Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2021 Jun 2;11:11542. doi: 10.1038/s41598-021-90985-y

Reproducibility of CT-based radiomic features against image resampling and perturbations for tumour and healthy kidney in renal cancer patients

Margherita Mottola 1,2,#, Stephan Ursprung 3,4, Leonardo Rundo 3,4, Lorena Escudero Sanchez 3,4, Tobias Klatte 5,6, Iosif Mendichovszky 3, Grant D Stewart 4,5, Evis Sala 3,4, Alessandro Bevilacqua 2,7,✉,#
PMCID: PMC8172898  PMID: 34078993

Abstract

Computed Tomography (CT) is widely used in oncology for morphological evaluation and diagnosis, commonly through visual assessments, often exploiting semi-automatic tools as well. Well-established automatic methods for quantitative imaging offer the opportunity to enrich the radiologist interpretation with a large number of radiomic features, which need to be highly reproducible to be used reliably in clinical practice. This study investigates feature reproducibility against noise, varying resolutions and segmentations (achieved by perturbing the regions of interest), in a CT dataset with heterogeneous voxel size of 98 renal cell carcinomas (RCCs) and 93 contralateral normal kidneys (CK). In particular, first order (FO) and second order texture features based on both 2D and 3D grey level co-occurrence matrices (GLCMs) were considered. Moreover, this study carries out a comparative analysis of three of the most commonly used interpolation methods, which need to be selected before any resampling procedure. Results showed that the Lanczos interpolation is the most effective at preserving original information in resampling, where the median slice resolution coupled with the native slice spacing allows the best reproducibility, with 94.6% and 87.7% of features, in RCC and CK, respectively. GLCMs show their maximum reproducibility when used at short distances.

Subject terms: Biomarkers, Kidney, Biomedical engineering, Statistics

Introduction

Computed Tomography (CT) is one of the most widely used technologies for morphological imaging and the standard of care adopted in oncology for diagnosis, staging and treatment follow-up. It is still based on visual lesion detection and morphological measurements (e.g. maximum diameter, size, etc.), often performed with the aid of software tools1. Nonetheless, since the early nineties2, radiologists have benefited from computer-aided systems exploiting a large number of features, developed by specialized research groups. The recent increase in high-performance computing resources in entry-level workstations and the growth of automatic tools for radiomic analyses, has made them popular and accessible to research groups. Accordingly, the number of radiomic studies has exploded, a large number of features is analysed to measure macroscopic tumour or tissue characteristics or to find latent properties3. The automatic quantification of tissue features based on radiomic approaches has shown improvements in terms of both reproducibility, discrimination and classification capability4, thus increasing the number of candidate imaging biomarkers3. However, the plethora of software packages available for radiomic analyses used by groups, with different degrees of expertise, has highlighted the urgent need for standardisation of methodology and measurements5,6. In fact, many factors are known to induce variability in radiomic features including noise7, heterogeneous voxel size8 and other CT parameter settings9, Region Of Interest (ROI) segmentation1013, as well as tumour phenotype14. Despite its importance, only a few studies perform a dedicated analysis of the robustness and reproducibility of radiomic studies. Some of them use phantoms to explore the effects of variable acquisition parameters, such as tube current7, or voxel size15. Other radiomic studies assess the effects of varying segmentations on first or second order texture features, for instance, in non-small cell lung cancer (NSCLC)1,16, and rectal cancer17. A different approach has been recently proposed18, where the authors test different perturbation chains on NSCLC and head and neck cancer datasets, to find the chain better reproducing the outcome of a test-retest procedure, to be used when such method is not applicable. All these studies, carried out on different tumours, analyse the reproducibility mainly against varying ROI segmentation, or a set of perturbations.

To the best of our knowledge, this is the first work assessing robustness of first order (FO) and 2D and 3D second order texture features in CT imaging of renal cell carcinoma (RCC) and normal kidney (CK), by addressing three types of perturbations induced by Added White Gussian Noise (AWGN) (N), different voxel-size (V) and varying ROI (R). In addition, we perform a comparative analysis to select the best interpolation methods to be preliminarily applied, if needed, before any feature extraction procedure. Finally, results can provide practical operating guidelines to choose the proper voxel size in case of datasets with heterogeneous in-plane resolutions and to aggregate information derived from grey level (GL) co-occurrence matrices (GLCMs)19, thus improving standardisation of radiomic studies.

Methods

Patient images

This study included 98 patients with RCC imaged at a single institution. CT acquisition parameters are provided in Table 1. Images were acquired with Siemens SOMATOM Definition AS/AS+ CT scanners, with iterative reconstruction kernel I30f\3. Scan resolution ranges from square voxel spacing vs=0.541 mm to vs=0.957 mm, with mean value equals to 0.740, and vz=5 mm-slice thickness. Mean values and ranges of tube voltage and exposure were 109 [100,140] KVp and 166 [137,535] mAs, respectively. Automatic tube voltage selection (CARE kV) and current modulation (CARE Dose) were employed to optimize the dose to patients, resulting in a mean and range of 109 [100,140] KVp and 166 [137,535] mAs, respectively. Image series of the corticomedullary and nephrographic phase, acquired at 35 and 100 s after the administration of the intravenous contrast agent (Omnipaque 300 mg I/ml, GE Healthcare) were included for 28 and 70 patients, respectively. This retrospective study was approved by Health Research Autority (HRA), University of Cambridge and Cambridge Research and Development (R&D) department that waived the written informed consent. This study was conducted according to relevant guidelines and regulations.

Table 1.

CT image acquisition parameters.

Parameter CT
Number of scans 98
Scanners Siemens SOMATOM AS/AS+
Tube voltage (kVp) 109 [100,140]
Exposure (mAs) 166 [137,535]
Reconstruction kernel I30f\3
Square voxel spacing ([x,y]-axes; mm) 0.740 [0.541,0.957]
Voxel spacing (z-axis; mm) 5
Image noise (σ; HU) 4 [2.9, 5.9]
SNR (dB) 42 [37, 45]

Segmentation

ROIs of RCC and CK were semi-automatically outlined using the MICROSOFT RADIOMICS TOOL (Version 1.0.30558.1, project InnerEye, https://www.microsoft.com/en-us/research/project/medical-image-analysis/) by a medical doctor and clinical researcher with three years experience in renal imaging. The structures were segmented in all slices at the original scan resolution (RCC volume was 196 cm3, on average). Polygonal ROIs were exported as DICOM RTSTRUCT and imported in MATLAB (Version R2019b, The MathWorks Inc, Natick, Massachusetts, https://it.mathworks.com/products/matlab.html/) to generate binary segmentation masks for RCC and CK. In particular, CK and RCC were segmented on the first and last slice where they were visible and contours for every other slice were interpolated20. Manual corrections were applied to sub-optimally segmented slices, leading to an iterative re-calculation of the remaining interpolated slices. The segmentation of CK included the renal cortex and medulla but not the the collecting system and hilar fat.

Image processing

Image processing for feature robustness analysis was performed according to the workflow reported in Fig. 1, where the main steps are outlined. For each block of the flowchart more details are provided in the reference section of the main manuscript.

Figure 1.

Figure 1

Workflow of CT image processing for feature robustness analysis. First, CT images were acquired (a) and ROIs segmented (b). Then, one out of three resampling methods was chosen based on the highest EME-IQ score (c), then CT images were perturbed by changing resolution (d), noise addition (e), and ROI variation (f). Finally, first order, and 2D and 3D texture features based on GLCMs (g) were extracted from original and perturbed images and robustness analysis was performed (h).

After CT image acquisition and segmentation, an image interpolation procedure was needed to standardize the different resolutions of the dataset. After discarding the highest and lowest 2.5% of grey values according to a simple and standard outlier removal procedure, three well-known interpolation methods were analysed and compared, to find the method best preserving the statistical properties of the original images, to be used for the image perturbation by voxel-size resampling. The comparison was carried out by exploiting the Enhancement Measurement Error-Image Quality (EME-IQ)21 score that measures local image contrast, amongst the most important information cues, so that higher EME-IQ values indicated the interpolation method better preserved sharpness and edges. After that, three different perturbations were applied to whole images and ROIs. The first perturbation considered was the AWGN, added to the original CT images using the same standard deviation (σ) as the original images. As the second perturbation, the original CT slices were resampled and interpolated along the z direction. Finally, to simulate inter-reader variability, ROI enlargement and shrinking were considered18. Lastly, 3D FO and 2D-3D texture features (from GLCMs) were computed on the original and perturbed CT images, and their reproducibility was studied. All the procedures were implemented in MATLAB.

Image interpolation

Different scan voxel size is a well-known source of variability for radiomic features in a heterogeneous image dataset that should be taken into account when performing radiomic studies8. Hence, applying a resampling procedure is needed, to achieve one voxel size15. However, any resampling method relies on interpolation techniques that may potentially alter latent image properties, thus affecting the reproducibility of the features extracted from the resampled image and from the original22. In radiomic analyses, the interpolation method must be conceived for quantitative imaging and preserve the original properties of the CT GL distributions, rather than to simply yield visually pleasant images. In this work, we analysed and compared three well-known interpolation methods, that is, linear interpolation, largely employed in most radiomic studies15 thanks to its low complexity and computational cost, Akima cubic Hermite spline23 and Lanczos24 interpolation, the latter mostly used in computer vision and quantitative imaging. The comparison was performed by resampling the original CT images at the best and worst resolutions, corresponding to the smallest (vss=0.541) and largest (vsl=0.957) original voxel size, and rounding the grey levels to the nearest integer. When employed for visualization purposes, the different interpolation methods are assessed through a forward-backward process, which compares the quality of original and restored image25. Here, we measured directly the quality of the interpolated images and adopted the EME-IQ score, a Non-Reference IQ measure that quantifies the level of local contrast21. For each patient, EME-IQ scores were computed and averaged on all CT slices, in both upsampling and downsampling. Then, the three interpolation methods were ranked according to their EME-IQ score. We chose the method that resulted in the best score for the highest number of patients for both upsampling and downsampling, and adopted it in the subsequent steps of V image perturbation.

