Skip to main content
Photoacoustics logoLink to Photoacoustics
. 2021 Jun 25;23:100282. doi: 10.1016/j.pacs.2021.100282

High-resolution functional photoacoustic monitoring of vascular dynamics in human fingers

Joongho Ahn 1, Jin Young Kim 1, Wonseok Choi 1, Chulhong Kim 1,*
PMCID: PMC8259315  PMID: 34258222

Abstract

Functional imaging of microvascular dynamics in extremities delivers intuitive information for early detection, diagnosis, and prognosis of vascular diseases. High-resolution and high-speed photoacoustic microscopy (PAM) visualizes and measures multiparametric information of microvessel networks in vivo such as morphology, flow, oxygen saturation, and metabolic rate. Here, we demonstrate high-resolution photoacoustic monitoring of vascular dynamics in human fingers. We photoacoustically monitored the position displacement of blood vessels associated with arterial pulsation in human fingers. Then, during and after arterial occlusion, we photoacoustically quantified oxygen consumption and blood perfusion in the fingertips. The results demonstrate that high-resolution functional PAM could be a vital tool in peripheral vascular examination for measuring heart rate, oxygen consumption, and/or blood perfusion.

Keywords: Photoacoustic microscopy, Blood vessel, Hemodynamics, Pulsation, Perfusion, Oxygen saturation, Digital vasculature

1. Introduction

An intricate and extensive network of vessels channels blood throughout the entire human body [1]. The blood circulating in the vessels transports oxygen and nutrients to cells and organs [2], so the vessels’ physical condition and circulatory ability are closely related to bodily dysfunctions. Insufficient delivery of oxygen or nutrients negatively affects tissues, impairing physiological functions [3]. Notably, the microvessels in the extremities, such as the fingers, and toes, are connected to the heart through macrovessels, and abnormalities of these macrovessels or of the heart itself can cause serious abnormalities of the microvessels [4]. When blood is not properly supplied to an extremity, it is exposed to vascular diseases and, in severe cases, may ultimately require amputation [5].

Different imaging techniques have been used to assess the structure and functioning of blood vessels. Both computed tomography (CT) angiography and magnetic resonance angiography provide wide-field images of primary arteries with high contrast [6,7]. However, these methods require ionizing radiation and/or contrast agents that are potentially harmful to some organs. Ultrasound (US) imaging, an agent-free and real-time imaging modality, provides anatomical information about the body, showing flow in blood vessels by using the Doppler effect [8]. However, all the above methods have low imaging resolution and so cannot diagnose impaired microcirculation [9]. Nailfold videocapillaroscopy (NVC) can image the capillaries with high resolution and observe blood flow at the microvessel level, but it can be applied only to the capillary loop of cuticles [10,11]. In addition, NVC cannot provide the oxygen saturation (sO2) of the microvessel, a key diagnostic indicator. Optical coherence tomography angiography (OCTA) can visualize the blood vessels by capturing more signal changes from the vessels over time than from static tissues [12]. OCTA can also estimate the flow and sO2 of the microvessels by using the Doppler effect and visible light spectroscopy [13,14]. However, although sO2 measurement using OCTA has been reported in human eye vessels [15], it has not yet been measured through the skin, due to the modality’s limited penetration depth [12]. Even when light passes through blood vessels and then bounces back from the tissues, OCTA registers the signal changes created by the blood vessels, which creates “false” blood vessels called OCTA projection artifacts that can affect clinical assessment [16].

As a superior alternative, photoacoustic (PA) imaging can directly visualize blood vessels because the hemoglobin in the blood is an endogenous absorber that generates PA waves [17]. Because oxyhemoglobin (HbO2) and deoxyhemoglobin (Hb) exhibit different absorption coefficients at different optical wavelengths, a spectroscopic technique can be applied to PA imaging to estimate sO2 [18]. In addition, because sound travels more slowly than light in tissue, depth-resolved 2D images can be created, stacks of which can be reconstructed to form 3D rendered images [19]. Exploiting the merits of the PA vascular imaging, previous studies have investigated structural and functional features such as vasomotion, sO2, and blood flow [[20], [21], [22]]. In addition, the recent development of high-repetition-rate pulsed lasers and high-speed scanners has significantly improved the imaging speed of photoacoustic microscopy (PAM) [[23], [24], [25], [26], [27], [28]]. Thanks to these advances, multi-wavelength and high-speed PAM has been developed and utilized in preclinical and clinical studies of blood vessels [[29], [30], [31], [32], [33], [34], [35], [36], [37], [38], [39]].

In this study, we use a dual-wavelength high-speed PAM system to capture the structural and functional dynamics of the micro-vessels in human fingers. We continuously imaged the 2D cross-sectional plane of the fingertips and found the position displacement of the blood vessels associated with arterial pulsation. Further, we monitored oxygen consumption during brachial cuffing and then observed blood perfusion after releasing the cuff. The results show that high-speed, high-resolution PAM can be invaluable in evaluating the vascular dynamics in human extremities.

