Design and Numerical Analysis of a Graphene-Coated SPR Biosensor for Rapid Detection of the Novel Coronavirus

Tarik Bin Abdul Akib 1, Samia Ferdous Mou 1, Md Motiur Rahman 1, Md Masud Rana 1, Md Rabiul Islam 2, Ibrahim M Mehedi 3, M A Parvez Mahmud 4,*, Abbas Z Kouzani 4
Editor: Alexander Star
PMCID: PMC8156410  PMID: 34067769

Abstract

In this paper, a highly sensitive graphene-based multiple-layer (BK7/Au/PtSe2/Graphene) coated surface plasmon resonance (SPR) biosensor is proposed for the rapid detection of the novel Coronavirus (COVID-19). The proposed sensor was modeled on the basis of the total internal reflection (TIR) technique for real-time detection of ligand-analyte immobilization in the sensing region. The refractive index (RI) of the sensing region is changed due to the interaction of different concentrations of the ligand-analyte, thus impacting surface plasmon polaritons (SPPs) excitation of the multi-layer sensor interface. The performance of the proposed sensor was numerically investigated by using the transfer matrix method (TMM) and the finite-difference time-domain (FDTD) method. The proposed SPR biosensor provides fast and accurate early-stage diagnosis of the COVID-19 virus, which is crucial in limiting the spread of the pandemic. In addition, the performance of the proposed sensor was investigated numerically with different ligand-analytes: (i) the monoclonal antibodies (mAbs) as ligand and the COVID-19 virus spike receptor-binding domain (RBD) as analyte, (ii) the virus spike RBD as ligand and the virus anti-spike protein (IgM, IgG) as analyte and (iii) the specific probe as ligand and the COVID-19 virus single-standard ribonucleic acid (RNA) as analyte. After the investigation, the sensitivity of the proposed sensor was found to provide 183.33°/refractive index unit (RIU) in SPR angle (θSPR) and 833.33THz/RIU in SPR frequency (SPRF) for detection of the COVID-19 virus spike RBD; the sensitivity obtained 153.85°/RIU in SPR angle and 726.50THz/RIU in SPRF for detection of the anti-spike protein, and finally, the sensitivity obtained 140.35°/RIU in SPR angle and 500THz/RIU in SPRF for detection of viral RNA. It was observed that whole virus spike RBD detection sensitivity is higher than that of the other two detection processes. Highly sensitive two-dimensional (2D) materials were used to achieve significant enhancement in the Goos-Hänchen (GH) shift detection sensitivity and plasmonic properties of the conventional SPR sensor. The proposed sensor successfully senses the COVID-19 virus and offers additional (1 + 0.55) × L times sensitivity owing to the added graphene layers. Besides, the performance of the proposed sensor was analyzed based on detection accuracy (DA), the figure of merit (FOM), signal-noise ratio (SNR), and quality factor (QF). Based on its performance analysis, it is expected that the proposed sensor may reduce lengthy procedures, false positive results, and clinical costs, compared to traditional sensors. The performance of the proposed sensor model was checked using the TMM algorithm and validated by the FDTD technique.

Keywords: biosensor, coronavirus, COVID-19, molecular detection, rapid detection, SARS-CoV-2, sensor, spike receptor-binding domain, surface plasmon resonance

1. Introduction

The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is a newly reported human transferrable virus [1]. To date, more than 160 million positive cases of SARS-CoV-2 have been identified globally, causing about 3.25 million deaths [1]. The World Health Organization (WHO) declared the outbreak of the human transferrable SARS-CoV-2 virus a global pandemic [1]. Typically, the SARS-CoV-2 is a single-stranded RNA virus containing four significant proteins: (i) spike (S) glycoprotein, (ii) envelope (E), (iii) membrane (M), and (iv) nucleocapsid (N) protein [2]. The spike proteins are two types of subunit (S1, S2) protein, where S1 binding with the host cell receptor human angiotensin-converting enzyme 2 (ACE2) and S2 is liable for membrane fusion [3,4]. ACE2 is found in the human lung, kidneys, heart, and various organs, which allows the virus spike protein to enter the cell [2]. The activation of the virus spike protein causes massive damages in the brain, lung, heart, and kidneys [5]. Thus, neutralization of SARS-CoV-2 virus spike protein is a vital target in clinical research for the invention of vaccines and the development of detection procedures to reduce the spread of infectious diseases. There have been more than 58 vaccines invented to neutralize the SARS-CoV-2 virus spike receptor-binding domain (RBD) in clinical trials [6]. Some vaccines have more than 90% effectiveness against the SARS-CoV-2 virus [6,7]. The COVAX (founded by WHO, CEPI, and GAVI) is working to accelerate the development, commercial manufacture, and equitable access of SARS-CoV-2 vaccines [6]. Yet, SARS-CoV-2 viruses are progressively growing—in the UK, a new variant of the coronavirus has been identified with 17 potential changes or mutations [8]. Vital changes in an N501Y mutation in the SARS-CoV-2 virus spike protein bind with the human ACE2 receptor, and this new variant virus is a 70% more effective human transferrable virus [8]. Since the vaccine is not commercially available everywhere and with unprecedented demand for rapid detection kits to detect the SARS-CoV-2 virus, early detection and supervision are vital for controlling the pandemic. It is essential to offer tremendously sensitive, quick-test results, and low-price analytical tools to monitor affected persons for effective quarantine and timely treatment.

At present, the real-time reverse-transcriptase-polymerase chain reaction (RT-PCR) technique is used as a reference method to detect the SARS-CoV-2 virus. The RT-PCR technique takes 1–3 days to get results (generally, the test period is about 4 h) [9]. The virus ribonucleic acid (RNA) extraction is a lengthy process that affects detection accuracy. To exceed the long process of RNA extraction, Zhao et al. established a pcMNPs technique for virus RNA extraction (the test period is about ~30 min) [10]. The pcMNPs for RNA extraction is willingly announced as the subsequent RT-PCR RNA extraction [10]. The RT-PCR sensitivity ranges among 45–60% for RNA extraction; the sensitivity is increased 90–95% using pcMNPs with the specificity of identification, and enhances scientific investigation [9,10]. In the rapid test method, the COVID-19 IgG/IgM rapid test kit is used clinically to detects antibodies (IgG/IgM), where IgG is sensible after 3–6 days and IgM sensible after 8 days in affected patient blood [11,12,13]. The SARS-CoV-2 antigen rapid test kit is also applied clinically to detect the COVID-19 virus but less sensitive than the reverse-transcriptase-polymerase chain reaction (RT-PCR) technique [14,15]. Additionally, the test results are confirmed after checking the chest computed tomography (CT) scan. The rapid test kit provides false results without proper handling [16,17]. In reference [18,19,20,21], an SPR-originated fast, accurate, and highly sensitive biosensor was proposed to detect the COVID-19 virus in the future. SPR is a conventional technique applied over the last three decades to detect biomolecular interactions for clinical or research purposes [22,23]. There are numerous industrially developed SPR sensors, such as compact SPR (CSPR) biosensor, optical sensing SPR (OSSPR) biosensor, and localized SPR (LSPR) biosensor, used in biosensing applications [24]. There is no crucial difference in operating principles between the biosensors. The SPR sensor is well-known as a bio-analytical process for real-time detection of ligand and analyte binding onto the sensing region [23]. The refractive index (RI) varies because of the binding of different concentrations of the analyte [25].

In this paper, a graphene-coated five-layer SPR biosensor was proposed for the rapid detection of coronavirus. The performance of the proposed senor was analyzed numerically, ensuring early-stage rapid detection and reduction in processing time. It is expected that the proposed sensor will be tested experimentally in the near future to identify the COVID-19 virus practically. The motive of the proposed sensor is to detect the SARS-CoV-2 virus spike protein rapidly without false reports. The performance of the proposed sensor has been investigated numerically with different ligand–analytes such as antibodies–spike protein, spike protein–anti-spike protein, and probe RNA–virus single-standard RNA. The graphene implicated specific ligands can be immobilized with the target analytes. The proposed sensor can specifically detect the COVID-19 virus spike protein, antibodies against the virus spike protein, and virus single-standard RNA. Moreover, by adding a graphene film onto the sensor surface, the sensor’s sensitivity is enhanced enormously. The modified graphene coating sheet with specific ligands or gene sequences sensing transduction offers an alternative optimistic result for the experimental SARS-CoV-2 diagnosis. The main contributions of this article are as follows:

  • A five-layer SPR biosensor is proposed for rapid detection of COVID-19 virus and to tackle the lengthy, expensive, false-positive results problems;

  • 2D materials are employed to achieve significant enhancement in the GH shift detection sensitivity and plasmonic properties of the proposed sensor;

  • A graphene layer is used to increase interaction with the ligands. The ligands can be immobilized with a graphene film through 1-pyrene butyric acid n-hydroxy-succinimide ester (PBSE) and can interact with target analytes;

  • The performance of the proposed sensor is investigated with different ligand–analytes to identify the COVID-19 virus. The proposed sensor can detect the COVID-19 virus spike protein and anti-spike protein rapidly. However, the detection of virus single-standard RNA is a lengthy process because of the RNA extraction process;

  • Finally, the comprehensive simulation results are analyzed during the adsorption of target analytes. The results show the detection accuracy, sensitivity, and superiority of the proposed sensor for the early-stage detection of COVID-19.

The paper is organized into five sections. The proposed sensor design model, fabrication process, FDTD method, and TMM algorithm for the proposed sensor are described in Section 2. The quality enhancement of the sensor is briefly described in Section 3. Simulation results and analysis for detection of the SARS-CoV-2 virus are represented in Section 4, and the conclusion is given in Section 5.

2. Methodology

Coronavirus samples (virus spike RBD or RNA) and anti-spike protein (IgG or IgM) can be collected from nasopharyngeal swabs (NPS) and blood, respectively. They are placed immediately into viral transport media (VTM) kits, as illustrated in Figure 1.

Figure 1.

Figure 1

Schematic diagram of diagnosis biomarkers: (a) SARS-CoV-2 virus spike protein RBD, (b) IgG or IgM antibodies, and (c) or RNA-oligonucleotides are collected from nasopharyngeal swabs or human blood.

The optimization method used in [12,26,27,28] can be used for the synthesis and purification of SARS-CoV-2 spike, and human monoclonal antibodies (IgG or IgM). The SARS-CoV-2 spike or antibodies (IgG or IgM) can flow with the phosphate-buffered saline (PBS) onto the sensor surface as an analyte [12,29]. The proposed sensor can detect the COVID-19 virus by utilizing the samples in three different modes: (i) the mAbs as ligand and the virus spike RBD as analyte, (ii) the virus spike RBD as ligand and the virus anti-spike protein as analyte and (iii) the specific probe as ligand and the virus single-standard RNA as analyte onto sensing region. In the rapid detection process of (i) and (ii), the probe ligand can be immobilized onto the graphene film through PBSE [30]. In the process of (iii), this sensor can identify the hybridization event between the probe and the virus RNA sequences. The virus RNA detection process is lengthy because of the RNA extraction process. The RI is increased due to the different concentrations of ligand–analyte binding events. The detection biomarker varies with the variation of RI and concentration. The change in reflectance (R), transmittance (T) intensity, a shift in SPR angle (θSPR), and SPRF is calculated to recognize the presence of the COVID-19 virus.

2.1. The Design of the Proposed Sensor

The proposed SPR sensor for the rapid detection of the SARS-CoV-2 virus is illustrated in Figure 2. The five-layer sensing region of the sensor is based on the Kretschmann-Raether configuration. The monochromatic He–Ne laser light incident at an acceptance angle onto prism (Bk7) and CCD (charge-coupled device) or CMOS (complementary metal oxide semiconductor) is a monitoring device. In the sensing region, the first layer is Bk7 and the second layer is a thin gold Au film; RI of Au is deliberated as nAu=0.1726+3.4218 i and layer thickness of dAu=50 nm [22,23]. Au film placed on Bk7 prism coupler, RI of Bk7 prism is nBK7= 1.5151. The third layer is platinum-di-selenide PtSe2, RI of PtSe2 is deliberated as nPtSe2= 2.9189+0.9593 i and coating thickness of dPtSe2= 2 nm [31]. The fourth layer is graphene, RI of the graphene layer is deliberated as ng= 3+1.1491 i and coating thickness of dg= 0.34×L nm, where L is the number of graphene layers [23]. The PBS (pH~7.4) can be used as a VTM, RI of PBS is deliberated as ns = 1.3348+Δn, where Δn is fluctuating, due to the ligand-analyte interaction on the sensing surface.

Figure 2.

Figure 2

Schematic diagram of the five-layered (Bk7/Au/PtSe2/Graphene/PBS) SPR biosensor for diagnosis of SARS-CoV-2 cultured virus; three operating modes are represented to detect the SARS-CoV-2 virus: (a) rapid recognition of whole virus spike RBD with immobilized human mAbs (mAbs as ligand and spike RBD as analyte), (b) rapid recognition of mAbs with immobilized virus spike RBD (spike RBD as ligand and mAbs as analyte), and (c) or recognition of the virus RNA sequence with immobilized probe sequence onto the graphene implicated sensor surface.

To be specific, the purpose of the BK7 is the highest accumulation of the incident light at the metal surface and to generate a surface plasmon wave (SPW) on the surface of the metal-dielectric interface. A gold layer and highly sensitive 2D materials (PtSe2 and Graphene) are employed to form plasmonic sensing substrates [32,33]. The PtSe2 layer is an emerging 2D group of 10 TMDC (transition metal dichalcogenides) that has intriguing optical attributes, tunable bandgap, phase transition, and superior electron mobility [34]. The graphene layer is also used for high conductivity, high carrier mobility, and superior interaction quality with the ligands. The PtSe2 and graphene layer synthesis enhanced the Goos-Hänchen (GH) shift detection sensitivity [34]. Therefore, the heterostructure metal interface (Au/PtSe2/Graphene) is used to achieve significant enhancement in detection sensitivity and plasmonic properties of the conventional SPR sensor [33,34,35].

2.2. The Fabrication Process of the Proposed Sensor

The possible fabrication technique of the proposed sensor for future development can be accomplished in two steps, which are illustrated in Figure 3. To demonstrate, in the first step, a piece of BK7 glass substrates is taken that has a x span of 7.95 μm, y span of 2 μm, and z span of 1 μm. The BK7 substrate is then washed in piranha solution (H2O2:3 H2SO4) to eliminate any pollutants. Then, the Au layer is grown on the top of the BK7 substrates using the physical vapor deposition (PVD) or sputtering technique [36,37]. The thickness of the Au layer relies on the particle sputtering deposition time [38]. The PVD method is usually renowned for representing corrosion, attire resistance, and esthetic features of the implicating films, which could be tuned on requirement [39].

Figure 3.

Figure 3

Schematic diagram of the possible multi-layer film fabrication process of the proposed SPR sensor for detection of the SARS-Cov-2 virus;—in step 1, the sensor Bk7/Au/PtSe2/graphene layers are deposited sequentially, the synthesis technique of PtSe2 and graphene film are also depicted. In step 2, the ligands;—(a) mAbs, (b) spike RBD, and (c) probe oligo are immobilized with the graphene layer through PBSE.

The third layer can be formed of PtSe2 by (i) sputtering the platinum (Pt) onto the SiO2 substrate and (ii) SiO2/Pt layer can be selenized with selenium (Se) steam (at 400 °C under 150 sccm of 10% H2/Ar flow) by thermally assisted conversion (TAC) for two hours [40,41,42,43]. (iii) The polymethyl methacrylate (PMMA) layer can be assisted with a SiO2/Pt stack in a spin-coated process to transfer the PtSe2 film onto the Au layer. (iv) The SiO2 layer can be removed by a wet etching process using 2 M NaOH at 25 °C. (v) The developed PMMA/PtSe2 stack can be cleaned in de-ionized (DI) water, transferred onto the BK7/Au substrate, and PMMA can be removed by applying acetone ((CH3)2CO) [40,44,45,46]. Then, the possible graphene film synthesis onto the PtSe2 layer includes the following steps [47,48,49,50,51]: (i) A high-quality graphene film (5 layers × 0.34 nm = 1.7 nm) can be grown onto the copper (Cu) foil by chemical vapor deposition (CVD) method at 1000 °C under a chamber pressure of 3.6 Torr by using methane (CH₄) gas as a carbon source for five minutes [47,48,49,50,51]. The CH₄ and Ar/H2 gas flow ratio can be maintained by operating the control valve [47]. (ii) The PMMA layer can be assisted with a Cu/graphene stack at a spin-coated process (spin rate 2000 r/min, at 175 °C) for five minutes to transfer the graphene film onto the target substrate. (iii) The Cu foil can be etched by the sulfuric acid (H2SO4) at 25 °C. (iv) The developed PMMA/graphene stack can be cleaned in DI water, transferred onto the BK7/Au/PtSe2 substrate, and PMMA can be removed by applying (CH3)2CO. In step 2, the possible immobilization process of the specific probe onto the graphene film for future development is as follows: A specific probe (sh) and the mismatching (wt) oligonucleotide solution can be taken, which has an equal concentration of 0.3 µM, and the target matching (mr) oligonucleotide solution has a concentration from 0.15 µM to 0.3 µM. The PBSE can be dissolved in (CH3)2SO (DMSO) to prepare a proper linker reagent that has a concentration of 5 mM. For possible immobilization of PBSE onto the graphene layer, the linker reagent can be added onto the graphene film for two hours at 25 °C. The immobilized graphene film by PBSE can be dissolved in probe oligonucleotide for four hours at 25 °C to confirm enough conjugation among the probe and the PBSE [52]. Similarly, the possible immobilization of antibody or spike protein onto the graphene film layer is as follows: The graphene film can be moistened in 2 mM PBSE in CH₃OH for one hour at 25 °C and cleaned many times with PBS (pH 7.4) solution with DI water. After that, the functionalized graphene film can be exposed to 250 μg/mL ligands (mAbs, or Spike RBD) for 4 h [30].

2.3. FDTD Technique for the Proposed Sensor

The finite-difference time-domain (FDTD) technique is used to evaluate electromagnetic field analysis for the proposed sensor by using commercial Lumerical FDTD software. FDTD technique is generally solving Maxwell’s equations using YEE-algorithms [53]. In this simulation, the non-uniform mesh is used to increase accuracy, minimizing numerical dispersion [54].

The sensor surface plasmon polaritons (SPPs) excitation is evaluated by applying angular interrogation technique at a 633 nm plane wavelength [53,55]. In the FDTD simulation region, the perfectly matched layer (PML) absorbing boundary conditions are used to absorb source light with minimum reflections, and the DFT transmission monitor is used to estimate the transmittance. The simulation time setting is set to 1000 (fs) at 26.5 °C temperature, background refractive index is set to 1.00, and incident angle sweep from 30° to 70° angle. In a three-layer sensor model, BK7 has a geometrical position from −8 μm to −50 nm (x span of 7.95 μm), Au layer from −50 m to 0 nm (x span of 50 nm) linear to the x-axis. In a four-layer model, the graphene layer is added with the three-layer model; it has a geometrical location from 0 nm to 1.7 nm (span of 1.7 nm) linear to the x-axis. In addition, the developed five-layer model has BK7 with a geometrical position from −8 μm to −50 nm (x span of 7.95 μm), Au layer from −50 m to 0 nm (x span of 50 nm), PtSe2 layer from 0 nm to 2 nm (x span of 2 nm), and graphene layer from 2 nm to 3.7 nm (x span of 1.7 nm) linear to the x-axis. The developed sensor design in FDTD software is depicted in Figure 4, and the simulation result is manifested in Section 3.

Figure 4.

Figure 4

FDTD simulation schematic for the proposed SPR sensor.

