LETTER
Molecular testing for infectious diseases is generally both very sensitive and specific (1). Well-designed PCR primers rarely cross-react with other analytes, and specificities seen during test validation are often 100% (2). However, analytical specificities measured during limited validation studies may not reflect real-world performance across the entire testing process. Surprisingly, there is exceedingly little literature on the specificity of high-throughput, sample-to-answer PCR testing, which critically informs debates on future SARS-CoV-2 screening algorithms.
Here, we examined the false-positivity rate of high-throughput, sample-to-answer nucleic acid amplification testing (NAAT) on three commercial assays: the Hologic Panther Fusion, Hologic Aptima, and Roche cobas 6800. Our University of Washington institutional review board (IRB)-approved study specifically used high-frequency testing in asymptomatic cohorts which confirmed any positive result on an orthogonal platform and attempted to recollect the patient/employee and retest within 24 h to confirm the initial positive result.
Our testing cohort comprised 7,242 results from 451 people (median age 27 years [interquartile range [IQR], 23 to 33 years]) repetitively sampled from May to November 2020 using nasal swabs collected by a health care worker. The median number of tests per individual was 10 (IQR 6 to 16), and the median number of days between consecutive tests was 2.1 days (IQR 2 to 5 days). During the study period there were 12 positive tests (0.17%) from 9 individuals (Table 1). Total and positive results by platform were as follows: Panther Fusion, 1,932 (26.7%, six positive); Aptima TMA, 1,526 (21.0%, four positive), and Roche cobas 6800, 3,784 (52.3%, two positive). Eight positive tests (0.11%, five individuals) were considered bona fide true positives based on repeat positives or outside testing and epidemiological data. One positive test had no follow-up testing and could not be adjudicated beyond it testing positive on an orthogonal platform, suggesting it was likely a true positive as well. Three positive tests from three separate individuals did not repeat as positive on a subsequent collection, nor did the original positive specimen test positive on an orthogonal platform (cases 2, 3, and 6). We consider these three tests false positives and estimate the overall false-positive rate of high-throughput, automated, sample-to-answer NAAT to be approximately 0.04% (3/7,242, 95% confidence interval [CI], 0.01 to 0.12%), yielding a specificity of 99.96% (95% CI, 99.88 to 99.99%). Two out of three of our false-positive cases occurred using the Aptima TMA assay with relative light unit (RLU) values of 614 and 615, respectively. The Washington State SARS-CoV-2 test positivity rate was 3.7% to 10.2% during the study period (https://www.doh.wa.gov/Emergencies/COVID19/DataDashboard). The main limitation to this work is that our measured false-positivity rate may not be generalizable outside our lab, and there is a potential that our false positives could constitute low-level blip shedding (3). Only one other study has examined the false-positive rate of SARS-CoV-2 testing, which found a similarly low rate (0.1%), albeit for the Thermo TaqPath assay extracted on the KingFisher Flex, which is neither a sample-to-answer platform nor random access (4). Overall, these very low false-positivity rates associated with high-throughput SARS-CoV-2 testing offer confidence in SARS-CoV-2 PCR test results and help laboratorians, epidemiologists, and regulators understand the specificity and positive predictive value associated with high-throughput NAAT.
TABLE 1.
Case | Previous result | Days from prior test | First test platform | Repeat positive on orthogonal? | CT first test primary platform | CT first test orthogonal platform | Next result | Days to next test | Second test platform | Clinical interpretation |
---|---|---|---|---|---|---|---|---|---|---|
1 | Negative | 14 | cobas | Yes | 32.5 | 34.6 | Positive | 1 | Unknown | True positiveb |
2 | NA | NA | Aptima | No | NA | Negative | Negative | 1 | Aptima | False positiveb |
3 | NA | NA | Aptima | No | NA | Negative | Negative | 1 | Aptima | False positiveb |
4 | Negative | 2 | Fusion | Yes | 37.5 | 34.72 | NA | NA | NA | True positivec |
5 | Inconclusive | 1 | Aptima | Yes | NA | 33.7 | NA | NA | NA | True positive |
6 | Negative | 5 | cobas | No | 31.8 | Negative | Negative | 7 | cobas | False positive |
7 | Negative | 0 | Fusion | Yes | 30.6 | 29.38 | Positive | 6 | Fusion | True positive |
8 | Negative | 2 | Fusion | Yes | 17.4 | 17.93 | Positive | 1 | Fusion | True positive |
9 | Negative | 10 | Aptima | Yes | NA | 30.4 | NA | NA | NA | Unknown |
Abbreviations: CT, threshold cycle; NA, not available.
Testing at an outside lab was performed within 1 day after our result; no CT value was available.
Treated as a clinical true positive. The individual developed symptoms within 1 day of test and had close contact with a person who was positive.
ACKNOWLEDGMENTS
We acknowledge Joshua Lieberman for his helpful insights as well as the staff and patients of UW Virology and the Department of Laboratory Medicine and Pathology at the University of Washington Medical Center.
A.L.G. reports contract funding from Abbott and Gilead for testing and research funding from Merck, all outside the submitted work.
Contributor Information
Alexander L. Greninger, Email: agrening@uw.edu.
Alexander J. McAdam, Boston Children’s Hospital
REFERENCES
- 1.Burd EM. 2010. Validation of laboratory-developed molecular assays for infectious diseases. Clin Microbiol Rev 23:550–576. doi: 10.1128/CMR.00074-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lieberman JA, Pepper G, Naccache SN, Huang M-L, Jerome KR, Greninger AL. 2020. Comparison of commercially available and laboratory developed assays for in vitro detection of SARS-CoV-2 in clinical laboratories. J Clin Microbiol 58:e00821-20. doi: 10.1128/JCM.00821-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Tsuboi H, Kasamatsu Y, Matsubara S, Sasao A, Kunimitsu K, Munakata N, Ito T, Tsuchido Y, Yamawaki M, Fujita N. 2020. Two cases of novel coronavirus infection (COVID-19) with transient viral elevation using semi-quantitative real-time reverse transcription PCR and symptom relapse after completion of 10 days of favipiravir treatment. J Infect Chemother doi: 10.1016/j.jiac.2020.12.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Lin L, Carlquist J, Sinclair W, Hall T, Lopansri BK, Bennett ST. 2021. Experience with false-positive test results on the TaqPath real-time reverse transcription-polymerase chain reaction coronavirus disease 2019 (COVID-19) testing platform. Arch Pathol Lab Med 145:259–261. doi: 10.5858/arpa.2020-0612-LE. [DOI] [PubMed] [Google Scholar]