INTRODUCTION
Shoulder pain affects over 30% of the global adult population, and costs Americans 7 billion dollars each year.1 Pathology that causes shoulder pain includes degenerative causes such as rotator cuff tendinopathy, osteoarthritis, adhesive capsulitis, and more acute causes including tendon tears and proximal humerus fractures. These directly lead to impairments in shoulder motion, daily activities, and decreased function which lead to disability. Conservative treatments are the mainstay of initial care and include over-the-counter and prescription pharmaceuticals.2–4 However, there are a myriad of public and individual health problems associated with analgesic use especially in older adults, including adverse effects, cost, non-compliance, and physical dependence with opioid analgesics.
A second common conservative treatment is physical therapy which incorporates physical modalities including ultrasound therapy. Ultrasound is widely used to treat soft tissue shoulder pathology including subacromial impingement syndrome, tendinitis, tendinosis, bursitis, calcific deposits, and myofascial tears.5–7 A recent multi-database review of RTCs in using therapeutic ultrasound to treat shoulder pathology examined the treatment parameters of frequency, duration, pulse/continuous wave mode and total amount of acoustic energy delivered to subjects.8 Alexander and colleagues8 found that the success of therapeutic ultrasound on the treatment of a range of soft tissue shoulder injuries was highly dependent on duration of ultrasound exposure per treatment, the total amount of acoustic energy delivered to the shoulder and the prevalence of treatment application. Specifically, the authors found that RTCs of ultrasound treatments which provided over 4,228 Joules of acoustic energy per treatment session over multiple sessions were beneficial to patients suffering from calcific tendinitis, supraspinatus tendinitis, adhesive capsulitis and subacromial bursitis. Ultrasound treatments for short durations with 2,019 Joules or less energy provided no significant benefit. The authors concluded that ultrasound application for the treatment of shoulder pathology holds promise but must focus on selecting optimal ultrasound treatment parameters that deliver ultrasound energy of 4,228 Joules per session or greater, and treatment schedules that expose tissues to ultrasound for sufficient amounts of time (i.e. greater than 5 hours). This scenario and treatment régime is particularly well suited for wearable long duration low-intensity therapeutic ultrasound (LITUS) that the team has developed and preliminarily evaluated in this Institutional Review Board (IRB) approved pilot study.
METHODS
Wearable Low Intensity Therapeutic Ultrasound Device
Most ultrasound devices with LITUS capability require wall-power, are bulky, and the battery-powered units are shoebox-sized and only deliver short 20–30 minute continuous treatments. All of the devices currently on the market cost more than $1200.00. The LITUS device developed by the investigative team is based on ultralow impedance design that allows for streamlined circuit architecture, optimized electroacoustic signal conduction, and construction of low-profile wearable systems.9–13 The wearable LITUS device is a diverging-wave system that operates at 2.95 MHz with 0.03–2 W/cm2 ultrasound intensity capability for 0.3–18 hrs of treatment depending on the output setting. This corresponds to 10,800 Joules of total acoustic energy delivered to the patient per treatment session.
The electronics of the LITUS device are constructed on a printed circuit board and housed in a PVC plastic enclosure with lithium-ion battery pack. The transducer of the device is housed in a waterproof, biocompatible ring with a polyurethane rubber boot and a 10° ultrasound diverging lens made from polystyrene. The active transducer element is a lead-zirconate-titanate (PZT-8), silver-plated piezocrystal that is air backed. The housing and lens in the LITUS device protects the electronics and piezocrystal from deterioration.
In comparison to the Rich-Mar clinical therapeutic ultrasound system shown in Figure 1, the LITUS device is less than 1/20th of the size, does not require wall power, and may provide the same treatment as is customary in current physical therapy. However, in this study the device was calibrated at 0.65 Watts and a nominal 90 mW/cm2 to deliver ultrasound for 5.5 hrs continuously per charge.
For the purpose of the study, each wearable LITUS device was packaged individually in small boxes with ultrasound gel, Tegaderm™ bandages, system charger, belt holder and instructions for use. All systems were acoustically calibrated and measured pre and post field deployment with an ultrasonic power meter and beam scanning system from Onda Inc. and Velmex Inc, respectively.
Clinical Pilot Study
The Institutional Review Board of Cayuga Medical Center (CMC), Ithaca, NY approved this study and informed consent was obtained from all subjects. Five subjects were instructed to wear the LITUS device alternating between anterior and posterior placement for 3–4 hours per day for 12 consecutive treatment sessions, and record their daily pain score on the visual analog scale (1 to 10) and global rate of health improvement scale (−7 to 7). Figure 2 demonstrates the device being applied to the shoulder with Tegaderm™ bandage and ultrasound gel. Inclusion criteria for the study consisted of subjects who: (1) were seeing their physician for shoulder pain due to rotator cuff impingement, tear, tendinosis or tendinitis, adhesive capsulitis, biceps tenosynovitis, proximal humerus fracture, (2) were 40 years or older in age, (3) reported taking pain medication for their shoulder pain during the week preceding enrollment, (4) reported functional limitations due to their shoulder pain, and (5) were deemed appropriate by their physician to participate. Exclusion criteria consisted of subjects who: (1) had known metastatic disease or infectious etiology about the shoulder (2) had a cervical radiculopathy (3) could not successfully demonstrate the ability to put on and take off the device, or (4) were cognitively impaired.
RESULTS
The data for mean VAS pain scores over the 12 consecutive treatment sessions is shown in Figure 3 with standard error. Across the total number of treatments the average reduction in pain from treatment start to treatment conclusion was found to be 30%. Figure 3 also shows the results of the mean percent improvement of GROC score as it relates to patients starting pain on the VAS. Across the total number of treatments the average GROC score was +3.22 with normalized improvement to starting VAS of 52%. When compared to starting VAS and GROC, the changes where significant p<0.05 for both measures.
For all subjects involved in this pilot, the wearable LITUS device provided some type of pain reduction and no pain increases. All participants were able to self-apply the ultrasound system in the convenience of their own homes after a short training session by a physical therapist.
CONCLUSION
In this clinical pilot study, a miniature long duration low-intensity therapeutic ultrasound system was preliminarily evaluated on rotator cuff tendinopathy. Though the pilot study was not placebo controlled, the results indicate the device may be applied successfully in the outpatient setting and also provide supportive evidence for a more formal larger placebo controlled study. Furthermore, the results of this study agreed with Alexander and colleagues clinical practice suggestion that daily therapeutic ultrasound providing over 4,228 Joules of acoustic energy per treatment session was beneficial to subjects with shoulder tendinopathy.
The wearable ultrasound system shows promise as an innovative therapeutic ultrasound platform for self-applied effective therapy, and as a possible replacement or adjuvant to pharmacotherapies for shoulder pain.
REFERENCES
- 1.Olshansky SJ, Goldman DP, Zheng Y, Rowe JW. Aging in America in the twenty-first century: demographic forecasts from the MacArthur Foundation Research Network on an Aging Society. Milbank Q. 2009;87(4):842–862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lin JC, Weintraub N, Aragaki DR. Nonsurgical treatment for rotator cuff injury in the elderly. J Am Med Dir Assoc. 2008;9(9):626–632. [DOI] [PubMed] [Google Scholar]
- 3.Meislin RJ, Sperling JW, Stitik TP. Persistent shoulder pain: epidemiology, pathophysiology, and diagnosis. Am J. Orthop. 2005;34(12 Suppl):5–9. [PubMed] [Google Scholar]
- 4.Chard MD, Hazleman BL. Shoulder disorders in the elderly (a hospital study). Ann. Rheum. Dis. 1987;46(9):684–687. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Robertson VJ, Spurritt D. Electrophysical Agents: Implications of their Availability and Use in Undergraduate Clinical Placements. Physiotherapy. 1998;84(7):335–344. [Google Scholar]
- 6.Pope G, Mockett S, Wright J. A Survey of Electrotherapeutic Modalities: Ownership and Use in the NHS in England. Physiotherapy. 1995;81(2):82–91. [Google Scholar]
- 7.Wong RA, Schumann B, Townsend R, Phelps CA. A survey of therapeutic ultrasound use by physical therapists who are orthopaedic certified specialists. Phys Ther. 2007;87(8):986–994. [DOI] [PubMed] [Google Scholar]
- 8.Alexander LD, Gilman DRD, Brown DR, Brown JL, Houghton PE. Exposure to low amounts of ultrasound energy does not improve soft tissue shoulder pathology: a systematic review. Phys Ther. 2010;90(1):14–25. [DOI] [PubMed] [Google Scholar]
- 9.Lewis GK, Olbricht WL. Wave Generating Apparatus, UPCT Patent Application No. PCT/US2009/50297. 2009.
- 10.Lewis GK. A portable therapeutic ultrasound system and methods of use thereof, UPCT Patent Application No. PCT/US2009/291732. 2009.
- 11.Lewis GK. Ultrasound coupling device and methods of use thereof, UPCT Patent Application No. PCT/US2009/291779. 2009.
- 12.Lewis GK. Low-profile ultrasound transducer and methods of use thereof, UPCT Patent Application No. PCT/US2009/291804. 2009.
- 13.Lewis GK, Olbricht WL. Development of a portable therapeutic and high intensity ultrasound system for military, medical, and research use. Rev Sci Instrum. 2008;79(11):114302. [DOI] [PMC free article] [PubMed] [Google Scholar]