Abstract
Cardiovascular disease is a major cause of death worldwide. New diagnostic tools are needed to provide early detection and intervention to reduce mortality and increase both the duration and quality of life for patients with heart disease. Seismocardiography (SCG) is a technique for noninvasive evaluation of cardiac activity. However, the complexity of SCG signals introduced challenges in SCG studies. Renewed interest in investigating the utility of SCG accelerated in recent years and benefited from new advances in low-cost lightweight sensors, and signal processing and machine learning methods. Recent studies demonstrated the potential clinical utility of SCG signals for the detection and monitoring of certain cardiovascular conditions. While some studies focused on investigating the genesis of SCG signals and their clinical applications, others focused on developing proper signal processing algorithms for noise reduction, and SCG signal feature extraction and classification. This paper reviews the recent advances in the field of SCG.
Keywords: seismocardiography, heart-induced vibrations, cardiovascular disease, signal processing, signal segmentation, noise removal, feature extraction, machine learning
1. Introduction
Cardiovascular disease results in one death every 40 s in the United States [1]. Improved diagnostic, surveillance, and intervention methods would help reduce mortality [2,3] and extend lives. Heart disease may be detected using many non-invasive methods including manual auscultation of heart sounds, which is a common component of physical examinations and known to provide useful diagnostic information. However, simple auscultation of heart sounds is of limited utility. Detection and processing of very-low-frequency heart sounds (“infrasounds”) below the limit of human ear detection may extend the diagnostic power of auscultation. New studies of cardiac-generated sounds using computational fluid dynamics [4,5] and advanced signal processing methods suggested increased potential to provide quantitative information that may be helpful for patient monitoring and diagnosis. Seismocardiography (SCG) is a noninvasive technique that measures cardiac-induced mechanical vibrations at the chest surface including those below the human hearing threshold. The reader is referred to previous SCG reviews [6,7] describing earlier studies, while this paper reviews more recent SCG studies including advances in instrumentation and signal processing that hold the promises of increased clinical utility. Since previous reviews of SCG [6,7] described early SCG studies, this article is more focused on the developments in the field during the last few years. Between September 2017 and March 2018, we conducted a search of the scientific journals and conferences using MEDLINE, as well as the Google Scholar search engine, for the following expressions: “SCG”, “seismocardiography”, and “seismocardiogram”. Reviewing the reference section of the initial results led to additional articles. More resources were added during the manuscript preparation and revision. The final sampling period includes studies that were published after [7] and before November 2018.
1.1. Definition of SCG Signals
The measurements of heart-induced motion, including displacement, velocity, and acceleration, were performed as early as the turn of the 20th century [8]. These approaches can be categorized into two classes [6,7]: (a) the measurement of whole-body recoil forces in response to cardiac ejection, usually termed ballistocardiography (BCG); and (b) the local chest surface measurement of cardiac-induced vibrations, typically referred to as SCG [9–13]. These vibrations are usually measured in the form of acceleration (m/s2). This article focuses on the latter.
1.2. Physiological Sources of SCG Signals
SCG signals are believed to be caused by cardiac mechanical processes including cardiac muscle contraction, cardiac valve movement, blood flow turbulence, and momentum changes. The characteristics of these signals are likely to contain useful information that correlate with cardiovascular physiologic [14] and pathologic processes [15]. Such information may powerfully complement methods of detecting heart electrical activity (such as electrocardiography), serologic testing, and imaging modalities (e.g., echocardiography, cardiac magnetic resonance imaging (MRI), and catheterization).
Early BCG studies [8] suggested that heart-induced motion may be used to estimate changes in cardiac output, and reported certain signal patterns in patients with myocardial infarction [16]. These signal patterns were also found to correlate with the strength of myocardial contractions [17,18] and contain detectable waveform changes during heart disease resolution [19]. After introducing SCG in early 1990s [13], recent studies suggest possible SCG utility for monitoring left-ventricle function, coronary blood flow during balloon angioplasty [9,10,20], heart and breathing rates [21–26], and ventricular filling, cardiac valve closure, and ventricular ejection preceding the onset of ischemic symptoms [10].
While the relationship between SCG waves and cardiac activity is not fully understood, several studies investigated this relationship. For example, SCG was reported to contain a low-frequency wave during atrial systole, a high-amplitude wave during ventricular systole, another wave during early ventricular filling, and some relatively high-frequency waves at the time of the first and second heart sounds [13]. Simultaneous recording of SCG and electrocardiogram (ECG) indicated that the peaks and valleys of the SCG correspond to known physiological events including mitral valve opening (MO) and closure (MC), isovolumetric contraction, ejection, aortic valve opening (AO) and closure (AC), and cardiac filling [7,27]. The utility of SCG in estimating cardiac intervals such as electromechanical systolic pre-ejection period (PEP) and left-ventricular ejection time (LVET) was also shown [28]. Multi-channel partially simultaneous SCG, ECG, and sonographic measurements were used to identify the feature points in a cardiac cycle corresponding to the four common valvular auscultation locations. Using these measurements, new feature points (including left-ventricular lateral wall contraction peak velocity, septal wall contraction peak velocity, trans-aortic valve peak flow, transpulmonary peak flow, trans-mitral ventricular relaxation peak flow, and trans-mitral atrial contraction peak flow) were reported.
Table 1 lists all the SCG feature points and cardiac time intervals (CTIs) that were reported in the literature, while Figure 1 shows a modified Wiggers diagram [29] where a sample SCG signal (in the dorso-ventral direction) is plotted along with aortic blood pressure, ventricular volume, and the electrocardiogram.
Table 1.
Feature Point | Reference |
---|---|
Peak of atrial systole (AS) | [10,14,30] |
Mitral valve closure (MC) | [10,14,28,30,31] |
Peak of rapid systolic ejection (RE) | [10,14,30,32] |
Peak of rapid diastolic filling (RF) | [10,14,30] |
Isovolumic contraction (IC) | [10] |
Mitral valve opening (MO) | [14,28,30,31] |
Aortic valve closure (AC) | [14,28,30–32] |
Aortic valve opening (AO) | [14,28,30–33] |
Isovolumic movement (IM) | [14] |
Rapid diastolic filling time | [14] |
Isotonic contraction (IC) | [14] |
Isovolumic relaxation time (IVRT) | [14,28,31] |
Left ventricular ejection time (LVET) | [14,28,31,32] |
Maximum acceleration in aorta (MA) | [28,32] |
Pre-ejection period (PEP) | [28,31,32] |
Total electromechanical systole period (QS2) | [28,31,32] |
Maximum blood injection (MI) | [28] |
Isovolumic contraction time (IVCT) | [28,31] |
Left ventricular lateral wall contraction peak velocity (LCV) | [34] |
Septal wall contraction peak velocity (SCV) | [34] |
Trans-aortic peak flow (AF) | [34] |
Trans-pulmonary peak flow (PF) | [34] |
Trans-mitral ventricular relaxation flow (MFE) | [34] |
Atrial contraction flow (MFA) | [34] |
During the cardiac cycle of healthy individuals, the apex and base rotate in opposite directions, which results in a twisting motion of the left ventricle [35] known to be affected by different factors such as aging and diastolic dysfunction. Investigating the rotational vibration induced by this heart twisting motion might provide complementary information to the current SCG analysis of uni- and triaxial accelerations. In recent studies [31,36,37], a three-axis micro electromechanical systems (MEMS) gyroscope and a three-axis accelerometer were used simultaneously to measure the rotational and axial components of chest vibrations. The potential utility of the combined analysis of axial and rotational heart-induced vibrations were suggested for the ECG-independent identification of systolic points (such as AO and AC) and cardiac time intervals (such as LVET and PEP) [38].
In summary, despite many studies conducted about SCG genesis, the relationship between SCG waves and cardiac activity is not yet fully understood. This is possibly because of the waveform variations in different studies and lack of understanding of the exact SCG waves sources. Thus, there is still a need for widely accepted universal labeling (i.e., valid for all/majority of patients) analogous to PQRST labeling in ECG.
1.3. Measurement Methods
New advances and availability of lightweight low-noise accelerometers improved the quality of recorded SCG signals. Different methods were used for SCG measurement in the recent studies, including the following:
Airborne ultrasound surface motion camera (AUSMC) [53].
Depending on the sensors that are used, SCG signals might consist of one or more axial and rotational components. For example, a uniaxial accelerometer can be used to measure SCG component in the dorso-ventral direction. However, combination of a triaxial accelerometer and triaxial gyroscope can provide information about axial and rotational heart-induced motion in three different directions. This review focuses on the dorso-ventral component of the SCG signal, unless otherwise stated.
Sensors are most commonly placed on (or directed to) the sternum or its left lower border. However, in some studies, other locations were used for SCG signal acquisition, including over the heart apex (lateral left lower chest) and the “aortic valve listening area” at the right upper sternal boarder [34,43,44]. Information about sensor type, model, and placement location in recent studies is summarized in Table 2 and Figure 2.
Table 2.