2. Materials and methods

2.1. High resolution, high-speed, spectroscopic photoacoustic microscopy

Fig. 1a shows the configuration of the high-resolution, high-speed, dual-wavelength PAM system, modified from a commercial high-speed PAM (OptichoM, Opticho, Republic of Korea). The system is equipped with two nanosecond pulsed lasers (AWAVE532-1W-10 K and MPA559, Advanced Optowave, NY, USA), which have 532 nm and 559 nm wavelengths, respectively. From the isosbestic point of 545 nm at which the absorption coefficients of oxyhemoglobin (HbO2) and deoxyhemoglobin (Hb) are equal, HbO2 absorbs 532 nm light more than Hb, and Hb absorbs 559 nm more than HbO2. Using the two wavelengths, the PAM can measure oxygen saturation using a spectral unmixing method [22]. The two laser beams are combined by a dichroic mirror (DMSP550R, Thorlabs, NJ, USA), but first the 532-nm beam is attenuated by a continuously variable neutral density (ND) filter wheel (NDC-50C-4 M, Thorlabs, NJ, USA) to balance the optical power of the two beams. The combined beam is coupled to a single mode fiber (P1-460B-FC-1, Thorlabs, NJ, USA) using a fiber optic coupler (Thorlabs, NJ, USA) to deliver a single-mode beam to the OR-PAM system. The beam from the fiber is collimated and focused by a reflective collimator and an objective lens (RC08FC-P01 and AC127-050-A, Thorlabs Inc., NJ, USA). The beam is focused on the target by a water-immersible micro-electro-mechanical system (MEMS) scanning module (OptichoM-MS, Opticho, Republic of Korea), which includes an opto-ultrasound beam combiner and a 1-axis MEMS scanning mirror (Fig. 1b). The scanning mirror operates from 30 Hz to 80 Hz. The PA waves generated on the target are returned to the beam combiner and acquired by a 50-MHz US transducer (V214-BC-RM, Olympus NDT, MA, USA) attached on the beam combiner. The acquired signals are amplified by a low-noise, high-gain amplifier (PE15A1013, Pasternack Enterprises, CA, USA) with a 50 dB gain. The amplified signals are digitized and transferred to a personal computer via a high-speed digitizer (ATS9350, Alazar Technologies, QC, Canada) with a 500 MS/s sampling rate. Additionally, two motorized linear stages (L-509, Physik Instrumente, Germany) are employed to expand the field of view (FOV). A multi-functional data acquisition (DAQ) board (PCIe-6321, National Instruments, TX, USA) used a fully synchronized timing sequence (Fig. 1c). The DAQ board generates a reference signal with the speed of the scanner. The trigger signals for the two lasers, digitizer, and scanner are initiated based on the timing of the positive edge of the reference signal. The 559 nm laser is started 4 μs late to prevent the PA signals from the two lasers from overlapping. The triggers of the lasers and digitizer for data acquisition are enabled only when the scanner scans unidirectionally. The data acquisition and storage were implemented and executed in LabVIEW (National Instruments, TX, USA), and post processing and data analysis were carried out in MATLAB (MathWorks, MA, USA).

Fig. 1.

Fig. 1

(a) Configuration of the high-speed optical-resolution photoacoustic microscopy (OR-PAM) system. (b) Optical paths of the two lasers as the MEMS scanner moves. (c) Fully synchronized timing diagram. MEMS, microelectromechanical system; FM, flat mirror; DM, dichroic mirror; FC, fiber collimator; RC, reflective collimator; OL, objective lens; UT, ultrasound transducer; CRL, correction lens; OUC, opto-ultrasound beam combiner; ACL, acoustic lens.

The lateral and axial resolutions of the OR-PAM system are 5 and 30 μm, respectively [36]. This lateral resolution is sufficient to image single blood vessels in cuticles and fingers without blurring. The axial size of the blood vessels may appear slightly thicker due to the relatively poor axial resolution. The PA B-mode imaging speed of 50 Hz is sufficient catch the heart rate, the dominant frequency of the circulatory system. The step sizes along fast and slow scanning axes were 5 μm and 10 μm. The pulse repetition frequency was 40 kHz and 10 kHz for single- and multi-wavelength imaging, respectively. The signal-to-ratio (SNR), the maximum PA value of the blood vessels over the standard deviation of the background noise, was calculated to be 28 dB in human in vivo experiments described later.

2.2. In vivo PA imaging of human fingers with brachial cuffs

For human experiments, all experimental procedures were approved by the Institutional Review Board (IRB) of Pohang University of Science and Technology. We recruited a healthy volunteer as the subject of the in vivo experiments. Before the experiments, we completely explained the experimental procedure and received his informed consent. The operator and subject wore laser safety glasses and flame-retardant clothes for protection from the laser beam. For the experiments that required brachial pressure, the subject was first fitted with brachial cuff (REF CUF-F-A, A&D Company, Japan). Then, the subject’s right hand was placed on a customized finger holder fixed on a motorized vertical transition stage (8MVT188-20, Standa, Lithuania). After the little finger was covered with ultrasound gel, the vertical stage was moved to contact the finger against the thin membrane under the water tank. The water tank was then filled with enough water to submerge the MEMS scanning module. After the above preparation was completed, the subject took comfortable position, then the PA imaging experiments were performed. If brachial pressure was required, air was injected into the cuff using a digital blood pressure monitor (UA-651BLE, A&D Company, Japan). A pressure of over 150 mmHg was applied to ensure arterial occlusion. The measured laser pulse energies for the 532 and 559 nm beams were 530 nJ and 630 nJ, respectively, which correspond to 14.2 mJ/cm2 and 16.6 mJ/cm2, respectively, on the skin. Because the pulse widths of the beams are 10 ns and 18 ns, the conversion efficiencies for the beams from the pulse energy to the PA signal are different. To get the similar PA signal level from the two beams, we fixed the optical power of the 532-nm beam and adjusted the optical power of 559-nm beam by rotating the continuously variable ND filter wheel, while monitoring the PA signals on the blood vessels. The resulting optical exposures were below the maximum permissible exposure (MPE) level for skin, 20 mJ/cm2 set by the American National Standards Institute (ANSI) Z136.1 standard. We did not observe any laser-induced burns on the skin during or after the experiments.

3. Results and discussion

3.1. Photoacoustic monitoring of arterial pulsation

We continuously imaged a cross-sectional plane of the subject’s finger, at 50 Hz. Fig. 2a and b show PA B-mode images of the blood vessels during subsequent systole and diastole conditions, respectively. Supplementary Video S1 shows real-time PA monitoring of the systole and diastole dynamics. The vascular movement was quantified with following steps: (1) a region of interest (ROI) that includes the single blood vessel was selected in the first image. (2) A pixel that has the highest photoacoustic signal in the ROI was found. (3) The above steps were repeated on the consecutive images. (4) The pixels’ axial positions were accumulated. The axial position displacement over time shows heartbeat and motion (Fig. 2c). The blood vessels quickly oscillate with approximately 24 μm (red dotted lines in Fig. 2c) and slowly moves (black dotted line) by the heartbeat and motion, respectively. This movement by the heartbeat could be observed, because our PAM system has about 3-um axial step size from 500 MS/s sampling frequency, assuming that the speed of sound in water is 1500 m/s. The heartbeat exhibits a periodic movement with a frequency of 1.21 Hz (Fig. 2d), equivalent to 72.6 pulses per minute, and the motion has lower frequency components than those of the heartbeat. The periodic movement matches well with the subject’s heart rate as measured by a sensor on a smartphone (Galaxy Note9, Samsung Electronics, Republic of Korea). Thus, it can be concluded that the axial position displacement is considered as surrounding blood vessels' deformation caused by arterial pulsation located below the imaging area. Further, this consideration agrees well with the previously reported result using an ultrasound-array-based PAM and its validation using a commercial US machine [40].

Fig. 2.

Fig. 2

Photoacoustic monitoring of arterial pulsation based on the axial position displacement of the blood vessels in a finger. Photoacoustic B-mode of the finger during systole (a) and diastole (b). (c) Axial position displacement of the blood vessels over time. (d) Frequency response of the axial position displacement. The dominant frequency, 1.21 Hz, represents a pulsation at 72.6 beats per minute. All scale bars are 200 μm.

3.2. Photoacoustic monitoring of blood perfusion after releasing a brachial cuff

To investigate dynamic changes, we continuously monitored the finger blood vessels and blood perfusion while releasing the brachial cuff. Fig. 3a and b are PA maximum amplitude projection (MAP) images of the finger during and after brachial cuffing, respectively. With the cuff, less blood vessels are observed than without the cuff. Fig. 3a1-2 and b1-2 are PA B-mode images cut along the green dashed lines (i) and (ii) in Fig. 3a, respectively. The difference in vascular density with and without cuff conditions is also confirmed in the cross-sectional B-mode images. Fig. 3c and d are PA depth-encoded images of Fig. 3a and b, respectively. After cuffing, petechiae, small areas of bleeding near the skin caused by the elevated blood pressure, are seen in Fig. 3d (yellow arrows). Representative PA B-mode images of the finger (along the white dashed line (iii) in Fig. 3a) with and without the cuff are shown in Fig. 3e and f, respectively. The change in vessel density is obvious when comparing these two B-mode images.

Fig. 3.

Fig. 3

Real-time photoacoustic monitoring of blood perfusion in a finger during brachial cuffing and after release. PA MAP images (a) during and (b) after branchial cuffing. (a1, a2) PA B-mode images cut along lines (i) and (ii) in (a), respectively, and (b1, b2) PA B-mode images cut along lines (i) and (ii) in (b), respectively. (c, d) PA depth-encoded images of (a) and (b), respectively. Depth refers to the distance from the skin surface. (e, f) PA B-mode images with and without the cuff respectively, during continuous PA B-mode imaging near line (iii) in (a). (g) Relative change in vessel density over time. The cuff was released at 15 s. After the cuff is released, the vessel density increases for 10.6 s, and stays the same thereafter. All scale bars are 200 μm. PA, photoacoustic, and MAP, maximum amplitude projection.

Additionally, we quantified the change in vessel density in Fig. 3g and Supplementary Video S2. Initially, the PA signals from the skin surface were removed using the skin profile estimator in 3D Photoacoustic Visualization Studio (3D PHOVIS) [41]. Then, after thresholding the PA B-mode images using a value four times higher than the noise level to clearly separate the PA signals from the blood vessels, we counted the number of filtered pixels. Finally, the relative vessel density was calculated by normalizing the number of the blood vessels without the cuff to that with the cuff (Fig. 3g). The blood vessels are about 2.8 times denser without the cuff than with the cuff. Further, we fitted the relative vessel density to the sigmoidal curve using a Boltzmann fitting, and the curve was linearly approximated at the center of the curve. The blood perfusion time is 10.6 s from the cuff release to recovery, measured by calculating the time difference between the 10 % and 90 % points of the linearized slope. These results are consistent with previously reported results on humans with arterial occlusion. This rapid and exaggerated reperfusion is reactive hyperemia that significantly reduces microvascular resistance and quickly supplies blood to oxygen-starved tissues in response to temporary ischemia [42]. The amount of reperfusion is proportional to the microvascular ability to dilate blood vessels, and reactive hyperemia predicts cardiovascular events because it evaluates impaired vasodilatory function [43].

A previously reported study on human palms [44] compared relative changes in the averaged value of PA signals on the capillary bed. This study could not investigate single blood vessels because the experimental instrument had low spatial resolutions, 70-μm lateral and 54-μm axial resolutions. Further, the low temporal resolution provided by the motor-driven B-mode imaging speed of about 1 Hz made it difficult to provide an accurate perfusion time. Compared to the previous study, our PAM's improvement on spatial and temporal resolutions enables to visualize single capillaries and observe vascular density in real time. We believe that we can estimate the blood perfusion time more accurately than in previous studies, thanks to our system's high resolution and high-speed imaging capability. A previous study [45] using the OCTA technique in the human nailfold with direct pressure, not a brachial cuff, observed a reduced signal from capillary regions and the occluded blood vessels when pressure was applied, and confirmed reperfusion after pressure was released. These OCTA results agree well with those of our PA imaging. However, while the OCTA study applied pressure directly to the skin area within the imaging FOV, we used brachial cuffing to block arterial and venous flows to prevent other tissues from being deformed. Other studies [46,47] have used using clinical PA/US systems to measure changes in PA signals from muscle and arteries/veins in the human forearm during the occlusion. Clinical systems can image deep tissues and vessels by sacrificing spatial resolution. Although these studies targeted larger blood vessels than we did, their results are consistent with ours: PA signals decreased during occlusion. In the future, the combination of the PAM and clinical PA/US systems, with their different imaging resolutions and depths [48], may allow us to better understand blood perfusion from microvessels to macrovessels.

Additionally, the high-resolution and high-speed PAM system could be a new tool for counting petechiae, such as those shown in Fig. 3d, during the tourniquet test for capillary fragility [49]. According to World Health Organization (WHO), the tourniquet test is considered positive for capillary fragility when 20 or more petechiae per 2.5 cm2 (1 inch square) can be observed by the naked eye [50]. In our case, over 20 petechiae per 2 mm2 can be seen, because the high-resolution volumetric PAM could detect small petechiae, invisible to the naked eye, on and under the skin. Compared to conventional visual examination, high-resolution and high-speed capabilities enable the PAM to detect up to 10-um petechiae in seconds, both on the skin and under the skin. We found that the petechiae on and under the skin can be photoacoustically detected, which suggests that the high-resolution and high-speed PAM system could be a new quantification tool to count the number of the petechiae during the tourniquet test.

3.3. Change in oxygen saturation during arterial occlusion

To investigate the functional imaging capability of our system, we monitored the sO2 change in finger vessels during arterial occlusion. Six consecutive sets of volumetric PA images were acquired, over 400 s, at 532 and 559 nm. The three steps of the sO2 image and signal processing procedures are shown schematically in Fig. 4(a). In the first step, the PA B-mode images were pre-processed as follows: (1) The skin layer on the raw PA data was removed using the skin profile estimator in 3D PHOVIS. (2) The PA signal levels from focused and unfocused regions were equalized, using the background PA signals to compensate for optical and acoustic attenuation [51]. (3) The maximum values at each position were projected to get PA MAP images. In the second step, the PA MAP images were pre-processed as follows: (4) The PA MAP images were blurred to minimize differences in the PA signals due to laser fluctuation and MEMS scanning. (5) To resharpen the blood vessels’ boundaries, a high-pass filter was applied to the blurred images. In the third and final step, we used spectral unmixing to obtain the PA sO2 image [52]. (6) The sO2 values were calculated pixel by pixel. (7) Using the mask created by thresholding the PA MAP images, the valid sO2 values were extracted and made into a PA sO2 image.

Fig. 4.

Fig. 4

Change in oxygen saturation (sO2) in a finger due to arterial occlusion. (a) sO2 processing procedure. (b) Consecutive (i) sO2, (ii) vessel mask, and (iii) masked sO2 images during 400 s with a brachial cuff. The blue and red vessel masks indicate nail- and hand-side blood vessels, respectively. (c) Change in sO2 values of the nail- and hand-side blood vessels over time. The error bar represents the standard error of the averaged sO2 of the blood vessels in the vessel masks.

Fig. 4b(i) shows consecutive PA sO2 images during arterial occlusion. The overall color of the images gradually changes from red to blue, indicating decreasing sO2 values due to the curtailed oxygen supply. In the PA sO2 images, we selected big blood vessels with high SNR for reliable analysis. For blood vessels that are small or out of focus, the variability of their sO2 values was severe due to the low SNR, so these blood vessels were excluded from the analysis. To extract reliable sO2 values, we identified well-connected vessels using the connected-component labeling function in the Image Processing Toolkit in MATLAB, and then we manually picked out ten blood vessels [53]. The selected blood vessels were classified into two groups, nail-side and hand-side, and used as a mask to obtain vessel-masked PA sO2 images (Fig. 4b(ii) and (iii)). We calculated the mean and standard error of the averaged sO2 values of each blood vessel group. The sO2 values of the nail- and hand-side blood vessels decreased to 11.8 % and 9.2 % from the baseline, respectively, with the same slope (Fig. 4c). In this experiment, we quantitatively confirmed that the finger continued to consume oxygen during the arterial occlusion. The rate of sO2 decrease became smaller at approximately 5 min after the arterial occlusion. This saturation phenomenon was also observed in a previous study [44], which reported the relative PA signal ratio during arterial occlusion using wavelengths of 561 and 570 nm, on which deoxy- and oxy-hemoglobin have higher absorption coefficient than the other wavelength, respectively. The amount of the sO2 variation is similar to the sO2 difference measured in animals during the physiological states of normoxia and hypoxia [54].

A previously reported study used near-infrared spectroscopy (NIRS) imaging [55] to measure sO2 changes in human hands. The regional sO2 change in tissues during arterial occlusion showed a similar trend to that of the blood vessels in our study. However, because NIR light scatters in the tissue volume as it travels from source to detector [56], NIRS imaging cannot separate the effects of different tissues and visualize blood vessels, so it simply provides an average sO2 value for a specific area. Despite its relatively poor resolution, NIRS imaging is relatively unaffected by the subject’s motion, which enables long monitoring, including before, during, and after an occlusion. In our study, on the other hand, we could continuously monitor single blood vessels and their sO2 only during the occlusion. Due to unintended motion at the moment of occlusion or release, it was difficult to maintain the same finger position and to continue monitoring the same blood vessels before, during, and after the occlusion. To minimize the effect of inevitable motions, a finger-mounted structure such as a fingertip pulse oximeter could be an alternative. By capitalizing on the different principles and strengths of PA and NIRS, dual mode imaging techniques can provide sO2 information from capillaries and local tissues, advancing our understanding of the microcirculation.

4. Discussion and conclusion

Using high-resolution, high-speed, and spectroscopic PAM, we obtained quantitative physiological information about blood vessels in human fingers in vivo. We also extracted the heart rate from the axial position displacement of the blood vessels. Currently, photoplethysmography (PPG) is frequently used to obtain the heart rate [57], directly measuring variations in blood volume caused by arterial pulsation [58]. Pursuing a different approach, the PA imaging in our study indirectly extracted the axial position displacement induced by arterial pulsation. Because arterial volume variations and the accompanying displacement of the surrounding blood vessels are interrelated, analyzing these physical changes requires simultaneous PA imaging and PPG sensing. In addition, the amount of the axial position displacement, such as the 24 μm shown in Fig. 2, may be related to the amplitude of the PPG signals and/or blood pressure.

During the arterial occlusion, the perfusion time and the sO2 variation were also investigated. These results for the perfusion time and sO2 variation are consistent with those in previous non-imaging-based studies. In addition, previous clinical NIRS studies also showed similar results for perfusion time and sO2 variation in healthy volunteers, although not providing imaging results [59,60]. From a wider perspective, by using multiple wavelengths, the PPG technique can also estimate the sO2 in arteries [61]. PA imaging and PPG sensing provide the non-arterial and arterial sO2, respectively, so knowing the sO2 in different types of blood vessels at a time would be useful in analyzing oxygen transfer in the blood vessels. Furthermore, the combination of the above-mentioned imaging and sensing techniques with PA, PPG and NIRS could help understand the blood circulation on both macroscopic and microscopic levels. In addition, the simultaneous imaging with multiple sources implemented in this paper allows multi-resolution imaging (e.g., optical- and acoustic-resolution PA imaging) and/or multi-modal imaging (e.g., US/PA imaging), which can visualize deeper blood vessels or surrounding anatomical structures [62,63]. The PAM could be a vascular visualization tool for peripheral vascular disease because it provides vascular and functional imaging. In addition to the applicability of the PAM itself, it can be easily combined with optical and ultrasonic technologies. The PAM study using a transparent ultrasonic transducer [64] is accelerating the combination of these technologies, which would help to diagnose PAD by providing multiple clinical information.

In conclusion, we visualized and measured multiparametric information about human finger microvessel networks in vivo, using high-resolution, high-speed, spectroscopic PAM. The high-resolution and spectroscopic capabilities enabled visualizing single blood vessels and measuring the sO2 in the blood vessels. With the cost-effective configuration of the single US transducer and MEMS scanner, the high-speed imaging capability of the PAM system made it possible to measure the clinical indicators, the heart rate, perfusion time, and sO2 variation. These results show that vascular dynamics and some clinical indicators can be measured and evaluated via photoacoustic monitoring. Nevertheless, several improvements are easily envisioned. Either finger-mounted or handheld devices, or perhaps auxiliary tools would allow conducting experiments in a comfortable position that minimizes inevitable motions [[65], [66], [67]]. The heartrate and vascular density change were quantified on 2D images due to slow 3D imaging speed. However, these in-plane quantification may be inaccurate because it is affected by out-of-plane motions. For more reliable results, it is necessary to develop better laser sources with faster repetition rates, less fluctuation, and more wavelengths. Motion correction can be applied to mitigate the inevitable motions [68,69], and vascular enhancement can be adopted to achieve high-quality blood vessel images [70,71]. From the improvement, if real-time 3D imaging is available, the heartbeat and vascular density could be more accurate. Further investigations of the pressure, position, and application time of the brachial cuff for occlusion are needed to obtain clinically meaningful results. With such technical improvements, it is essential to conduct comparative clinical studies on the proposed vascular functions at the distal ends between different groups. (e.g., young vs. elderly or patients vs. controls). In the future, the high-resolution, high-speed, spectroscopic PAM could provide more reliable physiological results and become a fully realized tool for screening, diagnosing, and predicting the progress of vascular conditions.

Declaration of Competing Interest

Chulhong Kim and Jin Young Kim have financial interests in OPTICHO, which, however, did not support this research.

Acknowledgments

This research was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education and the Ministry of Science and ICT (2020R1A6A1A03047902, 2019R1A2C2006269, 2020R1C1C1013549), by National R&D Program through the National Research Foundation of Korea (NRF) funded by Ministry of Science and ICT (2020M3H2A1078045), by Korea Institute for Advancement of Technology (KIAT) grant funded by the Korea Government (MOTIE) (P0008763), and by BK21 FOUR project. The 559 nm pulsed laser was supported by Advanced Optowave.

Biographies

graphic file with name fx1.jpg

Joongho Ahn received his B.S in Electronics Engineering at Kyungbook National University and pursues Ph.D. degree in the Department of Convergence IT Engineering at Pohang University of Science and Technology. His research interests include photoacoustic imaging systems, and its integration with other imaging and sensing technique, such as ultrasound imaging, optical coherent tomography, and photoplethysmography, and their preclinical and clinical uses.

graphic file with name fx2.jpg

Dr. Jin Young Kim received his Ph.D. degree and post-doctoral training from the Departments of Mechanical Engineering and Convergence IT Engineering at Pohang University of Science and Technology (POSTECH), Republic of Korea. He is currently a Research Assistant Professor in the Department of Convergence IT Engineering at POSTECH, working on novel photoacoustic imaging, and multimodal optical imaging systems, and their clinical applications and translations.

graphic file with name fx3.jpg

Dr. Wonseok Choi received the B. S degree (summa cum laude) in 2012, and the Ph.D. degree in 2020, both from the Department of Electrical Engineering in Pohang University of Science and Technology (POSTECH) in Republic of Korea. Currently, he is a researcher in the Medical Device Innovation Center at POSTECH, working on ultrasound/photoacoustic image-guided focused ultrasound therapy and clinical photoacoustic/ultrasound imaging systems.

graphic file with name fx4.jpg

Dr. Chulhong Kim studied for his Ph.D. degree and post-doctoral training at Washington University in St. Louis, St. Louis, Missouri. He currently holds Mueunjae Chaired Professorship and a Professor of Electrical Engineering, Convergence IT Engineering, and Mechanical Engineering at Pohang University of Science and Technology (POSTECH) in Republic of Korea. He is also the Director of Medical Device Innovation Center at POSTECH supported by Ministry of Education. He was the recipients of the 2017 IEEE EMBS Early Career Achievement Award, the 2017 Korean Academy of Science and Technology Young Scientist Award, the 2016 Nightingale Award from IFMBE, and the 2017 KOSOMBE Young Investigator Award “Contributions to multi-scale photoacoustic imaging from super-resolution atomic force photoactivated microscopy for research to systems for clinical applications.

Footnotes

Appendix A

Supplementary material related to this article can be found, in the online version, at doi:https://doi.org/10.1016/j.pacs.2021.100282.

Contributor Information

Joongho Ahn, Email: joongho.ahn@postech.ac.kr.

Jin Young Kim, Email: ronsan@postech.ac.kr.

Wonseok Choi, Email: zoids1215@postech.ac.kr.

Chulhong Kim, Email: chulhong@postech.edu.

Appendix A. Supplementary data

The following are Supplementary data to this article:

Download video file (1.3MB, mp4)
Download video file (7.1MB, mp4)

References

  • 1.Cabrera Fischer E.I. Biomechanical Modeling of the Cardiovascular System. IOP Publishing; 2019. Structural basis of the circulatory system; pp. 1-1–1-18. [Google Scholar]
  • 2.Padsalgikar A.D. Cardiovascular system: structure, assessment, and diseases. In: Padsalgikar A.D., editor. Plastics in Medical Devices for Cardiovascular Applications. William Andrew Publishing; 2017. pp. 103–132. [Google Scholar]
  • 3.Guven G., Hilty M.P., Ince C. Microcirculation: physiology, pathophysiology, and clinical application. Blood Purif. 2020;49(1–2):143–150. doi: 10.1159/000503775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Otah K.E., Otah E., Clark L.T., Salifu M.O. Relationship of lower extremity skin blood flow to the ankle brachial index in patients with peripheral arterial disease and normal volunteers. Int. J. Cardiol. 2005;103(1):41–46. doi: 10.1016/j.ijcard.2004.08.043. [DOI] [PubMed] [Google Scholar]
  • 5.Beckman Joshua A., Duncan Meredith S., Damrauer Scott M., Wells Quinn S., Barnett Joey V., Wasserman David H., Bedimo Roger J., Butt Adeel A., Marconi Vincent C., Sico Jason J., Tindle Hilary A., Bonaca Marc P., Aday Aaron W., Freiberg Matthew S. Microvascular disease, peripheral artery disease, and amputation. Circulation. 2019;140(6):449–458. doi: 10.1161/CIRCULATIONAHA.119.040672. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Fleischmann D., Hallett R.L., Rubin G.D. CT angiography of peripheral arterial disease. J. Vasc. Interv. Radiol. 2006;17(1):3–26. doi: 10.1097/01.RVI.0000191361.02857.DE. [DOI] [PubMed] [Google Scholar]
  • 7.Ersoy H., Rybicki F.J. MR Angiography of the lower extremities. Am. J. Roentgenol. 2008;190(6):1675–1684. doi: 10.2214/AJR.07.2223. [DOI] [PubMed] [Google Scholar]
  • 8.Evans D.H., Jensen J.A., Nielsen M.B. Ultrasonic colour Doppler imaging. Interface Focus. 2011;1(4):490–502. doi: 10.1098/rsfs.2011.0017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Lin E., Alessio A. What are the basic concepts of temporal, contrast, and spatial resolution in cardiac CT? J. Cardiovasc. Comput. Tomogr. 2009;3(6):403–408. doi: 10.1016/j.jcct.2009.07.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Karbalaie A., Emrani Z., Fatemi A., Etehadtavakol M., Erlandsson B.-E. Practical issues in assessing nailfold capillaroscopic images: a summary. Clin. Rheumatol. 2019;38(9):2343–2354. doi: 10.1007/s10067-019-04644-9. [DOI] [PubMed] [Google Scholar]
  • 11.Cutolo M., Cerinic M. Nailfold videocapillaroscopy for the early diagnosis of systemic sclerosis in Raynaud’s phenomenon. Future Rheumatol. 2006;1:41–51. [Google Scholar]
  • 12.Liu M., Drexler W. Optical coherence tomography angiography and photoacoustic imaging in dermatology. Photochem. Photobiol. Sci. 2019;18(5):945–962. doi: 10.1039/c8pp00471d. [DOI] [PubMed] [Google Scholar]
  • 13.Braaf B., Gräfe M.G.O., Uribe-Patarroyo N., Bouma B.E., Vakoc B.J., de Boer J.F., Donner S., Weichsel J. OCT-based velocimetry for blood flow quantification. In: Bille J.F., editor. High Resolution Imaging in Microscopy and Ophthalmology: New Frontiers in Biomedical Optics. Springer International Publishing; Cham: 2019. pp. 161–179. [Google Scholar]
  • 14.Shau Poh Chong C.W.M., Leahy Conor, Srinivasan Vivek J. Cerebral metabolic rate of oxygen (CMRO2) assessed by combined Doppler and spectroscopic OCT. Biomed. Opt. Express. 2015;6(10):3941–3951. doi: 10.1364/BOE.6.003941. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Song W., Shao W., Yi W., Liu R., Desai M., Ness S., Yi J. Visible light optical coherence tomography angiography (vis-OCTA) and local microvascular retinal oximetry in human retina. bioRxiv. 2020 doi: 10.1364/BOE.395843. 2020.04.17.047514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Spaide R.F., Fujimoto J.G., Waheed N.K. Image artifacts in optical coherence tomography angiography. Retina. 2015;35(11):2163–2180. doi: 10.1097/IAE.0000000000000765. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Wang L.V., Yao J. A practical guide to photoacoustic tomography in the life sciences. Nat. Methods. 2016;13(8):627–638. doi: 10.1038/nmeth.3925. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hsun-Chia H., Lidai W., Lihong V.W. In vivo photoacoustic microscopy of human cuticle microvasculature with single-cell resolution. J. Biomed. Opt. 2016;21(5):1–8. doi: 10.1117/1.JBO.21.5.056004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Baik J.W., Kim J.Y., Cho S., Choi S., Kim J., Kim C. Super wide-field photoacoustic microscopy of animals and humans in vivo. IEEE Trans. Med. Imaging. 2020;39(4):975–984. doi: 10.1109/TMI.2019.2938518. [DOI] [PubMed] [Google Scholar]
  • 20.Junjie Yao K.I.M., Shi Yunfei, Taber Larry A., Wang Lihong V. In vivo photoacoustic imaging of transverse blood flow by using Doppler broadening of bandwidth. Opt. Lett. 2010;35(9):1419–1421. doi: 10.1364/OL.35.001419. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Song Hu K.M., Wang Lihong V. Noninvasive label-free imaging of microhemodynamics by optical-resolution photoacoustic microscopy. Opt. Express. 2009;17(9):7688–7693. doi: 10.1364/oe.17.007688. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Liu C., Liang Y., Wang L. Single-shot photoacoustic microscopy of hemoglobin concentration, oxygen saturation, and blood flow in sub-microseconds. Photoacoustics. 2020;17 doi: 10.1016/j.pacs.2019.100156. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kim J.Y., Lee C., Park K., Lim G., Kim C. Fast optical-resolution photoacoustic microscopy using a 2-axis water-proofing MEMS scanner. Sci. Rep. 2015;5(1):7932. doi: 10.1038/srep07932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Lan B., Liu W., Wang Y.-C., Shi J., Li Y., Xu S., Sheng H., Zhou Q., Zou J., Hoffmann U., Yang W., Yao J. High-speed widefield photoacoustic microscopy of small-animal hemodynamics. Biomed. Opt. Express. 2018;9(10):4689–4701. doi: 10.1364/BOE.9.004689. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Kim J., Kim J.Y., Jeon S., Baik J.W., Cho S.H., Kim C. Super-resolution localization photoacoustic microscopy using intrinsic red blood cells as contrast absorbers. Light Sci. Appl. 2019;8(1):103. doi: 10.1038/s41377-019-0220-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lee C., Kim J.Y., Kim C. Recent progress on photoacoustic imaging enhanced with microelectromechanical systems (MEMS) technologies. Micromachines. 2018;9(11) doi: 10.3390/mi9110584. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Kim J.Y., Lee C., Park K., Han S., Kim C. High-speed and high-SNR photoacoustic microscopy based on a galvanometer mirror in non-conducting liquid. Sci. Rep. 2016;6(1):34803. doi: 10.1038/srep34803. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Kim J.Y., Lee C., Park K., Lim G., Kim C. A PDMS-Based 2-Axis waterproof scanner for photoacoustic microscopy. Sensors. 2015;15(5) doi: 10.3390/s150509815. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Liu W., Yao J. Photoacoustic microscopy: principles and biomedical applications. Biomed. Eng. Lett. 2018;8(2):203–213. doi: 10.1007/s13534-018-0067-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Bi R., Dinish U.S., Goh C.C., Imai T., Moothanchery M., Li X., Kim J.Y., Jeon S., Pu Y., Kim C., Ng L.G., Wang L.V., Olivo M. In vivo label-free functional photoacoustic monitoring of ischemic reperfusion. J. Biophotonics. 2019;12(7) doi: 10.1002/jbio.201800454. [DOI] [PubMed] [Google Scholar]
  • 31.Aguirre J., Hindelang B., Berezhnoi A., Darsow U., Lauffer F., Eyerich K., Biedermann T., Ntziachristos V. Assessing nailfold microvascular structure with ultra-wideband raster-scan optoacoustic mesoscopy. Photoacoustics. 2018;10:31–37. doi: 10.1016/j.pacs.2018.02.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Upputuri P.K., Pramanik M. Fast photoacoustic imaging systems using pulsed laser diodes: a review. Biomed. Eng. Lett. 2018;8(2):167–181. doi: 10.1007/s13534-018-0060-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Jeon S., Kim J., Lee D., Baik J.W., Kim C. Review on practical photoacoustic microscopy. Photoacoustics. 2019;15 doi: 10.1016/j.pacs.2019.100141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Choi W., Park E.-Y., Jeon S., Kim C. Clinical photoacoustic imaging platforms. Biomed. Eng. Lett. 2018;8(2):139–155. doi: 10.1007/s13534-018-0062-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Chen J., Zhang Y., He L., Liang Y., Wang L. Wide-field polygon-scanning photoacoustic microscopy of oxygen saturation at 1-MHz A-line rate. Photoacoustics. 2020;20 doi: 10.1016/j.pacs.2020.100195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Moothanchery M., Bi R., Kim J.Y., Balasundaram G., Kim C., Olivo M.C. High-speed simultaneous multiscale photoacoustic microscopy. J. Biomed. Opt. 2019;24(8):1–7. doi: 10.1117/1.JBO.24.8.086001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Park S.M., Kim D.Y., Cho S.-W., Kim B.-M., Lee T.G., Kim C.-S., Lee S.-W. Quickly alternating green and red laser source for real-time multispectral photoacoustic microscopy. Photoacoustics. 2020;20 doi: 10.1016/j.pacs.2020.100204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Zhang W., Li Y., Yu Y., Derouin K., Qin Y., Nguyen V.P., Xia X., Wang X., Paulus Y.M. Simultaneous photoacoustic microscopy, spectral-domain optical coherence tomography, and fluorescein microscopy multi-modality retinal imaging. Photoacoustics. 2020;20 doi: 10.1016/j.pacs.2020.100194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Ogunlade O., Ho J.O.Y., Kalber T.L., Hynds R.E., Zhang E., Janes S.M., Birchall M.A., Butler C.R., Beard P. Monitoring neovascularization and integration of decellularized human scaffolds using photoacoustic imaging. Photoacoustics. 2019;13:76–84. doi: 10.1016/j.pacs.2019.01.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Liang S., Konstantin I.M., Shung K.K., Lihong V.W. Ultrasound-array-based real-time photoacoustic microscopy of human pulsatile dynamics in vivo. J. Biomed. Opt. 2010;15(2):1–4. doi: 10.1117/1.3333545. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Cho S., Baik J., Managuli R., Kim C. 3D PHOVIS: 3D photoacoustic visualization studio. Photoacoustics. 2020;18 doi: 10.1016/j.pacs.2020.100168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Philpott A., Anderson Todd J. Reactive hyperemia and cardiovascular risk. Arterioscler. Thromb. Vasc. Biol. 2007;27(10):2065–2067. doi: 10.1161/ATVBAHA.107.149740. [DOI] [PubMed] [Google Scholar]
  • 43.Rosenberry R., Nelson M.D. Reactive hyperemia: a review of methods, mechanisms, and considerations. Am. J. Physiol. Regul. Integr. Comp. Physiol. 2020;318(3):R605–R618. doi: 10.1152/ajpregu.00339.2019. [DOI] [PubMed] [Google Scholar]
  • 44.Christopher P.F., Lihong V.W., Lynn A.C. In vivo functional photoacoustic microscopy of cutaneous microvasculature in human skin. J. Biomed. Opt. 2011;16(2):1–5. doi: 10.1117/1.3536522. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Woo June C., Hequn W., Ruikang K.W. Optical coherence tomography microangiography for monitoring the response of vascular perfusion to external pressure on human skin tissue. J. Biomed. Opt. 2014;19(5):1–9. doi: 10.1117/1.JBO.19.5.056003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Karlas A., Kallmayer M., Fasoula N.-A., Liapis E., Bariotakis M., Krönke M., Anastasopoulou M., Reber J., Eckstein H.-H., Ntziachristos V. Multispectral optoacoustic tomography of muscle perfusion and oxygenation under arterial and venous occlusion: a human pilot study. J. Biophotonics. 2020;13(6) doi: 10.1002/jbio.201960169. [DOI] [PubMed] [Google Scholar]
  • 47.Yang J., Zhang G., Chang W., Chi Z., Shang Q., Wu M., Pan T., Huang L., Jiang H. Photoacoustic imaging of hemodynamic changes in forearm skeletal muscle during cuff occlusion. Biomed. Opt. Express. 2020;11(8):4560–4570. doi: 10.1364/BOE.392221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Attia A.B.E., Balasundaram G., Moothanchery M., Dinish U.S., Bi R., Ntziachristos V., Olivo M. A review of clinical photoacoustic imaging: current and future trends. Photoacoustics. 2019;16 doi: 10.1016/j.pacs.2019.100144. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Grande A.J., Reid H., Thomas E., Foster C., Darton T.C. Tourniquet test for dengue diagnosis: systematic review and meta-analysis of diagnostic test accuracy. PLoS Negl. Trop. Dis. 2016;10(8):e0004888. doi: 10.1371/journal.pntd.0004888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.World Health O. World Health Organization; Geneva: 1997. Dengue Haemorrhagic Fever: Diagnosis, Treatment, Prevention and Control. [Google Scholar]
  • 51.Oraevsky A.A., Clingman B., Zalev J., Stavros A.T., Yang W.T., Parikh J.R. Clinical optoacoustic imaging combined with ultrasound for coregistered functional and anatomical mapping of breast tumors. Photoacoustics. 2018;12:30–45. doi: 10.1016/j.pacs.2018.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Li M., Tang Y., Yao J. Photoacoustic tomography of blood oxygenation: a mini review. Photoacoustics. 2018;10:65–73. doi: 10.1016/j.pacs.2018.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Haralick R.M., Shapiro L.G. Addison-Wesley; 1992. Computer and Robot Vision. [Google Scholar]
  • 54.Zhang H.F., Maslov K., Sivaramakrishnan M., Stoica G., Wang L.V. Imaging of hemoglobin oxygen saturation variations in single vessels in vivo using photoacoustic microscopy. Appl. Phys. Lett. 2007;90(5) [Google Scholar]
  • 55.Gargani L., Bruni C., Barskova T., Hartwig V., Marinelli M., Trivella M.G., Matucci-Cerinic M., L’Abbate A. Near-infrared spectroscopic imaging of the whole hand: a new tool to assess tissue perfusion and peripheral microcirculation in scleroderma. Semin. Arthritis Rheum. 2019;48(5):867–873. doi: 10.1016/j.semarthrit.2018.08.002. [DOI] [PubMed] [Google Scholar]
  • 56.Ferrari M., Mottola L., Quaresima V. Principles, techniques, and limitations of near infrared spectroscopy. Can. J. Appl. Physiol. 2004;29(4):463–487. doi: 10.1139/h04-031. [DOI] [PubMed] [Google Scholar]
  • 57.Castaneda D., Esparza A., Ghamari M., Soltanpur C., Nazeran H. A review on wearable photoplethysmography sensors and their potential future applications in health care. Int. J. Biosens. Bioelectron. 2018;4(4):195–202. doi: 10.15406/ijbsbe.2018.04.00125. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Tamura T. Current progress of photoplethysmography and SPO2 for health monitoring. Biomed. Eng. Lett. 2019;9(1):21–36. doi: 10.1007/s13534-019-00097-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Kyle B., Litton E., Ho K.M. Effect of hyperoxia and vascular occlusion on tissue oxygenation measured by near infra-red spectroscopy (InSpectra™): a volunteer study. Anaesthesia. 2012;67(11):1237–1241. doi: 10.1111/j.1365-2044.2012.07265.x. [DOI] [PubMed] [Google Scholar]
  • 60.Guoqiang Y., Turgut D., Gwen L., Chao Z., Britton C., Emile R.M.M.D., Arjun G.Y. Time-dependent blood flow and oxygenation in human skeletal muscles measured with noninvasive near-infrared diffuse optical spectroscopies. J. Biomed. Opt. 2005;10(2):1–12. doi: 10.1117/1.1884603. [DOI] [PubMed] [Google Scholar]
  • 61.Abay T.Y., Kyriacou P.A. Photoplethysmography for blood volumes and oxygenation changes during intermittent vascular occlusions. J. Clin. Monit. Comput. 2018;32(3):447–455. doi: 10.1007/s10877-017-0030-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Choi S., Kim J.Y., Lim H.G., Baik J.W., Kim H.H., Kim C. Versatile single-element ultrasound imaging platform using a water-proofed MEMS scanner for animals and humans. Sci. Rep. 2020;10(1):6544. doi: 10.1038/s41598-020-63529-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Park J., Jeon S., Meng J., Song L., Lee J., Kim C. Delay-multiply-and-sum-based synthetic aperture focusing in photoacoustic microscopy. J. Biomed. Opt. 2016;21(3):1–10. doi: 10.1117/1.JBO.21.3.036010. [DOI] [PubMed] [Google Scholar]
  • 64.Park J., Park B., Kim T.Y., Jung S., Choi W.J., Ahn J., Yoon D.H., Kim J., Jeon S., Lee D., Yong U., Jang J., Kim W.J., Kim H.K., Jeong U., Kim H.H., Kim C. Quadruple ultrasound, photoacoustic, optical coherence, and fluorescence fusion imaging with a transparent ultrasound transducer. Proc. Natl. Acad. Sci. U. S. A. 2021;118(11) doi: 10.1073/pnas.1920879118. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Dangi A., Agrawal S., Datta G.R., Srinivasan V., Kothapalli S.R. Towards a low-cost and portable photoacoustic microscope for point-of-care and wearable applications. IEEE Sens. J. 2020;20(13):6881–6888. doi: 10.1109/jsen.2019.2935684. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Tang J., Dai X., Jiang H. Wearable scanning photoacoustic brain imaging in behaving rats. J. Biophotonics. 2016;9(6):570–575. doi: 10.1002/jbio.201500311. [DOI] [PubMed] [Google Scholar]
  • 67.Park K., Kim J.Y., Lee C., Jeon S., Lim G., Kim C. Handheld photoacoustic microscopy probe. Sci. Rep. 2017;7(1):13359. doi: 10.1038/s41598-017-13224-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Zhao H., Chen N., Li T., Zhang J., Lin R., Gong X., Song L., Liu Z., Liu C. Motion correction in optical resolution photoacoustic microscopy. IEEE Trans. Med. Imaging. 2019;38(9):2139–2150. doi: 10.1109/TMI.2019.2893021. [DOI] [PubMed] [Google Scholar]
  • 69.Schwarz M., Garzorz-Stark N., Eyerich K., Aguirre J., Ntziachristos V. Motion correction in optoacoustic mesoscopy. Sci. Rep. 2017;7(1):10386. doi: 10.1038/s41598-017-11277-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Zhao H., Ke Z., Yang F., Li K., Chen N., Song L., Zheng C., Liang D., Liu C. Deep learning enables superior photoacoustic imaging at ultralow laser dosages. Adv. Sci. 2021;8(3) doi: 10.1002/advs.202003097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Zhao H., Li K., Chen N., Zhang K., Wang L., Lin R., Gong X., Song L., Liu Z., Liu C. Multiscale vascular enhancement filter applied to in vivo morphologic and functional photoacoustic imaging of rat ocular vasculature. IEEE Photonics J. 2019;11(6):1–12. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Download video file (1.3MB, mp4)
Download video file (7.1MB, mp4)

Articles from Photoacoustics are provided here courtesy of Elsevier

RESOURCES