Image perturbation

Additive noise (AWGN)

CT images are known to be mostly affected by quantum noise, arising from the effects of the variability of electronic density of tissue voxels26, statistically represented by a random Gaussian process27. Therefore, we perturbed CT images by AWGN where, for each patient, σ is given by the average of the standard deviation of each slice, estimated according to the method proposed by Ikeda et al.28.

Changing voxel size

Original CT images consisted of anisotropic voxels, with different in-plane resolutions, but one slice spacing that was on average one order of magnitude bigger. As regards slice resolution, we investigated three different resampling strategies, that are: (1) upsampling the whole dataset to vss, (2) downsampling to vsl, (3) resampling at the median resolution (vsM=0.741 mm). Although working with isotropic voxels would be advisable, resampling to the z-axis resolution for isotropy would introduce an unrecoverable signal loss. Therefore, besides keeping the original scan resolution (vz=5 mm), we limited the highest resolution to vz=1 mm, exploring intermediate values, with 1-mm steps. In total, combining three voxel sizes with five slice thicknesses, we tested 15 different voxel sizes.

Segmentation perturbation

One of the causes affecting the clinical reliability of radiomic features as predictive or prognostic biomarkers is the lack of reproducibility of quantitative measurements, depending on the variability of intra- and inter-observer ROI segmentation1. Similarly to what done in29 and18, we simulated such variability, considering volume variations up to 20%, by ROI enlargement and shrinking. Actually, while ROI erosion just implies missing some tissue of the same type, ROI dilatation means including different tissues. Consequently, ROIs were shrunk by 10%, 15% and 20% or dilated by 10%. This procedure was carried out through binary morphological dilation and erosion, with a 3 × 3 pixel square structuring element (SE), according to a pixel-based random contourization procedure. Of course, it is unlikely to achieve the exact percentage variation, therefore the exceeding pixels were randomly removed to attain the expected percentage.

Feature extraction

Radiomic features commonly employed to depict tumour heterogeneity can be grouped into first, second, and higher order statistical descriptors. In particular, FO features measure the statistical properties of the GL distribution and ignore its spatial relationships within the ROIs, whilst texture features (i.e. second and higher order ones) investigate the relationships between neighbouring grey levels at pixel- or region-level30. FO and GLCM-based texture features are very attractive, also because of their low computational complexity, they are computed in all radiomic packages31 and also employed for building predictive models in renal diseases4. In this study, we included 13 FO and 19 GLCM features computed in both 2D (GLCM2D) and 3D (GLCM3D), since no agreement exists yet on how to aggregate GLCM information to extract single representative features6. Hence, GLCM2D were computed in four directions, θ=0, 45, 90, 135, and GLCM3D were extended in 13 directions32, with five odd distances, from δ=1 to δ=9. The features were extracted after intensity-based outlier removal was performed on CT images at the 2.5% threshold at both left and right tails of GL distributions. Based on a preliminary analysis of our CT dataset, the commonly used choice17,33 of 32 quantization levels was adopted for the GLCM computation. GLCMs were also symmetrized and direction-weighted. In GLCM2D, features were first computed on each slice and then averaged. In all, 108 radiomic features were computed on the original and perturbed CT images, for RCC and CK separately.

Table 2 lists all FO and GLCM features, while a detailed mathematical formulation of the radiomic features extracted is also provided in Supplementary Note 1.

Table 2.

List of first order (n=13) and GLCM2D-3D (n=19) features.

First order GLCM2D-3D
Mean (m) Autocorrelation (autoc)
Median (M) Correlation (corr)
Skewness (s) Cluster prominence (cprom)
Maximum value (max) Homogeneity (homom)
m of last decile (m90th) Maximum probability (maxpr)
M of last decile (M90th) Contrast (contr)
Standard deviation (std) Cluster shade (cshade)
M absolute deviation (MAD) Variance (sosvh)
Interquartile range (iqr) Dissimilarity (dissi)
Local coefficient of variation (lCV) Energy (energ)
Uniformity (u) Entropy (entro)
Entropy (e) Difference variance (dvarh)
Kurtosis (k) Difference entropy (denth)
Information measure of corr (inf1h)
Inverse difference normalized (indnc)
Inverse difference moment normalized (idmnc)
Sum average (savgh)
Sum variance (svarh)
Sum entro (senth)

Robustness analysis

All the extracted features were analysed for both RCC and CK and robustness was assessed using the Intraclass Correlation Coefficient (ICC)(1,1) with 95% confidence interval (CI)18. Radiomic features were considered as being robust (r) if ICC 95% CI 0.90, non-robust (nr) if CI < 0.90, and with indeterminate robustness (i) otherwise (i.e. with 0.90 strictly included in CI). In total, 29 perturbations were assessed, one arising from N, 24 combinations of V, and four from R, as detailed in Table 3.

Table 3.

Descriptions of the 29 perturbations assessed.

Number Type Perturbation Description
1 N N AWGN
15 V vs-vz 15 voxel-sizes by combining vss, vsM, vsl with vz in [1, 5]
1 V Global assessment of the all 15 voxel-sizes
1 vss Resolution vss kept fixed and vz in [1, 5]
1 vsM Resolution vsM kept fixed and vz in [1, 5]
1 vsl Resolution vsl kept fixed and vz in [1, 5]
1 Z1 Resolution vz=1 kept fixed for all vss, vsM, vsl
1 Z2 Resolution vz=2 kept fixed for all vss, vsM, vsl
1 Z3 Resolution vz=3 kept fixed for all vss, vsM, vsl
1 Z4 Resolution vz=4 kept fixed for all vss, vsM, vsl
1 Z5 Resolution vz=5 kept fixed for all vss, vsM, vsl
4 R R+10 Dilation, volume variation equals to +10%
R-10 Erosion, volume variation equals to -10%
R-15 Erosion, volume variation equals to -15%
R-20 Erosion, volume variation equals to -20%

First, the robustness of all radiomic features together was investigated against all perturbations, to have an overview of features behaviour depending on the tissue phenotype (i.e. RCC or CK) only. Mean percentage of r, nr, and i features were reported for each perturbation type. Moreover, the proportional contribution given by each feature class to the global robustness was investigated, together with the contribution of the single features. In practice, robustness was assessed (i) for all feature classes (i.e. FO, GLCM2D, GLCM3D) against all image perturbations, (ii) for each feature class against each perturbation type (i.e. N, V, R), and (iii) for each feature against all image perturbations.

Finally, this study assessed the real need for having GLCMs computed at multiple δ distances, because of their known high correlation. To this end, we performed this analysis by adopting the voxel size resulting as the most reliable from analyses at step (i). In conclusion, the correlation of features computed at all δ was measured through the linear Pearson coefficient (ρ) and the statistical significance of the differences was assessed by the ANOVA test (p-value 0.001).

Results

Interpolation methods

Figure 2 reports the comparison of Linear, Akima and Lanczos interpolation methods based on the EME-IQ score. When resampling at vss (Fig. 2a), the linear method achieved the best result for only 4% of patients, Akima for 18% and Lanczos for 78% of patients. This ranking was confirmed when resampling at vsl (Fig. 2b), where linear and Akima methods reached 1% and 36%, respectively, whilst Lanczos still proved to be the best one for 63% of patients. On the original CT images, mean EME-IQ score was 2.87, ranging between [1.36,6.72], whilst on upsampled images, mean and range EME-IQ values increased, being 2.99 [1.61,6.48] for linear interpolation, 3.09 [1.68,6.71] for Akima, and 3.12 [1.70,6.80] for Lanczos method. When downsampling, EME-IQ scores decreased with respect to the original CT images, with 2.71 [1.53,5.40] for linear, 2.84 [1.63, 5.66] for Akima, and 2.85 [1.65, 5.64] for Lanczos. To allows readers to assess the visual differences of these three methods, some exemplifying images are shown in Supplementary Figure S1 for three patients where linear, Akima, and Lanczos methods were respectively the best methods, in either upsampling (vss) and downsampling (vsl), or both. Therefore, hereafter Lanczos is chosen as the reference interpolation method.

Figure 2.

Figure 2

Comparison of linear, Akima, and Lanczos interpolation methods, based on the percentage of patients receiving the highest EME-IQ score, for both upsampling (a) and downsampling (b) respectively at the smallest (vss) and the largest (vsl) resolutions of the CT image dataset. The Lanczos method yielded the highest EME-IQ score in 78% and 63% of patients, respectively.

Robustness of feature classes against image perturbation

Figure 3a,b reports the percentage of r, nr, and i features against each perturbation for RCC (Fig. 3a) and CK (Fig. 3b), respectively (detailed robustness performance of each feature class against all image perturbations are reported in Supplementary Figure S2). Summarizing the information of Fig. 3a, in RCC there are, on average against all perturbations, 65.6% of r features, 18.0% nr, and 16.4% i. Similarly in Fig. 3b, in CK 39.0% was r, 42.9% nr, the remaining 18.0% i. In both RCC and CK, the highest percentage of r features is achieved with the N perturbation (last columns of Fig. 3a,b), where practically all features were r (r-RCC: 100%, r-CK: 99.6%), or at worst i (i-CK: 0.4%). As regards V perturbations, whilst in RCC the percentage of r and nr features was 73.0% and 14.5%, in CK values are lower, with r-CK: 50.8% and nr-CK: 30.3%. Despite this difference, RCC and CK showed a common behaviour against all perturbations at fixed vz values (i.e. [Z1–Z5]), both having a low percentage of r features (on average, 56.7% for RCC and 40.8% for CK) if compared with those at fixed vs resolutions (i.e. vss, vsM, vsl), that is on average 68.6% for RCC and 48.1% for CK. In addition, if considering vz, most of the 15 combinations had more than 60% of r features in RCC, mainly referred to vsM coupled with multiple vz values, with the highest percentage (94.6%) achieved with vsM-Z5, that is, without interpolating along the z-direction. This couple was also the best in CK, with 87.7% of r features. As regards R perturbations, whilst in RCC there was 94.6%, 76.9%, and 70.4% of r features at R-10, R-15, and R-20, respectively, in CK they were 8.9%, 4.9%, 2.5%, respectively. Finally, as regards R+10, RCC showed 19% of r features, while CK had only 4% of them.

Figure 3.

Figure 3

Overall percentage of robust (r), non-robust (nr), and with indeterminate robustness (i) features against image perturbation for RCC (a) and CK (b), separately. Feature robustness is also reported per each feature class (i.e. FO, GLCM2D, and GLCM3D) for RCC (c) and CK (d).

Figure 3c,d focuses on feature robustness of each feature class per perturbation type (values are also reported in Table 4).

Table 4.

Feature robustness against image perturbations referred to Fig. 3c,d.

RCC CK
FO GLCM2D GLCM3D FO GLCM2D GLCM3D
N V R N V R N V R N V R N V R N V R
r (%) 100 100 94.9 100 59.7 83.2 100 59.8 76.1 100 75.6 38.5 98.9 39.4 3.9 100 37.3 2.5
nr (%) 0 0 0 0 14.9 0.7 0 28.7 10.9 0 8.3 53.8 0 45.4 88.4 0 37.0 91.9
i (%) 0 0 5.1 0 25.4 16.1 0 11.5 13 0 16.1 7.7 1.1 15.2 7.7 0 25.7 5.6

FO features hold the highest percentage of r features, in both RCC and CK, for V (RCC: 100.0%, CK: 75.6%), R (RCC: 94.9%, CK: 38.5%), and N (RCC and CK: 100.0%) perturbation types. GLCM2D and GLCM3D features achieved comparable results in both RCC and CK. In fact, the percentage of r features averaged over all perturbations was GLCM2D: 80.9% and GLCM3D: 78.6% in RCC, and GLCM2D: 47.4% and GLCM3D: 46.4% in CK. In particular, robustness against V and R was always higher in RCC than in CK. In particular, R perturbation showed the greatest difference, with GLCM2D-R: 83.2% and GLCM3D-R: 76.1% in RCC and GLCM2D-R: 3.9% and GLCM3D-R: 2.5% in CK. Finally, analysing how each of the 29 perturbations affected each feature class (Supplementary Figure S2), one can see that while for CK there were no differences between GLCM2D and GLCM3D (the same 3 perturbations showed at least 60% of r features), in RCC 13 and 10 perturbations showed at least 60% of r features, in GLCM2D and GLCM3D, respectively, but only 7 of them were shared.

Robustness analysis of single features

All the FO features were r features in RCC in at least 60% of perturbations, and 9 of them (m, M, max, m90th, M90th, iqr, u, e, and lCV) were confirmed in CK too. As far as second order features are concerned, 48 GLCM2D and 53 GLCM3D resulted r features in RCC in at least 60% of perturbations, and 44 of them were in both classes. In CK, 28 GLCM2D and 24 GLCM3D resulted r features in at least 60% of perturbations and 21 of them were shared. Finally, as regards r features shared between RCC and CK, they were autoc, entro, savgh, sentro, sosvh, svarh. It is worth mentioning that cprom, cshad, energ, inf1h, and maxpr were found in RCC only.

GLCM features have an intrinsic redundancy, since they were computed at multiple δ, and these measures are often highly correlated. This is true also in this study, with ρ0.90 for all the selected features and ANOVA tests yielding p-values>0.03 for almost all δ, that is far above the established significance threshold, meaning they perform the same. For exemplification purposes, we show in Fig. 4a–e histograms for a representative feature, GLCM2D-sosvh, computed in RCC at multiple δ, where no relevant difference in distributions can be detected. This becomes more explicit in Fig. 4f, showing the boxplots of GLCM2D-sosvh for all δ, where ANOVA test confirmed their statistically equivalence (p-value=0.97). Actually, the equivalence between distances weakens as they shorten. For instance, the features entro and sentro representing local tissue heterogeneity showed significant differences between δ=1 and δ=3 in both RCC and CK (p-value10-6), and between δ=3 and δ=5 in CK only. Analogously, the features cprom, maxpr, and inf1h in RCC were statistically equivalent for δ3, and different from δ=1 (p-value10-4).

Figure 4.

Figure 4

Histograms of GLCM2D-sosvh computed from δ=1 to δ=9 (ae) did not show any relevant difference. In addition, boxplots (f) confirmed no statistical significance at ANOVA test (p-value=0.97).

Discussion

Assessing the robustness of radiomic features has become necessary to determine feature reproducibility before translating predictive and prognostic radiomic biomarkers into clinical practice. Feature robustness depends on the tumour phenotype and is not generalizable14, hence, this study focuses on the need for analysing feature robustness on RCC in CT using one of the largest datasets22. In addition, the analysis is extended to CK to determine which features might be robust enough to assess, for instance, diffuse renal diseases. In particular, in this study we analyse the robustness of radiomic features against some of the most frequent sources of variability, which are noise, heterogeneous scan voxel size and varying segmentation. Moreover, this study offers an in-depth analysis of three known interpolation methods aiming at supporting researchers in choosing the most appropriate one when resampling CT images. Results show that Lanczos interpolation outperforms the other methods in both upsampling and downsampling procedures. In particular, this study highlights one of the major limits of linear interpolation, probably the method most widely employed in radiomic studies for feature stability assessment in heterogeneous datasets and suggested by the Imaging Biomarker Standardization Initiative (IBSI)6. More specifically, if on the one hand resampling based on linear interpolation improves visual images’ perception, on the other hand it smooths tissue edges and texture variation, thus limiting quantitative information34. Some authors are aware of the importance of the interpolation methods, which may influence feature robustness. For instance, Whybra et al.30 carried out a comparison of feature robustness, after linear and spline interpolations. The authors concluded that the two methods were equivalent since no difference existed in terms of feature stability, albeit in the presence of large numerical variations. However, although those features may be reproducible, this does not ensure that the features are correctly representing the original CT image information. For this purpose, we recommend a preliminary analysis to assess that the resampling procedure does not affect the properties of the GLs distribution. It is worth noting that the upsampling procedure, although adding artificial information, improves the original image quality (EME-IQ = 2.87) with all methods considered (EME-IQ = 2.99 for linear, EME-IQ = 3.09 for Akima, and EME-IQ = 3.12 for Lanczos). Even more relevant, when downsampling, while linear interpolation degrades (EME-IQ = 2.71) original image quality, Lanczos, performing the best (EME-IQ = 2.85), preserves the EME-IQ score of the original image.

