Abstract
The motivation for this study was to develop a microdevice for the precise rheological characterization of biofluids, especially blood. The method presented was based on the principles of rheometry and fluid mechanics at the microscale. Traditional rheometers require a considerable amount of space, are expensive, and require a large volume of sample. A mathematical model was developed that, combined with a proper experimental model, allowed us to characterize the viscosity of Newtonian and non-Newtonian fluids at different shear rates. The technology presented here is the basis of a point-of-care device capable of describing the nonlinear rheology of biofluids by the fluid/air interface front velocity characterization through a microchannel. The proposed microrheometer uses a small amount of sample to deliver fast and accurate results, without needing a large laboratory space. Blood samples from healthy donors at distinct hematocrit percentages were the non-Newtonian fluid selected for the study. Water and plasma were employed as testing Newtonian fluids for validation of the system. The viscosity results obtained for the Newtonian and non-Newtonian fluids were consistent with pertinent studies cited in this paper. In addition, the results achieved using the proposed method allowed distinguishing between blood samples with different characteristics.
Keywords: rheometer, microrheometer, rheology, hemorheology, viscosity, blood, plasma
1. Introduction
The study of the viscosity of biofluids, such as blood, has been used to understand their physiological mechanisms and biological functions [1]. Due to their complex properties, most biological fluids have a non-Newtonian behavior. While a Newtonian fluid shows a viscosity that is independent of the shear rate, non-Newtonian fluids’ viscosity shows a dependence on the applied shear rate.
As illustrated in Figure 1, blood exhibits shear-thinning behavior. Shear-thinning fluid viscosity is characterized by a monotonically decreasing dependence on the shear rate. Such behavior is closely related to the dynamics and interactions of red blood cells (RBCs), the main cellular component of blood. Blood viscosity variation as a function of shear rate due to rheological properties of RBCs is shown in Figure 1 [2,3,4,5]. RBCs have a biconcave disklike shape at rest; they aggregate and form rouleaux structures that can, reversibly and continuously, disaggregate to single-flowing discocytes for increasing shear rates. This change in microstructure implies a viscosity variation. When RBCs circulate through the vascular system, deformability and dynamics of RBCs account for a further decrease in blood viscosity. Furthermore, blood viscosity value strongly depends on other parameters such as temperature, whole blood cellular components (for example hematocrit value), and hemoglobin content [6].
Figure 1.
Viscosity profiles of whole blood and plasma. Adapted from Rosencranz [7].
The gold standard for viscosity measurements comprises a wide range of types of rheometers (instruments that measure the stress and deformation history of a material). The specific kind of rheometer that can only measure the steady shear viscosity function of a material is called a viscometer [8]. Rheometers may be differentiated on the type of flow they induce; i.e., drag flows and pressure-driven flows. Typical drag-flow rheometers are the coaxial cylinder Couette flow and coaxial cone-plate, while capillary rheometers are pressure-driven. The disadvantages of using this type of equipment include their high operation costs and the large size of samples they require. The rheometrical properties of complex fluids have been traditionally studied using such macroscopic devices.
In general, macroscopic rheometers provide less accuracy than their microfluidic counterparts. In contrast to classical methods, the study of rheology through microfluidic techniques is a widely used method to analyze the viscosity of complex fluids. Several microfluidic platforms have been developed to reach high confinement levels [9,10,11] (and thus, high precision), such as broader shear rate crystal microbalance (QCM) [12], laser-induced capillary wave [13], and the use of multiple microfluidic channels. This relatively new field has come up with some viscometers that measure viscosity as a function of shear rate and temperature, although most of them are not applied in biofluids. Novel microfluidic systems’ accuracy is equivalent to traditional methods, being also preferable for measuring complex viscosity behavior [14] and microflows [15]. A substantial number of viscometers for biofluid characterization exist [16,17,18]. However, they still present disadvantages in terms of space and sample amount required.
Some new applications use lab-on-a-chip structures from diverse materials aiming for miniaturization, such as glass and polydimethylsiloxane (PDMS) [19,20]. Microfluidic techniques that use shear as a mean to understand viscosity include pressure sensing, flow-rate sensing, surface tension, coflowing streams, and diffusion-based and velocimetry-based sensing [1]. The microrheometer proposed by Solomon [21] used a mobile phone camera to detect the fluid-front advancement through a microchannel. Srivastava [22], on the other hand, proposed a microrheometer in which the measurement of viscosity was based on monitoring with a camera the capillary pressure-driven movement of fluid sample, and shear rate varied with time. The device fabrication was not simple, and required much more time and specialized equipment and materials to produce, than our PDMS on glass microchannels. By contrast, the method of detection in our device is based on an array of electrodes that automatically detect the advancement of the fluid. By adding electronic sensing, we do not need further video processing and analysis. Additionally, our method has been demonstrated to measure viscosity of blood and differentiate between samples at distinct hematocrit levels.
The scale reduction allows for an increase in sensitivity and accuracy. Other factors such as hematocrit level, plasma, and coagulation could affect blood viscosity. On that matter, point-of-care devices such as thromboelastograph (TEG) (Haemoscope Corporation, Niles, IL, USA) and the ROTEM thromboelastometer (Pentapharm GmbH, Munich, Germany) [23,24] have led the research toward successful results for blood and plasma [25].
The technology presented here consists of a front microrheometer mainly composed of three parts: a microfluidic consumable that comprises a microchannel with an inlet and an outlet; the electronic setup, including pump, electronic detection, and data acquisition; and a mathematical model to properly convert obtained data into viscosity measurements. The electronic detection is performed with a sensor array (of metallic electrodes) located along the microchannel. This technology allows us to characterize the rheological properties of the fluid as a function of the fluid-front advancement, using a small amount of sample, reducing space required, and finally minimizing time and material consumption. The tested samples come into contact only with the walls of the microfluidic channel; therefore, there is neither a need for cleaning protocols, nor for the use of cleaning substances. Unlike other rheometers that require extensive cleaning of their components, the microfluidic chip is intended to be used as a consumable. The consumable is disposed of after each experiment. Thanks to the number of pairs of electrodes (24), situated at each side of the microchannel, it is possible to obtain a high number of data points, thus enhancing the accuracy of measurements. Finally, the integration of electronic detection into microfluidic devices improves the reliability of measurements.
The general goal of this study was to develop the method and model for a microrheometer capable of characterizing the viscosity of biofluids by analyzing the fluid-front advancement, and reducing the required sample’s amount, space, and time consumption. First, we characterized the viscosity of healthy blood samples at distinct hematocrit levels. The viscosity of blood obtained at was 12.2 mPa·s for the sample at 50% Ht; 10.67 mPa·s for 42% Ht; and 8.81 mPa·s for 35% Ht. These results were consistent with results obtained by other authors [26,27,28]. This constitutes a necessary benchmark to, in the near future, compare and detect rheological anomalies associated with hematological diseases, since the shape and biomechanical properties of RBCs can be altered in several diseases related to hematological disorders [29]. Second, we validated the new technology by comparing its results to data obtained using traditional equipment. We compared our viscosity results to results obtained using a rotational rheometer (Malvern Kinexus Pro+, Malvern Instruments Limited, Worcestershire, UK). The viscosity value obtained for blood at 42% Ht using this macrorheometer for comparison was
2. Materials and Methods
2.1. Mathematical Model
The development of mathematical models describing the Newtonian and non-Newtonian characteristics of blood as a complex fluid [30,31] are very important for understanding viscosity behavior at the microscale, and determine how different blood conditions affect the viscosity values. Deformation of RBCs at high shear rates and the formation of aggregates in the form of rouleaux at low shear rates are essential aspects to keep in focus when developing models for blood [32,33].
The linear displacement of a fluid layer with respect to another in the time interval allows obtaining the rate of displacement (or velocity). If we consider a fluid flow moving into the direction between two plates, the only nonzero velocity component is , which varies in the z direction due to the interaction with the planes, being . The shear rate is the rate of change of velocity at which one layer of fluid passes over an adjacent layer. The shear produced between the layers depends on the z dimension, and is:
| (1) |
An easy method to study fluid flow consists of the study of the fluid-front (fluid–air interface) advancement. In a microfluidic channel, the front velocity, is the change in position of the fluid, through time along the microfluidic channel. The term is defined as the average position of the front, where is the fluid-front position with respect to z, as shown in Figure 2. One of the advantages of working with the mean front velocity is that its value is independent of the z position. The theoretical model used assumes that velocity inside the microchannel behaves according to Darcy’s law. In past experiments [34], Trejo C. observed that this theoretical description of the velocity of the fluid inside the microchannel held for microchannels of 1 mm width and heights over 150 um. To test distinct aspect ratios, the authors considered microchannels of different heights: b = 300 μm, b = 200 μm, b = 150 μm, and b = 50 μm; width w = 1 mm; and length lc = 4 cm. For heights lower than 150 um the model is not longer valid.
Figure 2.
The velocity is the x component of velocity inside the microfluidic channel as a function of height indicates the mean position of the fluid front. The channel dimensions satisfy the relation b.
By studying the fluid/air interface, the shear rate can be defined as the normalization of the mean front velocity, according to the microchannel height, b, as defined by Equation (1) (see Figure 2).
| (2) |
The viscosity of non-Newtonian fluids can be described using different models according to the characteristics of each fluid. The most commonly used are the Carreau–Yasuda and power-law models [35]. The power-law model, also known as the Oswaldt–De Waele fluid model, describes the viscosity of a fluid on dependence of its shear rate:
| (3) |
where the prefactor m corresponds to the value of viscosity at . The exponent n indicates the nature of the viscosity of the studied fluid. Depending on the value of n, the behavior of a fluid is characterized as Newtonian or non-Newtonian. Values of correspond to shear-thinning behavior, while corresponds to shear-thickening behavior. On the other hand, Newtonian fluids have a value of n = 1.
Using the Stokes equation, it is possible to calculate the flow through a known geometry as a function of the average velocity. In most cases, microfluidic systems are composed of pressure sources and microfluidic channels connected by cylindrical tubing, creating coupled fluidic systems. The flows passing through two coupled geometries are equivalent due to the mass conservation principle. The mathematical deductions concerning these calculations are detailed in Appendix A.1.
In a closed system composed of a pressure source, a fluid reservoir, and the coupling of tubing with a microchannel of rectangular cross-section, the pressure inside the rectangular microchannel, ΔP, is the summation of all the pressures in the system, as shown in Figure 3.
Figure 3.
Diagram of the pressures involved in the system.
As properly deduced in Appendix A.2, the effective pressure () is defined as the pressure resulting from the combination of the pumping system pressure, the capillary pressure, and the hydrostatic pressure into the whole system shown in Figure 3 (see Equation (A25) deduced in Appendix A.2). Effective pressure has an algebraic dependence with :
| (4) |
where all the independent variables are grouped in K, which depends on the fluid properties (m and n) and the geometrical parameters of the system shown in Equation (4):
| (5) |
According to the mass conservation principle, in our system, (Appendix A.2). This is why we can write velocity inside the tubing as a function of the velocity measured inside the microchannel, :
| (6) |
2.2. Experimental Model
2.2.1. General Case: Non-Newtonian Fluids
We used a coupled system in which a tube was connected to a microchannel structure. The effective pressure, , acting in the system (see Figure 3) caused fluid to flow through the coupled system. Through electronic detection, the mean front velocity was measured along the microfluidic channel. By calculating velocity values and the channel dimensions, shear-rate values were obtained. By plotting vs. shear rate, and fitting a curve for the relation between and , we obtained:
| (7) |
Terms A and n can be obtained from Equation (6). By equating experimental data in Equation (5) with the mathematical model in Equations (3) and (6), we could calculate the value of m as follows:
| (8) |
Once n and m were obtained, the viscosity values of a fluid were calculated using the power-law model in Equation (2).
2.2.2. Newtonian Fluids
A Newtonian fluid has a constant viscosity. In opposition to non-Newtonian fluids, it does not depend on the applied pressure. Analogously, as it has been done in the general case, the mathematical model for Newtonian fluids is shown in Appendix A.3.
Pressure for the entire system is described in Equation (8), which has a linear dependence on the mean front velocity.
| (9) |
For the Newtonian case, can also be expressed in terms of the geometry of the system and the experimentally obtained front velocity:
| (10) |
Along the same lines, the experimental data of the shear rate values obtained from applying a known pressure to a Newtonian fluid have the linear relation shown in Equation (10):
| (11) |
Using the experimental results alongside the mathematical model, m can be calculated as:
| (12) |
In Newtonian fluids for which n = 1, the power-law equation for viscosity in Equation (2) can be re-written as:
| (13) |
2.3. Experimental Method
The experiments were performed using microfluidic chips made of polydimethylsiloxane (PDMS) attached to glass substrates by using plasma bonding; schematics of the final structures are shown in Figure 4a. The glass substrates had a printed pattern of 24 pairs of gold electrodes, which were in contact with the flowing fluid. The electrodes had a separation in between, and were in the form of 4 groups of 6 pairs located along the length of the channel. The distance between each set was 8.5 mm (Figure 4b). The channel width was , the channel height was, and the length from inlet to outlet was . Commonly, PDMS, glass, and Cole Parmer Tygon tubing are employed due to their good biocompatibility, low cost, and high adaptability.
Figure 4.
Schematics of the microfluidic consumable. (a) The microfluidic channel made of PDMS sealed on a glass substrate. The channel dimensions are , and (b) Gold electrodes printed on the glass surface and beneath the PDMS layer.
The experimental setup used to carry out the experiments was composed of a pressure pump (Dolomite Fluika) connected to a closed fluid reservoir. From the reservoir, Tygon tubing with an internal radius and length cm was connected to the microchannel. The pressure exerted on the fluid was set using a simple graphic interface, which controlled the pressure pump’s performance. A schematic representation of the experimental setup described is presented in Figure 5. Different pressures for the Fluika Pump were set to run the experiment, ranging from 500 Pa to 5000 Pa. Each electrode was connected via a pin array box to a National Instruments control card. As soon as the pressure began to be applied, electronic reading was activated by the controller of a myRIO National Instruments card. This tool communicated with the fluidic pump and the electronic reading pins through a computer. The sample fluid came out from the reservoir at the set pressure and went directly into the channel. As the fluid came into contact with the electrodes, the time required for the fluid front to reach each electrode pair inside the microfluidic channel was obtained. Using the time data and the geometry of the microfluidic system, it was possible to calculate the fluid-front velocity between electrode pairs through the microchannel.
Figure 5.
Experimental setup for laboratory use.
Ten samples of 500 μL from each different test fluid were used during this study: (fluid standard MGVS60, deionized water (DI water), plasma, and blood from healthy donors). We selected MGVS60, a viscosity standard tested in strict accordance with ASTM D2162, the primary method for viscosity standards calibration that is typically used for the calibration and verification of viscosity-measuring equipment. Its viscosity is at 24 °C. Deionized water was selected because its viscosity is constant and known: [36]. Blood samples and their respective plasma were provided in 10 mL tubes by Banc de Sang i Teixits (Barcelona, Spain), from 10 anonymous healthy donors. The use of these samples was authorized by the Bioethics Committee of the University of Barcelona (IRB 00003099).
To obtain the plasma samples, we centrifuged whole blood samples at 2500 rpm for 5 min. The plasma that rested at the top was collected with a pipette. Samples were prepared at three different hematocrit concentrations, Ht (%) of 50%, 42%, and 35%. These were prepared by taking plasma previously separated from whole blood and then adding to it the desired volume of RBCs corresponding to each hematocrit percentage. The hematocrit selected for further comparison to a benchtop device was 42% Ht, consistent with the levels found in healthy adults. All the tests were performed at room temperature (25 °C) and, for the case of blood, within two days of extraction, to avoid cell damage from aging [28]. At the beginning of each experiment, the DI water was evaluated to guarantee that the setup was properly calibrated.
3. Experimental Results and Discussion
For a correct rheological characterization of the fluid, it was necessary to properly determine the pressures inside the system (Figure 3). The pressure set for the pumping source was automatically recorded. Considering that a sample of 500 μL inside a reservoir supposes a height,, the contribution generated by the hydrostatic pressure can be obtained. Parameters for the different tested fluids at room temperature can be found in Table 1. The material used for the microchannel was hydrophobic, meaning that it opposed the fluid-front advancement. Fluid-front images at different pressures were obtained by using a camera and an inverted microscope. The contact angle, θ, was extracted optically via Image J free software. The surface tension (τ) of the standard fluid was experimentally measured with a capillary tube with an internal diameter of 1.15 mm [37,38]. By using Equation (A20) deduced in Appendix A.2, it was possible to calculate for a rectangular cross-section channel at each measured pressure by analyzing the fluid-front contact angle. The contact angle also depended on the fluid being analyzed and the total pressure applied. The curvature of the fluid–air interface was expected to be almost constant for water, standard MGVS60, and plasma; however, it changed as the pressure was increased when measuring blood (Table 2).
Table 1.
Characteristic values for water, MGVS60 plasma, and blood.
| Sample |
at 25 °C |
at H = 5 mm |
at 25 °C |
|
|---|---|---|---|---|
| Water | 102.42 | |||
| MGVS60 | 55.56 | 0.107 | 106.90 | |
| Plasma | 102.27 | |||
| Blood | * (Table 2) |
* Contact angle for blood was nonconstant at different pressures. See Table 2.
Table 2.
Contact angle and capillary pressure at each pump pressure applied for healthy blood at 42% hematocrit and RT (25 °C). was obtained using Equation (A20).
| 500 | 103.24 | 115.11 |
| 1000 | 106.94 | 146.46 |
| 1500 | 109.60 | 168.62 |
| 2000 | 111.76 | 186.35 |
| 2500 | 113.62 | 201.37 |
| 3000 | 115.27 | 214.54 |
| 3500 | 116.76 | 226.35 |
| 5000 | 120.64 | 256.22 |
When studying blood-front behavior, dependence between the applied pressure and the dynamic contact angle measured was observed. Hence, depended on the applied pressure, . To obtain the contact angle of blood inside our microfluidic channel, we conducted a separate experiment for blood at 42% Ht. By using a microscope (Optika XDS-3, Optika SRL, Bergamo, Italy) and a high-speed camera (Photron Fastcam SA3, Photron Limited, Tokyo, Japan), we captured images of the fluid front at different set pressures, . We measured the contact angle using the angle-measurement tool in Image J, as depicted in Figure A2. Calculation of was done as per Appendix A.2. The obtained contact angle when changing the applied pressure and thus, the capillary pressure obtained can be seen in Table 2. The contact angles of blood used in this paper have been considered independent of the Ht (%) value.
To validate our setup and the use of electronic detection as a proper method to determine viscosity for Newtonian fluids, two different strategies were followed: first, a Newtonian standard fluid and DI water with reported viscosity value were evaluated. Second, the obtained results with our technology were compared to the results obtained with a commercial macrorheometer, the Malvern Kinexus Pro+ (Malvern Instruments Limited, Worcestershire, United Kingdom).
From the data obtained (applied pressure and time), knowing the device geometry, resistances, and the pressures added to the system, it was possible to plot the effective pressure as a function of shear rate for each sample.
A linear fit of the data obtained when plotting the applied pressure with an external pressure pump as a function of or is shown in Figure 6. It was used to determine for Newtonian fluids, as detailed in Equation (A33), as:
| (14) |
Figure 6.
Effective pressure vs. shear rate obtained for the Newtonian calibration fluids studied. Newtonian fluids: DI water and MGVS60: (a) linear scale and (b) log-log scale.
As shown in Figure 6 and detailed in Table 3, in both cases, n was very close to 1, which indicated that the fluid was Newtonian, compatible with the behavior expected for both.
Table 3.
Viscosity obtained using the developed microrheometer (with the corresponding n values obtained) and a commercial macrorheometer and electronic detection methods. The samples analyzed were DI water and MGVS60, a commercial standard fluid.
| Sample | Viscosity (mPa·s) Microrheometer |
Viscosity (mPa·s) Macrorheometer |
|
|---|---|---|---|
| DI water | 1.08 ± 0.03 | 1.017 | 1.02 ± 0.01 |
| MGVS60 | 5.87 ± 0.05 | 1.065 | 5.88 ± 0.02 |
From these data, as previously explained, the viscosity could be obtained by using Equation (13). These results were compared to those obtained with a commercial macrorheometer (see Figure 7).
Figure 7.
Viscosity vs. shear rate using the microrheometer for DI water (green) and standard MGVS60, and using the macrorheometer: (a) linear scale and (b) log–log scale.
Water has been found to exhibit a constant viscosity of η ≈ 1.002 mPa·s [39]. The same results were obtained with both rheometers (differing from each other by 3%). The viscosity obtained with the standard MGVS60 was around using the microrheometer, while the macrorheometer delivered a viscosity of 5.8 mPa·s. The expected value for this fluid was.
These results validated that the accuracy of the new electronic sensing method was adequate (equivalent to standard macrorheometer measures) regarding the measurement of viscosity in an automatic, fast, and effective way.
Once the measurement system was validated for Newtonian fluids with known characteristics, the experiment was repeated with plasma and blood at different % Ht values. The results are shown in Figure 8.
Figure 8.
Effective pressure vs. shear rate of blood, a non-Newtonian fluid, at 50% Ht, 42% Ht, 35% Ht; and plasma, a Newtonian fluid.
The behavior of the shear rate as Peff applied to the fluid variance is shown in Figure 8. Non-Newtonian fluids (in this case, blood as a shear-thinning fluid) showed a nonlinear relation between the pressure and their shear-rate response, while plasma showed a Newtonian behavior. Using the experimentally obtained values alongside Equations (6) and (7) for the non-Newtonian case, and Equations (10) and (11) for the Newtonian cases, the value of K could be calculated. Then, m and n were obtained. The values for each are shown in Table 4 (see references in Appendix B); water and MGVS60 values are also included.
Table 4.
Values for prefactor m and n for Newtonian and non-Newtonian fluids.
| Sample | m (Pa·s) | n | |
|---|---|---|---|
| *Blood at 50% Ht | 0.0122 | 0.8500 | |
| *Blood at 42% Ht | 0.0107 | 0.8519 | |
| *Blood at 35% Ht | 0.0088 | 0.8720 | |
| MGVS60 | 0.0058 | 1.0174 ± 0.002 | |
| Plasma | 0.0018 ± 0.0002 | 1.0131 ± 0.028 | |
| Water | 0.0010 ± 00101 | 1.0651 ± 0651 | |
*Blood viscosity values presented at .
With the effective pressure, and after obtaining m and n (Table 4), viscosity plots were obtained. Viscosity was calculated using Equation (2) (for non-Newtonian fluids: blood at different Ht (%). levels) and Equations (11) and (12) (for Newtonian fluids: plasma).
As depicted in Figure 9, plasma exhibited Newtonian behavior with a constant viscosity, between 1.50–1.81 mPa·s [40]. Blood exhibited shear-thinning behavior, as its viscosity decreased as the shear increased. Note that the viscosity of blood at 42% Ht (see Figure 9) varied from 12 to 7 mPa·s within the range evaluated, with a value of viscosity at blood at 50% Ht had a viscosity of ; and blood at 35% Ht had the lowest viscosity of the three tested hematocrit levels, . At a constant shear rate, the viscosity of blood increased as the hematocrit grew. This also demonstrated that our new device could be used to distinguish between different hematocrit levels. The hematocrit selected for further comparison to a benchtop device was 42%, consistent with the levels found in healthy adults. The viscosity obtained by this method was 10.69 mPa·s. for blood at 42% Ht. This can be compared to results also obtained under room-temperature conditions by consulted authors, who situated the viscosity of blood at between 9–10 mPa·s [40]. Dispersion in the measurements at the low shear-rate values observed in Figure 9 would be reduced by increasing the number of samples studied. With a sufficiently large N, a general curve and validity intervals for healthy blood could be obtained, which would indicate what the behavior of the viscosity must be as a function of the shear rate in a healthy patient (at a constant hematocrit value).
Figure 9.
(a) Viscosity vs. shear rate of blood, a non-Newtonian fluid, at 50% Ht, 42% Ht, and 35% Ht. Blood at 42% Ht using the macrorheometer (white circles). Viscosity vs. shear rate of plasma, a Newtonian fluid. (b) Viscosity vs. shear rate of non-Newtonian and Newtonian fluids in log–log scale.
Regarding the rheological properties of RBCs, blood showed higher resistance to flow at low values. In contrast, at higher values of , blood showed a more Newtonian behavior, which was reflected in the nearly flat region on the viscosity vs. shear rate plot. This distinct shear-thinning curve behavior could only be properly observed when shear rates as low as were reached. For this reason, it was important to choose a range of pressure that allowed for this regime to be available. Residual fit for both methods was a useful way to determine the accuracy differences between our technology and a commercial rheometer (see Figure 10).
Figure 10.
(a). Experimentally obtained viscosity values vs. predicted values for both devices using blood. (b) Residuals for macro and microrheometers.
Residuals, R, for the micro- and macrorheometer were calculated as the difference between experimentally obtained viscosity, η, and predicted viscosity, (Equation (14)).
| (15) |
The theoretically predicted viscosity of blood, , for both methods was obtained by calculating viscosity at each shear rate, using the fit of viscosity vs. shear rate from plots in Figure 9. For the micro- and macrorheometer, the calculation was made using Equations (15) and (16) respectively:
| (16) |
| (17) |
A good correlation was observed when plotting experimentally obtained values for viscosity and theoretically predicted values using the power-law model shown in Figure 10a, thus confirming that the experimentally obtained data was close enough to theoretically expected values. In addition, residuals’ plotted vs. predicted values in Figure 10b show that the macrorheometer residuals were larger than those obtained with our microrheometer. The accuracy of the prediction for the macrorheometer in this experiment was lower than the accuracy of our microrheometer by three orders of magnitude. Residuals for our microrheometer were closer to zero. At these low shear rates, from 1 to 25 , our technology performed with higher accuracy. This could be explained by the fact that blood is a complex fluid of the shear-thinning type. We decided to work on this window to obtain the nonlinear viscosity of blood and to avoid hemolysis. Our microrheometer had the advantage of being able to operate under low shear conditions with high stability. This makes it suitable for applications such as the analysis of blood, where low-shear-rate performance is necessary. At the shear-rate values required to observe the non-Newtonian behavior of blood, the microrheometer results presented herein had a smaller deviation. This can be attributed to the high stability at low shear rates. These low shear rates could be reached thanks to the size reduction achieved by using a microchannel. Moreover, a benchtop rheometer for laboratory use also requires a rather large sample of blood. For the macrorheometer experiments, the sample amount was 1.19 mL for each run. The microrheometer required samples of approximately 250 μL; this constituted 21% of the sample required for the benchtop device.
4. Conclusions
The technology presented here was based on the detection of the advancement of the fluid front using electrodes printed underneath a microfluidic channel. The accuracy of the method was tested by measuring the viscosities of plasma, DI water, and a commercial viscosity standard. Experiments to measure human blood viscosity also were performed. Unlike other microfluidic capillary viscometers, the microrheometer presented here was based on the detection of the fluid/air interface as it advanced through a microchannel. Through the experimental method described, it was possible to produce a series of shear rates that constituted a proper range to observe how the viscosity of non-Newtonian fluid changed as the applied pressure increased. Effective pressure, the pressure acting on the fluid is composed of all the interactions between the pressures acting in the system.
The characterization of the non-Newtonian behavior was obtained by applying the power-law model in Equation (2). The experimentally obtained results included a shear-rate window range of less than up to for both Newtonian and non-Newtonian fluids. The values obtained for viscosity—1.05 mPa·s for water and 1.8 mPa·s for plasma—fell in line with the values found in the literature [36,39]. Alternatively, we used the calibration standard MGVS60 to test the accuracy of the setup and obtained a viscosity value of showing a difference of less than 2% from the target viscosity indicated by the manufacturer at 24 °C. On the other hand, blood results showed shear thinning behavior, due to the presence of mainly red blood cells, under normal morphological conditions [2]. This confirmed that the experimental conditions were adequate to observe the shear-thinning behavior and the action of red blood cells on the viscosity of the whole fluid. The viscosity of blood at was 12.2 mPa·s for the sample at 50% Ht; 10.67 mPa·s for 42% Ht; and 8.81 mPa·s for 35% Ht. In addition, the viscosity value obtained for blood at 42% Ht using a macrorheometer for comparison was . As expected, as the hematocrit level increased, so did the value of viscosity. We therefore concluded that the new method proposed in this article is a reliable and accurate device that exhibits a degree of sensitivity capable of differentiating between these distinct hematocrit percentages. In addition, we measured the viscosity of DI water, MGVS60, and blood using a macroscopic rheometer for comparison purposes. Focusing on non-Newtonian fluids, we compared the behavior of the residuals of blood viscosity obtained with both rheometers.
Throughout this study, it has been demonstrated that the microrheometer using electronic detection of the fluid front can deliver viscosity results comparable to those obtained using traditional equipment, while saving time and reducing the sample size and space required. At the same time, this technique permits a better control at low shear rates for shear-thinning fluids, such as blood. The development of this experimental and theoretical method for blood will be useful in the near future to determine the presence of RBC abnormalities through the rheological characterization of blood. It could be useful for medical diagnosis in the study of diseases associated with changes in blood viscosity, operating in a wide range of shear rates [41].
Acknowledgments
Lourdes Méndez-Mora and Josep Ferré-Torres acknowledge support from Generalitat de Catalunya under the program Doctorat Industrial (2018 DI 068) and (2018 DI 064). T.A. acknowledges the Spanish MINECO for funding under grant MTM2015-71509-C2-1-R. Tomas Alarcón acknowledges further support from the Ministry of Economy and Competitivity (MINECO) for funding awarded to the Barcelona Graduate School of Mathematics under the “Maria de Maeztu” program; grant number MDM-2014-0445. Tomas Alarcón has been partially funded by the CERCA Programme of the Generalitat de Catalunya. Aurora Hernandez-Machado acknowledges support from MINECO (Spain) under project FIS2016-78883-C2-1P, Ministerio de Ciencia e Innovación (Spain) under project PID2019-106063GB-100, and AGAUR (Generalitat de Catalunya) under project 2017 SGR-1061. Claudia Trejo-Soto and Aurora Hernandez-Machado acknowledge partial support from ANID/PCI CONICYT (Chile) under project MEC80180021. Along the same lines, the authors extend their gratitude toward Rheo Diagnostics S. L. for making this research possible.
Appendix A
Appendix A.1. Flow Inside a Known Geometry
The flow passing through a known geometry can be obtained as the product of the mean velocity of the fluid multiplied by the cross-section area. The model described herein corresponds to a structure formed by a tubing of length, , and internal radius, connected into a microfluidic channel of dimensions and. The resistance inside a known geometry is defined by the relation . The pressure loss inside the tubing is:
| (A1) |
According to the mass conservation principle, in our system, ; therefore, we write velocity inside the tubing as a function of the velocity measured inside the microchannel:
| (A2) |
The flow can also be obtained from the integration process of mean velocity along with the geometry; a cylinder (the flow into a tube or pipe) or between parallel walls (a rectangular cross-section channel that satisfies the relation). Using cylindrical coordinates of the Stokes [42] equation, an expression for can be obtained, utilizing the average flow velocity inside the tube, :
| (A3) |
where and is the pressure inside the cylindrical geometry. Defining shear rate as and using the power-law model in Equation (2) (, the previous equation can be written as follows:
| (A4) |
By integrating this expression while imposing a non-slip boundary condition, , we obtain the velocity component
| (A5) |
Now, an equation for the flow rate, can be defined:
| (A6) |
To determine the flow in a rectangular microchannel, we first need to know the velocity of a plane Poiseuille flow, , using the non-slip boundary condition at the position of the solid plates :
| (A7) |
where is the pressure inside the microchannel and h(t) is the position of the fluid interface inside the microchannel.
The resulting equation for the flow rate inside the microchannel is:
| (A8) |
where is the flow inside the channel and is the flow inside the tubing. The term is the tubing length and is its internal radius. is the pressure inside the rectangular geometry and is the pressure inside the cylindrical geometry. These parameters can be seen in Figure A1a,b.
Figure A1.
(a) Schematic representation of a Poiseuille flow between two parallel walls separated at a distance b. The mean front velocity, , and the flow inside the microfluidic channel is . (b) Schematic representation of Poiseuille flow in a cylindrical tube of radius and length . The velocity, is the mean velocity of the fluid inside the cylindrical tube and is the flow inside the tube ).
Given current micromanufacturing techniques, the microchannels are mainly rectangular. Microfluidic systems often need pneumatic systems to apply pressure. To make the connections between the pneumatic systems and the microchannels, tubes are used, with different radii depending on their purpose. For this reason, this study defined a mathematical model describing the flow inside a coupled structure (cylindrical tube–rectangular channel). Due to the mass conservation principle, the flows passing through two coupled geometries are equivalent. This equivalency can be expressed as:
| (A9) |
where and are the width and height of the channel, respectively; and is the velocity inside the tubing.
From Equation (A9), the velocity inside the tubing can be written as:
| (A10) |
Appendix A.2. Pressure Drop Inside a Microfluidic System
In a closed system composed of a pressure source, a fluid reservoir, and the coupling of tubing with a microchannel of rectangular cross-section (geometries illustrated in Figure A1a,b), the pressure inside the rectangular microchannel, , is the summation of all the pressures in the system, as shown in Figure 3, is expressed in Equation (A11):
| (A11) |
where indicates the values set for the pumping source, is the pressure drop due to the coupled tubing, is the hydrostatic pressure, and is the capillary pressure. All these concepts are introduced below.
The hydrostatic pressure is the one that originates from the fluid height (the difference between the fluid reservoir height and the measurement site height). It can be calculated by multiplying the fluid height inside the reservoir relative to the microfluidic channel, density of the fluid, and gravity: .
Capillary pressure, is defined as the resistance of the hydrophobic channel walls to the fluid advancement. is dependent on the contact angle and the surface tension of the fluid, (which is dependent on temperature). Considering the geometrical parameters of a rectangular section channel, with height and width, as shown in Figure A2, can be deduced.
Figure A2.
Top view of the contact angle of a fluid–air interface inside a microfluidic channel of width ω and height b.
With the fluid front in Figure A2 expressed in terms of two radii: shown in the top view, and, the lateral view counterpart, we get:
| (A12) |
This is needed to determine the value of the curvature radii, and . Following the geometrical representation of the rectangular cross-section channel in Figure A2, the tangent curve to the radius on a circumference has an angle of 90°, and the contact angle of the front can be written as:
| (A13) |
Then:
| (A14) |
| (A15) |
| (A16) |
| (A17) |
Considering the geometrical parameters of a rectangular section channel, with height and width, can be written as:
| (A18) |
Since the contact angle on the side of the surface is also θ, and using the same trigonometric as before, is:
| (A19) |
Finally, replacing and in Equation (A12), capillary pressure can be calculated using:
| (A20) |
The value of capillarity is always positive, since the surface is hydrophobic, and the contact angle is always θ = π/2+α.
The term can be also understood as the product of the hydraulic resistance inside the microchannel, S (dependent on m, n and the channel geometric parameters and) and the channel flow, . Using Equation (A8), is:
| (A21) |
Considering the flow dependence on front velocity :
| (A22) |
, similarly to , can be expressed as the product of and. is the hydraulic resistance represented by the tubing; it depends on m, n, and the tubing parameters , . is the flow inside the tubing:
| (A23) |
Taking into account the flow dependence on flow velocity inside the tube and the relation seen in Equation (A10), for the coupled system in this study, can be written as follows in Equation (A24). Due to mass conservation inside the coupled system as shown in Figure 5, depends on m, n, the tubing, and the microchannel parameters:
| (A24) |
According to mass conservation, and taking Equation (A21) and Equation (A23) into account, an effective pressure () defined as the pressure drop caused by the pumping system, the capillary pressure, and the hydrostatic pressure into the whole system can be determined by:
| (A25) |
Moreover, Equation (A25) allows for the evaluation of the case in which one of the main terms (relative to the coupled structures of the tubing and microchannel) is much larger than the other (relative to the microchannel only). is constant and not time-dependent. Then, when the relation in Equation (A26) is satisfied,
| (A26) |
The term on the right side can be neglected and the expression for is no longer dependent on time. Now, Equation (A25) can be simplified to:
| (A27) |
was described in Equation (1) as the ratio between the mean front velocity, , and channel height, .
All the independent variables can now be expressed in the form of a constant, defined as:
| (A28) |
This depends on the fluid properties and ) and geometrical parameters of the system. Now, Equation (A27) can be written as follows, where depends on the geometrical properties of the setup and the fluid viscosity value:
| (A29) |
Appendix A.3. Newtonian Fluids
For the case of Newtonian fluids, Equation (A22) and Equation (A24) from Appendix A.2 are simplified to:
| (A30) |
| (A31) |
Due to mass conservation, the pressure for the entire system can be described (as it was in Equation (A25)) as follows:
| (A32) |
As evaluated in the non-Newtonian case, Equation (A32) allows evaluating when the term relative to the coupled geometry is much larger than the term that only depends on the microchannel, ; then, the time dependence of the equation also can be neglected in the case of the Newtonian fluids. This allows for Equation (A32) to be expressed as:
| (A33) |
Just as before, and now for the Newtonian case, can be expressed in terms of the geometry of the system and the experimentally obtained front velocity:
| (A34) |
The relation between the pressure and the mean front velocity follows a linear behavior. The equivalent expression for Equation (A29) for Newtonian fluids is:
| (A35) |
Appendix B. Viscosity Data
Author Contributions
Conceptualization, A.H.-M.; investigation, L.M.-M.; supervision, C.T.-S., A.H.-M. and T.A.; validation, C.R.-L.; visualization, J.F.-T. and S.L.; writing—original draft, L.M.-M.; writing—review and editing, M.C.-F. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by Generalitat de Catalunya, under grants number 2018 DI 068 and 2018 DI 064; Ministry of Economy and Competitivity (MINECO) under grants number MTM2015-71509-C2-1-R, MDM-2014-0445 and FIS2016-78883-C2-1P; Ministerio de Ciencia e Innovación (Spain) under grant number PID2019-106063GB-100; AGAUR (Generalitat de Catalunya) under project 2017 SGR-1061; ANID/PCI CONICYT (Chile) under project MEC80180021.
Institutional Review Board Statement
The study was conducted according to the guidelines of the Declaration of Helsinki, and approved by the University of Barcelona’s Bioethics Commission (CBUB) (Project 160016 and date of approval 1 March 2016).
Informed Consent Statement
All the samples were obtained by Banc de Sang I Teixits with the informed consent of the subjects. All the donors are anonymous and healthy. The samples were analytical surplus blood without information or personal data. The blood samples were going to be extracted for routine analysis purposes regardless of this study. The results herein obtained would not affect clinical decisions about the individuals’ care.
Data Availability Statement
Data is contained within the article.
Conflicts of Interest
The authors declare no conflict of interest.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Gupta S., Wang W.S., Vanapalli S.A. Microfluidic viscometers for shear rheology of complex fluids and biofluids. Biomicrofluidics. 2016;10:043402. doi: 10.1063/1.4955123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Merrill E.W., Gilliand E.R., Cokelet G., Shin H., Britten A., Wells R.E. Rheology of human blood, near and at zero flow. Effects of temperature and hematocrit level. Biophys. J. 1963;3:199–213. doi: 10.1016/S0006-3495(63)86816-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Merrill E.W. Rheology of blood. Physiol. Rev. 1969;49:863–888. doi: 10.1152/physrev.1969.49.4.863. [DOI] [PubMed] [Google Scholar]
- 4.Chien S., Usami S., Bertles J.F. Abnormal rheology of oxygenated blood in sickle cell anemia. J. Clin. Investig. 1970;49:623–634. doi: 10.1172/JCI106273. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Thurston G.B. Rheological parameters for the viscosity viscoelasticity and thixotropy of blood. Biorheology. 1979;16:149–162. doi: 10.3233/BIR-1979-16303. [DOI] [PubMed] [Google Scholar]
- 6.Wells R., Schmid-Schönbein H. Red cell deformation and fluidity of concentrated cell suspensions. J. Appl. Physiol. 1969;27:213–217. doi: 10.1152/jappl.1969.27.2.213. [DOI] [PubMed] [Google Scholar]
- 7.Rosencranz R., Bogen S.A. Clinical Laboratory Measurement of Serum, Plasma, and Blood Viscosity. Pathol. Patterns Rev. 2006;125:S78–S86. doi: 10.1309/FFF7U8RRPK26VAPY. [DOI] [PubMed] [Google Scholar]
- 8.Tropea C. Springer Handbook of Experimental Fluid Mechanics. Springer; Berlin/Heidelberg, Germany: 2007. [Google Scholar]
- 9.Fåhræus R., Lindqvist T. The Viscosity of The Blood in Narrow Capillary Tubes. Am. J. Physiol. Content. 1931;96:562–568. doi: 10.1152/ajplegacy.1931.96.3.562. [DOI] [Google Scholar]
- 10.Lázaro G.R., Hernández-Machado A., Pagonabarraga I. Rheology of red blood cells under flow in highly confined microchannels. II. Effect of focusing and confinement. Soft Matter. 2014;10:7207–7217. doi: 10.1039/C4SM01382D. [DOI] [PubMed] [Google Scholar]
- 11.Lázaro G.R., Hernández-Machado A., Pagonabarraga I. Collective behavior of red blood cells in confined channels. Eur. Phys. J. E. 2019;42 doi: 10.1140/epje/i2019-11805-0. [DOI] [PubMed] [Google Scholar]
- 12.Doy N., McHale G., Newton M.I., Hardacre C., Ge R., MacInnes J.M., Kuvshinov D., Allen R.W. Small volume laboratory on a chip measurements incorporating the quartz crystal microbalance to measure the viscosity-density product of room temperature ionic liquids. Biomicrofluidics. 2010;4:014107. doi: 10.1063/1.3353379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Oba T., Kido Y., Nagasaka Y. Development of laser-induced capillary wave method for viscosity measurement using pulsed carbon dioxide laser. Int. J. Thermophys. 2004;25:1461–1474. doi: 10.1007/s10765-004-5751-z. [DOI] [Google Scholar]
- 14.Pipe C.J., Majmudar T.S., McKinley G.H. High shear rate viscometry. Rheol. Acta. 2008;47:621–642. doi: 10.1007/s00397-008-0268-1. [DOI] [Google Scholar]
- 15.Choi S., Park J.K. Microfluidic rheometer for characterization of protein unfolding and aggregation in microflows. Small. 2010;6:1306–1310. doi: 10.1002/smll.201000210. [DOI] [PubMed] [Google Scholar]
- 16.Kang Y., Yeom E., Lee S.J. A microfluidic device for simultaneous measurement of viscosity and flow rate of blood in a complex fluidic network. Biomicrofluidics. 2013;7 doi: 10.1063/1.4823586. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Chevalier J., Ayela F. Microfluidic on chip viscometers. Rev. Sci. Instrum. 2008;79 doi: 10.1063/1.2940219. [DOI] [PubMed] [Google Scholar]
- 18.Clasen C., Gearing B.P., McKinley G.H. The flexure-based microgap rheometer (FMR) J. Rheol. 2006;50:883–905. doi: 10.1122/1.2357190. [DOI] [Google Scholar]
- 19.Lee T.A., Liao W.H., Wu Y.F., Chen Y.L., Tung Y.C. Electrofluidic Circuit-Based Microfluidic Viscometer for Analysis of Newtonian and Non-Newtonian Liquids under Different Temperatures. Anal. Chem. 2018;90:2317–2325. doi: 10.1021/acs.analchem.7b04779. [DOI] [PubMed] [Google Scholar]
- 20.Del Giudice F. Simultaneous measurement of rheological properties in a microfluidic rheometer. Phys. Fluids. 2020;32:052001. doi: 10.1063/5.0006060. [DOI] [Google Scholar]
- 21.Solomon D.E., Abdel-Raziq A., Vanapalli S.A. A stress-controlled microfluidic shear viscometer based on smartphone imaging. Rheol. Acta. 2016;55:727–738. doi: 10.1007/s00397-016-0940-9. [DOI] [Google Scholar]
- 22.Srivastava N., Burns M.A. Analysis of non-Newtonian liquids using a microfluidic capillary viscometer. Anal. Chem. 2006;78:1690–1696. doi: 10.1021/ac0518046. [DOI] [PubMed] [Google Scholar]
- 23.Jackson G.N.B., Ashpole K.J., Yentis S.M. The TEG® vs the ROTEM® thromboelastography/thromboelastometry systems. Anaesthesia. 2009;64:212–215. doi: 10.1111/j.1365-2044.2008.05752.x. [DOI] [PubMed] [Google Scholar]
- 24.Thakur M., Ahmed A.B. A review of thromboelastography. Int. J. Perioper. Ultrasound Appl. Technol. 2012;1:25–29. doi: 10.5005/jp-journals-10027-1006. [DOI] [Google Scholar]
- 25.Bandey H.L., Cernosek R.W., Lee W.E., Ondrovic L.E. Blood rheological characterization using the thickness-shear mode resonator. Biosens. Bioelectron. 2004;19:1657–1665. doi: 10.1016/j.bios.2004.01.001. [DOI] [PubMed] [Google Scholar]
- 26.Kang Y.J., Yoon S.Y., Lee K.H., Yang S. A highly accurate and consistent microfluidic viscometer for continuous blood viscosity measurement. Artif. Organs. 2010;34:944–949. doi: 10.1111/j.1525-1594.2010.01078.x. [DOI] [PubMed] [Google Scholar]
- 27.Yilmaz F., Gundogdu M. A critical review on blood flow in large arteries; relevance to blood rheology, viscosity models, and physiologic conditions. Korea-Aust. Rheol. J. 2008;20:197–211. [Google Scholar]
- 28.Trejo-Soto C., Costa-Miracle E., Rodriguez-Villarreal I., Cid J., Castro M., Alarcon T., Hernandez-Machado A. Front microrheology of the non-Newtonian behaviour of blood: Scaling theory of erythrocyte aggregation by aging. Soft Matter. 2017;13:3042–3047. doi: 10.1039/C6SM02412B. [DOI] [PubMed] [Google Scholar]
- 29.Diez-Silva M., Dao M., Han J., Lim C.T., Suresh S. Shape and biomechanics characteristics of human red blood cells in health and disease. MRS Bull. 2010;35:382–388. doi: 10.1557/mrs2010.571. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Apostolidis A.J., Moyer A.P., Beris A.N. Non-Newtonian effects in simulations of coronary arterial blood flow. J. Nonnewton. Fluid Mech. 2016;233:155–165. doi: 10.1016/j.jnnfm.2016.03.008. [DOI] [Google Scholar]
- 31.Van Kempen T.H.S., Peters G.W.M., Van De Vosse F.N., Van Kempen T.H.S., Van De Vosse F.N., Peters G.W.M. A constitutive model for the time-dependent, nonlinear stress response of fibrin networks. Biomech. Model. Mechanobiol. 2015;14:995–1006. doi: 10.1007/s10237-015-0649-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Armstrong M., Horner J., Clark M., Deegan M., Hill T., Keith C., Mooradian L. Evaluating rheological models for human blood using steady state, transient, and oscillatory shear predictions. Rheol. Acta. 2018;57:705–728. doi: 10.1007/s00397-018-1109-5. [DOI] [Google Scholar]
- 33.Ramaswamy B., Yeh Y.T.T., Zheng S.Y. Microfluidic device and system for point-of-care blood coagulation measurement based on electrical impedance sensing. Sens. Actuators B Chem. 2013;180:21–27. doi: 10.1016/j.snb.2011.11.031. [DOI] [Google Scholar]
- 34.Trejo-Soto C., Costa-Miracle E., Rodriguez-Villarreal I., Cid J., Alarcón T., Hernández-Machado A. Capillary filling at the microscale: Control of fluid front using geometry. PLoS ONE. 2016;11 doi: 10.1371/journal.pone.0153559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bruus H. Theoretical Microfluidics. Oxford University Press; Oxford, UK: 2006. [Google Scholar]
- 36.Baskurt O. Handbook of Hemorheology and Hemodynamics. IoS Press; Amsterdam, The Neetherlands: 2007. [Google Scholar]
- 37.Adamson A.W., Gast A.P., Wiley J. Physical Chemistry of Surfaces. 6th ed. Interscience Publishers; New York, NY, USA: 1967. [Google Scholar]
- 38.Murillo C. In: Fisicoquímica. 5th ed. Levine I.N., editor. Volume 2 McGraw Hill; Madrid, Spain: 2004. [Google Scholar]
- 39.Bruus H. Understanding Rheology. Oxford University Press; New York, NY, USA: 2008. [Google Scholar]
- 40.Thurston G.B., Henderson N.M. Effects of flow geometry on blood viscoelasticity. Biorheology. 2006;43:729–746. [PubMed] [Google Scholar]
- 41.Colomer J., Hernández-Machado A., Alarcón Cor T. Method and Apparatus for Measuring Rheological Properties of Newtonian and Non-Newtonian Fluids. EP3093647A1. International Patent. 2016 Nov 17;
- 42.Guyon E., Hulin J.-P., Petit L., Mitescu C.D. Physical Hydrodynamics. 2nd ed. Oxford University Press; Oxford, UK: 2015. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data is contained within the article.












