SARS-CoV-2 antigen-detecting rapid diagnostic tests (Ag-RDTs) provide laboratory-independent results at the point of care and are powerful tools for public health interventions. Clinical and analytical studies, published in 2021, showed SARS-CoV-2 Ag-RDT detection thresholds related to the presence of infectious virus in symptomatic SARS-CoV-2 infections.1, 2 However, the majority of Ag-RDT validation studies were done before SARS-CoV-2 variants of concern (VOC) or interest (VOI) emerged, with the VOCs currently outcompeting earlier lineages.3 To date, data on routine diagnostic performance for VOCs and VOIs are sparse.4, 5 Furthermore, clinical validation studies comparing multiple VOCs in parallel are hardly feasible.
We investigated the analytical sensitivity of nine commercially available Ag-RDTs using cultured SARS-CoV-2, comparing lineage B.1.610 (first COVID-19 pandemic wave in Europe) with VOCs B.1.1.7, B.1.351, and P.1, and VOI P.2.
Infectious titres and RNA copies of virus stocks grown in Vero E6 were quantified by plaque titration (for infectious titres) and RT-PCR (E gene). Isolates were tested in serial dilutions, starting with 5·44 log10 PFU/mL, except for P.1, which had a maximum titre of log10 4·24 PFU/mL. An infectious titre of 5·44 log10 PFU/mL corresponded to 10.26, 12.11, 9.86, and 11·23 log10 RNA copies per mL for B.1.610, B.1.1.7, B.1.351 and P.2. For P.1, the infectious titre of 4·24 log10 PFU/mL corresponded to 11·81 log10 RNA copies per mL.
Ag-RDT assays were done according to the manufacturers' instructions, with the exception that 5 μL of virus dilution was directly added to the proprietary buffer, and then applied to the Ag-RDT in duplicates under BSL3 conditions. Results were read independently by two individuals. Any visible test band in the presence of a visible control band was considered as positive. Ag-RDT buffer without virus was used as negative control.
When analysing results normalised to PFU/mL, comparable or better performance to the early-pandemic lineage was observed for B.1.1.7, B.1.351, P.1, and P.2 for all assays (appendix). Overall sensitivity and specificity for individual isolates varied between Ag-RDTs, with the best-performing assay positive at dilutions as low as 2·43 log10 PFU/mL and the lower-sensitive assays positive at 4·54 log10 PFU/mL. Consistently, the highest sensitivity was seen for P.1 and P.2. Although testing for analytical sensitivity with cultured virus cannot fully replace clinical data, our data provide reassuring results for the use of Ag-RDTs to diagnose VOCs. Phenotypic properties, such as a large difference in the RNA–infectious virus ratio, could hint at production of defective viral particles and their effect on diagnostic test performance should be further investigated.
This work was supported by the Swiss National Science Foundation (grant number 196383), the Fondation Ancrage Bienfaisance du Groupe Pictet, and the Foundation for Innovative New Diagnostics (FIND). The Swiss National Science Foundation and the Fondation Ancrage Bienfaisance du Groupe Pictet had no role in data collection, analysis, or interpretation. Antigen-detecting rapid diagnostic tests were provided by FIND and FIND was involved in methodology, data analysis, interpretation and writing. JAE and CE are employees of FIND. We declare no competing interests.
Supplementary Material
References
- 1.Nordgren J, Sharma S, Olsson H. SARS-CoV-2 rapid antigen test: high sensitivity to detect infectious virus. J Clin Virol. 2021;140 doi: 10.1016/j.jcv.2021.104846. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Corman VM, Haage VC, Bleicker T. Comparison of seven commercial SARS-CoV-2 rapid point-of-care antigen tests: a single-centre laboratory evaluation study. Lancet Microbe. 2021 doi: 10.1016/S2666-5247(21)00056-2. published online April 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.WHO Weekly epidemiological update on COVID-19. May, 11 2021. https://www.who.int/publications/m/item/weekly-epidemiological-update-on-covid-19---11-may-2021
- 4.Jungnick S, Hobmaier B, Mautner L. Detection of the new SARS-CoV-2 variants of concern B.1.1.7 and B.1.351 in five SARS-CoV-2 rapid antigen tests (RATs), Germany, March 2021. Euro Surveill. 2021;26 doi: 10.2807/1560-7917.ES.2021.26.16.2100413. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Public Health England Guidance SARS-CoV-2 lateral flow antigen tests: evaluation of VUI-202012/01. Dec 23, 2020. https://www.gov.uk/government/publications/sars-cov-2-lateral-flow-antigen-tests-evaluation-of-vui-20201201/sars-cov-2-lateral-flow-antigen-tests-evaluation-of-vui-20201201
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.