Consultation on kidney stones, Copenhagen 2019: aspects of intracorporeal lithotripsy in flexible ureterorenoscopy

Søren Kissow Lildal 1, Kim Hovgaard Andreassen 2, Joyce Baard 3, Marianne Brehmer 4,5, Matthew Bultitude 6, Ylva Eriksson 4,5, Khurshid R Ghani 7, Helene Jung 8, Guido Kamphuis 3, Peter Kronenberg 9, Ben Turney 10, Olivier Traxer 11, Øyvind Ulvik 12, Palle Jörn Sloth Osther 8,
PMCID: PMC8217045  PMID: 33067728

Abstract

Purpose

To summarize current knowledge on intracorporeal laser lithotripsy in flexible ureterorenoscopy (fURS), regarding basics of laser lithotripsy, technical aspects, stone clearance, lithotripsy strategies, laser technologies, endoscopes, and safety.

Methods

A scoping review approach was applied to search literature in PubMed, EMBASE, and Web of Science. Consensus was reached through discussions at the Consultation on Kidney Stones held in September 2019 in Copenhagen, Denmark.

Results and conclusions

Lasers are widely used for lithotripsy during fURS. The Holmium laser is still the predominant technology, and specific settings for dusting and fragmenting have evolved, which has expanded the role of fURS in stone management. Pulse modulation can increase stone ablation efficacy, possibly by minimizing stone retropulsion. Thulium fibre laser was recently introduced, and this technology may improve laser lithotripsy efficiency. Small fibres give better irrigation, accessibility, and efficiency. To achieve optimal results, laser settings should be adjusted for the individual stone. There is no consensus whether the fragmentation and basketing strategy is preferable to the dusting strategy for increasing stone-free rate. On the contrary, different stone scenarios call for different lithotripsy approaches. Furthermore, for large stone burdens, all laser settings and lithotripsy strategies must be applied to achieve optimal results. Technology for removing dust from the kidney should be in focus in future research and development. Safety concerns about fURS laser lithotripsy include high intrarenal pressures and temperatures, and measures to reduce both those aspects must be taken to avoid complications. Technology to control these parameters should be targeted in further studies.

Keywords: Laser lithotripsy, Ureteroscopy, Kidney stones, Endoscopes, Safety, Technology

Introduction

Flexible ureterorenoscopy (fURS) has evolved as one of the major modalities for upper urinary tract stone management. This development is largely the result of continuing advancements in laser technology for intracorporeal lithotripsy, which is the focus of this review. Most daily clinical practices in the field of kidney stones, including lithotripsy techniques, are not supported by randomized trials and meta-analyses. This is not due to low research activity in this area, but rather to difficulties in designing meaningful trials that reflect daily clinical practice, because stone disease is so diverse. In this context, it is particularly important for practising experts to share their knowledge on clinical applications. The concept of the meeting “Consultation on Kidney Stones”, on which this review is based, was to create a forum for transfer and development of clinical expertise.

Methods

To evaluate lithotripsy in fURS regarding the balance between existing evidence, expert opinions, and safety and efficacy of new technological improvements, key opinion leaders in the field were invited to assess and discuss existing evidence at the 2-day meeting entitled “Consultation on Kidney Stones: Aspects of Intracorporeal Lithotripsy” held in Copenhagen, Denmark, in September 2019. The participating experts were assigned different topics and prepared presentations according to scoping reviews using PubMed, EMBASE, and Web of Science to search the literature. The first day of the meeting was open only to the experts, who individually presented their topics, which were then discussed by the whole group. The presentations were subsequently adjusted if necessary. The second day was an open meeting at which all presentations were given to an international audience, and this was followed by free discussions.

Basics of laser lithotripsy

Laser fibre size

The deflection range of a flexible ureteroscope decreases with increased size of the laser fibre inserted [13]. The size of the fibre also has an inverse effect on irrigation [1, 2]. An in vitro study of the ablation efficiency of laser fibres showed that large fibres (550 μm) produced wider fissures than small fibres (200 μm), but the fissures made by the small fibres were deeper [4]. Another investigation found that using a larger fibre created more retropulsion of the stone during lithotripsy [5]. Comparison of total fragmentation volume indicated little difference between fibres except at very low pulse energies (0.2 J), at which small fibres were more efficient [4]. These observations suggest that smaller fibres provide the following: better irrigation and thus improved visibility; better deflection and hence increased accessibility; and less retropulsion and thus higher ablation efficiency, and also an overall reduction in operating time [2, 6].

Laser fibre tips

Laser fibres can have a standard flat tip or a polished or ball-shaped tip. The ball-tip fibre is designed to reduce damaging friction forces generated within the working channel of the ureteroscope. It has been shown that this type of fibre can be safely passed through a deflected ureteroscope without causing liner perforation. The same advantage can be achieved by cleaving a standard fibre in such a manner that the coating protects the scope from the laser core [7]. Flat-tip fibres require greater insertion force at all angles and therefore can cause the ureteroscope liner to leak if it is deflected 45°or more [8].

An in vitro study comparing lithotripsy performance of different types of laser fibres found that the standard fibre functioned just as well as specially designed fibres [9]. Kronenberg et al. [10] noted that, compared to fibres stripped off the coating material, coated fibres achieved significantly higher ablation volumes, and there was no difference in performance between coated fibres cleaved with metal or ceramic scissors. However, Aldoukhi et al. [11] observed that single-use fibres and cleaved reusable fibres performed better than fibres with tips cut using Mayor scissors. The difference in efficiency between stripped and coated fibres may be due to the fibre itself being damaged during the stripping off the coating, leading to a dispersion of laser energy [10, 11]. Another downside of such stripping is that a fibre devoid of its coloured coating is difficult to discern in front of the ureteroscope. The above-mentioned factors have not been scientifically evaluated in vivo, and individual surgeons may have different preferences.

Laser settings

Various laser systems have been developed that offer different laser settings, such as high frequency and long-pulse duration. The total power (W) used is equal to the pulse energy (J) multiplied by the pulse frequency (Hz). In an in vitro study, a combination of low frequency and high pulse energy was more effective than the opposite combination with the same total power level, and pulse energy and ablation volume were linearly correlated with the width and depth of fissures observed in the stone material [4]. However, high pulse energy resulted in more stone retropulsion, larger stone fragments, and laser fibre burn-back, all leading to prolonged operating time [4, 12]. Lowering the pulse energy and increasing the frequency give rise to less retropulsion and fibre burn-back, which results in dusting of the stone. Regarding the pulse length, preliminary data from an in vitro study showed that short pulse settings led to significantly higher ablation volumes than long-pulse settings [13], whereas long-pulse lithotripsy was associated with less fibre burn-back, smaller fragments, and more dust. Other investigations have been unable to demonstrate a significant relationship between pulse duration and stone ablation [14, 15].

