Central Message.
Late calcific rupture of ePTFE is rare but reported. Combining repair techniques and more neochordae may increase durability of mitral repair.
See Article page 34.
There are multiple surgical techniques to treat leaflet prolapse in degenerative mitral valve repair. Classic Carpentier leaflet resection and sliding leaflet plasty,1 limited leaflet resections, folding leaflet plasty, chordal transfers and shortenings, papillary muscle repositioning, and the addition of expanded polytetrafluoroethylene (ePTFE) (Gore-Tex; W.L. Gore & Associates, Inc, Newark, Del) chordae.
PTFE is a linear nonabsorbent polymer that is electronegative with a surface charge that mimics normal endothelium and less thrombogenicity. ePTFE has a porous microstructure with high breaking strength. Since testing of ePTFE in sheep2 and its introduction to the field of mitral valve surgery,3,4 this material has been used widely. ePFTE has been used as a single or multiple chordae, premeasured loops, loop-in-loop. It has been used in combination with leaflet resection or as sole modality. Their use is more convenient than complex resection techniques in less invasive surgical approaches to treat mitral valve prolapse. Excellent long-term durability of mitral valve repair has been reported using ePTFE chords.5,6
Early or late rupture of ePTFE chords is described in the literature but the overall incidence seems to be low. Causes of early rupture have been described as mechanical damage from instruments, clips, or friction; and use of CV5 ePTFE instead of CV3. Late rupture due calcific degeneration was first described by Butany and colleagues.7 They reported on histologic study that the neochordae were laminated by host fibrosa and endothelium and calcium penetrated the interstices with possible stiffening leading to potential calcification and fracture.
Bortolotti and colleagues8 reported a case of possible fatigue-induced late rupture of neochordae, and reviewed 4 cases of late neochordae rupture due to calcific degeneration.
Luthra and colleagues9 describe a case of late rupture of neochord due to calcific degeneration, although there is no pathology of the ruptured ePTFE chord.
Given the low but definitive possibility of neochordae rupture, one could consider blending more than one technique to reinforce the repair and/or using more ePTFE chordae for the length of prolapsing segment for a durable mitral valve repair.9,10 With different neochordae implantation techniques and use or lack of concurrent mitral valve repair techniques, long-term results of multiple surgeons/groups need to be studied to better understand the long-term durability of neochordae. If limited or no resection of the leaflet has been performed along with use of ePTFE chords at primary repair, re-repair more likely will be possible for recurrent mitral regurgitation due to rupture of neochordae because there could be more preserved leaflet tissue to work with.
Footnotes
Disclosures: Dr McCarthy has received consultant fees and royalties from Edwards Lifesciences, is a member of the Advisory Board for Abbott, and has received an honorarium from Atricure. Dr Pawale has nothing to disclose with regard to commercial support.
References
- 1.Carpentier A. Cardiac valve surgery—the “French correction”. J Thorac Cardiovasc Surg. 1983;86:323–337. [PubMed] [Google Scholar]
- 2.Vetter H.O., Burack J.H., Factor S.M., Malacuso F., Frater W.M. In: Biologic and Bioprosthetic Valves. Bodnar E., Yacoub M., editors. Yorke Medical Books; New York: 1986. Replacement of chordae tendineae of the mitral valve using the new expanded PTFE suture in sheep; pp. 772–785. [Google Scholar]
- 3.David T.E. Replacement of chordae tendineae with expanded polytetrafluoroethylene sutures. J Card Surg. 1989;4:286–290. doi: 10.1111/j.1540-8191.1989.tb00291.x. [DOI] [PubMed] [Google Scholar]
- 4.Zussa C., Frater R.W., Polesel E., Galloni M., Valfre C. Artificial mitral valve chordae: experimental and clinical experience. Ann Thorac Surg. 1990;50:367–373. doi: 10.1016/0003-4975(90)90476-m. [DOI] [PubMed] [Google Scholar]
- 5.David T.E., David C.M., Lafreniere-Roula M., Manlhiot C. Long-term outcomes of chordal replacement with expanded polytetrafluoroethylene sutures to repair mitral leaflet prolapse. J Thorac Cardiovasc Surg Tech. August 30, 2019 doi: 10.1016/j.jtcvs.2019.08.006. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 6.Hata H., Fujita T., Shimahara Y., Sato S., Ishibashi-Ueda H., Kobayashi J. A 25-year study of chordal replacement with expanded polytetrafluoroethylene in mitral valve repair. Interact Cardiovasc Thorac Surg. 2015;20:463–468. doi: 10.1093/icvts/ivu441. [DOI] [PubMed] [Google Scholar]
- 7.Butany J., Collins M.J., David T.E. Ruptured synthetic expanded polytetrafluoroethylene chordae tendinae. Cardiovasc Pathol. 2004;13:182–184. doi: 10.1016/S1054-8807(04)00006-7. [DOI] [PubMed] [Google Scholar]
- 8.Bortolotti U., Celiento M., Pratali S., Anastasio G., Pucci A. Recurrent mitral regurgitation due to ruptured artificial chordae: case report and review of the literature. J Heart Valve Dis. 2012;21:440–443. [PubMed] [Google Scholar]
- 9.Luthra S., Ismail A., Tsang G. Calcific degeneration and late fracture expanded polytetrafluoroethylene neochords after mitral valve repair. J Thorac Cardiovasc Surg Tech. 2020;1:34–36. doi: 10.1016/j.xjtc.2019.11.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.De Paulis R. Commentary: if your pants are too loose, use both belt and suspenders. J Thorac Cardiovasc Surg Tech. September 21, 2019 doi: 10.1016/j.jtcvs.2019.08.120. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]