Abstract
The serial crystallography (SX) technique enables the determination of the room-temperature structure of a macromolecule while causing minimal radiation damage, as well as the visualization of the molecular dynamics by time-resolved studies. The fixed-target (FT) scanning approach is one method for SX sample delivery that minimizes sample consumption and minimizes physical damage to crystals during data collection. Settling of the crystals on the sample holder in random orientation is important for complete three dimensional data collection. To increase the random orientation of crystals on the sample holder, we developed a polyimide mesh-based sample holder with irregular crystal mounting holes for FT-SX. The polyimide mesh was fabricated using a picosecond laser. Each hole in the polyimide mesh has irregularly shaped holes because of laser thermal damage, which may cause more crystals to settle at random orientations compared to regular shaped sample holders. A crystal sample was spread onto a polyimide-mesh, and a polyimide film was added to both sides to prevent dehydration. Using this sample holder, FT-SX was performed at synchrotron and determined the room-temperature lysozyme structure at 1.65 Å. The polyimide mesh with irregularly shaped holes will allow for expanded applications in sample delivery for FT-SX experiments.
Subject terms: X-ray crystallography, Nanocrystallography
Introduction
Serial crystallography (SX) using an X-ray free electron laser or synchrotron X-ray is an emerging technique for determining room-temperature structures with minimal radiation damage1–3. These room-temperature structures exhibit more biologically relevant structural flexibility compared to traditional cryocrystallography, and can provide reliable structural information on radiation-sensitive proteins4–6. Moreover, the SX technique enables visualization of time-resolved structural dynamics in pump-probe studies using optical laser or diffusion-based approaches7–9. Therefore, SX can be used to improve the current understanding of the biological and chemical mechanisms of macromolecules compared to conventional X-ray crystallography.
In general SX experiments, X-ray exposes to each crystal sample only once or to renew volumes each time on a large crystal sample10. Therefore, large number of crystals are required for collecting complete diffraction data, which is one of the obstacles in SX experiments. Accordingly, to reduce the efforts in sample preparation, minimizing the consumption of crystal samples during SX data collection is one of the key factors. Various sample delivery techniques such as injectors11–13, syringes14,15, injection with viscous medium12,14,16–21, fixed target scanning22–28, microfluidic devices29, capillaries5,30, and conveyer belts31 have been developed and applied to SX experiments at XFEL facilities or synchrotrons. Among these, the fixed target sample scanning method has a great advantage for low sample consumption and for minimizing the physical damage to the crystal during the data collection when compared with other sample delivery methods22,25,26. Various types of sample holders such as silicon nitride22,32, graphene-based chips33, and nylon mesh26,28 have been developed to stably fix crystals to the sample holder. These crystal supports are regularly arranged with holes of a specific shape and size. During data collection, X-rays are passed through the hole in which the crystal in the sample holder is placed. Transmission of X-rays to crystals through regular holes is advantageous for lowering the background scattering but requires precise alignment of the X-rays and holes. In contrast, in a nylon mesh-based sample holder composed of nylon and polyimide, which can transmit X-rays and generate low background scattering, data is collected by scanning the nylon mesh holes as well as the nylon mesh material on which the crystals can be placed26,28. Methods for fixing crystals between Mylar films34 and for placing crystals in polyimide tubing35 have been applied in FT-SX data collection. Although these methods have been successfully applied in SX experiments, crystals spread onto regular-shaped holes on the sample support or settle in an identical plane in the sample holder, and thus settle in a preferred orientation depending on the shape of the crystals and crystal supporting materials used. If most of the crystals are in the preferred orientation in the sample holder, it may not be possible to collect the complete diffraction dataset or it may be necessary to collect more images compared to a normal dataset.
To obtain crystals in various orientations within the sample holder, a method for supporting crystals embedded in a viscous material between two films was applied for FT-SFX27. In this method, lysozyme and glucose isomerase crystals were embedded in agarose and gelatin, respectively, and placed at a thickness of 350 µm between polyimide films27. Although this method was effective, it may be difficult to apply when the crystals are physically damaged during the mixing step with viscous medium27. On the contrary, FT-SSX with oscillation has been developed to collect the complete dataset by introducing oscillations in the sample holder for a short time when the fixed sample is exposed to X-rays36. The advantage of this method is that it can yield diffraction data for a wide area of reciprocal space of the crystal during data collection36. However, additional devices are required to apply this system to an existing beamline hutch and the previous beamline equipment needs to be reinstalled.
To enhance the random orientation of crystals on the sample holder, we developed a polyimide mesh-based crystal support with irregular holes for SX experiments. The irregular hole in the polyimide film was fabricated using a picosecond laser. Using this crystal support, we performed fixed-target serial synchrotron crystallography (FT-SSX) experiments using lysozyme crystals of 30–40 µm and determined its room-temperature structure at 1.65 Å. This polyimide mesh-based crystal support with irregular holes can be used with the sample holder in FT-SX experiments.
Results
Fabrication of polyimide mesh with irregular holes
We used the following criteria to develop a new crystal support for FT-SX: (1) The crystal support material should transmit X-rays. (2) X-ray background scattering from the crystal support should not affect data analysis. (3) During data collection, precise alignment between the X-ray and sample holder should not be required. (4) To increase random orientations of the crystal, the hole in which the crystal is mounted and its surrounding environment, should be irregular. (5) The sample holder should maintain the hydrated environment of the crystal sample.
Polyimide film was selected as the crystal-supporting material because it satisfied our sample holder development criteria. Initially, we considered fabricating holes with various shapes in a single polyimide film by using a laser drilling method to provide a random orientation when the crystals are mounted on the holes in the polyimide film. During the initial laser drilling experiment on the polyimide film, we observed that thermal damage occurred around the location where the laser penetrated (Fig. 1A). Even if lasers with the same spacing and power were used to penetrate the polyimide layer, drilling resulted in irregular shaped holes because of the inherent thermal damage at the laser drilling point. We considered that this feature was very useful for enabling the random orientation of crystals on the polyimide mesh. Next, laser drilling in a rectangular shape was performed on the polyimide film. During laser drilling of the polyimide film, when the drilling interval was narrow, the film spacer between the drilling holes was thermally damaged. To prevent this, a polyimide mesh with an irregular hole shape was manufactured by controlling the laser energy intensity and laser transmission position (Fig. 1B). Finally, the intensity and exposure position of the laser were adjusted so that the hole size was maintained at approximately 350 × 50 μm (Fig. 1C). As observed under a microscope, the holes were rectangular with rounded corners (Fig. 1D). The shape of each of the pores in the polyimide mesh was varied due to thermal damage, which would allow the crystals to be settled in a relatively random manner compared to a sample holder with holes of uniform shape. The thermally damaged film showed an irregular depth of approximately 20–30 μm in the rear direction through which the laser penetrated (Fig. 1E). These crystal supporting holes formed an irregularly shaped funnel-like structure. When spreading the crystals on the polyimide mesh-based sample support with irregular holes, the crystals could be settled on the polyimide mesh surface and inside the hole. The distance of the crystal position from the surface of the polyimide mesh to the inside of the hole was less than 50 μm, which was in a range that does not critically affect the distance from the sample to the detector during data processing.
Assembly of the sample holder
When crystals are spread on a polyimide mesh and then exposed to room temperature in the absence of hydration, a dehydrated environment is created by evaporation of the crystallization solution. This may cause various problems, such as changes in the crystal space group, collapse of the crystal lattice, and deterioration of data quality because of salt crystallization from the evaporated crystal solution26,37. To prevent these effects, it is very important to maintain the hydrated environment of the crystals settled on the polyimide mesh. In this experiment, the method of enclosing the mounted crystal sample using a polyimide film was the same as that for the previously used nylon mesh-based sample holder26,28. The enclosing polyimide film was attached to the PVC frame for easy handling (Fig. 2A and Supplementary Figure S1). When a crystal-mounted polyimide mesh was placed between the polyimide films, the bottom and upper films were enclosed using double-sided adhesive polyimide film (Fig. 2A,B and Supplementary Figure S2). This sample holder was attached to a magnetic base and then mounted on the goniometer for data collection by the raster scanning method (Fig. 2C).
Application of the polyimide-mesh based sample holder to FT-SSX
To demonstrate the application of the polyimide mesh-based sample holder with irregular crystal mounting holes, FT-SSX experiments were performed using lysozyme as a model sample on an X-ray synchrotron. Data were collected by raster scanning with oscillation at room temperature (Table 1). Although the X-ray size at the sample position was less than 10 μm (full-width half maximum) in both the horizontal and vertical directions, to avoid radiation damage to neighboring crystals around the X-ray penetration point, the interval of raster scanning points was selected as 50 μm in both the vertical and horizontal directions38. During raster scanning, the sample holder was moved both vertically and horizontally by 6750 μm. Accordingly, there were 135 scan points in the both vertical and horizontal directions, and 36,450 images were collected in 80 min using two sample holders. At each raster scanning point, the X-ray exposure time was 100 ms with an oscillation of 0.011°. The average diffracted weight dose and the average dose to the exposed area of the crystal for the X-ray used in this experiment were calculated as 0.607264 MGy and 0.309161 MGy, respectively. Further, the recent experimentally determined dose limit was 0.38 MGy when performing the room-temperature SSX experiment38.
Table 1.
Data collection | Lysozyme |
---|---|
Energy (eV) | 12,659 |
Expose time (ms) | 100 |
Space group | P43212 |
Cell dimension (Å) | |
a | 79.45 |
b | 79.45 |
c | 38.47 |
Collected images | 36,450 |
Hits images | 11,958 |
Indexed images | 7288 |
Indexed pattern | 7423 |
Resolution (Å) | 80.0–1.65 (1.71–1.65) |
Unique reflections | 15,398 (1498) |
Completeness (%) | 100.0 (99.93) |
Multiplicity | 83.7 (57.7) |
SNR | 6.58 (1.58) |
CC | 0.9836 (0.5635) |
CC* | 0.9958 (0.8490) |
Rsplit (%)a | 10.69 (70.26) |
Wilson B factor (Å2) | 31.76 |
Refinement | |
Resolution (Å) | 56.18–1.65 (1.70–1.65) |
Rworkb | 17.85 (27.74) |
Rfreec | 20.87 (27.92) |
R.m.s. deviations | |
Bond length (Å) | 0.013 |
Bond angle (°) | 1.618 |
B factors (Å) | |
Protein | 31.23 |
Water | 34.03 |
Ramachandran (%) | |
Preferred | 98.43 |
Allowed | 1.57 |
Outliers | 0.00 |
Highest resolution shell is shown in parentheses.
aRsplit = .
bRwork = Σ||Fobs| − |Fcalc||/Σ|Fobs|, where Fobs and Fcalc are the observed and calculated structure-factor amplitudes respectively.
cRfree was calculated as Rwork using a randomly selected subset of unique reflections not used for structure refinement.
In total, 11,958 images included Bragg peaks with a hit rate of 32.80%. Through indexing, 7423 lysozyme diffraction patterns were obtained from all 7288 hit images. The indexing and multi-crystal hit rates were 60.94% and 1.85%, respectively. Data were collected up to 1.65 Å with a completeness of 100%, and the redundancy, CC, CC*, SNR, and Rsplit were 83.7, 0.9836, 0.9958, 6.58, and 10.69, respectively. Structure refinement was performed to 1.65 Å, and the Rwork and Rfree of the final model were 17.85% and 20.87%, respectively. A clear electron density map was observed from Lys19 to Leu145 (Fig. 3A). Electron density map analysis showed no significant negative Fo-Fc electron density, considered as radiation damage, at the disulfide bond sites (Csy24-Csy145, Csy48-Csy133, Cys82-Cys98, and Cys94-Cys112) (Fig. 3B), which are relatively more radiation-sensitive than other amino acids. Superimposition of the lysozyme delivered in the polyimide mesh with the previously reported lysozyme structure at room temperature using a nylon mesh (PDB code 6IRJ), polyacrylamide (6I6G), shortening (6KCB), wheat starch (7BVM), and lard injection matrix (7CJZ) revealed a high similarity with a root mean squared deviation of 0.106–0.231 Å for all Cα atoms.
Next, we re-processed the collected data to analyze whether minimal images were required to obtain a reliable structure. Following processing with 1.65 Å resolution in 1000 units (Table 2), when 3000 diffraction patterns were used, the overall completeness, SNR, CC, CC*, and Rsplit were 99.96, 4.25, 0.9277, 0.9810, and 21.50, respectively. Although the quality of the data statistics was low compared to that of the complete data, we obtained a dataset suitable for structure determination from at least 3000 diffraction patterns. In further data processing, datasets merged from 1000, 2000, and 3000 images provide the solution for molecular replacement. Although all data showed a reliable electron density map (Fig. 4), only the dataset merged from 3000 images provided the appropriate model with a Rwork/Rfree of 0.2162/0.2599.
Table 2.
Data processing | No. of indexed images | ||
---|---|---|---|
1000 | 2000 | 3000 | |
Resolution (Å) | 80.0–1.65 (1.71–1.65) | ||
Unique reflections | 14,827 (1413) | 15,355 (1490) | 15,398 (1499) |
Completeness (%) | 96.29 (94.26) | 99.72 (99.40) | 99.96 (99.80) |
Multiplicity | 8.7 (5.9) | 17.8 (12.1) | 28.9 (19.7) |
SNR | 3.00 (1.13) | 3.43 (1.28) | 4.25 (1.61) |
CC | 0.8116 (0.3654) | 0.8792 (0.4508) | 0.9277 (0.5570) |
CC* | 0.9465 (0.7316) | 0.9673 (0.7883) | 0.9810 (0.8458) |
Rsplit (%) | 37.53 (110.47) | 28.69 (81.67) | 21.50 (67.10) |
Molecular replacement | |||
Top LLG | 3361.874 | 5053.713 | 6122.978 |
Top TFZ | 53.6 | 60.1 | 62.4 |
Refinement | |||
Resolution (Å) | 56.18–1.65 (1.73–1.665) | ||
Rwork | 0.2993 (0.4063) | 0.2500 (0.3357) | 0.2162 (0.3104) |
Rfree | 0.3563 (0.4110) | 0.2935 (0.3977) | 0.2599 (0.3463) |
Measurement of background scattering
Polyimide is well-known to generate very low background scatter by X-rays; in this experiment, polyimide showed negligible levels of X-ray scattering during data processing39. Nevertheless, analysis of background scattering from polyimide generated during data collection is essential for further SX applications. In this experiment, the thickness of the polyimide mesh and polyimide film used for enclosing the film was 25 μm. As the data were collected by raster scanning, X-rays penetrated the polyimide mesh pores, as well as the polyimide mesh, and the thicknesses through which X-rays were transmitted were 50 and 75 µm, respectively. Because both materials were composed of the same polyimide, they both exhibited circular background scattering at 15 Å (Fig. 5A). The thickness of the 50- and 75-μm polyimides showed 7–9 analogue-to-digital units (ADUs) (Fig. 5A). We consider this value to indicate a very low background scattering that is negligible in data processing. Next, we compared background scattering of the polyimide mesh-based sample holder with the previously reported nylon mesh and viscous-based background scattering. The background scattering of nylon mesh showed 10 ADU at 3.4 and 4.5 Å, respectively, which included 7 ADU at 15 Å generated from the enclosing polyimide film (Fig. 5B). For the crystal support using agarose, background scattering from agarose showed 55 ADU at 90 Å, and diffuse background scattering for the solution contained ~ 15 ADU at 3.2 Å (Fig. 5C). This support was also enclosed by a polyimide film and showed ~ 13 ADU at 15 Å (Fig. 5C). Although background scattering from the nylon mesh and from the viscous materials did not significantly affect data processing, the polyimide mesh-based sample holder showed the least background scattering.
Discussion
We developed a new crystal support using a polyimide mesh-based sample holder with irregular crystal mounting holes for application in FT-SX. Using this sample holder, we successfully determined the room-temperature structure of lysozyme FT-SSX. General sample holders have the same size and shape of holes in which the crystal is mounted; we fabricated a sample holder with holes showing irregular shapes based on thermal damage that occurs when the laser penetrates the polyimide film. It was not possible to directly compare and analyze the extent to which the polyimide mesh with irregular holes conferred significant advantages in data collection. However, when a crystal is placed in a regular-shaped hole and the sample holder is placed vertically for data collection, it will be placed in the preferred orientation according to the shape of the crystal and the hole of the sample holder when the crystal is settled at the bottom of the hole by gravity. In contrast, for the irregular hole applied in this experiment, even if the crystal sinks to the bottom of the hole via gravity, it settles in a random orientation because the bottom of the holes is irregular.
In previously reported FT-SX, sample mounting methods such as a sandwich and sheet-on-sheet approach do not require precise alignment between the sample holder and X-ray23,34; therefore, they do not require technical operation and allow more data collection during beamtime. As all the material in the sample holder used in this study is composed of polyimide, through which X-rays can penetrate, it is not necessary to precisely align the X-rays and sample holders, in contrast to a sample holder in which the X-rays pass through a specific hole. In addition, the X-ray background scattering generated from the sample holder occurs in the low-resolution region, and thus does not significantly affect data resolution; further, the intensity of X-ray background scattering is extremely low. Accordingly, this sample holder may be applicable for crystal samples with a weak diffraction limit as well as in phasing experiments, where low background scattering is an important factor. In this study, we created irregular holes in the polyimide film using a picosecond laser; however, crystal mounting holes of various shapes can also be made by precision microfabrication or other laser sources. The rectangular crystal mounting hole had a size of 50 × 350 μm, which is suitable for a crystal of 20–50 μm in size. In future application studies, when fabricating a polyimide mesh with irregular holes, the size of the mesh pores should be adjusted considering the purpose of the study and the size of the crystal. Additionally, because the sample holder is composed only of polyimide film, it can be reused without causing physical damage during synchrotron X-ray experiments; in contrast, when applied with an X-ray free electron laser, the polyimide mesh is damaged.
In conclusion, the proposed polyimide mesh-based sample support with irregular crystal mounting holes can be used for future FT-SX studies. However, since this experiment did not quantitatively investigate the effectiveness of the sample holder with irregular holes is for data collection, future studies need to compare the diffraction data quality to that of a traditional sample holder with a flat surface. Specifically, it is necessary to compare the random orientation of the crystals in the two sample holders using a low symmetry crystal, and to compare the number of indexed images required to determine the structure with a certain level of quality (completeness balanced with redundancy). This further study would provide insight into the data collection efficiency of a sample holder with irregular holes by performing a repeated data collection on a number of mounts.
Methods
Fabrication of polyimide mesh
The polyimide film (25 μL) was purchased from Covalue Youngjin Co. (Daegu, Republic of Korea). Laser drilling of polyimide to create crystal mounting holes was performed using a picosecond laser at UV 355 nm (Kortherm Science Co., Ltd., Incheon, Korea). The marking speed and delay were 300 mm/s and 310 μs, respectively. The jump speed and delay were 583.3 mm/s and 500 μs, respectively. To remove the dust generated during laser drilling, the polyimide-mesh was rinsed with water and ethanol.
Protein crystal preparation
Lysozyme from chicken egg white was purchased from Sigma-Aldrich (L6876; St. Louis, MO, USA). Lysozyme was crystallized by the microtube batch crystallization method as reported previously18. Briefly, lysozyme powder was dissolved in a buffer containing 10 mM Tris–HCl, pH 8.0, and 200 mM NaCl, and the final lysozyme solution concentration was 50 mg/mL. The protein solution (200 μL) was mixed in a microtube containing crystallization solution (200 µL) composed of 0.1 M Na-acetate, 2 M NaCl, and 8% (w/v) PEG8000. The mixed samples were immediately vortexed for 30 s and then incubated overnight at 20 °C. The crystal size of the lysozyme was 30–40 μm.
Preparation of sample holder
To facilitate the handling of the enclosing film, a 0.3-mm-thick PVC frame was attached to a polyimide film (13 × 13 mm) with double-sided adhesive polyimide tape. A polyimide-mesh with irregular holes (8 × 8 mm) was placed on the first enclosing film. The lysozyme crystal suspension (30 μm) was transferred onto the polyimide mesh prepared with irregular holes using a pipette. After spreading the crystals with a pipette tip, 15 μL of crystallization solution was removed from the corner of the sample holder, and the second enclosing film was immediately covered with double-sided adhesive polyimide tape. The sample holder was then mounted on the goniometer and fixed to the magnetic base using clay.
Data collection
The FT-SSX experiment using the polyimide mesh with irregular holes was performed at the 11C beamline at the Pohang Light Source (PLS-II, Republic of Korea). The X-ray energy and photon flux were 12.659 keV and 1.3 × 1012 photons/s, respectively. The beam size at the sample position was 4.5 (vertical) × 8 (horizontal) μm2 (full-width half maximum). Data were collected as previously reported for FT-SSX using a nylon mesh-based sample holder28. Briefly, raster scanning was performed at 50-µm intervals in both the horizontal and vertical directions. The X-ray was exposed to each raster scanning point for 100 ms with 0.011° oscillation. The average diffraction weighted dose and average dose in the exposed region of the crystal were calculated using RADDOSE-3D40. Data were collected at room temperature (24–25 °C) and diffraction patterns were recorded on a Pilatus 6 M with a 10-Hz readout (Dectris, Baden-Daettwil, Switzerland).
Data processing and structure determination
Hit images containing diffraction patterns were filtered using Cheetah41 and then processed with CrystFEL42. The phasing problem was solved by molecular replacement using phase-MR in Phenix43, with lysozyme (PDB code 7CJZ)18 used as the search model. Manual model building was performed using Coot44. Structure refinement was performed using Phenix.refinement in PHENIX43. The geometry of the final model was validated using MolProbity45. Structure figures were generated using PyMOL (https://pymol.org/). The coordinates and structure factors have been deposited to Protein Data Bank under the accession code 7DTB. Diffraction images have been deposited to CXIDB under ID 163.
Background scattering analysis
The polyimide mesh with irregular holes, nylon mesh-based sample holder, and viscous medium-based crystal supporting sample holder were exposed to an X-ray for 100 ms. The polyimide mesh (25 µm), nylon-mesh (pore size: 60 µm), and gelatin viscous medium (thickness: < 350 µm) were enclosed in two polyimide films (total thickness: 50 µm). The X-ray energy and photon flux were 12.659 keV and 1.3 × 1012 photons/s, respectively. Among the collected data, 20 images were randomly extracted, and background scattering was analyzed using ADXV (https://www.scripps.edu/tainer/arvai/adxv.html).
Supplementary Information
Acknowledgements
We thank the staff at the 11C beamline at Pohang Accelerator Laboratory for their assistance with data collection. The authors thank the Global Science experimental Data hub Center (GSDC) at Korea Institute of Science and Technology Information (KISTI) for providing computational support. This work was funded by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MOE) (NRF-2017M3A9F6029736).
Author contributions
K.H.N. and Y.C. conceived the experiments. K.H.N. and J.K. conducted the experiments. K.H.N. wrote the manuscript. All authors have revised the manuscript.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
The online version contains supplementary material available at 10.1038/s41598-021-92687-x.
References
- 1.Martin-Garcia JM, Conrad CE, Coe J, Roy-Chowdhury S, Fromme P. Serial femtosecond crystallography: A revolution in structural biology. Arch. Biochem. Biophys. 2016;602:32–47. doi: 10.1016/j.abb.2016.03.036. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Schlichting I. Serial femtosecond crystallography: The first five years. IUCrJ. 2015;2:246–255. doi: 10.1107/S205225251402702X. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Nam KH. Approach of serial crystallography. Crystals. 2020;10:854. doi: 10.3390/cryst10100854. [DOI] [Google Scholar]
- 4.Weinert T, et al. Serial millisecond crystallography for routine room-temperature structure determination at synchrotrons. Nat. Commun. 2017;8:542. doi: 10.1038/s41467-017-00630-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Stellato F, et al. Room-temperature macromolecular serial crystallography using synchrotron radiation. IUCrJ. 2014;1:204–212. doi: 10.1107/S2052252514010070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Nam KH. Room-temperature structure of xylitol-bound glucose isomerase by serial crystallography: Xylitol binding in the M1 site induces release of metal bound in the M2 site. Int. J. Mol. Sci. 2021;22:3892. doi: 10.3390/ijms22083892. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Mehrabi P, et al. Liquid application method for time-resolved analyses by serial synchrotron crystallography. Nat. Methods. 2019;16:979–982. doi: 10.1038/s41592-019-0553-1. [DOI] [PubMed] [Google Scholar]
- 8.Schmidt M. Time-resolved macromolecular crystallography at pulsed X-ray sources. Int. J. Mol. Sci. 2019;20:1401. doi: 10.3390/ijms20061401. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Nam KH. Molecular dynamics—From small molecules to macromolecules. Int. J. Mol. Sci. 2021;22:3761. doi: 10.3390/ijms22073761. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Martiel I, Muller-Werkmeister HM, Cohen AE. Strategies for sample delivery for femtosecond crystallography. Acta Crystallogr. D Struct. Biol. 2019;75:160–177. doi: 10.1107/S2059798318017953. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.DePonte DP, et al. Gas dynamic virtual nozzle for generation of microscopic droplet streams. J. Phys. D: Appl. Phys. 2008;41:195505. doi: 10.1088/0022-3727/41/19/195505. [DOI] [Google Scholar]
- 12.Weierstall U, et al. Lipidic cubic phase injector facilitates membrane protein serial femtosecond crystallography. Nat. Commun. 2014;5:3309. doi: 10.1038/ncomms4309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Roessler CG, et al. Acoustic injectors for drop-on-demand serial femtosecond crystallography. Structure. 2016;24:631–640. doi: 10.1016/j.str.2016.02.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Sugahara M, et al. Grease matrix as a versatile carrier of proteins for serial crystallography. Nat. Methods. 2015;12:61–63. doi: 10.1038/nmeth.3172. [DOI] [PubMed] [Google Scholar]
- 15.Park SY, Nam KH. Sample delivery using viscous media, a syringe and a syringe pump for serial crystallography. J. Synchrotron Radiat. 2019;26:1815–1819. doi: 10.1107/S160057751900897X. [DOI] [PubMed] [Google Scholar]
- 16.Conrad CE, et al. A novel inert crystal delivery medium for serial femtosecond crystallography. IUCrJ. 2015;2:421–430. doi: 10.1107/S2052252515009811. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Nam KH. Sample delivery media for serial crystallography. Int. J. Mol. Sci. 2019;20:1094. doi: 10.3390/ijms20051094. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Park J, et al. Polyacrylamide injection matrix for serial femtosecond crystallography. Sci. Rep. 2019;9:2525. doi: 10.1038/s41598-019-39020-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Nam KH. Shortening injection matrix for serial crystallography. Sci. Rep. 2020;10:107. doi: 10.1038/s41598-019-56135-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Nam KH. Polysaccharide-based injection matrix for serial crystallography. Int. J. Mol. Sci. 2020;21:3332. doi: 10.3390/ijms21093332. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Nam KH. Lard injection matrix for serial crystallography. Int. J. Mol. Sci. 2020;21:5977. doi: 10.3390/ijms21175977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Hunter MS, et al. Fixed-target protein serial microcrystallography with an x-ray free electron laser. Sci. Rep. 2014;4:6026. doi: 10.1038/srep06026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Coquelle N, et al. Raster-scanning serial protein crystallography using micro- and nano-focused synchrotron beams. Acta Crystallogr. D Biol. 2015;71:1184–1196. doi: 10.1107/S1399004715004514. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Roedig P, et al. High-speed fixed-target serial virus crystallography. Nat. Methods. 2017;14:805–810. doi: 10.1038/nmeth.4335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Owen RL, et al. Low-dose fixed-target serial synchrotron crystallography. Acta Crystallogr. D Biol. Crystallogr. 2017;73:373–378. doi: 10.1107/S2059798317002996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Lee D, et al. Nylon mesh-based sample holder for fixed-target serial femtosecond crystallography. Sci. Rep. 2019;9:6971. doi: 10.1038/s41598-019-43485-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Lee K, et al. Viscous-medium-based crystal support in a sample holder for fixed-target serial femtosecond crystallography. J. Appl. Crystallogr. 2020;53:1051–1059. doi: 10.1107/S1600576720008663. [DOI] [Google Scholar]
- 28.Park S-Y, Choi H, Eo C, Cho Y, Nam KH. Fixed-target serial synchrotron crystallography using nylon mesh and enclosed film-based sample holder. Crystals. 2020;10:803. doi: 10.3390/cryst10090803. [DOI] [Google Scholar]
- 29.Monteiro DCF, et al. A microfluidic flow-focusing device for low sample consumption serial synchrotron crystallography experiments in liquid flow. J. Synchrotron. Radiat. 2019;26:406–412. doi: 10.1107/S1600577519000304. [DOI] [PubMed] [Google Scholar]
- 30.Nam KH. Stable sample delivery in viscous media via a capillary for serial crystallography. J. Appl. Crystallogr. 2020;53:45–50. doi: 10.1107/S1600576719014985. [DOI] [Google Scholar]
- 31.Beyerlein KR, et al. Mix-and-diffuse serial synchrotron crystallography. IUCrJ. 2017;4:769–777. doi: 10.1107/S2052252517013124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Murray TD, et al. A high-transparency, micro-patternable chip for X-ray diffraction analysis of microcrystals under native growth conditions. Acta Crystallogr. D Biol. Crystallogr. 2015;71:1987–1997. doi: 10.1107/S1399004715015011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Sui S, et al. Graphene-based microfluidics for serial crystallography. Lab Chip. 2016;16:3082–3096. doi: 10.1039/C6LC00451B. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Doak RB, et al. Crystallography on a chip—without the chip: Sheet-on-sheet sandwich. Acta Crystallogr. D Struct. Biol. 2018;74:1000–1007. doi: 10.1107/S2059798318011634. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Lee D, et al. Application of a high-throughput microcrystal delivery system to serial femtosecond crystallography. J. Appl. Crystallogr. 2020;53:477–485. doi: 10.1107/S1600576720002423. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Wierman JL, et al. Fixed-target serial oscillation crystallography at room temperature. IUCrJ. 2019;6:305–316. doi: 10.1107/S2052252519001453. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Lobley CM, et al. A generic protocol for protein crystal dehydration using the HC1b humidity controller. Acta Crystallogr. D Struct. Biol. 2016;72:629–640. doi: 10.1107/S2059798316003065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.de la Mora E, et al. Radiation damage and dose limits in serial synchrotron crystallography at cryo- and room temperatures. Proc. Natl. Acad. Sci. USA. 2020;117:4142–4151. doi: 10.1073/pnas.1821522117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Thorne RE, Stum Z, Kmetko J, O'Neill K, Gillilan R. Microfabricated mounts for high-throughput macromolecular cryocrystallography. J. Appl. Crystallogr. 2003;36:1455–1460. doi: 10.1107/S0021889803018375. [DOI] [Google Scholar]
- 40.Bury CS, Brooks-Bartlett JC, Walsh SP, Garman EF. Estimate your dose: RADDOSE-3D. Protein Sci. 2018;27:217–228. doi: 10.1002/pro.3302. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Barty A, et al. Cheetah: Software for high-throughput reduction and analysis of serial femtosecond X-ray diffraction data. J. Appl. Crystallogr. 2014;47:1118–1131. doi: 10.1107/S1600576714007626. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.White TA, et al. Recent developments in CrystFEL. J. Appl. Crystallogr. 2016;49:680–689. doi: 10.1107/S1600576716004751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Liebschner D, et al. Macromolecular structure determination using X-rays, neutrons and electrons: Recent developments in Phenix. Acta Crystallogr. D Biol. Crystallogr. 2019;75:861–877. doi: 10.1107/S2059798319011471. [DOI] [PMC free article] [PubMed] [Google Scholar]