The Moses technology introduces a pulse-shaped modulation that optimizes energy delivery through water to the target stone. The Moses platform has two settings: Moses contact (MC) mode, for operation at close range (< 2 mm); Moses distance (MD) mode, for lithotripsy at a distance of 2 mm. In a preclinical study, the Moses technology resulted in more efficient laser lithotripsy with significantly reduced stone retropulsion [16]. In an automated in vitro dusting model, compared to long-pulse lithotripsy, Moses holmium:yttrium–aluminum–garnet (Ho:YAG) (Ho:YAG) laser technology provided greater ablation of soft stones when in contact with the stone surface [17]. In an investigation using a three-dimensional positioning system, researchers examined the impact of laser fibre working distance on fragmentation when altering pulse width or modulation [18], which showed that holmium laser lithotripsy was significantly affected by the fibre tip to stone working distance, with the greatest ablation volume obtained with the fibre in contact with the stone.

Considering optimal settings for the lithotriptor, different stone compositions require different settings to achieve the desired effect, and the desired effect may vary in individual situations. The recommendation is to start with a test setting of low pulse energy and low frequency to initially determine how the stone reacts, and thereafter adjust settings accordingly to improve efficiency of the lithotripsy.

Technical aspects of laser lithotripsy and stone clearance during fURS

Various methods for laser lithotripsy have been described, and combining them can lead to more efficient ablation and clearance of stones with different characteristics [19]. The techniques described include dusting (dancing/painting), chipping, fragmentation, popcorning, dustmenting, and popdusting, and all these strategies can be combined in different ways to achieve the desired outcome. Dancing/painting is done by moving the laser fibre from side to side over the surface of the stone, using dusting settings. Chipping entails lasering off small pieces from the edges of the stone leaving small fragments for spontaneous passage, whereas fragmenting involves cutting a stone into larger fragments for subsequent basket removal. The fragmentation and basketing strategy use a setting of low frequency (4–10 Hz) and high pulse energy (0.6–2.0 J); dusting uses a setting of high frequency (15–80 Hz) and low pulse energy (0.2–0.5 J); popcorning stones requires settings of moderate-to-high frequencies (10–20 [40] Hz) and moderate pulse energy (1–1.5 J)[2023].

Popcorning refers to a non-contact technique in which the laser energy whirls around fragments that come in contact with the laser and breaks them into smaller pieces. Chawla et al. [20] conducted an in vitro investigation of popcorning with different laser settings and observed that 1.5 J and 40 Hz produced the greatest mean decrease in stone burden per amount of time used; however, such a high-power level cannot be recommended in vivo. In that study, settings of 1.0 J and 20 Hz were most efficient with regard to mean stone weight loss per total amount of energy used. Emiliani et al. [24] suggested that a good compromise for popcorning would be using a long pulse of 1.5 J and 20 Hz with a 270 μm laser fibre and taking as much time as possible (> 4 min) to produce clinically insignificant fragments. In another in vitro model, Aldoukhi et al. found that the popcorn te

chnique was more effective when the laser fibre was positioned in contact with the stone as compared to at a distance of 2 mm from the stone, and when performed in a small (11 mm) rather than a larger calyceal model [25]. These in vitro studies do not replicate the clinical environment, and, working in a limited space, surgeons should always aim to choose a safe power level based on the irrigation flow applied and other factors, including outflow (access sheath) and temperature of the irrigation fluid.

Humphreys et al. [26] conducted a non-randomized prospective clinical comparative trial considering dusting versus fragmenting, and they found no statistically significant difference in stone-free rate (SFR) and no difference in complication rates between the two patient groups. However, although stones were significantly larger in the dusting arm (96 vs 63 mm2, p < 0.001) the mean operative time was significantly shorter in that group (36 vs 67 min, p < 0.001). This agrees with clinical data showing that, for large stones, it is easier to control the complete stone burden during dusting, whereas fragmenting may result in fragments moving into multiple different calyces in which they must be dealt with, potentially prolonging OR time and reducing SFR.

Ureteroscopic lasering of lower pole stones can be challenging. The anatomy of the lower pole must be considered, because a long infundibular length, a narrow infundibular width, and a steep infundibulo-pelvic angle may have a negative impact on stone clearance, especially if more than one of these aspects is unfavourable [2730]. Using a basket/grasper to move the stone to another location can be a viable alternative that can facilitate lithotripsy, reduce scope damage, and aid fragment passage [31, 32]. Schuster et al. [33] compared SFR in patients with lower pole stones treated in situ and subjects with relocation of the stone before lithotripsy. For stones > 1 cm in size, complete clearance was obtained in 100% of cases by relocation versus 29% of cases in situ (p = 0.005); the difference between the two groups was not significant for stones of < 1 cm. Placing the patient in Trendelenburg position during surgery can facilitate stone relocation to the upper pole from other sites by creating a dependent upper pole; two randomized trials have demonstrated that this method results in shorter OR time, better SFR, less flexible ureteroscope manipulation, and less stone migration into the lower pole [34, 35]. If the stone cannot be relocated due to size or anatomy, or if lithotripsy in situ is preferred, a ball-tip laser fibre can be safely introduced through a deflected scope.

The 2020 EAU Urolithiasis Guidelines recommend that only baskets made of Nitinol be used for fURS. A range of different baskets are available, such as triangular baskets, 4-wire baskets, smaller baskets for narrow locations, and larger baskets for strength. Baskets can also vary regarding penetration force, radial dilation force, and opening dynamics, and add different degrees of resistance to scope deflection [36]. Bach et al. [37] tested various sizes of baskets in five different endoscopes and found that, in contrast to laser fibres, the size of the basket did not influence the deflection of the scope, which made relocation of a stone easier than in in situ lithotripsy. However, the size of the basket had an inverse effect on irrigation flow. Conversely, Patel et al. [36] showed some limitation of deflection, which was greater for a 2.2 Fr compared to a 1.5 Fr basket. In practice, these differences will vary depending on the type of ureteroscope employed and how long it has been in use, and the specific basket that is chosen.

Methods for the evacuation of small residual fragments include the glue-clot technique and irrigation/suction systems. In the glue-clot technique, autologous venous blood is injected into a calyx, where it agglutinates with stone fragments to form a clot that can be extracted with a basket [38]. In a randomized pilot study of 47 patients, an automated irrigation/suction pump system was tested and found to reduce mean operating time by 35% (p = 0.04) and to increase SFR from 69 to 92% (p = 0.048) compared with the standard pressurized technique [39].

Comparing clinical outcome of different laser lithotripsy strategies

Comparing fragmenting and dusting with regard to clinical outcomes has proven difficult, because the objective metrics for such comparison vary. The studies performed have differed regarding rates of pre-stenting and staged procedures, and ways of reporting complication grades, emergency room visits, intensive care admission, hospitalization, and re-intervention. Reports concerning follow-up also differ with regard to definitions of stone free, imaging modalities, and timing of the follow-up and repeat intervention rate. Stone free has been defined as zero fragment, < 2 mm, or > 4 mm residual fragment. Imaging modality for follow-up varies between computed tomography (CT), kidney–ureter–bladder radiograph (KUB), ultrasound (US), or fluoroscopy at operation, and a large proportion of patients apparently have no follow-up imaging at all [40]. An assessment of administrative data in the United States showed that 39% of patients had no imaging at 12 months post-surgery[41].

A retrospective study demonstrated that fragments > 4 mm in size were associated with significantly higher rates of stone growth leading to complications and need for re-intervention, whereas fragments of > 2 mm were likely to grow, but were not associated with complications or re-intervention [42]. Another retrospective investigation reported that the cumulative repeat surgery rate at 5 years after fURS was proportional to the size of residual fragments on post-operative CT [43]: with no visible residual fragments, fragments < 4 mm, or fragments > 4 mm, the rates were 3.5%, 8.2%, and 46.2%, respectively. CT has been reported to have a sensitivity of 95–100% for residual fragments [44].

Eleven studies comparing spontaneous passage and fragmentation with active retrieval [45] were identified, but most of them could not be compared with each other due to differences in imaging modalities and definitions of SFR. Only three comparative, non-randomized studies of the various disintegration strategies in the treatment of renal stones were pinpointed (Table 1) [26, 46, 47]. Two of those assessments used CT for follow-up, whereas the third used KUB and US. Only one of those three studies reported the power settings for dusting and fragmentation, respectively.

Table 1.

Comparative studies regarding SFR with spontaneous passage vs active retrieval during lithotripsy

Cohort Lee et al. 2016 Humphreys et al. 2018 El-Nahas et al. 2019
Spontaneous Retrieval Spontaneous Retrieval Spontaneous Retrieval
Study design Retrospective Retrospective Prospective Prospective Retrospective Retrospective
Study period 2010–2015 2010–2015 2013–2016 2013–2016 2015–2017 2015–2017
Cases, n 76 172 68 82 51 56
Mean stone size (SD) 11 mm (± 5.2) 11 mm (± 4.8) 11 mm (± 4.3) 8.8 mm (± 3.5) NS NS
Laser power (setting) NS NS 30–100 W (NS) 30–100 W (NS) 20–60 W (0.3 J/20 Hz) 20–60 W (1.0 J/10 Hz)
OR time 82 min 83 min 36 min 67 min 76 min 91 min
Stenting rate 100% 100% 100% 100% 100% 100%
Complications 11%* 11%* 13.2%* 19.5%** 8%* 9%**
Definition of SFR < 3 mm < 3 mm No fragment No fragment No fragment and < 4 mm No fragment and < 4 mm
Imaging used CT CT KUB, US KUB, US CT CT
Imaging timing 4 weeks 4 weeks 4–6 weeks 4–6 weeks 2 months 2 months
Stone free rate 87% 89% 58% 74% 59% and 86% 79% and 89%

SD standard deviation, OR operating room, SFR stone-free rate, CT computed tomography, KUB kidney–ureter–bladder radiograph, US ultrasound

In a randomized, double-blinded clinical trial comparing regular and Moses modes of Ho:YAG laser lithotripsy, Ibrahim et al. showed that the Moses technology was associated with significantly lower fragmentation/pulverization (stone dusting) and procedural times, which the authors explained by less retropulsion of stones during Moses mode lithotripsy [48].

Conclusive evidence for comparing spontaneous passage and active retrieval strategies is limited, and it is difficult to compare existing studies, because they vary regarding disintegration strategy, SFR criteria, and time and modality of imaging at follow-up. Moreover, the investigations have not systematically reported stone density and composition and post-operative complications.

Endoscopes and laser lithotripsy

Ureteroscopes vary with respect to location of working channel (Fig. 1), deflection ability and irrigation flow with laser fibre inserted, distance of laser tip from the tip of the endoscope, and optical imaging quality and illumination of the surgical field. All of the listed features have a potential impact on laser lithotripsy, although evidence for this is based mainly on in vitro studies and expert opinions, and very little has been published regarding effects on clinical outcome.

Fig. 1.

Fig. 1

Different positions of working channel depending on manufacturer and model of ureteroscope

The position of the working channel of the endoscope can influence irrigation flow and thus create differences in the flushing and movement of stone and tissue. The position also dictates location of the laser fibre/basket inserted through the channel and thereby affects the ability to reach stones in certain positions within the kidney [49].

In vitro comparison of single-use and reusable flexible ureteroscopes showed that the latter had bettervision characteristics [50]. It appears that scopes differ regarding occurrence of light flashes during laser lithotripsy, over-illumination of the field, and the speed at which light intensity can be regulated. These phenomena may occur due to automatic light intensity regulation being affected by laser energy flashes or by stone fragments or dust that is flushed around by irrigation and laser pulse pressure waves. The cited in vitro comparison also demonstrated that the flow varied between different flexible ureteroscopes equipped with the same type of laser fibre [50]. The problem of working channel damage can be almost completely overcome by employing single-use flexible ureteroscopes.

Safety aspects of laser lithotripsy during fURS

Direct contact of the laser can cause local burns on the urothelium. Furthermore, high-flow irrigation for optimal visualization and cooling leads to increased intraluminal pressure, although such irrigation is necessary.

It should also be noted that incorrect handling of lasers may damage the ureteroscope and even harm the surgeon and OR personnel. A safety distance between the tip of the fibre and the tip of the scope should be maintained to minimize the risk of damaging endoscopes. Talso et al. [51] observed that the length of the fibre projecting from ureteroscopes was 1–2 mm at first appearance in the camera image, whereas it was 3–4 mm when reaching one-fourth of the screen, and in that position, the laser bubble was never in contact with the tip of the ureteroscope and was therefore protected from damage.

The incidence of ureteral perforation during ureterorenoscopy has been reported to be 0.4–6.3%, and the risk factors identified included size of ureteroscope, prolonged operation time, surgeon’s lack of experience, stone characteristics, lack of pre-stenting, and high laser energy [52, 53]. Possible complications of perforation are extravasation, haematoma, sepsis, pain, obstruction, stricture, and loss of kidney function [5255]. Use of laser during intracorporeal lithotripsy has been observed to be an independent risk factor, with a 3.6% risk of Post-ureteroscopic Lesion Scale (PULS; [56] grade 1 lesions and 3.1% risk of PULS grade 2 lesions. To avoid ureteral perforation, it is recommended that optimal visualization be secured by irrigation and that a safety distance of at least 1 mm be maintained between laser tip and urothelium.

