Skip to main content
HHS Author Manuscripts logoLink to HHS Author Manuscripts
. Author manuscript; available in PMC: 2021 Jul 20.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2021 Feb 25;68(3):376–388. doi: 10.1109/TUFFC.2020.3037999

Hydrophone Spatial Averaging Correction for Acoustic Exposure Measurements from Arrays II: Validation for ARFI and Pulsed Doppler Waveforms

Keith A Wear 1, Anant Shah 2, Aoife M Ivory 3, Christian Baker 4
PMCID: PMC8290933  NIHMSID: NIHMS1714644  PMID: 33186103

Abstract

This article reports experimental validation of a method for correcting underestimates of peak compressional pressure (pc), peak rarefactional pressure (pr), and pulse intensity integral (pii) due to hydrophone spatial averaging effects that occur during output measurement of clinical linear and phased arrays. Pressure parameters (pc, pr, and pii), which are used to compute acoustic exposure safety indexes such as mechanical index (MI) and thermal index (TI), are often not corrected for spatial averaging because a standardized method for doing so does not exist for linear and phased arrays. In a companion article (Part I), a novel analytic, inverse-filter method was derived to correct for spatial averaging for linear or nonlinear pressure waves from linear and phased arrays. In the present article (Part II), the inverse filter is validated on measurements of acoustic radiation force impulse (ARFI) and pulsed Doppler waveforms. Empirical formulas are provided to enable researchers to predict and correct hydrophone spatial averaging errors for membrane-hydrophone-based acoustic output measurements. For example, for a 400-μm membrane hydrophone, inverse filtering reduced errors (means ± standard errors for 15 linear array / hydrophone pairs) from about 34% (pc), 22% (pr), and 45% (pii) down to within 5% for all three parameters. Inverse filtering for spatial averaging effects significantly improves accuracy of estimates of acoustic pressure parameters for ARFI and pulsed Doppler signals.

Keywords: Acoustic output measurement, exposimetry, hydrophone, spatial averaging

I. INTRODUCTION

A. Hydrophone Measurements of Acoustic Exposure

The goal of this work was to experimentally validate an inverse-filter method to improve accuracy of measurements of acoustic exposure safety parameters from clinical linear and phased arrays. Mechanical Index (MI) and thermal index (TI) characterize likelihood of mechanical and thermal bioeffects respectively [1, 2] and are partly derived from pressure measurements performed with hydrophones [38]. However, hydrophones can underestimate pressure due to spatial averaging of the pressure field across the hydrophone sensitive element surface during output measurement.

Hydrophones integrate an incident pressure wave over the surface of a sensitive element to produce an output voltage. Choice of hydrophone geometrical sensitive element size involves trade-offs. As sensitive element size increases, sensitivity tends to increase, but spatial resolution decreases and therefore spatial averaging artifacts increase. The optimum size depends on application. Common hydrophones (and geometrical sensitive element diameters) for biomedical applications include membrane (200–1000μm), needle (40–1500 μm), capsule (85–1000 μm), and fiber-optic (10–100 μm) hydrophones.

B. Distortions in Hydrophone Measurements due to Frequency-Dependent Sensitivity and Spatial Averaging

Accurate reconstruction of an input pressure wave from a hydrophone output voltage waveform entails consideration of two effects. The first effect to consider is frequency-dependent sensitivity, which is the ratio of the output voltage spectrum to the input pressure spectrum (e.g., in V/MPa) for a normally-incident quasi-planar wave. If the hydrophone sensitivity is constant over the frequency band of the pressure wave spectrum, then it is appropriate to simply divide the voltage signal by a sensitivity scale factor (e.g., in V/MPa) to convert voltage to pressure. Otherwise, it is more accurate to deconvolve the voltage signal with the sensitivity [912]. Fig. 1 shows how hydrophone sensitivity magnitudes can vary over a range from 1 MHz to 60 MHz.

Fig. 1.

Fig. 1.

Sensitivity magnitudes, |ML(f)|, for five hydrophones.

The second effect to consider is spatial averaging. If the pressure wave appears quasi-planar and perpendicularly incident across the sensitive element, then there is no need to correct for spatial averaging. Otherwise, for beams that are obliquely-incident, highly-focused or highly-nonlinear, it may be necessary to correct for spatial averaging. Nonlinear beams, such as those used in acoustic radiation force impulse (ARFI) or pulsed Doppler applications, are particularly susceptible to spatial averaging artifacts because they can contain many harmonic components with harmonic beam widths that decrease as harmonic frequency increases. Even if the fundamental beam is much broader than the hydrophone sensitive element, some harmonic components may have beam widths that are on the order of or smaller than the hydrophone sensitive element and therefore are susceptible to spatial averaging. The spatial averaging filter is a low-pass filter.

The effects of spatial averaging may be found by integrating the free field (i.e., the field in the absence of a hydrophone) over the “effective” hydrophone sensitive element surface. The effective sensitive element size depends on frequency and may be inferred from measurements of directivity [13]. Frequency-dependent effective sensitive element sizes have been reported for membrane [1418], needle [16, 19], and fiber-optic [20] hydrophones.

C. Methods for Spatial Averaging Correction

Many methods for correction for hydrophone spatial averaging have been derived for transducers with circular symmetry. One early method, which is recognized by International Electrotechnical Commission 62127–1, involves measuring a frequency-independent scale factor that is valid for linear pressure waves [7, 21]. However, nonlinear pressure waves require a frequency-dependent scale factor because of their broad bandwidth associated with multiple harmonic components. Numerical approaches have been developed that apply to linear and nonlinear pressure waves [2226]. An analytic, inverse-filter spatial averaging correction for linear or nonlinear pressure waves was derived [27] and validated for nonlinear signals measured with 1) needle hydrophones at diagnostic pressure levels [28], 2) membrane hydrophones at diagnostic pressure levels [29, 30], and 3) needle and fiber-optic hydrophones at therapeutic pressure levels [31]. While these methods have been demonstrated to be effective for sources with circular symmetry, most diagnostic ultrasound transducers have rectangular rather than circular symmetry. Empirical evidence suggests that hydrophone spatial averaging effects are significant for clinical ARFI [32] and pulsed Doppler [33] signals generated by linear arrays.

D. Outline of this Article

In a companion article (Part I), the first spatial averaging correction method for linear and phased arrays transmitting pressure waves with arbitrary linearity / nonlinearity was derived and tested using simulation [34]. In the present article (Part II), this spatial averaging correction method is validated with experimental data acquired using clinical arrays transmitting ARFI and pulsed Doppler waveforms. First, data acquisition and analysis methods using three clinical array transducers and five hydrophones are described. Second, plots of harmonic beam widths and effective hydrophone sensitive element diameters versus harmonic frequency are presented to illustrate how hydrophone spatial averaging effects increase with harmonic frequency. Third, theoretical and experimental spatial averaging filters for all transducer / hydrophone combinations are given. Fourth, a method described in the companion article (Part I) to produce simulated pressure waveforms based on parameters in regulatory acoustic output reporting tables is validated with measurements. (This simulation was used in Part I to assess the effects of hydrophone spatial averaging on MI and thermal index with bone at focus (TIB) for diagnostic ultrasonic imaging systems.) Fifth, spatial averaging errors in peak compressional pressure (pc), peak rarefactional pressure (pr), and pulse intensity integral (pii) (see 5.4.3–1 in [35]), as functions of hydrophone nominal geometrical sensitive element diameter are shown. Finally, some concluding remarks are given.

II. Methods

A. Hydrophones

Table I lists the hydrophones used. The reference hydrophone was a high-resolution capsule hydrophone (HGL-0085, Onda Corp., Sunnyvale, CA), which had a nominal geometrical sensitive element diameter of 85 μm and was expected to exhibit minimal spatial averaging effects (which will be validated in Section III.C). The frequency-dependent effective sensitive element size of the HGL-0085 was calculated from directivity measurements using methods described previously [17, 19]. Four membrane hydrophones with nominal geometrical sensitive element diameters ranging from 200 to 1000 μm (spanning the typical range for medical ultrasound) were used in order to investigate the effect of geometrical sensitive element diameter on spatial averaging artifacts.

TABLE I.

Hydrophones

Manufacturer Model Notes Nominal Geometrical Sensitive Element Diameter dg (μm)
Onda (C) HGL-0085 85
Precision Acoustics (M) D1202 differential 200
NTR (M) HMA-0500 500
Precision Acoustics (M) UT1606 600
Marconi (M) IP039 bilaminar 1000

C: capsule; M: membrane

B. Transmitted Beams

Each probe was driven with a Verasonics (Kirkland, WA) Vantage 128 system. Driving frequency, pulse duration, transmit voltage, and array-azimuthal-focusing distance were all programmed using a MATLAB (Natick, MA) script. Measurements were performed with transducer driven in two conditions: one representing ARFI mode (30 cycles) and the other representing Doppler mode (3 cycles). Although ARFI pulses can contain up to 1000 or more cycles [36], 30 cycles was considered sufficient to achieve steady state waveforms suitable for hydrophone measurement [37].

Table II lists the transducers used. The linear array transducers (with driving frequencies and transmit focal distances) included an ATL (Bothell, WA) L7–4 (5.2 MHz, 3.8 cm and 5.8 cm), a Verasonics L11–4v (7.3 MHz, 5.8 cm), and an ATL L12–5 (7.8 MHz, 4.1 cm). These transducers had frequencies and sizes that are typical for medical imaging applications. Transducers were immersed in freshly deionized water. The average water temperature was measured with a digital thermometer to be 19 ± 1° C.

TABLE II.

Transducer and Wave Properties

Name L7-4 shallow L7-4 deep L11-4 L12-5
Frequency f1 (MHz) 5.2 5.2 7.3 7.8
Lx (mm) 38.4 38.4 38.4 51.2 (25.6)
Ly (mm) 7 7 7 7
Number of Elements 128 128 128 256 (128)
Elevational focus (mm) 25 25 20 20
Programmed Focal Length (mm) 38.4 58 58 41
Measured zppsi (mm) 38 ± 1 58 ± 1 58 ± 1 41 ± 1
F# 1.0 1.5 1.5 1.6
Transmit Voltage (ARFI) (V) 27.0 30.1 28.2 30.0
Transmit Voltage (pulsed Doppler) (V) 8.3 9.9 9.9 9.9
Number of cycles (ARFI) 30 30 30 30
Number of cycles (pulsed Doppler) 3 3 3 3
qxnx = σ1x / nqx) (ARFI) 0.72 ± 0.07 0.74 ± 0.07 0.79 ± 0.08 0.82 ± 0.08
qyny = σ1y / nqy) (ARFI) 0.55 ± 0.05 0.50 ± 0.07 0.28 ± 0.03 0.31 ± 0.03
qx (pulsed Doppler) 0.66 ± 0.07 0.79 ± 0.08 0.81 ± 0.08 0.76 ± 0.08
qy (pulsed Doppler) 0.58 ± 0.06 0.49 ± 0.05 0.33 ± 0.03 0.46 ± 0.06
Focal compressional pressure pc (MPa) (ARFI) 7.57 ± 0.11 7.51 ± 0.42 6.73 ± 0.05 5.97 ± 0.11
Focal rarefactional pressure pr (MPa) (ARFI) 2.65 ± 0.02 2.09 ± 0.03 1.89 ± 0.03 1.55 ± 0.08
Focal compressional pressure pc (MPa) (pulsed Doppler) 8.61 ± 0.37 9.07 ± 0.30 6.57 ± 0.08 5.62 ± 0.11
Focal rarefactional pressure pr (MPa) (pulsed Doppler) 2.59 ± 0.02 2.14 ± 0.01 1.75 ± 0.01 1.57 ± 0.01
Spectral Index SI (ARFI) 0.24 ± 0.01 0.38 ± 0.01 0.42 ± 0.01 0.42 ± 0.01
Spectral Index SI (pulsed Doppler) 0.25 ± 0.01 0.40 ± 0.01 0.39 ± 0.01 0.35 ± 0.01
σm (ARFI) 1.31 ± 0.04 2.08 ± 0.06 2.57 ± 0.08 1.79 ± 0.05
σm (pulsed Doppler) 1.42 ± 0.04 2.32 ± 0.07 2.15 ± 0.06 1.40 ± 0.04
σq (ARFI) 1.68 ± 0.05 2.66 ± 0.08 3.45 ± 0.10 2.28 ± 0.07
σq (pulsed Doppler) 1.82 ± 0.05 3.02 ± 0.09 2.92 ± 0.09 1.84 ± 0.06

The lateral dimension of the array is x, and the elevation dimension is y. Although the L12-5 has 256 elements (spanning 51.2 mm), only the central 128 elements were used (spanning 25.6 mm). Values of q are determined from curve fits σn= σ1 / nq or alternatively FWHMn = FWHM1 / FWHMq for each of the x and y directions where n is harmonic number. zppsi is the position of the maximum measured pulse-pressure-squared integral.

C. Data Acquisition

To determine the measurement position, each hydrophone was scanned along the beam axis. Measurements were performed at the position of the maximum pulse-pressure-squared integral (zppsi) in water (not derated). For each transducer, all five hydrophone measurements were performed at the same transducer driving voltage. Vertical and horizontal scans between ±2 mm at increments of 50 μm were performed for each transducer / hydrophone combination.

Signals were acquired at 1 GHz with a calibrated Tektronix (Beaverton, OR) DPO-7254 digital phosphor oscilloscope. Data acquisition time was 10 μs.

D. Measurement of Spatial Averaging Filters

A reference time-domain voltage response up1(t) from a nonlinear tone burst pressure wave (ARFI or pulsed Doppler) was measured using the high-resolution HGL-0085 (nominal geometric sensitive element radius ag = 42.5 μm) placed at the maximum zppsi of the transducer. The membrane hydrophone voltage response up2(t) was measured by replacing the HGL-0085 with a membrane hydrophone and repeating the measurement with the same source configuration and driving voltage. This measurement was performed for each of the four membrane hydrophones in Table I. Spectra Up1(f) and Up2(f) were estimated using the Fast Fourier Transform (FFT). Spatial averaging filters Sp2(f) were evaluated using a substitution method as described previously [28].

Sp2f=Up2f/ML2(f)Up1f/ML1fSp1(f) (1)

Hydrophone sensitivities ML1(f) and ML2(f) in (1) were measured at the National Physical Laboratory (NPL) and are shown in Fig. 1. The quotients of functions in the numerator and denominator, Upi(f) / MLi(f), may be regarded as pressure spectra that have been corrected for frequency-dependent hydrophone sensitivities [7, 3840]. The sensitivity correction bandwidth was limited by the maximum frequency for which the hydrophone sensitivities were calibrated, which was 60 MHz.

The reference hydrophone spatial averaging filter Sp1(f) in (1) was computed using (15) from Part I [34], which depends on the frequency-dependent effective sensitive element size aeff(f). The form for aeff(f) for the HGL-0085 was obtained from a parametric fit to a rigid piston model [4144] for a needle hydrophone, [aeff(f) - ag] / ag = Aexp(−Bkag), where A = 1.85, B = 1.05, k = 2π/λ [28] and ag = 42.5 μm. The validity of this model for the HGL-0085 will be experimentally tested in this article (see Section III.A).

As discussed in Part I [34], signals were low-pass filtered and rarefactional filtered [45] in order to suppress common potential errors associated with inverse filtering [34].

Harmonic beam width parameters σnx and σny (where n = harmonic number) were obtained from Gaussian fits of the form exp[-x2/(2σnx2)] and exp[-y2/(2σny2)] (where x is the lateral coordinate and y is the elevational coordinate) to harmonic components extracted from FFTs of transverse scans using the high-resolution HGL-0085 hydrophone. The exponents qx and qy, where σnx = σ1x / nqx and σny = σ1y / nqy, were estimated by fitting harmonic full-width half maxima (FWHM) to a power law function of harmonic number n [27]. FWHMnx = σnx22ln2 ≈ 2.35σnx. FWHMny ≈ 2.35σny [27].

E. Validation of Simulation

A simulation was used in Part I [34] to assess the effects of hydrophone spatial averaging on MI and TIB for diagnostic ultrasound imaging systems. The simulation reconstructed radio-frequency (RF) waveforms from four parameters: driving frequency (f1), pr, pulse duration (PD), and pii (information available from acoustic output reporting tables prepared for regulatory purposes [34]). The simulation was experimentally tested in the present article (Part II) on the complete dataset of RF waveforms acquired for all transducer / hydrophone combinations. Waveforms were not deconvolved for sensitivity or corrected for hydrophone spatial averaging (because these operations had not been applied for the data in acoustic output reporting tables).

First, the four parameters (f1, pr, PD, and pii) were directly measured from the experimentally-acquired RF waveform for each transducer / hydrophone combination. Second, the simulation was used to reconstruct a waveform based on the four-parameter set for each transducer / hydrophone combination. By design, simulated waveforms matched f1 and pr exactly. Simulation performance was measured by computing means and standard deviations of errors in PD and pii (simulation vs. experiment). In addition, linear regressions of simulated vs. experimental waveforms were computed. Simulation performance was measured by correlation coefficients (ideally, close to one), regression intercepts (ideally, close to zero) and regression slopes (ideally, close to one).

F. Nonlinearity Indexes

Signal nonlinearity was characterized by the spectral index SI [46], the nonlinear propagation parameter σm [47], and the local distortion parameter σq [48]. SI is the fraction of the power spectrum contained in frequencies above the fundamental frequency [46, 49].

III. Results

A. Frequency-Dependent Effective Sensitive Element Diameter for HGL-0085

Fig. 2 shows measurements of frequency-dependent effective sensitive element diameter for the high-resolution HGL-0085 hydrophone, obtained from directivity measurements as described previously [17, 19]. As frequency increases, the effective sensitive element diameter asymptotically approaches the geometrical sensitive element diameter of 85 μm. Fig. 2 also shows the prediction derived from the rigid piston model directivity [43]. The rigid piston model has previously been shown to accurately predict directivity and frequency-dependent effective sensitive element size of needle [19] and reflectance-based [20] hydrophones.

Fig. 2.

Fig. 2.

Frequency-dependent effective sensitive element diameter for the HGL-0085 hydrophone. The prediction derived from the rigid piston (RP) model directivity is also shown.

B. Harmonic-Frequency-Dependent Beam Widths

Fig. 3 shows focal plane lateral beam profiles for harmonics from the L7–4 transducer, measured with the high-resolution HGL-0085 hydrophone, at two different zoom levels. Gaussian fits are also shown. Fig. 3 shows that the harmonic lateral beam profiles were approximated well by Gaussians (as assumed by the theory) at least over a range of axial distances corresponding to a nominal geometrical sensitive element diameter of 500 μm.

Fig. 3.

Fig. 3.

Focal plane lateral beam profiles for harmonics from the L7–4 transducer. The right panel is a zoomed version of the left panel. Error bars denote ± one standard deviation. Gaussian fits are shown in black. The vertical dashed red lines show the geometrical extent of a hydrophone with a sensitive element diameter of 500 μm. Harmonic numbers are denoted by n.

Fig. 4 shows lateral and elevational FWHM (measured in scans with the high-resolution HGL-0085 hydrophone) plotted vs. harmonic frequency. Fits for frequency-dependent FWHM inversely proportional to nq (where n is harmonic number) are also shown. The fitting parameters are denoted by qx and qy in the lateral and elevational directions respectively and are reported in Table II. The root-mean-squared (RMS) differences between frequency-dependent FWHM and fits inversely proportional to nq in the main spatial averaging direction (lateral direction) over the harmonic frequencies shown in Fig. 4 were 40 μm ± 17 μm (ARFI) and 38 μm ± 18 μm (pulsed Doppler). The dashed lines show frequency-dependent effective sensitive element diameters deff(f) for five membrane hydrophones with geometrical sensitive element diameters dg ranging from 200 to 1000 μm based on an empirical model [17]. At high frequencies, deff(f) asymptotically approaches a value close to dg but at low frequencies, deff(f) > dg.

Fig. 4.

Fig. 4.

Lateral (blue asterisks) and elevational (black x’s) full-width half maxima (FWHM) plotted vs. harmonic frequency. Error bars denote ± one standard deviation. Fits for frequency-dependent FWHM inversely proportional to nq (where n is harmonic number) are also shown. The dashed lines show frequency-dependent effective sensitive element diameters deff(f) for five membrane hydrophones with geometrical sensitive element diameters dg ranging from 200 to 1000 μm.

Fig. 4 shows that spatial averaging is rarely a concern in the elevational dimension because FWHMelev(f) is almost always much greater than deff(f) (except sometimes at high frequencies for hydrophones with large sensitive elements). However, spatial averaging is almost always a concern in the lateral dimension because FWHMlat(f) is often comparable to or even smaller than deff(f), even for hydrophones with 200-μm-geometrical-diameter sensitive elements.

C. Spatial Averaging Filters and Estimates of pc, pr, and pii

Figs. 5 and 6 show spatial averaging filters (SAFs) for all combinations of transducers and hydrophones for ARFI (Fig. 5) and pulsed Doppler (Fig. 6) signals. The red curves are theoretical SAFs based on two-dimensional integrals over the frequency-dependent sensitive element surfaces, (10) in Part I [34]. The black dashed curves are analytic forms based on “equivalent square” hydrophones, (15) in Part I. The measurements are in good agreement with both theoretical forms. Averaged over all hydrophones in Table I and all transducers in Table II, the root-mean-squared errors (RMSE) between theoretical and experimental spatial averaging filters were 2.8% ± 1.3% (ARFI) and 3.2% ± 1.9% (pulsed Doppler). Figs. 5 and 6 illustrate the low-pass behavior of SAFs.

Fig. 5.

Fig. 5.

Spatial averaging filters (SAF) for ARFI measurements for all transducer / hydrophone combinations. Error bars denote ± one standard deviation. Red curves are theoretical SAFs based on two-dimensional integrals over frequency-dependent sensitive element surfaces. Black dashed curves are analytic forms based on “equivalent square” hydrophones.

Fig. 6.

Fig. 6.

Spatial averaging filters (SAF) for Doppler measurements for all transducer / hydrophone combinations. Error bars denote ± one standard deviation. Red curves are theoretical SAFs based on two-dimensional integrals over frequency-dependent sensitive element surfaces. Black dashed curves are analytic forms based on “equivalent square” hydrophones.

Figs. 7 and 8 show effects of spatial averaging on measurements of peak compressional pressure (pc), peak rarefactional pressure (pr), and pulse intensity integral (pii) at the focal plane (and not derated) for ARFI (Fig. 7) and pulsed Doppler (Fig. 8) signals measured with all transducer / hydrophone combinations. The plots show 1) raw data (“Raw RF”), 2) data deconvolved for hydrophone sensitivity only (“Decon Sens”), and 3) data deconvolved for hydrophone sensitivity and inverse filtered using spatial averaging filters (“Decons Sens + SAF”). Data are plotted as functions of hydrophone nominal geometrical sensitive element diameter, dg. Exponential and linear fits to data are also shown. Figs. 7 and 8 show that inverse filtering with the spatial averaging filter reduces the dependence of measurements on hydrophone sensitive element size. Scatter in Figs. 7 and 8 may be due to uncertainties in hydrophone sensitivities and positioning.

Fig. 7.

Fig. 7.

Effects of spatial averaging on measurements of peak compressional pressure (pc), peak rarefactional pressure (pr), and pulse intensity integral (pii) for ARFI signals measured with all transducer / hydrophone combinations. Names of measured parameters are placed at the top of each column (instead of as y-axis labels) to support matrix format for presenting results. The plots show 1) raw data (“Raw RF”), 2) data deconvolved for hydrophone sensitivity only (“Decon Sens”), and 3) data deconvolved for hydrophone sensitivity and inverse filtered using spatial averaging filters (“Decons Sens + SAF”). Data are plotted as functions of hydrophone nominal geometrical sensitive element diameter, dg. Error bars (which are so small they are difficult to discern) denote standard deviations. Exponential and linear fits to data are also shown.

Fig. 8.

Fig. 8.

Effects of spatial averaging on measurements of peak compressional pressure (pc), peak rarefactional pressure (pr), and pulse intensity integral (pii) for Doppler signals measured with all transducer / hydrophone combinations. Names of measured parameters are placed at the top of each column (instead of as y-axis labels) to support matrix format for presenting results. The plots show 1) raw data (“Raw RF”), 2) data deconvolved for hydrophone sensitivity only (“Decon Sens”), and 3) data deconvolved for hydrophone sensitivity and inverse filtered using spatial averaging filters (“Decons Sens + SAF”). Data are plotted as functions of hydrophone nominal geometrical sensitive element diameter, dg. Error bars (which are so small they are difficult to discern) denote standard deviations. Exponential and linear fits to data are also shown.

Figs. 7 and 8 show that values for pc, pr, and pii measured with the high-resolution reference HGL-0085 hydrophone (dg = 85 μm) that were not corrected for spatial averaging (leftmost red point in each panel, “Decon Sens”) were close to values that were corrected for spatial-averaging (leftmost blue point in each panel, “Decon Sens + SAF”). The RMS differences (uncorrected vs. corrected for hydrophone spatial averaging) in parameters measured with the HGL-0085 hydrophone were 6% (ARFI pc), 3% (ARFI pr), 6% (ARFI pii), 4% (Doppler pc), 4% (Doppler pr), and 5% (Doppler pii). These small values imply that the reference SAF (Sp1(f) in (1)) did not have a big effect for the HGL-0085.

The fits may be used to predict performance of the spatial averaging filter for a typical membrane hydrophone with 400-μm geometrical sensitive element diameter, which was the most common hydrophone in the FDA database for ARFI and pulsed Doppler submissions between 2015 and 2019 [34]. Table III shows that inverse filtering with the spatial averaging filter reduced spatial averaging error magnitudes from about 34% (pc), 22% (pr), and 45% (pii) down to within 5% for all three parameters. The latter two parameters, pr and pii, are typically included in acoustic output reporting tables for regulatory purposes, while pc is not.

TABLE III.

Percent Errors in Acoustic Output Before and After Inverse Filtering to Compensate for Spatial Averaging

Pressure Parameter ARFI Pulsed Doppler
Before After Before After
pc −34% ± 2% 4% ± 3% −34% ± 2% 0% ± 3%
pr −23% ± 4% −2% ± 7% −21% ± 2% −4% ± 5%
pii −46% ± 5% 5% ± 6% −43% ± 2% 3% ± 2%

Values assume a typical membrane hydrophone with 400 μm geometrical sensitive element diameter. Values are means ± standard errors.

D. Validation of Simulation

Fig. 9 shows measured and simulated pulsed Doppler waveforms. Simulation errors were computed as differences between parameters or waveforms measured experimentally vs. those reconstructed from only input values for f1, pr, PD, and pii.

Fig. 9.

Fig. 9.

Pulsed Doppler waveform (blue solid line) generated by the L7–4 transducer and measured with the NTR hydrophone. Simulated waveform (red dashed line) reconstructed from acoustic output reporting parameters: f1, pr, pulse duration (PD), and pii.

For simulated ARFI waveforms, the error in PD was −1.3% ± 4.4% (mean ± standard deviation) and the error in pii was 0.2% ± 3.8% for all transducer / hydrophone combinations. The correlation coefficient between simulated and experimental waveform values was 0.95 ± 0.02 (mean ± standard deviation). The intercept and slope of linear regression fits of simulated to experimental waveform values were −0.04 ± 0.02 MPa and 0.96 ± 0.02 (means ± standard deviations).

For pulsed Doppler waveforms, the error in PD was 0.1% ± 4.3% and the error in pii was 2.5% ± 8.1% for all transducer / hydrophone combinations. The correlation coefficient between simulated and experimental waveform values was 0.90 ± 0.03. The intercept and slope of linear regression fits of simulated to experimental waveforms were −0.04 ± 0.01 MPa and 0.92 ± 0.03 (means ± standard deviations).

The ARFI and pulsed Doppler results suggest that the simulation could reliably reconstruct realistic waveforms based only on information available in acoustic output reporting tables. This supports the assessment in Part I of the effects of hydrophone spatial averaging on MI and TIB for diagnostic ultrasonic imaging systems

E. Empirical Formulas for Spatial Averaging Errors

Fig. 10 shows experimental (from Part II) and simulated (from Part I) errors in pressure parameters as functions of dg/(λ1F/#) for ARFI (left column) and pulsed Doppler (right column) waveforms. The product λ1F/# is an index of the fundamental focal spot width. FWHM1x ≈ 1.2λ1F/# [50]. Experimental and simulated data follow similar trends. This level of agreement supports some of the underlying assumptions of the simulation, including the functional form of the pulse and the dynamic aperture algorithm (i.e., the method by which the number of transmit elements, and therefore the effective array width, is reduced for shallow imaging depths).

Fig. 10.

Fig. 10.

Experimental and simulated errors in pressure parameters for ARFI (left column) and pulsed Doppler (right column) waveforms. Experimental data are fit to functions of the form (–100%)[1 – exp(−αx)] where x = dg / (λ1F/#) and α is a fitting parameter. Standard deviations for measurements are less than one percent. The abscissa is the ratio of the membrane hydrophone geometrical sensitive element diameter to the product of the fundamental wavelength λ1 and the F/# (ratio of focal length to array width). The product λ1F/# is an index of the fundamental focal spot width. FWHM1x ≈ 1.2λ1F/#. All experimental and simulation data were inverse filtered to deconvolve the effects of hydrophone frequency-dependent sensitivity.

Empirical formulas for relative errors in pressure parameters were generated to provide correction factors for membrane-hydrophone-based acoustic output measurements, which may be useful particularly when aeff(f) for the specific hydrophone used (e.g., obtained from directivity measurements) is unavailable. The experimental data in Fig. 10 were fit to exponential functions of the form (−100%)[1 – exp(−αx)] where x = dg / (λ1F/#). Table IV gives α values that may be used to correct for spatial averaging errors in pressure parameters for a given set of experimental parameters (dg, λ1, and F/#). Table IV also gives RMSD values between data and fits.

TABLE IV.

Spatial Averaging Error Coefficients

Pressure Parameter ARFI Pulsed Doppler
α RMSD (%) α RMSD (%)
pc 0.47 7.5 0.49 8.0
pr 0.22 8.9 0.22 7.9
pii 0.63 8.2 0.62 7.9

Spatial Averaging Error = (−100%) [1 − exp(−αx)], where x = dg / (λ1F/#). RMSD is root-mean-squared difference between experimental data points in Fig. 10 and function fits. These coefficients are for pressure measurements performed with membrane hydrophones.

Simulations based on circular transducers suggest that errors in pr are essentially independent of local distortion parameter σq (see Table IV and Figs. 12 and 13 in [30]). This is because pr is primarily determined by low-frequency spectral content [51] while σq is primarily an indicator of harmonic spectral content. (See Fig. 5 in [34]).

Simulations based on circular transducers suggest that errors in pii increase with σq (see Table IV and Figs. 12 and 13 in [30]). However, if dg is chosen to be small enough so that dg/(λ1F/#) < 1, then errors in pii show little dependence on σq, provided that σq ≤ 2, which covers a great range of signals (as can be seen in Fig. 5 of [30]). For the data shown in Table II, σq = 2.5 ± 0.7 (ARFI) and 2.4 ± 0.7 (pulsed Doppler (means ± standard deviations).

IV. Conclusion

An inverse-filter method for correcting for hydrophone spatial averaging artifacts derived in Part I [34] has been validated with experimental data acquired using three clinical linear array transducers and five hydrophones. Without correction for spatial averaging artifacts, errors in MI and TIB can be substantial for ARFI and pulsed Doppler waveforms using typical transducers and hydrophones. A simulation method to obtain time-domain RF waveforms from data in acoustic output reporting tables (described in Part I) has been validated with experimental data acquired using three clinical linear array transducers and five hydrophones. The experimental validation in Part II reinforces the predictions in Part I regarding quantitative estimates for spatial averaging errors in MI and TIB for clinical diagnostic ultrasound systems.

Acknowledgements

The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. The authors are grateful for funding support from the FDA Office of Women’s Health and the U.K. Department for Business, Energy & Industrial Strategy.

This work was supported by the U.S. Food and Drug Administration Office of Women’s Health and the U.K. Department for Business, Energy & Industrial Strategy.

Contributor Information

Keith A. Wear, U.S. Food and Drug Administration, Silver Spring, MD, 20993.

Anant Shah, National Physical Laboratory, Teddington, U.K..

Aoife M. Ivory, National Physical Laboratory, Teddington, U.K.

Christian Baker, National Physical Laboratory, Teddington, U.K..

REFERENCES

  • [1].Apfel RE and Holland CK, “Gauging the likelihood of cavitation from short-pulse, low-duty cycle diagnostic ultrasound,” Ultrasound Med Biol, vol. 17, no. 2, pp. 179–85, 1991. [DOI] [PubMed] [Google Scholar]
  • [2].Bigelow TA et al. , “The thermal index: its strengths, weaknesses, and proposed improvements,” J Ultrasound Med, vol. 30, no. 5, pp. 714–34, May 2011. [DOI] [PubMed] [Google Scholar]
  • [3].Harris GR, Preston RC, and Lewin PA, “Special Issue on Ultrasound Exposimetry,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 35, no. 2, pp. 87–269, 1988.18290135 [Google Scholar]
  • [4].Harris GR, “Hydrophone Measurements in Diagnostic Ultrasound Fields,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 35, no. 2, pp. 87–101, March 1988. [DOI] [PubMed] [Google Scholar]
  • [5].Harris GR, “Progress in medical ultrasound exposimetry,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 52, no. 5, pp. 717–36, May 2005. [DOI] [PubMed] [Google Scholar]
  • [6].Szabo TL, “Ultrasonic Exposimetry and Acoustic Measurements,” in Diagnostic Ultrasound Imaging: Inside Out 2nd ed. San Diego, CA, USA: Academic Press, 2014, pp. 565–604. [Google Scholar]
  • [7].IEC 62127–1 Ultrasonics – Hydrophones – Part 1: Measurement and characterization of medical ultrasonic fields up to 40 MHz, 2013. [Google Scholar]
  • [8].IEC 62359 Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017. [Google Scholar]
  • [9].Hurrell A, “Voltage to Pressure Conversion: Are you getting “phased” by the problem,” (in English), AMUM 2004: Advanced Metrology for Ultrasound in Medicine 2004, vol. 1, no. 1, pp. 57–62, 2004. [Google Scholar]
  • [10].Wilkens V and Koch C, “Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone,” (in English), J. Acoust. Soc. Am., vol. 115, no. 6, pp. 2892–2903, June 2004. [Google Scholar]
  • [11].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris GR, “Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 61, no. 1, pp. 62–75, January 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Martin E and Treeby B, “Investigation of the repeatability and reproducibility of hydrophone measurements of medical ultrasound fields,” J. Acoust. Soc. Am., vol. 145, no. 3, pp. 1270–1282, 2019. [DOI] [PubMed] [Google Scholar]
  • [13].IEC 62127–3 Ultrasonics – Hydrophones – Part 3: Properties of hydrophones for ultrasonic fields up to 40 MHz, 2013. [Google Scholar]
  • [14].Shombert DG, Smith SW, and Harris GR, “Angular Response of Miniature Ultrasonic Hydrophones,” (in English), Medical Physics, vol. 9, no. 4, pp. 484–492, 1982. [DOI] [PubMed] [Google Scholar]
  • [15].Harris GR and Shombert DG, “A Pulsed near-Field Technique for Measuring the Directional Characteristics of Acoustic Receivers,” (in English), IEEE Trans Son. Ultrason., vol. 32, no. 6, pp. 802–808, 1985. [Google Scholar]
  • [16].Radulescu EG, Lewin PA, Nowicki A, and Berger WA, “Hydrophones’ effective diameter measurements as a quasi-continuous function of frequency,” (in English), Ultrasonics, vol. 41, no. 8, pp. 635–641, November 2003. [DOI] [PubMed] [Google Scholar]
  • [17].Wear KA, Baker C, and Miloro P, “Directivity and Frequency-Dependent Effective Sensitive Element Size of Membrane Hydrophones: Theory vs. Experiment,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 11, pp. 1723–1730, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Wilkens V and Molkenstruck W, “Broadband PVDF membrane hydrophone for comparisons of hydrophone calibration methods up to 140 MHz,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 54, no. 9, pp. 1784–1791, September 2007. [DOI] [PubMed] [Google Scholar]
  • [19].Wear KA, Baker C, and Miloro P, “Directivity and Frequency-Dependent Effective Sensitive Element Size of Needle Hydrophones: Predictions from Four Theoretical Forms Compared with Measurements,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 65, no. 10, pp. 1781–1788, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Wear KA and Howard SM, “Directivity and frequency-dependent effective sensitive element size of a reflectance-based fiber optic hydrophone: predictions from theoretical models compared with measurements,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 65, no. 12, pp. 2343–2348, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Preston RC, Bacon DR, and Smith RA, “Calibration of medical ultrasonic equipment-procedures and accuracy assessment,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 35, no. 2, pp. 110–21, 1988. [DOI] [PubMed] [Google Scholar]
  • [22].Radulescu EG, Lewin PA, Goldstein A, and Nowicki A, “Hydrophone spatial averaging corrections from 1 to 40 MHz,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 48, no. 6, pp. 1575–1580, November 2001. [DOI] [PubMed] [Google Scholar]
  • [23].Radulescu EG, Lewin PA, and Nowicki A, “1–60 MHz measurements in focused acoustic fields using spatial averaging corrections,” Ultrasonics, vol. 40, no. 1–8, pp. 497–501, May 2002. [DOI] [PubMed] [Google Scholar]
  • [24].Zeqiri B and Bond AD, “The Influence of Wave-Form Distortion on Hydrophone Spatial-Averaging Corrections - Theory and Measurement,” (in English), J. Acoust. Soc. Am., vol. 92, no. 4, pp. 1809–1821, October 1992. [Google Scholar]
  • [25].Cooling MP, Humphrey VF, and Wilkens V, “Hydrophone area-averaging correction factors in nonlinearly generated ultrasonic beams,” J. Phys. Conf. Series, Adv. Metro for Ultrasound in Med., vol. 279, pp. 1–6, 2011. [Google Scholar]
  • [26].Bessonova O and Wilkens V, “Investigation of Spatial Averaging Effect of Membrane Hydrophones for Working Frequencies in the Low MHz Range,” DAGA 2012 - Darmstadt, pp. 937–938, 2012. [Google Scholar]
  • [27].Wear KA, “Considerations for choosing sensitive element size for needle and fiber-optic hydrophones I: Theory and graphical guide,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 2, pp. 318–339, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Wear KA and Liu Y, “Considerations for choosing sensitive element size for needle and fiber-optic hydrophones II: Experimental validation of spatial averaging model,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 2, pp. 340–347, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Wear K, Shah A, and Baker C, “Correction for Spatial Averaging Artifacts for Circularly-Symmetric Pressure Beams Measured with Membrane Hydrophones,” in IEEE Intl. Ultrason. Symp., 2020, pp. 978–1-7281–5448-0/20, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Wear KA, Shah A, and Baker C, “Correction for Hydrophone Spatial Averaging Artifacts for Circular Sources,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. in press, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Wear KA and Howard SM, “Correction for spatial averaging artifacts in hydrophone measurements of high intensity therapeutic ultrasound: an inverse filter approach,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 9, pp. 1453–1464, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Wear K, Shah A, Ivory AM, and Baker C, “Hydrophone Spatial Averaging Artifacts for ARFI Beams from Array Transducers “ in IEEE Intl. Ultrason. Symp., 2020, pp. 978–1-7281–5448-0/20, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Wear K, Shah A, Ivory AM, and Baker C, “Hydrophone Spatial Averaging Artifacts for Pulsed Doppler Beams from Array Transducers “ in IEEE Intl. Ultrason. Symp., 2020, pp. 978–1-7281–5448-0/20, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Wear KA, “Hydrophone Spatial Averaging Correction for Acoustic Exposure Measurements from Arrays I: Theory and Impact on Diagnostic Safety Indexes,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. in press, 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [35].Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, 2004. [Google Scholar]
  • [36].Deng Y, Rouze NC, Palmeri ML, and Nightingale KR, “Ultrasonic Shear Wave Elasticity Imaging Sequencing and Data Processing Using a Verasonics Research Scanner,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 64, no. 1, pp. 164–176, January 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Deng Y, Palmeri ML, Rouze NC, Rosenzweig SJ, Abdelmalek MF, and Nightingale KR, “Analyzing the Impact of Increasing Mechanical Index and Energy Deposition on Shear Wave Speed Reconstruction in Human Liver,” Ultrasound Med Biol, vol. 41, no. 7, pp. 1948–57, July 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [38].Hurrell AM and Rajagopal S, “The Practicalities of Obtaining and Using Hydrophone Calibration Data to Derive Pressure Waveforms,” (in English), IEEE Trans. Ultrason., Ferroelectr. Freq. Control, vol. 64, no. 1, pp. 126–140, January 2017. [DOI] [PubMed] [Google Scholar]
  • [39].Eichstadt S, Wilkens V, Dienstfrey A, Hale P, Hughes B, and Jarvis C, “On challenges in the uncertainty evaluation for time-dependent measurements,” (in English), Metrologia, vol. 53, no. 4, pp. S125–S135, August 2016. [Google Scholar]
  • [40].Liu Y, Wear KA, and Harris GR, “Variation of High-Intensity Therapeutic Ultrasound (HITU) Pressure Field Characterization: Effects of Hydrophone Choice, Nonlinearity, Spatial Averaging and Complex Deconvolution,” Ultrasound Med Biol, vol. 43, no. 10, pp. 2329–2342, October 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [41].Jones DS, “The Scattering of a Scalar Wave by a Semi-Infinite Rod of Circular Cross Section,” (in English), Philosophical Transactions of the Royal Society of London Series a-Mathematical and Physical Sciences, vol. 247, no. 934, pp. 499–528, 1955. [Google Scholar]
  • [42].Harris GR and Gammell PM, “Sensitivity measurements of piezoelectric polymer hydrophones from 0.2–2 MHz using a broadband-pulse technique,” (in English), Journal of the Acoustical Society of America, vol. 105, no. 2, pp. 725–731, February 1999. [Google Scholar]
  • [43].Krucker JF, Eisenberg A, Krix M, Lotsch R, Pessel M, and Trier HG, “Rigid piston approximation for computing the transfer function and angular response of a fiber-optic hydrophone,” (in English), J. Acoust. Soc. Am., vol. 107, no. 4, pp. 1994–2003, April 2000. [DOI] [PubMed] [Google Scholar]
  • [44].Wear KA, Liu Y, and Harris GR, “Pressure pulse distortion by needle and fiber-optic hydrophones due to nonuniform sensitivity,” IEEE Trans Ultrason Ferroelectr. Freq Control, vol. 65, no. 2, pp. 137–148, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [45].Wear KA, Liu YB, Gammell PM, Maruvada S, and Harris GR, “Correction for Frequency-Dependent Hydrophone Response to Nonlinear Pressure Waves Using Complex Deconvolution and Rarefactional Filtering: Application With Fiber Optic Hydrophones,” (in English), IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 62, no. 1, pp. 152–164, January 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46].Duck FA, “Estimating in situ exposure in the presence of acoustic nonlinearity,” J Ultrasound Med, vol. 18, no. 1, pp. 43–53, January 1999. [DOI] [PubMed] [Google Scholar]
  • [47].Bacon DR, “Finite-Amplitude Distortion of the Pulsed Fields Used in Diagnostic Ultrasound,” (in English), Ultrasound in Medicine and Biology, vol. 10, no. 2, pp. 189–195, 1984. [DOI] [PubMed] [Google Scholar]
  • [48].Duncan T, Humphrey VF, and Duck FA, “A numerical comparison of nonlinear indicators for diagnostic ultrasound fields,” Proceedings World Congress of Ultrasonics, pp. 85–88, 2003. [Google Scholar]
  • [49].Bigelow TA and O’Brien WD, “Experimental evaluation of indicators of nonlinearity for use in ultrasound transducer characterizations,” (in English), Ultrasound in Medicine and Biology, vol. 28, no. 11–12, pp. 1509–1520, November-December 2002. [DOI] [PubMed] [Google Scholar]
  • [50].Szabo TL, “Beamforming,” in Diagnostic Ultrasound Imaging - Inside Out Second ed. Oxford, U.K.: Elsevier, 2014. [Google Scholar]
  • [51].Harris GR, “Are current hydrophone low frequency response standards acceptable for measuring mechanical/cavitation indices?,” (in English), Ultrasonics, vol. 34, no. 6, pp. 649–654, August 1996. [DOI] [PubMed] [Google Scholar]

RESOURCES