Abstract
Background:
Continuous glucose monitoring (CGM) and continuous subcutaneous insulin infusion (CSII) are the standard of care for type 1 diabetes in children. There is little reported on device-related skin complications and treatment options. This study documents cutaneous reactions to CGM and CSII devices in children and young adults with type 1 diabetes.
Methods:
One hundred and twenty-one subjects (3-25 years) with type 1 diabetes and CGM and/or CSII use were recruited over a three-month period from the Naomi Berrie Diabetes Center at Columbia University Irving Medical Center. A five-question survey was completed for each subject detailing demographic data, diabetes management, and device-related skin complications.
Results:
Sixty percent of subjects reported skin complications related to CGM and/or CSII use. Terms most frequently used to describe cutaneous reactions were “red,” “itchy,” “painful,” and “rash.” Subjects who used both CGM and CSII were more likely to report skin problems than those who used only CSII (odds ratio 2.9, [95% confidence interval: 1.2-6.7]; P = .015). There were no associations between skin complications and sex or race/ethnicity. Twenty-two percent of subjects with adverse skin event(s) discontinued use of a device due to their skin problem. Seven percent were evaluated by a dermatologist. Eighty-one percent used a range of products to treat their symptoms, with variable perceived clinical outcomes.
Conclusions:
Skin complications related to CSII or CGM devices are commonly reported in pediatric patients with type 1 diabetes and may lead to interruption or discontinuation of device use. Future studies are needed to elucidate the causes of these reactions and determine the best methods for prevention.
Keywords: acrylates, continuous glucose monitoring, continuous subcutaneous insulin infusion, cutaneous reactions, insulin pumps
Introduction
Continuous glucose monitoring (CGM) and continuous subcutaneous insulin infusion (CSII) have become the standard of care for management of type 1 diabetes in pediatric patients. These medical devices are applied to the skin of patients and allow for integrated monitoring and regulation of glucose metabolism. Use of CSII and CGM devices—either separately or as hybrid closed-loop systems—results in streamlined diabetes management. Both CSII and CGM have demonstrated improved glycemic outcomes, as well as improvements in quality of life, compared to multiple daily insulin injections and fingerstick blood glucose monitoring, respectively.1-4 These devices have been widely adopted by patients in recent years as they have advanced in convenience and efficacy.
Although serious adverse events such as hypoglycemic seizures or hospitalization related to the use of CGM and CSII devices are rare, skin reactions to these devices have been reported, ranging from mild to severe.5,6 Cutaneous adverse events described include lipoatrophy, lipohypertrophy, scarring, local erythema, subcutaneous infection, exacerbation of pre-existing inflammatory skin conditions, and irritant or allergic contact dermatitis (ACD).7-9 Recently, there have been increasing case reports of ACD related to components of both CGM and CSII devices and/or device adhesives, particularly acrylates.10-12 Skin reactions can have a significant impact on quality of life, glycemic control, and treatment adherence.7 Cutaneous adverse events have been reported as a reason for device discontinuation.13-16
The incidence and prevalence of device-related skin reactions among patients using CGM and CSII devices are not known. A small observational study reported a frequency of adverse dermatological effects of CSII use in pediatric patients as high as 43%.9 A systematic review of cutaneous complications with interstitial glucose monitoring found a low incidence of one cutaneous event per eight weeks of device use based on 19 trials.16 However, observational studies suggest a high prevalence among CGM users.14,16,17 There are a number of reasons that trials may underestimate the rate of cutaneous adverse events. A history of skin reactions is often part of the exclusion criteria when investigating the safety profile of these devices. Sensitization to an allergen can take months to years to develop, and therefore a cutaneous hypersensitivity reaction may not be captured in trials with short-term follow-up.13
We propose that skin complications related to the use of glucose regulating and monitoring devices represent an underreported barrier to daily use of these devices in a subset of patients, leading to interrupted or discontinued use in some cases. Here we present the results of a questionnaire designed to capture skin reactions to CGM and CSII devices as well as their treatments, including perceived efficacy, in a cohort of pediatric patients at a comprehensive diabetes care center.
Methods
Subjects with type 1 diabetes and present or past use of CGM and/or CSII device(s) were recruited from the Naomi Berrie Diabetes Center at Columbia University Irving Medical Center when they presented for scheduled clinical appointments over a three-month period in Spring 2017. Consent/assent was obtained from subjects and/or their caregiver. The study design, survey, and consent/assent forms were approved by the Institutional Review Board of Columbia University Irving Medical Center.
Each subject and/or caregiver completed one qualitative five-question standardized survey. Subjects were asked to report demographic information, including age, sex, race, and ethnicity. They were then asked to indicate each device they had ever used, including CSII infusion set types and CGM type, and to report whether they had experienced any skin problems related to their devices. Additional questions involved further characterization of reactions, if present, and their treatment, along with perceived efficacy. Subjects and families had the option to free-text any details regarding clarification of reactions and results of treatment.
Qualitative analysis of free-text responses characterizing skin reactions was performed in an iterative manner. Free-text responses were reviewed for each subject, and descriptive words and phrases were categorized by clinical relevance that was agreed upon by the study authors, including both clinical endocrinologists and dermatologists. For example, if subjects listed antibiotic use or drainage, the reaction would be classified as an “infection.” Statistical analysis was performed using Stata 14 (2015). For categorical variables, we compared proportions with skin complications, using the Pearson chi-squared test or Fisher’s exact test. To examine the association between the outcome of reporting a skin problem and using CGM and CSII vs CSII alone, adjusting for other predictors, we fit multiple logistic regression models.
Results
One hundred and twenty-one subjects aged 3-25 years (mean age = 13.9 ± 4.8 years) responded to the survey (Table 1). Sixty percent (72/121) of subjects reported skin problems related to the use of a device. There were no significant associations found between skin complications and sex or race/ethnicity.
Table 1.
Sex | Patients (%) |
---|---|
Female | 47% (57/121) |
Male | 53% (64/121) |
Ethnicity/race | |
Non-Hispanic white | 65% (79/121) |
Non-Hispanic black | 5% (6/121) |
Hispanic | 21% (25/121) |
Other | 9% (11/121) |
Among the clinically relevant terms used by subjects to describe skin problems, the most common were “red,” “itchy,” and “painful” (Table 2). Of note, 32% of subjects who reported skin complications (n = 72) specifically reported an adverse event related to a CGM device, whereas 18% attributed a reaction to an adhesive used to attach a device or sensor to the skin.
Table 2.
Reaction | Percent of patients reporting reaction (n = 72) |
---|---|
Red | 35% (25/72) |
Itchy | 31% (22/72) |
Painful | 22% (16/72) |
Rash | 22% (16/72) |
Skin change | 13% (9/72) |
Infection | 11% (8/72) |
Psoriasis exacerbation | 3% (2/72) |
Eczema exacerbation | 1% (1/72) |
Thirty-six subjects reported using CSII devices only, 83 reported using both CSII and CGM devices, and 2 reported using CGM devices only. Among subjects using both CGM and CSII devices, 69% reported skin problems, compared to 39% of subjects using CSII devices alone (P = .002). Adjusting for age, those who used both devices had 2.9 times the odds of reporting a skin problem (P = .015, 95% confidence interval [1.2, 6.7]), compared to those using only one device.
Eighty-one percent of subjects with skin problems used a range of both prescribed and over the counter medications and dressings to treat their symptoms. The most commonly used products reported were adhesive barrier wipes, transparent film dressings, and topical steroids. Additional products include adhesive bandages, topical antibiotics, adhesive removers, adhesive glue, and hydrocolloid adhesive pads. Common brands and specific types of products were also listed out for ease of identification by the subject. Within the survey, the subjects had the option of identifying if a product resulted in improvement, worsening, or no change of a skin reaction (Table 3). For those that used topical steroids, hydrocolloid adhesive pads, or topical antibiotics, there was an overall consensus of improvement with use (77%, 69%, and 82%, respectively). Within the product categories, there were further varied responses. For example, for subjects that used multiple brands of adhesive barrier wipes, they may have noted improvement with one brand and either no change or worsening with a second.
Table 3.
