Abstract
Cardiovascular implantable electronic devices (CIED) are effective and important components of modern cardiovascular care. Despite the dramatic improvements in the functionality and reliability of these devices, over time patients are at risk for developing several morbidities, the most feared of which are local and systemic infections. Despite significant financial investment and aggressive therapy with hospitalization, intravenous antibiotics, and transvenous lead extraction, the outcomes include a 1-year mortality rate as high as 25%. This risk of infection has increased over time, likely due to the increased complexity of the surgical interventions required to insert and replace these devices. The only way to reduce this morbidity and mortality is to prevent these infections, and other than preoperative antibiotics, there were little data supporting effective therapy until the WRAP-IT trial provided randomized data showing that pocket infections can be reduced by 60% at 12 months and major CIED infections reduced by 40% at 1 year with the use of the absorbable antibiotic eluting envelope in patient CIED procedures at high risk of infection. Not all CIED procedures are at high risk of infection and justify the use of the envelope, but cost-effectiveness data support the use of the antibiotic envelope particularly in patients with defibrillator replacements, revisions, and upgrades, such as to a resynchronization device and in patients with prior CIED infection, history of immunocompromise, two or more prior procedures, or a history of renal dysfunction.
Keywords: Cardiovascular implantable electronic devices, Pacemaker, Implantable defibrillator, Cardiac resynchronization, Infection, Antibiotic envelope
What’s new?
Prospectively determined cardiovascular implantable electronic device major infection rate is 1.2% over 12 months.
Absorbable antibiotic envelope use reduces major infections 40% and pocket infections 60% over 12 months.
Cost effective use of the absorbable antibiotic envelope was observed in procedures in high-risk populations.
High-risk populations include patients with prior cardiovascular implantable electronic devices infection, history of immunocompromise, two or more prior procedures for those with an implantable defibrillator or resynchronization defibrillator, history of renal dysfunction, or patients undergoing a revision or upgrade.
Introduction
The first fully implantable pacemaker (PPM) was placed in 1958 and ushered in an era of rapid development of cardiac implantable electronic devices (CIED). Pacemakers (PPM), internal cardiac-defibrillators (ICD), and later cardiac resynchronization pacemakers (CRT-P) and defibrillators (CRT-D) provided life changing and lifesaving therapy for disease processes for which previously there were few therapeutic options. Over the ensuing decades, the number of CIED implants rose dramatically. Improvements in surgical technique and available tools led to a decrease in many complications associated with CIED implantation, however, infection rates have increased despite advances. Infection remains one of the most difficult challenges associated with CIED implantation and is associated with high morbidity and mortality and costly treatment. While the incidence of CIED infections has increased, surprisingly few advances have been made to help prevent these infections. For decades after the first pacemaker implantation, the only proven prophylactic measure against CIED infection was preoperative antibiotic administration.1
In recent years, however, we have finally seen progress in the effort to prevent CIED infections. Advances in device technology such as subcutaneous ICDs and leadless pacemakers allow device implantation without the need for endovascular leads or a subcutaneous pocket, respectively. While these technologies show promise in smaller studies and in select patient populations, their ability to impact infection rates at a large scale in their current state seems unlikely.2,3 In contrast, the antibiotic eluting envelope (AE) appears poised to make an impact immediately. In the largest cardiac device trial to date, the WRAP-IT investigators demonstrated a significant reduction in CIED infections with the use of the TYRX (Medtronic, TM, USA) AE.1 It remains to be seen whether the AE coupled with other measures such as strict surgical techniques and the use of preoperative antibiotics will finally begin to reverse the decades-long trend of increasing CIED infection rates.