During laser lithotripsy, visualization can be impaired by stone dust and debris. Increasing irrigation to solve this problem unequivocally raises the intrarenal pressure, potentially leading to tubular, venous, and lymphatic backflow. Baseline intrarenal pressure is approximately 10 mmHg, and the threshold for backflow is 30–35 mmHg. Intrarenal pressure levels during fURS and holmium laser usage have been noted to be as high as 50–350 mmHg [57, 58]. Intrarenal pressures above 30 mmHg for more than 10 min have been shown to significantly increase the incidence of septicaemia in percutaneous nephrolithotomy [59]. Although several studies have suggested that complications such as sepsis and post-operative pain leading to longer hospitalization are related to high intrarenal pressure, no definitive causal data exist [6063]. Intrarenal pressure during fURS may be reduced by the use of a ureteral access sheath (UAS) [64]. If a UAS is to be applied, it must have a larger diameter than the scope, which make the UAS a double-edged sword: on one hand, it reduces intrarenal pressure and potential septic complications, and on the other hand, it may increase the risk of strain-induced lesions to the ureter [65]. Therefore, it is important to use the smallest UAS compatible with the ureteroscope chosen to ensure both safe placement without significant friction and sufficient outflow, and thereby achieve both low intrarenal pressure and sufficient flow for visualization during lithotripsy [66].

High intrarenal temperatures have been observed during laser lithotripsy, especially at high total power settings (> 30 W) [67]. High temperatures can lead to thermal ablation of kidney tissue, resulting in permanent kidney damage. Overall, such tissue damage is dependent on temperature, time, and blood flow [68]. In the ureter, the threshold for tissue damage has been shown to be around 43 °C for 120 min, which is equivalent to approximately 50 °C for 56 s or 56 °C for 0.9 s [6971]. In an in vitro model, Wollin et al. achieved temperatures of > 43 °C at all laser settings within 1 min without irrigation, and the maximum temperature reached was 100 °C at fragmentation settings [69]. Aldoukhi et al. recorded temperatures in a pig kidney during high-power holmium laser lithotripsy at 0.5 J and 80 Hz and different irrigation flows, and found that only high irrigation flow (38–40 ml/min) could keep the caliceal temperature at ≤ 43 °C [72]. To reduce heating, it is important to have secured continuous flow irrigation and consider use of room temperature or even cooled irrigation fluid. Lasering should be carried out intermittently to allow cooling in low-flow areas.

Comparison of different laser technologies for intracorporeal lithotripsy

The past and current standard for laser lithotripsy is the Ho:YAG laser, which has revolutionized retrograde stone management. Introduction of high-power Ho:YAG laser systems that utilize laser settings of high frequency and low pulse energy (HiFr-LoPE), varying pulse lengths, and Moses technology for reduction of retropulsion and improvement of fragmentation have expanded the surgeon’s ability to ablate urinary tract stones and thereby opened a new era for the dusting technique. Reducing the need for fragment retrieval and the potential hazards of this technique (UAS usage and multiple scope passages) may be advantageous [73]. Recently, a new laser technology for stone management has emerged—the thulium fibre laser (TFL). The TFL emits light with a wavelength of 1940 nm, compared to 2100 nm with Ho:YAG [7479]. Comparing laser setting options, TFL offers frequencies of 5–2200 Hz, pulse energies of 0.025–6 J, and pulse durations of 200–12,000 microsec [76, 8083], whereas Ho:YAG offers 5–100 Hz, 0.2–6 J, and 50–1300 microsec [81, 84]. These qualities of the TFL laser give this technology the potential to produce 3–4 times more dust compared to high-power Ho-YAG lasers [78, 85, 86]. Furthermore, due to the novel laser beam generation by laser diodes in TFL, smaller fibres (50 µm) can be utilized, which may result in higher energy intensity at the tip of the fibre, better scope deflection, and better irrigation, all of which can potentially increase stone ablation volume at comparable settings [74, 80, 81]. Initial reports of TFL performance in the clinical setting are promising but still limited [87], and further evaluation, including consideration of safety aspects (temperature) and ability of the laser to embrace all lithotripsy strategies, will be necessary to define the role of TFL. To date, no clinical comparative studies of holmium and thulium fibre lasers have been conducted. Nevertheless, both the HiFr-LoPE Ho:YAG and the new TFL technology show promise in expanding the role of flexible ureteroscopic stone management by approaching larger stones and more complex stone scenarios.

Conclusions

Lasers have revolutionized the scenario of intracorporeal lithotripsy during fURS. Ho:YAG is still the predominant laser for stone management, and, with this technology, specific settings for dusting and fragmenting have evolved, which have expanded the role of fURS in endoscopic stone management. Recently, the TFL was introduced, and this new laser source may prove to further increase the efficiency of laser lithotripsy. There is still debate as to whether the fragmentation and basketing strategy are preferable to the dusting strategy for increasing SFR. However, one does not exclude the other, because different stone scenarios call for different lithotripsy approaches, and, for large stone burdens, all laser lithotripsy settings (dusting, fragmenting, and popcorning) may often be necessary for optimizing the final result. Technology for removing dust from the kidney should be a focus for future research and development. Safety aspects of fURS laser lithotripsy include high intrarenal pressures and temperatures. Inasmuch as measures to reduce both pressure and temperature are mandatory to avoid sepsis, bleeding, and nephron loss, technology to control these parameters should be targeted in further investigations.

Acknowledgements

We are grateful to Boston Scientific, Olympus, Coloplast, Cook Medical, EMS, Lumenis, Richard Wolf, Karl Storz Endoscope, Apsu Medical, BL Medical, Opiva Nordic, and Rocamed for supporting the Copenhagen Consultation on Kidney Stones meeting 2019. We thank chief assistant Katrine Lindholtz for outstanding administrative management, Stockholm Data Design AB for managing the website, and Patricia Ödman for language revision.

Author contributions

SKL: data analysis and manuscript writing. KA: data analysis and manuscript editing. JB: data analysis and manuscript editing. MBr: project co-ordination, presentation, manuscript editing. MBu: data collection and analysis, presentation, and manuscript editing. YE: data analysis and manuscript editing. KG: data collection and analysis, presentation, and manuscript editing. HJ: data collection and analysis, presentation, and manuscript editing. GK: data collection and analysis, presentation, and manuscript editing. PK: data collection and analysis, presentation, and manuscript editing. BT: data analysis and manuscript editing. OT: data collection and analysis, presentation, and manuscript editing. ØU: data analysis and manuscript editing. PO: project co-ordination, data analysis, presentation, manuscript writing, and editing.

Compliance with ethical standards

Conflict of interest

Søren K. Lildal: no conflicts of interest. Kim Andreassen: consultant for AMBU. Joyce Baard: no conflicts of interest. Marianne Brehmer: no conflicts of interest. Matthew Bultitude: consultant for Boston Scientific. Ylva Eriksson: no conflicts of interest. Khurshid Ghani: consultant for Boston Scientific, Lumenis, and Olympus; Grant from Boston Scientific. Helene Jung: no conflicts of interest. Guido Kamphuis: consultant for Boston Scientific and Olympus. Peter Kronenberg: consultant for Olympus. Ben Turney: no conflicts of interest. Olivier Traxer: consultant for Boston Scientific, Olympus, Rocamed, and Coloplast. Øyvind Ulvik: consultant for Olympus. Palle J S Osther: consultant for Olympus, Boston Scientific, Bonvisi AB, and Coloplast.

Informed consent

Informed consent is not relevant.

Human and animal rights

This is a review article involving no studies of human participants and/or animals.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Kronenberg P, Traxer O. Update on lasers in urology 2014: current assessment on holmium:yttrium-aluminum-garnet (Ho:YAG) laser lithotripter settings and laser fibers. World J Urol. 2015;33(4):463–469. doi: 10.1007/s00345-014-1395-1. [DOI] [PubMed] [Google Scholar]
  • 2.Kronenberg P, Traxer O. The truth about laser fiber diameters. Urology. 2014;84(6):1301–1307. doi: 10.1016/j.urology.2014.08.017. [DOI] [PubMed] [Google Scholar]
  • 3.Pasqui F, Dubosq F, Tchala K, Tligui M, Gattegno B, Thibault P, et al. Impact on active scope deflection and irrigation flow of all endoscopic working tools during flexible ureteroscopy. Eur Urol. 2004;45(1):58–64. doi: 10.1016/j.eururo.2003.08.013. [DOI] [PubMed] [Google Scholar]
  • 4.Kronenberg P, Traxer O. In vitro fragmentation efficiency of holmium: yttrium-aluminum-garnet (YAG) laser lithotripsy—a comprehensive study encompassing different frequencies, pulse energies, total power levels and laser fibre diameters. BJU Int. 2014;114(2):261–267. doi: 10.1111/bju.12567. [DOI] [PubMed] [Google Scholar]
  • 5.Kronenberg P, Traxer O. V1718 laser fibers, pulse energy and retropulsion—what we can see and what we can’t. J Urol. 2013;189(4s):e707. [Google Scholar]
  • 6.Kronenberg P, Somani B. Advances in lasers for the treatment of stones-a systematic review. Curr Urol Rep. 2018;19(6):45. doi: 10.1007/s11934-018-0807-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Baghdadi M, Emiliani E, Talso M, Servián P, Barreiro A, Orosa A, et al. Comparison of laser fiber passage in ureteroscopic maximum deflection and their influence on deflection and irrigation: Do we really need the ball tip concept? World J Urol. 2017;35(2):313–318. doi: 10.1007/s00345-016-1873-8. [DOI] [PubMed] [Google Scholar]
  • 8.Carlos EC, Li J, Young BJ, Radvak D, Wollin DA, Winship BB, et al. Let’s get to the point: comparing insertion characteristics and scope damage of flat-tip and ball-tip holmium laser fibers. J Endourol. 2019;33(1):22–26. doi: 10.1089/end.2018.0229. [DOI] [PubMed] [Google Scholar]
  • 9.Kronenberg P, Traxer O. Lithotripsy performance of specially designed laser fiber tips. J Urol. 2016;195(5):1606–1612. doi: 10.1016/j.juro.2015.10.135. [DOI] [PubMed] [Google Scholar]
  • 10.Kronenberg P, Traxer O. Are we all doing it wrong? Influence of stripping and cleaving methods of laser fibers on laser lithotripsy performance. J Urol. 2015;193(3):1030–1035. doi: 10.1016/j.juro.2014.07.110. [DOI] [PubMed] [Google Scholar]
  • 11.Aldoukhi AH, Knudsen BE, Black KM, Hall TL, Roberts WW, Ghani KR. Are we cutting ourselves short? Laser lithotripsy performance based on differences in fiber-tip preparation. Urology. 2019;134(1):79–83. doi: 10.1016/j.urology.2019.08.027. [DOI] [PubMed] [Google Scholar]
  • 12.Sea J, Jonat LM, Chew BH, Qiu J, Wang B, Hoopman J, et al. Optimal power settings for holmium:YAG lithotripsy. J Urol. 2012;187(3):914–919. doi: 10.1016/j.juro.2011.10.147. [DOI] [PubMed] [Google Scholar]
  • 13.Kronenberg P, Traxer O. Ultra-short, short, medium and long-pulse laser lithotripsy performance. Eur Urol Suppl. 2016;15(3):e1035. [Google Scholar]
  • 14.Sroka R, Pongratz T, Scheib G, Khoder W, Stief CG, Herrmann T, et al. Impact of pulse duration on Ho:YAG laser lithotripsy: treatment aspects on the single-pulse level. World J Urol. 2015;33(4):479–485. doi: 10.1007/s00345-015-1504-9. [DOI] [PubMed] [Google Scholar]
  • 15.Bader MJ, Pongratz T, Khoder W, Stief CG, Herrmann T, Nagele U, et al. Impact of pulse duration on Ho:YAG laser lithotripsy: fragmentation and dusting performance. World J Urol. 2015;33(4):471–477. doi: 10.1007/s00345-014-1429-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Elhilali MM, Badaan S, Ibrahim A, Andonian S. Use of the Moses technology to improve holmium laser lithotripsy outcomes: a preclinical study. J Endourol. 2017;31(6):598–604. doi: 10.1089/end.2017.0050. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Winship B, Wollin D, Carlos E, Li J, Peters C, Simmons WN, et al. Dusting efficiency of the Moses holmium laser: an automated in vitro assessment. J Endourol. 2018;32(12):1131–1135. doi: 10.1089/end.2018.0660. [DOI] [PubMed] [Google Scholar]
  • 18.Aldoukhi AH, Roberts WW, Hall TL, Ghani KR. Watch your distance: the role of laser fiber working distance on fragmentation when altering pulse width or modulation. J Endourol. 2019;33(2):120–126. doi: 10.1089/end.2018.0572. [DOI] [PubMed] [Google Scholar]
  • 19.Hecht SL, Wolf JSJ. Techniques for holmium laser lithotripsy of intrarenal calculi. Urology. 2013;81(2):442–445. doi: 10.1016/j.urology.2012.11.021. [DOI] [PubMed] [Google Scholar]
  • 20.Chawla SN, Chang MF, Chang A, Lenoir J, Bagley DH. Effectiveness of high-frequency holmium:YAG laser stone fragmentation: the “popcorn effect”. J Endourol. 2008;22(4):645–650. doi: 10.1089/end.2007.9843. [DOI] [PubMed] [Google Scholar]
  • 21.Weiss B, Shah O. Evaluation of dusting versus basketing—can new technologies improve stone-free rates? Nat Rev Urol. 2016;13(12):726–733. doi: 10.1038/nrurol.2016.172. [DOI] [PubMed] [Google Scholar]
  • 22.Doizi S, Keller EX, De Coninck V, Traxer O. Dusting technique for lithotripsy: what does it mean? Nat Rev Urol. 2018;15(11):653–654. doi: 10.1038/s41585-018-0042-9. [DOI] [PubMed] [Google Scholar]
  • 23.Aldoukhi AH, Roberts WW, Hall TL, Ghani KR. Holmium laser lithotripsy in the new stone age: dust or bust? Front Surg. 2017;4:1–6. doi: 10.3389/fsurg.2017.00057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Emiliani E, Talso M, Cho S-Y, Baghdadi M, Mahmoud S, Pinheiro H, et al. Optimal settings for the noncontact holmium:YAG stone fragmentation popcorn technique. J Urol. 2017;198(3):702–706. doi: 10.1016/j.juro.2017.02.3371. [DOI] [PubMed] [Google Scholar]
  • 25.Aldoukhi AH, Roberts WW, Hall TL, Teichman JMH, Ghani KR. Understanding the popcorn effect during holmium laser lithotripsy for dusting. Urology. 2018;122:52–57. doi: 10.1016/j.urology.2018.08.031. [DOI] [PubMed] [Google Scholar]
  • 26.Humphreys MR, Shah OD, Monga M, Chang Y-H, Krambeck AE, Sur RL, et al. Dusting versus basketing during ureteroscopy-which technique is more efficacious? A prospective multicenter trial from the EDGE research consortium. J Urol. 2018;199(5):1272–1276. doi: 10.1016/j.juro.2017.11.126. [DOI] [PubMed] [Google Scholar]
  • 27.Elbahnasy AM, Shalhav AL, Hoenig DM, Elashry OM, Smith DS, McDougall EM, et al. Lower caliceal stone clearance after shock wave lithotripsy or ureteroscopy: the impact of lower pole radiographic anatomy. J Urol. 1998;159(3):676–682. [PubMed] [Google Scholar]
  • 28.Inoue T, Murota T, Okada S, Hamamoto S, Muguruma K, Kinoshita H, et al. Influence of pelvicaliceal anatomy on stone clearance after flexible ureteroscopy and holmium laser lithotripsy for large renal stones. J Endourol. 2015;29(9):998–1005. doi: 10.1089/end.2015.0071. [DOI] [PubMed] [Google Scholar]
  • 29.Jessen JP, Honeck P, Knoll T, Wendt-Nordahl G. Flexible ureterorenoscopy for lower pole stones: influence of the collecting system’s anatomy. J Endourol. 2014;28(2):146–151. doi: 10.1089/end.2013.0401. [DOI] [PubMed] [Google Scholar]
  • 30.Danilovic A, Rocha BA, Torricelli FCM, Marchini GS, Batagello C, Vicentini FC, et al. Size is not everything that matters: preoperative CT predictors of stone free after RIRS. Urology. 2019;132:63–68. doi: 10.1016/j.urology.2019.07.006. [DOI] [PubMed] [Google Scholar]
  • 31.Kourambas J, Delvecchio FC, Munver R, Preminger GM. Nitinol stone retrieval-assisted ureteroscopic management of lower pole renal calculi. Urology. 2000;56(6):935–939. doi: 10.1016/s0090-4295(00)00821-9. [DOI] [PubMed] [Google Scholar]
  • 32.Auge BK, Dahm P, Wu NZ, Preminger GM. Ureteroscopic management of lower-pole renal calculi: technique of calculus displacement. J Endourol. 2001;15(8):835–838. doi: 10.1089/089277901753205852. [DOI] [PubMed] [Google Scholar]
  • 33.Schuster TG, Hollenbeck BK, Faerber GJ, Wolf JSJ. Ureteroscopic treatment of lower pole calculi: comparison of lithotripsy in situ and after displacement. J Urol. 2002;168(1):43–45. [PubMed] [Google Scholar]
  • 34.Pan J, Xue W, Xia L, Zhong H, Zhu Y, Du Z, et al. Ureteroscopic lithotripsy in Trendelenburg position for proximal ureteral calculi: a prospective, randomized, comparative study. Int Urol Nephrol. 2014;46(10):1895–1901. doi: 10.1007/s11255-014-0732-z. [DOI] [PubMed] [Google Scholar]
  • 35.Zhou R, Han C, Hao L, Chen B, Zang G, Fan T, et al. Ureteroscopic lithotripsy in the Trendelenburg position for extracting obstructive upper ureteral obstruction stones: a prospective, randomized, comparative trial. Scand J Urol. 2018;52(4):291–295. doi: 10.1080/21681805.2018.1492966. [DOI] [PubMed] [Google Scholar]
  • 36.Patel N, Akhavein A, Hinck B, Jain R, Monga M. Tipless nitinol stone baskets: comparison of penetration force, radial dilation force, opening dynamics, and deflection. Urology. 2017;103:256–260. doi: 10.1016/j.urology.2017.01.010. [DOI] [PubMed] [Google Scholar]
  • 37.Bach T, Geavlete B, Herrmann TRW, Gross AJ. Working tools in flexible ureterorenoscopy—influence on flow and deflection: what does matter? J Endourol. 2008;22(8):1639–1643. doi: 10.1089/end.2008.0184. [DOI] [PubMed] [Google Scholar]
  • 38.Cloutier J, Cordeiro ER, Kamphuis GM, Villa L, Letendre J, de la Rosette JJ, et al. The glue-clot technique: a new technique description for small calyceal stone fragments removal. Urolithiasis. 2014;42(5):441–444. doi: 10.1007/s00240-014-0679-7. [DOI] [PubMed] [Google Scholar]
  • 39.Lechevallier E, Luciani M, Nahon O, Lay F, Coulange C. Transurethral ureterorenolithotripsy using new automated irrigation/suction system controlling pressure and flow compared with standard irrigation: a randomized pilot study. J Endourol. 2003;17(2):97–101. doi: 10.1089/08927790360587423. [DOI] [PubMed] [Google Scholar]
  • 40.Ghani KR, Wolf JSJ. What is the stone-free rate following flexible ureteroscopy for kidney stones? Nat Rev Urol. 2015;12(5):281–288. doi: 10.1038/nrurol.2015.74. [DOI] [PubMed] [Google Scholar]
  • 41.Ahn JS, Holt SK, May PC, Harper JD. National imaging trends after ureteroscopic or shock wave lithotripsy for nephrolithiasis. J Urol. 2018;199(2):500–507. doi: 10.1016/j.juro.2017.09.079. [DOI] [PubMed] [Google Scholar]
  • 42.Chew BH, Brotherhood HL, Sur RL, Wang AQ, Knudsen BE, Yong C, et al. Natural history, complications and re-intervention rates of asymptomatic residual stone fragments after ureteroscopy: a report from the EDGE research consortium. J Urol. 2016;195(4 Pt 1):982–986. doi: 10.1016/j.juro.2015.11.009. [DOI] [PubMed] [Google Scholar]
  • 43.Portis AJ, Laliberte MA, Heinisch A. Repeat surgery after ureteroscopic laser lithotripsy with attempted complete extraction of fragments: long-term follow-up. Urology. 2015;85(6):1272–1278. doi: 10.1016/j.urology.2015.03.019. [DOI] [PubMed] [Google Scholar]
  • 44.Bhojani N, Paonessa JE, El Tayeb MM, Williams JCJ, Hameed TA, Lingeman JE. Sensitivity of noncontrast computed tomography for small renal calculi with endoscopy as the gold standard. Urology. 2018;117:36–40. doi: 10.1016/j.urology.2018.03.041. [DOI] [PubMed] [Google Scholar]
  • 45.Fayad AS, Elsheikh MG, Ghoneima W. Tubeless mini-percutaneous nephrolithotomy versus retrograde intrarenal surgery for lower calyceal stones of 2 cm: a prospective randomised controlled study. Arab J Urol. 2017;15(1):36–41. doi: 10.1016/j.aju.2016.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Lee YJ, Bak DJ, Chung J-W, Lee JN, Kim HT, Yoo ES, et al. Is it necessary to actively remove stone fragments during retrograde intrarenal surgery? Investig Clin Urol. 2016;57(4):274–279. doi: 10.4111/icu.2016.57.4.274. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.El-Nahas AR, Almousawi S, Alqattan Y, Alqadri IM, Al-Shaiji TF, Al-Terki A. Dusting versus fragmentation for renal stones during flexible ureteroscopy. Arab J Urol. 2019;17(2):138–142. doi: 10.1080/2090598X.2019.1601002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Ibrahim A, Elhilali MM, Fahmy N, Carrier S, Andonian S. Double-blinded prospective randomized clinical trial comparing regular and moses modes of holmium laser lithotripsy. J Endourol. 2020;34(5):624–628. doi: 10.1089/end.2019.0695. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Villa L, Ventimiglia E, Proietti S, Giusti G, Briganti A, Salonia A, et al. Does working channel position influence the effectiveness of flexible ureteroscopy? Results from an in vitro study. BJU Int. 2020;125(3):449–456. doi: 10.1111/bju.14923. [DOI] [PubMed] [Google Scholar]
  • 50.Dragos LB, Somani BK, Keller EX, de Coninck VMJ, Herrero MR-M, Kamphuis GM, et al. Characteristics of current digital single-use flexible ureteroscopes versus their reusable counterparts: an in-vitro comparative analysis. Transl Androl Urol. 2019;8(Suppl 4):S359–S370. doi: 10.21037/tau.2019.09.17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Talso M, Emiliani E, Haddad M, Berthe L, Baghdadi M, Montanari E, et al. Laser fiber and flexible ureterorenoscopy: the safety distance concept. J Endourol. 2016;30(12):1269–1274. doi: 10.1089/end.2016.0209. [DOI] [PubMed] [Google Scholar]
  • 52.De Cruz JE, Morcillo E, Folkersma LR, Alvarez S, Margallo FMS, Soria F. Complications of endourologic procedures: a review on iatrogenic ureteral perforation. ARC J Urol. 2017;2(2):17–24. [Google Scholar]
  • 53.Cornu JN, Herrmann T, Traxer O, Matlaga B. Prevention and management following complications from endourology procedures. Eur Urol Focus. 2016;2(1):49–59. doi: 10.1016/j.euf.2016.03.014. [DOI] [PubMed] [Google Scholar]
  • 54.Paiva MM, da Silva RD, Jaworski P, Kim FJ, Molina WR. Subcapsular hematoma after ureteroscopy and laser lithotripsy. Can J Urol. 2016;23(4):8385–8387. [PubMed] [Google Scholar]
  • 55.Srougi V, Padovani GP, Marchini GS, Vicentini FC, Mazzucchi E, Srougi M. Outcomes of surgical treatment of ureteral strictures after laser ureterolithotripsy for impacted stones. Can J Urol. 2015;22(6):8079–8084. [PubMed] [Google Scholar]
  • 56.Schoenthaler M, Buchholz N, Farin E, Ather H, Bach C, Bach T, et al. The Post-Ureteroscopic Lesion Scale (PULS): a multicenter video-based evaluation of inter-rater reliability. World J Urol. 2014;32(4):1033–1040. doi: 10.1007/s00345-013-1185-1. [DOI] [PubMed] [Google Scholar]
  • 57.Jung H, Osther PJS. Intraluminal pressure profiles during flexible ureterorenoscopy. Springerplus. 2015;4:373. doi: 10.1186/s40064-015-1114-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Wilson W, Preminger G. Intrarenal pressures generated during flexible deflectable ureterorenoscopy. J Endourol. 1990;4(2):135–141. [Google Scholar]
  • 59.Guo H-Q, Shi H-L, Li X-G, Gan W-D, Zeng L-Q, Liu G-X, et al. Relationship between the intrapelvic perfusion pressure in minimally invasive percutaneous nephrolithotomy and postoperative recovery. Zhonghua Wai Ke Za Zhi. 2008;46(1):52–54. [PubMed] [Google Scholar]
  • 60.Alsyouf M, Abourbih S, West B, Hodgson H, Baldwin DD. Elevated renal pelvic pressures during percutaneous nephrolithotomy risk higher postoperative pain and longer hospital stay. J Urol. 2018;199(1):193–199. doi: 10.1016/j.juro.2017.08.039. [DOI] [PubMed] [Google Scholar]
  • 61.Whitehurst LA, Somani BK. Perirenal hematoma after ureteroscopy: a systematic review. J Endourol. 2017;31(5):438–445. doi: 10.1089/end.2016.0832. [DOI] [PubMed] [Google Scholar]
  • 62.Uchida Y, Takazawa R, Kitayama S, Tsujii T. Predictive risk factors for systemic inflammatory response syndrome following ureteroscopic laser lithotripsy. Urolithiasis. 2018;46(4):375–381. doi: 10.1007/s00240-017-1000-3. [DOI] [PubMed] [Google Scholar]
  • 63.Fan S, Gong B, Hao Z, Zhang L, Zhou J, Zhang Y, et al. Risk factors of infectious complications following flexible ureteroscope with a holmium laser: a retrospective study. Int J Clin Exp Med. 2015;8(7):11252–11259. [PMC free article] [PubMed] [Google Scholar]
  • 64.Ng YH, Somani BK, Dennison A, Kata SG, Nabi G, Brown S. Irrigant flow and intrarenal pressure during flexible ureteroscopy: the effect of different access sheaths, working channel instruments, and hydrostatic pressure. J Endourol. 2010;24(12):1915–1920. doi: 10.1089/end.2010.0188. [DOI] [PubMed] [Google Scholar]
  • 65.Osther PJS. Risks of flexible ureterorenoscopy: pathophysiology and prevention. Urolithiasis. 2018;46(1):59–67. doi: 10.1007/s00240-017-1018-6. [DOI] [PubMed] [Google Scholar]
  • 66.Al-Qahtani SM, Letendre J, Thomas A, Natalin R, Saussez T, Traxer O. Which ureteral access sheath is compatible with your flexible ureteroscope? J Endourol. 2014;28(3):286–290. doi: 10.1089/end.2013.0375. [DOI] [PubMed] [Google Scholar]
  • 67.Hein S, Petzold R, Suarez-Ibarrola R, Müller P-F, Schoenthaler M, Miernik A. Thermal effects of Ho:YAG laser lithotripsy during retrograde intrarenal surgery and percutaneous nephrolithotomy in an ex vivo porcine kidney model. World J Urol. 2020;38(3):753–760. doi: 10.1007/s00345-019-02808-5. [DOI] [PubMed] [Google Scholar]
  • 68.Brehmer M, Svensson I. Heat-induced apoptosis in human prostatic stromal cells. BJU Int. 2000;85(4):535–541. doi: 10.1046/j.1464-410x.2000.00473.x. [DOI] [PubMed] [Google Scholar]
  • 69.Wollin DA, Carlos EC, Tom WR, Simmons WN, Preminger GM, Lipkin ME. Effect of laser settings and irrigation rates on ureteral temperature during holmium laser lithotripsy, an in vitro model. J Endourol. 2018;32(1):59–63. doi: 10.1089/end.2017.0658. [DOI] [PubMed] [Google Scholar]
  • 70.Aldoukhi AH, Ghani KR, Hall TL, Roberts WW. Thermal response to high-power holmium laser lithotripsy. J Endourol. 2017;31(12):1308–1312. doi: 10.1089/end.2017.0679. [DOI] [PubMed] [Google Scholar]
  • 71.Sapareto SA, Dewey WC. Thermal dose determination in cancer therapy. Int J Radiat Oncol Biol Phys. 1984;10(6):787–800. doi: 10.1016/0360-3016(84)90379-1. [DOI] [PubMed] [Google Scholar]
  • 72.Aldoukhi AH, Hall TL, Ghani KR, Maxwell AD, MacConaghy B, Roberts WW. Caliceal fluid temperature during high-power holmium laser lithotripsy in an in vivo porcine model. J Endourol. 2018;32(8):724–729. doi: 10.1089/end.2018.0395. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Tracey J, Gagin G, Morhardt D, Hollingsworth J, Ghani KR. Ureteroscopic high-frequency dusting utilizing a 120-W holmium laser. J Endourol. 2018;32(4):290–295. doi: 10.1089/end.2017.0220. [DOI] [PubMed] [Google Scholar]
  • 74.Blackmon RL, Irby PB, Fried NM. Thulium fiber laser lithotripsy using tapered fibers. Lasers Surg Med. 2010;42(1):45–50. doi: 10.1002/lsm.20883. [DOI] [PubMed] [Google Scholar]
  • 75.Blackmon RL, Irby PB, Fried NM. Comparison of holmium:YAG and thulium fiber laser lithotripsy: ablation thresholds, ablation rates, and retropulsion effects. J Biomed Opt. 2011;16(7):071403. doi: 10.1117/1.3564884. [DOI] [PubMed] [Google Scholar]
  • 76.Hardy LA, Gonzalez DA, Irby PB, Fried NM (2018) Fragmentation and dusting of large kidney stones using compact, air-cooled, high peak power, 1940-nm, thulium fiber laser. In: Kang HW, Chan KF (eds) Therapeutics and diagnostics in urology 2018. SPIE; 2018, pp 57–61
  • 77.Ventimiglia E, Doizi S, Kovalenko A, Andreeva V, Traxer O. Effect of temporal pulse shape on urinary stone phantom retropulsion rate and ablation efficiency using holmium:YAG and super-pulse thulium fibre lasers. BJU Int. 2020;126(1):159–167. doi: 10.1111/bju.15079. [DOI] [PubMed] [Google Scholar]
  • 78.Andreeva V, Vinarov A, Yaroslavsky I, Kovalenko A, Vybornov A, Rapoport L, et al. Preclinical comparison of superpulse thulium fiber laser and a holmium:YAG laser for lithotripsy. World J Urol. 2020;38(2):497–503. doi: 10.1007/s00345-019-02785-9. [DOI] [PubMed] [Google Scholar]
  • 79.White MD, Moran ME, Calvano CJ, Borhan-Manesh A, Mehlhaff BA. Evaluation of retropulsion caused by holmium:YAG laser with various power settings and fibers. J Endourol. 1998;12(2):183–186. doi: 10.1089/end.1998.12.183. [DOI] [PubMed] [Google Scholar]
  • 80.Fried NM. Recent advances in infrared laser lithotripsy [Invited] Biomed Opt Express. 2018;9(9):4552–4568. doi: 10.1364/BOE.9.004552. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81.Traxer O, Keller EX (2019) Thulium fiber laser: the new player for kidney stone treatment? A comparison with holmium:YAG laser. World J Urol. 10.1007/s00345-019-02654-5 [DOI] [PMC free article] [PubMed]
  • 82.Ergakov D, Martov AG, Guseynov M. The comparative clinical study of Ho: YAG and SuperPulse Tm fiber laser lithotripters. Eur Urol Suppl. 2018;17(2):e1391. [Google Scholar]
  • 83.Yaroslavsky I, Vinnichenko V, McNeill T, Novoseltseva A, Perchuk I, Vybornov A et al (2018) Optimization of a novel Tm fiber laser lithotripter in terms of stone ablation efficiency and retropulsion reduction. In: procspie. 2018, p 104680H (Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series; vol. 10468)
  • 84.Fried NM, Irby PB. Advances in laser technology and fibre-optic delivery systems in lithotripsy. Nat Rev Urol. 2018;15(9):563–573. doi: 10.1038/s41585-018-0035-8. [DOI] [PubMed] [Google Scholar]
  • 85.Traxer O, Rapoport L, Tsarichenko D, Dymov A, Enikeev D, Sorokin N, et al. V03-02 first clinical study on superpulse thulium fiber laser for lithotripsy. J Urol. 2018;199(4S):e321–e322. [Google Scholar]
  • 86.De Coninck VMJ, Keller EX, Kovalenko A, Vinnichenko V, Traxer O. Dusting efficiency comparison between Moses technology of Ho:YAG laser and superpulse thulium fiber laser. Eur Urol Suppl. 2019;18(1):e1757–e1758. [Google Scholar]
  • 87.Enikeev D, Taratkin M, Klimov R, Alyaev Y, Rapoport L, Gazimiev M et al (2020) Thulium-fiber laser for lithotripsy: first clinical experience in percutaneous nephrolithotomy. World J Urol. 10.1007/s00345-020-03134-x [DOI] [PubMed]

Articles from World Journal of Urology are provided here courtesy of Springer

RESOURCES