Abstract
Background/purpose
The orthodontic aligner becomes popular worldwide in orthodontic therapy as an esthetic alternative to fixed labial braces. This study evaluated orthodontic tooth movement behavior using different aligner materials and attachment shapes for the movement of a single tooth.
Materials and methods
First bicuspid extracted resin typodont models were printed with a 3D printer. Three type of attachments, an ellipsoid shape (thick and thin) and a bar, were designed to fit the canine crown surface. Three types of aligner materials, Polyethylene Terephthalate enhanced with glycol (BIOSTAR) Polyethylene Terephthalate (BenQ), and Thermoplastic polyurethanes (TPU) were used to fabricate different aligners. The typodonts with aligners were sunk in a water bath to simulate canine distal movement in vivo. The canine crown, root movement, and long axis angle changes in each step were calculated and recorded. The data were analysed using a oneway ANOVA statistical method.
Results
Comparing the three aligners, the changes the long axis of the canine showed that the BENQ group had a smaller change in the long axis angle. The BENQ group canine involved bodily movement, but the canine movement of the BIOSTAR and TPU group involved tipping. Comparing the three attachments, the bar type attachment had more canine crown tipping in the BIOSTAR and TPU groups. The thick and thin ellipsoid-shaped attachments showed no statistical differences in tooth movement.
Conclusion
Attachment shape or size had little influence on the bodily movement of the tooth. A high modulus material may thus be suitable for clinical applications.
Keywords: Attachment, Aligner, Polyethylene terephthalate, Polyethylene terephthalate glycol, Thermoplastic polyurethanes
Introduction
Digital dentistry has arrived and now covers all fields of dental treatment. In particular, digitally aligned orthodontic treatment can be performed by orthodontists or general dentists. For the sake of the esthetic demands by patients, it has become popular worldwide in orthodontic therapy as an esthetic alternative to fixed labial braces.1 The aligner is highly accepted by most of the treatment population except patients with allergic reactions.
Numerous aligner companies have appeared in the market with similar treatment philosophies. The thermoplastic materials used by aligner manufacturers currently include polyethylene terephthalate glycol-modified (PET-G), polypropylene, polycarbonate (PC), thermoplastic polyurethanes (TPU), ethylene vinyl acetate, etc.2 Materials should be transparent, have a low degree of hardness, good elasticity, high resilience, and should be biocompatible and effective in terms of correcting tooth positions.3 The most important aspects are comfort and aesthetics.
Aligner performance is strongly influenced by the material construction. In the early hours of wearing an aligner, 50% of the initial stress value can be released. After 24 h, orthodontic loads on the aligner and changes in stress influence the programed tooth movement.4 Although aligner material compositions are different, the thickness of the plates range from 0.5 mm to 1. 5 mm.5 The thickness of the material can affect the biomechanical properties associated with tooth movement. Among the different aligner materials, thick materials deliver higher forces than those made from thin materials.6,7
It is difficult to correct tooth torque and rotation through the use of an aligner treatment.8 Attachments have been designed to apply force systems to teeth. The different attachment shapes are designed to enhance retention and facilitate complex orthodontic tooth movement. Optimized attachment shapes are increasing in complexity and are proving to be clinically better in terms of controlling tooth movements.9 The use of beveled attachments increases retention significantly, but ellipsoid attachments have been found to have no significant effects on retention.10
Thermoplastic materials on the market have very different mechanical characteristics. A 0.75-mm thick single-layer materials stress relaxation study, where assays showed that PETG materials led to a higher velocity of stress relaxation (62% in 24 h) as compared to other materials. The TPU material started with very low initial stress values. The present study was aimed toward examining the behavior of orthodontic aligners during the entire course of treatment. The variables include various aligner materials combined with various attachments of different shapes.
Material and methods
Testing model setup
The typodont had standard metal aligned teeth (mandible first molar, second bicuspid, canine, and later incisor) seeded in typodont wax (Fig. 1). The wax pattern including the teeth was impressed and transferred to a stone model. After the stone model was set, it was scanned with a 3 Shape Trios oral scanner (3 shape Co. Copenhagen, Denmark). The canine tooth was extracted using the 3Shape Ortho system software. A supporting box (w: 15.8 mm, L: 43.3 mm, H: 41.2 mm) was added as a typodont stent, which allowed the canine and first bicuspid spaces to be empty in the software (Fig. 2).
Attachment designs and teeth alignment
The attachments used in present study included a thin ellipsoid shape (w: 1 mm, L: 3 mm, H: 0.75 mm), a thick ellipsoid shape (w: 3 mm, L: 1 mm, H: 1 mm), and a bar shape (w: 1 mm, L: 5 mm, H: 1 mm) (Fig. 1). The attachments were added to the canine labial surface using the software. Then, the original testing resin models with and without attachments were printed out using the FORM 2 (Form Lab. Somerville, MA. USA) SLA 3D laser printer (Figure 3, Figure 4).