With regard to feature robustness, there are many more r features in RCC (65.6%) than in CK (39.0%), although both RCC and CK show an excellent robustness against N perturbation (N-RCC: 100%, N-CK: 99.6%). This agrees with the outcome of Zwanenburg et al.18, which similarly found that the highest percentage of robust features was for N perturbations. Instead, substantial differences of r features between RCC and CK are found under R and V perturbations. In fact, results show that against R- perturbations, while in RCC r features are never lower than 70%, in CK a very low percentage of features are reproducible if changes in volumes are higher than 10%. As expected, r features have a much worse performance against R+10 perturbation, this suggesting that when segmenting it is always better performing a “safe” contouring, that is, underestimating rather than overestimating the ROI.

When resampling a heterogeneous CT dataset, the goal is to minimize interpolation artefacts. Our results show that choosing the median resolution (vsM) does this, with a greater effect in CK rather than in RCC. In fact, among the different CT voxel sizes, vsM achieves the highest percentage of r features, this suggests that resampling at the median voxel size is strongly recommended. In addition, focussing on the different vs-vz couples, vsM performs best when no interpolation in the z-direction is carried out between slices. This is somewhat expected, since the large difference between vs (higher) and vz (lower) voxel sizes makes the interpolation along the z-direction introduce a low reliable signal, if compared with the information in the original CT slices. Accordingly, while resampling along z-axis should be carefully evaluated, especially in case of a large slice spacing, preservation of the original vz resolution could be in most cases the best choice.

Our robustness analysis finds the FO features are definitely the most reproducible ones, confirming what was reported in the review of the most recent research works regarding feature repeatability and reproducibility by Traverso et al.22. In addition, all r features in CK (9/13) are robust in RCC as well. Besides the well-established statistical descriptors (e.g. m, M, etc.), there are both lCV and e, two common indicators for measuring local heterogeneity or irregularity, that is also one of the changing properties of normal tissues while shifting into tumour ones35. The remaining 4/13 FO features resulted robust in RCC only, thus showing a higher specificity for tumour tissues, which could be useful for specific tumour-related clinical questions. It is worth noting that also all GLCM features (both 2D and 3D) that prove to be robust in CK, are robust in RCC too, while other features, measuring local asymmetries of GLCMs, are more tumour-specific.

This research also investigates the well-known phenomenon of the high correlation of GLCM-based texture features computed at different distances, to see whether and to what extent using higher distances is worth. In practice, almost all features are shown to be equivalent when computed at distances from 3 to 9. Five r features show a difference in RCC when computed at δ=1 and δ=3 and two only in CK at δ=3 and δ5. This evidence is yet more relevant if considering that even in CK, having really wide ROIs, distances higher than δ=3 are most of times equivalent. This could suggest that computing textures at distance δ=3 should be general enough, thus allowing feature selection to be simpler and more effective, besides reducing computational burden. As a general remark, the recent literature lacks comparative studies between 2D and 3D texture feature robustness, and even when features are compared on the basis of their capabilities (e.g., predictive ability, and so on) the outcomes are controversial36. Our results show that the overall robustness of GLCM features computed in 2D or 3D is similar. However, our findings show a higher number of r features for GLCM2D and, at the same time, a higher number of perturbations not affecting robustness of GLCM2D features. Therefore, GLCM2D texture features should be preferred.

The main limitation of this study is that only first and second order features are considered. However, these are the first features whose robustness has been analysed in radiomic studies based on CT images of renal disease. In addition, our methodological approach can be exploited to extend the study to include more feature classes. An additional limitation arises from having considered 10% as the lowest volume variation bound. This threshold might yield overestimated inaccuracies of radiologist’s segmentations. Therefore, the results reported in this study can be considered as the worst scenario. Finally, another limitation is the analysis of all RCC subtypes together. However, an analysis of feature robustness across different RCC subtypes would have been beyond the scope of this study.

This work aimed at assessing the robustness of radiomic features against some of the most common sources of variability. Our findings allowed drawing some concluding remarks that could be useful guidelines for radiomic studies. In particular, texture features should be used at very short distances, heterogeneous CT datasets have to be resampled at the median slice resolution, whilst should not be interpolated along the cranial-caudal direction and, Lanczos should be used as the interpolation method.

