Abstract
Fluidic larger-scale integration (LSI) resides at the heart of comprehensive sample-to-answer automation and parallelization of assay panels for frequent and ubiquitous bioanalytical testing in decentralized point-of-use/point-of-care settings. This paper develops a novel “digital twin” strategy with an emphasis on rotational, centrifugo-pneumatic flow control. The underlying model systematically connects retention rates of rotationally actuated valves as a key element of LSI to experimental input parameters; for the first time, the concept of band widths in frequency space as the decisive quantity characterizing operational robustness is introduced, a set of quantitative performance metrics guiding algorithmic optimization of disc layouts is defined, and the engineering principles of advanced, logical flow control and timing are elucidated. Overall, the digital twin enables efficient design for automating multiplexed bioassay protocols on such “Lab-on-a-Disc” (LoaD) systems featuring high packing density, reliability, configurability, modularity, and manufacturability to eventually minimize cost, time, and risk of development and production.
Keywords: centrifugal microfluidics, Lab-on-a-Disc, centrifugo-pneumatic flow control, integration, sample-to-answer automation, multiplexing, parallelization, reliability, tolerances, design-for-manufacture, digital twin, event triggering, timing
1. Introduction
There is a strong market need for providing rather complex bioanalytical testing capabilities that are currently carried out by well-trained personnel in professional facilities, for manifold decentralized use cases. Main applications are found in biomedical point-of-care and global diagnostics, liquid handling automation for the life sciences, process analytical techniques, and cell line development for biopharma, as well as monitoring the environment, infrastructure, industrial processes, and agrifood. Since the early 1990s, microfluidic “Lab-on-a-Chip”, also—now quite synonymously—referred to as Micro Total Analysis Systems (“µTAS”), have been proposed to cater for these strong demands [1,2,3,4,5].
Since these pioneering days, centrifugal microfluidic technologies have played a significant role in the scientific community [6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28,29], and, even more, in commercially directed initiatives [30,31,32,33,34,35,36,37,38,39,40,41,42]; as common in pioneering technologies, many of these companies have experienced significant success, while others were absorbed or discontinued in the meantime. The development of such “Lab-on-a-Disc” (LoaD) systems involves the well-coordinated development of fluidic, manufacturing, instrumentation, software, detection, and, last but not least, bioassay capabilities. The latter two elements of the LoaD technology stack are often closely derived from established reagent kits, and optical or electrochemical sensing techniques. However, the degrees of freedom for fluidic layouts are significantly restricted, chiefly owing to the typical, single-use character of the carrier chip as the multi-branched channel networks can typically not be regenerated.
To still mitigate the cost, the disposable chips are usually produced in polymer materials by (compression-)injection molding, which is a well-established technique in optical data storage manufacturing, and thus often adopt geometries such as Compact DiscTM (CD), Digital Video Disc (DVD), or Blu-ray formats (or sections thereof) with their standard 12 cm diameter, thickness of about 1.2 mm, and a central, 15 mm diameter hole for coupling to the spindle. The disc shape is also commensurate with the rotational symmetry of the centrifugal field. While centrifugal microfluidic devices have been implemented on many other shapes, e.g., circular segments, semicircle, standard (rectangular) microscope slides, and even tubes, for the sake of this work, the popular term “disc” will represent all types of chip formats attached to a rotor.
Similar to a CD player or Discman, LoaD systems usually feature a modular setup with a rather rugged, possibly portable, and widely autonomous instrument taking up the fluidic carrier chip, which is equipped with the spindle motor, control, detection, display, and communication units. For immersion with most point-of-use scenarios outside professional laboratory infrastructure, all (liquid or dry) reagents need to be (pre-)stored on the chip so that only the (bio-)sample needs to be loaded by the end-user.
While the physical representation, i.e., pits and lands, on optical storage media are of micron- and sub-micron size, the channels and chamber in microfluidic systems commonly need to handle volumes in the upper microliter range, at least on the inlet side, corresponding to milli- to centimeter lateral dimensions. Such multi-scale manufacture of the low-aspect-ratio discs imposes significant challenges on manufacturing and assembly, which need to be addressed separately for each fluidic design and fabrication technology. Furthermore, as with other conventional Lab-on-a-Chip technologies, the development of LoaD applications goes through several stages, often starting from prototyping for initial proof-of-concept, pilot series, e.g., to establish sufficient statistics for fluidic and bioanalytical validation, and substantial optimization towards commercially viable mass production.
Each fabrication scheme, along with the scale-up, e.g., precision milling in prototyping and subsequent tool-based replication, involves restrictions on geometries, manufacturing tolerances, artefacts, and often entails considerable efforts for setting up the process [43]. Typically, any change in the fluidic layout, materials, manufacturing processes, detection concepts, instrumentation, software, reagents, and assay protocol may induce mutual interference. Consequently, diligently following Design-for-Manufacture (DfM) guidelines and proper interface management are thus key for reaching high Technology Readiness Levels (TRLs) within reasonable techno-economical boundary conditions.
Most centrifugal microfluidic platforms operate in a batch-wise “stop-and-go” fashion, meaning that Laboratory Unit Operations (LUOs) for sample preconditioning, such as metering/aliquoting [44,45,46], removal of bioparticles [29,47,48,49,50,51], incubation, purification/concentration/extraction [52,53], homogenization [54,55], resuspension of dry, and mixing with liquid reagents [56,57,58,59] are gated by normally closed valves. Upon completion of each LUO, the liquid is forwarded to the next step. These centrifugal LUOs and their connected downstream detection methods have been surveyed before [7,8,12,19,20,21,22,23,24,25,26,27,60,61,62,63,64,65,66,67].
Techniques for such valving can be broadly distinguished by their retention and release mechanisms [68,69]. Initial LoaD concepts were mainly based on siphoning [31,32] and capillary burst valves [6], typically requiring the precise definition of geometrical features or surface coatings. Concepts based on sacrificial materials, for instance, wax [70,71,72,73], membranes [74], or stick-packaging [75], provide a liquid and vapor barrier, thus making them suitable for longer-term onboard storage of liquid buffers or reagents; yet, they either require an externally powered unit [67,76] or very high, often poorly defined release rates. Rudimentary centrifugo-pneumatic (CP) flow control mechanisms were introduced [77,78], which tend to require large, rotationally induced pressure heads for opening; therefore, they are primarily suitable for valving the penultimate LUO in a liquid handling sequence.
Subsequent work refined the release mechanism of these CP valving techniques by venting their compression chamber. Other than removing the seal by instrument-based units such as lasers [74] or xurography [79], (water) dissolvable films (DFs) covering outlets that are pneumatically coupled to the compression chamber were opened by ancillary liquids [80]. With this strategy, complex, highly multiplexed networks offering “handshake” as well as logical flow control, such as AND, OR, and IF conditionals, could be developed, which operate widely independently of the spin protocol [49,81,82,83,84].
This paper investigates key design features gearing centrifugo-pneumatically coupled valving networks towards larger-scale integration (LSI) of parallelized, multi-step bioassay panels. Note that the notion of fluidic LSI has previously been used by Quake et al. [85,86] in a different context of a pneumatically multiplexed liquid handling on a stationary, elastomeric platform, which is to be distinguished from the centrifugal technology be presented here. For the first time, the approach presented here also factors in experimental tolerances to deliver high predictability for batch-mode operated microfluidic LoaD platforms [68]. In pursuit of this novel objective, a “digital twin” [87,88], i.e., a model-based representation that serves as the virtual counterpart of a physical object or process [68,87,88], is developed for optimizing the fluidic performance, robustness, packing density, and manufacturability of rotationally controlled valving schemes for LoaD platforms.
First, the basic principles of centrifugal microfluidic liquid handling are briefly reviewed to establish the mathematical framework of the digital twin. Then a set of performance metrics is introduced along which fluidic layouts can be systematically optimized for given design objectives. The next section elaborates on the fluidic multiplexing in the context of real space and band widths in the frequency domain induced by standard deviations experimental input parameters. The method is illustrated by an exemplary bioassay. In the next section, the digital twin concept is applied to advanced flow control schemes allowing enhanced multiplexing and flexible timing before concluding.
2. Basics of Rotational Flow Control
2.1. Centrifugal Field
According to classical mechanics, an object of (uniform) density and volume with a center of mass located at a distance R from the axis of rotation experiences a centrifugal field density
(1) |
at the angular spin rate . Note that in typical microfluidic scenarios, a (bio-) particle suspended in a liquid of density experiences buoyancy, so the in (1) would refer to the difference between the densities of the object and its surrounding medium.
While not relevant to the immediate context of this paper focusing on valving, the rotationally induced Coriolis (pseudo) force acts on objects traveling at a speed in the non-inertial frame spinning at . In common LoaD systems, is directed opposite to the direction of motion in the plane of the disc, and has been used for agitating advective mixing [57,59] and routing [89].
2.2. Pressure Contributions
Rather than forces, the response of liquids as continuous media to a field (1) is better captured by pressures , i.e., force per unit (cross-sectional) area . In the context of the essentially hydrostatic approximation underlying the mainstay of this paper, flow-induced effects relating to the viscosity of the liquid are neglected. A valve is represented by a geometrical structure which is composed of chambers and interconnecting channels holding a liquid of total volume .
On LoaD systems, the distribution of the (contiguous) liquid within forms under the impact of the centrifugal pressure head
(2) |
scaling with the square of the spin rate and the radial product
(3) |
composed of its mean radial position
(4) |
and its radial extension
(5) |
which is (radially) confined by its inner and outer menisci and , respectively. For typical values , , and , spin rates and , rotationally induced pressure heads (2) of roughly and are obtained, the latter corresponding to about of the standard atmospheric pressure .
The net counterpressure in the axial -direction opposing can be compounded to with (the gradients of) the pressures pointing parallel and opposite to the main (axial) direction of the channel. Based on the hydrostatic equilibrium
(6) |
and the continuity of liquid volume
(7) |
the two (radial) positions and of its confining menisci, and thus the distribution throughout , can hence, either algebraically or numerically, be calculated.
The contributions and might be generated externally [11,74,76,83,90,91,92,93], e.g., by compressed air bottles, onboard pumps, or mechanical valves (corresponding to ); for more common, rotationally controlled valving mainly investigated in the mainstay of this work, pneumatic pressures
(8) |
linked to the compression of a gas volume enclosed by the liquid from their original size at the ambient atmospheric pressure to , e.g., by mechanical work or temperature change. Through shrinking , (8) may, theoretically, be made randomly large.
Capillary action is associated with a pressure
(9) |
governed by the diameter of the channel (with a round cross section) and the contact angle of its inner wall with a liquid of surface tension . have been implemented. Typical values , and yield (9), which corresponds to rather low spin rates in the regime of .
2.3. Angular Acceleration
Certain LUOs, e.g., for mixing by chaotic advection [56] and or mechanical cell lysis [94] with obstacles or (magnetic) beads, benefit from rotational accelerations . Assuming a solid disc of mass and radius , and thus a moment of inertia and resulting angular momentum , the steepness of the ramps is limited by the maximum torque of the spindle motor. Such angular accelerations might be approximated by an inertially induced pressure term
(10) |
for a fluid volume of density traveling at a speed at a radial position through a channel of cross-section .
Note that LoaD systems feature cavities, i.e., channels and chambers, through which the liquids move along the assay protocol; the moment of inertia thus increases over time with the centrifugally outbound motion of (the center of mass of) the liquid distribution
2.4. Critical Spin Rate
The common, normally closed valving schemes of LoaD systems can be described by a first retention phase when the outer meniscus of at is retained upstream of a critical point in . Subsequently, the rotationally induced release of (part of) the liquid volume to an outlet is prompted through shifting the spin protocol across a critical frequency . For spin-rate independent axial pressure heads and , and thus also , we derive a critical frequency
(11) |
by combining (2) and (6). High- and low-pass valves distinguish by whether the release is triggered by lifting or lowering across (11), respectively.
As an alternative to modulating , the hydrostatic equilibrium (6) may also be altered by adding a liquid volume to the inner end of for elevating (and possibly also ) in (2). Furthermore, in centrifugo-pneumatic valves delineated by and (8), the outer compression chamber may be enlarged or vented, i.e., , for prompting liquid release.
Note that for highly dynamic systems, inertial effects will have to be included in calculating the frequency (11), e.g., by introducing an inertial pressure term (10) in , which would thus lower the effective retention rate compared to the quasi-hydrostatic case mainly considered in the course of this work.
2.5. Siphon Valving
From the early days of LoaD systems, siphoning principles have been utilized for centrifugal flow control. In these geometries, , a (usually larger) inlet reservoir is connected by an outlet channel featuring inbound and outbound segments with an inner crest point at to a receiving chamber. During retention, the front meniscus at resides in the inbound segment where the centrifugal pressure (2) points antiparallel to the direction of , thus effectively stabilizing the retention of the distribution against perturbations, such as inertial effects or droplet break-off. This phase is characterized by a between the front meniscus in the inbound section and the liquid level in the inlet reservoir at , so that (3). The axial direction of determines the sign of during retention of , i.e., whether is elevated or depressed with respect to .
In many cases, siphons connect two vented (or very large) chambers, thus neutralizing the impact of the ambient pressure applying at either end. In macroscopic setups, priming of the siphon channel is then triggered by the addition of a sufficient volume to the inlet for the liquid distribution to progress (sufficiently) past the crest point at . While such release mechanisms can also be deployed for miniaturized systems, LoaD technologies have mainly resorted to priming by capillary action, i.e., (9), energized by a hydrophilically coated siphon outlet with . Furthermore, pneumatic pumping through compression of a volume in a side chamber connected to the inlet has been explored [20,28]. In both of these low-pass constructs, either an overflow mechanism, or a continuous liquid “pulley” effect [95] sets in once the condition is met between the menisci in the inner reservoir and the outbound segment; at this stage, also (2) works in tandem with to drag the liquid radially outwards, thus further increasing in a self-amplifying pumping mechanism.
In centrifugo-pneumatic (CP) siphon valving (Figure 1), and (8), i.e., , so that is needed to drive the liquid past the inner crest point at all the way to the outlet chamber against shrinking compression volume , and thus growing (8). Instead of a self-propelling pulley effect promoted by a narrow outlet, advancement of the outbound meniscus requires a steady increase of the spin rate , and, as long as there is a positive , the centrifugal pressure (2) can, at least mathematically, match any counteracting (8).
For a sufficiently narrow outlet channel, the integrity of the liquid plug might be preserved by the surface tension along the entire path until the front meniscus at reaches the entrance of the outer chamber at the radial position ; thereafter, the liquid is transferred. However, for larger cross-sections of the outbound segment of the siphon and high spin rates, the liquid will creep or drip after passing the crest segment into the outer chamber until , rather than showing a quasi-solid, piston-like behavior. There are thus critical frequencies and (11) characterizing the retention and release into the outer chamber, which depend on whether the liquid plug is cut at or . In either case, a volume remains in the upstream part of the valve.
Note that the critical frequency (11) of CP valves, i.e., their basic as well as their siphon variants, can be widely configured by the original compression volume ; its permanently gas-filled parts can be located at “anywhere”, e.g., where disc real estate is still available, if the fractional volume of its sub-compartments remains pneumatically interconnected.
2.6. Centrifugo-Pneumatic Siphon Valving with Dissolvable-Film Membranes
In order to supersede the prerequisite for external actuation to induce venting, CP valves have integrated sacrificial, fluidically impermeable [51] membranes in their final receiving chamber (Figure 1); these seal the compression chamber until they come into contact with their specific solvent, usually an aqueous bioliquid. The opening of these dissolvable films (DFs) requires the transfer of sufficient liquid to raise the filling level above its location.
3. Performance Metrics
In this section, quantitative characteristics for rotationally controlled valving are specified to play a pivotal role in fluidic LSI requiring high packing density that is tightly interwoven with system-level reliability. These metrics are run along with the example of CP-DF siphon valving (Figure 1), but can also be readily employed for further valve types.
3.1. High Field Strengths for LUOs
LUOs, such as plasma separation, become most efficient at high field strengths (1). Given a medium of (differential) density retained by a valve of critical spin rate (11) at a given position , we obtain
(12) |
for the maximum field strength available for the LUO. As shifting in (12) mostly impacts in the radial product (11), the primary parameter towards maximizing (12) for a given retention rate (11) is , i.e., the liquid distribution extending between and , which can be trimmed by the geometry for a given . The dimensionless metric
(13) |
references (12) to the largest available field strength at the maximum radial position and (safe) rotational frequency that can be produced by the spindle motor.
3.2. Radial Space
Due to the unidirectional nature of the centrifugal field (1), (the center of gravity of) a liquid distribution must always advance in the radially outbound direction. At least in the absence of externally powered centripetal pumping, the processing chambers of subsequent LUOs must thus be staggered radially.
As part of such serial arrangement, the outwards pointing field (1) puts a price tag on the radial position
(14) |
echoing that a central placement is most favorable, i.e., (14) minimized, to keep sufficient room for downstream LUOs. Note that small (14) might interfere with requirements on maximizing the local field strength (13) for LUOs. Further, the overall consumption of radial space
(15) |
with the inner and outer boundaries and of the structure , respectively, should be taxed towards an optimization of the packing density.
3.3. Spatial Footprint
The real estate available for patterning on a discoid carrier is limited, e.g., between outside the central hole and to leave an outer rim for bonding a lid, e.g., with a radial extension of 5 mm, for chip geometries derived from conventional optical data storage formats. This makes central real estate, where space is the smallest and most precious.
If we consider a function of the full (azimuthal) width of along the radial direction , a simple measure of the spatial footprint would be the entire surface area enclosed by its outer contours confined by and . Considering the scarcity of central space, we introduce the metric
(16) |
which relates the spatial footprint of at with respect to the total area contained within the ring of (infinitesimal) radial extension at a radial location .
3.4. Volume Definition and Loss
Liquid transfer through an LUO and its connected downstream valve may be associated with a volume left behind in the structure after completion. Depending on the fluidic process of emptying a CP-DF siphon valve, the lost amount of liquid
(17) |
might vanish for a perfect “pulley” mechanism, which maintains a continuous liquid column until the entire liquid in is dragged through the crest channel, and in a pure “overspill” mode until (Figure 1d). For serial processing, the volume arriving at the next valve can be factored into its retention rate , and also into the mixing ratio of a quantitative bioassay.
The metric
(18) |
is useful to quantify volume loss with respect to the loaded volume . For siphoning, where a significant fraction of the liquid remains in , may be minimized through reduction of and . As any uncertainty directly impacts the spread of the retention rate , minimizing the relative fluctuation
(19) |
is critical for assuring reliability towards fluidic LSI.
3.5. Reliability and Band Width
In practice, all experimental input parameters impacting the liquid distribution , and thus (11), are subject to random spreads of standard deviations delineated by a Gaussian distribution. This statistical spread implies that at a given spin speed , the actual (radial) position of the meniscus will statistically vary around its target value . So, to make sure that the conditions for operationally reliable flow control are guaranteed, e.g., during retention, the spin rate has to stay outside the corridor ; in other words, has to be lifted or lowered through this entire frequency band for assuring proper retention and release for high- and low-pass valves, respectively.
The factor represents the degree of (component-level) reliability, with probabilities (and “erf” representing the error function) that the actual values are found in this frequency interval. Above and , reliability of this (single) valving step is in the range of 10 and 1 defects per million opportunities (DPMO); for , faults are practically absent.
At least in case that the input parameters are (widely) independent and their spreads comparatively small, Gaussian error propagation provides an analytical formula to calculate the standard deviation
(20) |
of (11), which is a very suitable method for optimization routines. Figure 2 reveals the scaling of and its standard deviation with three impact parameters defined in Figure 1. For a given , the volume of the main compression compartment (a) and the radial position (b) are varied; in (c), the field strength . (1) is kept constant while shifted, e.g., to mimic translocating a particle sedimentation structure requiring a minimum to a radial position where space is still available in a multiplexed layout.
Alternatively, Monte Carlo (MC) methods can be interpreted as virtual manufacturing and testing by running a large number of iterations, where a set of values is obtained from a random number generator assuming Gaussian distributions of of standard deviations around their nominal values and inserted into Equation (11) for obtaining a histogram for in the simulated experiment.
The resulting band widths (20), which are associated with concurrently loaded valves, and thus change over the course of an LUO sequence, need to “fit” into the practically available envelope of spin rates between and; these boundaries are, on the lower end at , constrained by a minimum centrifugal field (1) or pressure (2) required to stabilize a liquid distribution , e.g., to suppress unwanted capillary motion (9); the upper value at is imposed by factors such as the limited torque of the spindle motor (10), the pressure tightness of the (bonded) lid, and operational safety.
Towards multiplexing, the metric
(21) |
rewards minimum use of band width (20) with respect to the available frequency envelope.
4. Multiplexing
4.1. System-Level Robustness
Towards microfluidic LSI, many such serial and parallelized valving steps need to be coordinated on the LoaD device; each component failure might affect the overall reliability. For independently operating, concurrently loaded valves in a given step , the probability for consistent system-level function amounts to ; for the example of , and thus , drops from approximately 90% to 86%, 77% and 60% for , respectively, in case the bands must be avoided for properly controlling each valve . For loaded valves, extending the protected frequency intervals from to would significantly improve the (fluidic) reliability at the system level from to .
Note that the above considerations apply to each step separately, i.e., after opening a particular valve of the cohort at a point in time , a new group of loaded valves emerges; this fresh set consists of the valves that remained closed in step , plus the ones next in line to receive the released liquid, excluding “unvalved”, e.g., end-point chambers. The protected areas in the frequency domain thus change over the course of the centrifugally multiplexed assay protocol .
4.2. Frequency Corridor
We consider serial valving in steps with critical spin rates that are activated at points in time ; each controls a subset of concurrently loaded and opened valves . Retention is assured while the spin rate does neither surpass nor drop below the (aggregate) band widths centered at the critical frequencies and where for high- and low-pass valves, respectively. Figure 3a displays the scenario in frequency space for rotational actuation. An initial retention corridor between and successively broadens as further valves are triggered, with and .
For the case of actuating (high-pass) CP valves by venting at with in the exemplary scenario portrayed in Figure 3b, the retention zone for the spin rate curve is also defined by the innermost limits of the allowed bands of the valves still loaded at a given point in time Nevertheless, the order of release is merely imposed by the venting sequence, and does hence not necessitate steadily rising or decreasing critical retention rates and at , respectively.
4.3. Configuration in Real and Frequency Space
Theoretically, the geometry can always be modified to minimize the radial product (3), thus allowing any randomly large retention rate (11) for a given radial position and liquid volume . Yet, its standard deviation (20) might become too broad to still be able to sufficiently multiplex flow control within the confinement of the practical spin rate envelope between and (Figure 2 and Figure 3). Furthermore, there would be additional restrictions to parameter spaces, e.g., on the smallest feature sizes and tolerances (20) associated with the manufacturing technology, or minimum liquid volumes, for instance, as prescribed by the loading and metering technique, the bioassay protocol, or the detection technology.
It also needs to be considered that, especially towards concentrated packing of LoaD substrates, it is wise to leave the liquid carrying parts of geometries , that may have already been thoroughly validated in terms of design-for-manufacture (DfM), fluidic and assay performance, untouched. For CP valving techniques, it is therefore advantageous to only alter larger, uncritical features, e.g., the main volume of the permanently gas-filled part of the compression chamber (Figure 2a), for tuning (11). Different may simply be accommodated by adjusting the width of the inlet reservoir (Figure 1). Nonetheless, the paucity of central disc space may necessitate shifting the radial location , while the retention rate (11) still needs to comply with the planned valving sequence (Figure 4).
Figure 4 shows distinctive scenarios for CP-DF siphon valves commonly encountered when seeking the best compromise in real and frequency space. We again investigate simultaneous actuation of valves at times {}, and sequential firing with isoradial as well as radially inbound and outbound staggering on the given disc. The ordering of (11) is accomplished by adjusting the main compression volume , while leaving all other parts of and untouched. Note, again, that without non-rotationally powered pumping mechanisms, (the center of gravity of) a given liquid distribution can only migrate radially outbound, i.e., .
Figure 5 displays the distribution of the frequency bands in the different rotational actuation scenarios for simultaneous and sequential release (Figure 4) through adjusting the main volume of the compression chamber . Note that the requirement to create non-overlapping bands for serial opening even leads to a slight increase in towards the highest spin rate (Figure 5b), and a huge widening of frequency bands in the radially outbound release sequence (Figure 5d).
4.4. Exemplary Bioassay Panel
Figure 6 illustrates the rotational automation of a typical bioassay protocol. Plasma P is separated from a (blood) sample S before mixing with a first, onboard stored liquid reagent L1. This mixture P&L1 is then released to a final detection chamber, where it is successively combined with reagents L2 and L3. The liquid handling is coordinated by high-pass valves opening after their respective bands are crossed, thus creating an - corridor with a successively growing upper limit over time . Intermediate mixing steps are implemented by zig-zag patterns in . The domains available in real and frequency space are confined by and , and and , respectively.
The unidirectional nature of the centrifugal field (1) requires that the positions of the serial processing of the sample through loading, separation, and detection chambers need to follow , while the release rates of the high-pass valves need to continuously increase along , so that overlap of their associated bands is avoided. In the portrayed exemplary assay protocol, high quality and speedy plasma separation is linked to a large field strength (1), which is limited by (11), thus making an outer position most favorable. On the other hand (Figure 2a), high retention rates (11) for the onboard reagents within the narrow corridor are best achieved for their more central placement (when estimating and ).
Overall, the structures presenting LUOs with their downstream valves need to meet several, partially contradictory demands on the maximum retention rate (11), disc real estate (16), maximum field strength (12), and footprints in radial (14), (15), and frequency space (21).
Figure 7 shows the release rates for given and (at ) of CP-DF siphon valves with the same default geometries , except for their (main) compression volumes , which are varied to guarantee disjunct tolerance intervals for a given . The two distributions represent sets of identical with their radial positions shifted by as a possible way to increase the functional integration density on a given disc cartridge. Note that the rotational actuation of these high-pass valves “squeezes” the release rates for the onboard reagents in the narrow strip within (see also Figure 6b).
4.5. Assay Parallelization
A straight-forward method to replicate several simultaneously processed, identical assay protocols on the same LoaD carrier, e.g., for testing different samples for the same analytes, or for redundancies for controls, calibration, or statistics, is to set equal release frequencies , which are crossed synchronously . The packing density might be augmented by the radial staggering of the structures , for which Figure 2 shows the required adjustments of the main compression volumes , and their impact of the total band width . Concurrent processing of homologous assay protocols with the (primary) variation of inlet volumes may be readily achieved by adjusting the cross-sections of the inlet reservoirs (Figure 1). In the general case of diverging multiplexing protocols, each release step must comply with the frequency corridor required to reliably shut the concurrently loaded valves.
5. Advanced Rotational Flow Control
We now expand on the actuation mode of CP-DF siphon valves by venting (Figure 1d). Instead of the various mechanical, radiation-induced, or thermal methods to open the compression chamber, a second (ancillary) liquid volume wets a DF placed at a remote location to effectuate . First, a handshake mechanism called “event triggering” is introduced, which also enables fluidic conditionals logical flow control. Afterward, the timing of event triggering, e.g., to comply with mixing or incubation periods of bioassays, through delay elements and rotational pulsing, is outlined.
5.1. Event Triggering
Two DFs are attached to the compression chamber of CP-DF siphon valves (Figure 8). In an exemplary assembly of upper and lower layers displaying the microfluidic conduits sandwiching a sealing sheet containing a set of DFs, the first “Load Film” (LF) initially blocks the outlet for the liquid through a vertical via into a lower layer; a second “Control Film” (CF) is placed in a pneumatically connected, distal compartment [81]. Below a critical spin rate (11), the pneumatic counterpressure (8) prevents the liquid from reaching the LF.
The remote CF can be accessed by another ancillary liquid through a channel, e.g., situated in the lower layer, to open the DF, to thus massively expand the enclosed gas volume . With , the main liquid advances into the chamber containing the LF, which then disintegrates for release through the exit channel.
Event triggering (Figure 8) refers to the constellation when a liquid received from a preceding stage of an assay protocol takes on the role of the ancillary liquid to establish a “handshake” type flow control; instead of rotational cues or external power, the sequential release follows a “domino effect” that proceeds broadly independent of the spin rate . The frequency corridor (Figure 3), and especially the maximum field strength (12) that is available for LUOs along the assay procedure, are thus only restricted by the (minimum) retention rate (11) of concurrently loaded (high-pass) valves, or .
Figure 9 illustrates the benefits of event-triggered flow control regarding configurability and packing density in real and frequency space. As opposed to the implementation of the exemplary bioassays with the rotationally actuated high-pass CP-DF siphon valves (Figure 7), the liquid reagents may all be released with respect to the plasma separation at . Overall, the speed (and quality) of plasma separation in the initial stages of common assay protocols may hence be significantly improved.
5.2. Logical Flow Control
Such an event-triggered relay may also be considered as a logical “IF” condition, in a way that a (subsequent) valve only fires “if” another liquid has arrived to dissolve the CF in a designated pneumatic compartment [81]. This concept has been further refined by placing the CF at critical radial positions (Figure 1a) in a way that a minimum liquid volume must arrive in the DF chamber to dissolve the membrane, and thus vent and open the next valve. Based on such a logical “IF” condition, closed-loop “handshake” flow control has been implemented. Alternatively, the DF in the receiving chamber may be placed at a radial position that is only reached after a minimum amount of liquid has entered from (at least) two incoming channels, thus establishing an “AND” element. In a similar way, a logical “OR” may be defined by positioning the DF so that it is wetted after the arrival of a minimum liquid volume from a defined subset of incoming channels.
5.3. Delay Elements
Bioassay protocols usually require certain time spans, e.g., to account for reaction kinetics between bioentities that are dissolved, suspended, or bound to surfaces on cells and beads of walls. However, event triggering “hard-wires” process intervals in the network of valving structures in a direct “knock-on” fashion. To this end, delay elements have been incorporated into event-triggered fluidic systems for bioassay integration [96]. Examples are hydrodynamic flow resistances imposed by narrow and long channel segments, or paper strips, which have also been utilized for other purposes [15,97]. Through imbibition at defined wicking speed, these timers slow down the transfer of the main or ancillary liquids compared to open channels. While, in theory, also reducing the spin speed during transfer periods would be an option, such action would require rather complex “closed-loop” coordination between monitoring liquid flow on the disc and the frequency of the spindle motor .
5.4. Pulsing of Spin Rate
Commonly, the aggregate time spans for dissolution and liquid transfer range on the second- to minute scale, while bioassays normally require processing times of minutes to an hour. So, while such physical delay elements prolong the intervals available for LUOs, the timing still faces upper limits, and it is also ‘inscribed’ in the layout and material characteristics, such as the dissolution time of the sacrificial membrane or the imbibition speed of the paper strip. In addition, the requirement of the rotational actuation mode to steadily increase the critical spin rates along successive steps managed by high-pass CP valves compromises the flexibility that may be required for enabling microfluidic LSI.
For situations when liquid volumes are synchronously transferred between serial steps at and , the concept of rotational pulsing (Figure 10) has been presented [82]. This flow control mechanism exploits that within the interval , the subsequent valve is not fully loaded with , yet, thus intermittently increasing is retention rate (Figure 10c) to allow jumps in the spin rate that only need to stay inside the corridor required to retain the (other) concurrently loaded liquids; furthermore, the spin rate may be reduced below , thus alleviating the requirement for a steady increase (Figure 7) or constant spin rate (Figure 8 and Figure 9) while sequentially progressing the release steps .
In a similar way as ‘venting’, rotational pulsing allows retaining liquid menisci at sufficiently “safe” spacing at from the DF during retention. It can also be used to synchronize parallel processes, e.g., when the same protocol needs to be run on several patient samples. However, in heterogeneous assay panels where each channel would run a different protocol, the height of the rotational pulse for the CP-DF valve to be released is confined by the lowest critical frequency of the concurrently loaded valves that need to stay closed, i.e., .
6. Conclusions and Outlook
“Lab-on-a-Disc” (LoaD) systems are highly promising candidates for enabling cost-efficient and robust commercial solutions for decentralized testing, especially in low-resource settings and minimally trained operators. Underlying sample-to-answer automation and parallelization of bioassay protocols roots in fluidic large(r)-scale integration (LSI) of Laboratory Unit Operations (LUOs) and their control valves at tight packing densities.
This work has elaborated the basic principles of centrifugal flow control to directly connect experimental parameters such as liquid volumes, geometrical dimensions, and the ambient atmospheric pressure to the retention and release rates of rotationally actuated valves at the pivot of LSI. In the wake of such “digital-twin” modeling, frequency bands are calculated from tolerances in the input variables to quantify operational robustness. As the practically limited real and frequency spaces of LoaD systems severely restrict fluidic LSI, a suite of metrics has been elaborated to algorithmically guide performance optimization. Furthermore, the principles of advanced rotational valving schemes enabling conditional, closed-loop flow control and random timing by “digital” pulsing were systematically explored. With the availability of extended fabrication capabilities that are beyond the reach of most academic laboratories, larger numbers of LoaD devices can be provided for statistically relevant validation and refinement of the digital-twin model.
Future work will expand the design tool to allow for the simulation of mixed layouts synergistically combining various types of high- and low-pass valves, for refined geometries exhibiting rounded edges and contours, for incorporating dynamic fluid effects and bioassay kinetics, for onboard storage and release of reagents, for bead-based multiplexing techniques [24], and for externally powered, onboard pumping and control modules [76]. The degree of multiplexing might be markedly enhanced by multi-disc stack architectures.
Furthermore, manufacturing processes may be virtualized, e.g., through mold flow analysis for commercial polymer replication techniques, which commonly indicates the need for fairly even distributions of fluidic structures and flow resistances (for the centrally injected hot melt), minimum draft angles, and surface roughness to facilitate proper demolding. Design guidelines are also proving to be important for assembly and bonding.
With its high predictive power, the presented work and the suggested follow-ups can enable the computational generation and validation of disc layouts regarding fluidic functionality and compliance with design-for-manufacture (DfM) guidelines and related scale-up of mass fabrication. Such software might then create an effective interface for foundry services as a hallmark of mature industrial supply chains in task-sharing economies. Superior control and knowledge of manufacturing tolerances can also empower anti-counterfeit strategies protecting genuine medical products and their users [98].
Most foundational patents on centrifugal microfluidics that were primarily filed throughout the 1990s and early 2000s have expired, thus creating substantial wiggle room in the intellectual property (IP) space. Apart from classical exploitation, such opportunity might also resonate well with the present highly disruptive socio-economic and technological trends in the technological ecosystem of LoaD, such as Maker/Fab Labs, hackathons, science as a service (SaaS), cloud computing, Machine Learning (ML)/Artificial Intelligence (AI), Big Data, Internet of Things (IoT), and decentralization. Overall, this development may enable novel open platform concepts, e.g., to usher a new era of blockchain-backed participatory research models and crowdsourcing of ideas, expertise, skills, workforce, infrastructure, and equipment may be established [99,100,101,102].
Appendix A. Default Geometry of CP-DF Siphon Valves
The structure , loaded liquid volumes , and radial positions can be varied across a multi-dimensional parameter space, e.g., to tune retention rates of other key performance indicators. Table A1 gives of overview or generic values which can be used to initiate the optimization.
Table A1.
Funding
This research received no external funding.
Conflicts of Interest
The author declares no conflict of interest.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Manz A., Graber N., Widmer H. Miniaturized total chemical analysis systems: A novel concept for chemical sensing. Sensors Actuators B Chem. 1990;1:244–248. doi: 10.1016/0925-4005(90)80209-I. [DOI] [Google Scholar]
- 2.Reyes D.R., Iossifidis D., Auroux P.-A., Manz A. Micro Total Analysis Systems. 1. Introduction, Theory, and Technology. Anal. Chem. 2002;74:2623–2636. doi: 10.1021/ac0202435. [DOI] [PubMed] [Google Scholar]
- 3.Auroux P.-A., Iossifidis D., Reyes D.R., Manz A. Micro Total Analysis Systems. 2. Analytical Standard Operations and Applications. Anal. Chem. 2002;74:2637–2652. doi: 10.1021/ac020239t. [DOI] [PubMed] [Google Scholar]
- 4.Whitesides G.M. The origins and the future of microfluidics. Nat. Cell Biol. 2006;442:368–373. doi: 10.1038/nature05058. [DOI] [PubMed] [Google Scholar]
- 5.Yager P., Edwards T., Fu E., Helton K., Nelson K., Tam M.R., Weigl B.H. Microfluidic diagnostic technologies for global public health. Nat. Cell Biol. 2006;442:412–418. doi: 10.1038/nature05064. [DOI] [PubMed] [Google Scholar]
- 6.Madou M.J., Kellogg G.J. LabCD: A centrifuge-based microfluidic platform for diagnostics; Proceedings of the BiOS ‘98 International Biomedical Optics Symposium; San Jose, CA, USA. 27–29 January 1998; pp. 80–93. [DOI] [Google Scholar]
- 7.Ducrée J., Haeberle S., Lutz S., Pausch S., Von Stetten F., Zengerle R. The centrifugal microfluidic Bio-Disk platform. J. Micromech. Microeng. 2007;17:S103–S115. doi: 10.1088/0960-1317/17/7/S07. [DOI] [Google Scholar]
- 8.Gorkin R., Park J., Siegrist J., Amasia M., Lee B.S., Park J.-M., Kim J., Kim H., Madou M., Cho Y.-K. Centrifugal microfluidics for biomedical applications. Lab Chip. 2010;10:1758–1773. doi: 10.1039/b924109d. [DOI] [PubMed] [Google Scholar]
- 9.Park J.-M., Cho Y.-K., Lee B.-S., Lee J.-G., Ko C. Multifunctional microvalves control by optical illumination on nanoheaters and its application in centrifugal microfluidic devices. Lab Chip. 2007;7:557–564. doi: 10.1039/b616112j. [DOI] [PubMed] [Google Scholar]
- 10.Watts A.S., Urbas A.A., Moschou E., Gavalas V.G., Zoval J.V., Madou M., Bachas L. Centrifugal Microfluidics with Integrated Sensing Microdome Optodes for Multiion Detection. Anal. Chem. 2007;79:8046–8054. doi: 10.1021/ac0709100. [DOI] [PubMed] [Google Scholar]
- 11.Clime L., Geissler M., Veres T., Brassard D. Active pneumatic control of centrifugal microfluidic flows for lab-on-a-chip applications. Lab Chip. 2015;15:2400–2411. doi: 10.1039/C4LC01490A. [DOI] [PubMed] [Google Scholar]
- 12.Strohmeier O., Keller M., Schwemmer F., Zehnle S., Mark D., von Stetten F., Zengerle R., Paust N. Centrifugal microfluidic platforms: Advanced unit operations and applications. Chem. Soc. Rev. 2015;44:6187–6229. doi: 10.1039/C4CS00371C. [DOI] [PubMed] [Google Scholar]
- 13.Thompson B.L., Birch C., Nelson D.A., Li J., Duvall J.A., Le Roux D., Tsuei A.-C., Mills D.L., Root B.E., Landers J.P. A centrifugal microfluidic device with integrated gold leaf electrodes for the electrophoretic separation of DNA. Lab Chip. 2016;16:4569–4580. doi: 10.1039/C6LC00953K. [DOI] [PubMed] [Google Scholar]
- 14.Smith S., Mager D., Perebikovsky A., Shamloo E., Kinahan D., Mishra R., Delgado S.M.T., Kido H., Saha S., Ducrée J., et al. CD-Based Microfluidics for Primary Care in Extreme Point-of-Care Settings. Micromachines. 2016;7:22. doi: 10.3390/mi7020022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Hwang H., Kim S.-H., Kim T.-H., Park J.-K., Cho Y.-K. Paper on a disc: Balancing the capillary-driven flow with a centrifugal force. Lab Chip. 2011;11:3404–3406. doi: 10.1039/c1lc20445a. [DOI] [PubMed] [Google Scholar]
- 16.Kim T.-H., Hwang H., Gorkin R., Madou M., Cho Y.-K. Geometry effects on blood separation rate on a rotating disc. Sens. Actuators B Chem. 2013;178:648–655. doi: 10.1016/j.snb.2013.01.011. [DOI] [Google Scholar]
- 17.Hwang H., Kim H.-H., Cho Y.-K. Elastomeric membrane valves in a disc. Lab Chip. 2011;11:1434–1436. doi: 10.1039/c0lc00658k. [DOI] [PubMed] [Google Scholar]
- 18.Roy E., Stewart G., Mounier M., Malic L., Peytavi R., Clime L., Madou M., Bossinot M., Bergeron M.G., Veres T. From cellular lysis to microarray detection, an integrated thermoplastic elastomer (TPE) point of care Lab on a Disc. Lab Chip. 2014;15:406–416. doi: 10.1039/C4LC00947A. [DOI] [PubMed] [Google Scholar]
- 19.Kong L.X., Perebikovsky A., Moebius J., Kulinsky L., Madou M. Lab-on-a-CD. J. Lab. Autom. 2016;21:323–355. doi: 10.1177/2211068215588456. [DOI] [PubMed] [Google Scholar]
- 20.Hess J.F., Zehnle S., Juelg P., Hutzenlaub T., Zengerle R., Paust N. Review on pneumatic operations in centrifugal microfluidics. Lab Chip. 2019;19:3745–3770. doi: 10.1039/C9LC00441F. [DOI] [PubMed] [Google Scholar]
- 21.Aeinehvand M.M., Magaña P., Aeinehvand M.S., Aguilar O., Madou M.J., Martinez-Chapa S.O. Ultra-rapid and low-cost fabrication of centrifugal microfluidic platforms with active mechanical valves. RSC Adv. 2017;7:55400–55407. doi: 10.1039/C7RA11532F. [DOI] [Google Scholar]
- 22.Aeinehvand M.M., Weber L., Jiménez M., Palermo A., Bauer M., Loeffler F.F., Ibrahim F., Breitling F., Korvink J., Madou M., et al. Elastic reversible valves on centrifugal microfluidic platforms. Lab Chip. 2019;19:1090–1100. doi: 10.1039/C8LC00849C. [DOI] [PubMed] [Google Scholar]
- 23.Van Nguyen H., Nguyen V.D., Nguyen H.Q., Chau T.H.T., Lee E.Y., Seo T.S. Nucleic acid diagnostics on the total integrated lab-on-a-disc for point-of-care testing. Biosens. Bioelectron. 2019;141:111466. doi: 10.1016/j.bios.2019.111466. [DOI] [PubMed] [Google Scholar]
- 24.Rombach M., Hin S., Specht M., Johannsen B., Lüddecke J., Paust N., Zengerle R., Roux L., Sutcliffe T., Peham J.R., et al. RespiDisk: A point-of-care platform for fully automated detection of respiratory tract infection pathogens in clinical samples. Analyst. 2020;145:7040–7047. doi: 10.1039/D0AN01226B. [DOI] [PubMed] [Google Scholar]
- 25.Miyazaki C., Carthy E., Kinahan D. Biosensing on the Centrifugal Microfluidic Lab-on-a-Disc Platform. Processes. 2020;8:1360. doi: 10.3390/pr8111360. [DOI] [Google Scholar]
- 26.Madadelahi M., Acosta-Soto L.F., Hosseini S., Martinez-Chapa S.O., Madou M.J. Mathematical modeling and computational analysis of centrifugal microfluidic platforms: A review. Lab Chip. 2020;20:1318–1357. doi: 10.1039/C9LC00775J. [DOI] [PubMed] [Google Scholar]
- 27.Homann A.R., Niebling L., Zehnle S., Beutler M., Delamotte L., Rothmund M.-C., Czurratis D., Beller K.-D., Zengerle R., Hoffmann H., et al. A microfluidic cartridge for fast and accurate diagnosis of Mycobacterium tuberculosis infections on standard laboratory equipment. Lab Chip. 2021;21:1540–1548. doi: 10.1039/D1LC00035G. [DOI] [PubMed] [Google Scholar]
- 28.Zehnle S., Schwemmer F., Bergmann R., von Stetten F., Zengerle R., Paust N. Pneumatic siphon valving and switching in centrifugal microfluidics controlled by rotational frequency or rotational acceleration. Microfluid. Nanofluidics. 2015;19:1259–1269. doi: 10.1007/s10404-015-1634-9. [DOI] [Google Scholar]
- 29.Zehnle S., Rombach M., Zengerle R., Von Stetten F., Paust N. Network simulation-based optimization of centrifugo-pneumatic blood plasma separation. Biomicrofluidics. 2017;11:024114. doi: 10.1063/1.4979044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Abaxis. [(accessed on 14 June 2021)]; Available online: https://www.abaxis.com/
- 31.Schembri C.T., Ostoich V., Lingane P.J., Burd T.L., Buhl S.N. Portable Simultaneous Multiple Analyte Whole-Blood Analyzer for Point-of-Care Testing. Clin. Chem. 1992;38:1665–1670. doi: 10.1093/clinchem/38.9.1665. [DOI] [PubMed] [Google Scholar]
- 32.Schembri C.T., Burd T.L., Kopf-Sill A.R., Shea L.R., Braynin B. Centrifugation and capillarity integrated into a multiple analyte whole blood analyser. J. Autom. Chem. 1995;17:99–104. doi: 10.1155/S1463924695000174. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Gyros Protein Technologies. [(accessed on 14 June 2021)]; Available online: https://www.gyrosproteintechnologies.com/
- 34.Shea M., Ommert S., Gleich L., Towle T., Haspel H., Schmid N., Jindal H., Kellogg G., Carvalho B., Contarino M., et al. ADMET Assays on Tecan’s LabCD-ADMET System. J. Lab. Autom. 2003;8:74–77. doi: 10.1016/S1535-5535-04-00260-6. [DOI] [Google Scholar]
- 35.Biosurfit S.A. [(accessed on 14 June 2021)]; Available online: https://www.biosurfit.com/
- 36.Burstein Technologies, Inc. [(accessed on 31 May 2021)];2006 Available online: https://web.archive.org/web/20061209052345/http://www.bursteintechnologies.com.
- 37.Mian A., Kieffer-Higgins S.G., Corey G.D. US6319469B1–Devices and methods for using centripetal acceleration to drive fluid movement in a microfluidics system. [(accessed on 14 June 2021)];1996 Available online: https://patents.google.com/patent/US6319469B1/en.
- 38.Radisens Diagnostics. [(accessed on 14 June 2021)]; Available online: www.radisens.com.
- 39.LaMotte Chemical Products Co. [(accessed on 14 June 2021)]; Available online: https://www.lamotte.com.
- 40.Blusense Diagnostics. [(accessed on 14 June 2021)]; Available online: https://blusense-diagnostics.com/
- 41.Spindiag GmbH. [(accessed on 14 June 2021)]; Available online: http://www.spindiag.de.
- 42.RotaPrep Inc. [(accessed on 19 April 2021)]; Available online: https://rotaprep.com/
- 43.Ducrée J. Efficient Development of Integrated Lab-On-A-Chip Systems Featuring Operational Robustness and Manufacturability. Micromachines. 2019;10:886. doi: 10.3390/mi10120886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Mark D., Haeberle S., Metz T., Lutz S., Ducrée J., Zengerle R., von Stetten F. Aliquoting structure for centrifugal micro-fluidics based on a new pneumatic valve; Proceedings of the MEMS 2008, 21st IEEE International Conference on Micro Electro Mechanical Systems; Tucson, AZ, USA. 13–17 January 2008. [Google Scholar]
- 45.Schwemmer F., Hutzenlaub T., Buselmeier D., Paust N., Von Stetten F., Mark D., Zengerle R., Kosse D. Centrifugo-pneumatic multi-liquid aliquoting—Parallel aliquoting and combination of multiple liquids in centrifugal microfluidics. Lab Chip. 2015;15:3250–3258. doi: 10.1039/C5LC00513B. [DOI] [PubMed] [Google Scholar]
- 46.Keller M., Wadle S., Paust N., Dreesen L., Nuese C., Strohmeier O., Zengerle R., von Stetten F. Centrifugo-thermopneumatic fluid control for valving and aliquoting applied to multiplex real-time PCR on off-the-shelf centrifugal thermocycler. RSC Adv. 2015;5:89603–89611. doi: 10.1039/C5RA16095B. [DOI] [Google Scholar]
- 47.Haeberle S., Brenner T., Zengerle R., Ducrée J. Centrifugal extraction of plasma from whole blood on a rotating disk. Lab Chip. 2006;6:776–781. doi: 10.1039/B604145K. [DOI] [PubMed] [Google Scholar]
- 48.Steigert J., Brenner T., Grumann M., Riegger L., Lutz S., Zengerle R., Ducrée J. Integrated siphon-based metering and sedimentation of whole blood on a hydrophilic lab-on-a-disk. Biomed. Microdevices. 2007;9:675–679. doi: 10.1007/s10544-007-9076-0. [DOI] [PubMed] [Google Scholar]
- 49.Kinahan D.J., Kearney S.M., Kilcawley N.A., Early P.L., Glynn M., Ducrée J. Density-Gradient Mediated Band Extraction of Leukocytes from Whole Blood Using Centrifugo-Pneumatic Siphon Valving on Centrifugal Microfluidic Discs. PLoS ONE. 2016;11:e0155545. doi: 10.1371/journal.pone.0155545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Dimov N., Gaughran J., McAuley D., Boyle D., Kinahan D.J., Ducree J. Centrifugally automated solid-phase purification of RNA; Proceedings of the 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS); San Francisco, CA, USA. 26–30 January 2014; pp. 260–263. [DOI] [Google Scholar]
- 51.Gaughran J., Boyle D., Murphy J., Kelly R., Ducrée J. Phase-selective graphene oxide membranes for advanced microfluidic flow control. Microsystems Nanoeng. 2016;2:16008. doi: 10.1038/micronano.2016.8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Strohmeier O., Keil S., Kanat B., Patel P., Niedrig M., Weidmann M., Hufert F., Drexler J., Zengerle R., von Stetten F. Automated nucleic acid extraction from whole blood, B. subtilis, E. coli, and Rift Valley fever virus on a centrifugal microfluidic LabDisk. RSC Adv. 2015;5:32144–32150. doi: 10.1039/C5RA03399C. [DOI] [Google Scholar]
- 53.Brassard D., Geissler M., Descarreaux M., Tremblay D., Daoud J., Clime L., Mounier M., Charlebois D., Veres T. Extraction of nucleic acids from blood: Unveiling the potential of active pneumatic pumping in centrifugal microfluidics for integration and automation of sample preparation processes. Lab Chip. 2019;19:1941–1952. doi: 10.1039/C9LC00276F. [DOI] [PubMed] [Google Scholar]
- 54.Karle M., Miwa J., Röth G., Zengerle R., Von Stetten F. A Novel Microfluidic Platform for Continuous DNA Extraction and Purification using Laminar Flow Magnetophoresis; Proceedings of the 2009 IEEE 22nd International Conference on Micro Electro Mechanical Systems; Sorrento, Italy. 25–29 January 2009; pp. 276–279. [DOI] [Google Scholar]
- 55.Kido H., Micic M., Smith D., Zoval J., Norton J., Madou M. A novel, compact disk-like centrifugal microfluidics system for cell lysis and sample homogenization. Colloids Surfaces B Biointerfaces. 2007;58:44–51. doi: 10.1016/j.colsurfb.2007.03.015. [DOI] [PubMed] [Google Scholar]
- 56.Grumann M., Geipel A., Riegger L., Zengerle R., Ducrée J. Batch-mode mixing on centrifugal microfluidic platforms. Lab Chip. 2005;5:560–565. doi: 10.1039/b418253g. [DOI] [PubMed] [Google Scholar]
- 57.Ducrée J., Brenner T., Haeberle S., Glatzel T., Zengerle R. Multilamination of flows in planar networks of rotating microchannels. Microfluid. Nanofluidics. 2006;2:78–84. doi: 10.1007/s10404-005-0056-5. [DOI] [Google Scholar]
- 58.Burger R., Kinahan D.J., Cayron H., Reis N., Fonseca J., Ducrée J. Siphon-Induced Droplet Break-Off for Enhanced Mixing on a Centrifugal Platform. Inventions. 2019;5:1. doi: 10.3390/inventions5010001. [DOI] [Google Scholar]
- 59.Ducrée J., Haeberle S., Brenner T., Glatzel T., Zengerle R. Patterning of flow and mixing in rotating radial microchannels. Microfluid. Nanofluidics. 2005;2:97–105. doi: 10.1007/s10404-005-0049-4. [DOI] [Google Scholar]
- 60.Lutz S., Mark D., Roth G., Zengerle R., von Stetten F. Centrifugal Microfluidic Platforms for Molecular Diagnostics. Clin. Chem. Lab. Med. 2011;49:S608. [Google Scholar]
- 61.Tang M., Wang G., Kong S.-K., Ho H.-P. A Review of Biomedical Centrifugal Microfluidic Platforms. Micromachines. 2016;7:26. doi: 10.3390/mi7020026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Maguire I., O’Kennedy R., Ducrée J., Regan F. A review of centrifugal microfluidics in environmental monitoring. Anal. Methods. 2018;10:1497–1515. doi: 10.1039/C8AY00361K. [DOI] [Google Scholar]
- 63.Burger R., Amato L., Boisen A. Detection methods for centrifugal microfluidic platforms. Biosens. Bioelectron. 2016;76:54–67. doi: 10.1016/j.bios.2015.06.075. [DOI] [PubMed] [Google Scholar]
- 64.Duffy D.C., Gillis H.L., Lin J., Sheppard N.F., Kellogg G.J. Microfabricated Centrifugal Microfluidic Systems: Characterization and Multiple Enzymatic Assays. Anal. Chem. 1999;71:4669–4678. doi: 10.1021/ac990682c. [DOI] [Google Scholar]
- 65.Azimi-Boulali J., Madadelahi M., Madou M.J., Martinez-Chapa S.O. Droplet and Particle Generation on Centrifugal Microfluidic Platforms: A Review. Micromachines. 2020;11:603. doi: 10.3390/mi11060603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Brennan D., Coughlan H., Clancy E., Dimov N., Barry T., Kinahan D., Ducrée J., Smith T.J., Galvin P. Development of an on-disc isothermal in vitro amplification and detection of bacterial RNA. Sensors Actuators B Chem. 2017;239:235–242. doi: 10.1016/j.snb.2016.08.018. [DOI] [Google Scholar]
- 67.Delgado S.M.T., Kinahan D.J., Sandoval F.S., Julius L.A.N., Kilcawley N.A., Ducrée J., Mager D. Fully automated chemiluminescence detection using an electrified-Lab-on-a-Disc (eLoaD) platform. Lab Chip. 2016;16:4002–4011. doi: 10.1039/C6LC00973E. [DOI] [PubMed] [Google Scholar]
- 68.Ducrée J. Systematic review of centrifugal valving based on digital twin modelling towards highly integrated Lab-on-a-Disc systems. Nat. Microsyst. Nanoeng. 2021 doi: 10.20944/preprints202105.0683.v2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Ducrée J. Design optimization of centrifugal microfluidic “Lab-on-a-Disc” systems towards fluidic larger-scale integration. Appl. Sci. 2021 doi: 10.20944/preprints202105.0282.v2. [DOI] [Google Scholar]
- 70.Abi-Samra K., Hanson R., Madou M., Iii R.A.G. Infrared controlled waxes for liquid handling and storage on a CD-microfluidic platform. Lab Chip. 2011;11:723–726. doi: 10.1039/C0LC00160K. [DOI] [PubMed] [Google Scholar]
- 71.Kong L.X., Parate K., Abi-Samra K., Madou M. Multifunctional wax valves for liquid handling and incubation on a microfluidic CD. Microfluid. Nanofluidics. 2015;18:1031–1037. doi: 10.1007/s10404-014-1492-x. [DOI] [Google Scholar]
- 72.Al-Faqheri W., Ibrahim F., Thio T.H.G., Moebius J., Joseph K., Arof H., Madou M. Vacuum/Compression Valving (VCV) Using Parrafin-Wax on a Centrifugal Microfluidic CD Platform. PLoS ONE. 2013;8:e58523. doi: 10.1371/journal.pone.0058523. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Wang Y., Li Z., Huang X., Ji W., Ning X., Liu K., Tan J., Yang J., Ho H.-P., Wang G. On-board control of wax valve on active centrifugal microfluidic chip and its application for plasmid DNA extraction. Microfluid. Nanofluidics. 2019;23:112. doi: 10.1007/s10404-019-2278-y. [DOI] [Google Scholar]
- 74.Garcia-Cordero J.L., Kurzbuch D., Benito-Lopez F., Diamond D., Lee L.P., Ricco A.J. Optically addressable single-use microfluidic valves by laser printer lithography. Lab Chip. 2010;10:2680–2687. doi: 10.1039/c004980h. [DOI] [PubMed] [Google Scholar]
- 75.Stumpf F., Schwemmer F., Hutzenlaub T., Baumann D., Strohmeier O., Dingemanns G., Simons G., Sager C., Plobner L., von Stetten F., et al. LabDisk with complete reagent prestorage for sample-to-answer nucleic acid based detection of respiratory pathogens verified with influenza A H3N2 virus. Lab Chip. 2015;16:199–207. doi: 10.1039/C5LC00871A. [DOI] [PubMed] [Google Scholar]
- 76.Delgado S.M.T., Kinahan D.J., Julius L.A.N., Mallette A., Ardila D.S., Mishra R., Miyazaki C.M., Korvink J., Ducrée J., Mager D. Wirelessly powered and remotely controlled valve-array for highly multiplexed analytical assay automation on a centrifugal microfluidic platform. Biosens. Bioelectron. 2018;109:214–223. doi: 10.1016/j.bios.2018.03.012. [DOI] [PubMed] [Google Scholar]
- 77.Iii R.G., Nwankire C.E., Gaughran J., Zhang X., Donohoe G.G., Rook M., O’Kennedy R., Ducrée J. Centrifugo-pneumatic valving utilizing dissolvable films. Lab Chip. 2012;12:2894–2902. doi: 10.1039/c2lc20973j. [DOI] [PubMed] [Google Scholar]
- 78.Godino N., Iii R.G., Linares A.V., Burger R., Ducrée J. Comprehensive integration of homogeneous bioassays via centrifugo-pneumatic cascading. Lab Chip. 2013;13:685–694. doi: 10.1039/C2LC40722A. [DOI] [PubMed] [Google Scholar]
- 79.Kinahan D.J., Early P.L., Vembadi A., MacNamara E., Kilcawley N.A., Glennon T., Diamond D., Brabazon D., Ducrée J. Xurography actuated valving for centrifugal flow control. Lab Chip. 2016;16:3454–3459. doi: 10.1039/C6LC00568C. [DOI] [PubMed] [Google Scholar]
- 80.Mishra R., Gaughran J., Kinahan D., Ducrée J. Functional Membranes for Enhanced Rotational Flow Control on Centrifugal Microfluidic Platforms. Ref. Modul. Mater. Sci. Mater. Eng. 2017 doi: 10.1016/b978-0-12-803581-8.04041-8. [DOI] [Google Scholar]
- 81.Kinahan D.J., Kearney S.M., Dimov N., Glynn M.T., Ducrée J. Event-triggered logical flow control for comprehensive process integration of multi-step assays on centrifugal microfluidic platforms. Lab Chip. 2014;14:2249–2258. doi: 10.1039/C4LC00380B. [DOI] [PubMed] [Google Scholar]
- 82.Kinahan D.J., McConville K., Henderson B., McCaul M., McNamara E., Diamond D., Ducrée J. Digital pulse actuated flow control on a centrifugal disc towards multiparameter water quality monitoring. In: Pamme N., Ducrée J., editors. Proceedings of the 20th International Conference on Miniaturized Systems for Chemistry and Life Sciences (μTAS 2016); Dublin, Ireland. 9–13 October 2016; Dublin, Ireland: CBMS; 2016. pp. 871–872. [Google Scholar]
- 83.Kinahan D.J., Renou M., Kurzbuch D., Kilcawley N.A., Bailey Éanna, Glynn M., McDonagh C., Ducrée J. Baking Powder Actuated Centrifugo-Pneumatic Valving for Automation of Multi-Step Bioassays. Micromachines. 2016;7:175. doi: 10.3390/mi7100175. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Mishra R., Reilly G., Agnew M., Garvey A., Rogers C., Andrade E., Ma H., Fitzgerald S., Zapatero J., O’Kennedy R., et al. Laser-actuated centrifugo-pneumatic flow control towards ‘sample-to-answer’ integrated detection of multi-marker panels at the point-of-care; Proceedings of the 2018 IEEE Micro Electro Mechanical Systems (MEMS); Dublin, Ireland. 21–25 January 2018; pp. 1185–1188. [DOI] [Google Scholar]
- 85.Unger M.A., Chou H.-P., Thorsen T., Scherer A., Quake S.R. Monolithic Microfabricated Valves and Pumps by Multilayer Soft Lithography. Science. 2000;288:113–116. doi: 10.1126/science.288.5463.113. [DOI] [PubMed] [Google Scholar]
- 86.Thorsen T., Maerkl S.J., Quake S.R. Microfluidic Large-Scale Integration. Science. 2002;298:580–584. doi: 10.1126/science.1076996. [DOI] [PubMed] [Google Scholar]
- 87.Digital Twin. [(accessed on 25 May 2021)];2021 Available online: https://en.wikipedia.org/wiki/Digital_twin.
- 88.Marr B. What Is Digital Twin Technology—And Why Is It So Important? [(accessed on 25 May 2021)];2017 Available online: https://www.forbes.com/sites/bernardmarr/2017/03/06/what-is-digital-twin-technology-and-why-is-it-so-important/
- 89.Brenner T., Glatzel T., Zengerle R., Ducrée J. Frequency-dependent transversal flow control in centrifugal microfluidics. Lab Chip. 2004;5:146–150. doi: 10.1039/B406699E. [DOI] [PubMed] [Google Scholar]
- 90.Clime L., Daoud J., Brassard D., Malic L., Geissler M., Veres T. Active pumping and control of flows in centrifugal microfluidics. Microfluid. Nanofluidics. 2019;23:29. doi: 10.1007/s10404-019-2198-x. [DOI] [Google Scholar]
- 91.Noroozi Z., Kido H., Madou M.J. Electrolysis-Induced Pneumatic Pressure for Control of Liquids in a Centrifugal System. J. Electrochem. Soc. 2011;158:P130–P135. doi: 10.1149/2.028111jes. [DOI] [Google Scholar]
- 92.Abi-Samra K., Clime L., Kong L., Gorkin R., Kim T.-H., Cho Y.-K., Madou M. Thermo-pneumatic pumping in centrifugal microfluidic platforms. Microfluid. Nanofluidics. 2011;11:643–652. doi: 10.1007/s10404-011-0830-5. [DOI] [Google Scholar]
- 93.Haeberle S., Schmitt N., Zengerle R., Ducrée J. Centrifugo-magnetic pump for gas-to-liquid sampling. Sens. Actuators A Phys. 2007;135:28–33. doi: 10.1016/j.sna.2006.09.001. [DOI] [Google Scholar]
- 94.Siegrist J., Gorkin R., Bastien M., Stewart G., Peytavi R., Kido H., Bergeron M., Madou M. Validation of a centrifugal microfluidic sample lysis and homogenization platform for nucleic acidextraction with clinical samples. Lab Chip. 2009;10:363–371. doi: 10.1039/B913219H. [DOI] [PubMed] [Google Scholar]
- 95.Soroori S., Kulinsky L., Kido H., Madou M. Design and implementation of fluidic micro-pulleys for flow control on centrifugal microfluidic platforms. Microfluid. Nanofluidics. 2014;16:1117–1129. doi: 10.1007/s10404-013-1277-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Kinahan D.J., Kearney S.M., Faneuil O.P., Glynn M.T., Dimov N., Ducrée J. Paper imbibition for timing of multi-step liquid handling protocols on event-triggered centrifugal microfluidic lab-on-a-disc platforms. RSC Adv. 2014;5:1818–1826. doi: 10.1039/C4RA14887H. [DOI] [Google Scholar]
- 97.Godino N., Vereshchagina E., Gorkin R., Ducrée J. Centrifugal automation of a triglyceride bioassay on a low-cost hybrid paper-polymer device. Microfluid. Nanofluidics. 2014;16:895–905. doi: 10.1007/s10404-013-1283-9. [DOI] [Google Scholar]
- 98.Ducrée J. Anti-counterfeit technologies for centrifugal microfluidic “Lab-on-a-Disc” systems: An analysis. 2021 in press. [Google Scholar]
- 99.Ducrée J., Etzrodt M., Bartling S., Walshe R., Harrington T., Wittek N., Posth S., Wittek K., Ionita A., Prinz W., et al. Unchaining Collective Intelligence for Science, Research, and Technology Development by Blockchain-Boosted Community Participation. Front. Blockchain. 2021;4:631648. doi: 10.3389/fbloc.2021.631648. [DOI] [Google Scholar]
- 100.Ducrée J. Research—A blockchain of knowledge? Blockchain Res. Appl. 2020;1:100005. doi: 10.1016/j.bcra.2020.100005. [DOI] [Google Scholar]
- 101.Ducrée J., Gravitt M., Walshe R., Bartling S., Etzrodt M., Harrington T. Open Platform Concept for Blockchain-Enabled Crowdsourcing of Technology Development and Supply Chains. Front. Blockchain. 2020;3:386525. doi: 10.3389/fbloc.2020.586525. [DOI] [Google Scholar]
- 102.Ducrée J., Etzrodt M., Gordijn B., Gravitt M., Bartling S., Walshe R., Harrington T. Blockchain for Organizing Effective Grass-Roots Actions on a Global Commons: Saving the Planet. Front. Blockchain. 2020;3:33. doi: 10.3389/fbloc.2020.00033. [DOI] [Google Scholar]