2.4. Transfer Matrix Algorithm

According to the best performance of the proposed SPR sensor at a wavelength (λlight) of 633 nm, monochromatic laser light is applied [16,56]. The reflection intensity of p-polarized incident light can be denoted as [22,53,57,58]:

Rp=rp2 (1)
where,rp=F11+F12nNn1 F21+F22nNF11+F12nNn1 +F21+F22nN (2)

The proposed sensor works on the principle of the TIR technique; however, for analyzing the multi-layer sensor numerically or to determine the reflection and transmission intensity of the proposed sensor, the TMM algorithm is encrypted in MATLAB software. For the multi-layer coated sensor, Fij represents the transfer matrix features as follows [22,57,58]:

Fij=K=2N1cosβkisinβknkinksinβkcosβkij=F11F12F21F22 (3)
where,βk=2πdkλεkn12 sin2θ11/2θk=cos1εkn12 sin2θ11/2ϵqk=μkεk1/2

In (3), εk is the dielectric constant, βk is the random phase constant, nk is the RI, θk is the angle of entrance, and dk is the depth of kth layer. The propagation constant (kspw) of SPW is changed with the immobilization of analytes in the sensing region. This propagation constant can be expressed as [25,56,59]:

kspw=2πλlightnAu2nPtSe22ng2nAu2+nPtSe22+ng23 (4)

The propagation constant point is shifted due to binding with the analyte as a result of the impact on surface plasmon resonance frequency (SRF). The Fresnel concept is applied in this proposed five-layered SPR sensor model to estimate transmittance intensity versus the SPRF curve. SPRF can be denoted as [22,56]:

SRF= C×kspw2π nAu2nPtSe22ng23 (5)

The ratio of reflectance and incidence angle is familiar as the SPR curve. The incident angle is recognized as the θSPR, which can be expressed as [23,60]:

θSPR=sin11nBK7nAu2nPtSe22ng2nAu2+nPtSe22+ng23 (6)

In (4)–(6), De Moivre’s method [61] is applied to calculate the nth roots of a complex number. In (7), ΔθSPR and Δns is the change of SPR angle and RI, respectively. The resonance wavelength Δλres and the spectral resolution Δλdr is used to calculate Δn [62]. The most commonly used mathematical formulas are as follows:

Δn=Δns×ΔλdrΔλresS=ΔθSPRΔns= ΔλresΔns FWHM=Δθ0.5 DA=1Δθ0.5FOM= S ×DASNR= ΔθSPRFWHMQF=SNR×S (7)

Finally, the proficiency of the SPR sensors is categorized on the basis of sensitivity (S), full width at half maximum (FWHM), DA, FOM, SNR, and QF parameters [31,63].

3. Quality Enhancement of the Proposed Sensor

In the beginning, the proposed hybrid sensor performance was analyzed at different excitation wavelengths. Commercially available SPR biosensors such as Biacore-3000, SPRm-200, and Spreeta-2000 operate at a wavelength of 690 nm, 780 nm, and 830 nm, respectively [64,65]. The RI of sensor materials changes during the different excitation wavelengths, which is tabulated in Table 1 [34]. In Figure 5, the proposed sensor is operated at four different widely used wavelengths of 532 nm, 632.8 nm, 780 nm, and 1152 nm. The sensor SPR (θSPR~Rmin), SRF SRF~Tmax curve characteristics, and sensitivity at different operating wavelengths are tabulated in Table 2. The table highlights that the maximum sensitivity of the sensor is 200°/RIU and 1000THz/RIU at a 632.8 nm wavelength. Therefore, the proposed sensor is operated at a wavelength of 632.8 nm He-Ne laser light.

Table 1.

RI of sensor materials at different wavelengths.

RI λ1 (532.0 nm) λ2 (632.8 nm) λ3 (780.0 nm) λ4 (1152 nm)
BK7 [34] 1.5195 1.5151 1.5112 1.5055
Au [34,66] 0.0990 + 2.9952i 0.1377 + 3.6183i 0.2063 + 4.5133i 0.4417 + 6.7308i
PtSe2 [34,67] 2.5917 + 1.1332i 2.9029 + 0.8905i 2.8229 + 0.5277i 2.7986 + 0.2878i
Graphene [34,68] 3 + 0.9658i 3 + 1.1487i 3 + 1.4160i 3 + 2.0913i

Figure 5.

Figure 5

Schematic diagram of: (a) SPR and (b) SPRF curve characteristics of Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene(1.7 nm)/PBS (RI = 1.3348) layered model with variation in incident light wavelength λlight.

Table 2.

Data evaluation of ΔRminλ, ΔθSPRλ, ΔTmaxλ and ΔSRFFreqλ, and sensitivity (S) attributors with change of incident wavelength λlight in the proposed SPR sensor.

Incident Light Wavelength, (λ) [nm] Rminλ [%] θSPRλ [deg.] Tmaxλ [dB] SRFfreqλ[THz] ΔRminλ [%]=Rmin532 nmRminλ ΔθSPRλ [deg.]=θSPR532 nmθSPRλ ΔTmaxλ [dB]=Tmax532 nmTmaxλ ΔSRFFreqλ [THz]=SRFFreq532 nmSRFFreqλ Sensitivity (S)
deg/RIU THz/RIU
λ1= 532.0 nm 74.56 80.70 13.689 1691.2 0.00 0.00 0.0000 0.00 50 100
λ2= 632.8 nm 31.22 76.25 3.7420 1399.6 43.34 4.45 9.9470 291.60 200 1000
λ3= 780.0 nm 15.04 70.05 1.6288 1098.6 59.52 10.65 12.0602 592.60 100 900
λ4= 1152 nm 19.12 65.70 2.1217 721.29 55.44 15.00 11.5673 969.91 100 570

As a second factor, the proposed SPR sensor is compared with other conventional sensors, as shown in Figure 6. In addition, Table 3 shows the evaluated results for minimum reflectance (Rmin), SPR angle (θSPR), maximum transmittance Tmax, surface resonance frequency (SRF), sensitivity, FWHM, DA, FOM, and QF for three different structures of the SPR sensor model (SM). In a three-layer model (Bk7/Au/PBS) without a graphene layer, Rmin is 0.54%, θSPR is 72.60°, SRF is 104.19 THz, and Tmax is 0.054 dB. Similarly, in a four-layer graphene-coated model (Bk7/Au/Graphene/PBS) where Rmin is 14.8%, θSPR is 75.05°, SRF is 1391.5 THz, and Tmax is 1.601 dB. Eventually, the proposed sensor will be covered with a total of five layers containing PtSe2 and a graphene film layer (Bk7/Au/PtSe2/Graphene/PBS), where Rmin is 35.08%, θSPR is 78.10°, SRF is 1409.3 THz and Tmax is 4.320 dB. Equation (7) is used to calculate the sensitivity, where Δns = 0.03 RIU. The proposed sensor’s sensitivity is enhanced to 183.3 deg RIU−1. The proposed five-layer (S5) sensor sensitivity is 6.50 times higher than that of the three-layer (S3) conventional sensor and the calculation as in (8).

SM=ΔθSPRΔns              =ΔθSPRS3+ΔθSPRS5Δns                         =ΔθSPRS3+5.50×ΔθSPRS3Δns                =1+5.50×ΔθSPRS3Δns           =6.50×S3 (8)

Figure 6.

Figure 6

Schematic diagram of (a) SPR and (b) SPRF curve characteristics of the Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene (1.7 nm) sensor with other conventional sensors with PBS analyte (RI = 1.3348).

Table 3.

Evaluation of the proposed sensor model with other conventional models with regards to sensitivity (S), FWHM, detection accuracy (DA), the figure of merit (FOM), and quality factor (QF).

Structure (SM) RminS
%
θSPRS
deg
TmaxS
dB
SRFfreqS
THz
ΔRminS % ΔθSPRS deg ΔTmaxS dB ΔSRFfreqS
THz
S
deg/RIU
FW-HM
deg
DA
1/deg
FOM
[RIU1]
QF
deg/RIU
Bk7/Au(50 nm)/PBS 0.54 72.60 0.054 104.19 - - - - - - - - -
Bk7/Au(50 nm)/Graphene (5 layer)/PBS 14.80 75.05 1.601 1391.5 14.26 2.46 1.547 1287.31 82.0 1.30 0.77 63.07 155.17
Bk7/Au(50 nm)/PtSe2 (2 nm)/Graphene (5 layer)/PBS 35.08 78.10 4.320 1409.3 34.54 5.50 4.266 1305.11 183.3 3.90 0.26 47.00 258.50

In addition, the proposed sensor’s sensitivity was increased by adding the graphene film layer (0.34 nm × L), where L is the number of additions of the graphene layer [30]. The reflectance and transmittance spectra with the addition of graphene layers is shown in Figure 7, and the data is tabulated in Table 4. The sensor’s sensitivity is enhanced by the increase in graphene layers, as previously described [69,70,71]. The sensitivity (SL) calculation with the addition of a different number of graphene layers (L = 1, 2, …, 5) is shown in Equation (9). The sensitivity with one graphene layer is 18.33 deg RIU−1 and FOM is 48.24 RIU−1. The sensitivity with five graphene layers is 88.33 deg RIU−1 and FOM is 38.24 RIU−1. The proposed sensor sensitivity is enhanced (1 + 0.55) × L times with the addition of graphene layers. An observation in Table 4 points out that the sensitivity, FWHM, and QF are increased but FOM and DA are decreased with the addition of graphene layers. The ideal number of graphene layers was between 2 and 5 graphene layers for optimum sensitivity, QF, as well as DA and FOM [55]. In the proposed sensor, a maximum of five-layer graphene film was added to achieve maximum efficiency [31].

SL=ΔθSPRΔns              =ΔθSPRL=0+ΔθSPRL=1Δns                         =ΔθSPRL=0+0.55×ΔθSPRL=0Δns                 =1+0.55×ΔθSPRL=0Δns                       =1+0.55×S0 (9)

Figure 7.

