Abstract
Today, engineered nanomaterials are frequently used. Nanosized titanium dioxide (TiO2) has been extensively used for many years and graphene is one type of emerging nanomaterial. Occupational airborne exposures to engineered nanomaterials are important to ensure safe workplaces and to extend the information needed for complete risk assessments. The main aim of this study was to characterize workplace emissions and exposure of graphene nanoplatelets, graphene oxide, TiO2 nanofibers (NFs) and nanoparticles (NPs) during down-stream industrial handling. Surface contaminations were also investigated to assess the potential for secondary inhalation exposures. In addition, a range of different sampling and aerosol monitoring methods were used and evaluated. The results showed that powder handling, regardless of handling graphene nanoplatelets, graphene oxide, TiO2 NFs, or NPs, contributes to the highest particle emissions and exposures. However, the exposure levels were below suggested occupational exposure limits. It was also shown that a range of different methods can be used to selectively detect and quantify nanomaterials both in the air and as surface contaminations. However, to be able to make an accurate determination of which nanomaterial that has been emitted a combination of different methods, both offline and online, must be used.
Keywords: Occupational exposure, Electron microscopy, Thermal-optical carbon analysis, Direct-reading instruments, PIXE, Aerosol
Introduction
Due to their novel and valuable properties compared with bulk materials, the use of engineered nanomaterials is increasing. Graphene, a 2D carbon nanomaterial, has a robust, but also flexible, structure rendering it useful in a variety of applications [1]. It can exist in different structures such as graphene, graphene oxide, and graphene nanoplatelets [2, 3]. Its electrical and thermal properties makes graphene useful in for example transistors [2, 4] and chemical sensors [5], and the optical properties can be used in biological sensors [6]. Another application of graphene is as metal surface coatings to inhibit corrosion [7, 8] and to reduce wear and friction on sliding metal surfaces [9, 10].
Multiple reviews [11–15] have generally concluded that graphene toxicity depends on the physiochemical properties of the nanomaterial. However, one of the most widely recognized mechanism for graphene-nanomaterial-induced toxicity in living systems is the induction of oxidative stress and production of reactive oxygen species (ROS) [15]. In a more recent review by Fadeel et al. [16], the authors highlighted the need for standardized graphene characterizations, and of robust and validated toxicological assays in order to advance the field of graphene toxicity.
Studies of occupational exposure to nanomaterials are needed to make complete risk assessments. Basinas et al. [17] showed that many exposure assessments have been done for carbon nanotubes (CNTs), carbon nanofibers (CNFs), and titanium dioxide nanoparticles (TiO2 NPs), but far less for other engineered nanomaterials, such as graphene. Sanchez et al. [12] stated that there is a need for measurements of airborne graphene exposure levels at both research laboratories and full-scale manufacturing facilities. This was further pointed out by Arvidsson et al. [13], who requested that workplace emissions and exposures of graphene should be investigated. Only a few emission and exposure measurements during production [18–21] and handling [18] of graphene nanomaterials have since then been conducted. These studies showed low levels of exposures to graphene. However, according to Lee et al. [18], monitoring of other work tasks including down-stream graphene handling processes is needed for a full understanding of the exposure situation. To our knowledge, no studies of emission and exposure measurements during down-stream handling processes of graphene nanomaterials, such as manufacturing of ink and surface coatings containing graphene, have previously been conducted.
Another common nanomaterial is TiO2. It is used in paints and sunscreens [22, 23] as well as in transistors [24], biosensors [25], cancer treatment [26], and different surface coatings [27, 28]. Nanosized TiO2 is commonly found as spherical NPs, but can also be produced in other shapes including NFs [29] and nanowires [30]. The toxicity of nanosized TiO2, especially spherical ones, has been studied to a greater extent than graphene. As with graphene, the different physiochemical properties of TiO2 NPs have a strong influence on the toxicity [31]. Generally though, only moderate effects have been observed, including pulmonary inflammation [32–34] and pathological neural changes [35] after inhalation/instillation in rodents. Induction of DNA damage has also been observed in lung cell studies [36, 37]. A few studies have been carried out on exposed workers. Pelclova et al. [38] showed for example that the leukotriene levels in exhaled breath condensate were elevated in workers exposed to TiO2 NPs. The toxicity of TiO2 NFs has not been as thoroughly investigated. Hurbánková et al. [39] showed the development of serious lung inflammatory and cytotoxic processes after intratracheal instillation of TiO2 NFs in rats and Medina-Reyes et al. [40] observed cell cytotoxicity and genomic instability of TiO2 NF exposure on alveolar epithelial cells. In addition, Allegri et al. [41] performed a comparative exposure study on alveolar epithelial cells showing that TiO2 NFs were more toxic than TiO2 NPs.
Occupational airborne exposure to TiO2 NPs has been extensively studied. Production of TiO2 NPs, bagging and handling of the NP dry powder, as well as incorporation of the NPs into other products has been investigated. Different handling tasks, both with NP dry powder and NP containing liquid, have been shown to constitute an occupational exposure risk [42–45]. The review by Debia et al. [46] further strengthen this conclusion. However, studies of TiO2 NF exposure in occupational settings have, to our knowledge, not yet been conducted.
The Organisation for Economic Cooperation and Development (OECD) have suggested a harmonized three-tiered approach for nanomaterial emission and exposure assessments [47] and recommendations for measurement strategies and instrument use [48].
This study aims to generate new knowledge about emissions and exposures of nanomaterials not extensively studied previously. Emissions and exposures, with a focus on different graphene nanomaterials (both nanoplatelets and oxide) and different TiO2 nanomaterials (both NFs and NPs), were characterized with a multi-metric approach during down-stream industrial handling. Emission and exposures of carbon black (CB) and copper (Cu) were also measured in a few cases. Different sampling and aerosol monitoring methods were evaluated to be able to recommend methods to be used specifically for these nanomaterials to complement the different tiers and measurement strategies described by OECD. An additional aim was to assess the potential of secondary inhalation exposure, caused by resuspension of particles deposited on surfaces.
Methods
Facilities
Measurements were performed 2016 and 2017 at two different workplaces, with 20 and 15 employees respectively, hereafter “Study A” and “Study B”. As recommended by the OECD [47], initial contextual information was gathered (tier 1) and basic exposure assessments (tier 2) were performed at both workplaces prior to conducting the expert exposure assessment (tier 3) described herein.
During Study A, different nanomaterials including graphene nanoplatelets, spherical TiO2 NPs, CB, and Cu were handled. The nanomaterials were used in ink formulations for printing electronics, sensors, and labels. Measurements were conducted in a chemistry laboratory and a printing laboratory. In Study B, the nanomaterials handled included graphene oxide and TiO2 NFs for use in friction and wear reducing surface coatings. Measurements were conducted in a chemistry laboratory and a test laboratory. Both workplaces were equipped with general ventilation and process ventilation systems such as fume hoods. The amount of TiO2 handled during the two studies differed by three orders of magnitude (from about 5 kg per day during Study A to about 5 g per day during Study B). The amount of graphene nanomaterial handled was a few grams per day during both Study A and B. Similar processes were investigated during both studies.
Work tasks
Different work tasks were performed during the two studies and a thoroughly written logbook documented the specific activities carried out. Table 1 shows these work tasks with detailed descriptions. In Study A, three workers performed the work tasks and in Study B only one.
Table 1.
Work task number | Work task | Work task description | Location/ Engineering controls | PPE |
---|---|---|---|---|
Study A | ||||
A1 | Preparation of graphene nanoplatelets ink (weighing, mixing). Note that this is not the same ink printed with in A3! | Weighing of graphene nanoplatelet powder, addition of liquid and mixing. | Chemistry laboratory/Fume hood | Half-face respirator (A + P3), lab coat, protective gloves of nitrile, goggles |
A2—day 1 | Preparation of titanium dioxide nanoparticle ink (weighing, mixing) | Weighing of titanium dioxide nanoparticle powder, weighting of liquid. Addition of powder by sieving into the liquid and mixing. | ||
A2—day 2 | Preparation of titanium dioxide nanoparticle ink (weighing, mixing) | |||
A3 | Screen printing with graphene ink (HDPlas Graphene Ink), 10 sheets | Transfer of ink to the screen printer, start of printing process and a number of sheets was printed. The printed sheets were placed in the drying oven. After the printing process the ink was removed and the screen printer and its equipment was cleaned. | Printing laboratory/process ventilation | Clean room lab coat, hair net, protective gloves of nitrile |
A4 | Screen printing with carbon black ink (C740), 10 sheets | |||
A5 | Screen printing with carbon black ink (CXT0641), 10 sheets | |||
A6—day 1 | Screen printing with carbon black ink (C7102), 20 sheets | |||
A6—day 2 | Screen printing with carbon black ink (C7102), 22 sheets | |||
A7 | Screen printing with copper ink (CP-PLS-010715-R1A), 10 sheets | |||
Study B | ||||
B1 | Preparation of graphene oxide coating (weighing, mixing) | Weighing of graphene powder, addition of liquid to the powder and mixing. Transferring of the liquid to another container followed by addition of glass beads and more mixing. Finally, 2 steps of filtration of the liquid to remove the glass beads. | Chemistry laboratory/Parts in open fume hood (weighing, mixing), parts on open bench (filtration steps) | Half-face respirator (A + P3), lab coat, protective gloves of nitrile, goggles |
B2 | Spraying and curing graphene oxide coating, and cleaning/washing up after | Transfer of the liquid coating to a paint sprayer, cleaning of metal plates, paint spraying of metal plates, placing the coated plates in a furnace. Cleaning of paint sprayer in the spray booth followed by washing up mixing equipment in the fume hood and at the sink. | Test laboratory/parts in spray booth, parts in open fume hood and sink (cleaning and washing up) | |
B3 | Abrasion test with a metal brush on graphene oxide coating | The coated plate was placed in the abrasion testing equipment. A brush, rotated with a certain speed, was placed to have contact with the plate. When the surface was abraded the test was ended. | Test laboratory/none | Laboratory coat, protective gloves of nitrile, goggles |
B4 | Abrasion test with a nylon brush on graphene oxide coating | |||
B5 | Preparation of titanium dioxide nanofiber coating (weighing, mixing) | Weighing of titanium dioxide nanofiber powder, addition of liquid to the powder and mixing. Transferring of the liquid to another container followed by addition of glass beads and more mixing. Finally, 2 steps of filtration of the liquid to remove the glass beads. | Chemistry laboratory/parts in open fume hood (weighing, mixing), parts on open bench (filtration steps) | Half-face respirator (A + P3), lab coat, protective gloves of nitrile, goggles |
B6 | Spraying and curing titanium dioxide nanofiber coating, and cleaning/washing up after | Transfer of the liquid coating to a paint sprayer, cleaning of metal plates, paint spraying of metal plates, placing the coated plates in a furnace. Cleaning of paint sprayer in the spray booth followed by washing up mixing equipment in the fume hood and at the sink. | Test laboratory/parts in spray booth, parts in open fume hood and sink (cleaning and washing up) | |
B7 | Abrasion test with a metal brush on titanium dioxide nanofiber coating | The coated plate was placed in the abrasion testing equipment. A brush, rotated with a certain speed, was placed to have contact with the plate. When the surface was abraded the test was ended. | Test laboratory/None | Laboratory coat, protective gloves of nitrile, goggles |
B8 | Abrasion test with a nylon brush on titanium dioxide nanofiber coating |
Engineering controls and personal protective equipment
Different types of exposure control techniques and enclosures, as well as different types of personal protection equipment (PPE), were used during the different work tasks (Table 1).
Sampling strategy
Time resolved and filter based measurements of airborne particles were conducted in four different spatial zones, a methodology described in detail in Isaxon et al. [49]. The measurement zones included: (1) emission zone (EZ)—no more than a few centimeters from a potential source, (2) personal breathing zone (PBZ)—within a radius of 30 cm from a worker’s nose and mouth, (3) background zone (BZ)—at least 2–3 m away from any potential particle source, and (4) supply air (SA)—in the inflowing air from the general ventilation system. Figure 1 shows the schematics of the facilities in Study A and B including where the different work tasks were performed and where the different measurement zones were located.
Air sampling methods and analyses
Filter sampling for elemental composition
Time-integrated emission zone, personal breathing zone, and background zone samples were collected by open-face sampling on 25-mm filters mounted in filter cassettes. Quartz filters (SKC Inc., USA) were used to collect samples for elemental carbon (EC) analysis. Filters made of polycarbonate (SKC Inc., USA) were used for analysis of titanium (Ti), and of cellulose (pore size 0.45 µm, SKC Inc., USA) for elemental analysis of Cu. Pumps (Escort ELF, MSA, USA) with a flow rate of 2.3 l/min for the polycarbonate filters and 3.0 l/min for the quartz and cellulose filters—checked before and after sampling—were used for all sample collection.
Quantification of EC of the graphene nanomaterials was conducted according to the NIOSH NMAM 5040 protocol for thermal-optical analysis (DRI Model 2001 OC/EC Carbon Analyzer, Atmoslytic Inc., USA) [50]. Temperature steps for the EC fraction were: 680 °C (EC1), 750 °C (EC2), and 900 °C (EC3). The method was modified with an extended oxidation time, 150 s instead of 30 s, at the highest temperature, 900 °C, in order to achieve complete oxidation of all carbonaceous nanomaterials [51]. The limit of detection (LOD) for EC was determined to be 0.06 µg C/cm2 (corresponding to 4-h sampling with a carbon airborne concentration of 0.5 µg/m3).
Quantification of Ti was conducted by Particle-Induced X-ray Emission (PIXE) analysis [52]. In PIXE, a 2.55 MeV proton beam is focused on the filter specimen. This renders the atoms in a state of high excitation, which causes inner shell vacancies. The characteristic X-ray emission lines are caused by the quickly occurring transition to a state of lower energy. When protons are used, the cross section for the creation of an inner shell vacancy is very high, thereby the sensitivity is very high; for instance, the LOD for Ti was <6 ng/cm2.
The elemental quantification of Cu was performed by digestion with 1 ml concentrated nitric acid (Nitric Acid, Trace metal grade, Fisher Chemicals) in an oven (60 °C) for 16 h, followed by dilution to 10 ml with Milli-Q water to a stock solution. Analysis was performed by inductively coupled plasma-mass spectrometry (ICP-MS, iCAP Q, Thermo Scientific, Germany). The LOD for Cu was three times the standard deviation of blank filters i.e. <0.01 µg/sample. All results were blank filter corrected.
Filter sampling for SEM analysis
To be able to morphologically characterize the workplace air, regarding engineered NPs and NFs and their aggregates and agglomerates, during down-stream handling processes, total dust fraction on filters according to Nilsson et al. [53] and Vaquero et al. [21] were collected. Time-integrated samples were collected at 2.3 l/min (sampling pump Escort ELF, MSA, USA) by open-face sampling on 25-mm polycarbonate filters (pore size 0.4 µm, SKC Inc., USA). The filters were analyzed by Scanning Electron Microscopy (SEM) using a Hitachi SU8010 Cold Field Emission SEM (Hitachi, Japan) with an acceleration voltage of 10 kV. A sputtering tool (Q150T ES, Quorum, UK) was used to coat the sample with 10 nm of platinum:palladium (Pt:Pd, 80:20). A minimum of 10 random 1.25 × 10−5 cm2 areas of the filter were imaged and used for quantifying the number of particles. During Study A, 17 filter samples were collected and the highest LOD was determined to be 0.47 cm−3. During Study B, 13 filter samples were collected and the highest LOD was determined to be 1.91 cm−3. In addition to the areas imaged for particle quantification, a larger part of the surface was investigated with lower resolution in order to identify any engineered nanomaterial.
Direct reading instruments
Several different direct-reading time resolved instruments were used in the four different measurement zones (Table 2). The aerodynamic particle diameter size distribution in the range 0.5–20 µm was obtained by two Aerodynamic Particle Sizers (APS, model 3321, TSI Inc., USA) with a time resolution of 5 s. Two condensation particle counters (CPC, model 3775 and 3010, respectively, TSI Inc., USA) were used to measure the total number concentration of particles > 0.007 µm, with a time resolution of 1 s. One APS and one CPC measured in the emission zone, while the second APS and CPC were measuring in the background zone. An aethalometer (model AE33, Magee Scientific, USA) was used to measure the black carbon (BC) mass concentration (as a proxy for EC) in the background zone, with a time resolution of 1 min. During Study A, the measurements in the emission zone were supplemented with a fast aerosol mobility size spectrometer (DMS Model 500 MkII, Cambustion, UK) and a DustTrak (model DRX 8533, TSI Inc., USA). The DMS measured the particle number size distribution in the size range 0.005–1 µm, with a time resolution of 1 s. The DustTrak measured the particle mass concentration in four size fractions: PM1, PM2.5, respirable and PM10, with a time resolution of 1 s. During Study B, a portable aethalometer (model AE51, AethLabs, USA) was used to measure the BC mass concentration in the emission zone, with a time resolution of 10 s.
Table 2.
Personal breathing zone | Emission zone | Background zone | Supply air | |
---|---|---|---|---|
Study A | Partector, portable aethalometer | APS, CPC, DMS, DustTrak | APS, CPC, Aethalometer | P-Trak, DustTrak |
Study B | Partector, portable aethalometer | APS, CPC, portable aethalometer | APS, CPC, Aethalometer | P-Trak, DustTrak |
During both studies, two instruments were carried by the workers to measure in the personal breathing zone. A similar second portable aethalometer to measure the BC mass concentration, with a time resolution of 10 s, and an aerosol dosimeter (Partector, Naneos, Switzerland) to measure the lung deposited surface area (LDSA) concentration, with a time resolution of 1 s.
For particle measurements in the supply air, a P-Trak (model 8525, TSI Inc., USA) and a DustTrak (model DRX 8534, TSI Inc., USA) were used to assess the particle number concentration (0.02–1 µm) and the particle mass concentration (0.1–15 µm), respectively, with a time resolution of 1 s.
Surface sampling method and analysis
During Study B, tape samples from different surfaces were collected according to a tape stripping method described by Hedmer et al. [54]. Tape samples were collected at the end of the workdays from surfaces in the near-field zone of the exposure source (<1 m). Surface contaminations in the far-field zone (>1 m), including two offices and a conference room, were also studied (Table 4). Two field blank tape samples were also obtained. The tape samples were prepared and analyzed with the same SEM method as the air filter samples.
Table 4.
Sampling location | Room | Related to work task number | Surface characteristics | SEM analysis | ||
---|---|---|---|---|---|---|
Material | Assessed indication of roughness | Detection of TiO2 nanofiber (Yes/No) | Detection of graphene oxide particles (Yes/No) | |||
Work areas | ||||||
Work areas in fume hood | Chemistry laboratory | B1, B5 | Stainless steel | Smooth | No | No |
Cover on work area | B1, B5 | Cardboard | Smooth | No | No | |
Sink | B1–2, B5–6 | Stainless steel | Smooth | Yes | Yes | |
Desk next to abrasion testing machine | Test laboratory | B3–4, B7–8 | Laminate | Smooth | No | Yes |
Floors | ||||||
Floor in front of fume hood | Chemistry laboratory | B1, B5 | Epoxy coated concrete | Smooth | No | No |
Floor in front of spray booth | Test laboratory | B2, B6 | Epoxy coated concrete | Smooth | No | No |
Floor | Conference room | – | Parquet | Smooth | No | No |
Floor | Office 1 | – | Laminate | Smooth | No | No |
Door sill | Office 2 | – | Wood | Smooth | No | No |
Other surfaces | ||||||
Balance in fume hood | Chemistry laboratory | B1, B5 | Plastic | Smooth | No | Yes |
Cover under the brush in the abrasion testing machine | Test laboratory | B3–4, B7–8 | Metal | Smooth | Yes | No |
Plate with graphene based coating, sampled after abrasion test with metal brush | B3 | Metal with graphene based coating | Smooth | No | a | |
Plate with graphene based coating, sampled after abrasion test with nylon brush | B4 | Metal with graphene based coating | Smooth | No | a | |
Plate with TiO2 nanofiber based coating, sampled after abrasion test with metal brush | B7 | Metal with TiO2 nanofiber based coating | Smooth | No | No | |
Plate with TiO2 nanofiber based coating, sampled after abrasion test with nylon brush | B8 | Metal with TiO2 nanofiber based coating | Smooth | Yes | No |
–Far-field surfaces.
aNot possible to determine which type of particles that are detected .
Results
After the completion of tier 1 (contextual information gathering) and tier 2 (basic exposure assessment), we concluded that tier 3 expert exposure assessments were needed at both companies. The tier 3 measurements are the ones described herein.
Filter sampling
Table 3 shows the results from the filter samples collected during work tasks using CB, graphene nanomaterials and TiO2 NPs and NFs. Note that some of the filters have been sampled for more than one work task (according to Table 1). During the one work task where Cu was used (A7), no concentration of Cu was detected in either the personal breathing zone or the emission zone.
Table 3.
Work task number | Nanomaterials handled during the work task | Sampling time (min) | SEM analysis | Elemental carbon (µg/m3) | Metal conc. Ti (µg/m3) | |
---|---|---|---|---|---|---|
Detection of nanomaterial (Yes/No)/Type | Number conc. (cm−3) | |||||
Study A (2016) | ||||||
Personal breathing zone | ||||||
A1 | Graphene nanoplatelets | 42 | a | a | <LOD | – |
A2—day 1 | TiO2 NPs | 83 | Yes/TiO2 NPs | 25 | – | 7.5 |
A2—day 2 | TiO2 NPs | 45 | Yes/TiO2 NPs | 1.8 | – | 2.1 |
A3-6—day 1 | CB, graphene | 106 | Yes/CB | a | 5.6 | – |
A6—day 2 | CB | 39 | – | – | <LOD | – |
Emission zone | ||||||
A1 | Graphene nanoplatelets | 43 | Yes/graphene nanoplatelets | a | 26 | – |
A2—day 1 | TiO2 NPs | 83 | Yes/TiO2 NPs | 25 | – | 70 |
A2—day 2 | TiO2 NPs | 57 | Yes/TiO2 NPs | 7.0 | – | 28 |
A3 | Graphene | 13 | a | a | <LOD | – |
A4 | CB | 15 | Yes/CB | a | <LOD | – |
A5 | CB | 15 | Yes/CB | a | 98 | – |
A6—day 1 | CB | 31 | Yes/CB | a | 8.2 | – |
A6—day 2 | CB | 37 | – | – | <LOD | – |
Study B (2017) | ||||||
Personal breathing zone | ||||||
B1–4 | Graphene oxide | 172 | No | <LOD | 1.3 | – |
B5–6 | TiO2 NFs | 100 | Yes/TiO2 NFs | a | – | <LOD |
B7-8 | TiO2 NFs | 60 | No | <LOD | – | <LOD |
Emission zone | ||||||
B1–2 | Graphene oxide | 207 | Yes/graphene oxide | a | 1.9 | – |
B3 | Graphene oxide | 32 | No | <LOD | <LOD | – |
B4 | Graphene oxide | 111 | a | a | <LOD | – |
B5 | TiO2 NFs | 54 | Yes/TiO2 NFs | a | – | 2.2 |
B6 | TiO2 NFs | 129 | No | <LOD | – | <LOD |
B7 | TiO2 NFs | 23 | No | <LOD | – | <LOD |
B8 | TiO2 NFs | 123 | No | <LOD | – | 0.2 |
Background zone | ||||||
B1–4 | Graphene oxide | 480 | No | <LOD | 0.2 | |
B5–6 | TiO2 NFs | 202 | No | <LOD | – | <LOD |
B7–8 | TiO2 NFs | 173 | No | <LOD | – | <LOD |
–Not sampled.
aNot possible to determine.
Graphene detection and quantification
Table 3 shows that graphene nanomaterials were detected (Fig. 2) on the emission zone filters at both workplaces. Figure 2a shows the raw graphene nanoplatelet material used during Study A, and Fig. 2b shows the sampled material found in the emission zone during handling of graphene nanoplatelet powder (work task A1). Figure 2c shows the raw graphene oxide material used during Study B, and Fig. 2d shows the sampled material found in the emission zone during handling of graphene oxide powder (work task B1). The amounts of graphene nanomaterial in the emission zones (26 and 1.9 µg/m3 for Study A and B, respectively) were quantified as EC with thermal-optical analysis. During Study B, EC (most likely from graphene oxide) was also quantified in the personal breathing zone (1.3 µg/m3). Two of the direct reading instruments (APS and portable aethalometer) showed clear particle concentration peaks in the emission zone during the 1-min weighing event of graphene oxide powder (performed twice) during Study B (work task B1), see Fig. 3. The coarse particle number size distribution (count median aerodynamic diameter (CMD) of ~2 µm) of the initial graphene oxide particle concentration peak (seen in Fig. 3) is shown in Fig. 4. From the SEM images (Fig. 2c, d), the size of the graphene oxide particles is estimated to be about 10–20 µm.
Titanium dioxide nanofiber detection and quantification
As shown in Table 3, TiO2 NFs, handled at the workplace during Study B, could be detected (Fig. 5) both in the emission zone and in the personal breathing zone during powder handling (work task B5). Figure 5a shows the raw TiO2 NF material used during Study B, and Fig. 5b, c shows examples of sampled material found in the emission zone and personal breathing zone, respectively. The sampled fiber length was assessed to be 1–15 µm. TiO2 NF concentration in the emission zone was 2.2 µg/m3, quantified as Ti with PIXE. The APS in the emission zone showed particle concentration peaks during the two-minute weighing of the TiO2 NF powder, see Fig. 6. The coarse particle number size distribution of the concentration peaks seen in Fig. 6 revealed a CMD of ~0.7 µm (not shown).
Titanium dioxide nanoparticle detection and quantification
Spherical TiO2 NP agglomerates, handled at the workplace during Study A, were detected (Fig. 7) on filters from both the emission zone and the personal breathing zone during handling of the powder on two separate days (work task A2—day 1 and A2—day 2). The concentration reached 70 and 28 µg/m3 in the emission zone for the first and second day, respectively (quantified with the PIXE), and 7.5 and 2.1 µg/m3 in the personal breathing zone the first and second day, respectively. Particle concentration peaks during weighing and mixing of the TiO2 NP powder were detected by APS and DustTrak in the emission zone during both the first and second day, see Fig. 8. Similar to the PIXE results, the particle concentrations measured with the APS and DustTrak were lower during the second day. The Partector in the personal breathing zone showed a particle exposure reaching a peak lung deposited surface area concentration of 92 µm2/cm3 during the first day and of only 9 µm2/cm3 the second day (not shown).
Surface contamination
Workplace surfaces in the chemical laboratory, test laboratory, offices, and conference room were tape sampled in Study B. In total, 15 tape samples were collected, and TiO2 NFs were detected in 20% (n = 3) and graphene oxide in 20% (n = 3) (Table 4). Surface contamination of both TiO2 NFs and graphene oxide was found on only one surface, at the sink in the chemical laboratory which was related to work task B1–2 and B5–6. A SEM image of the nanomaterial-related surface contamination can be seen in Fig. 9. The length of the TiO2 NFs was assessed to be 10–50 µm. TiO2 NFs were also detected on the plate with TiO2 NF based coating after the abrasion test with the nylon brush. No surface contamination of nanomaterial was found outside the chemical and test laboratories.
Discussion
This study aimed to characterize workplace emissions and exposure of graphene nanoplatelets, graphene oxide, TiO2 NFs, and NPs, as well as CB and Cu during industrial handling. The results showed that powder handling, regardless of nanomaterial, generates the highest (among the investigated processes) particle emissions and exposures. It was shown that a range of different methods successfully can be used to selectively detect and quantify nanomaterials both in the air and on surfaces, and that, to be able to make an accurate determination of which nanomaterial that has been emitted when, a combination of different methods, both offline and online, must be used.
Measurements of airborne graphene nanomaterials
The results show that graphene can be detected and quantified with several different methods. With the offline method consisting of filter sampling and SEM analysis we were, by comparing with the original nanomaterial, able to identify both graphene nanoplatelets used during Study A, and graphene oxide used during Study B (Fig. 2). Lee et al. [18] and Vaquero et al. [21] have also reported similar graphene-like structures in workplace air samples. According to OECD [48], both SEM and TEM are frequently used for determination of nanomaterial present in the workplace air.
In Fig. 4, the peak particle number size distribution during handling of graphene oxide powder (seen in Fig. 3) showed a CMD of ~2 µm. In the SEM images of both collected graphene nanoplatelets and graphene oxide, larger particles were found (10–20 µm). This is not contradictive since 2D materials such as graphene may have an aerodynamic diameter much smaller than its geometrical dimensions [55], a fact that indicates its ability to reach the alveolar region of the lung when inhaled.
The EC concentration (assumed to originate from graphene) was quantified with thermal-optical analysis (OC/EC) of filter samples. EC analysis is much less labor intensive than the SEM/TEM method. In previous workplace studies of graphene, Lee et al. [18] and Vaquero et al. [21] also used EC as an exposure metric for graphene, while exposure data of EC are missing in the studies by Spinazze et al. [19] and Boccuni et al. [20]. In the current study, EC concentrations collected as total dust fractions were detected both in emission zone and personal breathing zone filter samples, for example during weighing and mixing of graphene oxide. Lee et al. [18] collected respirable fractions of EC while Vaquero et al. [21] collected total dust fraction, but none of these studies detected any airborne EC concentrations during manufacturing of graphene nanomaterial.
We used on-line measurements of equivalent BC (eBC) as a proxy for EC by using a portable aethalometer. This allowed for highly time-resolved measurements and identification of exposure peaks for several work tasks. The combination of OC/EC analysis and portable aethalometer for graphene quantification has previously been used by Lee et al. [18] at a graphene nanoplatelets manufacturer. They found the clearest emission peaks in terms of eBC during graphene weighing. We have shown that these methods can be used also at down-stream handling facilities. The portable aethalometer and the APS (both time-resolved instruments) were used together with a thoroughly written logbook. By this approach we could determine that weighing and mixing the dry powder material generated the highest particle emission and exposure levels. The importance of documenting the activities to be able to match the measured concentration profiles have previously been demonstrated by e.g., Ham et al. [44], Hedmer et al. [51], and Isaxon et al. [49].
Both OC/EC analysis and eBC measurements of graphene nanomaterials is subject to cross sensitivity by other carbonaceous nanomaterials such as CB and CNTs, but also from soot particle sources, such as diesel exhaust, that may be present in workplace air. Both EC and eBC concentrations were substantially elevated in the emission zone compared with simultaneous measurements in the background zone during Study B, suggesting that cross sensitivity to background EC or eBC was low.
The eBC mass concentration was reported assuming the standard instrument settings. This may have affected the accuracy of the method as an offset of the results if the ENMs have an altered instrument response, for example by a different mass absorption cross section compared with the standard value that is based on atmospheric soot. This uncertainty can be tolerated when it comes to ENM exposures, where the aim often is to link exposure peaks with specific work tasks. In the literature, an eBC/EC ratio of 0.14 was found during simulated powder handling of CNTs [56].
Even though a combination of offline and online methods is recommended, different methods can be used individually for specific needs, depending on what type of information that is needed. For graphene, the following can be used: (1) SEM analysis for accurate identification of the nanomaterial, (2) EC analysis of filter samples for an accurate assessment of higher carbon concentrations (relatively high LOD), (3) direct reading portable aethalometer for highly time-resolved eBC assessment, and (4) direct reading APS for low-concentration (although less specific) detection of particle emissions (low LOD). This means that with access to one of these methods, a preliminary emission and exposure assessment should be possible to conduct.
Measurements of airborne titanium dioxide nanofibers and nanoparticles
NFs are another type of nanomaterials that are of concern, especially if they are long (>5 µm) and insoluble, due to their potential to cause adverse health effects, including frustrated phagocytosis and accumulation over time finally causing lung disease [57]. During Study B, TiO2 NFs were detected both in the emission zone and the personal breathing zone with the offline filter sampling method followed by SEM analysis (Fig. 5). They appear to be present in the workplace air as bundles, and some of the fibers detected by SEM during Study B were at least 5 µm and could therefore constitute a risk if inhaled. Bianchi et al. [58] have recently shown that long (around 10 µm) TiO2 NFs induce cell cytotoxicity in vitro and inflammation in vivo, while shortened (around 2 µm) TiO2 NFs seem to mitigate the toxic effects, even without macrophages present in the in vitro cultures. SEM-analysis can also be used to quantify the particle concentration in the air. However, it is a time consuming process, and during Study B it was shown to be difficult to distinguish the different fibers due to agglomeration and therefore to count them. The PIXE analysis was a valid alternative, even though it gives the concentration as a different metric (mass concentration, µg/m3). Detection and quantification can also be conducted with direct reading measurements (APS could be used to determine what specific handling process generated the particles (Fig. 6); weighing the dry powder material), but to confirm fiber emissions, time-integrated filter sampling followed by SEM is necessary.
General NP dry powder handling has previously been shown to constitute a potential source for worker exposure by e.g., Huang et al. [42] and Curwin and Bertke [43]. Lee et al. [59] found substantial total mass concentrations during powder collection of TiO2 NPs in the pigment industry (500–5000 µg/m3), however, no elemental analysis of Ti was carried out. In the current study, this was confirmed, were elevated concentrations of TiO2 NPs were found in both the emission zone and the personal breathing zone (with SEM, PIXE, APS, DustTrak, and Partector) during weighing and mixing of dry TiO2 NP powder. Interestingly, the mass concentrations of Ti (measured with PIXE) were about a factor 10 lower in the personal breathing zone than in the emission zone. This reduction factor was most likely achieved by the use of engineering controls in the emission zone (fume hood) and such a factor could be used for other assessments of personal exposure where only measurements in the emission zone can be conducted or vice versa.
Another observation was that the particle concentrations measured with PIXE, APS, DustTrak, and Partector were all lower during the second day compared with the first day of weighing and mixing of the TiO2 NP powder. As shown in Fig. 8, the procedure of mixing the powder with the liquid had been adjusted from the first to the second day. During the first day, the TiO2 NP powder were weighed first and then the liquid was added to form the printing ink. The second day, this procedure had been adjusted so that the TiO2 NP powder were directly mixed in with the liquid. This clearly shows that with an easy change in the handling procedure, the particle emissions and exposures can be lowered by as much as a factor 10 (for example seen with the Partector where the lung deposited surface area went from 92 µm2/cm3 during the first day to only 9 µm2/cm3 during the second day). The Partector has also previously been shown to be an important tool for personal exposure assessments to improve workplace monitoring [60–62].
Quantification of Ti from TiO2 nanomaterials by ICP-MS is not possible due to that TiO2 is a poorly soluble oxide. Laser-ICP-MS could be one alternative [45] and ICP-OES (optical emission spectrometer) [42] another. The current study has highlighted PIXE as a valid and reliable method for titanium quantification, which has also been shown by Relier et al. [63].
Surface contamination
Surface contamination of TiO2 NFs and graphene oxide was for the first time studied at down-stream handling processes. Surface contaminations of both TiO2 NFs and graphene oxide were found on one surface (sink) in the near-field zone in the chemical laboratory in Study B. The detected surface contamination can probably be related to washing of equipment after preparations of both types of coatings. The surface contamination of TiO2 NFs and graphene oxide could probably be resuspended into the workplace air, which if so would cause a risk of secondary inhalation exposure. The percentage of TiO2 NF surface contamination in the collected tape samples were lower compared with what was previously found in a similar study [54]. This indicates that there was no widespread nanomaterial contamination at workplace B. An interesting finding was that TiO2 NFs could be detected on a tape sample from one of the abrasion-tested metal plates with TiO2 NF based coating. During the abrasion tests, dust containing manufactured nanomaterials is generated and could be emitted to the workplace air, especially since this process was openly performed at the workplace without any engineering controls.
Proposed occupational exposure limits
Mihalache et al. [64] have reviewed the proposed occupational exposure limits (OELs) for a number of different nanomaterials. For TiO2 NPs, a range between 17 and 2000 µg/m3 have been proposed, with probably the most established one by NIOSH of 300 µg/m3 as an 8 h average airborne exposure [65]. The highest TiO2 concentration in the personal breathing zone during the current study was 7.5 µg/m3, corresponding to an 8 h average of 1.3 µg/m3, well below the suggested OEL. For TiO2 NFs, there is no specific OEL, but the BSI [66] suggested OEL for fibrous nanomaterials of 0.01 fibers/cm3 could be a suitable OEL to consult. However, in the current study it was found to be difficult to distinguish the different fibers by SEM due to agglomeration and therefore challenging to count them. The detected fibers were, furthermore, only found during the low resolution scanning-to-identify investigation and none were found during the quantification imaging, rendering the fiber number concentration below the LOD. Tsang et al. [67] demonstrated that a probabilistic approach can be used in the risk assessment of exposure scenarios involving production of TiO2 NPs. Thus, they could identify one out of seven exposure scenarios with statistically significant level of risk.
Neither for graphene nanomaterials are there any legal binding OELs yet, but e.g., Lee et al. [68] have based on data from a subchronical inhalation toxicity study calculated a recommended OEL for graphene nanomaterial to be 18 µg/m3. Also, health based guidance value for occupational inhalation exposure to graphene nanoplatelets was estimated to 212 µg/m3 by Spinazze et al. [69]. However, so far there are only few in vivo inhalation exposure studies reported in the literature. Thus, according to Pelin et al. [3] OEL for graphene cannot be determined based on the available data. The highest EC concentration in the personal breathing zone, in the current study, was 5.6 µg/m3, corresponding to an 8 h average of 1.2 µg/m3, well below the suggested OELs.
Recommendations for occupational hygienists
As a general recommendation for occupational hygienists, the portable aethalometer could be one option for a small and easy-to-use instrument for personal exposure monitoring of carbon-based nanomaterials. However, from the current study, data from the personal sampling portable aethalometer could not be used due to too much noise, most likely arising from the short sampling time. This could possibly be addressed by longer averaging times [70] as well as noise reduction treatment during the data post-processing [71]. As a stationary instrument in the emission zone, the portable aethalometer worked well and can be an alternative to bulkier, more expensive and advanced instruments. For non-carbon-based nanomaterials, a suitable personal exposure instrument option would instead be a lung deposited surface area instrument. However, when it is critical with an accurate determination of which nanomaterial that is emitted, SEM analysis must be performed as a complement to the direct reading measurements. Open-face filter sampling was used in the current study to morphologically characterize the real emission and exposure situations of all released NPs and NFs and their aggregates and agglomerates. Chemical analysis is also a good option, especially when aiming to compare the result with existing OEL, if one can ensure that no other sources of that chemical compound are present in the workplace. For carbon-based materials (such as graphene) filter sampling with following EC analysis can be used. In the current study, open-face filter sampling was used, but it could have been possible to instead use a respirable cyclone to get the respirable EC concentrations. Then, it would be possible to compare measured EC concentrations with the recommended exposure limit of respirable EC set to 1 µg/m3 as an 8 h time weighted average concentration for CNTs and CNFs [72]. For TiO2 nanomaterials, filter sampling with following PIXE analysis would be the recommended chemical analysis to perform.
Conclusions
Down-stream industrial handling of graphene nanomaterials and TiO2 NFs were, for the first time, investigated for particle emissions and exposures into the workplace air. We showed that weighing the dry powder material generated particle emissions, even though the exposure levels were low compared with proposed OELs. Surface contaminations of both TiO2 NFs and graphene were found on the sink in the chemical laboratory at workplace B. A range of different sampling and aerosol monitoring methods was used and evaluated. For a fast and reliable workplace emission and exposure assessment of graphene and TiO2 nanomaterials, a combination of different methods, both offline and online, must be used to ensure the detected emissions contain the specific nanomaterials. If a first exposure assessment should be performed (according to tier 2 of the OECD guidelines), access to only one or a few of these methods is enough. When a tier 3 exposure assessment is needed, a combination of multiple of the mentioned methods should be used together with additional direct reading instruments such as CPC and APS.
Acknowledgements
The authors would like to acknowledge the following persons: Louise Gren for helping out with the on-site workplace measurements during Study B. Linus Ludvigsson for helping out with the SEM analyses. Jan Pallon for conducting the PIXE analysis. Thomas Lundh for conducting the metal analysis with ICP-MS. The project was funded by AFA insurance (Dnr 130122) and NanoLund (p38-2013), and was carried out within the framework of NanoLund at Lund University.
NanoLund
Karin Lovén1, Sara M. Franzén2, Christina Isaxon1, Maria E. Messing2, Anders Gudmundsson1, Joakim Pagels1, Maria Hedmer4
Funding
The project was funded by AFA insurance (Dnr 130122) and NanoLund (p38-2013), and was carried out within the framework of NanoLund at Lund University.
Compliance with ethical standards
Conflict of interest
The authors declare no conflict of interest. The project sponsors have only contributed financially, and have not participated in preparing the research material, writing, reviewing, or approving the submitted paper.
Footnotes
Members of the NanoLund are listed below Acknowledgements.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Karin Lovén, Email: Karin.loven@design.lth.se.
NanoLund:
Karin Lovén, Sara M. Franzén, Christina Isaxon, Maria E. Messing, Anders Gudmundsson, Joakim Pagels, and Maria Hedmer
References
- 1.Castro Neto AH, Guinea F, Peres NMR, Novoselov KS, Geim AK. The electronic properties of graphene. Rev Mod Phys. 2009;81:109–62. doi: 10.1103/RevModPhys.81.109. [DOI] [Google Scholar]
- 2.Balandin AA. Thermal properties of graphene and nanostructured carbon materials. Nat Mater. 2011;10:569–81. doi: 10.1038/nmat3064. [DOI] [PubMed] [Google Scholar]
- 3.Pelin M, Sosa S, Prato M, Tubaro A. Occupational exposure to graphene based nanomaterials: risk assessment. Nanoscale. 2018;10:15894–903. doi: 10.1039/C8NR04950E. [DOI] [PubMed] [Google Scholar]
- 4.Chen ZH, Lin YM, Rooks MJ, Avouris P. Graphene nano-ribbon electronics. Physica E. 2007;40:228–32. doi: 10.1016/j.physe.2007.06.020. [DOI] [Google Scholar]
- 5.Schedin F, Geim AK, Morozov SV, Hill EW, Blake P, Katsnelson MI, et al. Detection of individual gas molecules adsorbed on graphene. Nat Mater. 2007;6:652–5. doi: 10.1038/nmat1967. [DOI] [PubMed] [Google Scholar]
- 6.Loh KP, Bao QL, Eda G, Chhowalla M. Graphene oxide as a chemically tunable platform for optical applications. Nat Chem. 2010;2:1015–24. doi: 10.1038/nchem.907. [DOI] [PubMed] [Google Scholar]
- 7.Rafiee J, Mi X, Gullapalli H, Thomas AV, Yavari F, Shi YF, et al. Wetting transparency of graphene. Nat Mater. 2012;11:217–22. doi: 10.1038/nmat3228. [DOI] [PubMed] [Google Scholar]
- 8.Ding R, Li WH, Wang X, Gui TJ, Li BJ, Han P, et al. A brief review of corrosion protective films and coatings based on graphene and graphene oxide. J Alloy Compd. 2018;764:1039–55. doi: 10.1016/j.jallcom.2018.06.133. [DOI] [Google Scholar]
- 9.Berman D, Erdemir A, Sumant AV. Few layer graphene to reduce wear and friction on sliding steel surfaces. Carbon. 2013;54:454–9. doi: 10.1016/j.carbon.2012.11.061. [DOI] [Google Scholar]
- 10.Mutyala KC, Wu YMA, Erdemir A, Sumant AV. Graphene—MoS2 ensembles to reduce friction and wear in DLC-Steel contacts. Carbon. 2019;146:524–7. doi: 10.1016/j.carbon.2019.02.047. [DOI] [Google Scholar]
- 11.Jastrzebska AM, Kurtycz P, Olszyna AR. Recent advances in graphene family materials toxicity investigations. J Nanopart Res. 2012;14:1320. doi: 10.1007/s11051-012-1320-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Sanchez VC, Jachak A, Hurt RH, Kane AB. Biological interactions of graphene-family nanomaterials: an interdisciplinary review. Chem Res Toxicol. 2012;25:15–34. doi: 10.1021/tx200339h. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Arvidsson R, Molander S, Sanden BA. Review of potential environmental and health risks of the nanomaterial graphene. Hum Ecol Risk Assess. 2013;19:873–87. doi: 10.1080/10807039.2012.702039. [DOI] [Google Scholar]
- 14.Hu XG, Zhou QX. Health and ecosystem risks of graphene. Chem Rev. 2013;113:3815–35. doi: 10.1021/cr300045n. [DOI] [PubMed] [Google Scholar]
- 15.Guo XQ, Mei N. Assessment of the toxic potential of graphene family nanomaterials. J Food Drug Anal. 2014;22:105–15. doi: 10.1016/j.jfda.2014.01.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Fadeel B, Bussy C, Merino S, Vazquez E, Flahaut E, Mouchet F, et al. Safety assessment of graphene-based materials: focus on human health and the environment. ACS Nano. 2018;12:10582–620. doi: 10.1021/acsnano.8b04758. [DOI] [PubMed] [Google Scholar]
- 17.Basinas I, Jimenez AS, Galea KS, van Tongeren M, Hurley F. A Systematic review of the routes and forms of exposure to engineered nanomaterials. Ann Work Expo Heal. 2018;62:639–62. doi: 10.1093/annweh/wxy048. [DOI] [PubMed] [Google Scholar]
- 18.Lee JH, Han JH, Kim JH, Kim B, Bello D, Kim JK, et al. Exposure monitoring of graphene nanoplatelets manufacturing workplaces. Inhal Toxicol. 2016;28:281–91. doi: 10.3109/08958378.2016.1163442. [DOI] [PubMed] [Google Scholar]
- 19.Spinazze A, Cattaneo A, Campagnolo D, Bollati V, Bertazzi PA, Cavallo DM. Engineered nanomaterials exposure in the production of graphene. Aerosol Sci Technol. 2016;50:812–21. doi: 10.1080/02786826.2016.1195906. [DOI] [Google Scholar]
- 20.Boccuni F, Ferrante R, Tombolini F, Lega D, Antonini A, Alvino A, et al. Workers’ exposure to nano-objects with different dimensionalities in R&D laboratories: measurement strategy and field studies. Int J Mol Sci. 2018;19:349. doi: 10.3390/ijms19020349. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Vaquero C, Wendelbo R, Egizabal A, Gutierrez-Cañas C, López de Ipiña J. Exposure to graphene in a pilot production plant. J Phys Conf Series. 2019;1323.
- 22.Khataee A, Moradkhannejhad L, Heydari V, Vahid B, Joo SW. Self-cleaning acrylic water-based white paint modified with different types of TiO2 nanoparticles. Pigm Resin Technol. 2016;45:24–9. doi: 10.1108/PRT-09-2014-0070. [DOI] [Google Scholar]
- 23.Sharma S, Sharma RK, Gaur K, Torres JFC, Loza-Rosas SA, Torres A, et al. Fueling a hot debate on the application of TiO2 nanoparticles in sunscreen. Materials. 2019;12:2317. doi: 10.3390/ma12142317. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Dastan D, Gosavi SW, Chaure NB. Studies on electrical properties of hybrid polymeric gate dielectrics for field effect transistors. Macromol Symp. 2015;347:81–6. doi: 10.1002/masy.201400042. [DOI] [Google Scholar]
- 25.Victorious A, Clifford A, Saha S, Zhitomirsky I, Soleymani L. Integrating TiO2 nanoparticles within a catecholic polymeric network enhances the photoelectrochemical response of biosensors. J Phys Chem C. 2019;123:16186–93. doi: 10.1021/acs.jpcc.9b02740. [DOI] [Google Scholar]
- 26.Ou G, Li ZW, Li DK, Cheng L, Liu Z, Wu H. Photothermal therapy by using titanium oxide nanoparticles. Nano Res. 2016;9:1236–43. doi: 10.1007/s12274-016-1019-8. [DOI] [Google Scholar]
- 27.Pi JK, Yang HC, Wan LS, Wu J, Xu ZK. Polypropylene microfiltration membranes modified with TiO2 nanoparticles for surface wettability and antifouling property. J Membr Sci. 2016;500:8–15. doi: 10.1016/j.memsci.2015.11.014. [DOI] [Google Scholar]
- 28.Tang YJ, Hu XL, Zhang XQ, Guo DL, Zhang JH, Kong FO. Chitosan/titanium dioxide nanocomposite coatings: rheological behavior and surface application to cellulosic paper. Carbohyd Polym. 2016;151:752–9. doi: 10.1016/j.carbpol.2016.06.023. [DOI] [PubMed] [Google Scholar]
- 29.Nalbandian MJ, Greenstein KE, Shuai DM, Zhang ML, Choa YH, Parkin GF, et al. Tailored synthesis of photoactive TiO2 nanofibers and Au/TiO2 nanofiber composites: structure and reactivity optimization for water treatment applications. Environ Sci Technol. 2015;49:1654–63. doi: 10.1021/es502963t. [DOI] [PubMed] [Google Scholar]
- 30.Safajou H, Khojasteh H, Salavati-Niasari M, Mortazavi-Derazkola S. Enhanced photocatalytic degradation of dyes over graphene/Pd/TiO2 nanocomposites: TiO2 nanowires versus TiO2 nanoparticles. J Colloid Interface Sci. 2017;498:423–32. doi: 10.1016/j.jcis.2017.03.078. [DOI] [PubMed] [Google Scholar]
- 31.Wang JX, Fan YB. Lung injury induced by TiO2 nanoparticles depends on their structural features: size, shape, crystal phases, and surface coating. Int J Mol Sci. 2014;15:22258–78. doi: 10.3390/ijms151222258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Grassian VH, O’Shaughnessy PT, Adamcakova-Dodd A, Pettibone JM, Thorne PS. Inhalation exposure study of titanium dioxide nanoparticles with a primary particle size of 2 to 5 nm. Environ Health Perspect. 2007;115:397–402. doi: 10.1289/ehp.9469. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Pujalte I, Dieme D, Haddad S, Serventi AM, Bouchard M. Toxicokinetics of titanium dioxide (TiO2) nanoparticles after inhalation in rats. Toxicol Lett. 2017;265:77–85. doi: 10.1016/j.toxlet.2016.11.014. [DOI] [PubMed] [Google Scholar]
- 34.Lauvas AJ, Skovmand A, Poulsen MS, Kyjovska ZO, Roursgaard M, Goericke-Pesch S, et al. Airway exposure to TiO2 nanoparticles and quartz and effects on sperm counts and testosterone levels in male mice. Reprod Toxicol. 2019;90:134–40. doi: 10.1016/j.reprotox.2019.07.023. [DOI] [PubMed] [Google Scholar]
- 35.Wang JX, Liu Y, Jiao F, Lao F, Li W, Gu YQ, et al. Time-dependent translocation and potential impairment on central nervous system by intranasally instilled TiO2 nanoparticles. Toxicology. 2008;254:82–90. doi: 10.1016/j.tox.2008.09.014. [DOI] [PubMed] [Google Scholar]
- 36.Falck GCM, Lindberg HK, Suhonen S, Vippola M, Vanhala E, Catalan J, et al. Genotoxic effects of nanosized and fine TiO2. Hum Exp Toxicol. 2009;28:339–52. doi: 10.1177/0960327109105163. [DOI] [PubMed] [Google Scholar]
- 37.Di Bucchianico S, Cappellini F, Le Bihanic F, Zhang YN, Dreij K, Karlsson HL. Genotoxicity of TiO2 nanoparticles assessed by mini-gel comet assay and micronucleus scoring with flow cytometry. Mutagenesis. 2017;32:127–37. doi: 10.1093/mutage/gew030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Pelclova D, Zdimal V, Kacer P, Fenclova Z, Vlckova S, Komarc M, et al. Leukotrienes in exhaled breath condensate and fractional exhaled nitric oxide in workers exposed to TiO2 nanoparticles. J Breath Res. 2016;10:036004. doi: 10.1088/1752-7155/10/3/036004. [DOI] [PubMed] [Google Scholar]
- 39.Hurbankova M, Cerna S, Kovacikova Z, Wimmerova S, Hraskova D, Marcisiakova J, et al. Effect of TiO2 nanofibres on selected bronchoalveolar parameters in acute and subacute phase—experimental study. Cent Eur J Publ Heal. 2013;21:165–70. doi: 10.21101/cejph.a3881. [DOI] [PubMed] [Google Scholar]
- 40.Medina-Reyes EI, Delgado-Buenrostro NL, Deciga-Alcaraz A, Freyre-Fonseca V, Flores-Flores JO, Hernandez-Pando R, et al. Titanium dioxide nanofibers induce angiogenic markers and genomic instability in lung cells leading to a highly dedifferentiated and fibrotic tumor formation in a xenograft model. Environ Sci-Nano. 2019;6:286–304. doi: 10.1039/C8EN01078A. [DOI] [Google Scholar]
- 41.Allegri M, Bianchi MG, Chiu M, Varet J, Costa AL, Ortelli S, et al. Shape-related toxicity of titanium dioxide nanofibres. PLos ONE. 2016;11:e0151365. doi: 10.1371/journal.pone.0151365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Huang CH, Tai CY, Huang CY, Tsai CJ, Chen CW, Chang CP, et al. Measurements of respirable dust and nanoparticle concentrations in a titanium dioxide pigment production factory. J Environ Sci Heal A. 2010;45:1227–33. doi: 10.1080/10934529.2010.493792. [DOI] [PubMed] [Google Scholar]
- 43.Curwin B, Bertke S. Exposure characterization of metal oxide nanoparticles in the workplace. J Occup Environ Hyg. 2011;8:580–7. doi: 10.1080/15459624.2011.613348. [DOI] [PubMed] [Google Scholar]
- 44.Ham S, Yoon C, Lee E, Lee K, Park D, Chung E, et al. Task-based exposure assessment of nanoparticles in the workplace. J Nanopart Res. 2012;14:1–17. doi: 10.1007/s11051-012-1126-8. [DOI] [Google Scholar]
- 45.Xu HD, Zhao L, Chen ZF, Zhou JW, Tang SC, Kong FL, et al. Exposure assessment of workplace manufacturing titanium dioxide particles. J Nanopart Res. 2016;18:1–13. doi: 10.1007/s11051-015-3308-7. [DOI] [Google Scholar]
- 46.Debia M, Bakhiyi B, Ostiguy C, Verbeek JH, Brouwer DH, Murashov V. A systematic review of reported exposure to engineered nanomaterials. Ann Occup Hyg. 2016;60:916–35. doi: 10.1093/annhyg/mew041. [DOI] [PubMed] [Google Scholar]
- 47.OECD. Harmonized tiered approach to measure and assess the potential exposure to airborne emissions of engineered nano-objects and their agglomerates and aggregates at workplaces. Series on the safety of manufactured nanomaterials organisation for economic co-operation and development. J Nanopart Res: OECD; 2015. p. 55.
- 48.OECD. Strategies, techniques and sampling protocols for determining the concentrations of manufactured nanomaterials in air at the workplace. Series on the safety of manufactured nanomaterials organisation for economic co-operation and development. J Nanopart Res: OECD; 2017. p. 82.
- 49.Isaxon C, Lovén K, Ludvigsson L, Sivakumar S, Gudmundsson A, Messing ME, et al. Workplace emissions and exposures during semiconductor nanowire production, post-production, and maintenande work. Ann Work Expo Heal. 2020;64:38–54. doi: 10.1093/annweh/wxz088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Birch ME, Cary RA. Elemental carbon-based method for monitoring occupational exposures to particulate diesel exhaust. Aerosol Sci Technol. 1996;25:221–41. doi: 10.1080/02786829608965393. [DOI] [PubMed] [Google Scholar]
- 51.Hedmer M, Isaxon C, Nilsson PT, Ludvigsson L, Messing ME, Genberg J, et al. Exposure and emission measurements during production, purification, and functionalization of Arc-discharge-produced multi-walled carbon nanotubes. Ann Occup Hyg. 2014;58:355–79. doi: 10.1093/annhyg/met072. [DOI] [PubMed] [Google Scholar]
- 52.Johansson SAE. PIXE – A Novel Technique for Elemental Analysis. Endeavour. 1989;13:48–53. doi: 10.1016/0160-9327(89)90002-1. [DOI] [Google Scholar]
- 53.Nilsson TP, Isaxon C, Eriksson AC, Messing ME, Ludvigsson L, Rissler J, et al. Nano-objects emitted during maintenance of common particle generators: direct chemical characterization with aerosol mass spectrometry and implications for risk assessments. J Nanopart Res. 2013;15:1–16. doi: 10.1007/s11051-013-2052-0. [DOI] [Google Scholar]
- 54.Hedmer M, Ludvigsson L, Isaxon C, Nilsson PT, Skaug V, Bohgard M, et al. Detection of multi-walled carbon nanotubes and carbon nanodiscs on workplace surfaces at a small-scale producer. Ann Occup Hyg. 2015;59:836–52. doi: 10.1093/annhyg/mev036. [DOI] [PubMed] [Google Scholar]
- 55.Schinwald A, Murphy FA, Prina-Mello A, Poland CA, Byrne F, Movia D, et al. The threshold length for fiber-induced acute pleural inflammation: shedding light on the early events in asbestos-induced mesothelioma. Toxicol Sci. 2012;128:461–70. doi: 10.1093/toxsci/kfs171. [DOI] [PubMed] [Google Scholar]
- 56.Hashimoto N, Ogura I, Kotake M, Kishimoto A, Honda K. Evaluating the capabilities of portable black carbon monitors and photometers for measuring airborne carbon nanotubes. J Nanopart Res. 2013;15:1–17. doi: 10.1007/s11051-013-2033-3. [DOI] [Google Scholar]
- 57.Donaldson K, Murphy FA, Duffin R, Poland CA. Asbestos, carbon nanotubes and the pleural mesothelium: a review of the hypothesis regarding the role of long fibre retention in the parietal pleura, inflammation and mesothelioma. Part Fibre Toxicol. 2010;7:5. doi: 10.1186/1743-8977-7-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Bianchi MG, Campagnolo L, Allegri M, Ortelli S, Blosi M, Chiu M, et al. Length-dependent toxicity of TiO2 nanofibers: mitigation via shortening. Nanotoxicology. 2019;14:433–52. doi: 10.1080/17435390.2019.1687775. [DOI] [PubMed] [Google Scholar]
- 59.Lee JH, Kwon M, Ji JH, Kang CS, Ahn KH, Han JH, et al. Exposure assessment of workplaces manufacturing nanosized TiO2 and silver. Inhal Toxicol. 2011;23:226–36. doi: 10.3109/08958378.2011.562567. [DOI] [PubMed] [Google Scholar]
- 60.Geiss O, Bianchi I, Barrero-Moreno J. Lung-deposited surface area concentration measurements in selected occupational and non-occupational environments. J Aerosol Sci. 2016;96:24–37. doi: 10.1016/j.jaerosci.2016.02.007. [DOI] [Google Scholar]
- 61.Todea AM, Beckmann S, Kaminski H, Bard D, Bau S, Clavaguera S, et al. Inter-comparison of personal monitors for nanoparticles exposure at workplaces and in the environment. Sci Total Environ. 2017;605:929–45. doi: 10.1016/j.scitotenv.2017.06.041. [DOI] [PubMed] [Google Scholar]
- 62.Iavicoli I, Fontana L, Pingue P, Todea AM, Asbach C. Assessment of occupational exposure to engineered nanomaterials in research laboratories using personal monitors. Sci Total Environ. 2018;627:689–702. doi: 10.1016/j.scitotenv.2018.01.260. [DOI] [PubMed] [Google Scholar]
- 63.Relier C, Dubreuil M, Garcia OL, Cordelli E, Mejia J, Eleuteri P, et al. Study of TiO2 P25 nanoparticles genotoxicity on lung, blood, and liver cells in lung overload andnon-overload conditionsafter repeated respiratory exposure in rats. Toxicol Sci. 2017;156:527–37. doi: 10.1093/toxsci/kfx006. [DOI] [PubMed] [Google Scholar]
- 64.Mihalache R, Verbeek J, Graczyk H, Murashov V, van Broekhuizen P. Occupational exposure limits for manufactured nanomaterials, a systematic review. Nanotoxicology. 2017;11:7–19. doi: 10.1080/17435390.2016.1262920. [DOI] [PubMed] [Google Scholar]
- 65.NIOSH. US National Institute for Occupational Safety and Health. US: NIOSH TiO2 OEL 2011.pdf; 2011.
- 66.BSI. Nanotechnologies—Part 2: guide to safe handling and disposal of manufactured nanomaterials. PD 6699-2:2007 UK: British Standards Institution; 2007.
- 67.Tsang MP, Hristozov D, Zabeo A, Koivisto AJ, Jensen ACO, Jensen KA, et al. Probabilistic risk assessment of emerging materials: case study of titanium dioxide nanoparticles. Nanotoxicology. 2017;11:558–68. doi: 10.1080/17435390.2017.1329952. [DOI] [PubMed] [Google Scholar]
- 68.Lee YS, Sung JH, Song KS, Kim JK, Choi BS, Yu IJ, et al. Derivation of occupational exposure limits for multi-walled carbon nanotubes and graphene using subchronic inhalation toxicity data and a multi-path particle dosimetry model. Toxicol Res. 2019;8:580–6. doi: 10.1039/C9TX00026G. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Spinazze A, Cattaneo A, Borghi F, Del Buono L, Campagnolo D, Rovelli S, et al. Probabilistic approach for the risk assessment of nanomaterials: a case study for graphene nanoplatelets. Int J Hyg Envir Heal. 2019;222:76–83. doi: 10.1016/j.ijheh.2018.08.011. [DOI] [PubMed] [Google Scholar]
- 70.AethLabs. microAeth AE51 Operating Manual. US: AethLabs; 2016.
- 71.Hagler GSW, Yelverton TLB, Vedantham R, Hansen ADA, Turner JR. Post-processing method to reduce noise while preserving high time resolution in aethalometer real-time black carbon data. Aerosol Air Qual Res. 2011;11:539–46. doi: 10.4209/aaqr.2011.05.0055. [DOI] [Google Scholar]
- 72.NIOSH. Occupational exposure to carbon nanotubes and nanofibers. Current Intell Bull. 2013;65.