Abstract
Background
Considering the possibility of nasal self-sampling and the ease of use in performing SARS-CoV-2 antigen-detecting rapid diagnostic tests (Ag-RDTs), self-testing is a feasible option.
Objective
The goal of this study was a head-to-head comparison of diagnostic accuracy of patient self-testing with professional testing using a SARS-CoV-2 Ag-RDT.
Study design
We performed a manufacturer-independent, prospective diagnostic accuracy study of nasal mid-turbinate self-sampling and self-testing with symptomatic adults using a WHO-listed SARS-CoV-2 Ag-RDT. Procedures were observed without intervention. For comparison, Ag-RDTs with nasopharyngeal sampling were professionally performed. Estimates of agreement, sensitivity, and specificity relative to RT-PCR on a combined oro-/nasopharyngeal sample were calculated. Feasibility was evaluated by observer and participant questionnaires.
Results
Among 146 symptomatic adults, 40 (27.4%) were RT-PCR-positive for SARS-CoV-2. Sensitivity with self-testing was 82.5% (33/40; 95% CI 68.1–91.3), and 85.0% (34/40; 95% CI 70.9–92.9) with professional testing. At high viral load (≥7.0 log10 SARS-CoV-2 RNA copies/ml), sensitivity was 96.6% (28/29; 95% CI 82.8–99.8) for both self- and professional testing. Deviations in sampling and testing were observed in 25 out of the 40 PCR-positives. Most participants (80.9%) considered the Ag-RDT as easy to perform.
Conclusion
Laypersons suspected for SARS-CoV-2 infection were able to reliably perform the Ag-RDT and test themselves. Procedural errors might be reduced by refinement of the instructions for use or the product design/procedures. Self-testing allows more wide-spread and frequent testing. Paired with the appropriate information of the public about the benefits and risks, self-testing may have significant impact on the pandemic.
Keywords: COVID-19, SARS-CoV-2, Antigen-detecting rapid diagnostic test, Self-testing, Self-swab
1. Background
Antigen-detecting rapid diagnostic tests (Ag-RDTs) for SARS-CoV-2 have been widely recommended as a complement to RT-PCR [1]. Recent studies have shown the equivalence of anterior nasal, nasal mid-turbinate (NMT), and nasopharyngeal (NP) sampling, and established the feasibility of self-sampling for a WHO-listed SARS-CoV-2 Ag-RDT [2], [3], [4]. Considering the ease-of-use in performing the test, self-testing is a feasible option. In Germany, first SARS-CoV-2 Ag-RDTs received special approvals for self-administration by laypersons on 24 February 2021 [5].
2. Methods
We conducted a manufacturer-independent prospective study of diagnostic accuracy, user acceptability and feasibility of an Ag-RDT when performed by patients themselves, with using a self-collected NMT sample. For comparison, professional NP-sampling with testing on Ag-RDT was performed on the same participant as well as combined oropharyngeal (OP)/NP-sampling for RT-PCR as a reference [2].
The study took place at the ambulatory SARS-CoV-2 testing facility of Charité - Universitätsmedizin Berlin, from 30 November to 11 December 2020. Participants eligible for inclusion were adults with high suspicion of SARS-CoV-2 infection. Participants had to be able to understand the written instructions in German or English, defined as a minimum Common European Framework of Reference for Languages (CEFR) level of B2 (upper intermediate). Participants were enrolled according to laboratory capacity in a consecutive series.
The Ag-RDT evaluated was the STANDARD Q COVID-19 Ag Test (SD Biosensor, Korea), which is also distributed by Roche [6]. The test was chosen in consultation with the Foundation of Innovative New Diagnostics based on data on accuracy and ease-of-use. The NP-sampling and nasal sampling kits were provided with differing flocked swabs, whereby the nasal swab used for self-sampling is less flexible with a larger sampling surface [4]. The participants received written and illustrated instructions in German or English, adapted from the manufacturer's instructions for use. Participants performed the procedures in a separate room without time restrictions. The procedures were observed by a study physician, without answering questions or providing corrections. The ease of use was evaluated with participant and observer questionnaires.
NMT self-sampling (both sides) was followed by professional NP-sampling (through one nostril) for Ag-RDTs and combined OP/NP-sampling (through the other nostril) for RT-PCR. The Ag-RDTs were performed directly after sampling at point-of-care by participants and trained study physicians.
All Ag-RDT results were interpreted by two blinded study physicians with a semi-quantitative assessment, in addition to the participant's interpretation of the self-test [4]. A minimum of 30 positive NP or NMT samples according to Ag-RDT was set, which is recommended by the WHO Emergency Use Listing Procedure to demonstrate sample type equivalency [7].
3. Results
After exclusion of 4 participants (n = 3 not fulfilling the CEFR minimum language criterion, n = 1 lost PCR specimen), 146 adults were included in the analysis (see Supplementary Figure S1). Of these, 40 participants (27.4%) tested positive by RT-PCR. Mean age was 35 years (Standard Deviation [SD] 11.5), and 51.4% were female. All participants were symptomatic at the time of presentation, with a mean duration of 3.4 days (SD 2.0) post symptom onset. Previous use of any RDTs and/or laboratory and/or home-test experience was reported by 29 participants (20.4%). A higher education degree was present in 84 (59.6%) participants. Thirty-eight participants (26.6%) were not native German or English speakers (see Supplementary Table S2).
Self-testing (including self-read-out) yielded a sensitivity of 82.5% (33/40 RT-PCR positives detected; 95% CI 68.1–91.3) and a specificity of 100% (104/104; CI 96.5–100) compared to RT-PCR (Table 1 ). The sensitivity with professional Ag-RDT testing was 85.0% (34/40; CI 70.9–92.9) and specificity was 99.1% (105/106; CI 94.8–99.5). In patients with high viral load (≥7.0 log10 SARS-CoV-2 RNA copies/ml) the sensitivity was 96.6% (28/29; CI 82.8–99.8) for both self-testing and professional testing. The positive percent agreement between self-testing and professional testing on Ag-RDT was 91.4% (32/35; CI 77.6–97.0); the negative percent agreement was 99.1% (108/109; CI 95.0–100).
Table 1.
Antigen-RDT self-testing (NMT sample) with self-readout and professional readout of the result, as well as professional testing (NP sample) in RT-PCR positive outpatients from combined OP/NP swab. Ct-values and viral loads of the paired RT-PCR samples are shown as well as the duration of symptoms. Relevant protocol deviations in the self-test are noted.
| No. | Ag-RDT self-use (NMT) | Ag-RDT prof.-use (NP) | RT-PCR (OP/NP) | Days of Symptoms | Deviations in Ag-RDT self-use | ||
|---|---|---|---|---|---|---|---|
| self-readout | prof-readout | Ct value | Viral load1 | ||||
| 1 | pos. | pos. (+++) | pos. (+++) | 14.822 | 9.57 | 2 | sampling4 |
| 2 | pos. | pos. (+++) | pos. (+++) | 18.293 | 9.30 | 2 | extraction5 |
| 3 | pos. | pos. (+++) | pos. (+++) | 16.202 | 9.16 | 2 | extraction5 |
| 4 | pos. | pos. (+++) | pos. (+++) | 16.342 | 9.12 | 3 | extraction5 |
| 5 | pos. | pos. (+++) | pos. (+++) | 16.892 | 8.96 | 1 | sampling4, extraction5 |
| 6 | pos. | pos. (+++) | pos. (+++) | 19.513 | 8.94 | 2 | extraction5, volume6 |
| 7 | pos. | pos. (+++) | pos. (+++) | 17.432 | 8.80 | 2 | |
| 8 | pos. | pos. (+++) | pos. (+++) | 20.303 | 8.71 | 4 | |
| 9 | pos. | pos. (++) | pos. (+++) | 17.742 | 8.70 | 3 | |
| 10 | pos. | pos. (+++) | pos. (+++) | 20.353 | 8.69 | 4 | |
| 11 | pos. | pos. (+++) | pos. (+++) | 17.942 | 8.64 | 3 | extraction5, volume6 |
| 12 | pos. | pos. (+++) | pos. (+++) | 20.723 | 8.58 | 1 | |
| 13 | pos. | pos. (++) | neg. | 18.502 | 8.48 | 1 | sampling4, extraction5 |
| 14 | pos. | pos. (+++) | pos. (+++) | 21.343 | 8.40 | 3 | |
| 15 | pos. | pos. (+++) | pos. (+++) | 18.962 | 8.34 | 7 | extraction5, volume6 |
| 16 | pos. | pos. (+++) | pos. (+++) | 19.262 | 8.25 | 2 | sampling4, extraction5 |
| 17 | pos. | pos. (+++) | pos. (+++) | 19.292 | 8.24 | 3 | |
| 18 | pos. | pos. (+++) | pos. (+++) | 21.943 | 8.22 | 1 | extraction5 |
| 19 | pos. | pos. (+++) | pos. (+++) | 19.482 | 8.19 | 4 | |
| 20 | pos. | pos. (+++) | pos. (+++) | 19.492 | 8.18 | 4 | sampling4 |
| 21 | pos. | pos. (+++) | pos. (+++) | 19.522 | 8.17 | 4 | sampling4, extract.5, vol.6 |
| 22 | pos. | pos. (+++) | pos. (+++) | 19.742 | 8.11 | 2 | sampling4, extraction5 |
| 23 | pos. | pos. (+++) | pos. (+++) | 19.952 | 8.05 | 3 | |
| 24 | pos. | pos. (++) | pos. (+++) | 20.672 | 7.83 | 3 | extraction5 |
| 25 | pos. | pos. (+++) | pos. (+++) | 23.713 | 7.70 | 6 | sampling4 |
| 26 | pos. | pos. (++) | pos. (+++) | 23.993 | 7.62 | 3 | extraction5, volume6 |
| 27 | pos. | pos. (+++) | pos. (+++) | 21.982 | 7.44 | 7 | |
| 28 | pos. | pos. (++) | pos. (+++) | 24.893 | 7.35 | 8 | sampling4, extraction5 |
| 29 | neg. | pos. (++) | pos. (++) | 22.672 | 7.24 | 3 | read-out7 |
| 30 | pos. | pos. (++) | pos. (++) | 24.862 | 6.59 | 7 | sampling4 |
| 31 | neg. | neg. | pos. (+) | 25.072 | 6.52 | 5 | |
| 32 | pos. | pos. (+++) | pos. (++) | 28.583 | 6.26 | 8 | |
| 33 | neg. | neg. | neg. | 27.052 | 5.94 | 1 | |
| 34 | pos. | pos. (++) | pos. (+) | 29.712 | 5.15 | 5 | extraction5, volume6 |
| 35 | pos. | pos. (++) | pos. (+) | 30.032 | 5.05 | 7 | extraction5 |
| 36 | pos. | pos. (+) | pos. (++) | 30.282 | 4.98 | 7 | sampling4, extraction5 |
| 37 | neg. | neg. | neg. | 35.693 | 4.15 | 7 | |
| 38 | neg. | neg. | neg. | 33.292 | 3.91 | 14 | sampling4, extract.5, vol.6 |
| 39 | neg. | neg. | neg. | 34.542 | 3.71 | 8 | |
| 40 | neg. | neg. | neg. | 34.992 | 3.41 | 14 | extraction5 |
| Sensitivity 33/40 (82.5%) | Sensitivity 34/40 (85.0%) | Sensitivity 34/40 (85.0%) | |||||
log10 SARS-CoV-2 RNA copies/ml;
TIB Molbiol assay, E-gene target;
roche cobas SARS-CoV-2 assay (E-gene, T2 target);
sampling of possible reduced quality (unilateral, short, different position);
extraction of possible reduced quality related to stirring of the swab or squeezing of the buffer tube;
vol = volume: incorrect number of drops applied on the test device;
read-out: misinterpretation of the test-result.
One patient with a positive self-test had falsely interpreted his result as negative. Regarding divergent results, there was one participant yielding a false-negative result with self-testing who performed the test according to instructions, and one false-negative result with professional testing despite a high viral load. In the latter patient, NP-sampling was not optimal due to poor tolerance, which occurs rather frequently in clinical practice.
Inter-rater reliability for the double read-out of the self-test by the participant and the study physician was very high (kappa 0.98) and close to that of two study physicians interpreting the self-test (kappa 1) or the professionally performed test (kappa 0.98).
Deviations of self-sampling included a more vertically-directed angle for sampling (23%), incorrect depth of insertion of the swab (4% too superficial, 10% too deep), reduced intensity of swabbing (as to duration 24%, rotations 10% and rubbing 42%) and unilateral NMT-sampling (6%). One intervention by the study physician was necessary because of a possible risk of injury when the patient tried to insert the swab upside down into the nose.
Deviations of self-testing were observed for the specimen extraction (less stirring of the swab 4%, inadequate tube squeezing while stirring 35% and while removing the swab 33%). Furthermore, it proved difficult to apply exactly 4 drops to the sample well of the test device (27% applying more, 4% less). Several drops coming out at once was the main problem (see Supplementary Table S3).
Participants pointed out that their test performance may have been impaired by nervousness (3%), fever or dizziness, poor concentration, feeling cold, aversion to self-sampling, language barrier and limited fine motor skills (1% each). On a scale from 1 (easy) to 5 (difficult), 114 (81%) participants stated that the test was easy to perform (scale 1/2); 23 (16%) found it medium easy/difficult (scale 3), and 4 (3%) difficult (scale 4).
4. Discussion
The data presented here demonstrates the feasibility and accuracy of self-testing in an unselected population and lay the ground for potential broader use. Self-testing with Ag-RDTs could not only alleviate overstretched RT-PCR capacity and medical personnel, but also may result in increased access to frequent testing and significant impact on the pandemic [8,9]. If implemented for society at large, recent modeling data suggest that repeated screening for SARS-CoV-2, combined with immediate reporting, isolation, and quarantine, can greatly reduce viral transmission. In this model, test sensitivity is of minor importance [8].
However, self-testing should be accompanied by widespread public campaigns informing about limited sensitivity, the importance of complementary hygiene measures, e.g., mask use, physical distancing, and the necessity of self-quarantine in case of a positive test. Reporting requirements in positive cases should be clarified. As long as Ag-RDTs test capacities are limited, deployment must be prioritized, also with a global perspective. Individually, Ag-RDT self-testing could contribute to resumption of a certain degree of normal life, e.g., self-testing could be done before visiting nursing homes.
The study is limited as it was performed in a single center. Patients were rather young and educated. The generalizability of the findings and applicability to settings with different prevailing patient characteristics needs to be confirmed.
In conclusion, we demonstrate that symptomatic laypersons can reliably perform a SARS-CoV-2 lateral-flow Ag-RDT and test themselves. Procedural errors might be reduced by refinement of the Ag-RDTs for self-testing, such as modified instructions for use or product design/procedures. Paired with the appropriate information of the public about the benefits and risks, self-testing may have significant impact on the pandemic.
CRediT authorship contribution statement
Andreas K. Lindner: Conceptualization, Methodology, Writing - original draft, Writing - review & editing. Olga Nikolai: Methodology, Investigation, Formal analysis, Writing - review & editing. Chiara Rohardt: Methodology, Investigation, Writing - review & editing. Franka Kausch: Investigation, Writing - review & editing. Mia Wintel: Investigation, Writing - review & editing. Maximilian Gertler: Project administration, Writing - review & editing. Susen Burock: Investigation, Writing - review & editing. Merle Hörig: Investigation, Writing - review & editing. Julian Bernhard: Investigation, Writing - review & editing. Frank Tobian: Formal analysis, Writing - review & editing. Mary Gaeddert: Formal analysis, Writing - review & editing. Federica Lainati: Supervision, Writing - review & editing. Victor M. Corman: Investigation, Formal analysis, Writing - review & editing. Terry C. Jones: Investigation, Formal analysis, Writing - review & editing. Jilian A. Sacks: Supervision, Writing - review & editing. Joachim Seybold: Supervision, Project administration. Claudia M. Denkinger: Conceptualization, Methodology, Supervision, Writing - review & editing. Frank P. Mockenhaupt: Conceptualization, Methodology, Supervision, Writing - original draft, Writing - review & editing.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
Ethics
This study was approved by the ethics committee of Charité - Universitätsmedizin (EA1/371/20).
Data availability
All raw data and analysis code are available upon request to the corresponding author.
Funding statement
C.M. Denkinger reports grants from Foundation for Innovative New Diagnostics (FIND), and Ministry of Science, Research and Culture, State of Baden Wuerttemberg, Germany, to conduct of the study. J.A. Sacks reports grants from UK Department for International Development (DFID, recently replaced by FCMO) [grant number 300341–102], World Health Organization (WHO) and Unitaid [grant number: 2019-32-FIND MDR], to conduct of the study. FIND supplied the test kits for the study. The study was support by Heidelberg University Hospital and Charité University Hospital internal funds.
Acknowledgements
Heike Rössig, Elisabeth Linzbach, Katja von dem Busche, Stephanie Padberg, Melanie Bothmann, Zümrüt Tuncer, Stefanie Lunow, Beate Zimmer, Astrid Barrera Pesek, Sabrina Pein, Verena Haack, Oliver Deckwart, Birgit Zittlau, Andrea Junge.
Footnotes
Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.jcv.2021.104874.
Appendix. Supplementary materials
References
- 1.European Centre for Disease Prevention and Control. Options for the use of rapid antigen tests for COVID-19 in the EU/EEA and the UK. November 19 2020. https://www.ecdc.europa.eu/en/publications-data/options-use-rapid-antigen-tests-covid-19-eueea-and-uk. Date accessed: March 20, 2021.
- 2.Lindner A.K., Nikolai O., Kausch F., et al. Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with self-collected nasal swab versus professional-collected nasopharyngeal swab. Eur. Respir. J. 2020 doi: 10.1183/13993003.03961-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lindner A.K., Nikolai O., Rohardt C., et al. Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with professional-collected nasal versus nasopharyngeal swab. Eur. Respir. J. 2021 doi: 10.1183/13993003.04430-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Nikolai O., Rohardt C., Tobian F., et al. Anterior nasal versus nasal mid-turbinate sampling for a SARS-CoV-2 antigen-detecting rapid test: does localisation or professional collection matter? medRxiv. 2021 doi: 10.1101/2021.02.09.21251274. 2021.02.09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Federal Institute for Drugs and Medical Devices. Antigen Tests For SARS-CoV-2. https://www.bfarm.de/EN/MedicalDevices/AntigenTests/_node.html. Date last accessed: March 20 2021.
- 6.S.D. Biosensor. COVID-19 Ag STANDARD Q COVID-19 Ag Test. http://sdbiosensor.com/product/product_view?product_no=241. Date accessed: June 7, 2021.
- 7.WHO. Instructions and requirements for Emergency Use Listing (EUL) submission: in vitro diagnostics detecting SARS-CoV-2 nucleic acid and rapid diagnostics tests detecting SARS-CoV-2 antigens. https://extranet.who.int/pqweb/sites/default/files/documents/PQDx_347_NAT-antigen_instructions.pdf. Version 4, June 2020. Date last accessed March 20, 2021.
- 8.Larremore D.B., Wilder B., Lester E., et al. Test sensitivity is secondary to frequency and turnaround time for COVID-19 surveillance. medRxiv. 2020 doi: 10.1101/2020.06.22.20136309. 2020.06.22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Paltiel A.D., Zheng A., Walensky R.P. Assessment of SARS-CoV-2 Screening Strategies to Permit the Safe Reopening of College Campuses in the United States. JAMA Netw. Open. 2020;3(7) doi: 10.1001/jamanetworkopen.2020.16818. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All raw data and analysis code are available upon request to the corresponding author.