Supplementary Information

Acknowledgements

This work was partially supported by The University of Bologna and The Mark Foundation for Cancer Research and Cancer Research UK Cambridge Centre [C9685/A25177], the Wellcome Trust Innovator Award [RG98755] and the CRUK National Cancer Imaging Translational Accelerator (NCITA) [C42780/A27066]. Additional support was also provided by the National Institute of Health Research (NIHR) Cambridge Biomedical Research Centre (BRC-1215-20014). The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care. Microsoft Radiomics was provided to the Addenbrooke's Hospital (Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK) by the Microsoft InnerEye project.

Author contributions

All authors: conceptualization, writing—review and editing. M.M. and A.B. methodology, formal analysis; M.M. software, investigation, writing—original draft; A.B. validation, resources (computing resources and analysis tools); E.S. and G.D.S. resources (patients, ethics); S.U., L.R. and L.E. data curation (S.U. segmentation); A.B. and E.S. supervision and project administration.

Competing interests

G.D. Stewart has received educational grants from Pfizer, AstraZeneca, and Intuitive Surgical; consultancy fees from Pfizer, Merck, EUSA Pharma, and CMR Surgical; travel expenses from Pfizer; and speaker fees from Pfizer. Other authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Margherita Mottola and Alessandro Bevilacqua.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-021-90985-y.

References

  • 1.Parmar C, Velasquez ER, Leijenaar R, et al. Robust radiomics feature quantification using semiautomatic volumetric segmentation. PLoS ONE. 2014;9:e102107. doi: 10.1371/journal.pone.0102107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Doi K. Computer-aided diagnosis in medical imaging: Historical review, current status and future potential. Comput. Med. Imaging Graph. 2007;31:198–211. doi: 10.1016/j.compmedimag.2007.02.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Park J, Kim H. Radiomics as a quantitative imaging biomarker: Practical considerations and the current standpoint in neuro-oncologic studies. Nucl. Med. Mol. imaging. 2018;52:99–108. doi: 10.1007/s13139-017-0512-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ding J, Xing Z, Jiang Z, et al. CT-based radiomic model predicts high grade of clear cell renal cell carcinoma. Eur. J. Radiol. 2018;103:51–56. doi: 10.1016/j.ejrad.2018.04.013. [DOI] [PubMed] [Google Scholar]
  • 5.Larue R, Defraene G, De Ruysscher D, et al. Quantitative radiomics studies for tissue characterization: A review of technology and methodological procedures. Br. J. Radiol. 2017;90:20160665. doi: 10.1259/bjr.20160665. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Zwanenburg A, Valliéres M, Abdalah M, et al. The image biomarker standardization initiative: Standardized quantitative radiomics for high-throughput image-based phenotyping. Radiology. 2020;295:328–338. doi: 10.1148/radiol.2020191145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Mackin D, Ger R, Dodge C, et al. Effect of tube current on computed tomography radiomic features. Sci. Rep. 2018;8:2354–2363. doi: 10.1038/s41598-018-20713-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Shafiq-ul-Hassan M, Latifi K, Zhang G, et al. Voxel size and gray level normalization of CT radiomic features in lung cancer. Sci. Rep. 2018;8:10545. doi: 10.1038/s41598-018-28895-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Meyer M, Ronald J, Vernuccio F, et al. Reproducibility of ct radiomic features within the same patient: Influence of radiation dose and ct reconstruction settings. Radiology. 2019;293:583–591. doi: 10.1148/radiol.2019190928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Fotina I, Lütgendorf-Caucig C, Stock M, et al. Critical discussion of evaluation parameters for inter-observer variability in target definition for radiation therapy. Strahlentherapie Onkol. 2012;188:160–167. doi: 10.1007/s00066-011-0027-6. [DOI] [PubMed] [Google Scholar]
  • 11.Kalpathy-Cramer J, Mamomov A, Zhao B, et al. Radiomics of lung nodules: A multi-institutional study of robustness and agreement of quantitative imaging features. Tomography. 2016 doi: 10.18383/j.tom.2016.00235. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Qiu Q, Duan J, Duan Z, et al. Reproducibility and non-redundancy of radiomic features extracted from arterial phase CT scans in hepatocellular carcinoma patients: Impact of tumor segmentation variability. Quant. Imaging Med. Surg. 2019;9:453–464. doi: 10.21037/qims.2019.03.02. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Haarburger C, Müller-Franzes G, Weninger L, et al. Radiomics feature reproducibility under inter-rater variability in segmentations of CT images. Sci. Rep. 2020 doi: 10.1038/s41598-020-69534-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.van Timmere J, van Elmpt W, et al. Test-retest data for radiomics feature stability analysis: Generalizable or study-specific? Tomography. 2016;2:361–365. doi: 10.18383/j.tom.2016.00208. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Shafiq-ul-Hassan M, Zhang G, Latifi K. Intrinsic dependencies of CT radiomic features on voxel size and number of gray levels. Med. Phys. 2017;44:1050–1062. doi: 10.1002/mp.12123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Balagurunathan Y, Gu Y, et al. Test-retest reproducibility analysis of lung CT image features. J. Digit. Imaging. 2014;27:805–823. doi: 10.1007/s10278-014-9716-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Hu P, Wang J, Zhong H, et al. Reproducibility with repeat CT in radiomics study for rectal cancer. Oncotarget. 2016;7:71440–71446. doi: 10.18632/oncotarget.12199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Zwanenburg A, Leger S, Agolli L, et al. Assessing robustness of radiomic features by image perturbation. Sci. Rep. 2019;9:614–623. doi: 10.1038/s41598-018-36938-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Haralick R, Shanmugam K, Dinstein I. Texture features for image classification. IEEE Trans. Syst. Man Cybern. B. Cybern. 1973;3:610–621. doi: 10.1109/TSMC.1973.4309314. [DOI] [Google Scholar]
  • 20.Oktai O, Nanavati J, Schwaighofer A, et al. Evaluation of deep learning to augment image-guided radiotherapy for head and neck and prostate cancers. JAMA Netw. Open. 2020;3:e2027426. doi: 10.1001/jamanetworkopen.2020.27426. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Panetta K, Samani A, Agaian S. Choosing the optimal spatial domain measure of enhancement for mammogram images. Int. J. Biomed. Imaging. 2014 doi: 10.1155/2014/937849. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Traverso A, Wee L, Dekker A, et al. Repeatability and reproducibility of radiomic features: A systematic review. Int. J. Radiat. 2018;102:1143–1159. doi: 10.1016/j.ijrobp.2018.05.053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Song R, Zhang S, Li C, et al. Heart rate estimation from facial videos using a spatiotemporal representation with convolutional neural networks. IEEE Trans. Instrum. Meas. 2020 doi: 10.1109/TIM.2020.2984168. [DOI] [Google Scholar]
  • 24.Moraes M, Amorim P, Da Silva J, et al. Medical image interpolation based on 3d lanczos filtering. Comput. Methods Biomech. Biomed. Eng. Imaging Vis. 2019 doi: 10.1080/21681163.2019.1683469. [DOI] [Google Scholar]
  • 25.Umehara K, Ota J, Ishida T. Application of super-resolution convolutional neural network for enhancing image resolution in chest CT. J. Digit. Imaging. 2018;31:441–450. doi: 10.1007/s10278-017-0033-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Andria G, Attivissimo F, Lanzolla AML. A statistical approach for MR and CT images comparison. Measurement. 2013;46:57–65. doi: 10.1016/j.measurement.2012.05.016. [DOI] [Google Scholar]
  • 27.Koyuncu H, Ceylan R. Elimination of white gaussian noise in arterial phase CT images to bring adrenal tumours into the forefront. Comput. Med. Imaging Graph. 2018;65:46–57. doi: 10.1016/j.compmedimag.2017.05.004. [DOI] [PubMed] [Google Scholar]
  • 28.Ikeda M, Makino R, Imai K. A method for estimating noise variance of CT image. Comput. Med. Imaging Graph. 2010;34:642–650. doi: 10.1016/j.compmedimag.2010.07.005. [DOI] [PubMed] [Google Scholar]
  • 29.Baiocco S, Sah B, Mallia A, et al. Exploratory radiomic features from integrated 18f-fluorodeoxyglucose positron emission tomography/magnetic resonance imaging are associated with contemporaneous metastases in oesophageal/gastroesophageal cancer. Eur. J. Nucl. Med. Mol. Imaging. 2019;46:1478–1484. doi: 10.1007/s00259-019-04306-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Whybra P, Parkinson C, Foley K, et al. Assessing radiomic feature robustness to interpolation in 18F-FGD PET imaging. Sci. Rep. 2019;9:9649–9659. doi: 10.1038/s41598-019-46030-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Foy J, Robinson K, Li H, et al. Variation in algorithm implementation across radiomics software. J. Med. Imaging. 2018;5:044505. doi: 10.1117/1.JMI.5.4.044505. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Kovalev V, Kruggel F, Gertz H. Three-dimensional texture analysis of MRI brain datasets. IEEE Trans. Med. imaging. 2001;20:424–433. doi: 10.1109/42.925295. [DOI] [PubMed] [Google Scholar]
  • 33.Jin H, Kim J. Evaluation of feature robustness against technical parameters in CT radiomics: Verification of phantom study with patient dataset. J. Signal Process. Syst. 2020;92:277–287. doi: 10.1007/s11265-019-01496-z. [DOI] [Google Scholar]
  • 34.Tan, J. et al. Sharpness preserved sinogram synthesis using convolutional neural network for sparse-view CT imaging. Proc. SPIE 10949, Medical Imaging 2019: Image Processing, 109490E (15 March 2019)10.1117/12.2512894 (2019).
  • 35.Aerts H, Velazquez E, Leijenaar R, et al. Decoding tumour phenotype by noninvasive imaging using a quantitative radiomics approach. Nat. Commun. 2014;5:1–9. doi: 10.1038/ncomms5006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Liu Y, Liu S, Qu F, et al. Tumor heterogeneity assessed by texture analysis on contrast-enhanced CT in lung adenocarcinoma: Association with pathologic grade. Oncotarget. 2017;8:53664–53674. doi: 10.18632/oncotarget.15399. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES