Effect of Repeated Doses of Acetaminophen on a Continuous Glucose Monitoring System with Permselective Membrane

Douglas Denham 1,
PMCID: PMC8256057  PMID: 32865043

Continuous glucose monitoring systems use an electrochemical methodology for measuring glucose in interstitial fluid (ISF). Dexcom G6 uses a measurement voltage where other common compounds may get oxidized. The diffusion membrane can block some of the interfering compounds; however, not all compounds can be blocked due to their molecular mass. Dexcom G4 and G5 systems showed significant interference from acetaminophen.1-3 Dexcom introduced a permselective membrane in G6 that blocks the diffusion of acetaminophen to the electrode surface thereby minimizing its effect.4

In the clinical study performed by Dexcom, the study participants were asked not to take any acetaminophen from the day before sensor insertion until the in-clinic session, where study subjects were asked to take 1 g of acetaminophen and were followed for 8 hours. The study’s objective was met in that the observed mean interference effect was 3.1 ± 4.8 mg/dL, which was statistically significantly lower than the performance goal of 10 mg/dL. The authors concluded that this interference effect would have a negligible effect on bolus insulin administered. The authors argued that because the temporal profile of interference with G4 sensors follows that of ISF and plasma acetaminophen concentrations,2 studies of higher or repeated doses were unnecessary.

Since the interference from acetaminophen is reduced by using a permselective membrane, its effectiveness should be demonstrated by challenging the membrane with repeat doses of acetaminophen. We evaluated the effect of repeated doses of acetaminophen on G6 sensor performance in a clinical study. Fifteen adult subjects with diabetes wore one sensor in the abdomen. Data from 14 adult subjects (1 lost the sensor before in-clinic session) were collected over one 13-hour period during 3-7 days of the sensor wear. Each subject had a 1-hour baseline phase where venous blood was collected every 10 minutes. After this first hour, a dose of 1000 mg acetaminophen was given with a meal and venous samples were collected every 20 minutes for the next 4 hours. This step of acetaminophen ingestion and venous sample collection was repeated two more times in 4-hour intervals. Plasma venous samples were measured on a YSI2300 system. The maximum average bias from YSI and baseline after each administration of acetaminophen are presented in Table 1. A maximum average sensor bias of 7.2 mg/dL (normalized to the baseline offset) was observed around 4 hours after the first dose (the magnitude of the interference is consistent with the +5.2 mg/dL reported by Dexcom in the Instructions for Use).5 After the second dose, the maximum average sensor bias was approximately 8.6 mg/dL, which increased to 14.0 mg/dL after the third dose, suggesting that the permselective membrane may fail to effectively block acetaminophen after repeated doses.

Table 1.

The Maximum Average Bias from YSI and Baseline at Each Ingestion of 1000 mg of Acetaminophen Administered 4 Hours Apart on Dexcom G6.

Dosing period Bias from YSI, mg/dL Bias from baseline, mg/dL
Pre-dose (baseline) 15.1 0
Post first dose 22.3 7.2
Post second dose 23.7 8.6
Post third dose 29.1 14.0

The magnitude of interference from acetaminophen on G6 will have minimal impact on insulin dosing. However, the biases are large enough for the sensors to potentially miss hypoglycemic conditions, specifically clinically significant hypoglycemia. This study did not follow the subjects over multiple days and, therefore, the effect of higher doses or continued use of acetaminophen over multiple days on sensor performance is not known.

Acknowledgments

We are deeply indebted to the research participants. Our sincere thanks to all the staff of Clinical Trials of Texas.

Footnotes

Declaration of Conflicting Interests: The author declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: Study was funded by Abbott Diabetes Care.

References

  • 1. Basu A, Slama MQ, Nicholson WT, et al. Continuous glucose monitor interference with commonly prescribed medications: a pilot study. J Diabetes Sci Technol. 2017;11(5):936-941. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Basu A, Veettil S, Dyer R, Peyser T, Basu R. Direct evidence of acetaminophen interference with subcutaneous glucose sensing in humans: a pilot study. Diabetes Technol Ther. 2016;18(suppl 2):S243-S247. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Maahs DM, DeSalvo D, Pyle L, et al. Effect of acetaminophen on CGM glucose in an outpatient setting. Diabetes Care. 2015;38(10):e158-e159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Calhoun P, Johnson TK, Hughes J, Price D, Balo AK. Resistance to acetaminophen interference in a novel continuous glucose monitoring system. J Diabetes Sci Technol 2018;12(2):393-396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Dexcom, Dexcom G6 Continuous Glucose Monitoring System User Guide, 2020. (LBL014003 Ver 11). San Diego, CA. [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES