Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2021 May 27;16(5):e0252234. doi: 10.1371/journal.pone.0252234

Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy

Jaehyeon Park 1,2, Ji Woon Yea 1,2, Jae Won Park 1,2, Se An Oh 1,2,*
Editor: Stephen Chun3
PMCID: PMC8158872  PMID: 34043724

Abstract

The objective of this study was to analyze the difference in residual setup errors between 6D ExacTrac and 3D cone-beam computed tomography (CBCT) image-guided systems in spinal stereotactic body radiation therapy (SBRT). We investigated 76 patients with spinal tumors who received SBRT using Novalis Tx at our institution between January 2013 and September 2020. A Vac-lok (EZ-FIX®, Arlico Medical Company, South Korea) fixture and an assistive device, based on the region involved, were used to immobilize patients and to increase the inter-fractional setup reproducibility. The difference in the root mean square (RMS) between the 6D ExacTrac and 3D CBCT was -0.75 mm, 0.45 mm, 0.16 mm, and -0.03°; the RMS value was 1.31 mm, 1.06 mm, 0.87 mm, and 0.64°; and the standard deviation was 0.80 mm, 0.72 mm, 0.62 mm, and 0.44° for lateral, longitudinal, vertical, and yaw directions, respectively. The difference in the average RMS between ExacTrac and CBCT was <1.03 mm in the translation direction and <0.47° in the rotational direction; the results were statistically significant in the lateral, longitudinal, and vertical directions, but not in the yaw direction. Thus, it is necessary to verify the ExacTrac image according to the CBCT image.

Introduction

The clinical efficacy of stereotactic body radiation therapy (SBRT) for spinal tumors has been previously reported [16]. Precise delivery of high doses of radiation in spinal tumors with SBRT has shown potential clinical benefits. To perform sophisticated treatments such as stereotactic radiosurgery (SRS) and SBRT, the use of ExacTrac (BrainLAB, Feldkirchen, Germany) and cone-beam computed tomography (Varian Medical System, CA, USA) image guidance systems is essential [710].

SBRT delivers high doses per fraction; therefore, an extremely steep dose gradient is required to deliver minimum and maximum radiation doses to the normal organ and tumors, respectively [8]. To minimize the radiation dose to the normal organs, a minimal setup margin is required for the tumor [7]. In contrast, if the setup margin is too low, the uncertainty of the radiation dose delivered to the tumor may increase, leading to poor clinical results. Therefore, for an optimum amount of delivery, image guidance using ExacTrac and CBCT is crucial when treating spinal tumors with SBRT.

Chang et al. [11] reported the setup discrepancies with ExacTrac X-ray 6 degree of freedom (6D) and CBCT on a Novalis Tx system in both phantom and retrospective patient studies. In 16 patients who underwent spinal SBRT, the residual setup error for the root mean square (RMS) between ExacTrac and CBCT was <2.0 mm in patients and <1.0 mm in the phantom for the translational direction, and <1.5° in patients and <1.0° in the phantom for the rotational direction.

Similarly, in our institution [12], we analyzed the residual setup error between ExacTrac and CBCT in 107 brain tumor patients treated with SRS using Novalis Tx, from August 2012 to July 2016. The difference in the RMS on online matching between 6D ExacTrac and 3D CBCT was 1.01 in the translational direction and 0.82° in the rotational direction. However, the limitation of our study was that it only analyzed patients who received intracranial SRS.

Therefore, in this study, we aimed to analyze the difference in the residual setup error between 6D ExacTrac and 3D CBCT image guidance systems in patients with spinal tumors who underwent SBRT.

Materials and methods

Study overview

The retrospective data analysis of 76 patients with spinal tumors enrolled in this study was approved by the Institutional Review Board of the Yeungnam University Medical Center (YUMC 2020-09-068). The need for informed consent was waived by the approval of the Institutional Review Board of the Yeungnam University Medical Center (YUMC 2020-09-068), given that patient anonymity was ensured. All methods were carried out in accordance with relevant guidelines and regulations.