Figure 7

Schematic diagram of (a) SPR and (b) SPRF curve characteristics of Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene (0.34 nm × L) coated sensor reflectance and transmittance spectra with the thickness of the graphene layers (L = 0, 1, …, 5) before adsorption of the analyte.

Table 4.

Comparison of change in RminL,  θSPRL, TmaxL,  SRFfreqL, sensitivity, FWHM, DA, FOM, and QF with the variation of the graphene layer (L) in the proposed SPR biosensor.

Graphene layer thickness RminL % θSPRL 
deg
TmaxL 
dB
SRFfreqL
THz
ΔRminL [%] ΔθSPRL
deg
ΔTmaxL
dB
ΔSRFFreqL
THz
S
deg/RIU
FW-HM
deg
DA
[1/deg]
FOM
[RIU1]
QF
deg/RIU
L = 0 × 0.34 nm 15.20 75.45 1.6409 1394.2 - - - - - - - - -
L = 1 × 0.34 nm 19.05 76.00 2.1133 1397.6 3.85 0.55 0.4724 3.40 18.33 0.38 2.63 48.24 26.53
L = 2 × 0.34 nm 22.99 76.55 2.6124 1400.9 7.79 1.10 0.9715 6.70 36.67 0.78 1.28 47.01 51.72
L = 3 × 0.34 nm 27.00 77.10 3.1464 1404.0 11.8 1.65 1.5055 9.80 55.00 1.22 0.82 45.08 74.39
L = 4 × 0.34 nm 31.04 77.60 3.7156 1406.8 15.84 2.15 2.0747 12.60 71.67 1.73 0.58 41.43 89.07
L = 5 × 0.34 nm 35.08 78.10 4.3201 1409.3 19.88 2.65 2.6792 15.10 88.33 2.31 0.43 38.24 101.3

Further, the precise relevance of the SPR curve is to the SPR sensorgram, which has been utilized clinically over three decades to identify the presence of the target analyte. The real-time absorption curve (θA) in the sensorgram is obtained from the difference of SPR resonance angle (θR) and critical angle (θC), as shown in Figure 8 [72,73]. The sensorgram curve indicates the change of RI in the sensing region due to the interaction of the target analyte [74]. In this article, SPR curve characteristics are used to identify persons infected or uninfected by the COVID-19 virus.

Figure 8.

Figure 8

The Schematic of: (a) SPR, and (b) SPR sensorgram absorption curve.

Furthermore, electric field intensity is evaluated for the developed sensor, comparing it with other conventional sensors. In Lumerical FDTD, the angular interrogation technique is utilized to estimate the SPPs excitation near the sensor surface. The SPPs’ excitation produces an evanescent wave that decays exponentially with distance in a direction normal to the boundary [75]. The SPPs’ electromagnetic field intensity is maximum at a resonance angle of the incident light [75]. In Figure 9, the electric field intensity and angle of the incident concerning normal distance from the sensor interface are monitored for the three-layer, four-layer, and proposed five-layer SPR sensor. The strength of electric field intensity is enhanced near the surface in the proposed SPR sensor, compared to the other conventional sensor. Thus, the sensitivity of the proposed sensor is increased enormously for the detection of target analytes.

Figure 9.

Figure 9

The schematic of (a) the angle of incidence, and (b) the electric field intensity as a function of normal distance from the interface for (i) three-layer, (ii) four-layer, and (iii) five-layer SPR sensor configuration.

4. Result and Analysis

4.1. Rapid Detection of SARS-CoV-2 Virus Spike RBD

In rapid detection of the SARS-CoV-2 spike RBD, mAbs (IgG or IgM) can be immobilized onto the graphene film through PBSE as a probe ligand, and the whole virus can flow with PBS solution as an analyte. The RI of the sensing region is changed due to the interaction of different concentrations of the ligand-analyte. However, in numerical analysis, the sensing region RI is considered as, ns=1.3348+Δn, where the RI of PBS (pH 7.4) is 1.3348 and Δn is fluctuating due to the different concentrations of ligand-analyte interaction onto the sensing surface. The adaptive interaction distribution algorithm (AIDA) algorithm has been used to solve the Δn following reference [76,77,78]. To realize the interaction mechanism of the SARS-CoV-2 spike RBD, a rate constant distribution (RCD) algorithm with AIDA was established in reference [76,77]. The real experimental interaction mechanism of SARS-CoV-2 spike RBD with human ACE2 is the same or almost similar to the algorithm [76]. The affinity constant (KD), association (ka), and dissociation (kd) rate constants are varied due to variation in ligands and analytes [78,79]. Therefore, the rate constants value at the different concentration levels (concentration of 1.95 nM to 62.5 nM) of virus spike RBD is followed, as described in reference [76,77,78].

In Figure 10, the SPR and SPRF curve characteristics are illustrated with the different concentration levels of SARS-CoV-2 Spike RBDs flows as analyte. The θSPR~Rmin and SRF~Tmax characteristics are calculated in Table 5, with the variation of the analyte. In the simulation, the RI of PBS solution (concentration of 500 nM) without virus spike RBD was computed using the AIDA that has flowed onto the sensor surface, where Rmin is 36.25%, θSPR is 78.50°, SRF is 1411.3 THz and Tmax is 4.5017 dB calculated, respectively. Similarly, the RI of PBS solution with SARS-CoV-2 virus spike RBD (concentration of 1.95 to 62.5 nM) that flowed onto the sensor surface was computed. The RI is changed due to the interaction of different concentrations of mAbs and SARS-CoV-2 spike RBD. Due to the absorption of 1.953125nM SARS-CoV-2 spike RBD, the detection attributors shift right by 0.35° (from 78.50° to 78.85°) and 1.8 THz (from 1411.3 THz to 1413.1 THz). Furthermore, reflectance (R) and transmittance (T) also changed by 1.27% (from 36.25% to 37.52%) and 0.2018 dB (from 4.5017 dB to 4.7035 dB). Similarly, the detection attributors change with the absorption of SARS-CoV-2 spike RBDs; otherwise, the status would remain the same.

Figure 10.

Figure 10

Schematic diagram of: (a) SPR and (b) SPRF characteristics of Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene(1.7 nm) substrates with immobilized IgG (H014 or S309) as ligand and different concentration levels of SARS-CoV-2 spike RBDs (concentration of 1.953125 nM to 62.5 nM) as the analyte.

Table 5.

ΔRminCRBD,θSPRCRBD, TmaxCRBD and SRFfreqCRBD characteristics for different concentrations of SARS-Cov-2 Spike RBD as an analyte.

SARS-Cov-2 RBD Concentration (CRBD) with PBS (500 nM) [nM] RminCRBD [%] θSPRCRBD [deg.] TmaxCRBD [dB] SRFfreqCRBD[THz] ΔRminSC2R [%]= Rmin0 nMRminCRBD ΔθSPRSC2R [deg.]= θSPR0 nMθSPRCRBD ΔTmaxSC2R [dB] = Tmax0 nMTmaxCRBD ΔSRFfreqSC2R [THz]= SRFfreq0 nMSRFFreqCRBD
0 nM 36.25 78.50 4.5017 1411.3 0.00 0.00 0.0000 0.00
1.953125 nM 37.52 78.85 4.7035 1413.1 1.27 0.35 0.2018 1.80
3.90625 nM 38.91 79.25 4.9280 1414.9 2.66 0.75 0.4263 3.60
RBD 7.8125 nM 40.42 79.60 5.1792 1416.6 4.17 1.10 0.6775 5.30
15.625 nM 42.09 79.95 5.4627 1418.2 5.84 1.45 0.9610 6.90
31.25 nM 43.94 80.30 5.7880 1419.8 7.69 1.80 1.2863 8.50
62.5 nM 46.06 80.70 6.1728 1421.3 9.81 2.20 1.6711 10.00

The following (10) is employed for the diagnosis of COVID-19 patients:

ΔRmini %= Rmin0 nMRmin1.953125 nM=1.27%ΔθSPRi deg.= θSPR0 nMθSPR1.953125 nM=0.35 degΔTmaxi dB= Tmax0 nMTmax1.953125 nM=0.2018 dBΔSRFfreqi THz= SRFfreq0 nMSRFfreq1.953125 nM=1.8 THz (10)

4.2. Rapid Detection of SARS-CoV-2 Anti-Spike Protein

In this pandemic, numerous SARS-CoV-2 anti-spike proteins or mAbs have been screened out with time. mAbs was found to interact with virus spike RBD at the nM level and successfully deactivate the SARS-CoV-2 virus Spike RBD. The mAbs (3F11, BD-368-2, ab1, CB6, B38, H4, P2C-1F11, rRBD-15, 311mab–31B5, 311mab–32D4, CC12.1, COVA1-18, COVA2-15, S309, ADI-55689, REGN10989, H014, COV2-2196, COV2-2130) block the interaction between virus spike RBD and ACE2 [80]. For rapid detection, SARS-CoV-2 spike RBDs can be immobilized onto the graphene film with PBSE as a probe, and the mAbs (IgG;-H014) can flow with PBS as an analyte [81]. In Figure 11, the SPR and SPRF curve characteristics are shown with the different concentration levels of H014 flows as the analyte. The θSPR~Rmin and SRF~Tmax characteristics are calculated in Table 6.

Figure 11.

Figure 11

Schematic diagram of (a) SPR and (b) SPRF characteristics of Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene(1.7 nm) substrates with immobilized SARS-CoV-2 spike RBDs as the ligand and different concentration levels of IgG-H014 (concentration of 1.74 nM to 27.8 nM) as the analyte.

Table 6.

ΔRminCRBD,θSPRCRBD, TmaxCRBD and SRFfreqCRBD characteristics for different concentrations of IgG (H014) as an analyte.

IgG Concentration (CIgG) with PBS (500 nM) [nM] RminCIgG[%] θSPRCIgG
[deg.]
TmaxCIgG
[dB]
SRFfreqCIgG[THz] ΔRminIgG[%] =
Rmin0 nMRminCIgG
ΔθSPRIgG [deg.]= θSPR0 nMθSPRCigG ΔTmaxIgG[dB]=
Tmax0 nMTmaxCIgG
ΔSRFfreqIgG [THz]= SRFfreq0 nMSRFfreqCIgG
H014 0 nM 36.25 78.50 4.5017 1411.3 0.00 0.00 0.0000 0.00
1.74 nM 37.52 78.85 4.7034 1413.1 1.27 0.35 0.2017 1.80
3.47 nM 38.91 79.25 4.9277 1414.9 2.66 0.75 0.4260 3.60
6.94 nM 40.42 79.60 5.1784 1416.6 4.17 1.10 0.6767 5.30
13.9 nM 42.08 79.95 5.4608 1418.2 5.83 1.45 0.9591 6.90
27.8 nM 43.92 80.30 5.7837 1419.8 7.67 1.80 1.2820 8.50

In the simulation, the RI of PBS solution (concentration of 500 nM) without IgG (H014) was computed using the AIDA that had flowed onto the sensor surface, where Rmin is 36.25%, θSPR is 78.50°, SRF is 1411.3 THz, and Tmax is 4.5017 dB calculated, respectively. Similarly, the RI of the PBS solution with H014 (concentration of 1.74 to 27.8 nM, ref. [81]) that has flowed onto the sensor surface was computed. Due to the absorption of 1.74 nM H014, the detection attributors shift right by 0.35° (from 78.50° to 78.85°) and 1.8 THz (from 1411.3 THz to 1413.1 THz). While absorption increases, RminCIgG and TmaxCIgG attributors are also changed by 1.27% (from 36.25% to 37.52%) and 0.2017 dB (from 4.5017 dB to 4.7034 dB). Similarly, the detection parameters change with the absorption of H014; otherwise, the status would remain the same. The following (11) is employed for the diagnosis of COVID-19 patients:

ΔRminii %= Rmin0 nMRmin1.74 nM=1.27%ΔθSPRii deg.= θSPR0 nMθSPR1.74 nM=0.35 degΔTmaxii dB= Tmax0 nMTmax1.74 nM=0.2017 dBΔSRFfreqii THz= SRFfreq0 nMSRFfreq1.74 nM=1.8 THz (11)

4.3. Real-time Detection of SARS-CoV-2 RNA Sequence

Currently, the RT-PCR method is being applied to detect the SARS-CoV-2 virus. However, RNA extraction of the SARS-CoV-2 Spike is a long-time process. It takes more than one to three days to get a clinical report. To overcome this time-consuming procedure of RNA extraction, Zhao et al. developed a pcMNPs technique for RNA extraction (required ~30 min) [10]; it has been identified as a subsequent RT-PCR method. The sensor graphene layer can be attached to specific probe RNA sequences via PBSE to diagnose specific RNA sequences. This proposed sensor can detect hybridization events among the target SARS-CoV-2 spike RNA sequence and the probe RNA sequence. In Table 7, forward-reverse primers and probe genome sequences are specified with nucleotide position for detection of the RNA sequence of the COVID-19 virus [82,83,84]. In addition, the entirely matching (mr) sequence and mismatching (wt) sequence with probe (sh) linker is tabulated. In Figure 12, the SPR angle and SRF shifted right with the variation of different concentrated target RNA sequences. The detection attributor θSPR~Rmin and SRF~Tmax are calculated with probe (sh) sequence, where Rmin is 37.92%, θSPR is 78.95°, SRF is 1413.7 THz, and Tmax is 4.7670 dB. The attributors’ shifting parameters are tabulated in Table 8.

Table 7.

Primers and probe sequence orientation for recognition of SARS-CoV-2 virus RNA.

Name Target Genes Sequence 5′→3′ Nucleotide position Target Genes Sequence 5′→3′ Nucleotide Position
Forward Primer: ORF 1ab CCCTGTGGGTTTTACACTTAA 13342-13362 N gene GGGGAACTTCTCCTGCTAGAAT 28881-28902
Reverse Primer: ACGATTGTGCATCAGCTGA 13442-13460 CAG ACATTTTGCTCTCAAGCTG 28958-28979
Immobilized Mutant Probe (sh): FAM-CCGTCTGCGGTATGTGGAAAGGTTATGG-BHQ 13377-13404 FAM-TTGCTGCTGCTTGACAGATT-TAMRA 28934-28953
Matching type (mr) Sequence: TCCAACCTTTCCACATACCGCAGCGGA - AACTCTGTCAAGCAGCAGCAA -
Mismatching type (wt) Sequence: TCCAAGCAAACCCAATACCGCAGCGGA - AACTCTACTAATTAGCAGCAA -

Figure 12.

Figure 12

Schematic diagram of (a) SPR and (b) SPRF curve characteristics of Bk7/Au(50 nm)/PtSe2(2 nm)/Graphene (0.34 nm × L) substrates without probe (sh-Oligo), with sh-Oligo and different concentrated levels of oligonucleotide (wt or mr type) binding for recognition of SARS-Cov-2 virus RNA. θSPR angle and SPRF shift right due to binding with oligonucleotides.

Table 8.

RminCN,θSPRCN,TmaxCNandSRFfreqCN for different concentrations of the target oligonucleotides as an analyte.

Concentration (CN) with PBS (500 nM) [nM] RminCN [%] θSPRCN [deg.] TmaxCN [dB] SRFfreqCN[THz] ΔRminN [%]=RminshRminCN ΔθSPRN [deg.]=θSPRshθSPRCN ΔTmaxN[dB]=
TmaxshTmaxCN
ΔSRFFreqN [THz]=SRFfreqshSRFfreqCN
0 nM (no-sh-Oligo) 36.25 78.50 4.5017 1411.3 - - - -
150 nM (sh-Oligo) 37.92 78.95 4.7670 1413.7 0.00 0.00 0.0000 0.00
165 nM (wt-Oligo) 39.82 79.45 5.0778 1416.0 1.90 0.50 0.3108 2.30
180 nM (mr-Oligo) 41.97 79.95 5.4429 1418.1 4.05 1.00 0.6759 4.40
210 nM (mr-Oligo) 44.47 80.40 5.8827 1420.2 6.55 1.45 1.1157 6.50
240 nM (mr-Oligo) 47.34 80.90 6.4130 1422.1 9.42 1.95 1.6460 8.40
270 nM (mr-Oligo) 50.59 81.30 7.0506 1423.7 12.67 2.35 2.2836 10.00
300 nM (mr-Oligo) 54.21 81.70 7.8111 1425.1 16.29 2.75 3.0441 11.40

In the simulation, the mismatching type (wt) RNA oligonucleotide was seen to flow with 500 nM PBS and binding with sh-sequence, where Rmin is 39.82%, θSPR is 78.45°, SRF is 1416.0 THz, and Tmax is 5.0778 dB calculated, respectively. The attributors are shifted in minor levels during immobilization with a wt-type sequence than immobilization with an mr-type sequence. The entirely matching type (mr) RNA sequence immobilized with sh-sequence detection attributors are shifted right by 1.0° (from 78.95° to 79.95°) and 4.4 THz (from 1413.7 THz to 1418.1 THz). Meanwhile, R and T are also increased by 4.05% (from 37.92% to 41.97%) and 0.6759 dB (from 4.7670 dB to 5.4429 dB). In the sensing region, RI is increased with the increasing concentration of mr-type sequences. If a hybridization event takes place between the probe sequences and the mr-sequences, the attributors’ shift is equal to or greater than ΔRminc, ΔθSPRc, ΔTmaxc, and ΔSRFfreqc; if the event does not take place, the shift is smaller. For the detection of the virus, (12) is used:

ΔRminiii %= Rmin150 nMRmin180 nM=4.05%ΔθSPRiii deg.= θSPR150 nMθSPR180 nM=1.00 degΔTmaxiii dB= Tmax150 nMTmax180 nM=0.6759 dBΔSRFfreqiii THz=SRFfreq150 nMSRFfreq180 nM=4.4 THz (12)

In the numerical study, the SARS-CoV-2 virus spike RBD detection method is suggested by using the proposed sensor because monoclonal antibodies (IgG, IgM) take 3–8 days to be produced in the human body [85,86] and the RNA extraction period is a time-consuming process [10]. The detection of whole SARS-CoV-2 spike RBD and the anti-spike protein is done by a rapid method, compared to the virus RNA detection technique. In addition, the proposed sensor sensitivity provides 183.33° RIU−1 and 833.33 THz RIU−1 for detection of the whole SARS-Cov-2 virus spike; the sensor sensitivity offers 153.85° RIU−1 and 726.50 THz RIU−1 for detection of antibodies (IgG or IgM) against the virus spike protein, and the sensor sensitivity obtains 140.35° RIU−1 and 500.0THz RIU−1 for detection of virus RNA. The final decision or tested result will depend on the conditions of the attributor(θSPR~Rmin and SRF~Tmax). Eventually, the patient’s COVID-19 clinical test result is contingent on detection of attributors characteristics (Table 9). In rapid recognition of the whole SARS-CoV-2 virus spike RBD procedure, if ΔRmini~ΔθSPRi or ΔTmaxi~ΔSRFfreqi is larger or equivalent to ΔRminSC2R~ΔθSPRSC2R or ΔTmaxSC2R~ΔSRFfreqSc2R then the COVID-19 result is positive. Similarly, if ΔRmini~ΔθSPRi or ΔTmaxi~ΔSRFfreqi is smaller than ΔRminSC2R~ΔθSPRSC2R or ΔTmaxSC2R~ΔSRFfreqSc2R, then the COVID-19 result is negative. The detection of antibodies and RNA hybridization events against the SARS-CoV-2 virus follow the same procedure. In Table 9, the X and Y parameters are identified by the detection method. This sensor shows robust performance for the rapid detection of the SARS-CoV-2 virus.

Table 9.

COVID-19 test results based on the detection of attributors’ characteristics.

Pathology No. COVID-19 Test Result Making Condition Result
01 ΔRminX ΔRminY & ΔθSPRXΔθSPRY OR ΔTmaxXΔTmaxY  & ΔSRFfreqXΔSRFfreqY COVID-19 Positive
02 ΔRminX<ΔRminY & ΔθSPRX<ΔθSPRY ΔTmaxX<ΔTmaxY & ΔSRFfreqX<ΔSRFfreqY COVID-19 Negative
03 ΔRminX ΔRminY & ΔθSPRXΔθSPRY ΔTmaxXΔTmaxY & ΔSRFfreqXΔSRFfreqY Try again
04 ΔRminX ΔRminY & ΔθSPRXΔθSPRY ΔTmaxXΔTmaxY  & ΔSRFfreqXΔSRFfreqY Try again

Condition of X: Y (a) Rapid detection of SARS-CoV-2 Spike RBD: X = i and Y = SC2R; (b) Rapid detection of SARS-CoV-2 Anti-Spike Protein: X = ii and Y = IgG; (c) Real-time detection of SARS-CoV-2 RNA Sequence: X = iii and Y = N.

5. Conclusions

This paper proposes a graphene-based multi-layer coated SPR sensor for the early-stage detection of the COVID-19 virus. The proposed sensor was analyzed for rapid diagnosis and effectively differentiates between infected or uninfected persons. The BK7/Au/PtSe2/Graphene film-coated SPR sensor was found to have a superior sensitivity of 183.3°/RIU, compared to other conventional sensors. Graphene films have high conductivity and interaction quality with the ligand. The sensor sensitivity was found to be enhanced (1 + 0.55) × L times by increasing the number of graphene layers. The proposed sensor utilized the virus spike RBD, antibodies (IgG or IgM), or virus RNA to detect the virus. The performance of the proposed sensor was confirmed by the TMM algorithm and validated by the FDTD technique. Numerically, the proposed senor ensures early-stage detection, reducing processing time without false results. The proposed sensor is expected to be implemented commercially or clinically to identify COVID-19 patients.

Author Contributions

All authors contributed equally in preparing the manuscript. Conceptualization, T.B.A.A., S.F.M., and M.M.R. (Md. Motiur Rahman); methodology, T.B.A.A., S.F.M., M.M.R. (Md. Motiur Rahman), and M.M.R. (Md. Masud Rana); software, T.B.A.A., S.F.M., and M.M.R. (Md. Motiur Rahman); validation, T.B.A.A., M.M.R. (Md. Masud Rana), M.R.I., and M.A.P.M.; formal analysis, T.B.A.A., M.M.R. (Md. Masud Rana), M.R.I., and I.M.M.; investigation, T.B.A.A., M.M.R. (Md. Masud Rana), I.M.M., and M.A.P.M.; data curation, T.B.A.A. and M.M.R. (Md. Masud Rana); writing—original draft preparation, T.B.A.A.; writing—review and editing, M.M.R. (Md. Masud Rana), M.R.I., I.M.M., M.A.P.M., and A.Z.K.; supervision, M.M.R. (Md. Masud Rana), M.R.I., and A.Z.K. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.World Health Organization (WHO) Novel Coronavirus 2019 (COVID-19) World Health Organization; Geneva, Switzerland: 2019. [Google Scholar]
  • 2.Astuti I. Ysrafil Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2): An overview of viral structure and host response. Diabetes Metab. Syndr. 2020;14:407–412. doi: 10.1016/j.dsx.2020.04.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Wong S.K., Li W., Moore M.J., Choe H., Farzan M. A 193-Amino Acid Fragment of the SARS Coronavirus S Protein Efficiently Binds Angiotensin-converting Enzyme 2. J. Biol. Chem. 2004;279:3197–3201. doi: 10.1074/jbc.C300520200. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Huang Y., Yang C., Xu X.-F., Xu W., Liu S.-W. Structural and functional properties of SARS-CoV-2 spike protein: Potential antivirus drug development for COVID-19. Acta Pharmacol. Sin. 2020;41:1141–1149. doi: 10.1038/s41401-020-0485-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Gheblawi M., Wang K., Viveiros A., Nguyen Q., Zhong J.-C., Turner A.J., Raizada M.K., Grant M.B., Oudit G.Y. Angiotensin-Converting Enzyme 2: SARS-CoV-2 Receptor and Regulator of the Renin-Angiotensin System. Circ. Res. 2020;126:1456–1474. doi: 10.1161/CIRCRESAHA.120.317015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Knoll M.D., Wonodi C. Oxford–AstraZeneca COVID-19 vaccine efficacy. Lancet. 2021;397:72–74. doi: 10.1016/S0140-6736(20)32623-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Voysey M., Clemens S.A.C., A Madhi S., Weckx L.Y., Folegatti P.M., Aley P.K., Angus B., Baillie V.L., Barnabas S.L., Bhorat Q.E., et al. Safety and efficacy of the ChAdOx1 nCoV-19 vaccine (AZD1222) against SARS-CoV-2: An interim analysis of four randomised controlled trials in Brazil, South Africa, and the UK. Lancet. 2021;397:99–111. doi: 10.1016/S0140-6736(20)32661-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wise J. Covid-19: New coronavirus variant is identified in UK. BMJ. 2020;371:m4857. doi: 10.1136/bmj.m4857. [DOI] [PubMed] [Google Scholar]
  • 9.Cui F., Zhou H.S. Diagnostic methods and potential portable biosensors for coronavirus disease 2019. Biosens. Bioelectron. 2020;165:112349. doi: 10.1016/j.bios.2020.112349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhao Z., Cui H., Song W., Ru X., Zhou W., Yu X. A simple magnetic nanoparticles-based viral RNA extraction method for efficient detection of SARS-CoV-2. bioRxiv. 2020 doi: 10.1101/2020.02.22.961268. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Hou H., Wang T., Zhang B., Luo Y., Mao L., Wang F., Wu S., Sun Z. Detection of IgM and IgG antibodies in patients with coronavirus disease 2019. Clin. Transl. Immunol. 2020;9:e01136. doi: 10.1002/cti2.1136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Li Z., Yi Y., Luo X., Xiong N., Liu Y., Li S., Sun R., Wang Y., Hu B., Chen W., et al. Development and clinical application of a rapid IgM-IgG combined antibody test for SARS-CoV-2 infection diagnosis. J. Med. Virol. 2020;92:1518–1524. doi: 10.1002/jmv.25727. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lee H.-K., Lee B.-H., Seok S.-H., Baek M.-W., Lee H.-Y., Kim D.-J., Na Y.-R., Noh K.-J., Park S.-H., Kumar D.N., et al. Production of specific antibodies against SARS-coronavirus nucleocapsid protein without cross reactivity with human coronaviruses 229E and OC43. J. Vet. Sci. 2010;11:165–167. doi: 10.4142/jvs.2010.11.2.165. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Mak G.C., Cheng P.K., Lau S.S., Wong K.K., Lau C.S., Lam E.T., Chan R.C.W., Tsang D.N.C. Evaluation of rapid antigen test for detection of SARS-CoV-2 virus. J. Clin. Virol. 2020;129:104500. doi: 10.1016/j.jcv.2020.104500. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Krüttgen A., Cornelissen C.G., Dreher M., Hornef M.W., Imöhl M., Kleines M. Comparison of the SARS-CoV-2 Rapid antigen test to the real star Sars-CoV-2 RT PCR kit. J. Virol. Methods. 2021;288:114024. doi: 10.1016/j.jviromet.2020.114024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Korevaar D.A., Kootte R.S., Smits L.P., Aardweg J.G.V.D., Bonta P.I., Schinkel J., Vigeveno R.M., Berk I.A.V.D., Scheerder M.J., Lemkes B.A., et al. Added value of chest computed tomography in suspected COVID-19: An analysis of 239 patients. Eur. Respir. J. 2020;56:2001377. doi: 10.1183/13993003.01377-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Chen Z.-H., Li Y.-J., Wang X.-J., Ye Y.-F., Wu B.-L., Zhang Y., Xuan W.-L., Bao J.-F., Deng X.-Y. Chest CT of COVID-19 in patients with a negative first RT-PCR test. Medicine. 2020;99:e20837. doi: 10.1097/MD.0000000000020837. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Qiu G., Gai Z., Tao Y., Schmitt J., Kullak-Ublick G.A., Wang J. Dual-Functional Plasmonic Photothermal Biosensors for Highly Accurate Severe Acute Respiratory Syndrome Coronavirus 2 Detection. ACS Nano. 2020;14:5268–5277. doi: 10.1021/acsnano.0c02439. [DOI] [PubMed] [Google Scholar]
  • 19.Shrivastav A.M., Cvelbar U., Abdulhalim I. A comprehensive review on plasmonic-based biosensors used in viral diagnostics. Commun. Biol. 2021;4:1–12. doi: 10.1038/s42003-020-01615-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Samson R., Navale G.R., Dharne M.S. Biosensors: Frontiers in rapid detection of COVID-19. 3 Biotech. 2020;10:1–9. doi: 10.1007/s13205-020-02369-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Taha B.A., Al Mashhadany Y., Mokhtar M.H.H., Bin Zan M.S.D., Arsad N. An Analysis Review of Detection Coronavirus Disease 2019 (COVID-19) Based on Biosensor Application. Sensors. 2020;20:6764. doi: 10.3390/s20236764. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Hossain B., Islam M., Abdulrazak L.F., Rana M., Akib T.B.A., Hassan M. Graphene-Coated Optical Fiber SPR Biosensor for BRCA1 and BRCA2 Breast Cancer Biomarker Detection: A Numerical Design-Based Analysis. Photon Sens. 2019;10:67–79. doi: 10.1007/s13320-019-0556-7. [DOI] [Google Scholar]
  • 23.Akib T.B.A., Nazmuschayadat , Hossain B. Superior Performance of Surface Plasmon Resonance Biosensor for Recognizing of DNA Hybridization; Proceedings of the 2019 International Conference on Computer, Communication, Chemical, Materials and Electronic Engineering (IC4ME2); Rajshahi, Bangladesh. 11–12 July 2019; pp. 1–4. [Google Scholar]
  • 24.Peng W., Liu Y., Fang P., Liu X., Gong Z., Wang H., Cheng F. Compact surface plasmon resonance imaging sensing system based on general optoelectronic components. Opt. Express. 2014;22:6174–6185. doi: 10.1364/OE.22.006174. [DOI] [PubMed] [Google Scholar]
  • 25.Hossain M.B., Rana M.M. DNA Hybridization Detection Based on Resonance Frequency Readout in Graphene on Au SPR Biosensor. [(accessed on 31 December 2020)]; Available online: https://www.hindawi.com/journals/js/2016/6070742/
  • 26.Tee K.L., Jackson P.J., Scarrott J.M., Jaffe S.R., Johnson A.O., Johari Y., Pohle T.H., Mozzanino T., Price J., Grinham J., et al. Purification of recombinant SARS-CoV-2 spike, its receptor binding domain, and CR3022 mAb for serological assay. bioRxiv. 2020 doi: 10.1101/2020.07.31.231282. [DOI] [Google Scholar]
  • 27.Kim D., Lee J., Bal J., Chong C.-K., Lee J., Park H. Clinical Evaluation of an Immunochromatographic-Based IgM/IgG Antibody Assay (GenBody™ COVI040) for Detection of Antibody Seroconversion in Patients with SARS-CoV-2 Infection. Diagnostics. 2021;11:537. doi: 10.3390/diagnostics11030537. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Nikolayenko I.V., Galkin O.Y., Grabchenko N.I., Spivak M.Y. Preparation of highly purified human IgG, IgM, and IgA for immunization and immunoanalysis. Ukr. Bioorganica Acta. 2005;9:3–11. [Google Scholar]
  • 29.Perchetti G.A., Huang M.-L., Peddu V., Jerome K.R., Greninger A.L. Stability of SARS-CoV-2 in Phosphate-Buffered Saline for Molecular Detection. J. Clin. Microbiol. 2020;58:8. doi: 10.1128/JCM.01094-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Seo G., Lee G., Kim M.J., Baek S.-H., Choi M., Ku K.B., Lee C.-S., Jun S., Park D., Kim H.G., et al. Rapid Detection of COVID-19 Causative Virus (SARS-CoV-2) in Human Nasopharyngeal Swab Specimens Using Field-Effect Transistor-Based Biosensor. ACS Nano. 2020;14:5135–5142. doi: 10.1021/acsnano.0c02823. [DOI] [PubMed] [Google Scholar]
  • 31.Rahman M., Rana M., Rahman S., Anower M., Mollah A., Paul A.K. Sensitivity enhancement of SPR biosensors employing heterostructure of PtSe2 and 2D materials. Opt. Mater. 2020;107:110123. doi: 10.1016/j.optmat.2020.110123. [DOI] [Google Scholar]
  • 32.Lee J.-Y., Mai L.-W., Hsu C.-C., Sung Y.-Y. Enhanced sensitivity to surface plasmon resonance phase in wavelength-modulated heterodyne interferometry. Opt. Commun. 2013;289:28–32. doi: 10.1016/j.optcom.2012.10.005. [DOI] [Google Scholar]
  • 33.Anower S., Rahman M. Hybrid Heterostructures for SPR Biosensor. Biosensors. 2021 doi: 10.5772/intechopen.94932. [DOI] [Google Scholar]
  • 34.Guo Y., Singh N.M., Das C.M., Ouyang Q., Kang L., Li K., Coquet P., Yong K.-T. Two-dimensional PtSe2 Theoretically Enhanced Goos-Hänchen Shift Sensitive Plasmonic Biosensors. Plasmonics. 2020;15:1815–1826. doi: 10.1007/s11468-020-01204-9. [DOI] [Google Scholar]
  • 35.Jia Y., Li Z., Wang H., Saeed M., Cai H. Sensitivity Enhancement of a Surface Plasmon Resonance Sensor with Platinum Diselenide. Sensors. 2019;20:131. doi: 10.3390/s20010131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Chabot V., Miron Y., Grandbois M., Charette P.G. Long range surface plasmon resonance for increased sensitivity in living cell biosensing through greater probing depth. Sens. Actuators B Chem. 2012;174:94–101. doi: 10.1016/j.snb.2012.08.028. [DOI] [Google Scholar]
  • 37.Isaacs S., Abdulhalim I. Long range surface plasmon resonance with ultra-high penetration depth for self-referenced sensing and ultra-low detection limit using diverging beam approach. Appl. Phys. Lett. 2015;106:193701. doi: 10.1063/1.4921200. [DOI] [Google Scholar]
  • 38.Al-Janaby N., Al-Dergazly A. Fabrication of multi-mode tip fiber sensor based on surface plasmon resonance (SPR) Sustain. Eng. Innov. 2020;2:10–17. doi: 10.37868/sei.v2i1.27. [DOI] [Google Scholar]
  • 39.Fotovvati B., Namdari N., Dehghanghadikolaei A. On Coating Techniques for Surface Protection: A Review. J. Manuf. Mater. Process. 2019;3:28. doi: 10.3390/jmmp3010028. [DOI] [Google Scholar]
  • 40.Yim C., Lee K., McEvoy N., O’Brien M., Riazimehr S., Berner N.C., Cullen C.P., Kotakoski J., Meyer J.C., Lemme M.C., et al. High-Performance Hybrid Electronic Devices from Layered PtSe2 Films Grown at Low Temperature. ACS Nano. 2016;10:9550–9558. doi: 10.1021/acsnano.6b04898. [DOI] [PubMed] [Google Scholar]
  • 41.O’Brien M., McEvoy N., Motta C., Zheng J.-Y., Berner N.C., Kotakoski J., Elibol K., Pennycook T.J., Meyer J.C., Yim C., et al. Raman characterization of platinum diselenide thin films. 2D Mater. 2016;3:021004. doi: 10.1088/2053-1583/3/2/021004. [DOI] [Google Scholar]
  • 42.Yim C., O’Brien M., McEvoy N., Riazimehr S., Schäfer-Eberwein H., Bablich A., Pawar R., Iannaccone G., Downing C., Fiori G., et al. Heterojunction Hybrid Devices from Vapor Phase Grown MoS2. Sci. Rep. 2014;4:1. doi: 10.1038/srep05458. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Gatensby R., McEvoy N., Lee K., Hallam T., Berner N.C., Rezvani E., Winters S., O’Brien M., Duesberg G.S. Controlled synthesis of transition metal dichalcogenide thin films for electronic applications. Appl. Surf. Sci. 2014;297:139–146. doi: 10.1016/j.apsusc.2014.01.103. [DOI] [Google Scholar]
  • 44.Wang Y., Li L., Yao W., Song S., Sun J.T., Pan J., Ren X., Li C., Okunishi E., Wang Y.-Q., et al. Monolayer PtSe2, a New Semiconducting Transition-Metal-Dichalcogenide, Epitaxially Grown by Direct Selenization of Pt. Nano Lett. 2015;15:4013–4018. doi: 10.1021/acs.nanolett.5b00964. [DOI] [PubMed] [Google Scholar]
  • 45.Yan M., Wang E., Zhou X., Zhang G., Zhang H., Zhang K., Yao W., Lu N., Yang S., Wu S., et al. High quality atomically thin PtSe 2 films grown by molecular beam epitaxy. 2D Mater. 2017;4:045015. doi: 10.1088/2053-1583/aa8919. [DOI] [Google Scholar]
  • 46.Ultra-Fast, Ultra-Sensitive PtSe2 Gas Sensors. [(accessed on 4 March 2021)]; Available online: https://phys.org/news/2017-01-ultra-fast-ultra-sensitive-ptse2-gas-sensors.html.
  • 47.Jussila H., Yang H., Granqvist N., Sun Z. Surface plasmon resonance for characterization of large-area atomic-layer graphene film. Optica. 2016;3:151–158. doi: 10.1364/OPTICA.3.000151. [DOI] [Google Scholar]
  • 48.Ye S., Oh W.-C. Demonstration of enhanced the photocatalytic effect with PtSe2 and TiO2 treated large area graphene obtained by CVD method. Mater. Sci. Semicond. Process. 2016;48:106–114. doi: 10.1016/j.mssp.2016.03.001. [DOI] [Google Scholar]
  • 49.Yang X., Peng H., Xie Q., Zhou Y., Liu Z. Clean and efficient transfer of CVD-grown graphene by electrochemical etching of metal substrate. J. Electroanal. Chem. 2013;688:243–248. doi: 10.1016/j.jelechem.2012.09.025. [DOI] [Google Scholar]
  • 50.Van Ngoc H., Qian Y., Kil Han S., Kang D.J. PMMA-Etching-Free Transfer of Wafer-scale Chemical Vapor Deposition Two-dimensional Atomic Crystal by a Water Soluble Polyvinyl Alcohol Polymer Method. Sci. Rep. 2016;6:33096. doi: 10.1038/srep33096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Sattar S., Schwingenschlögl U. Electronic Properties of Graphene–PtSe2 Contacts. ACS Appl. Mater. Interfaces. 2017;9:15809–15813. doi: 10.1021/acsami.7b00012. [DOI] [PubMed] [Google Scholar]
  • 52.Gong W., Jiang S., Li Z., Li C., Xu J., Pan J., Huo Y., Man B., Liu A., Zhang C. Experimental and theoretical investigation for surface plasmon resonance biosensor based on graphene/Au film/D-POF. Opt. Express. 2019;27:3483–3495. doi: 10.1364/OE.27.003483. [DOI] [PubMed] [Google Scholar]
  • 53.Hossain B., Mehedi I.M., Moznuzzaman M., Abdulrazak L.F., Hossain A. High performance refractive index SPR sensor modeling employing graphene tri sheets. Results Phys. 2019;15:102719. doi: 10.1016/j.rinp.2019.102719. [DOI] [Google Scholar]
  • 54.Fazio R., Jannelli A., Agreste S. A Finite Difference Method on Non-Uniform Meshes for Time-Fractional Advection–Diffusion Equations with a Source Term. Appl. Sci. 2018;8:960. doi: 10.3390/app8060960. [DOI] [Google Scholar]
  • 55.Menon P.S., Said F.A., Mei G.S., Berhanuddin D.D., Umar A.A., Shaari S., Majlis B.Y. Urea and creatinine detection on nano-laminated gold thin film using Kretschmann-based surface plasmon resonance biosensor. PLoS ONE. 2018;13:e0201228. doi: 10.1371/journal.pone.0201228. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Hossain B., Akib T.B.A., Abdulrazak L.F., Rana M. Numerical modeling of graphene-coated fiber optic surface plasmon resonance biosensor for BRCA1 and BRCA2 genetic breast cancer detection. Opt. Eng. 2019;58:037104. doi: 10.1117/1.OE.58.3.037104. [DOI] [Google Scholar]
  • 57.Rahman M.S., Rikta K.A., Bin Bashar L., Anower M. Numerical analysis of graphene coated surface plasmon resonance biosensors for biomedical applications. Optica. 2018;156:384–390. doi: 10.1016/j.ijleo.2017.11.057. [DOI] [Google Scholar]
  • 58.Lin C., Chen S. Design of high-performance Au-Ag-dielectric-graphene based surface plasmon resonance biosensors using genetic algorithm. J. Appl. Phys. 2019;125:113101. doi: 10.1063/1.5066354. [DOI] [Google Scholar]
  • 59.Rahman M.S., Hossain B., Rana M. Sensitivity enhancement of porous silicon based SPR sensor using graphene-M0S2 hybrid structure; Proceedings of the 2nd International Conference on Electrical, Computer & Telecommunication Engineering (ICECTE); Rajshahi, Bangladesh. 8–10 December 2016; pp. 1–4. [Google Scholar]
  • 60.Islam M., Islam M., Shimul Y.C., Rahman A., Ruhe A.A., Hassan M., Hossain B. FDTD Analysis Fiber Optic SPR Biosensor for DNA Hybridization: A Numerical Demonstration with Graphene. J. Mater. Appl. 2019;8:13–19. doi: 10.32732/jma.2019.8.1.13. [DOI] [Google Scholar]
  • 61.De Moivre’s Formula. [(accessed on 9 March 2021)]; Available online: https://en.wikipedia.org/w/index.php?title=De_Moivre%27s_formula&oldid=1006955752.
  • 62.Cennamo N., Massarotti D., Galatus R.V., Conte L., Zeni L. Performance Comparison of Two Sensors Based on Surface Plasmon Resonance in a Plastic Optical Fiber. Sensors. 2013;13:721–735. doi: 10.3390/s130100721. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Meshginqalam B., Barvestani J. Performance Enhancement of SPR Biosensor Based on Phosphorene and Transition Metal Dichalcogenides for Sensing DNA Hybridization. IEEE Sens. J. 2018;18:7537–7543. doi: 10.1109/JSEN.2018.2861829. [DOI] [Google Scholar]
  • 64.Das C.M., Guo Y., Kang L., Ho H., Yong K. Investigation of Plasmonic Detection of Human Respiratory Virus. Adv. Theory Simulations. 2020;3:2000074. doi: 10.1002/adts.202000074. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Das C.M., Guo Y., Yang G., Kang L., Xu G., Ho H., Yong K. Gold Nanorod Assisted Enhanced Plasmonic Detection Scheme of COVID-19 SARS-CoV-2 Spike Protein. Adv. Theory Simul. 2020;3:2000185. doi: 10.1002/adts.202000185. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Ouyang Q., Zeng S., Jiang L., Hong L., Xu G., Dinh X.Q., Qian J., He S., Qu J., Coquet P., et al. Sensitivity Enhancement of Transition Metal Dichalcogenides/Silicon Nanostructure-based Surface Plasmon Resonance Biosensor. Sci. Rep. 2016;6 doi: 10.1038/srep28190. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Xie J., Zhang D., Yan X.-Q., Ren M., Zhao X., Liu F., Sun R., Li X., Li Z., Chen S., et al. Optical properties of chemical vapor deposition-grown PtSe 2 characterized by spectroscopic ellipsometry. 2D Mater. 2019;6:035011. doi: 10.1088/2053-1583/ab1490. [DOI] [Google Scholar]
  • 68.Bruna M., Borini S. Optical constants of graphene layers in the visible range. Appl. Phys. Lett. 2009;94:031901. doi: 10.1063/1.3073717. [DOI] [Google Scholar]
  • 69.Fu H., Zhang S., Chen H., Weng J. Graphene Enhances the Sensitivity of Fiber-Optic Surface Plasmon Resonance Biosensor. IEEE Sens. J. 2015;15:5478–5482. doi: 10.1109/JSEN.2015.2442276. [DOI] [Google Scholar]
  • 70.Shushama K.N., Rana M., Inum R., Hossain B. Graphene coated fiber optic surface plasmon resonance biosensor for the DNA hybridization detection: Simulation analysis. Opt. Commun. 2017;383:186–190. doi: 10.1016/j.optcom.2016.09.015. [DOI] [Google Scholar]
  • 71.Hossain B., Rana M. Graphene Coated High Sensitive Surface Plasmon Resonance Biosensor for Sensing DNA Hybridization. Sens. Lett. 2016;14:145–152. doi: 10.1166/sl.2016.3596. [DOI] [Google Scholar]
  • 72.Patching S.G. Surface plasmon resonance spectroscopy for characterisation of membrane protein–ligand interactions and its potential for drug discovery. Biochim. Biophys. Acta Biomembr. 2014;1838:43–55. doi: 10.1016/j.bbamem.2013.04.028. [DOI] [PubMed] [Google Scholar]
  • 73.Brogioni B., Berti F. Surface plasmon resonance for the characterization of bacterial polysaccharide antigens: A review. MedChemComm. 2014;5:1058–1066. doi: 10.1039/C4MD00088A. [DOI] [Google Scholar]
  • 74.Schasfoort R.B.M. Handbook of Surface Plasmon Resonance. Royal Society of Chemistry (RSC); Cambridge, UK: 2017. Chapter 1. Introduction to Surface Plasmon Resonance; pp. 1–26. [Google Scholar]
  • 75.Zhang J., Zhang L., Xu W. Surface plasmon polaritons: Physics and applications. J. Phys. D Appl. Phys. 2012;45:113001. doi: 10.1088/0022-3727/45/11/113001. [DOI] [Google Scholar]
  • 76.Forssén P., Samuelsson J., Lacki K., Fornstedt T. Advanced Analysis of Biosensor Data for SARS-CoV-2 RBD and ACE2 Interactions. Anal. Chem. 2020;92:11520–11524. doi: 10.1021/acs.analchem.0c02475. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Lan J., Ge J., Yu J., Shan S., Zhou H., Fan S., Zhang Q., Shi X., Wang Q., Zhang L., et al. Structure of the SARS-CoV-2 spike receptor-binding domain bound to the ACE2 receptor. Nature. 2020;581:215–220. doi: 10.1038/s41586-020-2180-5. [DOI] [PubMed] [Google Scholar]
  • 78.Forssén P., Multia E., Samuelsson J., Andersson M., Aastrup T., Altun S., Wallinder D., Wallbing L., Liangsupree T., Riekkola M.-L., et al. Reliable Strategy for Analysis of Complex Biosensor Data. Anal. Chem. 2018;90:5366–5374. doi: 10.1021/acs.analchem.8b00504. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Tian X., Li C., Huang A., Xia S., Lu S., Shi Z., Lu L., Jiang S., Yang Z., Wu Y., et al. Potent binding of 2019 novel coronavirus spike protein by a SARS coronavirus-specific human monoclonal antibody. Emerg. Microbes Infect. 2020;9:382–385. doi: 10.1080/22221751.2020.1729069. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Yu F., Xiang R., Deng X., Wang L., Yu Z., Tian S., Liang R., Li Y., Ying T., Jiang S. Receptor-binding domain-specific human neutralizing monoclonal antibodies against SARS-CoV and SARS-CoV-2. Signal Transduct. Target. Ther. 2020;5:1–12. doi: 10.1038/s41392-020-00318-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81.Lv Z., Deng Y.-Q., Ye Q., Cao L., Sun C.-Y., Fan C., Huang W., Sun S., Sun Y., Zhu L., et al. Structural basis for neutralization of SARS-CoV-2 and SARS-CoV by a potent therapeutic antibody. Science. 2020;369:1505–1509. doi: 10.1126/science.abc5881. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Wee S.K., Sivalingam S.P., Yap E.P.H. Rapid Direct Nucleic Acid Amplification Test without RNA Extraction for SARS-CoV-2 Using a Portable PCR Thermocycler. Genes. 2020;11:664. doi: 10.3390/genes11060664. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Xu Y., Li X., Zhu B., Liang H., Fang C., Gong Y., Guo Q., Sun X., Zhao D., Shen J., et al. Characteristics of pediatric SARS-CoV-2 infection and potential evidence for persistent fecal viral shedding. Nat. Med. 2020;26:502–505. doi: 10.1038/s41591-020-0817-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Li D., Zhang J., Li J. Primer design for quantitative real-time PCR for the emerging Coronavirus SARS-CoV-2. Theranostics. 2020;10:7150–7162. doi: 10.7150/thno.47649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Long Q.-X., Liu B.-Z., Deng H.-J., Wu G.-C., Deng K., Chen Y.-K., Liao P., Qiu J.-F., Lin Y., Cai X.-F., et al. Antibody responses to SARS-CoV-2 in patients with COVID-19. Nat. Med. 2020;26:845–848. doi: 10.1038/s41591-020-0897-1. [DOI] [PubMed] [Google Scholar]
  • 86.Denning D.W., Kilcoyne A., Ucer C. Non-infectious status indicated by detectable IgG antibody to SARS-CoV-2. Br. Dent. J. 2020;229:521–524. doi: 10.1038/s41415-020-2228-9. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)