Abstract
According to the statistics of the World Health Organization, an estimated 17.9 million people die from cardiovascular diseases each year, representing 31% of all global deaths. Continuous non-invasive arterial pressure (CNAP) is essential for the management of cardiovascular diseases. However, it is difficult to achieve long-term CNAP monitoring with the daily use of current devices due to irritation of the skin as well as the lack of motion artifacts suppression. Here, we report a high-performance skin-like optoelectronic system integrated with ultra-thin flexible circuits to monitor CNAP. We introduce a theoretical model via the virtual work principle for predicting the precise blood pressure and suppressing motion artifacts, and propose optical difference in the frequency domain for stable optical measurements in terms of skin-like devices. We compare the results with the blood pressure acquired by invasive (intra-arterial) blood pressure monitoring for >1500 min in total on 44 subjects in an intensive care unit. The maximum absolute errors of diastolic and systolic blood pressure were ±7/±10 mm Hg, respectively, in immobilized, and ±10/±14 mm Hg, respectively, in walking scenarios. These strategies provide advanced blood pressure monitoring techniques, which would directly address an unmet clinical need or daily use for a highly vulnerable population.
Keywords: skin-like devices, optoelectronics, blood pressure monitor, optical measurement
We demonstrate skin-like optoelectronic systems that can monitor continuous non-invasive arterial pressure, which would directly address a clinical need or daily use for the population.
INTRODUCTION
The most reliable method at present for monitoring continuous blood pressure is to insert a catheter sensor into a human artery, which results in massive suffering to the patient [1]. An alternative solution is continuous non-invasive arterial pressure (CNAP) measurement [2–4]. In the 1970s CNAP measurement was already being explored. Until now, some CNAP technologies [5–7] have been proposed and explored [8–13], such as pulse wave velocity [14–18]. These technologies have been exhaustively investigated and optimized over the past 50 years for CNAP measurement. However, there are still no suitable devices for long-term and precise CNAP monitoring in daily use due to the irritation to skin and suppression of motion artifact [19].
Devices for CNAP monitoring will be mounted on human skin for a long time, which requires them to not cause discomfort to the skin [20–23]. Some commercial devices are released for measuring CNAP. These devices need to be bundled or pinched to both fingers and arms [24,25], which results in irritation to human skin. Skin-like devices [26–35], which can be firmly mounted on and deformed with the skin, offer an alternative solution for long-term vital signs in daily use [36–38]. There are some skin-like devices for detecting blood oxygen saturation and pulse rate [39,40]. However, existing strategies for skin-like sensors cannot apply for CNAP monitoring. The certain relationship between CNAP and the pulse transit time (PTT) recording can be set up by the Moens–Korteweg (M–K) equation. PTT refers to the time it takes a pulse wave to travel between two arterial sites [41]. Photoplethysmogram (PPG) detected by optical methods can be used for PTT measurement [42,43], while only one parameter PTT cannot give both systolic and diastolic blood pressure [11,44]. Previous works introduce physical models of the blood circulation system to solve this problem [42,45]. However, these physical models did not consider the influence of the device deformation and motion artifacts. The PPG signals will be mixed with noises caused by the deformation of the device or tissue. It is necessary to build a new physical model for skin-like devices to measure precise PTT.
On the other hand, the PTT is very challenging to measure, as even the ‘onset’ of each pulse is hard to detect without external counter-pressure methods. Generally, the propagating speed of pulse wave velocity (PWV) is over 10 m/s, and the ‘onset’ of each period is difficult to detect due to the influence of motion artifacts [46,47]. The PPG or other pulse-related signals are not subject to the motion artifact or optical interference for measuring accurate PTT. Most artifacts originate in the movement of the distal arterial waveform detector in relation to the skin. There are some designs or algorithms for suppressing the motion artifacts (i.e. template matching, Masimo SET) [48–50]. However, all of them aim at solving motion artifacts for traditional devices. The motion artifacts of skin-like devices are quite different due to the deformable devices. The technologies for detecting distal arterial waveforms such as PPG are based on optical principles, and their accuracy relies on the stability of the optical path [51]. The deformation of skin-like devices will have huge impacts on the optical path. Meanwhile, the deformation of the skin can also change the optical path in the human tissue. False readings caused by shivering or other movements will be more serious in terms of skin-like optoelectronics.
In this work, we propose strategies for skin-like optoelectronic systems that can monitor precise CNAP in daily use. We introduce a theoretical analysis by the virtual work principle for considering the deformation of the skin-like system. The optical difference in the frequency domain is proposed for suppressing the motion artifacts. These features enable the skin-like systems to be mounted on the wrist without irritation and discomfort, and the long-term monitoring of CNAP using wireless data transmission. The assessment of the skin-like systems was made during comparison experiments with invasive (intra-arterial) blood pressure (IBP) monitoring in the intensive care unit (ICU) for >1500 min in total on 44 subjects. The results show that the absolute errors of diastolic and systolic blood pressure were only ±7/±10 mm Hg, respectively, over all subjects. Moreover, the absolute errors slightly increased to ±14 mm Hg in both diastolic and systolic blood pressure during walking.
RESULTS
Skin-like systems for precise CNAP monitoring
The optoelectronic device mainly consists of ultra-thin optoelectronics, watch-chain interconnects and biocompatible package. It is easy to integrate with a flexible circuit to compose a wearable skin-like system, as shown in Fig. 1A and Supplementary Fig. S1. Three inorganic light-emitting elements are in the center of the device, and four photodetectors are placed 1 cm away from both sides of the light-emitting elements, which creates a dual-channel structure for detecting the PPG signals at different location of the body (Fig. 1B). Due to enough distance between light sources and photodetectors, the optical shunt is avoided, which means the light reaches the photodetectors without passing through an arterial bed. The ultra-thin optical semiconductors hybrid-integrated here include infrared light (850 nm, GaAs based), red light (620 nm, GaAs based), green light (515 nm, Al2O3 based) and photodetector (400–1100 nm, silicon based). All of them are thinned by nano-diamond grinding technology, and the typical thickness is 11 μm/11.7 μm/11.3 μm/11.8 μm, respectively (Fig. 1C and Supplementary Fig. S2). The metal (Cu/Cr) with 160 nm thickness is patterned as a watch-chain shape to serve as electrical interconnects for power and signal transmission, as shown in Fig. 1D. This watch-chain shape has the advantages of low resistance and prominent stretchability. The function layer is packaged into a 50 μm-thick biocompatible material. Here, we utilize the transparent film dressing (W1624, Tegaderm, 3M, USA) as the package, which is widely used to cover and protect wounds and catheters. The device is imperceptible to the skin for long-term use. The thin, soft construction allows the device to make contact with skin only based on van der Waals interactions alone, as shown in Fig. 1E and F. Due to the ultra-thin structure and low modulus mechanics, the device can avoid any significant constraint on natural motions of the skin when undergoing pressing or stretching (Fig. 1G and H).
The skin-like device is easy to connect to a flexible circuit to realize signal processing and wireless transmission. Figure 1I and Supplementary Movie S1 show an example of the flexible system that is able to wirelessly transmit the signals to terminal devices such as a smart phone. Figure 1J illustrates the principle of the light-emitting and receiving sequence to realize the optical difference method. Three light-emitting elements with different wavelengths are derived by pulse light sequentially, which realize the optical difference method against deformation disturbance. The photodetectors can measure the waveforms of the PPG signals at the different locations of the artery (Fig. 1K), and then we can calculate the precise PTT value combined with the optical difference method.
Light-emitting elements and photodetectors are the inorganic semiconductors based on different materials. When their thickness reaches the level of 10 μm, the semiconductors will become very brittle and hard to be integrated. Here, we use the nano-diamond grinding process to fabricate the ultra-thin optoelectronics with high performance (Supplementary Fig. S3). The fabricating steps start with growing a AlGaInP/AlGaAs, InGaN/GaN quantum well structure epitaxial on a GaAs/sapphire substrate and doping impurity based on the silicon substrate, respectively. Nano-diamond with different sizes (diameter within ±50 nm absolute error) is sorted by centrifugation (Supplementary Table S1) and mixes to a turbid liquid with deionized water (mass ratio of 1:30). Then the optoelectronics are gradually thinned by the turbid liquid. To confirm the performance of thinned optoelectronics, we test the electroluminescence (EL) spectra of light-emitting elements and spectral responsibility of the photodetector, as shown in Fig. 2A and B. The threshold voltages of the light-emitting elements are 2.5, 1.8 and 1.3 V, respectively, and the spectral responsibility arrangement of the photodetector is 400–1200 nm. The critical parameters of the ultra-thin optoelectronics are the same as before the thinning process.
Generally, the shape of the interconnects determines the stretchability of the skin-like device. The most successful shape is the serpentine or serpentine-based pattern. However, the serpentine shape makes the actual length longer, which results in power consumption and signal attenuation. The interconnects for blood pressure monitoring transmit high-frequency signals, and the parasitic capacitance and large resistance will inevitably occur if we still use the serpentine structure. Here, we propose the watch-chain structure to guarantee both stretchability and low impedance. Finite element analysis (FEA) was utilized to analyze the strain distribution during deformation, as shown in Fig. 2C and Supplementary Fig. S4. Full 3D FEA is adopted to analyze the deformation behaviors of the ‘watch-chain’ interconnects under uniaxial stretching. The Cu electrode (thickness 150 nm) is supported by a thin layer of polyimide (PI, thickness 4 μm) and packaged by the film used in our device. The elastic modulus (E) and Poisson's ratio (ν) of Cu, PI and polyurethane film are 119 GPa/0.34, 2.5 GPa/0.34 and 4.68 MPa/0.5, respectively. Four-node shell and eight-node solid elements are used to analyze the interconnect wires and polyurethane film, and the refined meshes are adopted to ensure accuracy. In the condition of maximum pre-loaded strain up to ∼40%, the maximum strain in the interconnects is limited within 10%, which is also the fracture criteria of Cu [52]. The above analysis demonstrates that the watch-chain structure greatly improves the electronic and mechanical properties of the device. In addition, real-time tests show that the impedance of the watch-chain structure can maintain its impedance unchanged when subjected to a 40% strain. We compare the stretching results of the watch-chain and six parallel serpentine interconnects (Supplementary Fig. S5), as shown in Fig. 2D. Both structures are 1 cm in length and 150 nm in thickness. The total width of six parallel serpentines is the same as the watch-chain shape. The impedance of the watch-chain interconnect is 5.2 Ω, which is 27% smaller than 7.6 Ω of the six serpentine lines. Also, the maximum strain loaded on the watch-chain interconnect can reach 40%, which is larger than the maximum strain of the serpentine line. We repeat loading the applied strain of 0%–40%, and the skin-like device can be cycled for >300 times under this condition without any cracks (Fig. 2E).
Optical difference in the frequency domain
Most commonly, problems in PTT measurements arise from motion artifact. False readings of the PTT occur due to the irregular interference caused by movement. This problem is particularly prominent on skin-like devices due to the deformable character. Here, we propose an optical difference in the frequency domain to solve this problem, as shown in Fig. 3A. We list two examples of PPG signals and show the corresponding results processed by the optical difference method, including under the stable and motion artifact situation, respectively (Fig. 3B). The three light-emitting elements are sequentially illuminated for 100 μs at a 2 ms period (Supplementary Fig. S6), and PPG signals corresponding to different light elements are simultaneously detected by the time-multiplex method. The low-pass frequency filter is responsible for suppressing high-frequency noise in circuits, and its filter curve is shown in Supplementary Fig. S7. Generally, the original PPG signals are composed of alternating component (AC) and direct component (DC). AC is caused by the variable, pulsatile component of the light absorption by arterial blood, while DC is made up of the light absorption by tissue, venous blood and non-pulsatile arterial blood. Because the optical path of the light-emitting elements is similar, the significance of the motion artifact is also similar to each light. The AC intensities of each light are quite different when we set the DC at the same level by adjusting the intensity of incident light. Thus, the difference of the PPG intensities between green and red/infrared can keep the useful AC signals as well as effectively suppress the noise caused by motion. The frequency spectrum of each PPG signal is obtained via a fast Fourier transform, and then the amplitude difference between the green and red/infrared spectrum is calculated. At this point, the peaks on the spectrum corresponding to the pulse rate are clearly observed, which approximately equal to 1.35 Hz. Finally, the G–R and G–IR signals are converted from the frequency domain to the time domain by inverse fast Fourier transform. A homemade algorithm can achieve the optical difference method, and the flow chart is shown in Supplementary Fig. S8. We also use the sine wave signals with white noise to demonstrate efficiency. As shown in Supplementary Fig. S9, the original signals are unable to read the 1 Hz frequency even with a band-pass filter, while the optical difference in frequency can recover the periodic signal. Thus, the PPG signals are successfully separated from the motion artifact, as shown in Fig. 3C. Then we can easily obtain the precise time points of the pulse wave systolic peak (PWSP) and pulse wave end (PWE), which can be used to calculate the PTT value (Fig. 3D). The peak and valley count of the two signals are the same as the predicted values, as shown in Fig. 3E. The average periodic time of peak–peak, valley–valley and pulse are near 0.67 s as shown in Fig. 3F, which corresponds to the frequency of the pulse rate.
Assessments in ICU
We propose a new model using virtual work to connect PTT and CNAP. Blood vessels are subjected to three types of stresses that are independent of the radius of the vessels. They are the external strains by other tissues (e.g. muscles), blood pressure and the stresses resulting from the strain of blood vessels themselves. The relationship between these three stresses near PWSP and PWE can be expressed by the virtual work equation:
(1) |
In equation (1), Pbp is the instant blood pressure to the blood vessel; ρ, r0 and r are the blood density, initial and current vessel radius, respectively; Pex is the external pressure to the vessel by other tissues, including the pressure caused by motion artifact. Then we can calculate the blood pressure via equation (1) and the Beer–Lambert law:
(2) |
(3) |
where A1,2, B1,2, C1,2 and D1,2 are constants that can be calibrated by experiments, and the detailed theoretical derivation can be seen in the Methods section.
The skin-like system assessments are made on-site in Beijing Tsinghua Changgung hospital. The ICU is prospectively selected to record IBP. All the patients give written informed consent, which is approved by the ethics committee of the Tsinghua Changgung Hospital (Approval No. 18083-0-02). We use a closed-loop blood sampling system (682051, Argon, USA) and patient monitor (MX600, Philips) to record the IBP (Supplementary Fig. S10). This blood sampling system is certificated by the Gabarith test. The blood pressure is already monitored by an arterial cannula in the vessel. Drugs that may have a potential risk of changing the elasticity of blood vessels are not used during the experiments. Subjects keep the supine position, and both upper arm or wrist are at heart level. Thus, no potential energy affects the value of blood pressure. The information for each patient, including age, sex and breathing status, is recorded from their medical identifications, as shown in Supplementary Table S2. The skin-like system is mounted near the position of the radial artery, and IBP is recorded by the patient monitor for verifying the accuracy of the skin-like systems (Supplementary Fig. S11). The dual PPG signals monitored by our skin-like systems are synchronized with IBP by a time marker.
We recorded the IBP and CNAP monitored by our skin-like system on 44 subjects over 1500 min. The estimated systolic and diastolic blood pressures are summarized in Table 1 and Supplementary Table S3, respectively. In general, the absolute errors of the diastolic blood pressure (DBP) are smaller than those of systolic blood pressure (SBP), because the value of DBP is smaller than SBP. The correlation coefficients of SBP between estimated and invasive values range from 0.62 to 0.96, and their mean value and standard deviation (SD) are 0.82 and 0.08, respectively. The root mean square of estimation errors (RME) values of the 44 patients vary from 0.04% to 7.08%, and the mean value is 3.05%. The probability distribution of errors is also evaluated. Out of 50 000 heartbeats from 44 patients, 94% of the estimates were within the error range of 5 mm Hg, 4% were within 5–10 mm Hg, while only 1% of the absolute error was larger than 10 mm Hg.
Table 1.
Estimation error (%) | Error distribution (%)d | |||||
---|---|---|---|---|---|---|
Subject No. | CCa | RMEb | Meanc | <5% | <10% | ≥10% |
1 | 0.88 | 4.35 | 2.38 | 84.6 | 14.4 | 1 |
2 | 0.84 | 2.35 | 0.02 | 98 | 2 | 0 |
3 | 0.81 | 2.15 | –0.29 | 99 | 1 | 0 |
4 | 0.92 | 6.92 | 2.61 | 56.4 | 36.4 | 7.2 |
5 | 0.8 | 3.8 | 0.83 | 96.1 | 3.9 | 0 |
6 | 0.87 | 2.49 | –0.14 | 97.8 | 2.2 | 0 |
7 | 0.87 | 3.23 | –1.5 | 98 | 2 | 0 |
8 | 0.76 | 1.67 | –0.87 | 100 | 0 | 0 |
9 | 0.77 | 2.95 | –1.04 | 93.9 | 6.1 | 0 |
10 | 0.69 | 2.73 | –0.31 | 97.6 | 2.4 | 0 |
11 | 0.75 | 3.9 | –0.23 | 81.3 | 18 | 0.7 |
12 | 0.88 | 2.84 | –1.27 | 95.8 | 4.2 | 0 |
13 | 0.64 | 2.9 | 1.42 | 99.8 | 0.2 | 0 |
14 | 0.89 | 1.87 | 0.7 | 100 | 0 | 0 |
15 | 0.83 | 1.04 | –0.35 | 97.5 | 2.5 | 0 |
16 | 0.86 | 2.41 | –0.51 | 100 | 0 | 0 |
17 | 0.86 | 2.93 | 1.95 | 99.4 | 0.6 | 0 |
18 | 0.75 | 3.12 | 1.24 | 99.5 | 0.5 | 0 |
19 | 0.96 | 3.99 | 1.15 | 98.5 | 1.5 | 0 |
20 | 0.94 | 4.03 | –1.25 | 96.3 | 3.7 | 0 |
21 | 0.73 | 2.53 | 0.32 | 100 | 0 | 0 |
22 | 0.72 | 3.07 | –0.19 | 100 | 0 | 0 |
23 | 0.74 | 2.55 | –1.21 | 100 | 0 | 0 |
24 | 0.89 | 1.83 | 0.28 | 100 | 0 | 0 |
25 | 0.85 | 1.83 | –0.07 | 98.5 | 1.1 | 0.4 |
26 | 0.73 | 1.71 | 1.03 | 99.4 | 0.6 | 0 |
27 | 0.75 | 2.54 | 1.24 | 95.6 | 4.1 | 0.3 |
28 | 0.78 | 1.75 | 0.26 | 100 | 0 | 0 |
29 | 0.73 | 6.7 | 0.42 | 75.7 | 20.4 | 3.8 |
30 | 0.9 | 2.77 | 0.47 | 99.9 | 0.1 | 0 |
31 | 0.69 | 5.37 | 1.61 | 74.4 | 22.6 | 3 |
32 | 0.93 | 7.08 | –3.1 | 61.9 | 30.2 | 8 |
33 | 0.8 | 4.54 | 1.61 | 67.6 | 24.6 | 7.7 |
34 | 0.91 | 2.43 | 0.22 | 97.6 | 2.4 | 0 |
35 | 0.62 | 4.15 | 2.43 | 83.2 | 15.1 | 1.7 |
36 | 0.93 | 1.94 | 1.03 | 99.9 | 0.1 | 0 |
37 | 0.85 | 2.06 | 0.51 | 98.9 | 1.1 | 0 |
38 | 0.78 | 2.36 | –0.67 | 100 | 0 | 0 |
39 | 0.93 | 4.19 | 1.29 | 93.6 | 6.4 | 0 |
40 | 0.77 | 2.54 | –0.57 | 99.6 | 0.4 | 0 |
41 | 0.8 | 3.46 | –1.53 | 94.1 | 5.9 | 0 |
42 | 0.8 | 3.35 | 0.71 | 95.8 | 4.2 | 0 |
43 | 0.88 | 1.67 | –0.07 | 97.6 | 2.4 | 0 |
44 | 0.84 | 0.23 | 0 | 100 | 0 | 0 |
Maxe | 0.96 | 7.08 | 2.61 | 100.00 | 36.40 | 8.00 |
Minf | 0.62 | 0.23 | –3.10 | 56.40 | 0.00 | 0.00 |
Avgg | 0.82 | 3.05 | 0.24 | 93.70 | 5.53 | 0.77 |
SDh | 0.08 | 1.45 | 1.18 | 10.80 | 8.92 | 2.03 |
aCC, correlation coefficient between estimated and invasive values. bRME, root mean square of estimation errors. cMean, arithmetic means of estimation errors. dError distribution <5%, <10%, ≥10% represents range of normalized error within 5%, 10% and greater than 10%. eMax, maximum value. fMin, minimum value. gAvg, average value. hSD, standard deviation among 44 patients.
Figure 4A and B, and Supplementary Fig. S12 illustrate the CNAP monitored by the skin-like system and the IBP, respectively, as well as the estimation errors from a representative subject. The errors of SBP and DBP do not exceed 10 mm Hg in the comparison test over 1000 s. There is no significant difference between the CNAP and IBP. Figure 4C and D shows the CNAP distribution, the corresponding absolute errors, and the standard deviation (SD) of the error. The sampled SBP and DBP values range widely from 80/40 mm Hg to 200/110 mm Hg, respectively. The measuring range of the experiment covers the scale from low to ultra-high blood pressure as defined in the clinic. The absolute error and standard deviation are positively correlated with the blood pressure value, and the maximum average error is ∼10 mm Hg. Figure 4E and F shows the Bland–Altman plots corresponding to the monitored results, respectively. Here, the difference of over 55 000 values is plotted against the average values (in millimeters of mercury). Reference lines are marked to show ±1.96 SD. The ±1.96 SD is 11/7 mm Hg respectively, which means the construction of approximate 95% confidence intervals. In our tests, the overall correlation coefficient is 0.95, and the typical error is <10 mm Hg and <5 mm Hg, respectively.
The absolute errors of both SBP and DBP are less than ±10 mm Hg, and the data from skin-like systems follow a similar trend as those measured by the blood sampling system. Overall, only 2.05% points of systolic pressure (SP) and diastolic pressure (DP) are outside the 95% limits of agreement. The average difference between IBP and CNAP is <1 mm Hg, as shown in Fig. 4G and H. These results demonstrate the agreements of the two devices are also well.
Moreover, vital signs, including pulse rate, blood oxygen saturation (SpO2) and blood pressure, were monitored simultaneously in the walking scenario, as shown in Fig. 5. Also, we tested the performance with sitting and standing, as shown in Supplementary Fig. S13. In the experiments, one subject with the skin-like wearable systems mounted on the wrist was instructed to walk with arm movement. During walking, the PPG signals were recorded and wirelessly transmitted to a host terminal (Fig. 5A, Supplementary Movie S2). The original PPG signal, G–R signal, pulse rate, SpO2 and blood pressure are shown in Fig. 5B. Because the traditional devices, including fingertip oximeter and sphygmomanometer, cannot measure vital signs precisely during human walking, we monitored the SpO2 and blood pressure before walking as the standard value. During walking, the period of pulse is clearly shown in the G–R signal, which is 70 bpm. The SpO2 values by skin-like systems are near 98.8% during walking, and the maximum absolute error is less than ±1%. The monitored SP and DP fluctuate near 143 mm Hg and 66 mm Hg, respectively, and the maximum absolute error is less than ±14 mm Hg. For the blood pressure monitor, these are acceptable errors because the arm swing during walking yields a strong noise signal. The results demonstrate that the system can measure vital signs precisely and with no irritation to human activities.
DISCUSSION
CNAP has found its greatest utility as an early warning of cardiovascular disease. Our skin-like systems are an excellent tool. Here we propose skin-like optoelectronics systems for precise CNAP monitoring. The stable optical measurements for the skin-like optoelectronics system are investigated. We use the virtual work principle to analyze the stress on the blood vessel, and build up the relationship of blood pressure, PTT and PPG. This theoretical analysis is the foundation of precise CNAP monitoring. The optical difference in the frequency domain is an efficient solution for improving the performance of skin-like optoelectronics. The skin-like systems can be used for long-term monitoring in the emergency or daily settings without any irritation to the skin. Risks of allergy or other adverse reactions are minimized due to the use of compatible materials for the package. The skin-like systems are demonstrated by monitoring the blood pressure of 44 subjects over 1500 min, and the errors are acceptable in terms of clinical diagnosis.
CONCLUSION
The skin-like systems are highly versatile for applications such as monitoring ICU patients or evaluating physical conditions in daily life. Combined with a wireless many-to-one receiving terminal, the systems are convenient to report all patients’ situations to the doctor simultaneously. Thereby, the systems can serve as an early-warning functionality in terms of cardiovascular disease or potential patients. For example, the skin-like systems can be used as a precautionary monitor to prevent chronic or sudden diseases such as cardiac arrest. The same platforms can also be applied to an exercise system or sports competition to feedback every athletic physiologic parameter and guide earlier supplementation. In these scenarios or other interests, the monitoring data could be accumulated and provide information for big data analysis. The skin-like systems could be broadly utilized and change the quality of our life in different aspects, and open up a new prospect in biomedical engineering for both hospital diagnostic and daily healthcare.
METHODS
In human tissue, the blood vessels are always subjected to three forces no matter how the blood pressure changes, as shown in Fig. 6A. These forces are produced by blood pressure (Pbp), external pressure (Pex) by tissue, and the stress () in the blood vessel. The blood vessels can be treated as isotropic linear elastic material with a round shape. The blood vessel will be a stable state with a certain volume of internal blood. Under this situation, the virtual work equation can be expressed by:
(M1) |
where Pbp is real-time blood pressure, δr is the virtual change of the radius, s, ϵ and are the surface of the blood vessel, strain in the blood vessel, and volume of the blood vessel, respectively.
When the radius r changes to r + δr, the virtual work made by the three forces can be calculated as follows. The first term related to Pbp is:
(M2) |
where L is the length of the blood vessel. The δr2 is a second-order small value, and the equation (M2) can be shortened as follows:
(M3) |
The fundamental principle of the PTT-based method is based upon the pulse wave velocity (PWV) recording through the Moens–Korteweg (M–K) equation:
(M4) |
which relates PWV and PTT with the elastic modulus of the vessel wall E, blood density ρ, and arterial dimension properties, such as vessel thickness h and arterial diameter d. PWV is inversely related to PTT (PWV = K/PTT), where K is the distance between the two certain peripheral sites. The second term in equation (M1) can be expressed as follows:
(M5) |
In equation (M5), A and r0 represent the cross-sectional area of the blood vessel. The vessel is a film material and its thickness h is similar to constant when the r changes. Equation (M5) can be compacted as equation (M6):
(M6) |
The third term in equation (M1) is similar to the first term in equation (M6):
(M7) |
Now the equation (M1) can be written as equation (M8):
(M8) |
Here we define the direction away from the center of the blood vessel as positive. Equation (M8) can be compacted to equation (M9):
(M9) |
Each emitting light will be through a cutaneous vascular bed and then reflect to the photodetector during the detection of the PPG signal. The light absorption by blood and other tissue substances determines the pulse-add component of the signal, as shown in Fig. 6B. The fundamental principle of the optical absorption in tissue is based upon the Beer–Lambert law:
(M10) |
where Iout, Iin and IDC represent the output light intensity, the intensity of the incident light, and the tissue absorbed intensity, respectively; c1, ϵ1, and L1 are the average substance concentration in blood, index of light absorption, and light path in the blood. If the light source and photodetector are not very close, the light will totally cross the blood vessel and then turn back, as shown in Fig. 6D. The light path L1 will be proportional to the radius of the blood vessel:
(M11) |
C represents a certain constant in equation (M11). The radius r can be calculated by equation (M10) and (M11).
In a PPG period, the maximum and minimum intensity are corresponding to DBP and SBP. The Pbp can be obtained with equations (M9) and (M12), as shown in equation (M13):
(M12) |
(M13) |
The motion artifact will introduce the extra pressure to the external surface of the blood vessel. Figure 6C represents typical PPG signals with motion artifact. IDC fluctuations are particularly large. That means the main influences of the motion focus on the outside of the blood vessel. Thus, we can add extra pressure in equation (M13) to reflect the motion artifact:
(M14) |
where the added term is proportional to IDC, as shown in Fig. 6D.
Cm is constant. Then we can obtain the equations about SBP and DBP:
(M15) |
(M16) |
A 1,2, B1,2, C1,2 and D1,2 are certain constants that can be calibrated by experiments, and the SBP and DBP can be derived through equations (M15) and (M16).
Supplementary Material
FUNDING
This work was supported by the National Basic Research Program of China (2015CB351904) and the National Natural Science Foundation of China (11625207, 11320101001 and 11222220).
AUTHOR CONTRIBUTIONS
X. Feng, H. Li and Z. Liang designed the experiment. H. Li prepared the soft structure. H. Li and S. Cai thinned the optoelectronics. Z. Liang, Z. Wang, Y. Cao, Y. Xu, H. Zhou and H. Li monitored the vital signs. B. Lu and H. Li tested the signal. Z. Liang and Z. Han set up the model for Finite Element Method. X. Feng, H. Li, Y. Ma, Z. Liang and Y. Cao wrote the manuscript, which all authors read and commented on. X. Feng is responsible for coordinating the project.
Conflict of interest statement . None declared.
REFERENCES
- 1. Ding XR, Zhao N, Yang GZet al. . Continuous blood pressure measurement from invasive to unobtrusive: celebration of 200th birth anniversary of Carl Ludwig. IEEE J Biomed Health Inform 2016; 20: 1455–65. [DOI] [PubMed] [Google Scholar]
- 2. De Greeff A, Lorde I, Wilton Aet al. . Calibration accuracy of hospital-based non-invasive blood pressure measuring devices. J Hum Hypertens 2010; 24: 58–63. [DOI] [PubMed] [Google Scholar]
- 3. Nicklas JY, Beckmann D, Killat Jet al. . Continuous noninvasive arterial blood pressure monitoring using the vascular unloading technology during complex gastrointestinal endoscopy: a prospective observational study. J Clin Monit Comput 2019; 33: 25–30. [DOI] [PubMed] [Google Scholar]
- 4. Peng X, Schultz MG, Abhayaratna WPet al. . Comparison of central blood pressure estimated by a cuff-based device with radial tonometry. Am J Hypertens 2016; 29: 1173–8. [DOI] [PubMed] [Google Scholar]
- 5. Lillie JS, Liberson AS, Borkholder DA. Quantification of hemodynamic pulse wave velocity based on a thick wall multi-layer model for blood vessels. J Fluid Flow Heat Mass Transfer 2016; 3: 54–61. [Google Scholar]
- 6. Sharma M, Barbosa K, Ho Vet al. . Cuff-less and continuous blood pressure monitoring: a methodological review. Technologies 2017; 5: 21. [Google Scholar]
- 7. Wang C, Li X, Hu Het al. . Monitoring of the central blood pressure waveform via a conformal ultrasonic device. Nat Biomed Eng 2018; 2: 687–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Bilo G, Zorzi C, Ochoa Munera JEet al. . Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2015; 20: 291–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Ding X, Dai W, Luo Net al. . A flexible tonoarteriography-based body sensor network for cuffless measurement of arterial blood pressure. In: IEEE 12th International Conference on Wearable and Implantable Body Sensor Networks (BSN), Cambridge, MA, 2015. New York: IEEE, 2015, 1–4. [Google Scholar]
- 10. Fortin J, Marte W, Grüllenberger Ret al. . Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Comput Biol Med 2006; 36: 941–57. [DOI] [PubMed] [Google Scholar]
- 11. Kashif FM, Verghese GC, Novak Vet al. . Model-based noninvasive estimation of intracranial pressure from cerebral blood flow velocity and arterial pressure. Sci Transl Med 2012; 4: 129ra44. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Pereira T, Correia C, Cardoso J. Novel methods for pulse wave velocity measurement. J Med Biol Eng 2015; 35: 555–65. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Turner JR, Viera AJ, Shimbo D. Ambulatory blood pressure monitoring in clinical practice: a review. Am J Med 2015; 128: 14–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Mukkamala R, Hahn J-O, Inan OTet al. . Toward ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans Biomed Eng 2015; 62: 1879–901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Buxi D, Redoute JM, Yuce MR. A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time. Physiol Meas 2015; 36: R1–26. [DOI] [PubMed] [Google Scholar]
- 16. Ding X, Yan BP, Zhang YTet al. . Pulse transit time based continuous cuffless blood pressure estimation: a new extension and a comprehensive evaluation. Sci Rep 2017; 7: 11554. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Salvi P, Magnani E, Valbusa Fet al. . Comparative study of methodologies for pulse wave velocity estimation. J Hum Hypertens 2008; 22: 669–77. [DOI] [PubMed] [Google Scholar]
- 18. Westerhof N, Stergiopulos N, Noble MIMet al. . Wave travel and pulse wave velocity. Snapshots of Hemodynamics: An Aid for Clinical Research and Graduate Education. Cham: Springer International Publishing, 2019, 165–73. [Google Scholar]
- 19. Zhang YL, Zheng YY, Ma ZCet al. . Radial pulse transit time is an index of arterial stiffness. Hypertens Res 2011; 34: 884–7. [DOI] [PubMed] [Google Scholar]
- 20. Luo N, Dai W, Li Cet al. . Flexible piezoresistive sensor patch enabling ultralow power cuffless blood pressure measurement. Adv Funct Mater 2016; 26: 1178–87. [Google Scholar]
- 21. Park SI, Brenner DS, Shin Get al. . Soft, stretchable, fully implantable miniaturized optoelectronic systems for wireless optogenetics. Nat Biotechnol 2015; 33: 1280–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Zhao J, Wang X, Liu Let al. . Human skin-like, robust waterproof, and highly breathable fibrous membranes with short perfluorobutyl chains for eco-friendly protective textiles. ACS Appl Mater Interfaces 2018; 10: 30887–94. [DOI] [PubMed] [Google Scholar]
- 23. Kim SH, Jung S, Yoon ISet al. . Ultrastretchable conductor fabricated on skin-like hydrogel–elastomer hybrid substrates for skin electronics. Adv Mater 2018; 30: 1800109. [DOI] [PubMed] [Google Scholar]
- 24. Hahn R, Rinösl H, Neuner Met al. . Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia. Br J Anaesth 2012; 108: 581–5. [DOI] [PubMed] [Google Scholar]
- 25. Jeleazcov C, Krajinovic L, Münster Tet al. . Precision and accuracy of a new device (CNAP™) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth 2010; 105: 264–72. [DOI] [PubMed] [Google Scholar]
- 26. Chen Y, Lu S, Zhang Set al. . Skin-like biosensor system via electrochemical channels for noninvasive blood glucose monitoring. Sci Adv 2017; 3: e1701629. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Chortos A, Liu J, Bao Z. Pursuing prosthetic electronic skin. Nat Mater 2016; 15: 937–50. [DOI] [PubMed] [Google Scholar]
- 28. Chung HU, Kim BH, Lee JYet al. . Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care. Science 2019; 363: eaau0780. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Hong Y, Lee B, Byun Jet al. . Key enabling technology for stretchable LED display and electronic system. SID Symposium Digest of Technical Papers 2017; 48: 253–6. [Google Scholar]
- 30. Li H, Xu Y, Li Xet al. . Epidermal inorganic optoelectronics for blood oxygen measurement. Adv Healthc Mater 2017; 6: 1601013. [DOI] [PubMed] [Google Scholar]
- 31. Someya T, Bao Z, Malliaras GG. The rise of plastic bioelectronics. Nature 2016; 540: 379–85. [DOI] [PubMed] [Google Scholar]
- 32. Yokota T, Zalar P, Kaltenbrunner Met al. . Ultraflexible organic photonic skin. Sci Adv 2016; 2: e1501856. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Zhang Y, Zheng N, Cao Yet al. . Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording. Sci Adv 2019; 5: eaaw1066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Kim Y, Chortos A, Xu Wet al. . A bioinspired flexible organic artificial afferent nerve. Science 2018; 360: 998–1003. [DOI] [PubMed] [Google Scholar]
- 35. Cai M, Nie S, Du YPet al. . Soft elastomers with programmable stiffness as strain-isolating substrates for stretchable electronics. ACS Appl Mater Interfaces 2019; 11: 14340–6. [DOI] [PubMed] [Google Scholar]
- 36. Yao S, Swetha P, Zhu Y. Nanomaterial-enabled wearable sensors for healthcare. Adv Healthc Mater 2018; 7: 1700889. [DOI] [PubMed] [Google Scholar]
- 37. Xue Z, Song H, Rogers JAet al. . Mechanically-guided structural designs in stretchable inorganic electronics. Adv Mater 2019; doi: 10.1002/adma.201902254. [DOI] [PubMed] [Google Scholar]
- 38. Liu Y, Pharr M, Salvatore GA. Lab-on-skin: a review of flexible and stretchable electronics for wearable health monitoring. ACS Nano 2017; 11: 9614–35. [DOI] [PubMed] [Google Scholar]
- 39. Kim J, Gutruf P, Chiarelli AMet al. . Miniaturized battery-free wireless systems for wearable pulse oximetry. Adv Funct Mater 2017; 27: 1604373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Khan Y, Han D, Pierre Aet al. . A flexible organic reflectance oximeter array. Proc Natl Acad Sci USA 2018; 115: E11015–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Smith RP, Argod J, Pepin JLet al. . Pulse transit time: an appraisal of potential clinical applications. Thorax 1999; 54: 452–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Kim EJ, Park CG, Park Jet al. . Relationship between blood pressure parameters and pulse wave velocity in normotensive and hypertensive subjects: invasive study. J Hum Hypertens 2007; 21: 141–8. [DOI] [PubMed] [Google Scholar]
- 43. Mukkamala R, Hahn JO, Inan OTet al. . Toward ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans Biomed Eng 2015; 62: 1879–901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Fung P, Dumont G, Ries Cet al. . Continuous noninvasive blood pressure measurement by pulse transit time. In: The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Francisco, CA, 2004. New York: IEEE, 2004, 738–41. [DOI] [PubMed] [Google Scholar]
- 45. Westerhof N, Lankhaar JW, Westerhof BE. The arterial Windkessel. Med Biol Eng Comput 2009; 47: 131–41. [DOI] [PubMed] [Google Scholar]
- 46. Benes J, Simanova A, Tovarnicka Tet al. . Continuous non-invasive monitoring improves blood pressure stability in upright position: randomized controlled trial. J Clin Monit Comput 2015; 29: 11–7. [DOI] [PubMed] [Google Scholar]
- 47. Ghosh S, Banerjee A, Ray Net al. . Continuous blood pressure prediction from pulse transit time using ECG and PPG signals. In: 2016 IEEE Healthcare Innovation Point-Of-Care Technologies Conference (HI-POCT), Cancun, 2016. New York: IEEE, 2016, 188–91. [Google Scholar]
- 48. Liu J, Yan B, Zhang Yet al. . Multi-wavelength photoplethysmography enabling continuous blood pressure measurement with compact wearable electronics. IEEE Trans Biomed Eng 2018; 66: 1514–25. [DOI] [PubMed] [Google Scholar]
- 49. Griksaitis MJ, Scrimgeour GE, Pappachan JVet al. . Accuracy of the Masimo SET® LNCS neo peripheral pulse oximeter in cyanotic congenital heart disease. Cardiol Young 2016; 26: 1183–6. [DOI] [PubMed] [Google Scholar]
- 50. Lim PK, Ng S-C, Lovell NHet al. . Adaptive template matching of photoplethysmogram pulses to detect motion artefact. Physiol Meas 2018; 39: 105005. [DOI] [PubMed] [Google Scholar]
- 51. Sinex JE. Pulse oximetry: principles and limitations. Am J Emerg Med 1999; 17: 59–66. [DOI] [PubMed] [Google Scholar]
- 52. Davis J. ASM Specialty Handbook: Copper and Copper Alloys. Metals Park: ASM International, 2008. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.