Preclinical evaluation of passive disinfection caps with a long-term catheter for the prevention of catheter-related bloodstream infection in pediatric cancer patients

Präklinische Bewertung von passiven Desinfektionskappen mit einem Langzeitkatheter zur Prävention von katheterbedingten Blutstrominfektionen bei pädiatrischen Krebspatienten

Arne Simon 1,*, Wolfgang Fahrendorf 2, Guido Hitschmann 3
PMCID: PMC8167456  PMID: 34113538

Abstract

The use of passive disinfection devices (disinfection caps) may be a beneficial part of a maintenance care bundle, aiming at the prevention of catheter-related bloodstream infections in pediatric cancer patients. This preclinical in vitro investigation tested the visual and mechanical integrity of a Broviac™ catheter hub after simulation testing with 122 3M™ CurosTM Stopper Disinfection Caps for Open Female Luers repeatedly attached and removed over 6 months. We found that these catheter hubs were compatible, fully operational, and airtight with use of 3M Curos stopper caps after 6 months of use with 122 caps per catheter hub.

Keywords: bloodstream infection, central venous catheter, pediatric cancer patients, scrub the hub, passive disinfection device

Introduction

Long-term central venous access devices (CVAD), such as a Broviac Catheter [1] or a Hickman Catheter [2] are indispensable tools in pediatric oncology centers (POC), facilitating the administration of cytotoxic chemotherapy, parenteral nutrition, and supportive medication, such as antibiotics, antifungals, analgesics, and blood products. In addition, these long-term catheters allow the frequent, painless sampling of blood for laboratory or microbiological investigations. The use of CVADs has been linked to catheter-associated bloodstream infections (CABSI), which may derive from the catheter entry site (extraluminal source) or the hub or inner surface of the catheter (intraluminal source) [3]. Most POCs have implemented maintenance-care bundles to prevent these adverse events [4], [5].

Frequent manipulation of the catheter hub or any other access point (such as 3-way stopcocks) increases the risk of catheter colonization and subsequent CABSI. In this regard, national [6] and international guidelines [7], [8] concerning maintenance care of vascular catheters recommend hand disinfection before any manipulation and emphasize the necessity of disinfecting the catheter hub (or any other access point) [9](9) before each manipulation. (Some POCs use needleless connection devices, but a significant impact of these devices on the prevention of CRBSI has not been confirmed and the membrane of the device still needs to be disinfected.)

Hub disinfection of a central venous access catheter can be accomplished by three different methods:

  1. Scrubbing the hub with a sterile cloth, which contains antiseptics (such as isopropanol with or without chlorhexidine or octenidine; scrub the hub) [10], [11], [12].

  2. Application of the antiseptic as a spray, while holding the hub on a sterile gauze pad [13].

  3. Using sterile caps which release IPA after they have been screwed on the Luer lock hub (passive disinfection devices).

To be effective, methods 1 and 2 need a defined dwell time for the antiseptic (e.g., at least 15 seconds). One caveat of infection prevention in clinical practice is that each procedure depends not only on knowledge (education) and skills (training), but also on adherence to recommendations/guidelines of the responsible healthcare worker [14]. Taking limited personnel and time into consideration, any time-consuming procedure may not be consistently followed in a busy unit [15].

External scrubbing of a female Luer lock with a cloth is not an effective method to disinfect the inner surface of the connection [16], [17]. In pediatric oncology, this is of particular interest, since health care workers often use central venous access catheters for blood sampling, and any blood residue must be removed thoroughly.

Method 2 is recommended by the German Commission for Hospital Hygiene and Infection Control (KRINKO) as one feasible alternative [18], but it may cause antiseptic inhalation exposure to the patient and the health care worker.

From a clinical perspective, the use of passive disinfection devices (method 3) intuitively appears to be an attractive alternative. Recent studies including cancer patients demonstrated the feasibility and effectiveness of this approach [15], [17], [19], [20], [21], [22], [23], [24], [25]. Referring to a recent meta-analysis [26], the available studies still are heterogeneous and all have certain limitations (e.g., interrupted time-series investigations instead of prospective randomization). The KRINKO advises attending physicians to consider the use of passive disinfection devices in high-risk clinical units with frequent catheter manipulation and high CABSI rates [18].

Before the implementation of method 3 in the Pediatric Hematology and Oncology unit of the Children’s Hospital Medical Center at Saarland University Hospital, a literature search and personal communication with the manufacturer of the Broviac™ catheters revealed that no preclinical data were available to confirm that the hub of these catheters tolerates the long-term use of a passive, IPA-containing disinfection device in terms of stability and mechanical integrity. In an individual patient, a central venous access catheter is often used for 6 to 12 months until intensive anticancer treatment is completed. Thus, the central venous access catheter must tolerate the use of the passive, IPA-containing disinfection device in the longer term. The preclinical investigation presented here was performed to confirm this before passive disinfection devices were integrated as a routine component of our preventive bundle [4].

Methods

In our unit, Broviac™ catheters are flushed (with ready-to-use syringes containing 10 ml of sterile 0.9% sodium chloride) and locked with heparin (2–3 ml; 100 IE/ml, 5 ml single use only vials) at least once a week. During their intensive treatment period, many patients must visit the pediatric cancer outpatient clinic twice a week for clinical examination and laboratory investigation.

In this regard, the in vitro study contained two changes of the passive disinfection cap (3M™ Curos™ Stopper Disinfecting Cap for Open Female Luers) per week. To simulate long-term use, the in vitro investigation was continued for 6 months. The materials used were three 6.6 F Broviac™ CV Catheter repair kits 1.0 mm Lumen (BARD Access Systems, Inc., Salt Lake City, UT, USA; Lot REDS2687, expiration date: 2022-04) and 3M™ Curos™ Stopper Disinfecting Cap (Teal) for Open Female Luers (3M Health Care, St. Paul, MN, USA; Lot 6523027, expiration date: 2021-06-03, 6542554, expiration date: 2021-09-26). The white hubs of three catheter repair kits were tested throughout this experiment. Curos stopper caps were placed on the hubs, following the instructions for use. Initially, the caps were changed five times per day, leaving at least 90 minutes between the changes. This sequence of events was performed Monday through Friday for two consecutive weeks. For an additional 24 consecutive weeks, the caps were changed three times per week (Mon, Wed, Fri). The caps were left on the catheter hubs between the changes. The test samples were stored at room temperature (20°–25° C). After the test period of 26 weeks, the catheter hubs to which Curos stopper caps were attached were visually inspected using a microscope to identify any type of mechanical degradation such as crazing, hairline cracks, chips, or warping. The catheter hubs were then exposed to 300 mbar air pressure and soaked in a water bath to verify air tightness. Eventually, each catheter hub tested had a total of 122 caps attached and removed during the 6-month test period.

Results

No visible differences were observed between catheter hubs treated with 3M Curos stopper disinfection caps. The catheter hub-cap connections were fully operational and airtight after 6 months of testing. In summary, from the results of this testing, it can be concluded that the catheter hubs were compatible with use of 3M Curos stopper caps through 6 months of use with 122 caps per catheter hub. The official test protocol is available as an online supplement to this article.

Discussion

This in vitro study confirmed that the mechanical integrity of the tested Broviac™ catheter model’s catheter hub was not compromised by regular use of 3M Curos stopper caps over a six-month period.

Formally, it should be acknowledged that the results reported here apply only to the mentioned catheter hubs with the specific lot number. On the other hand, there is no reason to assume any differences between the standard materials used in Broviac™ catheters of the same manufacturer with different lot numbers.

For clinical practice, this result is very important. These catheters are implanted by a pediatric surgeon in an operation theater and remain in place for 6 to 12 months or even longer. Theoretically, the availability of Broviac™ repair sets allows replacement of the most distal part of the catheter (including the hub). Unfortunately, this repair requires extensive manipulation, results in a diminished stability of the catheter against traction force, and the defects liability of the medical device extends from the manufacturer to the physician who performs the repair. Therefore, maintaining the integrity of the components of the Broviac™ catheter and avoiding repairs is advantageous.

The German Commission for Hospital Hygiene and Infection Control states that – in this clinical field – only materials that tolerate alcohol disinfection should be used (e.g., 3-way stopcocks) [18]. Other manufacturers may use the method presented here to investigate their medical devices accordingly. Finally, this investigation enabled us to include the IPA-containing passive disinfection cap into our routine maintenance bundle for the prevention of CABSI (4).

Notes

Competing interests

GH and WF are employees of 3M Deutschland GmbH, Neuss. 3M develops and sells medical devices for the prevention of catheter-related infections, including the passive disinfection device investigated in this study.

AS has received honoraria as a speaker and trainer in educational sessions concerning the prevention of catheter-related bloodstream infections (2019, 2020) and for the participation in an expert advisory board on this topic (2020) from 3M.

Since 2004, AS has been an appointed member of the Commission for Hospital Hygiene and Infection Control, affiliated with the Robert Koch Institute Berlin and coordinates the German recommendations for the prevention of vascular catheter-related infections. AS is a member of the management board (2nd chairman) of the German Society for Pediatric Infectious Diseases.

References

  • 1.Broviac JW, Cole JJ, Scribner BH. A silicone rubber atrial catheter for prolonged parenteral alimentation. Surg Gynecol Obstet. 1973;136(4):602–606. [PubMed] [Google Scholar]
  • 2.Hickman RO, Buckner CD, Clift RA, Sanders JE, Stewart P, Thomas ED. A modified right atrial catheter for access to the venous system in marrow transplant recipients. Surg Gynecol Obstet. 1979 Jun;148(6):871–875. [PubMed] [Google Scholar]
  • 3.Ammann RA, Laws HJ, Schrey D, Ehlert K, Moser O, Dilloo D, Bode U, Wawer A, Schrauder A, Cario G, Laengler A, Graf N, Furtwängler R, Simon A. Bloodstream infection in paediatric cancer centres--leukaemia and relapsed malignancies are independent risk factors. Eur J Pediatr. 2015 May;174(5):675–686. doi: 10.1007/s00431-015-2525-5. [DOI] [PubMed] [Google Scholar]
  • 4.Furtwängler R, Laux C, Graf N, Simon A. Impact of a modified Broviac maintenance care bundle on bloodstream infections in paediatric cancer patients. GMS Hyg Infect Control. 2015 Nov 16;10:Doc15. doi: 10.3205/dgkh000258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Simon A, Furtwängler R, Graf N, Laws HJ, Voigt S, Piening B, Geffers C, Agyeman P, Ammann RA. Surveillance of bloodstream infections in pediatric cancer centers - what have we learned and how do we move on? GMS Hyg Infect Control. 2016 May 12;11:Doc11. doi: 10.3205/dgkh000271. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Simon A, Furtwängler R, Laws HJ, Greiner J, Lehrnbecher T, Ammann R, Schilling AF, Graf N. Evidenz-basierte Empfehlungen zur Anwendung dauerhaft implantierter, zentralvenöser Zugänge in der pädiatrischen Onkologie. 5. Auflage. Wiesbaden: mhp-Verlag; 2017. [Google Scholar]
  • 7.Cellini M, Bergadano A, Crocoli A, Badino C, Carraro F, Sidro L, Botta D, Pancaldi A, Rossetti F, Pitta F, Cesaro S. Guidelines of the Italian Association of Pediatric Hematology and Oncology for the management of the central venous access devices in pediatric patients with onco-hematological disease. J Vasc Access. 2020 Nov; doi: 10.1177/1129729820969309. [DOI] [PubMed] [Google Scholar]
  • 8.O'Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, Lipsett PA, Masur H, Mermel LA, Pearson ML, Raad II, Randolph AG, Rupp ME, Saint S Healthcare Infection Control Practices Advisory Committee. Guidelines for the prevention of intravascular catheter-related infections. Am J Infect Control. 2011 May;39(4 Suppl 1):S1–34. doi: 10.1016/j.ajic.2011.01.003. [DOI] [PubMed] [Google Scholar]
  • 9.Holroyd JL, Vasilopoulos T, Rice MJ, Rand KH, Fahy BG. Incidence of central venous catheter hub contamination. J Crit Care. 2017 Jun;39:162–168. doi: 10.1016/j.jcrc.2017.02.035. [DOI] [PubMed] [Google Scholar]
  • 10.Simmons S, Bryson C, Porter S. “Scrub the hub”: cleaning duration and reduction in bacterial load on central venous catheters. Crit Care Nurs Q. 2011 Jan-Mar;34(1):31–35. doi: 10.1097/CNQ.0b013e3182048073. [DOI] [PubMed] [Google Scholar]
  • 11.Soothill JS, Bravery K, Ho A, Macqueen S, Collins J, Lock P. A fall in bloodstream infections followed a change to 2% chlorhexidine in 70% isopropanol for catheter connection antisepsis: a pediatric single center before/after study on a hemopoietic stem cell transplant ward. Am J Infect Control. 2009 Oct;37(8):626–630. doi: 10.1016/j.ajic.2009.03.014. [DOI] [PubMed] [Google Scholar]
  • 12.Horvath B, Norville R, Lee D, Hyde A, Gregurich M, Hockenberry M. Reducing central venous catheter-related bloodstream infections in children with cancer. Oncol Nurs Forum. 2009 Mar;36(2):232–238. doi: 10.1188/09.ONF.232-238. [DOI] [PubMed] [Google Scholar]
  • 13.Engelhart S, Exner M, Simon A. In vitro study on the disinfectability of two split-septum needle-free connection devices using different disinfection procedures. GMS Hyg Infect Control. 2015 Dec 9;10:Doc17. doi: 10.3205/dgkh000260. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Chittick P, Sherertz RJ. Recognition and prevention of nosocomial vascular device and related bloodstream infections in the intensive care unit. Crit Care Med. 2010 Aug;38(8 Suppl):S363–S372. doi: 10.1097/CCM.0b013e3181e6cdca. [DOI] [PubMed] [Google Scholar]
  • 15.Wright MO, Tropp J, Schora DM, Dillon-Grant M, Peterson K, Boehm S, Robicsek A, Peterson LR. Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection. Am J Infect Control. 2013 Jan;41(1):33–38. doi: 10.1016/j.ajic.2012.05.030. [DOI] [PubMed] [Google Scholar]
  • 16.Moureau NL, Flynn J. Disinfection of Needleless Connector Hubs: Clinical Evidence Systematic Review. Nurs Res Pract. 2015;2015:796762. doi: 10.1155/2015/796762. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Holroyd JL, Paulus DA, Rand KH, Enneking FK, Morey TE, Rice MJ. Universal intravenous access cleaning device fails to sterilize stopcocks. Anesth Analg. 2014 Feb;118(2):333–343. doi: 10.1213/ANE.0000000000000059. [DOI] [PubMed] [Google Scholar]
  • 18.Prävention von Infektionen, die von Gefäßkathetern ausgehen : Teil 1 – Nichtgetunnelte zentralvenöse Katheter Empfehlung der Kommission für Krankenhaushygiene und Infektionsprävention (KRINKO) beim Robert Koch-Institut. Bundesgesundheitsblatt Gesundheitsforschung Gesundheitsschutz. 2017 Feb;60(2):171–206. doi: 10.1007/s00103-016-2487-4. [DOI] [PubMed] [Google Scholar]
  • 19.Menyhay SZ, Maki DG. Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap. Infect Control Hosp Epidemiol. 2006 Jan;27(1):23–27. doi: 10.1086/500280. [DOI] [PubMed] [Google Scholar]
  • 20.Menyhay SZ, Maki DG. Preventing central venous catheter-associated bloodstream infections: development of an antiseptic barrier cap for needleless connectors. Am J Infect Control. 2008 Dec;36(10):S174. doi: 10.1016/j.ajic.2008.10.006. [DOI] [PubMed] [Google Scholar]
  • 21.Buchman AL, Spapperi J, Leopold P. A new central venous catheter cap: decreased microbial growth and risk for catheter-related bloodstream infection. J Vasc Access. 2009 Jan-Mar;10(1):11–21. doi: 10.1177/112972980901000103. [DOI] [PubMed] [Google Scholar]
  • 22.Maki DG. In vitro studies of a novel antimicrobial luer-activated needleless connector for prevention of catheter-related bloodstream infection. Clin Infect Dis. 2010 Jun;50(12):1580–1587. doi: 10.1086/652764. [DOI] [PubMed] [Google Scholar]
  • 23.Sweet MA, Cumpston A, Briggs F, Craig M, Hamadani M. Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line-associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit. Am J Infect Control. 2012 Dec;40(10):931–934. doi: 10.1016/j.ajic.2012.01.025. [DOI] [PubMed] [Google Scholar]
  • 24.Ramirez C, Lee A, Welch K. Central Venous Catheter Protective Connector Caps Reduce Intraluminal Catheter-Related Infection. Journal of Vascular Access. 2012;17(4):210–213. doi: 10.1016/j.java.2012.10.002. [DOI] [Google Scholar]
  • 25.Merrill KC, Sumner S, Linford L, Taylor C, Macintosh C. Impact of universal disinfectant cap implementation on central line-associated bloodstream infections. Am J Infect Control. 2014 Dec;42(12):1274–1277. doi: 10.1016/j.ajic.2014.09.008. [DOI] [PubMed] [Google Scholar]
  • 26.Voor In 't Holt AF, Helder OK, Vos MC, Schafthuizen L, Sülz S, van den Hoogen A, Ista E. Antiseptic barrier cap effective in reducing central line-associated bloodstream infections: A systematic review and meta-analysis. Int J Nurs Stud. 2017 Apr;69:34–40. doi: 10.1016/j.ijnurstu.2017.01.007. [DOI] [PubMed] [Google Scholar]

Articles from GMS Hygiene and Infection Control are provided here courtesy of German Medical Science

RESOURCES