Reference | Sensor Type | Sensor Model | Sensor Location |
---|---|---|---|
[54–56] | 3-Acc | SCA610-C21H1A, Murata Electronic | 1 cm above xiphoid |
[43,44] | 3-MEMS-Acc | MMA 7361, Freescale Semiconductor | Heart apex |
[57] | 3-MEMS-Acc | MMA 7361, Freescale Semiconductor | Above xiphoid |
[58] | 3-MEMS-Acc | Analog Devices | 2 cm above xiphoid |
[36,38,45] | 3-MEMS-Acc 3-MEMS-Gyr |
KXRB5-2042, Kionix MPU9150, Invensense |
Left sternal border along the 3rd rib |
[59] | 3-Acc | ViSi Mobile, Sotera Wireless | Chest wall |
[60,61] | 1-Acc 1-Acc |
4381, Brüel & Kjær 393C, PCB Piezotronics |
Above xiphoid |
[62,63] | 1-Acc | DS1104, DSPACE | Xiphoid process |
[64] | 3-Acc | ADXL 335, Analog Devices | Chest wall |
[46] | 3-SP-Acc | iPhone6, Apple | Midclavicular line and 4th intercostal space Belly above navel |
[65,66] | 3-Acc | 356A32, PCB Piezotronics | Left sternal border along the 4th intercostal space |
[67] | 3-Acc | X6-2mini, GCDC | Left sternal border along the 4th intercostal space |
[68] | 1-MEMS-Acc | SCA620, Murata Electronic |
Sternum—anterior chest |
[25,69,70] | 3-MEMS-Acc | MMA8451Q, Freescale Semiconductor | Sternum |
[34,71] | 1-Acc | LIS331DLH, STMicroelectronics | Mitral valve, tricuspid valve, aortic valve, pulmonary valve |
[72] | 3-MEMS-Acc | MMA 7361, Freescale Semiconductor | Left sternal border along the 3rd rib |
[73] | 3-MEMS-Acc | MMA8451Q, Freescale Semiconductor | Sternum, aortic valve, heart apex |
[74] | 3-Acc 1-Acc |
CXL01LF3, Crossbow Technology 7290-A, Endevco Microtron |
Manubrium Xiphoid |
[75–78] | 3-Acc | BMA280, Bosch Sensortec GmbH | Mid-sternum |
[79] | 3-MEMS-Acc | TSD109C, Biopac Systems | Left sternal border along the 3rd rib |
[41] | 3-Acc | 356A32, PCB Piezotronics | Sternum, upper and lower sternum |
[80] | 1-Acc | N/A | Sternum |
[81] | 3-MEMS-Acc 3-Gyr |
MMA8451Q, Freescale Semiconductor L3GD20, STMicroelectronics |
N/A |
[82] | 3-Acc | ADXL 335, Analog Devices | Mid-sternum, upper sternum, lower sternum Point of max impulse, below left clavicle, below right clavicle |
[83] | 3-MEMS-Acc 3-MEMS-Gyr |
SparkFun, Intel Edison | Sensor clipped on subjects clothes |
[50,51] | Microwave Doppler radar | ||
[47,48] | 3-SP-Acc | Xperia Z-series, Sony | Chest |
[49] | Laser Doppler vibrometer | PDV-100, Polytec | |
[37] | 3-MEMS-Acc 2-MEMS-Gyr |
LIS344ALH, STMicroelectronics LPY403AL, STMicroelectronics |
Heart apex Lower back of subject between 2nd and 3rd lumbar vertebrae |
[31] | 3-MEMS-Acc 3-MEMS-ARS |
MMA8451Q, Freescale Semiconductor MAX21000, Maxim Integrated |
Sternum |
[53] | AUSMC | Composed of the following sensors: - MA40S4S, Murata Electronics - FG-23629 Knowles microphone |
∼30 × 40 cm2 thoracic and abdominal surface |
In some applications, such as burn patients, highly infectious patients, and premature babies, attaching adhesive ECG electrodes or SCG sensors would not be feasible. Therefore, development of efficient contactless SCG detection techniques are under investigation. These techniques include laser Doppler vibrometry (LDV), microwave Doppler radar, and airborne ultrasound imaging [22,53,84]. A non-contact SCG measurement might also reduce skin coupling; artefacts that may tie present in the SCG signals acquired by the contact sensors attached to the skin.
The LDV approach compares the frequency shift between the outgoing and reflected laser beams and determines the corresponding vibration velocity of the surface that reflected the beam [85]. Considerations when using LDV for SCG measurements include the following: (1) the chest surface needs to be reasonably reflective for accurate LDV measurements, (2) the laser beam should be perpendicular to chest surface, and (3) chest movement due to respiration needs to be accounted for, since breathing causes the point of measurement to be displaced in the chest: plane. One solution to this issue is to develop an algorithm that can automatically have the beam follow a measurement point on the chest surface. Other LDV limitations include their cost and size.
Microwave Doppler radar is another non-contact method that can be used for SCG measurements. When recording the SCG signal using microwave Doppler radars, the SCG will exhibit in the phase variation of the microwave signal· SCG signals can then be extracted from this phase variation. Like LDV sensors, Doppler radar approaches have the benefit of contactless signal acquisition, but suffer from the reflection of background microwave signals (called radar clutter) lowering the signal-to-noise ratio [51].
The utility of ultrasound imaging in estimating non-contact two-dimensional (2D) SCG maps of the body surface was also investigated [53]. In addition to the advantages of other contactless measurement methods, this technique can collect SCG data from multiple locations through different channels resulting in a potentially higher reliability. However, this method requires a planar measurement surface that is parallel to the emission panel.
Piezoelectric and MEMS sensors are smaller and lighter than contactless sensors. Therefore, these sensors might be used in clinical settings for everyday and continuous screening subjects suspected of different cardiovascular diseases.
1.4. Parameters Affecting SCG Waveform
A main challenge in SCG studies is that SCG signal morphology appears to very significantly, not only by cardiovascular pathology, Cut also normal inter-subject variation. These changes are affected by several factors including respiratory cycle phases, gender, age, sensor chest location, health conditions, cardiac contractility, heart rhythm, and postural positions [23,25,86,87]. While these changes can lead to undesirable SCG variability, deeper understandings of these processes will enhance our understanding oi SCG signal s, help aggregate SCG cycles into groupings with similar SCG events to reduce SCG signal variability and noise, and hopefully lead to more accurate definition of SCG features for diagnoses and monitoring.
A few studies addressed SCG variability, e.g., the consistent effects of respiration [11,32]. One recent study [66] reported that the SCG morphology appeared to mainly depend on the lung volume (and, hence, possibly the intrathoracic pressure), rather than dependence on negative or positive airflow (i.e., inspiration or expiration). This SCG morphology variation can also be used to automatically identify the lung volume states and respiratory phases by employing machine learning [56]. Another study used support vector machines to classify the SCG cycles occurring during the high and low lung volumes [40]. Successful grouping of SCG cycles into two groups, where SCG events in each group are more similar to each other and dissimilar to the events in the alternate group, would improve the signal-to-noise ratio in calculating the SCG ensemble average. This would result in more accurate estimation of diagnostic information from the SCG ensemble average.
Subject motion and postural position were also shown to cause changes in SCG signals [7,75,88]. In an ongoing study, the effect of posture on the SCG signals was investigated for patients with heart failure (HF) [75]. The SCG signals were measured using a wearable unit in supine and seated positions. The SCG power spectral density (PSD) was estimated using Welch’s periodogram, and the means of PSD values were calculated in the 0–20-Hz band. The results showed that SCG signals contained high energy in bands greater than 8 Hz in the supine and seated postures. Identification of chest orientation (supine, 45°, or vertical), and, therefore, grouping of SCG signals according to chest orientation is possible with the use of certain triaxial accelerometers [88]. However, chest orientation measurement is not sufficient to account for all postural changes, as shown by the reported SCG differences between sitting upright and standing, which have the same chest orientation. Movement of the patient also produces a change in the SCG signal. The ability to filter noise originating from speaking, walking, and indistinct motions associated with workplace tasks was demonstrated [43]. Understanding the effects of posture and movement on the SCG waveform is a useful step toward continuous collection of SCG signals from a patient from a wearable sensor. However, this may not be needed if intermittent testing in a more controlled environment is performed at a fixed subject position.
Exercise and the following period of recovery was also demonstrated to produce changes in SCG signals. Not surprisingly, exercise is associated with an increase in the overall amplitude of the SCG signal, measured as the root-mean-square (RMS) power [89,90]. This increase in signal amplitude was shown to correlate with increased cardiac output observed during exercise [89]. This cardiac output increase is a result of increased heart rate and stroke volume. Exercise also produces changes in the left-ventricular ejection time (LVET) and the pre-ejection period (PEP) [88,90]. As exercise increases the heart rate (also seen as the R-R interval decrease in ECG), it generally causes a decrease in other measured time intervals such as LVET and PEP. LVET correlates with both heart rate and contractility and, hence, decreases with exercise. PEP is less affected by heart rate but does decrease during exercise due to the increased contractility (inotropy). These changes in LVET and PEP were detectable by SCG [88,91], and exercise-induced decrease in PEP was found to shift the SCG signal power spectrum toward higher frequencies [78].
Digestive state and mood may affect cardiac function through similar physiological mechanisms, thereby possibly affecting SCG signal morphology. Systematic investigations of these effects are lacking, and future studies are needed to determine the magnitude and nature of these effects on the SCG signal.
The sensitivity of the SCG signal to sensor location is well known and, therefore, needs to be taken into account when comparing results from different studies. Historically, investigators placed accelerometers at different anatomical locations, including the clavicle, the sternum, and various intercostal spaces [34,88,89]. A recent study [34] investigated the differences in SCG signals morphology at the common auscultation sites of the four heart valves (aortic, pulmonary, tricuspid, and mitral), and found significant differences in SCG morphology. That study also concluded, with the aid of sonographic measurements, that more feature points can be defined from multi-point SCG measurements.
Due to the sensitivity of SCG signals to the measurement location, unexperienced users might not be able to repeatably record the SCG at ideal locations. Hence, the SCG-based estimations of cardiac activity might change significantly due to sensor location errors. This can, in turn, result in inaccurate interpretations [41]. To overcome this issue, Ashouri and Inan [41] proposed a method to automatically detect when the sensor is not placed in a desired location by comparing the regression parameters from the acquired SCG and an SCG measured from a reference position.
