. 2021 May 4;35(12):e9098. doi: 10.1002/rcm.9098

Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns

Thomas H Walter 1,, Melvin Blaze M T 1, Cheryl Boissel 1
PMCID: PMC8244038  PMID: 33830546

Abstract

Rationale

Mixed‐mode reversed‐phase/anion exchange liquid chromatography is useful for separations of mixtures containing anions (e.g. ionized acids). However, when using this form of liquid chromatography with mass spectrometry detection, the bleed of amine‐containing hydrolysis products from the columns may cause ion suppression or enhancement.

Methods

Using electrospray ionization tandem quadrupole mass spectrometry detection, we determined the ion suppression or enhancement caused by column bleed for three mixed‐mode reversed‐phase/weak anion‐exchange columns containing stationary phases that differ in chemical structure. Two of the stationary phases are based on silica particles, while the third uses ethylene‐bridged hybrid organic/inorganic particles, which have improved hydrolytic stability. Mixtures of acidic and basic analytes were combined with the chromatography flow postcolumn, both with and without a column, and their mass spectrometry ion signal responses (peak areas) were determined. The ratio of signal response with and without a column is the matrix factor. Positive ion electrospray measurements were carried out using 0.1% formic acid (pH ~ 2.7) as a mobile phase additive, and 10mM ammonium formate (pH ~ 6.4) was used for negative ion electrospray detection.

Results

The matrix factors under both positive and negative ionization modes were closest to 1 (0.74–1.16) for the hybrid particle‐based columns, showing minimal ion suppression or enhancement. In contrast, the silica‐based columns gave matrix factors ranging from 0.04 to 1.86, indicating high levels of ion suppression or enhancement. These results may be explained by the differences in the structures of the stationary phases, which affect the relative amounts of hydrolysis products that elute from the columns.

Conclusions

The low levels of mass spectrometry ion suppression or enhancement caused by column bleed from the hybrid particle‐based columns should allow for accurate quantitative mass spectrometric detection combined with mixed‐mode reversed‐phase/weak anion‐exchange chromatography.

1. INTRODUCTION

Mixed‐mode reversed‐phase (RP)/anion‐exchange (AX) columns, which contain stationary phases having both hydrophobic and AX functionalities, provide a means to retain and separate anions (e.g. ionized acids) under RP conditions. 1 , 2 , 3 , 4 , 5 , 6 These columns have been most commonly used with optical detection, for example UV absorbance. When considering their use with electrospray ionization mass spectrometry (ESI‐MS), the hydrolytic stability of the stationary phase is an area of concern. Many mixed‐mode RP/AX stationary phases, particularly those based on silane‐bonded silica, suffer from relatively poor hydrolytic stability. This is reflected in the limited recommended pH ranges for these columns. For example, one brand of RP/AX columns is recommended for use in the pH range 1.5–4, 7 while a second has a range of 2.5–7.5. 8 Organic polymer‐based mixed‐mode columns have been reported to be stable over a wide pH range (e.g. 1–14), but suffer from low efficiencies. 9 In recent work to improve the hydrolytic stability of silica‐based mixed‐mode RP/AX materials, a polymer‐coating approach was used. 10 The resulting stationary phase was shown to have good stability when used with a pH 5 mobile phase at 60°C. To improve both acid and base stability without sacrificing column efficiency, a mixed‐mode RP/AX stationary phase based on ethylene‐bridged hybrid organic/inorganic particles 11 was recently developed. This stationary phase has been shown to be stable from pH 2 to 10. 12

Ion suppression or enhancement is an important concern in quantitative liquid chromatography (LC) ESI‐MS. 13 , 14 , 15 , 16 The primary cause is linked to components that co‐elute with the analytes and affect the ionization process. These components may originate from the sample, from mobile phase additives or from contaminants introduced during sample preparation. 13 , 15 , 16 Hydrolysis products eluted from the column stationary phase, referred to as column bleed, are another potential source. Under acidic conditions, silane bonded phases hydrolyze to form organosilanols, which may condense to form organosiloxanes. 17 Under basic conditions, not only is the bonded phase susceptible to hydrolysis, but the underlying silica particles may begin to be attacked, forming soluble silicates. 18 , 19 The stability of different bonded phases varies over a wide range, depending on the attachment chemistry, the structure and surface concentration of the bonded organic groups, and the chemical and physical properties of the underlying particles. 20 Elution of bonded‐phase hydrolysis products containing amine groups from mixed‐mode RP/AX columns has been shown to cause problems due to their strong MS response. These problems include a high background signal in total ion current mode 21 and low ion transmission in high‐resolution MS. 22

Here, we compare the magnitude of ion suppression or enhancement due to column bleed for hybrid particle‐based mixed‐mode columns relative to two silica‐based mixed‐mode RP/AX columns. These three stationary phases employ different bonding chemistries as well as different particles. We describe these differences. The column bleed profiles of the three materials were determined using both an acidic mobile phase with positive ionization ESI (ESI+) and a neutral mobile phase with negative ionization ESI (ESI−). To quantify ion suppression or enhancement caused by hydrolysis products eluted from these columns, we monitored seven small‐molecule analytes under ESI+ and five small‐molecule analytes under ESI− using analyte‐specific selected reaction monitoring (SRM) transitions. The analyte mixtures were combined with the LC mobile phase flow postcolumn, both in the presence and in the absence of a column (using a union instead of a column). The ratio of signal response with and without a column is the matrix factor.

2. METHODS

2.1. Chemicals

LC‐MS grade acetonitrile was purchased from Honeywell (Muskegon, MI) and MS grade formic acid (FA) was purchased from Fisher Scientific (Hampton, NH). Ammonium formate and all analytes were purchased from Millipore‐Sigma (Burlington, MA). Deionized water was produced using a Millipore Milli‐Q system.

2.2. Sample and mobile phase preparation

The sample for the ESI+ test contained 10 μg/mL thiourea, 20 μg/mL thymine, 10 μg/mL adenine, 2 μg/mL 5‐fluorocytosine, 5 μg/mL thiamine, 15 μg/mL tryptophan and 1.5 μg/mL niflumic acid in 1:1 acetonitrile/0.1% (v/v) FA in water. For the ESI− test, the sample contained 10 μg/mL citric acid, 10 μg/mL malic acid, 10 μg/mL guanosine‐5′‐monophosphate disodium salt hydrate, 10 μg/mL thymidine‐5′‐monophosphate disodium salt hydrate and 1.5 μg/mL niflumic acid in 1:1 acetonitrile/0.1% (v/v) FA in water. The aqueous and organic mobile phases for the ESI+ test were 0.1% (v/v) FA in water and in acetonitrile, respectively, and for the ESI− test, the aqueous and organic mobile phase were 10mM ammonium formate in water and acetonitrile, respectively.

2.3. Instrumentation and columns

All analyses were performed using two ACQUITY UPLC H‐Class systems coupled to Xevo TQD mass spectrometers (Waters Corporation, Milford, MA): one for ESI+ tests and the other for ESI− tests. The IntelliStart fluidics system built into the mass spectrometers was used to infuse the analyte solution by combining with the LC flow postcolumn into the MS probe. Both UPLC systems were equipped with a quaternary solvent manager, a flow‐through needle sample manager and a column manager. Data acquisition and analysis were performed using MassLynx V4.2. Atlantis PREMIER BEH C18 AX columns (column W) (1.7 μm, 2.5 μm and 5 μm, 2.1 × 150 mm) were also obtained from Waters Corporation (Milford, MA). Primesep B columns (column P) (5 μm, 2.1 × 150 mm) were obtained from SiELC Technologies (Wheeling, IL) and Acclaim Mixed‐Mode WAX‐1 columns (column A) (5 μm, 2.1 × 150 mm) were purchased from Thermo Fisher Scientific (Waltham, MA).

2.4. Experimental plan

To evaluate ESI‐MS signal suppression or enhancement due to column stationary phase bleed, seven different analytes under ESI+ and five different analytes under ESI− were monitored using analyte‐specific SRM transitions. The analytes chosen are readily detectable under ESI+ and/or ESI− modes of ionization and include acidic, basic and neutral analytes with molecular weights ranging from 76 to 363 Da. The LC and MS conditions are given in Tables 1 and 2. The analyte mixtures were infused postcolumn into the LC mobile phase flow before entering the ESI‐MS setup, as depicted in Figure 1A. A mobile phase gradient was carried out, with the acetonitrile content increasing from 10% to 90%. A double gradient was executed five times for each column, and the results for the fifth run were used for the calculations. The infusions of the analyte mixtures were performed at 2.5 min intervals at a flow rate of 200 μL/min for 0.5 min while the acetonitrile content was held at 90% and 10% (see Figures 1B and 1C). For the ESI+ measurements, five replicate infusions were performed during both the 10% and 90% organic portions of the gradient, and the results for the second through the fifth infusions were used in the calculations. For the ESI− measurements, five replicate infusions were performed during the 90% organic portion of the gradient, and four infusions during the 10% organic portion, and the results for the third through the fifth (90% organic) and second through the fourth infusions (10% organic) were used in the calculations. During the second portion of the gradient (from 43.0 to 93.0 min), full‐scan mass spectra (m/z 50–1000) of the column bleed were recorded during the 90% and 10% acetonitrile portions of the gradient. The ESI‐MS signal responses for the postcolumn‐infused analytes were measured in the presence and absence of the column for three W columns (1.7 μm, 2.1 × 50 mm; 2.5 μm, 2.1 × 50 mm; and 5 μm, 2.1 × 150 mm), one P column (5 μm, 2.1 × 150 mm) and one A column (3 μm, 2.1 × 150 mm). Separate columns were used for ESI+ and ESI− experiments. The matrix factors imparted by the column stationary phase bleed on these analytes were calculated using the following equation for each individual measurement:1

2.4.

TABLE 1.

LC conditions

Mobile phase A (ESI+) 0.1% (v/v) formic acid in water
Mobile phase B (ESI+) 0.1% (v/v) formic acid in acetonitrile
Mobile phase A (ESI−) 10mM ammonium formate in water
Mobile phase B (ESI−) Acetonitrile
Flow rate: 0.2 mL/min
Column temperature 50°C
Time (min) %A %B Curve
Initial 90 10 Initial
7.5 10 90 6
22.5 10 90 6
24.0 90 10 6
46.5 90 10 6
54.0 10 90 6
69.0 10 90 6
70.5 90 10 6
93.0 90 10 6
96.0 90 10 6

TABLE 2.

MS conditions

Source conditions
Capillary voltage (kV) ESI+, 2.90; ESI−, 2.0
Cone voltage (V) 45
Desolvation temperature 250°C
Desolvation gas (L/h) 600
SRM transitions, ESI+
Compound name Precursor ion (m/z) Product ion (m/z) Collision energy (V)
Thiourea 76.91 43.05 80
Thymine 126.94 53.11 26
5‐Fluorocytosine 129.97 57.95 32
Adenine 136.02 93.99 46
Tryptophan 205.06 145.99 20
Thiamine 265.07 121.95 16
Niflumic acid 282.92 144.96 40
SRM transitions, ESI−
Compound name Precursor ion (m/z) Product ion (m/z) Collision energy (V)
Malic acid 133.00 71.00 14
Citric acid 191.10 111.00 14
Niflumic acid 281.10 237.20 35
Thymidine‐5′‐monophosphate 321.10 194.90 22
Guanosine‐5′‐monophosphate 362.10 78.90 28

FIGURE 1.

FIGURE 1

(A) Schematic of the instrument setup for the matrix factor study. (B) Chart showing the mobile phase double gradient (% acetonitrile versus time). (C) Representative SRM signal response showing the timing of the postcolumn infusions relative to the mobile phase gradient. The second portion of the gradient (43–93 min) was used to record mass spectra from m/z 50 to 1000

MatrixFactor(MF)=AnalytepeakareainthepresenceofcolumnAnalytepeakareaintheabsenceofcolumn

3. RESULTS AND DISCUSSION

3.1. Chemical and physical properties of the three stationary phases

The properties of the three stationary phases are summarized and compared in Table 3. The stationary phase of column W is based on ethylene‐bridged hybrid particles and contains both C18 groups and tertiary alkylamine moieties, the latter creating a positive surface charge below approximately pH 8–9. 12 , 23 This stationary phase is endcapped to reduce the concentration of residual silanols and to improve the stability of the stationary phase to basic mobile phases. The stationary phases for columns P and A are based on silica particles. The structures of the bonded phases for these columns have previously been reported 1 and are shown in Figure 2. The surface concentrations of these bonded phases were calculated based on their carbon contents and surface areas. 24 The nature of the attachment to the silica surface for these bonded phases was established using 29Si cross‐polarization magic angle spinning NMR spectroscopy. 25 , 26 The results indicate that the stationary phase of column A is monofunctionally bonded, while that of column P is trifunctionally bonded. Unlike column W, the stationary phases for columns A and P contain bonded groups that incorporate both hydrophobic and AX moieties together in the same structure. In contrast, the stationary phase of column W is bonded with separate hydrophobic and AX groups, with the AX group surface concentration being approximately five times lower than that of the C18 groups.

TABLE 3.

Chemical and physical properties of stationary phases evaluated

Particle properties Bonded‐phase properties
Column Material Pore size (Å) Pore volume (cm3/g) Surface area (m2/g) %C Surface concentration of primary group (μmol/m2) Endcapping
W Ethylene‐bridged hybrid 95 0.7 270 16.83 1.6 Yes
P Silica 100 0.9 300 6.96 1.4 No
A Silica 120 0.9 300 13.47 2.6 Unknown

FIGURE 2.

FIGURE 2

Schematic representations of bonded phases for columns A, P and W

3.2. Effect of column bleed on ESI+ MS signal response

Elution of bonded phase hydrolysis products from columns can affect the signal intensity for co‐eluting analytes when using ESI‐MS. Bonded‐phase hydrolysis products containing ionizable groups, such as amines, are particularly problematic because of their strong MS response. Full‐scan ESI+ mass spectra (m/z 50–1000) obtained using 0.1% FA (pH ~ 2.7) as a mobile phase additive with columns A and P showed very intense MS peaks that are attributed to bonded‐phase hydrolysis products (see Figures 3 and 4). The mass spectra obtained at a mobile phase composition of 10/90 acetonitrile/aqueous 0.1% FA are shown in Figure 3, while the mass spectra in Figure 4 were recorded at a composition of 90/10 acetonitrile/aqueous 0.1% FA. For column P, for both mobile phase compositions a strong peak saturating the detector was observed at m/z 305.9. This is consistent with the protonated form of the trisilanol hydrolysis product expected to be formed from the bonded‐phase structure shown in Figure 2, with an elemental formula of [C15H36NO3Si]+. The less intense peak at m/z 474.6 observed in 90/10 acetonitrile/aqueous 0.1% FA (Figure 4) is tentatively ascribed to the protonated form of a related bonded‐phase hydrolysis product having two dodecyl groups attached to the nitrogen atom, with an elemental formula of [C27H60NO3Si]+. This appears plausible considering the synthetic chemistry believed to be used to make the stationary phase of column P. 27 For column A, for both mobile phase compositions an intense peak saturating the detector was observed at m/z 345.1. This is attributed to the protonated form of the silanol hydrolysis product expected to be formed from the structure shown in Figure 2, with an elemental formula of [C18H41N2O2Si]+. Column W showed a few peaks with intensities 25–35 times lower than those of the other two columns. This is most evident in the mass spectrum observed in 90/10 acetonitrile/aqueous 0.1% FA (see Figure 4).

FIGURE 3.

FIGURE 3

Representative ESI+ mass spectra of column bleed for the three columns and the union at a mobile phase composition of 10:90 acetonitrile/0.1% (v/v) FA in water. The major peaks are labelled with the proposed elemental formulas. Peaks that saturated the MS detector are marked with asterisks. The vertical scales were magnified tenfold for the bottom two mass spectra

FIGURE 4.

FIGURE 4

Representative ESI+ mass spectra of column bleed for the three columns and the union at a mobile phase composition of 90:10 acetonitrile/0.1% (v/v) FA in water. The major peaks are labelled with the proposed elemental formulas. Peaks that saturated the MS detector are marked with asterisks. The vertical scales were magnified tenfold for the bottom two mass spectra

Figure 5 shows a comparison of the matrix factors for the representative analytes measured under ESI+ with 0.1% FA mobile phase at 90% acetonitrile (Figure 5A) and at 10% acetonitrile (Figure 5B) for the three columns. A matrix factor value of 1 indicates the signal responses in the presence and the absence of the column are the same, signifying no matrix effect. A matrix factor value less than 1 indicates that the MS signal response in the presence of the column is lower than that in the absence of the column, signifying suppression of the analyte MS signal response primarily caused by column bleed. A matrix factor greater than 1 indicates the MS signal response in the presence of the column is greater than in the absence of column, signifying enhancement in the analyte MS signal response primarily caused by column bleed. Columns P and A (black and red bars, respectively) exhibited significant signal suppression for many of the analytes, as shown by matrix factors below 0.6. This was the case at both 10% and 90% acetonitrile. In contrast, column W (blue bars) showed matrix factors ranging from 0.74 to 0.95, indicating little signal suppression. The results for the latter column represent the grand average of the values obtained using one 1.7 μm, one 2.5 μm and one 5 μm column with five replicate measurements for each column, and the error bars show ± one pooled standard deviation. The results for columns A and P represent the average of five measurements each and the error bars show ± one standard deviation. We attribute the lower matrix factors for column W to a lower concentration of amine‐containing bonded‐phase hydrolysis products.

FIGURE 5.

FIGURE 5

Comparison of matrix factors from column bleed for column P (black bars with vertical lines), column A (red bars with horizontal lines) and column W (blue solid bars). The error bars show one standard deviation for five measurements each on one column A and one column P and fifteen measurements on three W columns. (A) 0.1% FA mobile phase additive and ESI+ detection at 10% acetonitrile; (B) 0.1% FA mobile phase additive and ESI+ detection at 90% acetonitrile

3.3. Effect of column bleed on ESI− MS signal response

In the ESI− experiments performed using a 10mM ammonium formate (pH ~ 6.4) mobile phase, the mass spectra of the column bleed showed a number of moderate intensity peaks. The mass spectrum obtained for column P at the 10/90 acetonitrile/aqueous mobile phase composition (Figure 6) shows a series of peaks that match the known m/z values of silicate oligomers containing from three to ten silicon atoms. 28 These likely arise from hydrolysis of the silica particles. The mobile phase pH of approximately 6.4 used for this experiment was above the recommended limit for column P. The mass spectrum observed for column A at the same mobile phase composition does not show these peaks, but one peak has a m/z value (421.4) consistent with a bonded‐phase hydrolysis product bonded to a silicate monomer, with an elemental formula of [C18H41N2O5Si2] (Figure 6). This peak is more intense in the 90/10 acetonitrile/aqueous composition, and a peak at m/z 259.2 is observed that is consistent with a bonded‐phase hydrolysis product where the amide group has been hydrolyzed to form a carboxylate group, giving a species with an elemental formula of [C13H27O3Si] (Figure 7). Column W showed only a few peaks with intensities more than 10 times lower than those of the other two columns.

FIGURE 6.

FIGURE 6

Representative ESI− mass spectra of column bleed for the three columns and the union at a mobile phase composition of 10:90 acetonitrile/10mM ammonium formate in water. The major peaks are labelled with the proposed elemental formulas. The vertical scales were magnified tenfold for the bottom two mass spectra

FIGURE 7.

FIGURE 7

Representative ESI− mass spectra of column bleed for the three columns and the union at a mobile phase composition of 90:10 acetonitrile/10mM ammonium formate in water. The major peaks are labelled with the proposed elemental formulas. The vertical scales were magnified tenfold for the bottom two mass spectra

Shown in Figure 8 is a comparison of the matrix factors for the three columns for the representative analytes measured under ESI− with 10mM ammonium formate mobile phase additive at 10% acetonitrile (Figure 8A) and at 90% acetonitrile (Figure 8B). Column P (black bars) showed matrix factors significantly less than 1 for all the analytes, indicating ion suppression. In contrast, column A (red bars) showed matrix factors slightly greater than 1 for many of the analytes, indicating ion enhancement. While similar trends were observed at 10% and 90% acetonitrile, more severe suppression for column P and enhancement for column A was observed at 90% acetonitrile. In contrast, at both 10% and 90% acetonitrile, column W (blue bars) showed matrix factors ranging from 0.9 to 1.3, indicating little to no suppression or enhancement from column bleed. We attribute this to the greater hydrolytic stability of the hybrid particle‐based stationary phase for column W.

FIGURE 8.

FIGURE 8

Comparison of matrix factors from column bleed for column P (black bars with vertical lines), column A (red bars with horizontal lines) and column W (blue solid bars). The error bars show one standard deviation for four (A) or five (B) measurements each on one column A and one column P and twelve (A) or fifteen (B) measurements on three W columns. (A) 10mM ammonium formate and ESI− detection at 10% acetonitrile; (B) 10mM ammonium formate and ESI− detection at 90% acetonitrile

4. CONCLUSIONS

These results show that the silica‐based RP/AX columns (columns A and P) exhibited significant MS signal suppression or enhancement, which appears to be caused by hydrolysis products from the bonded phases and/or the silica particles. In contrast, the hybrid particle‐based column W exhibited minimal MS signal suppression or enhancement caused by column bleed. This is due to the improved hydrolytic stability of the stationary phase for column W, which is a consequence of its ethylene‐bridged hybrid particles and bonded‐phase structure, containing separate C18 and AX groups, both attached using stable bonding chemistries. This improved stability makes column W well suited for LC/MS applications, particularly for quantitative methods.

PEER REVIEW

The peer review history for this article is available at https://publons.com/publon/10.1002/rcm.9098.

ACKNOWLEDGMENTS

The authors thank Yuehong Xu, Xiang‐jin Song, Bei Niu, Erik Goulet, Stephen McBride and ZhiLin Zhang for carrying out the analytical tests for the stationary phases and Maureen Deloffi and Mike Dion for packing the Atlantis PREMIER BEH C18 AX columns used in this work.

Walter TH, Blaze M. T. M, Boissel C. Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns. Rapid Commun Mass Spectrom. 2021;35:e9098. 10.1002/rcm.9098

DATA AVAILABILITY STATEMENT

The data that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1. Zhang K, Liu X. Mixed‐mode chromatography in pharmaceutical and biopharmaceutical applications. J Pharm Biomed Anal. 2016;130:19‐34. [DOI] [PubMed] [Google Scholar]
  • 2. Liu X, Pohl C. A weak anion‐exchange/reversed‐phase mixed‐mode HPLC column and its applications. Am Lab. 2007;39:22‐25. [Google Scholar]
  • 3. Lämmerhofer M, Richter M, Wu J, Nogueira R, Bicker W, Lindner W. Mixed‐mode ion exchangers and their comparative chromatographic characterization in reversed‐phase and hydrophilic interaction chromatography elution modes. J Sep Sci. 2008;31(14):2572‐2588. [DOI] [PubMed] [Google Scholar]
  • 4. Zhang L, Dai Q, Qiao X, Yu C, Qin X, Yan H. Mixed‐mode chromatographic stationary phases: Recent advancements and its applications for high‐performance liquid chromatography. Trends Anal Chem. 2016;82:143‐163. [Google Scholar]
  • 5. Sýkora D, Rezanka P, Záruba K, Král V. Recent advances in mixed‐mode chromatographic stationary phases. J Sep Sci. 2019;42(1):89‐129. [DOI] [PubMed] [Google Scholar]
  • 6. Mansour FR, Danielson ND. Multimodal liquid chromatography of small molecules. Anal Methods. 2013;5(19):4955‐4972. [Google Scholar]
  • 7. https://www.sielc.com/primesep-b.html. Accessed February 25, 2021.
  • 8. https://www.thermofisher.com/order/catalog/product/064984#/064984. Accessed February 25, 2021.
  • 9. Eleveld JT, Claessens HA, Ammerdorffer JL, van Herk AM, Cramers CA. Evaluation of mixed‐mode stationary phases in liquid chromatography for the separation of charged and uncharged oligomer‐like model compounds. J Chromatogr A. 1994;677(2):211‐227. [Google Scholar]
  • 10. Bäurer S, Zimmermann A, Woiwode U, et al. Stable‐bond polymeric reversed‐phase/weak anion‐exchange mixed‐mode stationary phases obtained by simultaneous functionalization and crosslinking of a poly(3‐mercaptopropyl)methylsiloxane‐film on vinyl silica via thiol‐ene double click reaction. J Chromatogr A. 2019;1593:110‐118. [DOI] [PubMed] [Google Scholar]
  • 11. Wyndham KD, O'Gara JE, Walter TH, et al. Characterization and evaluation of C18 HPLC stationary phases based on ethyl‐bridged hybrid organic/inorganic particles. Anal Chem. 2003;75(24):6781‐6788. [DOI] [PubMed] [Google Scholar]
  • 12. Walter TH, Alden BA, Field JA, et al. Characterization of a highly stable mixed‐mode reversed‐phase/weak anion‐exchange stationary phase based on hybrid organic/inorganic particles. J Sep Sci. 2021;44(5):1005‐1014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Gosetti F, Mazzucco E, Zampieri D, Gennaro MC. Signal suppression/enhancement in high‐performance liquid chromatography tandem mass spectrometry. J Chromatogr A. 2010;1217(25):3929‐3937. [DOI] [PubMed] [Google Scholar]
  • 14. Zhou W, Yang S, Wang PG. Matrix effects and application of matrix effect factor. Bioanalysis. 2017;9(23):1839‐1844. [DOI] [PubMed] [Google Scholar]
  • 15. Antignac J‐P, de Wasch K, Monteau F, De Brabander H, Andre F, Le Bizec B. The ion suppression phenomenon in liquid chromatography‐mass spectrometry and its consequences in the field of residue analysis. Anal Chim Acta. 2005;529(1‐2):129‐136. [Google Scholar]
  • 16. Furey A, Moriarty M, Bane V, Kinsella B, Lehane M. Ion suppression: A critical review on causes, evaluation, prevention and applications. Talanta. 2013;115:104‐122. [DOI] [PubMed] [Google Scholar]
  • 17. Crowther JB, Fazio SD, Schiksnis R, Marcus S, Hartwick RA. Analysis of the acid hydrolysis products of monofunctional chemically bonded stationary phases for high‐performance liquid chromatography using capillary gas chromatography. J Chromatogr A. 1984;289:367‐375. [Google Scholar]
  • 18. Kirkland JJ, van Straten MA, Claessens HA. High pH mobile phase effects on silica‐based reversed‐phase high‐performance liquid chromatographic columns. J Chromatogr A. 1995;691(1‐2):3‐19. [DOI] [PubMed] [Google Scholar]
  • 19. Kirkland JJ, Henderson JW, DeStefano JJ, van Straten MA, Claessens HA. Stability of silica‐based, endcapped columns with pH 7 and 11 mobile phases for reversed‐phase high‐performance liquid chromatography. J Chromatogr A. 1997;762(1‐2):97‐112. [DOI] [PubMed] [Google Scholar]
  • 20. Claessens HA, van Straten MA. Review on the chemical and thermal stability of stationary phases for reversed‐phase liquid chromatography. J Chromatogr A. 2004;1060(1‐2):23‐41. [PubMed] [Google Scholar]
  • 21. Zimmermann A, Horak J, Sievers‐Engler A, et al. Surface‐crosslinked poly(3‐mercaptopropyl)‐methylsiloxane‐coatings on silica as a new platform for low‐bleed mass spectrometry‐compatible functionalized stationary phases synthesized via thiol‐ene click reaction. J Chromatogr A. 2016;1436:73‐83. [DOI] [PubMed] [Google Scholar]
  • 22. Schulze B, Bader T, Seitz W, Winzenbacher R. Column bleed in the analysis of highly polar substances: An overlooked aspect in HRMS. Anal Bioanal Chem. 2020;412(20):4837‐4847. [DOI] [PubMed] [Google Scholar]
  • 23. Kadlecová Z, Kalíková K, Folprechtová D, Tesařová E, Gilar M. Method for evaluation of ionic interactions in liquid chromatography. J Chromatogr A. 2020;1625:46130. [DOI] [PubMed] [Google Scholar]
  • 24. Berendsen GE, de Galan G. Preparation and chromatographic properties of some chemically bonded phases for reversed‐phase liquid chromatography. J Liq Chromatogr. 1978;1(5):561‐586. [Google Scholar]
  • 25. Sindorf DW, Maciel GE. Cross‐polarization magic‐angle‐spinning silicon‐29 nuclear magnetic resonance study of silica gel using trimethylsilane bonding as a probe of surface geometry and reactivity. J Phys Chem. 1982;86(26):5208‐5219. [Google Scholar]
  • 26. Sindorf DW, Maciel GE. Solid‐state NMR studies of the reactions of silica surfaces with polyfunctional chloromethylsilanes and ethoxymethylsilanes. J Am Chem Soc. 1983;105(12):3767‐3776. [Google Scholar]
  • 27. Orlovsky V, Zelechonok Y. Universal bonded phase material for chromatographic separation. US patent 7,045,059 B2, 2006.
  • 28. Pelster SA, Schrader W, Schüth F. Monitoring temporal evolution of silicate species during hydrolysis and condensation of silicates using mass spectrometry. J Am Chem Soc. 2006;128(13):4310‐4317. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.