Monitoring of QTc interval in patients with COVID-19. First experience with a portable ECG-recording device

María Abellas-Sequeiros 1, Cristina Lozano-Granero 1,, Cristina García-Sebastián 1, Eduardo Franco-Díez 1, Antonio Hernández-Madrid 1, Javier Moreno-Planas 1, Jaime Masjuan-Vallejo 2, Marcelo Sanmartín-Fernández 1, Jose Luis Zamorano-Gómez 1
PMCID: PMC8169181  PMID: 33843037

Treatment of coronavirus disease 2019 (COVID-19), as a new entity lacking sufficient evidence-based pharmacotherapy, was initially based in drugs with in-vitro activity against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) such as hydroxychloroquine, azithromycin and lopinavir/ritonavir. Each of these drugs is known to prolong QT interval [13]. This phenomenon is of utmost importance, as it is associated with early after-depolarizations, which can generate short-coupled premature action potentials that lead to ventricular fibrillation and sudden cardiac death [4]. Therefore, it seems mandatory to implement monitoring systems, in a scenario where the use of portable electrocardiographic-recording devices could be useful.

This is an observational prospective study in a tertiary hospital. Patients hospitalized for confirmed SARS-CoV-2 pneumonia and receiving specific treatment with either hydroxychloroquine, azithromycin or lopinavir/ritonavir (or combination) were included. Patients were classified into three groups: group 1 received one drug, group 2 received a combination of two drugs and group 3 received a combination of three drugs. Patients in which the electrocardiogram (ECG) tracing obtained with the mobile ECG recording device was of poor quality were excluded from the study.

Clinical information and 12-lead ECG recordings were collected. Baseline QT interval was measured in limb leads using the admission-ECG. The patients received pharmacological treatment according to center-specific guidelines. Verbal informed consent was obtained from patients and approval from the ethics committee of our center was received. A Kardia-Mobile 6L® (AliveCor, Inc., Mountain View, California) was used for ECG monitoring. This device, approved by the Food and Drug Administration for QT interval monitoring, consists of a lightweight 3-electrode hardware capable of registering a 6-lead ECG and a smartphone software application. ECGs were recorded periodically with the physicians’ criteria. A subset of patients at early discharge were offered one device to daily self-monitoring. Corrected QT (QTc) interval was calculated using the Bazett formula. Primary outcome was to describe the usefulness of tele-monitoring in COVID-19 patients for management of QT-prolonging drugs by assessing the proportion of patients with significant ECG changes that imply a change in management. Secondary outcome was to evaluate the inter-observer reproducibility of QT interval quantification using this gadget.

The Shapiro-Wilk test was used to assess normality in continuous variables. Continuous variables were compared using the Wilcoxon or Kruskal-Wallis tests, as appropriate. The Wilcoxon signed-rank test was used for paired measurements. Bilateral p values < 0.05 were considered statistically significant. To test inter-observer variability of QTc measurements, 120 random ECGs were analysed by two independent cardiologists. Intraclass correlation coefficient was obtained (two-way random, average measures, absolute agreement). Statistical analysis was performed using SPSS (IBM Inc., Armonk, NY).

A total of 70 patients fulfilled criteria. Tracings obtained with the portable device were of sufficient quality to provide an accurate QT interval measurement in 69 of them (98.6%). Mean follow-up was 6.2 ± 8.4 days. Characteristics of the different groups are shown in Table 1. An increase in QTc interval was observed in all treatment groups, being non-significant in group 1 (group 1: 340 [320.0–350.0] vs. 418 [384.5–460.5], p = 0.109). Significance was reached in both group 2 (408 [377.2–423.5] vs. 426 [412.0–450.0], p = 0.002) and group 3 (394 [373.0–422.5] vs. 435 [405.0–450.0], p = 0.001). Eighteen (25.7%) patients developed a QTc interval > 450 ms. These patients were significantly older (67.5 ± 14.9 vs. 55.0 ± 16.0 years, p = 0.004) and were more likely diagnosed of chronic obstructive pulmonary disease (33.3% vs. 5.9%, p = 0.009) and previous cardiopathy (50% vs. 13.7%, p = 0.007). Taking into consideration the information given by the ECG monitoring devices, physicians adopted a different approach in the management of 12 (17.4%) patients. Treatment was modified in 9 (13%) patients because of prolongation of the QTc interval (1 of them also developed first degree atrioventricular block, with a PR interval up to 280 ms); 2 (2.9%) patients required monitoring intensification (1 of them because of first degree atrioventricular block and the other because of prolongation of the QTc interval) and anticoagulation was started in one patient because of atrial fibrillation diagnosis.

Table 1.

Clinical characteristics and QTc interval evolution in the three treatment groups.

Group 1 (one drug)
N = 9 (13.0%)
Group 2 (two drugs)
N = 37 (53.6%)
Group 3 (three drugs)
N = 23 (33.3%)
P
Clinical characteristics
Age [years] 55.0 ± 18.3 66.0 ± 16.2 58.0 ± 15.8 0.248
Male sex 6 (66.7%) 25 (67.6%) 18 (78.3%) 0.643
Dyslipidemia 5 (55.6%) 9 (24.3%) 7 (30.4%) 0.749
Diabetes 7 (77.8%) 4 (10.8%) 5 (21.7%) 0.525
Hypertension 3 (33.3%) 16 (43.2%) 8 (34.8%) 0.387
Previous cardiopathy 6 (66.7%) 11 (29.7%) 3 (13.0%) 0.305
COPD 7 (77.8%) 6 (16.2%) 1 (8.7%) 0.679
Atrial fibrillation 1 (11.1%) 3 (8.1%) 0 0.297
Mortality 0 4 (10.8%) 0 0.159
QTc interval
QTc interval at admission [ms] 340.0 [320.0–350.0] 408.0 [377.2–423.5] 394.0 [373.0–422.5] 0.039
Maximal QTc interval [ms] 418.0 [384.5–460.5] 426.0 [412.0–450.0] 435.0 [405.0–450.0] 0.085
QTc > 450 ms 3 (33.3%) 9 (24.3%) 6 (26.1%) 0.859
ΔQTc > 60 ms 1 (11.1%) 4 (15.4%) 3 (9.7%) 0.818

Qualitative variables are shown as number (percentage); quantitative variables are shown as mean ± standard deviation or median [interquartile range], as appropriate; COPD — chronic obstructive pulmonary disease

The subgroup of 16 patients who were given a recorder to continue daily monitoring experienced also significant prolongation of the QTc interval (381.0 [372.0–401.5] at admission vs. 424.0 [406.0–436.7] at peak, p = 0.002), with mean time to reach the maximum QTc: 2 ± 1.8 days). By the end of the monitoring phase, QTc intervals were shorter than those at admission (381.0 [372.0–401.5] vs. 350.0 [334.2–369.0]; p = 0.019).

Inter-observer intraclass correlation coefficient was 0.824, 95% confidence interval 0.733–0.882 (good agreement).

Herein, is presented one of the first series of patients both hospitalized [5, 6] and discharged with COVID-19 with electrocardiographic surveillance using a portable ECG recorder [7]. The device proved to be useful for ECG monitoring in these patients, detecting ECG anomalies of enough importance to promote a change in management in 17.4% of them. These anomalies consisted not only in QTc interval prolongation, but also in PR interval prolongation and atrial arrhythmias. These small, light gadgets allow physicians to quickly perform an ECG to a high number of patients, including the monitoring of outpatients. These characteristics were determinant to ensure safe management in the first wave of the COVID-19 pandemic, when an enormous number of patients overtook available resources.

The intraclass correlation coefficient points a good agreement in the measurements of QTc interval using these portable recorders, supporting the solidness of this handheld device. Once confirmed the value and the reliability of these devices, ECG portable recorders are called to be part of the workaday armamentarium in our hospitals.

Footnotes

Conflict of interest: None declared

References

  • 1.Ray WA, Murray KT, Hall K, et al. Azithromycin and the risk of cardiovascular death. N Engl J Med. 2012;366(20):1881–1890. doi: 10.1056/NEJMoa1003833. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Chen CY, Wang FL, Lin CC. Chronic hydroxychloroquine use associated with QT prolongation and refractory ventricular arrhythmia. Clin Toxicol (Phila) 2006;44(2):173–175. doi: 10.1080/15563650500514558. [DOI] [PubMed] [Google Scholar]
  • 3.Anson B, Weaver J, Ackerman M, et al. Blockade of HERG channels by HIV protease inhibitors. Lancet. 2005;365(9460)(05):682–686. 70936–3. doi: 10.1016/s0140-6736. [DOI] [PubMed] [Google Scholar]
  • 4.Straus SM, Kors JA, De Bruin ML, et al. Prolonged QTc interval and risk of sudden cardiac death in a population of older adults. J Am Coll Cardiol. 2006;47(2):362–367. doi: 10.1016/j.jacc.2005.08.067. [DOI] [PubMed] [Google Scholar]
  • 5.Chang D, Saleh M, Gabriels J, et al. Inpatient use of ambulatory telemetry monitors for COVID-19 patients treated with hydroxychloroquine and/or azithromycin. J Am Coll Cardiol. 2020;75(23):2992–2993. doi: 10.1016/j.jacc.2020.04.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Guerrero LR, Ongay AG, Álvarez ÁC, et al. Protocolo de manejo hospitalario de alteraciones electrocardiográficas en pacientes con COVID-19 con un sistema portátil vinculado a smartphone. REC: CardioClinics. 2020;55(3):192–195. doi: 10.1016/j.rccl.2020.05.006.. [DOI] [Google Scholar]
  • 7.Strik M, Caillol T, Ramirez FD, et al. Validating QT-interval measurement using the apple watch ECG to enable remote monitoring during the COVID-19 pandemic. Circulation. 2020;142(4):416–418. doi: 10.1161/CIRCULATIONAHA.120.048253. [DOI] [PMC free article] [PubMed] [Google Scholar]