Abstract
An ability of lensless titanium-based antenna coupled microbolometers (Ti-bolometers) operating at room temperature to monitor precisely radiation patterns in terahertz time-domain spectroscopy (THz-TDS) systems are demonstrated. To provide comprehensive picture, two different THz-TDS systems and Ti-bolometers coupled with three different antennas—narrowband dipole antennas for 0.3 THz, 0.7 THz and a log-periodic antenna for wideband detection—were selected for experiments. Radiation patterns, spatial beam profiles and explicit beam evolution along the propagation axis are investigated; polarization-sensitive properties under various THz emitter power ranges are revealed. It was found that the studied Ti-bolometers are convenient lensless sensors suitable to discriminate and control THz radiation pattern features in various wideband THz-TDS systems.
Keywords: titanium-based microbolometeric sensors, THz time-domain spectroscopy systems, precise THz mode control, radiation pattern monitoring
1. Introduction
Terahertz time-domain spectroscopy (THz-TDS) is a distinctive example of a wideband electromagnetic radiation system invented more than three decades ago [1]. The THz spectroscopy is very powerful technique for the test and discrimination of different kind of materials [2,3,4]; it can successfully serve for a medical inspection [5,6], security aims [7,8], photovoltaics industry [9,10] as well as food control purposes [11,12].
From the point of view of THz imaging, it displays a possibility for multispectral (colored) THz imaging enabling thus discrimination of different materials without recording their spectral signatures directly [13,14,15] as “colored” images of the THz light contain information on chemical composition, spatial distributions and physical properties of tested objects. One needs to note that the quality of obtained images are highly sensitive to the radiation pattern of scanning beam [16]. This circumstance becomes of a particular importance in real-time inspection and sensing when a high capture rate of images is required, and detectors arrays need to be implemented in the recording set-up using collimated beams. Moreover, even in the case of bulky objects imaging, when extended focus [17] or bifocal (Fibonacci) focusing [18] helps to discriminate the content, control and monitoring of the beam shape along its propagation axis demands of special attention.
High quality images of optoelectronically generated radiation patterns covering 0.1–2 THz range were recorded by room-temperature antenna-coupled -silicon-based microbolometers array with integrated CMOS read-out electronics [19]. Thin film vanadium oxide (VOx)-based microbolometers arrays were demonstrated as promising tool for label-free detection of reaction of small molecules with proteins [20]. Titanium-based microbolometers (Ti-bolometers) were found to be useful and convenient instrument in security [21], plastic package inspection and spectroscopic-spatial analysis of materials proving the maturity of THz technology for medicine applications [22].
Very recently, we have demonstrated that optimised room temperature Ti-bolometers can serve as a convenient and easy-to-use imaging tool to accurately adjust low power THz sources without additional focusing components [23]. Radiation patterns and spatial modes profiles generated by electronic sources, frequency domain spectrometers, optically pumped molecular THz laser were measured and analyzed [23]. In addition, it was shown that these devices exhibiting room temperature sensitivity values of 200 kV/W, noise-equivalent power (NEP) of 20 pW/, and the response time in the 5 s range are sensitive and fast enough to monitor precisely coherent THz emitters in optoelectronic time-domain systems.
In this work, we present comprehensive experimental study dedicated to reveal features of Ti-bolometers for spatial mode control in THz TDS systems. Three lensless Ti-bolometer sensors with different antenna designs—narrowband dipole antenna for 0.3 THz, 0.7 THz and a log-periodic antenna for wideband detection—are explored in detail. Radiation patterns, spatial beam profiles and explicit beam evolution along the propagation axis was examined, polarization sensitive properties were revealed under different THz emitter power range. It was shown that the Ti-bolometers do not require additional focusing optics to discriminate spatial mode features in the wideband THz pulse. It was revealed that Ti-bolometer sensors and their arrays can be employed as easy-to-use imaging instruments in THz TDS systems for accurate optical alignment and quality control of their radiation patterns.
2. Materials and Methods
2.1. Design and Spectral Characterization of Ti-bolometer Sensors
The design of antenna-coupled Ti-bolometers and their frequency properties are presented in Figure 1. Each microbolometer was fabricated from titanium which was found to be the most suitable for silicon processing technology, temperature coefficient and reliability. The titanium bridge of 12 m and 2 m in length and width respectively was electrodeposited on the silicon nitride (SiN) membrane of thickness of 2 m. The aluminum antenna and interconnection paths of area of approximately 500 m × 500 m, were formed around the active part of Ti-bolometer (see in top of the panel (a) in Figure 1). To reduce the heat capacitance and to increase the speed of operation, the Ti-bolometer was suspended in the air by etching out the underlying SiN membrane and was coupled either to resonant double-dipole or to wideband log-periodic type THz antenna (see panel (a) in Figure 1) [22,23,24]. Sensitivity of the device was enhanced resonantly by adjustment of a dipole antenna and resonant-cavity design: the back side reflection mirror was placed at the quarter wavelength distance (Figure 1a).
The active gain of each antenna was numerically simulated and is shown in Figure 1b. Detectors’ response spectra at room temperature were also characterized experimentally employing a custom-made Fourier spectrometer. The reference spectrum was registered when interferometer was coupled with a thermo-acoustic Golay cell detector featuring the flat response in THz frequency range. Finally, in order to avoid the influence of emitter and the beamsplitter spectral response the bolometer detected spectrum was divided by the reference spectrum detected by the Golay cell. As it can be seen from the normalized response spectra given in Figure 1c, the resonant Ti-bolometers designed for 0.3 THz and 0.7 THz show the detection peaks around the frequencies of 0.3 THz and 0.65 THz, respectively. Additionally, at higher frequencies, lobes exhibiting lower amplitudes are observed which most probably come from higher dipole antenna resonances. The detector with log-periodic antenna demonstrates detection bandwidth in the range of 0.25–1.3 THz (Figure 1b). As is visible from Figure 1c,b, the obtained experimental data correlate well with the theoretical predictions.
The sensitivity and NEP values are given in Table 1. Additionally, here the geometrical bolometer antennae parameters such as length L, width W and polar angle of antenna fingers of the wideband detector are tabulated.
Table 1.
Detector Type | Sensitivity [V/W] | NEP [pW/] | L [m] | W[m] | [deg] |
---|---|---|---|---|---|
0.3 THz | 1000 | 20 | 674 | 16 | – |
0.7 THz | 1000 | 20 | 247 | 39 | – |
Wideband | 50 | 60 | 1042 | – | 45 |
2.2. Experimental Setup of THz Time-Domain System
The optical outputs of two different THz-TDS systems without reference beams schematically shown in the upper parts of Figure 2a,b were characterized using three different Ti-bolometer designs described above.
In THz-TDS System 1, a femtosecond fiber laser (Femtofiber Pro, Toptica Photonics Ag, Munich, Germany) providing pulses of 780 nm wavelength, 90 fs pulse duration and 150 mW output power at 80 MHz pulse repetition rate was used to excite photoconductive antenna made from 350 m-thick LT-GaAs wafer. A THz emitter antenna (Teravil Ltd., Vilnius, Lithuania) had the shape of a H-dipole made from Ti/Au with the width of 20 m and the gap of 50 m between electrodes. A 8 mm-thick and 12 mm in diameter high resistivity hyperhemispherical silicon lens (Lens 1) with a focal distance of 500 m was used to collimate the THz radiation. A reference THz spectrum measured with THz detector (Teravil Ltd., Vilnius, Lithuania) is shown in the bottom of Figure 2a. THz detector was a Hertzian dipole type photoconductive antenna made from LT-GaAs with a 6 m width gap. The emitter exhibits a wide emission spectrum ranging from 0 to 5 THz, as it can be seen from the panel (a) in Figure 2.
In THz-TDS System 2, femtosecond fiber laser (LightWare FF50, Ekspla, Vilnius, Lithuania) with a center wavelength of 1064 nm, 150 fs pulse duration, 60 mW output power and 30 MHz repetition rate was employed. The source of THz radiation was a photoconductive emitter made from LT-GaAsBi (Teravil Ltd., Vilnius, Lithuania) wafer with thickness of 500 m. For collimation, silicon hyperhemispherical lens (Lens 2) with the same specifications as in System 1 was used. A THz emitter/detector antenna was the H-dipole made of Ti/Au with dipole antenna gap size of 6 m. The reference THz spectrum measured with THz detector (Teravil Ltd., Vilnius, Lithuania) is given in the Figure 2b. It is seen that it extends up to nearly of 4 THz.
The investigated Ti-bolometers were mounted on a 3-dimensional x,y,z electric translation stage and mechanical rotation stage. The detector can be moved in the plane for a spatial mode profile detection; separately, a movement along axis enables testing the beam profile along the propagation direction. While dipole antenna equipped Ti-bolometers are sensitive to polarization, rotation with the angle as shown in the (a) panel of Figure 1 allows studying THz beam polarization properties.
The signal of radiation emitted by TDS source was captured by the bolometer and was read out using the conventional lock-in technique. The reference was taken from a mechanical chopper modulating the femtosecond laser beam at 930 Hz.
3. Results and Discussion
The radiation patterns of THz-TDS spectrometers were investigated via five diverse approaches. Initially, spatial modes profiles were examined in plane, i.e., perpendicular to the beam propagation direction within two distances—very close to the THz emitter, placing the Ti-bolometer at 10 mm from the source; then—recording it in a far-field, as far as 115 mm from the source. The next step was to explore spatial profile features along the propagation direction and to study polarization properties of the devices. Finally, collimating lenses were removed from the emission set-up aiming to illustrate abilities to resolve spatial modes of the emitter without any optical components. The scanning was performed moving the stage at continuous velocity of 25 mm/s with the lock-in averaging constant of 20 ms. The acquisition of 12 mm by 12 mm image took approximately 350 s for the given scanning velocity.
3.1. Features of the Radiation Patterns
The radiation patterns scanned in plane in the distance of 10 mm from the emitter lens tip using Ti-bolometers of three different antenna designs are depicted in Figure 3. The conventional femtosecond Ti-Sapphire system emitting wideband THz pulses in a frequency range of 0.1–2.5 THz with an average power of 1.2 W served as a THz source. The emitter was based on 500 m-thick LT-GaAs wafer, the lens parameters are the same as described above. As it can be seen, radiation patterns recorded with different antenna-coupled Ti-bolometer exhibits different features: 0.3 THz resonant Ti-bolometer displays the symmetric mode spot (Figure 3a), while distinctive feature of 0.7 THz frequency sensors is strongly expressed side lobes in intensity distribution (Figure 3b). Spatial beam pattern registered by the wideband detector reveals a bright central spot accompanied by well-pronounced side lobes as presented in Figure 3c. The spatial evolution of wideband THz pulses along an optical axis is shown in Figure 3d–f. Results demonstrate that spatial features-enriched beam profile evolves into almost symmetric Gaussian beam at the distance of 50 mm from the lens. This illustrates nicely that the Ti-bolometer is quite convenient and sensitive enough to characterize the performance of THz-TDS emitter with the hyperhemispherical lens attached. The recorded pattern structure can be understood as follows: at low frequencies a wide central part is formed, while at higher frequencies the central pattern part becomes narrower [25]. Regarding side fringes, their origin is related to internal reflections in the lens, and are emitted in smaller angles (bend closer to the central beam part). Consequently, only the wide central beam part is registered with 0.3 THz frequency sensor, while the 0.7 THz and the wideband bolometers display the narrowed central part and fringes around it.
A precise optical alignment of THz radiation in TDS spectrometer and monitoring of the beam pattern propagation can aid in improvement of system performance, e.g., by increasing its dynamical range and lowering the noise floor—features which are of particular importance for imaging and spectroscopy. Therefore, the Ti-bolometers were tested as possible instrument for these aims intending to provide full experimental characterization of THz beams evolution via amplitude distribution across the THz beam in a far-field varying the distance from 5 cm up to 12 cm range from the emitter.
The results of investigation for both THz TDS systems are given in Figure 4 starting at 75 mm and 55 mm distances, for System 1 and System 2, respectively. Such starting points were chosen according to the geometrical restrictions in investigated systems. The first feature that attracts attention is different beam spot sizes if compare starting recording data of System 1 and System 2—the beam spot is significantly smaller in the latter case. As was aforesaid, the hyperhemispherical lens of the same design was used to collimate the emitters beams profile. To do it effectively, one needs to place the emitter in its focal point. As the THz emitters in the experimental setups were of different thicknesses (350 m vs. 500 m), the observed differences unveil that emitter thickness of the System 1 was not well matched to lens focal distance resulting thus into a larger divergence of the emitted beam mode.
The beam pattern evolution along the propagation axis illustrated via five cross-sections of the mode profile registered every 10 mm is presented in Figure 4. As it can be seen from Figure 4, the investigated pattern does not experience essential changes with the distance. Note highly expressed central part of pattern decaying with the distance and absence of side-lobes in this far-field experiment. The beam is also of small divergence which was estimated rad at 0.7 THz in both experimental systems. Since , where is a wavelength, n denotes the refractive index and stands for a beam waist, we can easily find that mm for the above given divergence and radiation frequency in System 2.
These experimental observations are in a good agreement with theoretically calculated beam patterns in the distances far from the lens tip [25]. However, one can underline the variation of the central spot diameter of the detected pattern with frequency in both THz TDS systems. The beam profile recorded with the sensor designed for 0.7 THz frequency displays the smallest diameter 5.10 mm in System 1 versus 2.55 mm in System 2, the wideband Ti-bolometer—slightly larger beam diameter 7.95 mm vs 3.45 mm, while the profile recorded with the antenna-coupled Ti-bolometer for 0.3 THz amounts to 8.40 mm vs 4.95 mm. The accuracy here is in the order of 0.15 mm. As already mentioned earlier, the beam width at 0.3 THz is the largest because of the longer wavelength and larger dipole antenna dimensions (see Table 1). The wideband Ti-bolometer results (Figure 4c,f) show the beam spot size which is close to that for 0.3 THz in Figure 4a,d, but with higher intensity of the central part of the beam. Such increase in intensity is governed mainly by the Ti-bolometer’s responsivity bandwidth (see Figure 1) and by the fact that the lens concentrates higher frequencies in wideband THz emission closer to the optical axis.
3.2. Polarization Properties
Features of the presented Ti-bolometer design enables it to be sensitive to the polarization of the THz radiation—if the sensor axis is aligned in parallel to the THz field electrical component of the incident light, i.e., deg—the coupling should be efficient resulting therefore to the maximal heating of the bolometer’s titanium bridge, and, as a consequence, to the largest signal. Gradual rotation around optical axis ( deg) can serve hence as polarization-resolved tool permitting using the Ti-bolometer to monitor polarization properties in THz TDS spectrometer.
Investigations were performed at distances 75 mm and 55 mm from the lens tip (front x-y scan planes positions as given in Figure 4) in System 1 and System 2, respectively. The obtained polarization results for all the investigated bolometers are shown in Figure 5. To avoid possible misalignment artifacts, the angular dependencies of signal at the pattern center are normalized to the signal at deg and presented in panels (d) and (i) for different THz TDs spectrometers. As is seen, the dependencies decrease gradually with the rotation angle and flatten above 80 degrees, where the signal level approaches the noise floor.
One can indicate that an ellipticity of the beam spot and its smooth decrease down to the noise level is recorded by the Ti-bolometer along with change. This feature indicates that the sensor can be well-suited to monitor polarization state or cross-polarized angular emission patterns [26] of THz pulsed emission—it is rather important issue for the most of common real implementations of THz-TDS spectrometers or THz imaging systems.
3.3. The Radiation Patterns and Spatial Mode Profiles Recorded without Collimating Optical Components
An important concern in alignment of THz time-domain spectrometers is a possibility to adjust it in a convenient way without employment of focusing or collimating optics. In what follows, it is illustrated that Ti-bolometer can successfully be used to explore radiation patterns or quality of spatial modes without hyperhemispherical lens. The spatial beam characteristics (a scan in the plane) recorded with 0.7 THz Ti-bolometer without lens and System 1 are shown in Figure 6. As one can see, three lobes are still captured in a scanning area of 7 × 7 mm although signal-to-noise ratio was about 3.3. Careful inspection allows noting that these three constituents of the spatial mode do not diverge while going along z—direction indicating good alignment of the optical system. As is known, about 21 % of the incident power passes through silicon lens [25], on the other hand, without the lens radiation is dispersed over the large angles, therefore, only a small fraction of the total emitted power can be registered. Despite this fact, room temperature Ti-bolometers sensitivity of 200 kV/W including amplification circuit and NEP of 20 pW/ [23] is sufficient to register weak signal from photoconductive THz emitter even though the output beam is not effectively concentrated by the hyperhemispherical lens around the optical axis.
3.4. Time-Domain Versus Solid State-Based Solutions
Quality of the spatial beam profile and their control as well as propagation properties of the wideband THz beams was one of the prime interests since invention of THz time-domain spectrometers. Special attention was given to an angular distribution of the radiation emitted from a terahertz antenna system equipped with a truncated spherical silicon lens [25], where THz pulses were recorded by moving the photoconductive antenna in space and determining the intensity of the desired frequency at every point. Despite time consuming and complex experiments, authors successfully reconstructed THz beam profiles and found excellent agreement with theoretical estimates. To increase the THz emission power, antenna-biased system were investigated [27]; to optimise spectrometer operation, practical procedures for measuring the beam quality of the wideband THz radiation were introduced [28]. It is worth mentioning several other experimental techniques to evaluate THz-TDS beam profiles, for instance, electro-optical sampling using non-linear crystal [29], variable size aperture scanning [30] or fixed aperture raster scanning and Hartmann test [31]. These approaches allow deeper insight in both temporal and spatial THz electric field evolution, as well as emission pattern determination. However, they usually suffer of either diffraction limited resolution, either extensive time requirements or low detected signal levels. For example, variable aperture scanning usually experiences edge diffraction limited spatial resolution, at small aperture diameters in particular. Moreover, it measures only the integrated spectral power and is not suitable for a study of the wideband radiation since the low frequency portion can be filtered out at small apertures. Fixed aperture raster scanning features relatively low signal-to-noise ratio; Hartmann test technique is time consuming as it requires collecting at least twice as much data as fixed aperture raster scanning [31].
In real implementation schemes of THz-TDS spectrometers or THz imaging systems, beam quality control, pattern monitoring and optical alignment require much simpler and convenient-in-use solutions. Preference to use incoherent solid-state solutions in THz-TDS beam imaging exhibits lifted diffraction and signal level restrictions, as well as reduced recording time. Moreover, the devices are compact and reliable. As elegant examples can serve employment of CMOS-based field-effect transistors for THz autocorrelators development [32], coupled-charge device camera to measure spot size of the THz pulse generated by optical rectification in a mosaic organic crystal [33] or usage of high-speed sampling coupled to CMOS devices for hyperspectral imaging in THz time-domain system [34].
The presented study demonstrates that Ti-bolometers can also be important part of the family bringing together solid-state-based sensors for precise and convenient incoherent validation of THz-TDS radiation patterns; it can serve for alignment aims, spatial modes quality control, for waist and divergence estimation even without additional focusing or collimating optics.
4. Conclusions
Lensless titanium-based antenna coupled microbolometers operating at room temperature are demonstrated for the precise monitoring of radiation pattern and spatial mode profiles control in terahertz time-domain spectroscopy systems. To provide comprehensive picture, two different THz-TDS systems and three Ti-bolometers with individual antenna designs—narrowband dipole antenna for 0.3 THz, 0.7 THz and a log-periodic antenna for THz wideband detection—are used in the investigation.The radiation patterns, spatial mode control, and explicit beam evolution along the propagation axis were examined, polarization-sensitive properties were revealed under different THz emitter power range. To illustrate the beam quality control, the beam waist of 1.52 mm and divergence of 0.045 rad were estimated from the far-field measurements at 0.7 THz. It was shown that the Ti-bolometers exhibiting room temperature sensitivity values of 200 kV/W, noise-equivalent power of 20 pW/, and the response time in the 5 s range are sensitive and fast enough to monitor precisely coherent THz emitters in optoelectronic time-domain systems even without any focusing and collimating optics elements.
Acknowledgments
The authors would like to express sincere appreciations to Rimvydas Venckevičius and Ramūnas Adomavičius for the initial phase of the experiments; Arūnas Krotkus and Agnieszka Siemion for enlightening discussions. Special thanks goes to Aleksander Sešek for activities in Ti-bolometers preparation, to Kęstutis Ikamas and Daniil Pashnev for the kind assistance in evaluating spectral features of the device and to Domas Jokubauskis for his contribution in overcoming optical alignment issues in THz time-domain spectrometer.
Abbreviations
The following abbreviations are used in this manuscript:
THz | Terahertz |
THz-TDS | Terahertz time-domain spectroscopy |
bolometers | microbolometers |
LT-GaAs | Low temperture gallium arsenide |
Ti | Titanium |
NEP | Noise Equivalent Power |
Author Contributions
Conceptualization, G.V.; methodology, L.M., A.U., A.Š., I.K., G.V.; validation, J.T.; sensors development, J.T., A.Š.; investigation, L.Q., L.M., I.G.; formal analysis, G.V., L.M., I.G.; writing—original draft preparation, L.Q., I.G.; writing—review and editing, G.V., L.M., A.U., I.K., J.T.; visualization, L.Q., L.M., I.G.; supervision, G.V. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
Not applicable.
Conflicts of Interest
The authors declare no conflict of interest.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Cheung K., Auston D. A novel technique for measuring far-infrared absorption and dispersion. Infrared Phys. 1986;26:23–27. doi: 10.1016/0020-0891(86)90043-6. [DOI] [Google Scholar]
- 2.Neu J., Schmuttenmaer C.A. Tutorial: An introduction to terahertz time domain spectroscopy (THz-TDS) J. Appl. Phys. 2018;124 doi: 10.1063/1.5047659. [DOI] [Google Scholar]
- 3.Abina A., Puc U., Jeglič A., Zidanšek A. Structural characterization of thermal building insulation materials using terahertz spectroscopy and terahertz pulsed imaging. NDT E Int. 2016;77:11–18. doi: 10.1016/j.ndteint.2015.09.004. [DOI] [Google Scholar]
- 4.Xie L., Yao Y., Ying Y. The application of terahertz spectroscopy to protein detection: A review. Appl. Spectrosc. Rev. 2014;49:448–461. doi: 10.1080/05704928.2013.847845. [DOI] [Google Scholar]
- 5.Son J.H., Oh S.J., Cheon H. Potential clinical applications of terahertz radiation. J. Appl. Phys. 2019;125 doi: 10.1063/1.5080205. [DOI] [Google Scholar]
- 6.Sun Q., He Y., Liu K., Fan S., Parrott E.P., Pickwell-MacPherson E. Recent advances in terahertz technology for biomedical applications. Quant. Imaging Med. Surg. 2017;7:345–355. doi: 10.21037/qims.2017.06.02. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Trofimov V.A., Varentsova S.A. A possible way for the detection and identification of dangerous substances in ternary mixtures using thz pulsed spectroscopy. Sensors. 2019;19:2365. doi: 10.3390/s19102365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Puc U., Abina A., Rutar M., Zidanšek A., Jeglič A., Valušis G. Terahertz spectroscopic identification of explosive and drug simulants concealed by various hiding techniques. Appl. Opt. 2015;54:4495. doi: 10.1364/AO.54.004495. [DOI] [PubMed] [Google Scholar]
- 9.Miyagawa K., Nagai M., Yamashita G., Ashida M., Kim C., Akiyama H., Kanemitsu Y. Quantitative monitoring of the internal field in the depletion layer of a GaAs-based solar cell with terahertz radiation. Appl. Phys. Lett. 2018;113 doi: 10.1063/1.5037952. [DOI] [Google Scholar]
- 10.Minkevičius L., Suzanovičiene R., Balakauskas S., Molis G., Krotkus A., Valušis G., Tamošiunas V. Detection of tab wire soldering defects on silicon solar cells using terahertz time-domain spectroscopy. Electron. Lett. 2012;48:932–934. doi: 10.1049/el.2012.1995. [DOI] [Google Scholar]
- 11.Afsah-Hejri L., Hajeb P., Ara P., Ehsani R.J. A Comprehensive Review on Food Applications of Terahertz Spectroscopy and Imaging. Compr. Rev. Food Sci. Food Saf. 2019;18:1563–1621. doi: 10.1111/1541-4337.12490. [DOI] [PubMed] [Google Scholar]
- 12.Karaliūnas M., Nasser K.E., Urbanowicz A., Kašalynas I., Bražinskienė D., Asadauskas S., Valušis G. Non-destructive inspection of food and technical oils by terahertz spectroscopy. Sci. Rep. 2018;8:18025. doi: 10.1038/s41598-018-36151-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wan M., Healy J.J., Sheridan J.T. Terahertz phase imaging and biomedical applications. Opt. Laser Technol. 2020;122:105859. doi: 10.1016/j.optlastec.2019.105859. [DOI] [Google Scholar]
- 14.Aghamiri N.A., Huth F., Huber A.J., Fali A., Hillenbrand R., Abate Y. Hyperspectral time-domain terahertz nano-imaging. Opt. Express. 2019;27:24231–24242. doi: 10.1364/OE.27.024231. [DOI] [PubMed] [Google Scholar]
- 15.Puc U., Abina A., Jeglič A., Zidanšek A., Kašalynas I., Venckevičius R., Valušis G. Spectroscopic Analysis of Melatonin in the Terahertz Frequency Range. Sensors. 2018;18:4098. doi: 10.3390/s18124098. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kašalynas I., Venckevičius R., Tumonis L., Voisiat B., Seliuta D., Valušis G., Račiukaitis G. Reflective terahertz imaging with the TEM01 mode laser beam. Appl. Opt. 2013;52:5640–5644. doi: 10.1364/AO.52.005640. [DOI] [PubMed] [Google Scholar]
- 17.Minkevičius L., Jokubauskis D., Kašalynas I., Orlov S., Urbas A., Valušis G. Bessel terahertz imaging with enhanced contrast realized by silicon multi-phase diffractive optics. Opt. Express. 2019;27:36358. doi: 10.1364/OE.27.036358. [DOI] [PubMed] [Google Scholar]
- 18.Jokubauskis D., Minkevičius L., Karaliūnas M., Indrišiūnas S., Kašalynas I., Račiukaitis G., Valušis G. Fibonacci terahertz imaging by silicon diffractive optics. Opt. Lett. 2018;43:2795. doi: 10.1364/OL.43.002795. [DOI] [PubMed] [Google Scholar]
- 19.Oden J., Meilhan J., Lalanne-Dera J., Roux J.F., Garet F., Coutaz J.L., Simoens F. Imaging of broadband terahertz beams using an array of antenna-coupled microbolometers operating at room temperature. Opt. Express. 2013;21:4817–4825. doi: 10.1364/OE.21.004817. [DOI] [PubMed] [Google Scholar]
- 20.Oda N. Détecteur matriciel de type bolométrique à température ambiante et camera vidéo pour l’imagerie térahertz. Comptes Rendus Phys. 2010;11:496–509. doi: 10.1016/j.crhy.2010.05.001. [DOI] [Google Scholar]
- 21.Trontelj J., Sešek A. Electronic terahertz imaging for security applications. Terahertz Rf Millimeter-Submillimeter-Wave Technol. Appl. IX. 2016;9747:974713. doi: 10.1117/2.1201601.006300. [DOI] [Google Scholar]
- 22.Kašalynas I., Venckevičius R., Minkevičius L., Sešek A., Wahaia F., Tamošiūnas V., Voisiat B., Seliuta D., Valušis G., Švigelj A., et al. Spectroscopic terahertz imaging at room temperature employing microbolometer terahertz sensors and its application to the study of carcinoma tissues. Sensors. 2016;16:432. doi: 10.3390/s16040432. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Minkevičius L., Qi L., Siemion A., Jokubauskis D., Sešek A., Švigelj A., Trontelj J., Seliuta D., Kašalynas I., Valušis G. Titanium-based microbolometers: Control of spatial profile of terahertz emission in weak power sources. Appl. Sci. 2020;10:3400. doi: 10.3390/app10103400. [DOI] [Google Scholar]
- 24.Sešek A., Kašalynas I., Žemva A., Trontelj J. Antenna-coupled Ti-microbolometers for High-sensitivity Terahertz Imaging. Sens. Actuators A Phys. 2017;268:133–140. doi: 10.1016/j.sna.2017.11.029. [DOI] [Google Scholar]
- 25.Jepsen P.U., Keiding S.R. Radiation patterns from lens-coupled terahertz antennas. Opt. Lett. 1995;20:807–809. doi: 10.1364/OL.20.000807. [DOI] [PubMed] [Google Scholar]
- 26.Van Rudd J., Johnson J.L., Mittleman D.M. Cross-polarized angular emission patterns from lens-coupled terahertz antennas. J. Opt. Soc. Am. B. 2001;18:1524. doi: 10.1364/JOSAB.18.001524. [DOI] [Google Scholar]
- 27.Jepsen P.U., Jacobsen R.H., Keiding S.R. Generation and detection of terahertz pulses from biased semiconductor antennas. J. Opt. Soc. Am. B. 1996;13:2424. doi: 10.1364/JOSAB.13.002424. [DOI] [Google Scholar]
- 28.Bitman A., Lumer Y., Moshe I., Zalevsky Z. Characterization of spectrally broadband terahertz beam propagation. J. Opt. Soc. Am. B. 2012;29:1436. doi: 10.1364/JOSAB.29.001436. [DOI] [Google Scholar]
- 29.Gürtler A., Winnewisser C., Helm H., Uhd Jepsen P. Terahertz pulse propagation in the near field and the far field. J. Opt. Soc. Am. A. 2000;17:74. doi: 10.1364/JOSAA.17.000074. [DOI] [PubMed] [Google Scholar]
- 30.Molloy J.F., Naftaly M., Dudley R.A. Characterisation of Terahertz Beam Profile and Propagation through Complex Quasi-Optic Systems. Int. J. Terahertz Sci. Technol. 2011;4:99–103. doi: 10.11906/TST.099-103.2011.09.14. [DOI] [Google Scholar]
- 31.Molloy J.F., Naftaly M., Dudley R.A. Characterization of Terahertz Beam Profile and Propagation. IEEE J. Sel. Top. Quantum Electron. 2013;19:8401508. doi: 10.1109/JSTQE.2012.2205668. [DOI] [Google Scholar]
- 32.Ikamas K., Nevinskas I., Krotkus A., Lisauskas A. Silicon field effect transistor as the nonlinear detector for terahertz autocorellators. Sensors. 2018;18:3735. doi: 10.3390/s18113735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Chefonov O.V., Ovchinnikov A.V., Agranat M.B., Stepanov A.N. Terahertz beam spot size measurements by a CCD camera. Opt. Lett. 2019;44:4099. doi: 10.1364/OL.44.004099. [DOI] [PubMed] [Google Scholar]
- 34.Tsubouchi M., Nagashima K. High-speed terahertz color imaging using a 100 kHz line scan camera. Opt. Express. 2020;28:17820. doi: 10.1364/OE.394312. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Not applicable.