Patient selection

We investigated 76 patients with spinal tumors who received SBRT using the Novalis Tx in our institution between January 2013 and September 2020. The patient and treatment characteristics are described in Table 1. For all 76 patients (female, 40 [53%]; male, 36 [47%]; average age, 62 years) undergoing SBRT, image-guided verification was performed before the radiation treatment using BrainLAB 6D ExacTrac and CBCT. The treatment sites for spinal SBRT were cervical in 11 cases, thoracic in 41 cases, and lumbar in 24 cases. A total of 268 fractions were investigated: 27 Gy in 3 fractions (36 [47%] cases) and 32 Gy in 4 fractions (40 cases).

Table 1. Characteristics of patients and treatment.

Patient characteristics
Number of patients N = 76
Median age(Range) 62(30–82)
Gender(%)
Female 40(53)
Male 36(47)
Spinal region of used for the treatment (%)
Cervical 11(14)
Thoracic 41(54)
Lumbar 24(32)
Treatment characteristics
Number of fractions n = 268
Fraction schemes (dose/fraction) (%)
27 Gy in 3 fractions (7 Gy) 36(47)
32 Gy in 4 fractions (8 Gy) 40(53)

Immobilization and CT simulation

For all patients, Vac-lok (EZ-FIX®, Arlico Medical Company, South Korea) fixtures were used to minimize patient movement and to increase interfractional setup reproducibility. Depending on the treatment regions, various assistive devices were used to minimize movement during radiation. For instance, for most patients with tumors in the cervical spine, a head and neck thermal mask (DUONTM, Orfit Industries, Wijnegem, Belgium) was used for fixation. CT simulations, with a thickness of 2.5 mm, were performed using a Brilliance Big Bore CT simulator (Philips Inc., Cleveland, OH).

Treatment planning and delivery techniques

To increase the sophistication in images for contour delineation, axial T1, axial T2 gadolinium-enhanced T1 of 2 mm thickness, and magnetic resonance image sequences were fused. In the case of spine SBRT with our institution protocol, planning target volume (PTV) was created by extending the margin 3 mm in all directions from the clinical target volume. When major organs such as the spinal cord were adjacent, the PTV was reduced by radiation oncologist. The treatment plan was generated using nine-field static intensity-modulated radiotherapy and anisotropic analytical algorithm. The beam arrangement of the radiation was set in a direction to avoid critical organs, to the best possible extent, with the majority of them directed posteriorly. From April 2019, an upgraded version, Eclipse ARIA 15.6 (Varian Medical System, Palo Alto, CA, USA) instead of Eclipse ARIA 8.6 (Varian Medical System) was used for treatment planning. The utilized radiation treatment plan satisfied the TG-101 guidelines for the normal tissue tolerance dose [13].

Image registration and setup protocol

Fig 1 shows the positioning for the patient setup for T spinal stereotactic body radiation (SBRT) using BrainLAB Infrared (IR) Reflective Reference Star (BrainLAB, AG, Feldkirchen, Germany). Image registration was performed using BrainLAB’s 6D ExacTrac and Varian’s 3D CBCT. Using the images, shifts were calculated in the translational (ExacTrac and CBCT: lateral, longitudinal, and vertical) and rotational (ExacTrac: pitch, roll, and yaw direction; CBCT: yaw) directions. Fig 2 indicates image registration using ExacTrac tube 1 and the ExacTrac tube 2 with BrainLAB for thoracic SBRT. Image registration using Varian CBCT for thoracic SBRT is shown in Fig 3.

Fig 1. Positioning for the patient setup for thoracic spinal stereotactic body radiation therapy using BrainLAB Infrared (IR) Reflective Reference Star (BrainLAB, AG, Feldkirchen, Germany).

Fig 1

Fig 2. Image registration using ExacTrac tube 1 and ExacTrac tube 2 with BrainLAB for thoracic spinal stereotactic body radiation therapy.

Fig 2

(a) ExacTrac image; (b) digitally reconstructed radiography image; (c) registration image.

Fig 3. Image registration using Varian cone-beam computed tomography (CBCT) for thoracic spinal stereotactic body radiation therapy.

Fig 3

(a) planning computed tomography (CT); (b) CBCT; (c) registration image.

The setup protocol of spinal SBRT followed in our hospital was implemented. First, ExacTrac images were obtained using ExacTrac tubes 1 and 2, followed by image reconstruction by digitally reconstructed radiography simulator system using CT data. Second, the patient was couch-shifted, using the difference value in image registration between the measured ExacTrac image and reconstructed digitally reconstructed radiography image. Third, it was ensured that the patient had been properly moved using ExacTrac. Image setup tolerance limit was considered to be 0.5 mm for the translational direction and 0.5° for the rotational direction. However, if X-ray verification was performed more than three times and still exceeded the tolerance limit, then it was ignored. Fourth, 3D CBCT images were acquired from the patient’s exact position from which the ExacTrac images were obtained. Finally, after the radiation oncologist confirmed the patient’s setup position, it was finally corrected using the 3D CBCT and radiotherapy was performed.

Analysis of the residual setup errors between ExacTrac and CBCT

The difference in the residual setup error between ExacTrac and CBCT, for a total of 268 fractionations, was analyzed as RMS, standard deviation, and difference in the translational (lateral, longitudinal, and vertical) and rotational (yaw) directions. In addition, RMS values measured in each image were statistically analyzed through paired t-tests using the SPSS software (IBM Corp., Chicago, IL, USA). A p-value of less than 0.05 was considered statistically significant.

Results

In the exact patient position, before radiation, the average value obtained by subtracting the 6D ExacTrac value from the 3D CBCT value was -0.75 mm, 0.45 mm, 0.16 mm, and -0.03°; the RMS value was 1.31 mm, 1.06 mm, 0.87 mm, and 0.64°; and the standard deviation was 0.80 mm, 0.72 mm, 0.62 mm, and 0.44° for lateral, longitudinal, vertical, and yaw directions, respectively. Histograms and normalized curves for the translational and rotational directions between the 6D ExacTrac and 3D CBCT are shown in Fig 4.

Fig 4. Histogram and normalized curves for the translational and rotational directions between 6D ExacTrac and 3D cone-beam computed tomography.

Fig 4

(a) lateral; (b) longitudinal; (c) vertical; and (d) yaw directions.

Table 2 shows the residual setup errors between 6D ExacTrac and 3D CBCT for spinal SBRT. The present study results were compared with the results from a study by Chang et al. [11] on spinal tumors and those of the brain SRS study performed at our institution [12].

Table 2. Residual setup errors between 6D ExacTrac and 3D CBCT for spinal SBRT.

Region Number of patient Directions Setup error for 6D ExacTrac Setup error for 3D CBCT 6D ExacTrac vs 3D CBCT
Difference p-value of paired t-test
RMS SD RMS SD RMS SD
Chang et al. [11] Spine N = 11, n = 16 Translational
Lateral (x-axis) (mm) n/a n/a n/a n/a 1.22 1.25 0.875
Longitudinal (z-axis) (mm) n/a n/a n/a n/a 0.93 0.94 0.307
Vertical-(y-axis)(mm) n/a n/a n/a n/a 1.67 1.65 0.162
Rotational
Pitch (x-axis) (°) n/a n/a n/a n/a 0.54 0.53 0.271
Roll (z-axis) (°) n/a n/a n/a n/a 0.40 0.40 0.356
Yaw (y-axis) (°) n/a n/a n/a n/a 0.87 0.90 0.896
Oh et al. [12] Brain N = 107, n = 138 Translational
Lateral (x-axis) (mm) 0.20 0.20 0.97 0.65 1.01 0.60 <0.001a
Longitudinal (z-axis) (mm) 0.24 0.24 0.77 0.77 0.84 0.63 0.425
Vertical (y-axis) (mm) 0.20 0.20 0.76 0.75 0.76 0.56 0.028a
Rotational
Pitch (x-axis) (°) 0.18 0.18 n/a n/a n/a n/a n/a
Roll (z-axis) (°) 0.17 0.17 n/a n/a n/a n/a n/a
Yaw (y-axis) (°) 0.22 0.22 0.80 0.80 0.82 0.49 0.226
Present Study Spine N = 76, n = 268 Translational
Lateral (x-axis) (mm) 0.40 0.40 1.10 0.81 1.31 0.80 <0.001
Longitudinal (z-axis) (mm) 0.36 0.36 0.89 0.77 1.06 0.72 <0.001
Vertical (y-axis) (mm) 0.33 0.33 0.95 0.94 0.87 0.62 0.002
Rotational
Pitch (x-axis) (°) n/a n/a n/a n/a n/a n/a n/a
Roll (z-axis) (°) n/a n/a n/a n/a n/a n/a n/a
Yaw (y-axis) (°) 0.30 0.30 0.46 0.46 0.64 0.44 0.526

Residual setup errors between the 6D ExacTrac and 3D cone-beam computed tomography (CBCT) in spinal stereotactic body radiation therapy (SBRT). ap<0.05. N, number of patients; n, number of fractions; RMS, root mean square; SD, standard deviation.

The paired t-test of the RMS values of the residual setup error in 6D ExacTrac and 3D CBCT showed significant differences in the lateral, longitudinal, and vertical directions, but not in the yaw direction.

According to the region, the spine was divided into cervical, thoracic, and lumbar regions, and the difference in the residual setup error between 6D ExacTrac and CBCT images was analyzed. The results are shown in S1 Table. In particular, the cervical spine showed a significant difference between ExacTrac and CBCT in all directions (lateral, longitudinal, vertical, and yaw).

Discussion

We analyzed the images of patients treated with spine SBRT using ExacTrac and CBCT at the exact same location and determined the difference in residual setup errors between ExacTrac and CBCT images. In 76 patients who underwent SBRT, a total of 268 fractions of images were analyzed. The results showed that the difference in average RMS values between ExacTrac and CBCT was <1.03 mm in the translational direction and <0.47° in the rotational direction; the differences were statistically significant in the lateral, longitudinal, and vertical directions but not in the yaw direction. Based on these results, we suggest that ExacTrac image findings should be further confirmed using CBCT.

Recent imaging-related technologies have garnered a lot of attention recently; they have been used to improve the accuracy of radiotherapy. Particularly, the use of ExacTrac and CBCT has become widespread. Previously, various studies have evaluated the difference between ExacTrac and CBCT in phantom, intracranial, and spinal SRS.

The difference between the ExacTrac and CBCT images of the spine was studied by Chang et al. [11]. They evaluated 16 cases of spinal SBRT in 11 patients, and the difference in the RMS was found to be <2.0 mm in the translation direction and <1.5° in the rotational direction. They analyzed the cases of phantoms and patients separately. In the case of phantoms, there were significant differences in the ExacTrac and CBCT images in the vertical, lateral, and yaw directions, but not in the longitudinal, pitch, and roll directions. However, in the case of patients, there were no significant differences in RMS values between ExacTrac and CBCT in all directions (vertical, longitudinal, lateral, pitch, roll, and yaw). The authors believed that the differences between ExacTrac and CBCT images in spine SBRT, though small, were of clinical significance. Of note, their study’s major drawback was the limited number of cases. In addition, Chang et al. suggested that the following are possible sources of residual setup discrepancy between ExacTrac and CBCT: (1) inter-scan patient motion, (2) the difference in image fusion algorithm, (3) poorer visibility of anatomical structures in digitally reconstructed radiograph and X-ray images, and (4) discrepancy of geometric accuracy for the two systems.

