Abstract
Organ-dedicated PET scanners are becoming more prevalent because of their advantages in higher sensitivity, improved image quality, and lower cost. Detectors utilized in these scanners have finer pixel size with depth of interaction (DOI) capability. This work presents a LYSO(Ce) detector module with DOI capability which has the potential to be scaled up to a high-resolution small animal or organ-dedicated PET system. For DOI capability, a submodule with one LYSO block detector utilizing PETsys TOFPET2 application-specific integrated circuit (ASIC) was previously developed in our lab. We scaled up the submodule and optimized the configuration to allow for a compact housing of the dual-readout boards in one side of the blocks by designing a high-speed dual-readout cable to maintain the original pin-to-pin relationship between the Samtec connectors. The module size is 53.8 × 57.8 mm2. Each module has 2 × 2 LYSO blocks, each LYSO block consists of 4 × 4 LYSO units, and each LYSO unit contains a 6 × 6 array of 1 × 1 × 20 mm3 LYSO crystals. The four lateral surfaces of LYSO crystal were mechanically ground to W14, and the two end surfaces were polished. Two ends of the LYSO crystal are optically connected to SiPM for DOI measurement. Eight LYSO blocks performance including energy, timing, and DOI resolution is characterized with a single LYSO slab. The in-panel and orthogonal-panel spatial resolution of the two modules with 107.4 mm distance between each other are measured at 9 positions within the field of view (FOV) with a 22Na source. Results show that the average energy, timing, and DOI resolution of all LYSO blocks are 16.13% ± 1.01% at 511 keV, 658.03 ± 15.18 ps, and 2.62 ± 0.06 mm, respectively. The energy and timing resolution of two modules are 16.35% and 0.86 ns, respectively. The in-panel and orthogonal-panel spatial resolution of the two modules at the FOV center are 1.9 and 4.4 mm respectively.
Keywords: depth of interaction, lutetium-yttrium oxyorthosilicate (LYSO), silicon photomultipliers (SiPM), positron emission tomography, time, energy and spatial resolution
1. Introduction
Organ-dedicated positron emission tomography (PET) scanners can provide improved spatial resolution by using finer scintillator crystals, higher system sensitivity by placing the detectors closer to the organ, better image contrast recovery by reducing the noise from other organs, and lower cost by reducing scanner geometry (González et al 2018). To adapt to different imaging environment and offer the flexibility to place the detectors around the organ, limited angle geometry instead of complete ring-shaped geometry is utilized in systems dedicating to breast (MacDonald et al 2009, Raylman 2009), heart (Peng 2015), and prostate (Zorraquino et al 2013).
The measurement of the DOI is of critical importance for a PET scanner with compact geometry to simultaneously achieve high and uniform spatial resolution, and high photon sensitivity (Bailey et al 2003, Yang et al 2016). Developing a practical DOI-capable PET detector with balanced performance has been studied for over two decades, including but not limited to dual-ended readout method using two detectors coupled to the both ends of crystal (Moses and Derenzo 1994, Miyaoka et al 1998, Yang et al 2006, Ling et al 2007, Maas et al 2009), single-ended readout method using one detector coupled to one end of crystal (Berg et al 2016), side readout method by reading signal from crystal lateral surface (Yamaya et al 2006, Yeom et al 2014), light-sharing window-based method (Zhang et al 2019), monolithic scintillator detector based method (Schaart et al 2009), single-ended readout method with light guide on the top surface (Pizzichemi et al 2016). Among all DOI methods, dual-ended readout method has the advantages of a continuous DOI measurement and a balanced detector performance by increasing the total light output (Shao et al 2002, Burr et al 2004, Yang et al 2008, Kuang et al 2017).
The selection of photon sensors and their proper coupling to scintillators has a relatively large impact on the performance of final PET system. The conventional photomultiplier tube (PMT) is bulky and has a large insensitive area, and the gain across the photo-cathode of a PMT is non-uniform. The avalanche photodiode (APD) has a relatively large insensitive area at the edge and an intrinsically low signal-to-noise ratio (Yang et al 2008). Over the last few years, silicon photomultiplier (SiPM) has advanced tremendously in terms of cost and performance. Compared with PMT and APD, SiPM offers high amplification gain, compact size, small insensitive area, and low voltage bias, and has been widely used for dual-ended readout design (Herbert et al 2006, Schaart et al 2009, Yamaya et al 2011, Nadig et al 2019).
Our lab has developed a high-resolution PET detector submodule with dual-ended readout capability based on SiPMs and PETsys TOFPET2 application-specific integrated circuits (ASIC), which allows us to read out each SiPM channel independently (Li and Abbaszadeh 2019). In this study, the submodule is completed and configured to a full module that can be further scaled up to build high-resolution PET systems with DOI capabilities. To make the geometry compact, we optimized the previous lutetium-yttrium oxyorthosilicate (LYSO) block detector configuration by designing a high-speed dual-readout cable that accommodates the readout electronics on the same side of the scintillator. The energy, timing, and DOI resolution of eight detector blocks with dual-readout cable are characterized and are compared with previous design (Li and Abbaszadeh 2019). Four detector blocks formed a detector module with area of 53.8 × 57.8 mm2 and the energy and time resolution of the two detector modules were measured at 107.4 mm distance from each other. Spatial resolution at different positions within this FOV was measured.
2. Materials and methods
2.1. Detector module
The schematic design of the detector module is shown in figure 1. A module has 2 × 2 LYSO blocks. The details of LYSO block fabrication is shown in (Li and Abbaszadeh 2019). The analog output signals of the SiPMs were digitized by PETsys TOFPET2 ASIC (PETsys Electronics SA, Portugal) and acquired by PETsys SiPM Readout System. When making the detector module with dual-ended readout capability, the electronics would take space within the FOV and increase the panel distance of the detector module. To make the geometry compact, it can be seen that the TOFPET2 ASIC electronics were placed on one side of the LYSO block.
Figure 2 shows the picture of two detector modules developed in this work separated 107.4 mm from each other. The TOFPET2 ASICs and SiPM arrays of the detector module were cooled by a Peltier element (CUI Devices CP455535H) and a heat sink (Advanced Thermal Solutions ATS-52450P-C1-R0). The Peltier element was attached to a 3D printed lid with a thermally conductive adhesive (3M Electronic Specialty TC-2810-50ML), and the heat sink is on the top of the Peltier element. The whole lid was placed on the top of the detector module to keep the environment temperature around the detectors stable and to shield from environment light.
Figure 3 shows the specific components of TOFPET2 ASIC readout electronics, including PETsys FEB/S board and FEB/A board. FEB/S was the board coupled to SiPMs, and FEB/A was the board hosting the TOFPET2 ASIC. Both FEB/S and FEB/A boards have two LSHM connectors (Samtec Inc., US) for connecting. In order to keep FEB/S board connected to SiPM and both FEB/A boards on one side of LYSO blocks, we designed a dual-readout cable. Due to the performance advantages in high speed, wide bandwidth, small signal loss and small crosstalk, Samtec HLCD cables were used as the data cables in this application which host two LSHM connectors. Depending on the geometry requirement and the original pin-to-pin relationship, HLCD-40-06.30-TR-BL-2 and HLCD-40-06.30-TL-BR-2 (Samtec Inc., US) were chosen for this application. The picture of the two custom-made cables is shown in figure 3 bottom.
2.2. Flood histogram of LYSO unit
To characterize the detector module, the flood histogram of each LYSO unit was measured as a first step. As described in section 2.1, the PETsys TOFPET2 ASIC and PETsys SiPM Readout System were used to digitize and acquire the analog output signals from SiPM arrays boards. A 30-μCi Na-22 source (Eckert & Ziegler Inc., Germany) with an active diameter of 0.25 mm was used to irradiate the LYSO unit. During the experiments, the timing output and energy output would be read out and recorded, when a SiPM channel get triggered.
Figure 4 shows the flood histogram coordinate definition of one LYSO unit. The flood histogram was calculated using the position-encoding energy signals from the two 2 × 2 SiPM arrays as discussed in (Ren et al 2014),
(1) |
where A1, B1, C1 and D1 are the four energy outputs from the SiPM array on the one end of the LYSO unit and A2, B2, C2 and D2 are from the SiPM array on the other end. E1 and E2 are the total energy measured by the two SiPM arrays respectively as
(2) |
2.3. Energy, time and DOI resolution of LYSO block
A single 1 × 25.8 × 20 mm3 LYSO slab was used to characterize the energy resolution, timing resolution and DOI resolution of designed LYSO block by the experimental setup shown in figure 5. The LYSO slab was also read out by the same SiPM arrays as the LYSO block, but with only 8 channels aligned in a column. The same Na-22 source was used, located between the LYSO block and LYSO slab and had 20 mm distance from each of them. The Na-22 source and LYSO slab were on a translation stage, which could move along the depth direction of the LYSO block. The incident beam width was 1 mm in this experiment, which was determined by the width of the LYSO slab.
All 8 LYSO blocks within the two detector modules were measured individually by the experimental setup shown in figure 5. For the LYSO block, the side read out by the dual-readout cable is defined as depth 0 mm, and the side being directly read out is defined as depth 20 mm. Data were acquired at 5 depths (2, 6, 10, 14, 18 mm) for each detector, and each depth had a 5 min acquisition time. The SiPMs and ASICs were cooled down by a Peltier element and a fan. During the experiment, the SiPM temperature was maintained at 24.3 ± 1.9 °C.
The energy resolution (E) of an event was measured as
(3) |
The coincidence time (t) was estimated as
(4) |
where t1 to t8 are the earliest triggered timing outputs of 8 SiPM channels (corresponding to one LYSO unit) out of the 128 channels in a LYSO block (one block consists of 4 × 4 LYSO units), and t9 to t16 are the earliest triggered timing outputs of the of the LYSO slab. The DOI ratio was estimated as
(5) |
A DOI calibration curve was used to convert the DOI ratio to the interaction depth (presented in figure 9), which was obtained by a linear fit of the peak value of the DOI ratio histogram to the known depth of interaction (interaction depth = a × DOI ratio + b). The DOI resolution was acquired by a Gaussian fit of the depth histogram.
2.4. Spatial resolution
After completing the characteristics measuring of each LYSO block, all 8 LYSO blocks were assembled to two detector modules. The energy and timing resolution of the modules were characterized at 104.7 mm distance from each other. The spatial resolution within the FOV was calibrated by a Na-22 source. The spatial resolution was measured at the FOV center and 5, 10, 15, 20 mm away from the center along the in-panel (z axis) and orthogonal-panel (x axis) directions. Each source position had 300-s data acquisition time, and the temperature was controlled as 32.2 ± 0.9 °C. The increase in the temperature is due to operation of 4 LYSO blocks in the module with the same Peltier element as opposed to one LYSO block in the setup shown in figure 5.
3. Results and discussion
3.1. Flood histogram of LYSO unit
An example of the flood histogram of one LYSO unit is shown figure 6. All the 6×6 array of LYSO crystals within one LYSO unit could be resolved. The center LYSO crystals were brighter than the LYSO crystals near the sides, which indicates a higher detection efficiency in the center LYSO crystal. We described more details about crystal resolvability, light guide thickness, and how the flood histogram was obtained in (Li and Abbaszadeh 2019).
3.2. Energy, timing and DOI resolution of LYSO blocks
The energy resolution of all 8 LYSO blocks is shown in figure 7. As shown in figure 7 left, the energy resolution did not show an obvious dependence on the interaction depth over all LYSO blocks. Based on results of figure 7 right, the average energy resolution of all 8 LYSO blocks at all 5 depths was 16.13% ± 1.01% at 511 keV.
The timing resolution of all 8 LYSO blocks is shown in figure 8. The timing resolution got improved when the interaction was far away from the end that was read out by the dual-readout cable, which suggested that the dual-readout cable could slightly delay the timing trigger (either through delay of the cable itself or light traversing the LYSO to reach the side with SiPMs directly connected to the readout). Currently, we have not applied any calibration to consider the effect of this delay. The average timing resolution of all 8 LYSO blocks at all 5 depths was 658.03 ± 15.18 ps.
The DOI ratio histogram of one LYSO unit is shown in figure 9. The DOI resolution of all 8 LYSO blocks is shown in figure 10. The DOI resolution did not show an obvious dependence on the interaction depth, which was aligned with the energy resolution. The average DOI resolution of all 8 LYSO blocks at all 5 depths was 2.62 ± 0.06 mm.
Table 1 summarizes the design and performance of some dual-ended readout detectors for high-resolution PET applications. Compared with our previous direct-couple LYSO blocks detector design, using the dual-readout cables would slightly deteriorate the LYSO blocks detector performance. This is because the dual-readout cables prolong the signal transmission length, which would cause the delay and increase extra signal noise to our detector. However, compared with the other previously published results, our design still can achieve relatively good energy, timing and DOI resolution.
Table 1.
Citation | Crystal (mm3) | Detector Size (mm3) | Photo sensor | Energy (%) | Time (ns) | DOI (mm) |
---|---|---|---|---|---|---|
This paper | LYSO 1 × 1 × 20 | 25.8 × 25.8 × 20 | SiPM | 16.13 | 0.66 | 2.62 |
(Li and Abbaszadeh 2019) | LYSO 1 × 1 × 20 | 25.8 × 25.8 × 20 | SiPM | 15.66 | 0.60 | 2.33 |
(Godinez et al 2012) | LYSO 1.5 × 1.5 × 20 | 22 × 22 × 20 | PSPMT | 19 | 2.4 | 2.9 |
(Kolb et al 2014) | LSO 1.55 × 1.55 × 20 | 19.0 × 19.2 × 20 | G-APD | 12.8 | 1.14 | 2.9 |
(Shao et al 2014) | LYSO 1.9 × 1.9 × 30 | 16 × 16 × 30 | SSPM | 17.6 | 2.8 | 5.6 |
(Kuang et al 2018) | LYSO 0.97 × 0.97 × 20 | 11.6 × 11.6 × 20 | SiPM | 16.7 | 1.41 | 2.1 |
(Du et al 2018) | LYSO 0.95 × 0.95 × 20 | 46 × 46 × 20 | SiPM | 23.8 | 1.78 | 2.81 |
(Kuang et al 2019) | LYSO 0.5 × 0.5 × 20 | 10 × 10 × 20 | SiPM | 21 | 1.23 | 2.84 |
3.3. Characteristics of the two detector modules
The energy and timing resolution of the two detector modules were measured as 16.35% and 0.86 ns, respectively. The number of coincidence acquired at different source positions is shown in figure 11. It can be seen that given the off-center distance, the orthogonal-panel axis has a higher count than the in-panel axis.
Image was reconstructed with the gpurecon program (Cui et al 2013), using data collected from a single angle using the setup shown in figure 2. The voxel size was 0.5 × 0.5 × 0.5 mm3. The line profiles of all 9 source positions along the in-panel (z axis) and orthogonal-panel (x axis) directions are shown in figure 12.
The in-panel and orthogonal-panel spatial resolution were acquired by using a Gaussian function to fit the line profiles, as shown in figure 13. Specifically, the in-panel and orthogonal-panel spatial resolution at the FOV center were 1.94 mm and 4.44 mm, respectively. The results showed that spatial resolution gradually deteriorated when the source was moving away from the FOV center, and the deterioration was worse along the in-panel direction than the orthogonal-panel direction. The difference between the in-panel and the orthogonal-panel spatial resolution was due to the limited angle tomography of the two-panel geometry, which caused the incomplete angular sampling. Specifically, a pair of photons emitted approximately parallel to the panels along the in-panel axis are not likely to be detected, limiting the spatial information along the orthogonal-panel direction and causing the orthogonal-panel spatial resolution to be degraded.
Table 2 summarizes the design and spatial resolution of some breast- and brain-dedicated PET scanners at the FOV center. The comparison showed that this design could achieve a good in-panel spatial resolution, which indicates the potential of utilizing it for the high-resolution PET application.
Table 2.
Citation | Organ | Geometry | Crystal (mm3) | Resolution (mm) |
---|---|---|---|---|
This paper | — | Dual panel | 1 × 1 × 20 | IP 1.9, OP 4.4 |
(Freifelder et al 2001) | Breast | Dual plates | Continuous | IP 3.8, OP 4.6 |
(Murthy et al 2000) | Breast | Dual panel | Continuous | 2.8 |
(Baghaei et al 2000) | Breast | Partial ring | 2.7 × 2.8 × 19 | 2.8 |
(Doshi et al 2000) | Breast | Dual panel | 3 × 3 × 20 | 2.3 |
(Karimian et al 2005) | Breast | Ring | 3 × 5 × 20 | radial 2.8, axial 3.8 |
(Watanabe et al 2002) | Brain | Ring | 2.8 × 6.6 × 30 | radial 2.9, axial 2.9 |
(Yoshida et al 2006) | Brain | Ring | 2.9 × 2.9 × 7.5 | radial 3.1, axial 3.1 |
(Yamamoto et al 2011) | Brain | Ring | 4.9 × 5.9 × 7 | radial 4.0, axial 3.5 |
(Bauer et al 2016) | Brain | Ring | 1.5 × 1.5 × 10 | radial 2.8, axial 2.0 |
(Wienhard et al 2002) | Brain | Octagon | 2.1 × 2.1 × 7.5 | 2.4 |
(Tashima et al 2015) | Brain | Helmet-chin | 2.8 × 2.8 × 7.5 | 3.0 |
IP: in-panel. OP: orthogonal-panel.
The comparison of performance in tables 1 and 2 shows that our design system could achieve relatively good energy, timing, DOI resolution, and spatial resolution. Compared with other dual-ended readout PET detectors, our detector has the following advantages:
We carefully chose the surface roughness of LYSO crystal. The two end surfaces of LYSO crystal were polished to improve the light collection efficiency. The W14 roughness (roughness around 10–14 μm) of lateral surfaces is between the polished surface and the saw-cut surface. This design optimized a good dependency of light collection on depth of interaction with good energy and timing resolution.
We developed a high-speed dual-readout cable that allows for a compact housing of dual-readout boards in one side of detector blocks with low signal loss.
We used the TOFPET2 ASICs and PETsys electronics that are commercially available and scalable to high channel density.
4. Conclusions
In this work, a dual-ended readout detector module based on SiPM and TOFPET2 ASIC is characterized. The LYSO crystal size was 1 × 1 × 20 mm3 and the assembled module size is 53.8 × 57.8 mm2. To achieve a compact geometry, the ASICs and readout electronics are placed on the same side of the LYSO block by designing a high speed dual-readout cable. The energy, coincidence timing, and DOI resolution of the LYSO blocks were characterized as 16.13% ± 1.01%, 658.03 ± 15.18 ps and 2.62 ± 0.06 mm FWHM, respectively. Compared with previous direct-couple detector design, using the dual-readout cables would slightly deteriorate the detector performance. With no calibration for the cable delay, the time resolution of the two detector modules (4 LYSO blocks in coincidence with 4 LYSO blocks) was 860 ps. The spatial resolution of two detector modules is characterized by a Na-22 source. The in-panel and orthogonal-panel spatial resolution were measured as 1.9 and 4.4 mm. Compared with previous organ-dedicated PET systems, the sub-scanner shows good in-panel spatial resolution.
Acknowledgments
The authors would like to thank Dr. Negar Omidvari for the helpful discussions. This work is partially supported by the Cancer Scholars for Translational and Applied Research Program (C*STAR) and the NIH/NIBIB grant R01EB028091.
References
- Baghaei H, Wong WH, Uribe J, Li H, Zhang N and Wang Y 2000. IEEE Trans. Nucl. Sci 47 1080–4 [Google Scholar]
- Bailey D, Karp J and Surti S 2003. Physics and instrumentation in PET Positron Emission Tomography: Basic Science and Clinical ed Valk PE et al.
- Bauer CE, Brefczynski-Lewis J, Marano G, Mandich MB, Stolin A, Martone P, Lewis JW, Jaliparthi G, Raylman RR and Majewski S 2016. Brain and behavior 6 9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Berg E, Roncali E, Kapusta M, Du J and Cherry S 2016. Med. Phys 43 939–50 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Burr KC, Ivan A, Castleberry DE, LeBlanc JW, Shah KS and Farrell R 2004. IEEE Trans. Nucl. Sci 51 1791–8 [Google Scholar]
- Cui J, Pratx G, Meng B and Levin CS 2013. IEEE Trans. Med. Imaging 32 957–67 [DOI] [PubMed] [Google Scholar]
- Doshi NK, Shao Y, Silverman RW and Cherry SR 2000. Med. Phys 27 1535–43 [DOI] [PubMed] [Google Scholar]
- Du J, Bai X and Cherry SR 2018. Phys. Med. Biol 63 245019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Freifelder R, Cardi C, Grigoras I, Saffer J and Karp J 2001. 2001 IEEE Nuclear Science Symposium Conf. Record (Cat. No. 01CH37310) (IEEE) vol 3, pp 1241–5 [Google Scholar]
- Godinez F, Chaudhari AJ, Yang Y, Farrell R and Badawi RD 2012. Phys. Med. Biol 57 3435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- González AJ, Sánchez F and Benlloch JM 2018. IEEE Transactions on Radiation and Plasma Medical Sciences 2 388–403 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herbert DJ, Saveliev V, Belcari N, D’Ascenzo N, Del Guerra A and Golovin A 2006. IEEE Trans. Nucl. Sci 53 389–94 [Google Scholar]
- Karimian A, Thompson C, Sarkar S, Raisali G, Pani R, Davilu H and Sardari D 2005. Nucl. Instrum. Methods Phys. Res., Sect. A 545 427–35 [Google Scholar]
- Kolb A, Parl C, Mantlik F, Liu C, Lorenz E, Renker D and Pichler B 2014. Med. Phys 41 081916 (8 Part 1) [DOI] [PubMed] [Google Scholar]
- Kuang Z et al. 2018. Med. Phys 45 613–21 [DOI] [PubMed] [Google Scholar]
- Kuang Z et al. 2019. Nucl. Instrum. Methods Phys. Res., Sect. A 917 1–8 [Google Scholar]
- Kuang Z. et al. Phys. Med. Biol. 2017;62:5945. doi: 10.1088/1361-6560/aa71f3. [DOI] [PubMed] [Google Scholar]
- Li M and Abbaszadeh S 2019. Phys. Med. Biol 64 175008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ling T, Lewellen T and Miyaoka R 2007. Phys. Med. Biol 52 2213. [DOI] [PubMed] [Google Scholar]
- Maas MC, Schaart DR, van der Laan DJ, Bruyndonckx P, Lemaître C, Beekman FJ and van Eijk CW 2009. Phys. Med. Biol 54 1893. [DOI] [PubMed] [Google Scholar]
- MacDonald L, Edwards J, Lewellen T, Haseley D, Rogers J and Kinahan P 2009. J. Nucl. Med 50 1666–75 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miyaoka R, Lewellen T, Yu H and McDaniel D 1998. IEEE Trans. Nucl. Sci 45 1069–73 [Google Scholar]
- Moses W and Derenzo SE 1994. IEEE Trans. Nucl. Sci 41 1441–5 [Google Scholar]
- Murthy K, Aznar M, Thompson CJ, Loutfi A, Lisbona R and Gagnon JH 2000. J. Nucl. Med 41 1851–8 [PubMed] [Google Scholar]
- Nadig V, Schug D, Weissler B and Schulz V 2019. Evaluation Of The PETsys TOFPET2 ASIC In Multi-Channel Coincidence Experiments arXiv:1911.08156 [DOI] [PMC free article] [PubMed]
- Peng H 2015. Nucl. Instrum. Methods Phys. Res., Sect. A 779 39–46 [Google Scholar]
- Pizzichemi M, Stringhini G, Niknejad T, Liu Z, Lecoq P, Tavernier S, Varela J, Paganoni M and Auffray E 2016. Phys. Med. Biol 61 4679. [DOI] [PubMed] [Google Scholar]
- Raylman RR 2009. IEEE Trans. Nucl. Sci 56 620–4 [Google Scholar]
- Ren S, Yang Y and Cherry SR 2014. Med. Phys 41 072503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schaart DR, van Dam HT, Seifert S, Vinke R, Dendooven P, Löhner H and Beekman FJ 2009. Phys. Med. Biol 54 3501. [DOI] [PubMed] [Google Scholar]
- Shao Y, Meadors K, Silverman R, Farrell R, Cirignano L, Grazioso R, Shah K and Cherry SR 2002. IEEE Trans. Nucl. Sci 49 649–54 [Google Scholar]
- Shao Y, Sun X, Lan KA, Bircher C, Lou K and Deng Z 2014. Phys. Med. Biol 59 1223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tashima H et al. 2015. 2015 IEEE Nuclear Science Symposium and Medical Imaging Conf. (NSS/MIC) (IEEE) pp 1–3
- Watanabe M, Shimizu K, Omura T, Takahashi M, Kosugi T, Yoshikawa E, Sato N, Okada H and Yamashita T 2002. IEEE Trans. Nucl. Sci 49 634–9 [Google Scholar]
- Wienhard K et al. 2002. IEEE Trans. Nucl. Sci 49 104–10 [Google Scholar]
- Yamamoto S, Honda M, Oohashi T, Shimizu K and Senda M 2011. IEEE Trans. Nucl. Sci 58 668–73 [Google Scholar]
- Yamaya T, Hagiwara N, Obi T, Tsuda T, Kitamura K, Hasegawa T, Haneishi H, Inadama N, Yoshida E and Murayama H 2006. IEEE Trans. Nucl. Sci 53 1123–8 [Google Scholar]
- Yamaya T, Mitsuhashi T, Matsumoto T, Inadama N, Nishikido F, Yoshida E, Murayama H, Kawai H, Suga M and Watanabe M 2011. Phys. Med. Biol 56 6793. [DOI] [PubMed] [Google Scholar]
- Yang Y et al. 2016. J. Nucl. Med 57 1130–5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yang Y, Dokhale PA, Silverman RW, Shah KS, McClish MA, Farrell R, Entine G and Cherry SR 2006. Phys. Med. Biol 51 2131. [DOI] [PubMed] [Google Scholar]
- Yang Y, Qi J, Wu Y, James S, Farrell R, Dokhale PA, Shah KS and Cherry SR 2008. Phys. Med. Biol 54 433. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yang Y, Wu Y, Qi J, James SS, Du H, Dokhale PA, Shah KS, Farrell R and Cherry SR 2008. J. Nucl. Med 49 1132–40 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yeom J, Vinke R and Levin C 2014. Med. Phys 41 122501. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yoshida E, Kobayashi A, Yamaya T, Watanabe M, Nishikido F, Kitamura K, Hasegawa T, Fukushi M and Murayama H 2006. 2006 IEEE Nuclear Science Symposium Conf. Record (IEEE) vol 4, pp 2532–6 [Google Scholar]
- Zhang X, Siwei X, Jingwu Y, Fenghua W, Jianfeng X, Qiu H and Qiyu P 2019. Med. Phys 46 3385–9831107969 [Google Scholar]
- Zorraquino C, Bugalho R, Rolo M, Silva JC, Veckalns V, Silva R, Ortigão C, Neves J, Tavernier S and Varela J 2013. 2013 IEEE Nuclear Science Symposium and Medical Imaging Conf. (2013 NSS/MIC) (IEEE) pp 1–4 [Google Scholar]