Cardiac implantable electronic device infection—background
Cardiac implantable electronic device infections are one of the most common and devastating complications of CIED procedures with infection rates reported between 1% and 4%.1,4–6 Alarmingly, multiple studies have reported a staggering increase in the incidence of CIED infection over recent decades. Joy et al.7 reported an increase in the incidence of CIED infection from 1.45% in 2000 to 3.41% in 2008. The impact of these infections is profound. The 30-day mortality can be 5–8%, 1-year mortality up to 25%, and long-term mortality up to 1.5–2.4 times that of non-infected patients.8–10
Cardiac implantable electronic device infections are classified as pocket infections, limited to the pocket and surrounding tissue, or endovascular, including bacteraemia and endocarditis. Pocket infections are responsible for the majority of CIED infections and most commonly occur within 12 months of CIED surgery.11,12 Staphylococcal species are the responsible organism in about 70% of CIED infections while gram-negative bacilli account for another 9%.13,14 While multiple strategies of CIED infection prophylaxis have been investigated, until recently, only preoperative intravenous antibiotics have been shown to reduce the risk. Notably, local instillation of antibiotics into the pocket has not been demonstrated to reduce the risk of infection.15
Patient, procedural, and device factors have been associated with an increased risk of CIED infection (Table 1) and an increase in the presence of these risk factors in patients undergoing CIED implantation likely explains the increasing incidence.16–18 Patient factors which increase risk include end-stage renal disease, prior CIED infection, diabetes mellitus, and immunosuppression.15,19–21 Important procedural factors include haematoma formation, need for early re-operation, secondary procedures, and longer operative times. Device factors include CRT devices, high energy devices, and presence of multiple leads.13,20,22 While understanding these relevant risk factors may help guide appropriate use of emerging prophylactic options, it is important to note that most of these data come from relatively small observational studies.
Table 1.
Patient factors | Odds ratio | Source | Device factors | Odds ratio | Source |
---|---|---|---|---|---|
ESRD | 8.73 | 19,20 | Epicardial leads | 8.09 | 19,20 |
History CIED infectiona | 7.84 | 19,20 | Abdominal pocket | 4.01 | 19,20 |
Age ≥75 | 5.93 | 21 | CRTb | 2.87 | 15 |
Fever priora | 4.27 | 19,20 | ≥2 leads | 2.02 | 19,20 |
Immunosuppression | 3.44 | 19,20 | ICDa | 1.83 | 15 |
Renal insufficiencya | 1.48–3.02 | 19,20 | Dual chamber device | 1.45 | 19,20 |
COPD | 2.95 | 19,20 | |||
NYHA Class ≥2 | 2.47 | 19,20 | Procedural factors | Odds ratio | Source |
Skin disorder | 2.46 | 19,20 | <30 days reinterventiona | 16.29 | 21 |
Immunocompromised | 2.24 | 15 | Duration >1 h | 13.96 | 21 |
Malignancy | 2.23 | 19,20 | Haematoma | 8.46–4.95 | 19–21,35 |
Diabetes mellitusa | 2.08 | 19,20 | Revision/upgradeb | 6.46–4.16 | 15,21 |
Heparin bridging | 1.87 | 19,20 | Lead repositioning | 6.37 | 19,20 |
CHF | 1.65 | 19,20 | Replacementb | 4.93–1.7 | 19,20 |
Age <60 | 1.63 | 15 | ≥2 prior procedures | 3.37 | 15 |
Oral anticoagulantsa | 1.59 | 19,20 | Inexperienced operator | 2.85 | 19,20 |
Age 60–69 | 1.43 | 15 | Temporary pacing | 2.31 | 19,20 |
1 prior procedure | 1.51 | 15 |
CIED, cardiovascular implantable electronic devices; COPD, chronic obstructive pulmonary disease; CHF, congestive heart failure; CRT, cardiac resynchronization therapy; ESRD, endstage renal disease; ICD, implantable cardiac-defibrillator.
Prespecified risk factor for CIED infection in an observational AE trial.
Prespecified risk factor for CIED infection in a prospective, randomized AE trial.
Antibiotic envelopes
Given the profound impact of CIED infections and with little else in the way of effective prophylaxis, the TYRX AE was developed to fill this clinical need. The multifilament mesh envelope was also intended to minimize migration of the CIED. While the initial envelope was non-absorbable, the newest generation’s polymer comprised of glycolide, caprolactone, and trimethylene carbonate, is absorbed over the course of ∼9 weeks.23,24 An absorbable polyarylate polymer coating acts as a carrier for minocycline and rifampin.23 Antibiotic dose varies based upon the size of the AE, with the medium AE, intended for PPM implantation, containing 8.0 mg rifampin and 5.1 mg minocycline and the large AE, intended for ICD implantation, containing 11.9 mg rifampin and 7.6 mg minocycline.23 These antibiotics are eluted over a minimum of 7 days into the local tissue.23,24
Both rifampin and minocycline have a broad spectrum of activity, provide biofilm penetration, and are effective against most staphylococcal species making them choice candidates for prophylaxis against CIED pocket infections.25 Preclinical, in vitro trials demonstrated antimicrobial activity of the TYRX AE against organisms including methicillin-resistant and methicillin-sensitive Staphylococcus aureus, Staphylococcus epidermidis, and Escherichia coli among others.23 The TYRX AE also reduced infection after bacterial challenge in an animal model following CIED implantation.26
Early studies showed considerable promise for the non-absorbable AE with later studies supporting equal benefit from the absorbable envelope. Mittal et al.27 reported a reduction of CIED infection by 79% and 100% in medium- and high-risk groups, respectively, with the use of the non-absorbable AE. The combined Citadel and Centurion trials examined clinical outcomes with the use of the non-absorbable AE for patients undergoing replacement or upgrade with an ICD or CRT. The non-randomized registries demonstrated 12-month infection rates of 0.2% in the ICD group and 0.7% in the CRT cohort which compared favourably with the published benchmark of 2.2% for ICD infection and 1.3% for comorbidity-match controls for the CRT group.28 In a single-centre, retrospective cohort study, the non-absorbable and later the absorbable TYRX AEs were implanted in a total of 488 high-risk patients who were followed for a minimum of 300 days. Cardiac implantable electronic devices infection rates were 0% for the absorbable AE group, 0.3% for the non-absorbable AE group, and 3.1% for controls (P = 0.03 and 0.002, respectively).29 A single study reported a higher incidence of major infection in patients implanted with the non-absorbable AE compared with controls (5.4% and 1.1%, respectively, P = 0.048), but was a retrospective study in which the study group was at higher risk of infection given higher rates of chronic steroid use, longer hospitalizations, a greater proportion of devices with more than two intracardiac leads, and a higher rate of CIED replacement or revision for the index procedure.25
Given the initial promise of these non-randomized studies, the WRAP-IT trial was designed as the largest prospective, randomized CIED trial in history. 6983 patients were randomized in a 1 : 1 fashion to treatment with the absorbable TYRX AE or standard of care.1 The inclusion criteria specifying patients undergoing de novo CRT-D implantation, patients with an existing CIED undergoing generator replacement or upgrade or patients with an existing CIED undergoing system revision ensured a higher risk population.23 Notably, however, certain high-risk patients were excluded. Patients undergoing revision in whom the pocket had been opened in the previous 365 days were not included. Additionally, patients on chronic oral immunosuppression, on haemodialysis or peritoneal dialysis, patients requiring long-term vascular access, and those with prior CIED or endovascular infection within the prior 12 months were excluded.23 Exclusion of very high-risk populations may, in part, explain the low incidence of CIED infection seen in controls.
The WRAP-IT investigators reported that at 12 months, the primary endpoint of major infection occurred in 0.7% of patients implanted with the TYRX absorbable AE and 1.2% of patients in the control group which amounted to a 40% reduction in major infections.1,24 The benefit persisted in long-term follow-up with rates of major CIED infections of 1.3% in the envelope group and 1.9% in the control group for a hazard ratio of 0.64. The benefit of the envelope was driven by a 60% reduction in pocket infections which represented 75% of all major infections at 12 months. There was no reduction in the incidence of endocarditis or bacteremia.1 Within the subgroup receiving a high-power device (ICD or CRT-D), an approximate 50% reduction in major CIED infections was seen. No significant difference in major infections was observed in the low-power subgroup or in patients receiving a de novo CRT-D.1 The TYRX AE was successfully implanted in 99.7% of cases and there was no increased risk of procedure or system-related complications. Additionally, there were no reports of allergic reactions to the TYRX AE.24
Economic analyses have reported that the TYRX AE is cost-effective. In a cost analysis of the practice of implanting an AE as standard of care during CIED implantation, revision or replacement, one group estimated the cost of additional infections in the control group was $340 000 compared to the cost of the device in the AE group estimated at $320 000.30,31 The number needed to treat in the WRAP-IT trial was 200 while the base incremental cost-effectiveness of the TYRX AE was $112 603 per quality-adjusted life year.32,33 However, the cost-effectiveness of the AE was far more pronounced in certain subgroups such as those with prior CIED infection, history of immunocompromise, two or more prior procedures for those with a high power device, history of renal dysfunction, or undergoing a revision or upgrade33 (Figure 1).
Defining high risk
While no increased risk of procedure or device-related complications was seen in WRAP-IT or most of the other AE studies, not all patients derive equal benefit from the device. Not surprisingly, economic analyses suggest the patients that reap the greatest risk reduction from the TYRX AE are those at greatest risk for pocket infection. How we determine those at greatest risk for infection and most likely to benefit from AE implantation remains a challenge. Given its large size and prospective, randomized design, the WRAP-IT inclusion criteria, de novo CRT, generator change or upgrade, and CIED system revision, are the best supported indications for AE implantation.1,19 However, the trial excluded certain high-risk populations that are likely to benefit. For instance, early reoperation for CIED revision or upgrade has been shown to dramatically increase risk of CIED infection, yet patients with revision <365 days removed from prior CIED surgery were excluded to prevent confusion over the timing and source of the infection.21,34 While a host of other patient, device and procedural factors have been associated with increased risk of CIED infection (Table 1), these data come largely from observational studies with low event rates and the impact of AE implantation in the setting of many of these factors is unstudied.
Beyond the WRAP-IT inclusion criteria, data defining high-risk patients likely to benefit from AE implantation become murky. Other pre-specified risk factors from observational AE trials may be cautiously considered. A meta-analysis of AE studies found a significant reduction in the incidence of major infection for high-power devices, but no significant difference for low-power devices.30 Mittal et al.27 found implantation of the TYRX AE reduced infections by 79% and 100% in medium- and high-risk groups and defined risk using a multivariate logistic regression analysis to weight factors including early pocket re-exploration, male gender, diabetes mellitus, device upgrade, congestive heart failure, hypertension, and renal impairment. While a reduction in major CIED infection was reported by Kolek et al.29 in their single-centre retrospective study, AE implantation was reserved for patients deemed high risk by virtue of two or more risk factors for infection including diabetes mellitus, chronic renal disease, systemic anti-coagulation, chronic daily corticosteroid use, fever ≥100.5°F, or leucocytosis ≥11 000 WBC/μL 24 h prior to implantation, prior documented CIED infection, presence of three or more transvenous leads, pacemaker dependence, or early pocket reentry within 2 weeks of the original implantation. While these studies may suggest other high-risk patients likely to benefit from AE implantation, additional research is clearly needed.
The benefit of AE implantation in the setting of other risk factors is unproven. For some, the benefit of AE implantation would seem plausible from a mechanistic standpoint while for others, the benefit seems less likely. Pocket haematoma is associated with a seven-fold increase in risk of CIED infection, and while it seems reasonable that an AE would be of benefit in these cases, to date, there are no subgroup analyses examining AE efficacy in patients who develop a haematoma subsequent to their CIED procedure.16,20,29,35,36 The route of entry for most pocket infections is thought to be contamination during CIED surgery.19,37 Longer procedure times and operator inexperience may increase the risk of CIED infection. Perhaps longer exposure times or a less refined surgical prep or technique leads to an increased risk or burden of contamination. Again, the benefit of an AE implantation is such cases, while mechanistically plausible, is unproven. Age, young and old, has been variably reported as a risk factor for CIED infection, but the impact AE implantation in specific age groups is unknown.21,38 Finally, end-stage renal disease is one of the most powerful determinants of risk for CIED infection, but perhaps that risk will not be significantly impacted by an intervention with demonstrable effect against pocket infection, but no significant effect on endovascular infection.15,16,21,32,39 Future studies including planned analyses of the WRAP-IT data should further clarify the role of AEs.
Conclusion
Much work remains in the effort to eliminate CIED infections and their attendant morbidity, mortality risk, and economic burden. While the decades since the implantation of the first pacemaker saw incredible technological development and an explosion in implantation volumes, few advances were made in CIED prophylaxis. More recently, however, early retrospective and non-randomized studies of AE implantation showed promise with reductions in the incidence of CIED infection in high-risk patients.27–29,31 The WRAP-IT trial, the largest, randomized, prospective CIED trial to date, corroborated early findings, demonstrating a 40% reduction in CIED infections and a 60% reduction in major pocket infections.1,24 Numerous studies have demonstrated cost-effectiveness of the TYRX AE, but subgroup analyses reveal marked disparities with high-risk populations deriving the lion’s share of the benefit. Future studies should clarify which high-risk populations stand to benefit most from AE implantation during CIED surgery. In the interim, we will wait with great anticipation to see if this is the decade we finally see CIED infection rates begin to fall.
Funding
This article was published as part of a supplement supported by an educational grant from Medtronic.
Conflict of interest: B.L.W. has received consulting fees from Medtronic, Abbott and Philips. K.T. has received advisory board and consulting fees from Medtronic, Alivecor, Janssen, and Bristol Meyer Squibb. T.C. has received consultant fees from Biotronik.
References
- 1. Tarakji KG, Mittal S, Kennergren C, Corey R, Poole JE, Schloss E. et al. Antibacterial envelope to prevent cardiac implantable device infection. N Engl J Med 2019;380:1895–905. [DOI] [PubMed] [Google Scholar]
- 2. Koay A, Khelae S, Wei KK, Muhammad Z, Mohd Ali R, Omar R.. Treating an infected transcatheter pacemaker system via percutaneous extraction. Heart Rhythm Case Rep 2016;2:360–2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. El-Chami MF, Soejima K, Piccini JP, Reynolds D, Ritter P, Okabe T. et al. Incidence and outcomes of systemic infections in patients with leadless pacemakers: data from the Micra IDE study. Pacing Clin Electrophysiol 2019;42:1105–10. [DOI] [PubMed] [Google Scholar]
- 4. Clementy N, Carion PL, Leotoing L, Lamarsalle L, Wilquin-Bequet F, Brown B. et al. Infections and associated costs following cardiovascular implantable electronic device implantations: a nationwide cohort study. Europace 2018;20:1974–80. [DOI] [PubMed] [Google Scholar]
- 5. Ludwig S, Theis C, Brown B, Witthohn A, Lux W, Goette A.. Incidence and costs of cardiac device infections: retrospective analysis using German health claims data. J Comp Eff Res 2018;7:483–92. [DOI] [PubMed] [Google Scholar]
- 6. Goldenberg GR, Barsheshet A, Bishara J, Kadmon E, Omelchencko A, Strasberg B. et al. Effect of fibrotic capsule debridement during generator replacement on cardiac implantable electronic device infection risk. J Interv Card Electrophysiol 2020;58:113–8. [DOI] [PubMed] [Google Scholar]
- 7. Joy PS, Kumar G, Poole JE, London B, Olshansky B.. Cardiac implantable electronic device infections: who is at greatest risk? Heart Rhythm 2017;14:839–45. [DOI] [PubMed] [Google Scholar]
- 8. Lee DH, Gracely EJ, Aleem SY, Kutalek SP, Vielemeyer O.. Differences of mortality rates between pocket and nonpocket cardiovascular implantable electronic device infections. Pacing Clin Electrophysiol 2015;38:1456–63. [DOI] [PubMed] [Google Scholar]
- 9. Boersma L, Burke MC, Neuzil P, Lambiase P, Friehling T, Theuns DA. et al. Infection and mortality after implantation of a subcutaneous ICD after transvenous ICD extraction. Heart Rhythm 2016;13:157–64. [DOI] [PubMed] [Google Scholar]
- 10. El-Chami MF, Bonner M, Holbrook R, Stromberg K, Mayotte J, Molan A. et al. Leadless pacemakers reduce risk of device-related infection: review of the potential mechanisms. Heart Rhythm 2020;17:1393–7. [DOI] [PubMed] [Google Scholar]
- 11. Philippon F, O’Hara GE, Champagne J, Hohnloser SH, Glikson M, Neuzner J. et al. Rate, time course, and predictors of implantable cardioverter defibrillator infections: an analysis from the SIMPLE Trial. CJC Open 2020;2:354–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Uslan DZ, Sohail MR, St Sauver JL, Friedman PA, Hayes DL, Stoner SM. et al. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med 2007;167:669–75. [DOI] [PubMed] [Google Scholar]
- 13. Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, Council on Cardiovascular Surgery and Anesthesia et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation 2010;121:458–77. [DOI] [PubMed] [Google Scholar]
- 14. Sohail MR, Uslan DZ, Khan AH, Friedman PA, Hayes DL, Wilson WR. et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol 2007;49:1851–9. [DOI] [PubMed] [Google Scholar]
- 15. Birnie DH, Wang J, Alings M, Philippon F, Parkash R, Manlucu J. et al. Risk factors for infections involving cardiac implanted electronic devices. J Am Coll Cardiol 2019;74:2845–54. [DOI] [PubMed] [Google Scholar]
- 16. Balla C, Brieda A, Righetto A, Vitali F, Malagu M, Cultrera R. et al. Predictors of infection after “de novo” cardiac electronic device implantation. Eur J Intern Med 2020;77:73–8. [DOI] [PubMed] [Google Scholar]
- 17. Greenspon AJ, Patel JD, Lau E, Ochoa JA, Frisch DR, Ho RT. et al. 16-Year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008. J Am Coll Cardiol 2011;58:1001–6. [DOI] [PubMed] [Google Scholar]
- 18. Voigt A, Shalaby A, Saba S.. Continued rise in rates of cardiovascular implantable electronic device infections in the United States: temporal trends and causative insights. Pacing Clin Electrophysiol 2010;33:414–9. [DOI] [PubMed] [Google Scholar]
- 19. Blomstrom-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG. et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID), and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace 2019;22:515–49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Polyzos KA, Konstantelias AA, Falagas ME.. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace 2015;17:767–77. [DOI] [PubMed] [Google Scholar]
- 21. Sławek-Szmyt S, Araszkiewicz A, Grygier M, Szmyt K, Chmielewska-Michalak L, Seniuk W. et al. Predictors of long-term infections after cardiac implantable electronic device surgery- utility of novel PADIT and PACE DRAP scores. Circ J 2020;84:1754–63. [DOI] [PubMed] [Google Scholar]
- 22. Tarakji KG, Chan EJ, Cantillon DJ, Doonan AL, Hu T, Schmitt S. et al. Cardiac implantable electronic device infections: presentation, management, and patient outcomes. Heart Rhythm 2010;7:1043–7. [DOI] [PubMed] [Google Scholar]
- 23. Tarakji KG, Mittal S, Kennergren C, Corey R, Poole J, Stromberg K. et al. Worldwide Randomized Antibiotic EnveloPe Infection PrevenTion Trial (WRAP-IT). Am Heart J 2016;180:12–21. [DOI] [PubMed] [Google Scholar]
- 24. Mittal S, Wilkoff BL, Kennergren C, Poole JE, Corey R, Bracke FA. et al. The World-wide Randomized Antibiotic Envelope Infection Prevention (WRAP-IT) trial: long-term follow-up. Heart Rhythm 2020;17:1115–22. [DOI] [PubMed] [Google Scholar]
- 25. Hassoun A, Thottacherry ED, Raja M, Scully M, Azarbal A.. Retrospective comparative analysis of cardiovascular implantable electronic device infections with and without the use of antibacterial envelopes. J Hosp Infect 2017;95:286–91. [DOI] [PubMed] [Google Scholar]
- 26. Hansen LK, Brown M, Johnson D, Palme Ii DF, Love C, Darouiche R.. In vivo model of human pathogen infection and demonstration of efficacy by an antimicrobial pouch for pacing devices. Pacing Clin Electrophysiol 2009;32:898–907. [DOI] [PubMed] [Google Scholar]
- 27. Mittal S, Shaw RE, Michel K, Palekar R, Arshad A, Musat D. et al. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm 2014;11:595–601. [DOI] [PubMed] [Google Scholar]
- 28. Henrikson CA, Sohail MR, Acosta H, Johnson EE, Rosenthal L, Pachulski R. et al. Antibacterial envelope is associated with low infection rates after implantable cardioverter-defibrillator and cardiac resynchronization therapy device replacement: results of the citadel and centurion studies. JACC Clin Electrophysiol 2017;3:1158–67. [DOI] [PubMed] [Google Scholar]
- 29. Kolek MJ, Patel NJ, Clair WK, Whalen SP, Rottman JN, Kanagasundram A. et al. Efficacy of a bio-absorbable antibacterial envelope to prevent cardiac implantable electronic device infections in high-risk subjects. J Cardiovasc Electrophysiol 2015;26:1111–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Pranata R, Tondas AE, Vania R, Yuniadi Y.. Antibiotic envelope is associated with reduction in cardiac implantable electronic devices infections especially for high-power device—systematic review and meta-analysis. J Arrhythmia 2020;36:166–73. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Shariff N, Eby E, Adelstein E, Jain S, Shalaby A, Saba S. et al. Health and economic outcomes associated with use of an antimicrobial envelope as a standard of care for cardiac implantable electronic device implantation. J Cardiovasc Electrophysiol 2015;26:783–9. [DOI] [PubMed] [Google Scholar]
- 32. Barbar T, Patel R, Thomas G, Cheung JW.. Strategies to prevent cardiac implantable electronic device infection. J Innov Cardiac Rhythm Manage 2020;11:3949–56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Wilkoff BL, Boriani G, Mittal S, Poole JE, Kennergren C, Corey GR. et al. Cost-effectiveness of an antibacterial envelope for cardiac implantable electronic device infection prevention in the US Healthcare System From the WRAP-IT Trial. Circ Arrhythm Electrophysiol 2020;13:e008503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N. et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation 2007;116:1349–55. [DOI] [PubMed] [Google Scholar]
- 35. Essebag V, Verma A, Healey JS, Krahn AD, Kalfon E, Coutu B. et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION Study. J Am Coll Cardiol 2016;67:1300–8. [DOI] [PubMed] [Google Scholar]
- 36. de Oliveira JC, Martinelli M, Nishioka SAD, Varejão TNIA, Uipe DAVID, Pedrosa ANÍSIOALEXANDREANDRADE. et al. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol 2009;2:29–34. [DOI] [PubMed] [Google Scholar]
- 37. Da Costa A, Lelievre H, Kirkorian G, Celard M, Chevalier P, Vandenesch F. et al. Role of the preaxillary flora in pacemaker infections: a prospective study. Circulation 1998;97:1791–5. [DOI] [PubMed] [Google Scholar]
- 38. Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB.. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J 2019;40:1862–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Rennert-May E, Chew D, Lu S, Chu A, Kuriachan V, Somayaji R.. Epidemiology of cardiac implantable electronic device infections in the United States: a population-based cohort study. Heart Rhythm 2020;17:1125–31. [DOI] [PubMed] [Google Scholar]