Before this study, we analyzed intracranial SRS with 138 fractions in 107 patients [12]. A significant difference was noted only in the vertical and lateral directions, whereas no significant difference was observed in the longitudinal and yaw directions. However, the previous study was restricted to intracranial sites. Therefore, in this study, we investigated a higher number of cases of spinal SBRT, in contrast to the reduced number of cases reported by Chang et al. [11] and non-inclusion of extracranial sites in our previous study.

Our study analyzed residual setup errors according to the region of the spine involved, something that has not been studied previously; the spine was divided into cervical, thoracic, and lumbar regions to analyze the residual setup error between 6D ExacTrac and 3D CBCT. Notably, in the cervical spine region, there was a significant difference between ExacTrac and CBCT in all directions (lateral, longitudinal, vertical, and yaw). Therefore, particularly in the cervical spine, it is necessary to verify ExacTrac images using CBCT data.

This study has a few limitations. There may be patient movement between ExacTrac and CBCT, thus affecting the findings. In brain SRS, a frameless mask was used to immobilize the patient, whereas in spine SBRT, head and neck masks or Vac-lok were used, depending on the treatment site. The possibility of movement may be greater with the frameless masks for SRS than with the head and neck masks for SBRT. Moreover, this study analyzed the difference between the residual setup errors of 6D ExacTrac and 3D CBCT only through online review. In an offline review of 6D CBCT, it would have been possible to analyze 6D data (lateral, longitudinal, and vertical in the translational direction; and pitch, roll, and yaw in the rotational direction). However, as unwanted records were recorded on the server in offline review, they were excluded from this study.

Conclusions

Taken together, this study analyzed images of 268 fractions in 76 patients who received spine SBRT. Significant differences in the RMS were observed in the lateral, longitudinal, and vertical directions between ExacTrac and CBCT, whereas no significant difference was noted in the yaw direction. Therefore, for spine radiotherapy in clinical practice, verification with CBCT after ExacTrac image acquisition is essential.

Supporting information

S1 Table

(DOCX)

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

This work was supported by the 2020 Yeungnam University Research Grant. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Mantel F, Glatz S, Toussaint A, Flentje M, Guckenberger M. Long-term safety and efficacy of fractionated stereotactic body radiation therapy for spinal metastases. Strahlentherapie und Onkologie 2014;190:1141–1148. 10.1007/s00066-014-0706-1 [DOI] [PubMed] [Google Scholar]
  • 2.Zhuang H, Zhuang H, Lang N, Liu J. Precision Stereotactic Radiotherapy for Spinal Tumors: Mechanism, Efficacy, and Issues. Frontiers in Oncology 2020;10:826. 10.3389/fonc.2020.00826 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Guckenberger M, Mantel F, Gerszten PC, et al. Safety and efficacy of stereotactic body radiotherapy as primary treatment for vertebral metastases: a multi-institutional analysis. Radiation oncology 2014;9:1–8. 10.1186/1748-717X-9-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Gong Y, Xu L, Hongqing Z et al. Efficacy and safety of different fractions in stereotactic body radiotherapy for spinal metastases: A systematic review. Cancer medicine 2019;8:6176–6184. 10.1002/cam4.2546 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Redmond KJ, Lo SS, Fisher C, Sahgal A. Postoperative stereotactic body radiation therapy (SBRT) for spine metastases: a critical review to guide practice. International Journal of Radiation Oncology* Biology* Physics 2016;95:1414–1428. 10.1016/j.ijrobp.2016.03.027 [DOI] [PubMed] [Google Scholar]
  • 6.Greco C, Pares O, Pimentel N, et al. Spinal metastases: From conventional fractionated radiotherapy to single-dose SBRT. Reports of Practical Oncology & Radiotherapy 2015;20:454–463. 10.1016/j.rpor.2015.03.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Oh SA, Yea JW, Kang MK, Park JW, Kim SK. Analysis of the setup uncertainty and margin of the daily exactrac 6D image guide system for patients with brain tumors. PLoS One 2016;11, e0151709. 10.1371/journal.pone.0151709 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Oh SA, Kang MK, Kim SK, Yea JW. Comparison of IMRT and VMAT techniques in spine stereotactic radiosurgery with international spine radiosurgery consortium consensus guidelines. Progress in Medical Physics 2013;24:145–153. [Google Scholar]
  • 9.Kim J, Wen N, Jin JY, et al. Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT. Journal of Applied Clinical Medical Physics 2012;13:124–151. 10.1120/jacmp.v13i3.3729 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kim SH, Oh SA, Yea JW, Park JW. Prospective assessment of inter-or intra-fractional variation according to body weight or volume change in patients with head and neck cancer undergoing radiotherapy PLoS One 2019;14, e0216655. 10.1371/journal.pone.0216655 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Chang Z, Wang Z, Ma J, O’Daniel JC, Kirkpatrick J, Yin FF. 6D image guidance for spinal non-invasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilo-voltage cone-beam CT Radiotherapy and Oncology 2010;95:116–121 10.1016/j.radonc.2009.12.036 [DOI] [PubMed] [Google Scholar]
  • 12.Oh SA, Park JW, Yea JW, Kim SK. Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery PLoS One 2017;12, e0177798 10.1371/journal.pone.0177798 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Benedict SH, Yenice KM, Followill D et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101 Medical physics 2010;37:4078–4101 10.1118/1.3438081 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Stephen Chun

29 Apr 2021

PONE-D-21-03969

Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy

PLOS ONE

Dear Dr. Oh,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Jun 13 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Stephen Chun

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

  1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. . Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

3. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Although there was a previous study addressing a similar question, it was of much smaller scale and there was no region-specific (C-spine vsT-spine vs L-spine) data. Therefore, this study will provide very important info that will guide spine SBRT practice.

1.Vac-lok fixtures refers to just a cradle and not BodyFix, correct? From the very small picture, it looks like it was just a regular Vac-lok. Please clarify.

2. I presume that an Exac-Trac registration was done followed by a CBCT registration and the final shifts were based on CBCT. Please clarify.

3. Table 2- Please fix spacing of the table.

Reviewer #2: Manuscript is sound and presents valuable data that is very relevant to this widespread technology. It may be beneficial to include what margins were used for PTV expansion and whether the magnitude of uncertainty in the setup error would have any clinically significant effects on the dosimetry of the spine SBRT plan (PTV coverage, dose to OAR's, etc). Additionally, what factors are contributing to the residual setup discrepancy? Potential patient movement between ExacTrac and CBCT was mentioned...are there any other potential reasons (insufficient immobilization, anatomical motion, differences in image quality, etc)? Do these factors differ between cervical, thoracic and lumbar spine regions?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 May 27;16(5):e0252234. doi: 10.1371/journal.pone.0252234.r002

Author response to Decision Letter 0


7 May 2021

May 2021

Editorial Board

PLoS One

Dear Editor:

We would like to re-submit the attached manuscript entitled “Evaluation of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography on the image guide system of the Novalis-Tx for spine stereotactic body radiation therapy.” The manuscript ID is PONE-D-21-03969.

The manuscript has been carefully rechecked and appropriate changes (red text in the revised manuscript) have been made in accordance with the reviewers’ suggestions. The responses to their comments have been prepared and attached herewith.

We thank you and the reviewers for the thoughtful suggestions and insights, which have enriched the manuscript and produced a more balanced and better account of the research. We hope that the revised manuscript is now suitable for publication in your journal.

I look forward to your reply.

Sincerely,

Se An Oh

Department of Radiation Oncology

Yeungnam University College of Medicine

Daegu, South Korea

sean.oh5235@gmail.com

Attachment

Submitted filename: 2-1.Response_to_Reviewers_ExacTrac vs CBCT.docx

Decision Letter 1

Stephen Chun

12 May 2021

Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy

PONE-D-21-03969R1

Dear Dr. Oh,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Stephen Chun

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Stephen Chun

18 May 2021

PONE-D-21-03969R1

Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy

Dear Dr. Oh:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Stephen Chun

Academic Editor

PLOS ONE


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES