleitner M, et al. Serum neurofilament light chain withstands delayed freezing and repeated thawing. Sci. Rep. 2020;10:19982. doi: 10.1038/s41598-020-77098-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Haslacher H, Gerner M, Hofer P, Jurkowitsch A, Hainfellner J, Kain R, et al. Usage data and scientific impact of the prospectively established fluid bioresources at the hospital-based MedUni Wien biobank. Biopreserv. Biobank. 2018;16:477–482. doi: 10.1089/bio.2018.0032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Quanterix. Data Sheets and Technical Notes for NF-light. https://www.quanterix.com/products-technology/assays/nf-lightr-sr-x-version (Accessed September 2020).
- 14.Kramer MS, Feinstein AR. Clinical biostatistics: LIV. The biostatistics of concordance. Clin. Pharmacol. Ther. 1981;29:111–123. doi: 10.1038/clpt.1981.18. [DOI] [PubMed] [Google Scholar]
- 15.Khalil M, Pirpamer L, Hofer E, Voortman MM, Barro C, Leppert D, et al. Serum neurofilament light levels in normal aging and their association with morphologic brain changes. Nat. Commun. 2020;11:812. doi: 10.1038/s41467-020-14612-6. [DOI] [PMC free article] [PubMed] [Google Scholar]