High-spatial-resolution measurement of the SCG signal was carried out in a pilot study Here, a laser vibrometer was used to perform non-contact uniaxial SCG measurements in the dorso-ventral direction. The laser beam was <2 mm2 and the resulting SCG amplitude distribution is shown in Figure 3. These data suggest that, when the sensor location changed by 1 cm, the SCG amplitude can vary by >30%. An accelerometer with larger contact crea (3.5 cm2) was also used and it was found that, for a sensor location change of 1 cm, the SCG amplitude changed by about 5%. Due to this smaller change, the larger contact area may be beneficial in reducing SCG dependence on sensor placement. The effects of sensor contact area and SCG spatial distributions (including axial or rotational signals) need further investigations.
2. Signal Processing
SCG signal processing usually consists of several steps including preprocessing (e.g., down-sampling and denoising), signal segmentation, feature extraction, and classification (Figure 4). There were several recent studies that focused on noise removal, segmentation, and feature extraction of SCG signals. These studies are reviewed in this section.
2.1. Noise Reduction
While SCG signals can contain useful diagnostic information, they are often contaminated by noise from different sources including sensor mechano-electronics, motion artefacts, and environmental vibrations. This signal contamination might result in errors in calculating SCG features and eventually inaccurate signal classification, especially if automated SCG processing is performed (i.e., without human supervision). For example, a recent study [92] showed that, when determining the instantaneous frequency of SCG signals using different time-frequency distributions, estimation accuracy differed significantly with the signal-to-noise ratio. These results indicated that some time-frequency distributions performed poorly in noisy conditions and would lead to inaccurate time-frequency features. It was then concluded that feature extraction methods might fail or, at a minimum, perform inaccurately for low signal-to-noise ratio conditions.
Most research groups applied conventional band-pass filters to remove baseline wandering, body movements, and breathing artefacts from SCG signals [26,36,38,41,45,46,55,58–63,67,71,75,76,78–80,82,93]. A few studies utilized or proposed more advanced noise removal techniques [64,76,88,94–96]. A recent study [94] proposed a filtering algorithm based on the ensemble empirical mode decomposition (EEMD) to remove white Gaussian noise from SCG signals. This algorithm provided a higher signal-to-noise ratio than other filters such as Wiener filters. In a different study [76], a filtering algorithm based on empirical mode decomposition (EMD) was suggested to filter the SCG signals recorded during walking from a wearable device. This EMD-based denoising approach appeared to result in better estimations of PEP during walking. However, the EMD method generally suffers from mode mixing, and the EEMD algorithm was proposed to resolve this issue [97]. Thus, employment of EEMD in future studies might result in a more accurate denoising of SCG signals during walking, and eventually better estimation of cardiac time intervals such as PEP. Other noise reduction methods, including wavelet transform, adaptive filters, and morphological techniques (e.g., using top hat transform), were also used to remove noise from SCG signals [95].
Daily physical activities, such as walking, introduce noise into the recorded SCG signals and affect their morphology. Therefore, techniques that can remove noise from ambulatory SCG are essential. One study [98] used an evolving fuzzy neural network algorithm to identify the SCG cycles polluted by movement artefacts and remove them from the SCG signal. In another effort [72], a normalized least-mean-square (NLMS) adaptive filter was utilized to cancel the motion noise from SCG of ambulatory subjects. The results of that study showed that adaptive filtering was promising in denoising SCG signals captured during walking. To improve these results, another study [99] utilized a dual-sensor approach where the SCG signals from the anterior and posterior chest wall were acquired. An NLMS adaptive filter algorithm was then used to remove the motion artefact from SCG signal. The noise cancellation performance was calculated and compared for five different reference sensor placement spots around the chest wall. Results showed that using two SCG sensors can lead to a better motion noise cancellation than using a single sensor. Using multiple sensors, however, will increase system complexity.
Some studies pointed out the importance of assessing day-to-day variability when developing a robust system of SCG analysis. Pouyan et al. [100] proposed an algorithm based on a graph-mining technique, called graph similarity score, which was robust to noise and day-to-day variability and could be used to evaluate the risk of HF-related exacerbations for patients at home. A summary of noise removal techniques utilized for SCG denoising is listed in Table 3. More studies are needed that compare different filtering methods in clinical and ambulatory settings.
Table 3.
Method | Application | Reference |
---|---|---|
low-, band-, high-pass, notch filtering | Baseline wandering, breathing and body movement artefact removal | [26,36,38,41,45,46,55,58–63,67,71,75,76,78–80,82,93] |
Adaptive filtering | Motion artefact removal | [88,95] |
Averaging theory | Motion artefact removal | [101] |
Comb filtering | Removing respiration noise from radar signal | [50] |
Empirical mode decomposition | Baseline wandering, breathing and body movement artefact removal | [76,94,95] |
Independent component analysis | Motion artefact removal | [102] |
Median filtering | [96] | |
Morphological filtering | [95] | |
Polynomial smoothing | Motion artefact removal | [103] |
Savitzky–Golay filtering | Motion artefact removal | [83,103] |
Wavelet denoising | Segmentation of HSs and SCG | [64,95,96] |
Wiener filtering | [94] |
2.2. Segmentation
Signal segmentation is one of the first steps in the processing of SCG signals. Segmentation is the process of finding SCG events (i.e., cardiac cycles) in the SCG signal. Different methods and algorithms were used for SCG segmentation. For example, Jain and Tiwari [64] proposed a three-step algorithm where the signal was first filtered using a denoising algorithm based on discrete wavelet transform. The denoised signal peaks were then detected using an adaptive threshold based on Otsu’s method. The first and second components of SCG (corresponding to the first and second heart sounds, i.e., S1 and S2) were finally identified based on the signal energy. Other methods, such as matched filtering with a template consisting of previously identified SCG events, were also used for SCG segmentation [66,67].
2.3. Feature Extraction
Feature extraction is yet another step of SCG signal processing. Identifying the most significant signal features can result in efficient signal classification since these features are eventually the inputs to machine learning algorithms. Determining the most effective and accurate techniques to extract specific signal features is a necessary step that should be done before identification of useful features. For example, there are different methods for estimating the time-frequency distribution of the SCG signal. Every method has its own advantages and disadvantages, and might be suitable for certain types of signals or under certain conditions. Several studies were done to determine the most accurate methods for extracting time-frequency features of the SCG signals [67,104,105]. In these studies, different time-frequency distribution techniques were utilized, including short-time Fourier transform [67,104,105], polynomial chirplet transform (PCT) [67,105], wavelet transform with different mother functions [67], Wigner–Ville distribution, and smoothed pseudo Wigner–Ville distribution (SPWVD) [105]. PCT and SPWVD were found to have the most accurate time–frequency distribution estimations and appeared more suited for determining the frequency content of SCG signals. Using these methods, SCG signals of healthy subjects were found to contain three main spectral peaks below 100 Hz.
Historically, feature extraction of SCG signals mostly focused on the time domain and the frequency domain, separately. The time domain features include statistical features, such as mean, median, and standard deviation, and features related to cardiac mechanics, such as cardiac time intervals. The frequency domain features include statistical features and frequency coefficients obtained from fast Fourier transform (FFT).
2.3.1. Time-Domain Features
Statistical time-domain features include those based on the entire signal, and those from divided segments of the signal. Features from segments of the SCG signal were obtained by dividing the SCG signal into a specific number of equal-sized bins and calculating the arithmetic mean of each bin as a feature [56,106]. Similarly, one study divided the signal into bins; however, binning of the signal was performed discriminately, where the signal portions corresponding to higher variation received a higher concentration of bins [65]. That algorithm divided the bin corresponding the highest standard deviation in a recursive fashion, until some criteria, such as reaching the desired number of bins, was met. Other statistical time-domain features, such as mean, kurtosis, skewness, and standard deviation were also extracted from the SCG signal [41]. Time-domain features also included features related to cardiac mechanics, heart rate and heart rate variability, and turning point ratios [47,48]. In addition, when the ECG R and Q information is concurrently available with the SCG fiducial points (AO, AC, MO, and MC), certain intuitive time-domain features can be determined. These include CTIs (e.g., PEP, isovolumic contraction time (IVCT), LVET, and isovolumic relaxation time (IVRT)) and other metrics such as PEP/LVET ratio, (IVCT+IVRT)/LVET (also called myocardial performance index), and the LVET/R-R-interval ratio [81,107]. Amplitudes and slopes associated with the fiducial points, such as MC to AO slope, were used in some studies [61,107], as well as features of the SCG signal that do not depend on specific fiducial points, such as maxima, minima, and their associated widths of specific segments of the SCG signal [82].
2.3.2. Frequency-Domain Features
Statistical frequency-domain features include those obtained from various frequency bands, and across the entire available frequency spectrum. Features from the frequency bands of an averaged triaxial SCG signal were obtained by calculating the median of each band [78]. One study calculated the approximate and spectral entropy of the 0–11-Hz frequency band [47]. The average power of various frequency bands (0–3 Hz, 3–6 Hz, …, 15–18 Hz) was also utilized [41]. Various statistical metrics, such as arithmetic mean, median, standard deviation, skewness, kurtosis, mode, average power, sample entropy, spectral entropy, and the Kolmogorov complexity, were also calculated across the entire available frequency band [40,41].
Other frequency-domain features include frequency coefficients such as amplitudes and frequencies. Features were either obtained by taking the frequency amplitudes across a range of the frequency spectrum (0–512 Hz) [56], or by taking the frequencies and amplitudes at specific peaks of the spectrum, such as the first, second, and third peaks [41].
In summary, successful feature extraction from SCG signals results in a more efficient classification of these signals. Different studies that investigated the utility of various feature extraction methods/algorithms in both time and frequency domains were described in this section. However, more studies can possibly lead to improve the available methods and define more effective features. In addition, the features currently extracted from SCG signals can be categorized into intuitive (e.g., LVET) and non-intuitive (e.g., skewness) features. Future studies can also address the question of which intuitive or non-intuitive features can be more useful in classification of SCG signals. A summary of these features are listed in Table 4.
Table 4.
2.4. Machine Learning
Applications of predictive methods such as machine learning are increasingly being used in biomedical signal processing, including for SCG analysis. Much inter- and intra-subject variability exists in SCG signals and machine learning can be used to automatically recognize the underlying patterns. Some of the applications of machine learning techniques include detection of cardiovascular disease, cardiac mechanics, and parameters affecting SCG waveform such as respiration cycles.
Some studies sought to use classification tools such as support vector machines (SVM) and neural networks (NN) to automatically detect cardiovascular disease. An early study using NNs [107] classified patients based on their SCG as either having coronary artery disease (CAD) or as low risk/normal. They predicted CAD with a sensitivity of 80% and a specificity of 80%. Recent studies [47,48] sought to classify cardiovascular conditions with SCG signals obtained via a smartphone’s inertial measurement unit (IMU). A multi-class classifier was used [48] to classify subjects as either having ST-elevation myocardial infarction, having atrial fibrillation, being preferred for percutaneous coronary intervention procedure, or normal. The proposed classifier achieved classification accuracies between 70 and 79%. However, the same study [48] created a binary classifier (normal vs. atrial fibrillation) and achieved an accuracy of 98.7% using an SVM.
Classification was also used to detect the respiration cycles (inspiration and expiration) [56,106] and lung volume (high and low lung volume) [40,65]. In one study [56] classifying respiration cycles, two different training scenarios were implemented. The first was a leave-one-subject-out (LOSO) approach, which trained the SVM on all but one subject, and tested on the subject who was left out. The second was a subject-specific (SS) approach, which trained and tested on each subject individually. The average accuracies for LOSO and SS were 88.1% and 95.4% respectively. Other studies [40,65] sought to classify SCG signals according to the lung volume phases as opposed to inspiration/expiration.
Classification methods were also utilized to help identify fiducial points on the SCG signals [61], artefact presence in the SCG [98], and identification of the sensor location [41,82].
Other machine learning methods were used on SCG signals such as hidden Markov models (HMM) and graph similarity analysis. An HMM-based method was used in one study [108] to estimate the heart rate, heart rate variability, and CTIs from an SCG signal. A graph similarity analysis [78] was used in another study through the use of k-nearest neighbor graphs on SCG signals from HF patients to identify them as compensated (outpatient) or decompensated (hospitalized).
In summary, machine learning algorithms were used for different purposes in SCG studies, including SCG classification into different phases of respiratory cycle (e.g., high vs. low lung volume), determining fiducial points (e.g., IM and AO) and cardiac time intervals (e.g., PEP), and classification of subjects into patients and low risk/normal. A summary of the machine learning algorithms used for SCG analysis is listed in Table 5.
Table 5.
3. Recent Human Studies Suggesting Clinical Utilities
Early use of SCG for cardiac diagnosis faced obstacles such as the large instrumentation size and unclear understanding of the signal characteristics and inter- and intra-subject variabilities. However, recent advances in sensor technologies and signal processing methods led, at least in part, to new numerous studies that provided better insight into these issues. The high morbidity and mortality associated with cardiovascular disease and the high cost of care may have provided motivation to more studies that re-evaluated the feasibility and utility of seismocardiography for diagnosis and monitoring of cardiac function [77,111,112]. Some of the studies reviewed here focused on telemonitoring of cardiac time intervals and heart rates.
3.1. Portable Detection of SCG
Wearable technologies can continuously monitor cardiac activity outside clinics and hospitals. This continuous monitoring might help in early detection of serious cardiac conditions, which can enable timely intervention and potentially reduce healthcare costs. Most current wearable cardiac activity monitoring techniques are based on ECG measurements. However, recent studies proposed wearable SCG systems for the assessment of the mechanical aspects of cardiovascular function, including relative changes in cardiac output, contractility, and blood pressure [113]. SCG wearable monitors might be used to assess myocardial contractility via pre-ejection period (PEP) [81]. Another wearable system utilized triaxial accelerometers and gyroscopes to record all six axial and rotational components of the SCG signals [36]. The rotational vibration about the longitudinal (head-to-foot) axis showed a lower sensitivity to walking noise than other components, which might be useful for annotation of SCG signals in ambulant subjects [36].
Today, smartphones and smartwatches are common and can be used as part of telemedicine for real-time patient monitoring at a relatively low cost. Smartphones used SCG for continuous monitoring of heart rate variability [114], and cardiac activity of patients suffering from heart disease [115]. In a recent study, the feasibility and accuracy of measuring heart rate using a smartphone accelerometer was assessed in different postural positions [46] and suggested utility of SCG for heart rate estimation.
Wearable SCGs might be contaminated with different type of noise. Therefore, investigating the effective noise removal techniques for ambulatory subjects is needed. A few ongoing studies are addressing this question. These studies were described in Section 2.1.
3.2. Heart Rate Monitoring
Heart rate (HR) monitoring is a common way to monitor cardiovascular function, and can identify some abnormalities such as arrhythmia. Traditional HR estimation methods are mostly based on ECG signal processing. SCG signals can also be used for HR estimation. SCG-based HR estimation algorithms are not usually developed to replace the current HR monitoring methods. Instead, SCG-based HR estimations can be used as a feature in other studies that focus on the clinical utility of SCG, since recent studies revealed that SCG can reliably detect HR in the absence of other modalities such as ECG. For example, Cosoli et al. [26] suggested a general algorithm that can estimate HR from various signals, including SCG, ECG, phonocardiogram (PCG), and PPG. Considering the ECG signal as a gold standard, the SCG HR estimation was more accurate than the estimations from the PCG and PPG signals. Wahlstrom et al. [108] used an HMM to determine different stages of a cardiac cycle, which were used for estimating beat-to-beat intervals. The beat-to-beat intervals of the SCG signal were then utilized for HR and HR variability estimations. Mafi [116] suggested an algorithm based on empirical mode decomposition and empirical wavelet transform that can extract HR from SCG signals. Tadi et al. [25] used a Hilbert adaptive beat identification technique to determine the heartbeat timings and inter-beat time intervals from SCG signals. An android application was implemented based on this algorithm that could monitor the subject heart rate in real time using a smartphone accelerometer. Tadi et al. [69] proposed an algorithm based on S-transform, Shannon energy, and successive mean quantization transform to identify heartbeat and beat-to-beat interval from SCGs. The algorithms proposed in the latter two studies had a high agreement with the ECG HR. Taebi et al. [39] used SCG signals in the dorso-ventral direction to estimate the HR during different phases of respiration in real time. Their results showed that normal subjects have a different HR during high and low lung volumes. In a recent study [46], the heart rate was derived from a smartphone SCG signal, and compared to that extracted from ECG. Results showed that the HR provided by SCG, particularly in the dorso-ventral direction of the supine position, was equivalent to that provided by conventional ECG.
3.3. Pulse Transit Time
SCG is used to estimate different cardiovascular parameters such as cardiac time intervals, pulse transit time, and blood pressure. For example, non-contact SCG was used at different body locations for estimating central arterial pressure and carotid arterial pressure waveforms [49,117,118]. Pulse transit time might be estimated from the time difference between AO point on the xiphoid SCG and AO point on the carotid SCG [119]. Blood pressure changes can be monitored using pulse transit time. For this purpose, the pulse transit time, which was defined as the time required for the blood pressure wave to travel from one location to another [120], was first measured from the SCG signals [121]. The measured pulse transit time was then used to estimate the patient blood pressure [79]. Based on similar techniques, a wrist-watch, consisting of an accelerometer and an optical sensor, was developed to monitor blood pressure [122]. In this “SeismoWatch”, the blood pressure was estimated from the travel time of the micro-vibrations propagating from the heart to the wrist when the watch was held against the subject’s sternum. In a different study, Di Rienzo et al. [73] developed a system that measures SCG and PPG at multiple locations alongside the ECG signal. The pulse transit time may then be derived from the PPG.
3.4. Cardiac Time Intervals
Cardiac time intervals were used for a long time for cardiovascular disease diagnosis [123,124]. There are various SCG-based algorithms with different levels of accuracy that were proposed for estimating cardiac time intervals such as PEP, LVET, ICT, systolic time, and diastolic time in healthy subjects and patients with previous heart conditions [61,108]. For example, LVET might be estimated from SCG signals that are acquired using LDV and microwave Doppler radar [51,117]. The LVET value from non-contact SCG was similar to the value derived from a photoplethysmogram (PPG) [117]. Rivero et al. [125] proposed a new algorithm that uses continuous wavelet transform as a base to determine the aortic valve opening and isovolumic moment points on the SCG signal. The electro-mechanical window (EMw) is defined as the duration between the electrical and mechanical systole. EMw is a potential biomarker that can be utilized for diagnosing several cardiovascular diseases. ECG and PCG signals are conventionally used to determine EMw. However, Jain et al. [57] showed that SCG is a suitable alternative to PCG for estimating the EMw.
Analysis of SCG data recorded from the sleep patterns of a subject aboard the International Space Station (in microgravity) resulted in accurate identification of cardiac time intervals and SCG fiducial points (such as AO, AC, MO, MC, LVET, and PEP) with implications for future clinical application [81]. As described earlier, SCG morphology is affected by different factors such as the sensor location and respiration. Investigating the effect of these factors on the estimation of cardiac time intervals from SCG signals can possibly reveal clinically useful information.
3.5. SCG in Patients with Cardiac Conditions
In addition to human studies on healthy populations, there were several studies that focused on the application of SCG in patients with cardiovascular disease. SCG signals were used for diagnosis and monitoring of different clinical conditions such as atrial fibrillation [47,48,68,70], atrial flutter [55], heart valve disease [37,44,80], coronary artery disease and ischemia [9,10,48], myocardial infarction [126], heart failure [75,78,82,100,119], structural heart disease [80], and heart stress testing [58].
According to a 2017 report, the prevalence of any heart valve disease is 2.5% of the United States population [127]. Heart sounds that are believed to be generated by opening and closure of heart valves can be used as a diagnostic marker of these diseases. Stethoscope and PCG are the common conventional methods for heart sound monitoring. SCG signals were reported as a potential efficient alternative for PCG signals for monitoring of heart sound signals [44].
ECG is currently the main diagnostic method of atrial fibrillation (AF). A preclinical study [68] investigated the usefulness of SCG for AF detection. Results suggested that the amplitude of the SCG signal correlates to beat interval and significantly varies from beat to beat during AF. This study also suggested that the combination of SCG and ECG may reveal certain behavior in the electromechanical delay characteristic of AF, which may lead to extra indicators for early detection of AF.
Paukkunen et al. [55] showed that three-dimensional (3D) vector trajectory of SCG might be useful in diagnosing atrial flutter. The results of this study suggested that the intra-subject correlation of 3D SCGs was strong. However, the signals had a very weak inter-subject correlation. Future studies might prove the utility of SCG 3D vector trajectory for diagnosis of different cardiovascular disease and abnormalities.
4. Conclusions, Limitations, and Future Directions
Growth in the field of seismocardiography accelerated during the last decade. However, open issues and limitations hamper its clinical application. Reviewed here are some of the current limitations along with potential future work.
SCG variability: SCG morphology is affected by different factors such as respiration, sensor location, subject posture, the amount of chest surface soft tissue, and different heart diseases. Although studies investigated some of these factors, further research is needed to adequately account for SCG variations. The results of such investigations might improve utility for cardiac disease diagnosis and monitoring.
Lack of accepted standard for the cardio-mechanical SCG fiducial points: A small number of recent studies focused on robust documentation of the relationship between fiducial points and their physiological sources. In addition, SCG morphology changes with different factors (e.g., sensor location, patient posture, etc.). It would be useful to investigate the effect of these factors on the SCG signal fiducial points.
SCG genesis: Although several studies aimed to elucidate the physiological source(s) of the SCG signals, much work remains to be done. SCG signals are likely affected by extra-cardiac factors including respiration and intrathoracic pressure. Therefore, considering these parameters may further help elucidate SCG sources.
Computational models: Realistic computational simulations utilizing finite element and other modeling methodologies with realistic geometries and material properties might be helpful in predicting the effects of varied cardiac conditions on SCG features.
Library of SCG signals: A common comprehensive database would provide a basis for researchers interested in analyzing SCG and other biomedical signals. The MIT-BIH arrhythmia database is a good example of a biomedical signals database. This database plays an important role in stimulating both basic research and medical device development. A similar SCG database would attract more researchers to investigate and compare the performance of different algorithms and approaches.
Correlation between SCG and other electro-mechanical signals: Combining information from ECG, PCG, and BCG with that of SCG may lead to a hybrid modality with increased diagnostic utility of cardiac disease. This may result in more complex features that require increase use of machine learning approaches.
In conclusion, signal processing techniques and physiologic understandings rigorously applied may transform SCG signal analysis from a research interest to a powerful bedside or home monitoring tool.
Acknowledgments
Funding: This research was funded by the National Institutes of Health, grant number R01 EB012142.
Footnotes
Conflicts of Interest: The authors declare no conflicts of interest.
References
- 1.Mozaffarian D; Benjamin EJ; Go AS; Arnett DK; Blaha MJ; Cushman M; de Ferranti S; Despres J-P; Fullerton HJ; Howard VJ; et al. Heart Disease and Stroke Statistics—2015 Update: A Report From the American Heart Association. Circulation 2015,131, e29–e322. [DOI] [PubMed] [Google Scholar]
- 2.Lee L; Markwick A; Horsfall M; Sinhal A; Chew D Prognostic implications of moderate and severe mitral regurgitation in contemporary clinical care. Eurointervention 2012, 8, Q111. [Google Scholar]
- 3.Otto CM Timing of aortic valve surgery. Heart 2000, 84, 211–218. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Khalili F; Mansy HA Blood Flow through a Dysfunctional Mechanical Heart Valve. In Proceedings of the 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Orlando, FL, USA, 17–20 August 2016. [Google Scholar]
- 5.Khalili F; Gamage PPT; Mansy HA Hemodynamics of a Bileaflet Mechanical Heart Valve with Different Levels of Dysfunction. J. Appl. Biotechnol. Bioeng. 2017, 2, 00044. [Google Scholar]
- 6.Zanetti J; Tavakolian K Seismocardiography: Past, present and future. In Proceedings of the 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan, 3–7 July 2013; pp. 7004–7007. [DOI] [PubMed] [Google Scholar]
- 7.Inan OT; Migeotte PF; Park KS; Etemadi M; Tavakolian K; Casanella R; Zanetti J; Tank J; Funtova I; Prisk GK; et al. Ballistocardiography and Seismocardiography: A Review of Recent Advances. IEEE J. Biomed. Health inform. 2015,19,1414–1427. [DOI] [PubMed] [Google Scholar]
- 8.Starr I; Rawson J; Schroeder H; Joseph N Studies on the estimation of cardiac ouptut in man, and of abnormalities in cardiac function, from the heart’s recoil and the blood’s impacts; the Ballistocardiogram. Am. J. Physiol. 1939,127,1–28. [Google Scholar]
- 9.Salerno DM; Zanetti JM; Green LA; Mooney MR; Madison JD; Van Tassel RA Seismocardiographic changes associated with obstruction of coronary blood flow during balloon angioplasty. Am. J. Cardiol. 1991, 68, 201–207. [DOI] [PubMed] [Google Scholar]
- 10.Salerno DM; Zanetti J Seismocardiography for monitoring changes in left ventricular function during ischemia. Chest 1991,100, 991–993. [DOI] [PubMed] [Google Scholar]
- 11.Pandia K; Inan OT; Kovacs GT; Giovangrandi L Extracting respiratory information from seismocardiogram signals acquired on the chest using a miniature accelerometer. Physiol. Meas. 2012, 33,1643–1660. [DOI] [PubMed] [Google Scholar]
- 12.Bifulco P; Gargiulo GD; D’Angelo G; Liccardo A; Romano M; Clemente F; Cesarelli M; Angelo G; Liccardo A; Romano M; et al. Monitoring of respiration, seismocardiogram and heart sounds by a PVDF piezo film sensor. Measurement 2014,11, 786–789. [Google Scholar]
- 13.Zanetti J; Salerno D Seismocardiography: A new technique for recording cardiac vibrations. Concept, method, and initial observations. J. Cardiovasc. Technol. 1990, 9,111–118. [Google Scholar]
- 14.Crow RS; Hannan P; Jacobs D; Hedquist L; Salerno DM Relationship between Seismocardiogram and Echocardiogram for Events in the Cardiac Cycle. Am. J. Noninvasive Cardiol. 1994, 8, 39–46. [Google Scholar]
- 15.Taebi A; Mansy HA Time-frequency Analysis of Vibrocardiographic Signals. In Proceedings of the 2015 BMES Annual Meeting, Tampa, FL, USA, 7–10 October 2015. [Google Scholar]
- 16.Eddleman E Kinetocardiographic Changes in Ischemic Heart Disease. Circulation 1965, 32, 650–655. [DOI] [PubMed] [Google Scholar]
- 17.Starr I; Wood E Twenty-year studies with the ballistocardiograph, the relation between the amplitude of the first record of ‘health’ adults and eventual mortality and morbidity form heart disease. Circulation 1961, 23, 714–732. [Google Scholar]
- 18.Starr I; Noordegroaf A Ballistocardiography in Cardiovascular Research; Lippincott: Philadelphia, PA, USA, 1967. [Google Scholar]
- 19.Phibbs B; Lowe C; Holmes R The ultra lowfrequency force ballistocardiograph in acute cardiomyopathy. Circulation 1967, 36, 92–100. [DOI] [PubMed] [Google Scholar]
- 20.Scherhag AW; Pfleger S; Ceconi C; Voelker W; Gehring J; Staedt U; Heene DL Evaluation of signal-averaged cardiokymography for the detection of ischaemic left ventricular dysfunction. Int. J. Cardiol. 1997, 59, 305–312. [DOI] [PubMed] [Google Scholar]
- 21.Matsui T; Hagisawa K; Ishizuka T; Takase B; Ishihara M; Kikuchi M A Novel Method to Prevent Secondary Exposure of Medical and Rescue Personnel to Toxic Materials Under Biochemical Hazard Conditions Using Microwave Radar and Infrared Themography. IEEE Trans. Biomed. Eng. 2004, 51, 2184–2188. [DOI] [PubMed] [Google Scholar]
- 22.Morbiducci U; Scalise L; De Melis M; Grigioni M Optical vibrocardiography: A novel tool for the optical monitoring of cardiac activity. Ann. Biomed. Eng. 2007, 35, 45–58. [DOI] [PubMed] [Google Scholar]
- 23.Tavakolian K; Vaseghi A; Kaminska B Improvement of ballistocardiogram processing by inclusion of respiration information. Physiol. Meas. 2008, 29, 771–781. [DOI] [PubMed] [Google Scholar]
- 24.Mack DC; Patrie JT; Suratt PM; Felder RA; Alwan M Development and preliminary validation of heart rate and breathing rate detection using a passive, ballistocardiography-based sleep monitoring system. IEEE Trans. Inf. Technol. Biomed. 2009,13,111–120. [DOI] [PubMed] [Google Scholar]
- 25.Jafari Tadi M; Lehtonen E; Hurnanen T; Koskinen J; Eriksson J; Pankaala M; Teras M; Koivisto T A real-time approach for heart rate monitoring using a Hilbert transform in seismocardiograms. Physiol. Meas. 2016, 37,1885–1909. [DOI] [PubMed] [Google Scholar]
- 26.Cosoli G; Casacanditella L; Tomasini E; Scalise L Heart Rate assessment by means of a novel approach applied to signals of different nature. J. Phys. Conf. Ser. 2017, 778, 012001. [Google Scholar]
- 27.Gurev V; Tavakolian K; Constantino J; Kaminska B; Blaber AP; Trayanova NA Mechanisms underlying isovolumic contraction and ejection peaks in seismocardiogram morphology. J. Med. Biol. Eng. 2012, 32, 103–110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Tadi M; Koivisto T; Pankaala M; Paasio A; Knuutila T; Teras M; Hanninen P A new algorithm for segmentation of cardiac quiescent phases and cardiac time intervals using seismocardiography. In Proceedings of the Sixth International Conference on Graphic and Image Processing (ICGIP 2014), Beijing, China, 24–26 October 2015; p. 94432K. [Google Scholar]
- 29.Tavakolian K; Khosrow-Khavar F; Kajbafzadeh B; Marzencki M; Rohani S; Kaminska B; Menon C Seismocardiographic adjustment of diastolic timed vibrations. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS, San Diego, CA, USA, 28 August-1 September 2012; pp. 3797–3800. [DOI] [PubMed] [Google Scholar]
- 30.Zanetti JM; Poliac MO; Crow RS Seismocardiography: Waveform identification and noise analysis. In Proceedings of the Computers in Cardiology 1991, Venice, Italy, 23–26 September 1991; pp. 49–52. [Google Scholar]
- 31.Tadi MJ; Lehtonen E; Saraste A; Tuominen J; Koskinen J; Teräs M; Airaksinen J; Pänkäälä M; Koivisto T Gyrocardiography: A new non-invasive monitoring method for the assessment of cardiac mechanics and the estimation of hemodynamic variables. Sci. Rep. 2017, 7, 6823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Tavakolian K Charachterization and Analysis of Seismocardiogram for Estimation of Hemodynamic Parameters. Ph.D. Thesis, Simon Fraser University, Burnaby, BC, Canada, 2010. [Google Scholar]
- 33.Choudhary T; Sharma LN; Bhuyan MK Automatic Detection of Aortic Valve Opening using Seismocardiography in Healthy Individuals. IEEE J. Biomed. Health Inform. 2018. [DOI] [PubMed] [Google Scholar]
- 34.Lin W; Chou W; Chang P; Chou C; Wen M; Ho M; Lee M Identification of Location Specific Feature Points in a Cardiac Cycle Using a Novel Seismocardiogram Spectrum System. IEEE J. Biomed. Health Inform. 2018, 22, 442–449. [DOI] [PubMed] [Google Scholar]
- 35.Sengupta PP; Krishnamoorthy VK; Korinek J; Narula J; Vannan MA; Lester SJ; Tajik JA; Seward JB; Khandheria BK; Belohlavek M Left Ventricular Form and Function Revisited: Applied Translational Science to Cardiovascular Ultrasound Imaging. J. Am. Soc. Echocardiogr. 2007, 20, 539–551. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Yang C; Tang S; Tavassolian N Utilizing Gyroscopes Towards the Automatic Annotation of Seismocardiograms. IEEE Sens. J. 2017,17, 2129–2136. [Google Scholar]
- 37.Migeotte P-F; Mucci V; Delière Q; Lejeune L; van de Borne P Multi-dimensional kineticardiography a new approach for wearable cardiac monitoring through body acceleration recordings. In XIV Mediterranean Conference on Medical and Biological Engineering and Computing 2016; Springer: Cham, Switzerland, 2016; pp. 1125–1130. [Google Scholar]
- 38.Yang C; Tavassolian N Combined Seismo-and Gyro-cardiography: A More Comprehensive Evaluation of Heart-Induced Chest Vibrations. IEEE J. Biomed. Health Inform. 2017. [DOI] [PubMed] [Google Scholar]
- 39.Taebi A; Bomar AJ; Sandler RH; Mansy HA Heart Rate Monitoring During Different Lung Volume Phases Using Seismocardiography. In Proceedings of the IEEE SoutheastCon 2018, Saint Petersburg, FL, USA, 19–22 April 2018; pp. 1–5. [Google Scholar]
- 40.Solar BE; Taebi A; Mansy HA Classification of Seismocardiographic Cycles into Lung Volume Phases. In 2017 IEEE Signal Processing in Medicine and Biology Symposium (SPMB); IEEE: Philadelphia, PA, USA, 2017; pp.1–2. [Google Scholar]
- 41.Ashouri H; Inan OT Automatic Detection of Seismocardiogram Sensor Misplacement for Robust Pre-Ejection Period Estimation in Unsupervised Settings. IEEE Sens. J. 2017,17, 3805–3813. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Koegelenberg S Application of Laser Doppler Vibrocardiography for Human Heart Auscultation. Ph.D. Thesis, Stellenbosch University, Stellenbosch, South Africa, 2014. [Google Scholar]
- 43.Jain PK; Tiwari AK A novel method for suppression of motion artifacts from the seismocardiogram signal. In Proceedings of the 2016 IEEE International Conference on Digital Signal Processing (DSP), Beijing, China, 16–18 October 2016; pp. 6–10. [Google Scholar]
- 44.Jain PK; Tiwari AK; Chourasia VS Performance analysis of seismocardiography for heart sound signal recording in noisy scenarios. J. Med. Eng. Technol. 2016, 40,106–118. [DOI] [PubMed] [Google Scholar]
- 45.Yang C; Tang S; Tavassolian N Annotation of seismocardiogram using gyroscopic recordings. In Proceedings of the 2016 IEEE Biomedical Circuits and Systems Conference (BioCAS), Shanghai, China, 17–19 October 2016; pp. 204–207. [Google Scholar]
- 46.Landreani F; Martin-Yebra A; Casellato C; Frigo C; Pavan E; Migeotte P-F; Caiani EG Beat-to-beat heart rate detection by smartphone’s accelerometers: Validation with ECG. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 525–528. [DOI] [PubMed] [Google Scholar]
- 47.Lahdenoja O; Hurnanen T; Iftikhar Z; Nieminen S; Knuutila T; Saraste A; Kiviniemi T; Vasankari T; Airaksinen J; Pankaala M; et al. Atrial Fibrillation Detection via Accelerometer and Gyroscope of a Smartphone. IEEE J. Biomed. Health Inform. 2018,22,108–118. [DOI] [PubMed] [Google Scholar]
- 48.Iftikhar Z; Lahdenoja O; Tadi MJ; Hurnanen T; Vasankari T; Kiviniemi T; Airaksinen J; Koivisto T; Pänkäälä M Multiclass Classifier based Cardiovascular Condition Detection Using Smartphone Mechanocardiography. Sci. Rep. 2018, 8, 9344. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Metzler J; Kroschel K; Willersinn D Automatic detection of measurement points for non-contact vibrometer-based diagnosis of cardiac arrhythmias. In SPIE Medical Imaging. International Society for Optics and Photonics; Webster RJ, Fei B, Eds.; SPIE: Bellingham, WA, USA, 2017; p. 101351S. [Google Scholar]
- 50.Xia Z; Shandhi MDMH, Inan OT; Zhang Y Interference and Removal of Respiration Harmonics on Noncontact Seismo-cardiogram Signals. In Structural Health Monitoring 2017; DEStech Publications: Lancaster, PA, USA, 2017. [Google Scholar]
- 51.Shandhi MDMH; Xia Z; Inan OT; Zhang Y Clutter Effect on the Noncontact Seismocardiogram Signals Measured using Microwave Radars. In Structural Health Monitoring 2017; DEStech Publications: Lancaster, PA, USA, 2017. [Google Scholar]
- 52.Xia Z; Shandhi MMH; Inan OT; Zhang Y Non-Contact Sensing of Seismocardiogram Signals Using Microwave Doppler Radar. IEEE Sens. J. 2018,18, 5956–5964. [Google Scholar]
- 53.Shirkovskiy P; Laurin A; Jeger-Madiot N; Chapelle D; Fink M; Ing RK Airborne ultrasound surface motion camera: Application to seismocardiography. Appl. Phys. Lett. 2018,112, 213702. [Google Scholar]
- 54.Alamdari NT A Morphological Approach To Identify Respiratory Phases Of Seismocardiogram. Master’s Thesis, The University of North Dakota, Grand Forks, ND, USA, 2016. [Google Scholar]
- 55.Paukkunen M; Parkkila P; Hurnanen T; Pankaala M; Koivisto T; Nieminen T; Kettunen R; Sepponen R Beat-by-Beat Quantification of Cardiac Cycle Events Detected From Three-Dimensional Precordial Acceleration Signals. IEEE J. Biomed. Health Inform. 2016,20, 435–439. [DOI] [PubMed] [Google Scholar]
- 56.Zakeri V; Akhbardeh A; Alamdari N; Fazel-Rezai R; Paukkunen M; Tavakolian K Analyzing Seismocardiogram Cycles to Identify the Respiratory Phases. IEEE Trans. Biomed. Eng. 2017, 64,1786–1792. [DOI] [PubMed] [Google Scholar]
- 57.Jain PK; Tiwari AK; Bhagat OL Seismocardiography: An alternate method to estimate electro-mechanical window. In Proceedings of the 2016 International Conference on Systems in Medicine and Biology (ICSMB), Kharagpur, India, 4–7 January 2016; pp. 136–139. [Google Scholar]
- 58.Dinh A; Bui FM; Nguyen T An accelerometer based system to measure myocardial performance index during stress testing. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 4877–4880. [DOI] [PubMed] [Google Scholar]
- 59.Zhang G; Cottrell AC; Henry IC; McCombie DB Assessment of pre-ejection period in ambulatory subjects using seismocardiogram in a wearable blood pressure monitor. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 3386–3389. [DOI] [PubMed] [Google Scholar]
- 60.Khosrow-Khavar F Automatic and Non-Invasive Delineation of the Seismocardiogram Signal for the Estimation of Cardiac Time Intervals with Applications in Diastolic Timed Vibration and Early Stage Hemorrhage Detection. Ph.D. Thesis, Simon Fraser University, Burnaby, BC, Canada, 2016. [Google Scholar]
- 61.Khosrow-Khavar F; Tavakolian K; Blaber A; Menon C Automatic and Robust Delineation of the Fiducial Points of the Seismocardiogram Signal for Noninvasive Estimation of Cardiac Time Intervals. IEEE Trans. Biomed. Eng. 2017, 64,1701–1710. [DOI] [PubMed] [Google Scholar]
- 62.Shafiq G; Tatinati S; Veluvolu KC Automatic annotation of peaks in seismocardiogram for systolic time intervals. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 2672–2675. [DOI] [PubMed] [Google Scholar]
- 63.Shafiq G; Tatinati S; Ang WT; Veluvolu KC Automatic Identification of Systolic Time Intervals in Seismocardiogram. Sci. Rep. 2016, 6, 37524. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Jain PK; Tiwari AK An algorithm for automatic segmentation of heart sound signal acquired using seismocardiography. In Proceedings of the 2016 International Conference on Systems in Medicine and Biology (ICSMB), Kharagpur, India, 4–7 January, 2016; pp. 157–161. [Google Scholar]
- 65.Taebi A; Solar BE; Mansy HA An Adaptive Feature Extraction Algorithm for Classification of Seismocardiographic Signals. In Proceedings of the IEEE SoutheastCon 2018, Saint Petersburg, FL, USA, 19–22 April 2018; pp. 1–5. [Google Scholar]
- 66.Taebi A; Mansy HA Grouping Similar Seismocardiographic Signals Using Respiratory Information. In 2017 IEEE Signal Processing in Medicine and Biology Symposium (SPMB); IEEE: Philadelphia, PA, USA, 2017; pp.1–6. [Google Scholar]
- 67.Taebi A; Mansy HA Time-Frequency Distribution of Seismocardiographic Signals: A Comparative Study. Bioengineering 2017, 4, 32. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Pänkäälä M; Koivisto T; Lahdenoja O; Kiviniemi T; Saraste A; Vasankari T; Airaksinen J Detection of atrial fibrillation with seismocardiography. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 4369–4374. [DOI] [PubMed] [Google Scholar]
- 69.Tadi MJ; Lehtonen E; Lahdenoja O; Pankaala M; Koivisto T An adaptive approach for heartbeat detection based on S-transform in seismocardiograms. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 2370–2373. [DOI] [PubMed] [Google Scholar]
- 70.Hurnanen T; Lehtonen E; Tadi MJ; Kuusela T; Kiviniemi T; Saraste A; Vasankari T; Airaksinen J; Koivisto T; Pankaala M Automated Detection of Atrial Fibrillation Based on Time-Frequency Analysis of Seismocardiograms. IEEE J. Biomed. Health Inform. 2017, 21,1233–1241. [DOI] [PubMed] [Google Scholar]
- 71.Sahoo PK; Thakkar HK; Lin W-Y; Chang P-C; Lee M-Y On the Design of an Efficient Cardiac Health Monitoring System Through Combined Analysis of ECG and SCG Signals. Sensors 2018,18, 379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Yang C; Tavassolian N Motion Artifact Cancellation of Seismocardiographic Recording From Moving Subjects. IEEE Sens. J. 2016,16, 5702–5708. [Google Scholar]
- 73.Di Rienzo M; Lombardi P; Scurati D; Vaini E A new technological platform for the multisite assessment of 3D seismocardiogram and pulse transit time in cardiac patients. In Proceedings of the 2016 Computing in Cardiology Conference (CinC), Vancouver, BC, Canada, 11–14 September 2016; pp. 781–784. [Google Scholar]
- 74.Canino NK; Wang RR; Robinson CJ Using multiple placements of accelerometers to measure cardiovascular pulse transit times. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 4248–4251. [DOI] [PubMed] [Google Scholar]
- 75.Javaid AQ; Dowling S; Etemadi M; Heller JA; Roy S; Klein L; Inan OT Quantification of posture induced changes in wearable seismocardiogram signals for heart failure patients. In Proceedings of the 2016 Computing in Cardiology Conference (CinC), Vancouver, BC, Canada, 11–14 September 2016; pp. 777–780. [Google Scholar]
- 76.Javaid AQ; Ashouri H; Dorier A; Etemadi M; Heller JA; Roy S; Inan OT Quantifying and Reducing Motion Artifacts in Wearable Seismocardiogram Measurements During Walking to Assess Left Ventricular Health. IEEE Trans. Biomed. Eng. 2017, 64,1277–1286. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Chuo Y; Tavakolian K; Kaminska B Evaluation of a novel integrated sensor system for synchronous measurement of cardiac vibrations and cardiac potentials. J. Med. Syst. 2011, 35, 445–455. [DOI] [PubMed] [Google Scholar]
- 78.Inan OT; Baran Pouyan M; Javaid AQ; Dowling S; Etemadi M; Dorier A; Heller JA; Bicen AO; Roy S; De Marco T; et al. Novel Wearable Seismocardiography and Machine Learning Algorithms Can Assess Clinical Status of Heart Failure Patients. Circ. Heart Fail. 2018,11, e004313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Yang C; Tavassolian N Pulse Transit Time Measurement Using Seismocardiogram, Photoplethysmogram, and Acoustic Recordings: Evaluation and Comparison. IEEE J. Biomed. Health Inform. 2017. [DOI] [PubMed] [Google Scholar]
- 80.Yao J; Tridandapani S; Wick CA; Bhatti PT Seismocardiography-Based Cardiac Computed Tomography Gating Using Patient-Specific Template Identification and Detection. IEEE J. Transl. Eng. Health Med. 2017,5, 1–14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Di Rienzo M; Vaini E; Lombardi P An algorithm for the beat-to-beat assessment of cardiac mechanics during sleep on Earth and in microgravity from the seismocardiogram. Sci. Rep. 2017, 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Ashouri H; Hersek S; Inan OT Universal Pre-Ejection Period Estimation using Seismocardiography: Quantifying the Effects of Sensor Placement and Regression Algorithms. IEEE Sens. J. 2018,18,1665–1674. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Lee H; Lee H; Whang M An Enhanced Method to Estimate Heart Rate from Seismocardiography via Ensemble Averaging of Body Movements at Six Degrees of Freedom. Sensors 2018, 18, 238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Antognoli L; Marchionni P; Nobile S; Carnielli VP; Scalise L Assessment of cardio-respiratory rates by non-invasive measurement methods in hospitalized preterm neonates. In Proceedings of the 2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Rome, Italy, 11–13 June 2018; pp. 1–5. [Google Scholar]
- 85.Kon S; Oldham K; Horowitz R Piezoresistive and piezoelectric MEMS strain sensors for vibration detection. In Sensors and Smart Structures Technologies for Civil, Mechanical, and Aerospace Systems 2007; SPIE: Bellingham, WA, USA, 2007; Volume 6529, p. 65292V. [Google Scholar]
- 86.Libonati JR; Colby AM; Caldwell TM; Kasparian R; Glassberg HL Systolic and diastolic cardiac function time intervals and exercise capacity in women. Med. Sci. Sports Exerc. 1999, 31, 258–263. [DOI] [PubMed] [Google Scholar]
- 87.Korzeniowska-kubacka I; Kusmierczyk-droszcz B; Bilinska M; Barbara D-W; Mazurek K; Ryszard P Seismocardiography—A noninvasive method of assessing systolic and diastolic left ventricular function in ischaemic heart disease. Folia Cardiol. 2006,13, 319–325. [Google Scholar]
- 88.Di Rienzo M; Vaini E; Castiglioni P; Merati G; Meriggi P; Parati G; Faini A; Rizzo F Wearable seismocardiography: Towards a beat-by-beat assessment of cardiac mechanics in ambulant subjects. Auton. Neurosci. Basic Clin. 2013,178, 50–59. [DOI] [PubMed] [Google Scholar]
- 89.Castiglioni P; Faini A; Parati G; Di Rienzo M Wearable seismocardiography. In Proceedings of the 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France, 22–26 August 2007; pp. 3954–3957. [DOI] [PubMed] [Google Scholar]
- 90.Inan OT; Etemadi M; Paloma A; Giovangrandi L; Kovacs GTA Non-invasive cardiac output trending during exercise recovery on a bathroom-scale-based ballistocardiograph. Physiol. Meas. 2009, 30, 261–274. [DOI] [PubMed] [Google Scholar]
- 91.Di Rienzo M; Vaini E; Castiglioni P; Meriggi P; Rizzo F Beat-to-beat estimation of LVET and QS2 indices of cardiac mechanics from wearable seismocardiography in ambulant subjects. In Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan, 3–7 July 2013; pp. 7017–7020. [DOI] [PubMed] [Google Scholar]
- 92.Taebi A; Mansy HA Effect of Noise on Time-frequency Analysis of Vibrocardiographic Signals. J. Bioeng. Biomed. Sci. 2016, 6, 202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Abbasi-Kesbi R; Valipour A; Imani K Cardiorespiratory System Monitoring Using a Developed Acoustic Sensor. Healthc. Technol. Lett. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Taebi A; Mansy HA Noise Cancellation from Vibrocardiographic Signals Based on the Ensemble Empirical Mode Decomposition. J. Appl. Biotechnol. Bioeng. 2017, 2, 00024. [Google Scholar]
- 95.Sundar A; Pahwa V Evaluating the Performance of State of the Art Algorithms for Enhancement of Seismocardiogram Signals. In Proceedings of the First International Conference on Intelligent Computing and Communication; Springer: Singapore, 2017; pp. 37–45. [Google Scholar]
- 96.Choudhary T; Sharma LN; Bhuyan MK Heart Sound Extraction from Sternal Seismocardiographic Signal. IEEE Signal Process. Lett. 2018, 25, 482–486. [Google Scholar]
- 97.Wu Z; Huang NE Ensemble Empirical Mode Decomposition: A Noise-Assisted Data Analysis Method. Adv. Adapt. Data Anal. 2009,1,1–41. [Google Scholar]
- 98.Malcangi M; Quan H; Vaini E; Lombardi P; Di Rienzo M Applying the EFuNN Evolving Paradigm to the Recognition of Artefactual Beats in Continuous Seismocardiogram Recordings. In Proceedings of the 2017 International Conference on Engineering Applications of Neural Networks, Athens, Greece, 25–27 August 2017; pp. 256–264. [Google Scholar]
- 99.Yang C; Tavassolian N Motion noise cancellation in seismocardiogram of ambulant subjects with dual sensors. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 5881–5884. [DOI] [PubMed] [Google Scholar]
- 100.Pouyan MB; Dowling S; De Marco T; Klein L; Inan OT Quantifying the Accuracy of Heart Failure Decompensation Classification Using Wearable Seismocardiography and Graph Mining Algorithms. J. Card. Fail. 2017, 23, S125–S126. [Google Scholar]
- 101.Yang C; Tavassolian N Motion noise cancellation in seismocardiographic monitoring of moving subjects. In Proceedings of the 2015 IEEE Biomedical Circuits and Systems Conference: Engineering for Healthy Minds and Able Bodies (BioCAS), Atlanta, GA, USA, 22–24 October 2015. [Google Scholar]
- 102.Yang C; Tavassolian N An Independent Component Analysis Approach to Motion Noise Cancellation of Cardio-Mechanical Signals. IEEE Trans. Biomed. Eng. 2018. [DOI] [PubMed] [Google Scholar]
- 103.Pandia K; Ravindran S; Cole R; Kovacs G; Giovangrandi L Motion artifact cancellation to obtain heart sounds from a single chest-worn accelerometer. In Proceedings of the 2010 ICASSP, IEEE International Conference on Acoustics, Speech and Signal Processing, Dallas, TX, USA, 14–19 March 2010; pp. 590–593. [Google Scholar]
- 104.Taebi A; Mansy HA Time-frequency Description of Vibrocardiographic Signals. In Proceedings of the 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Orlando, FL, USA, 16–20 August 2016. [Google Scholar]
- 105.Taebi A; Mansy HA Analysis of Seismocardiographic Signals Using Polynomial Chirplet Transform and Smoothed Pseudo Wigner-Ville Distribution. In 2017 IEEE Signal Processing in Medicine and Biology Symposium (SPMB); IEEE: Philadelphia, PA, USA, 2017; pp. 1–6. [Google Scholar]
- 106.Zakeri V; Tavakolian K Identification of respiratory phases using seismocardiogram: A machine learning approach. In Proceedings of the 2015 Computing in Cardiology Conference, Nice, France, 6–9 September 2015; Volume 42, pp. 305–308. [Google Scholar]
- 107.Poliac MO; Zanetty JM; Salerno D; Wilcox GL Seismocardiogram (SCG) interpretation using neural networks. In Proceedings of the Fourth Annual IEEE Symposium on Computer-Based Medical Systems, Baltimore, MD, USA, 12–14 May 1991; pp. 288–295. [Google Scholar]
- 108.Wahlstrom J; Skog I; Handel P; Khosrow-khavar F; Tavakolian K; Stein PK; Nehorai A A Hidden Markov Model for Seismocardiography. IEEE Trans. Biomed. Eng. 2017. [DOI] [PubMed] [Google Scholar]
- 109.Taebi A Characterization, Classification, and Genesis of Seismocardiographic Signals. Ph.D. Thesis, University of Central Florida, Orlando, FL, USA, 2018. [Google Scholar]
- 110.Malcangi M; Quan H; Vaini E; Lombardi P; Di Rienzo M Evolving fuzzy-neural paradigm applied to the recognition and removal of artefactual beats in continuous seismocardiogram recordings. Evol. Syst. 2018,1–10. [Google Scholar]
- 111.Neary JP; MacQuarrie DS; Jamnik V; Gledhill N; Gledhill S; Busse EFG Assessment of mechanical cardiac function in elite athletes. Open Sport. Med. J. 2011, 5, 26–37. [Google Scholar]
- 112.Giorgis L; Hernandez AI; Amblard A; Senhadji L; Cazeau S; Jauvert G; Donal E Analysis of cardiac micro-acceleration signals for the estimation of systolic and diastolic time intervals in cardiac resynchronization therapy. In Proceedings of the 2008 Computers in Cardiology, Bologna, Italy, 14–17 September 2008; Volume 35, pp. 393–396. [Google Scholar]
- 113.Etemadi M; Inan OT Wearable Ballistocardiogram and Seismocardiogram Systems for Health and Performance. J. Appl. Physiol. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Siecinski S; Kostka P Determining Heart Rate Beat-to-Beat from Smartphone Seismocardiograms: Preliminary Studies. In Proceedings of the Conference on Innovations in Biomedical Engineering, Zabrze, Poland, 19–20 October 2017; pp. 133–140. [Google Scholar]
- 115.Sahoo PK; Thakkar HK; Lee MY A cardiac early warning system with multi channel SCG and ECG monitoring for mobile health. Sensors 2017,17, 711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Mafi M Signal Processing Methods for Heart Rate Detection Using the Seismocardiogram. Ph.D. Thesis, University of Saskatchewan, Saskatoon, SK, Canada, 2016. [Google Scholar]
- 117.Casacanditella L; Cosoli G; Casaccia S; Tomasini EP; Scalise L Indirect measurement of the carotid arterial pressure from vibrocardiographic signal: Calibration of the waveform and comparison with photoplethysmographic signal. In Proceedings of the 2016 IEEE 38th Annual International Conference of the Engineering in Medicine and Biology Society (EMBC), Orlando, FL, USA, 16–20 August 2016; pp. 3568–3571. [DOI] [PubMed] [Google Scholar]
- 118.Casacanditella L; Cosoli G; Casaccia S; Rohrbaugh JW; Scalise L; Tomasini EP Characterization and calibration of the central arterial pressure waveform obtained from vibrocardiographic signal. In AIP Conference Proceedings; AIP Publishing: Melville, NY, USA, 2016; p. 050006. [DOI] [PubMed] [Google Scholar]
- 119.Verma AK; Zanetti J; Fazel-Rezai R; Tavakolian K Pulse Transit Time Derivation using Xiphoidal and Carotid Seismocardiograms. In Proceedings of the 2017 Design of Medical Devices Conference, Minneapolis, MN, USA, 10–13 April 2017; p. V001T01A010. [Google Scholar]
- 120.Peter L; Noury N; Cerny M A review of methods for non-invasive and continuous blood pressure monitoring: Pulse transit time method is promising? IRBM 2014, 35, 271–282. [Google Scholar]
- 121.Yang C; Tavassolian N Pulse transit time measurement using seismocardiogram and in-ear acoustic sensor. In Proceedings of the 2016 IEEE Biomedical Circuits and Systems Conference (BioCAS), Shanghai, China, 17–19 October 2016; pp. 188–191. [Google Scholar]
- 122.Carek AM; Conant J; Joshi A; Kang H; Inan OT SeismoWatch: Wearable Cuffless Blood Pressure Monitoring Using Pulse Transit Time. Proc. ACM Interact. Mob. Wearable Ubiquitous Technol. 2017, 1, 40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Lombard WP; Cope OM The duration of the systole of the left ventricle of man. Am. J. Physiol.-Legacy Content 1926, 77, 263–295. [Google Scholar]
- 124.Weissler AM; Harris WS; Schoenfeld CD Systolic Time Intervals in Heart Failure in Man. Circulation 1968, 37,149–159. [DOI] [PubMed] [Google Scholar]
- 125.Pouymiro IR; Cordova EV; Perez FEV Robust Detection of AO and IM Points in the Seismocardiogram Using CWT. IEEE Eat. Am. Trans. 2016,14,4468–4473. [Google Scholar]
- 126.Tavakolian K; Khosrow-Khavar F; Kajbafzadeh B; Marzencki M; Blaber AP; Kaminska B; Menon C Precordial acceleration signals improve the performance of diastolic timed vibrations. Med. Eng. Phys. 2013, 35,1133–1140. [DOI] [PubMed] [Google Scholar]
- 127.Benjamin EJ; Blaha MJ; Chiuve SE; Cushman M; Das SR; Deo R; de Ferranti SD; Floyd J; Fornage M; Gillespie G; et al. Heart disease and stroke statistics—2017 update: A report from the American Heart Association. Circulation 2017,135, e146. [DOI] [PMC free article] [PubMed] [Google Scholar]