The mandible metal canines were placed into the above referenced 3D printed resin typodont and the empty spaces were filled with typodont wax (Fig. 3). A silicone impression material (Aquasil soft putty, Dentsply Co. Charlotte, NC, USA) was used to register the tooth position at the initial starting point. The register recording each test was set at the same start point. The first bicuspid was set to leave a 7 mm space. The typodont was scanned again. The data was manipulated using 3 Shape Ortho analyzer software to simulate the canine distal movement. The amount of canine distal movement at each step was set to 0.25 mm changes. A total of 28 sets of resin block working models were printed out using the Form 2 SLA printer.
The aligner fabrication
The aligners materials used in the present study included Duran (Scheu Dental GmbH Co., Iserlohn, Germany [abbreviation BIOSTAR], EasyDu (BenQ Co., Taipei, Taiwan [abbreviation BENQ]), and MaxFlex™ Coping sheet (MaxFlex Co., Taipei, Taiwan [abbreviation TPU]) (Table 1). The aligners were vacuum-thermoformed using the Biostar VI apparatus (Scheu-Dental, Germany). After separation from the working model, the gingival margins of the aligners were cut to the respective 2 mm gingival widths and then were trimmed and smoothed.
Table 1.
Brand | Composition | Thickness (mm) | Heating time (s) | Cooling time (s) |
---|---|---|---|---|
Duran (Scheu Dental, Iserlohn, Germany) | PETG | 0.75 | 30 | 20 |
EasyDU (BenQ Co, Taipei, Taiwan) | PET (PFb/PFc) | 0.75 | 20 | 20 |
MaxFlex™ coping sheet Maxflex Co, Taipei, Taiwan) |
TPU | 0.76 | 25 | 20 |
PETG:Polyethylene Terephthalate (PET) -glycol (G), test code: BIOSTAR
PET:Polyethylene Terephthalate, test code: BENQ.
TPU:Thermoplastic polyurethanes, test code: TPU.
Tooth movement simulation recording and movement calculation
The wearing aligner working model was placed into a 56 °C water bath for 90 s and then cooled in cold water for 90 s. The metal wire was prepared and embedded into a plastic sheet, where the metal wire was coordinated with the long axis of the tooth in the working model. The cooled working model was covered with a metal wire plastic sheet to take an X-ray. The radiation time was set 0.14 s. The above steps were repeated until the space was closed.
The tooth movement measurement points were recorded at the crown, root, and long axis of the canine on the X-ray film (Fig. 5). Each aligner was tested in triplicate. The data comparison included crown movement (mm), root movement (mm), and changes in the canine long axis (degrees). The data were analyzed using a one way ANOVA. All statistical tests were performed, with significance set at p < 0.05.
Results
Comparing the canine tooth movement without attachments
There were no differences in the canine crown movement in the three types of materials (Fig. 6A). There was a linear increase in the crown movement with the steps. The canine root movement in the BENQ group moved distally following the aligner steps. However, in the TPU and the BIOSTAR groups, the root stayed in the mesial region (Fig. 6B). The changes in the long axis of the canine indicated that the BENQ group had a smaller change in the long axis angle (1.2° ± 0.8°, p < 0.05). The TPU and BIOSTAR groups had greater changes in the long axis angle, but there was no statistically significant difference between them (p > 0.05, Fig. 6C).
Comparing canine tooth movement with and without attachments
There was no statistical difference in the canine crown movement between the attachment group and the group without attachments (p > 0.05, Fig. 7A). There was a linear increase in the crown movement with the aligner steps. There was no statistical difference between the bar and ellipsoid-shaped attachments in terms of crown movement (p > 0.05). The canine root movement in the groups with and without attachments demonstrated a statistically significant difference (p < 0.05). The thin ellipsoid-shaped attachment group showed less root mesial position than the thick type and bar type attachment groups after step 5 (p < 0.05 Fig. 7B).The changes in canine long axis angle showed that the bar type attachment group (27° ± 0.9°) had larger measurements than the group without attachments (26.3° ± 0.9°, p < 0.05). A comparison between the ellipsoid-shaped (thin: 25.5 ± 0.3°; thick: 26.1 ± 0.70) attachment group and the group without attachments showed no statistically significant difference in the long axis angle of the canine (p > 0.05, Fig. 7C).
There were no statistically significant differences in the canine crown movement for either the attachment groups or the groups without attachments (p > 0.05, Fig. 7D). There was a linear increase in the crown movement with aligner steps. The bar and ellipsoid-shaped attachments showed no statistically significant differences in terms of crown movement (p > 0.05). The canine root movement measurement with or without attachments group showed a statistically significant difference (p < 0.05). The bar type attachment group exhibited more root tip than the other group (p < 0.05). The thin ellipsoid-shaped attachment group showed less root tip than the other groups (p < 0.05, Fig. 7E). The changes in canine long axis angle showed that the bar type attachment group (27° ± 0.5°) had larger measurements than the group without attachments (24.7° ± 1°) (p < 0.05). A comparison of the TPU aligner for the ellipsoid-shaped (thin: 23.8 ± 1°; thick: 24.7 ± 0.80) attachment group and the group with no attachment showed no statistically significant difference (p > 0.05, Fig. 7F).
Comparison of the various aligner attachments
In the thin attachment comparison, the BIOSTAR and TPU groups showed no statistically significant difference in terms of the long axis movement (p > 0.05, Fig. 8A). In the thick attachment comparison, the BIOSTAR and TPU groups showed no statistically significant difference in terms of crown movement (p > 0.05), but there was a statistically significant difference in root movement after six steps (p < 0.05, Fig. 8B). The bar attachments applied on the BIOSTAR and TPU groups, in terms of both crown and root movement, showed no statistically significant differences (p > 0.05, Fig. 8C).
Discussion
The aligner tooth movement is accomplished by changing the aligner shape. Aligner discrepancies can create the pushing force necessary to move the tooth. This pushing force acts on the clinical crown but not through the tooth center resistance. Thus, the aligner tooth movement always induces movement of the crown tip.
The Young's moduli of the BIOSTAR, TPU and BENQ groups were 90, 99 and 134 (x106 Pa), respectively (Table 2), where a higher the modulus of the material represented the less material deformation. The present study showed that the canine crowns in the BIOSTAR and TPU groups underwent distal crown tip movement (Fig. 9A and B). However, the BENQ group canine crown showed bodily movement (Fig. 9C). It is thus suggested that a BENQ aligner with a higher modulus can wrap the tooth tightly and increase aligner retention.
Table 2.
Brand | BENQ | BIOSTAR | TPU |
---|---|---|---|
Young's modulus (E) × 106 Pa | 134 | 90 | 99 |
BIOSTAR: Duran (Scheu Dental GmbH Co., Iserlohn, Germany).
BenQ: EasyDu (BenQ Co., Taipei, Taiwan).
TPU: MaxFlex™ Coping sheet (MaxFlex Co., Taipei, Taiwan).
When the tooth is tipping, if the aligner material has a high degree of elasticity, the tooth can be rebound and upright. One may doubt the validity of the results of the present study, since the tests were conducted on a wax system and not under actual oral condition. However, when we reviewed the BIOSTAR and TPU results in this study, the canine underwent tipping movement and did not remain upright. This concurred with the clinical situation for the aligner (Fig. 6). Thus, we suggest that if the aligner modulus is high and can be retained in the mouth for a period of the, the tooth may undergo bodily movement.
To achieve better aligner retention, two factors need to be considered. One is the attachment shape and application, and the other is the aligner material selection. A well-defined geometric attachment is used to generate forces or movements, increasing the capacity of orthodontic aligners to move the tooth.10 When beveled attachments are compared with ellipsoid attachments or no attachment models, a hard aligner material requires a greater amount of force to remove an aligner from a cast.11 Retention depends on the material composition and does not necessarily correlate with the material thickness.12 The present study used the same thickness of aligner material to evaluate the attachment shape on tooth movement effects. The aligner bar-shaped attachment led to more crown tipping in the present study.
The BIOSTAR and TPU group aligner with and without were compared to determine attachment effects on tooth movement. The tooth movement results for the BIOSTAR and TPU groups showed no differences based on attachment (Fig. 7). This may have been because the Young's moduli of the aligners used in the BIOSTAR or TPU groups were low, so the aligner couldn't withstand the tooth tipping stress, which in turn led to tooth tipping.
Two shape attachments, ellipsoid and bar shape, were used in present study. Two contour thicknesses were used for the ellipsoid attachment, one thin and one thick. The ellipsoid attachment was placed on the distal occlusal site and on the mesial gingival site of the canine crown (Fig. 1). It was designed to press against the distal tip of the crown. The bar attachment was designed as a long rectangular shape, one half of the clinical length, to react against the distal movement force causing root mesial tipping. The difference between the ellipsoid and bar shape was two points of contact and one big area of contact to act against the aligner pushing force.
The canine showed crown distal tipping and root mesial toque in the BIOSTAR or TPU group aligners with both ellipsoid- and bar-shaped attachments (Figure 7, Figure 10). The BIOSTAR group showed that the bar type attachment caused a greater canine long axis angle than the ellipsoid attachment group. That is, more canine distal tipping occurred in the BIOSTAR bar attachment group. Using the thin or thick ellipsoid attachment on the BIOSTAR group canine led to tooth distal tipping but no statistically significant difference in the long axis angle of the canine (Figure 7, Figure 8, Figure 10). This indicated that ellipsoid- or bar-shaped attachment do not prevent crown tipping.
In the findings for the TPU group, the bar attachment exhibited larger measurements than the group with no attachment in terms on canine root movement and long axis angle measurements (Fig. 7). The canine root torque in the thin ellipsoid attachment group was less than that in thick ellipsoid attachment group (Figure 8, Figure 10). To prevent canine distal tipping, the ellipsoid attachment was preferable to the bar attachment in the present work. The attachment was unable to prevent canine tipping. This may have been due to the fact that the orthodontic force and moment delivered by the attachment may not meet his designed tooth movement.13
In an aligner system, the tooth tip often appears in the space close. To prevent side effects related to tooth tipping, in addition to using a highly elasticity aligner material and proper attachment design, order the aligner to upright by every four or five steps in space close stage may be another solution.
In conclusion, the purpose of the present study was to determine aligner behavior using an in vitro simulation. The attachments shape or size effects on tooth bodily movement were insignificant. The real behavior of orthodontic aligners during the course of treatment may require further research.
Declaration of competing interest
The authors have no conflicts of interest relevant to this article.
Acknowledgments
This work was supported by the Project of Chung Shan Medical University Hospital (CSH-2019-C-033), Taichung Taiwan.
References
- 1.Kravitz N.D., Kusnoto B., BeGole E., Obrez Am, Agran B. How well does Invisalign work? A prospective clinical study evaluating the efficacy of tooth movement with Invisalign. Am J Orthod Dentofacial Orthop. 2000;135:27–35. doi: 10.1016/j.ajodo.2007.05.018. [DOI] [PubMed] [Google Scholar]
- 2.Dupaix R.B., Boyce M.C. Finite strain behavior of poly (ethylene terephthalate) (PET) and poly(ethylene terephthalate)- glycol (PETG) Polymer. 2005;46:4827–4838. [Google Scholar]
- 3.Alexandros A., Youssef S.Al J., Spiros Z., Theodore E. Chemical and mechanical characteristics of contemporary thermoplastic orthodontic materials. Aust Orthod J. 2015;31:165–170. [PubMed] [Google Scholar]
- 4.Lombardo L., Martines E., Maxxanti V., Mollica F., Siciliani G. Stress relaxation properties of four orthodontic aligner materials: a 24-Hour in vitro study. Angle Orthod. 2017;87:11–18. doi: 10.2319/113015-813.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Guarneri M.P., Lombardo L., Gracco A., Siciliani G. vols. 15–24. Martina; Bologna, Italy: 2013. The state of the art of clean aligner technique. [Google Scholar]
- 6.Kohda N., Iijima M., Muguruma T., Brantley W.A., Ahluwalia K.S., Mizoguchi I. Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances. Angle Orthod. 2013;83:476–483. doi: 10.2319/052512-432.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kwon J.S., Lee Y.K., Lim B.S., Lim Y.K. Force delivery properties of thermoplastic orthodontic materials. Am J Orthod Dentofacial Orthop. 2008;133:228–234. doi: 10.1016/j.ajodo.2006.03.034. [DOI] [PubMed] [Google Scholar]
- 8.Ercoli F., Tepedino M., Parziale V., Luzi C. A comparative study of two different clear aligner systems. Prog Orthod. 2014;15:31–36. doi: 10.1186/s40510-014-0031-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Morton J., Derakhshan M., Kaza S., Li C. Design of the Invisalign system performance. Semin Orthod. 2017;23:3–11. [Google Scholar]
- 10.Barone S., Paoli A., Razionale A.V., Savignano R. Computational design and engineering of polymeric orthodontic aligners. Int J Numer Method Biomed Eng. 2017;33:e2839. doi: 10.1002/cnm.2839. [DOI] [PubMed] [Google Scholar]
- 11.Dasy H., Dasy A., Asatrian G., Rozsa N., Lee H.F., Kwak J.H. Effects of variable attachment shapes and aligner material on aligner retention. Angle Orthod. 2015;85:934–940. doi: 10.2319/091014-637.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Hahn W., Engelke B., Jung K. Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod. 2010;80:239–246. doi: 10.2319/033009-181.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kravitz N.D., Kusnoto B., Agran B., Viana G. Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign: a prospective clinical study. Angle Orthod. 2008;78:682–687. doi: 10.2319/0003-3219(2008)078[0682:IOAAIR]2.0.CO;2. [DOI] [PubMed] [Google Scholar]