Product category | Patients using products in category (n)a | Patients reporting clinical improvement (%) | Patients reporting no change (%) | Patients reporting clinical worsening (%) | Patients reporting discordant effects (%)b |
---|---|---|---|---|---|
Adhesive barrier wipe | 27 | 26% (7/27) | 48% (13/27) | 11% (3/27) | 15% (4/27) |
Transparent film dressing | 27 | 33% (9/27) | 48% (13/27) | 11% (3/27) | 7% (2/27) |
Topical steroid | 22 | 77% (17/22) | 18% (4/22) | 0% (0/22) | 5% (1/22) |
Adhesive bandage | 13 | 15% (2/13) | 54% (7/13) | 23% (3/13) | 8% (1/13) |
Hydrocolloid adhesive pad | 13 | 69% (9/13) | 15% (2/13) | 15% (2/13) | 0% (0/13) |
Topical antibiotic | 11 | 82% (9/11) | 18% (2/11) | 0% (0/11) | 0% (0/11) |
Adhesive remover | 10 | 50% (5/10) | 40% (4/10) | 10% (1/10) | 0% (0/10) |
Adhesive glue | 2 | 50% (1/2) | 0% (0/2) | 50% (1/2) | 0% (0/2) |
Among subjects with skin problems, 81% (58/72) reported using products to treat their symptoms. Many subjects reported using multiple products, and thus totals should not add up to 58 and percentages should not sum to 100%.
“Patients reporting discordant effects” refers to patients who used multiple supplemental products in a category with different perceived clinical effects.
Twenty-two percent of subjects who reported a skin problem discontinued the use of the device due to the adverse skin reactions listed in Table 2. Among all subjects who reported skin problems, 7% had been evaluated by a dermatologist.
Discussion
This study suggests that cutaneous reactions are a common adverse event for pediatric patients using CSII and CGM devices. The majority of subjects reported skin problems that they attributed to a reaction to such devices. These adverse skin reactions were variable in how they were defined and in their perceived response to treatments. A small percentage of subjects stated that the use of a device caused a flare of pre-existing eczema or psoriasis. Tolerability of devices for many patients may be dependent on skin sensitivity and the ability to find effective solutions for dermatological problems that arise.18
Given the relative safety of CGM and CSII devices, skin reactions may represent a significant barrier to uninterrupted use for a subset of pediatric patients. Younger patients often have less body surface area for device attachment, meaning that if a skin reaction occurs, there are fewer alternative sites for application. In addition, high levels of physical activity in young patients may compromise the ability of adhesives to stick to patients’ skin, requiring supplemental adhesives,13 which increases exposure to potentially irritating or sensitizing products. We did not assess for lack of adhesion with diabetes devices. However, many patients reported the use of extra adhesives to ensure that their devices remained in place.
Adhesives used to bind components of devices or attach them to skin are potential causes of contact dermatitis, both irritant and allergic. Eighteen percent of subjects who reported adverse events describe a rash that they attributed to a CGM sensor sticker or other adhesive. Irritant contact dermatitis (ICD) is a direct toxic effect at the site of exposure and could be caused by sweat and friction under an adhesive pad or other material, whereas ACD is immune-mediated response and requires prior sensitization. Patients may experience symptoms such as redness, itching, burning, and blistering. One study evaluating ACD in insulin pump users noted that earlier use of another wearable diabetes device may have contributed to new skin reactions through prior sensitization.19 As subjects using CGM and CSII together conferred a higher risk/odds ratio of reported skin problems compared to CSII alone in this study, it is possible that using multiple devices over time allows for increased opportunity to become sensitized to allergens.
Isobornyl acrylate (IBOA), found in plastic material used for the housing of some CGM devices and insulin pumps, has recently been identified as a major allergen causing ACD in patients using medical devices for diabetes therapy.19-25 Other allergens implicated include ethyl cyanoacrylate,26 N,N-dimethylacrylamide,27 and colophonium.11 Contact dermatitis, including both allergic and irritant, from repeated exposure to adhesives has been well described in the context of other medical devices or surgical materials, such as peristomal adhesive products28,29 and surgical glue.30-33 Further studies are needed to better characterize the prevalence of ACD related to medical devices for diabetes therapy. Patch testing can help confirm if a skin reaction is due to ACD or ICD as determined by the appearance, size, and duration of the resulting inflammation. Patients who present with ACD related to an insulin pump or CGM device may require patch testing with an expanded series of acrylates. IBOA is not included in the Food and Drug Administration standard patch testing series, and historically has not been included in common acrylate series used in screening by patch testing clinics.34
If a patient develops an adverse event such as ACD, therapeutic options include avoidance of the allergen by switching to an alternative device, or placement of a barrier to prevent direct contact with the skin.35,36 To improve tolerability of these devices, various supplemental products have been used by patients to improve skin adhesion and prevent rashes and other cutaneous reactions. In the absence of clear best practice recommendations, many patients use social media forums to seek solutions to skin-related reactions to their devices. Here we outlined commonly attempted product types used by patients, to inform clinicians.
The majority of subjects who reported adverse skin reactions used one or more products to attempt to improve their symptoms, with variable perceived clinical effects (Table 3). Although successful use of hydrocolloid adhesive pads in the setting of ACD related to acrylates has been reported,35,36 it is worth noting that using additional products has the potential to exacerbate an ACD or ICD.
Twenty-two percent of subjects who reported skin problems attributed device discontinuation to a cutaneous adverse event. This suggests that cutaneous reactions to CSII and/or CGM devices may prevent a subset of patients from continuing a beneficial treatment modality. Patients with cutaneous reactions may present first to the endocrinologist or pediatrician, as evidenced by the fact that only 7% of subjects with skin reactions reported being evaluated by a dermatologist. We recommend asking patients about potential skin reactions while they are using these devices and early referral to dermatology when indicated. With further characterization of these reactions and assessment of prevalence among pediatric patients with type 1 diabetes, we hope to develop a more uniform screening protocol. As the field of technology in diabetes management continues to advance, a standardized approach will be necessary to define, treat, and possibly prevent cutaneous reactions to ensure continued benefit from these devices. We also recommend at this time that patients with risk factors for ACD who are manifesting an adverse skin reaction without improvement to treatment be considered for patch testing to known components of the device and adhesives.
Limitations
The small sample size in this observational study precludes drawing conclusions regarding the prevalence of cutaneous reactions to CSII and CGM devices in a larger pediatric population with type 1 diabetes. The high rate of adverse cutaneous reactions reported may be in part due to response bias, as patients may have been more inclined to complete the study if they had a history of reactions. Subjects were not directly asked which devices or device components were associated with specific skin reactions. Subjects were asked to report historic and current cutaneous reactions and were not directly examined, precluding standard criteria documenting adverse events. Furthermore, information was not collected regarding prior dermatologic history, including prior cutaneous reactions to adhesives or history of atopy.
Conclusion
We report these findings to increase awareness of the potential impact of cutaneous adverse events related to CGM and CSII devices in type 1 diabetes patients. There may be a significant subset of pediatric and young adult patients who discontinue use of their device as a result of these reactions, even as the technology continues to become an important and advantageous treatment of type 1 diabetes. Future studies that survey a larger population, as well as patch testing of patients with suspected ACD, are recommended.
Acknowledgments
The authors thank Jimmy Duong of the Irving Institute for Clinical and Translational Research at Columbia University Irving Medical Center for his contributions to the development of the analytic plan and to data analysis.
Footnotes
Previous Publication as Abstract: This work was previously presented as in the General Poster Session at the ADA 78th Scientific Sessions and published in abstract form: Severin RK, Gandica R, Levin LE, Belsito DV, Garzon MC, and Williams KM. Cutaneous Reactions to Continuous Glucose Monitoring and Continuous Subcutaneous Insulin Infusion Devices in Type I Diabetes. Diabetes 2018 Jul; 67(Supplement 1).
Author Contributions: KMW is responsible for design, oversight, execution, interpretation of data, and management of collaborations for this study. MCG, RG, and KMW are responsible for conception of this study. RSR, RG, and KMW performed data collection and analysis. RSR and KMW wrote the manuscript and prepared displays with input from RG, LEL, DVB, and MCG.
Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.
ORCID iD: Kristen M. Williams https://orcid.org/0000-0002-8464-4537
References
- 1. Misso ML, Egberts KJ, Page M, O’Connor D, Shaw J. Continuous subcutaneous insulin infusion (CSII) versus multiple insulin injections for type 1 diabetes mellitus. Cochrane Database Syst Rev. 2010;(1):Cd005103. doi: 10.1002/14651858.CD005103.pub2. [DOI] [PubMed] [Google Scholar]
- 2. Rodbard D. Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes. Diabetes Technol Ther. 2017;19(S3):S25-S37. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Haviland N, Walsh J, Roberts R, Bailey TS. Update on clinical utility of continuous glucose monitoring in type 1 diabetes. Curr Diab Rep. 2016;16(11):115. [DOI] [PubMed] [Google Scholar]
- 4. Polonsky WH, Hessler D, Ruedy KJ, Beck RW, DIAMOND Study Group. The impact of continuous glucose monitoring on markers of quality of life in adults with type 1 diabetes: further findings from the DIAMOND randomized clinical trial. Diabetes Care. 2017;40(6):736-741. [DOI] [PubMed] [Google Scholar]
- 5. Englert K, Ruedy K, Coffey J, et al. Skin and adhesive issues with continuous glucose monitors: a sticky situation. J Diabetes Sci Technol. 2014;8(4):745-751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Pleus S, Ulbrich S, Zschornack E, Kamann S, Haug C, Freckmann G. Documentation of skin-related issues associated with continuous glucose monitoring use in the scientific literature. Diabetes Technol Ther. 2019;21(10):538-545. [DOI] [PubMed] [Google Scholar]
- 7. Jedlowski PM, Te CH, Segal RJ, Fazel MT. Cutaneous adverse effects of diabetes mellitus medications and medical devices: a review. Am J Clin Dermatol. 2019;20(1):97-114. [DOI] [PubMed] [Google Scholar]
- 8. Conwell LS, Pope E, Artiles AM, Mohanta A, Daneman A, Daneman D. Dermatological complications of continuous subcutaneous insulin infusion in children and adolescents. J Pediatr. 2008;152(5):622-628. [DOI] [PubMed] [Google Scholar]
- 9. Binder E, Lange O, Edlinger M, et al. Frequency of dermatological side effects of continuous subcutaneous insulin infusion in children and adolescents with type 1 diabetes. Exp Clin Endocrinol Diabetes. 2015;123(4):260-264. [DOI] [PubMed] [Google Scholar]
- 10. Herman A, de Montjoye L, Tromme I, Goossens A, Baeck M. Allergic contact dermatitis caused by medical devices for diabetes patients: a review. Contact Dermatitis. 2018;79(6):331-335. [DOI] [PubMed] [Google Scholar]
- 11. Hyry HSI, Liippo JP, Virtanen HM. Allergic contact dermatitis caused by glucose sensors in type 1 diabetes patients. Contact Dermatitis. 2019;81(3):161-166. [DOI] [PubMed] [Google Scholar]
- 12. Kamann S, Aerts O, Heinemann L. Further evidence of severe allergic contact dermatitis from isobornyl acrylate while using a continuous glucose monitoring system. J Diabetes Sci Technol. 2018;12(3):630-633. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Heinemann L, Kamann S. Adhesives used for diabetes medical devices: a neglected risk with serious consequences? J Diabetes Sci Technol. 2016;10(6):1211-1215. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Berg AK, Norgaard K, Thyssen JP, et al. Skin problems associated with insulin pumps and sensors in adults with type 1 diabetes: a cross-sectional study. Diabetes Technol Ther. 2018;20(7):475-482. [DOI] [PubMed] [Google Scholar]
- 15. Wong JC, Foster NC, Maahs DM, et al. Real-time continuous glucose monitoring among participants in the T1D exchange clinic registry. Diabetes Care. 2014;37(10):2702-2709. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Asarani NAM, Reynolds AN, Boucher SE, de Bock M, Wheeler BJ. Cutaneous complications with continuous or flash glucose monitoring use: systematic review of trials and observational studies. J Diabetes Sci Technol. 2019;14(2): 328-337. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Berg AK, Olsen BS, Thyssen JP, et al. High frequencies of dermatological complications in children using insulin pumps or sensors. Pediatr Diabetes. 2018;19(4):733-740. [DOI] [PubMed] [Google Scholar]
- 18. Wadwa RP, Fiallo-Scharer R, Vanderwel B, Messer LH, Cobry E, Chase HP. Continuous glucose monitoring in youth with type 1 diabetes. Diabetes Technol Ther. 2009;11(suppl 1):S83-S91. [DOI] [PubMed] [Google Scholar]
- 19. Raison-Peyron N, Mowitz M, Bonardel N, Aerts O, Bruze M. Allergic contact dermatitis caused by isobornyl acrylate in OmniPod, an innovative tubeless insulin pump. Contact Dermatitis. 2018;79(2):76-80. [DOI] [PubMed] [Google Scholar]
- 20. Corazza M, Scuderi V, Musmeci D, Foti C, Romita P, Borghi A. Allergic contact dermatitis caused by isobornyl acrylate in a young diabetic patient using a continous glucose monitoring system (freestyle libre). Contact Dermatitis. 2018;79(5):320-321. [DOI] [PubMed] [Google Scholar]
- 21. Herman A, Aerts O, Baeck M, et al. Allergic contact dermatitis caused by isobornyl acrylate in freestyle® libre, a newly introduced glucose sensor. Contact Dermatitis. 2017;77(6):367-373. [DOI] [PubMed] [Google Scholar]
- 22. Herman A, Baeck M, de Montjoye L, et al. Allergic contact dermatitis caused by isobornyl acrylate in the enlite glucose sensor and the paradigm minimed quick-set insulin infusion set. Contact Dermatitis. 2019;81(6):432-437. [DOI] [PubMed] [Google Scholar]
- 23. Mine Y, Urakami T, Matsuura D. Allergic contact dermatitis caused by isobornyl acrylate when using the freeStyle® libre. J Diabetes Investig. 2019;10(5):1382-1384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Oppel E, Hogg C, Summer B, Rueff F, Reichl FX, Kamann S. Isobornyl acrylate contained in the insulin patch pump OmniPod as the cause of severe allergic contact dermatitis. Contact Dermatitis. 2018;79(3):178-180. [DOI] [PubMed] [Google Scholar]
- 25. Oppel E, Kamann S, Reichl FX, Hogg C. The dexcom glucose monitoring system-an isobornyl acrylate-free alternative for diabetic patients. Contact Dermatitis. 2019;81(1):32-36. [DOI] [PubMed] [Google Scholar]
- 26. Peeters C, Herman A, Goossens A, Bruze M, Mowitz M, Baeck M. Allergic contact dermatitis caused by 2-ethyl cyanoacrylate contained in glucose sensor sets in two diabetic adults. Contact Dermatitis. 2017;77(6):426-429. [DOI] [PubMed] [Google Scholar]
- 27. Mowitz M, Herman A, Baeck M, et al. N,N-dimethylacrylamide-a new sensitizer in the freestyle libre glucose sensor. Contact Dermatitis. 2019;81(1):27-31. [DOI] [PubMed] [Google Scholar]
- 28. Jacobs M, Gilissen L, Vandevenne AS, Goossens A. Peristomal allergic contact dermatitis caused by stoma adhesive paste containing N-butyl monoester of polymethyl vinyl ether maleic acid. Contact Dermatitis. 2015;73(5):314-316. [DOI] [PubMed] [Google Scholar]
- 29. Cressey BD, Belum VR, Scheinman P, et al. Stoma care products represent a common and previously underreported source of peristomal contact dermatitis. Contact Dermatitis. 2017;76(1):27-33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Howard BK, Downey SE. Contact dermatitis from dermabond. Plast Reconstr Surg. 2010;125(6):252e-253e. [DOI] [PubMed] [Google Scholar]
- 31. Caton AM, Dauphine C. Allergic contact dermatitis after repeated exposure to dermabond. Am Surg. 2014;80(5):520-522. [DOI] [PubMed] [Google Scholar]
- 32. Lefevre S, Valois A, Truchetet F. Allergic contact dermatitis caused by dermabond®. Contact Dermatitis. 2016;75(4):240-241. [DOI] [PubMed] [Google Scholar]
- 33. Davis MD, Stuart MJ. Severe allergic contact dermatitis to dermabond prineo, a topical skin adhesive of 2-octyl cyanoacrylate increasingly used in surgeries to close wounds. Dermatitis. 2016;27(2):75-76. [DOI] [PubMed] [Google Scholar]
- 34. Christoffers WA, Coenraads PJ, Schuttelaar ML. Two decades of occupational (meth)acrylate patch test results and focus on isobornyl acrylate. Contact Dermatitis. 2013;69(2):86-92. [DOI] [PubMed] [Google Scholar]
- 35. Kamann S, Heinemann L, Oppel E. Usage of hydrocolloid-based plasters in patients who have developed allergic contact dermatitis to isobornyl acrylate while using continuous glucose monitoring systems [published online ahead of print September 20, 2019]. J Diabetes Sci Technol. doi: 10.1109/TBME.2017.2652062. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Kamann S, Oppel E. Hydrocolloid blister plaster decreases allergic contact dermatitis caused by freestyle libre and isobornyl acrylate. Contact Dermatitis. 2019;81(5):380-381. [DOI] [PubMed] [Google Scholar]