Phase contrast coronary blood velocity mapping with both high temporal and spatial resolution using triggered Golden Angle rotated Spiral k-t Sparse Parallel imaging (GASSP) with shifted binning

Dan Zhu 1,2, Gabriele Bonanno 2,3, Allison G Hays 2,3, Robert G Weiss 2,3, Michael Schär 2,*
PMCID: PMC8295217  NIHMSID: NIHMS1700594  PMID: 33977581

Abstract

Purpose:

High temporal and spatial resolutions are required for coronary blood flow measures. Current spiral breath-hold phase contrast (PC) MRI at 3T focus on either high spatial or high temporal resolution. We propose a golden angle (GA) rotated Spiral k-t Sparse Parallel imaging (GASSP) sequence for both high spatial (0.8mm) and high temporal (<21ms) resolutions.

Methods:

GASSP PC data are acquired in left anterior descending and right coronary arteries of 8 healthy subjects. Binning of GA rotated spiral data into cardiac frames may lead to large k-space gaps. To reduce those gaps, the binning window is shifted and a triggered GA scheme that resets the rotations angle every heartbeat is proposed. The gap reductions are evaluated in simulations and all subjects. Peak systolic velocity (PSV), peak diastolic velocity (PDV), coronary blood flow rate, and vessel area are validated against two reference scans, and repeatability / reproducibility are determined.

Results:

Shifted binning reduced the mean k-space gaps of the triggered GA scheme by 14°–22° in simulations and about 20° in vivo. The k-space gap across 3 cardiac frames was reduced with the triggered GA scheme compared to the standard GA scheme (35.3°±3.6° vs. 43°±13.7°, t-test P=0.04). PSV, PDV, flow rate and area had high intra-scan repeatability (0.92≤ICC≤0.99), and inter-scan (0.78≤ICC≤0.91) and intra-observer (0.91≤ICC≤0.98) reproducibility.

Conclusion:

GASSP enables single breath-hold coronary PC MRI with high temporal and spatial resolutions. Shifted binning and a triggered GA scheme reduce k-space gaps. Quantitative coronary flow metrics are highly reproducible, especially within the same scanning session.

Keywords: coronary flow, coronary velocity, coronary MRI, golden angle, spiral MRI, k-t sparse SENSE

Introduction

Phase contrast (PC) velocity-encoded MRI (1) enables blood velocity and flow measurements of the coronary arteries (2,3). These measurements are of clinical interest, because they enable the assessment of vascular physiology, to complement conventional anatomic lumen imaging. Among others, coronary PC MRI has been applied to assess coronary flow reserve (47), local coronary endothelial dysfunction (810), and more recently to measure the pressure gradients across a coronary artery stenosis (11).

Coronary PC MRI is challenging because the coronary arteries are small in diameter (<3–4 mm) and move due to cardiac contraction and respiration. High spatial resolution Δx ≤ 1mm is needed for accurate assessment of the mean flow (12) which includes the determination of cross-sectional areas that may benefit from even smaller voxel sizes (13). Furthermore, high temporal resolution is needed due to fast cardiac motion and to resolve temporal detail, especially for the right coronary artery (RCA), which is twice as mobile as the left anterior descending artery (LAD) (14). According to Marcus et al. the temporal resolution τ should be ≤58ms and ≤23ms for LAD and RCA, respectively (15). Additionally, PC MRI requires two acquisitions with identical parameters except for the velocity encoding (VE) gradients to subtract out background phase (1). As a result, PC MRI is twice as time-consuming as anatomical cine scans, which is especially challenging when using breath-holding to suppress respiratory motion.

Initial coronary PC MRI utilized single breath-held Cartesian segmented spoiled gradient echo techniques enabling acquisitions with Δx of 1.4–1.6mm and a τ of 120–160ms (2,3). Keegan et al. pioneered the use of highly efficient spiral readouts to improve the temporal resolution (1618). More recently, 3T-scanners have been applied to take advantage of the increased signal-to-noise ratio (SNR) enabling higher spatial resolution as well (8,18,19). Nevertheless, state-of-the-art breath-held spiral techniques at 3 T focus on either high spatial resolution (8,19) or high temporal resolution (18). Brandts et al use Δx = 0.8mm and show accurate and reproducible measurement of RCA peak velocity and flow volumes (19). To achieve such a small voxel size they use a long, 26-ms spiral readout window, leading to a temporal resolution of 33ms and, hence, potentially motion blurring. Moreover, the long readouts may lead to off-resonance induced blurring (20). Keegan et al. trade spatial resolution for high temporal resolution (τ = 20ms) to assess temporal patterns of coronary blood flow throughout the cardiac cycle (18). To achieve such a high temporal resolution, they shorten the spiral readout to 12 ms, forcing a rather large voxel size Δx of 1.4mm.

To achieve both high temporal and high spatial resolution in a single breath-hold, the data acquisition needs to be accelerated. Acquisitions with non-Cartesian trajectories can be accelerated with parallel imaging (PI) (21,22) and additionally with compressed sensing (CS) (23). In cine MRI, the temporal domain is transform sparse and suitable for CS reconstruction (24,25). Besides sparsity, incoherent undersampling artifacts are also essential for CS reconstructions (23). Incoherence in temporal domain can be achieved using golden angle (GA) rotation of radial or spiral readouts (26,27). K-t sparse SENSE combines PI and CS in temporal dimension and acceleration factors of 8 for structural cines (28) or 6 for phase-contrast cine (29) have been shown using Cartesian sampling. The combination of k-t sparse SENSE and GA rotation of either radial or spiral sampling is feasible in cine MRI as well (3032).

However, while the acquisition of consecutively GA-rotated radial spokes or spiral arms offers flexibility in terms of temporal resolution versus spatial resolution in dynamic imaging with uniform coverage of k-space (26), when binning data into non-consecutively acquired cardiac frames, the k-space coverage may not be uniform (3336). This non-uniformity creates gaps in azimuthal distribution and strongly depends on the number of cardiac frames and the heartrate (HR), which is unpredictable especially for stress studies. To achieve higher k-space uniformity in cine MRI, Han et al. proposed the use of a segmented golden ratio approach (36) but requires prospective triggering and prevents retrospective data rejection due to motion or arrhythmias.

The objective of this work is to combine k-t sparse SENSE with GA rotated spiral readouts to achieve single breath-hold coronary PC MRI with both high spatial (0.8mm) and high temporal (~ 21ms) resolutions. GA Spiral k-t Sparse Parallel imaging (GASSP) (37) is combined with interleaved two-sided VE (38) and electrocardiogram (ECG) based retrospective cardiac gating (39). To reduce k-space gaps of binned cine data, two new methods are tested (40). First, the binning window is shifted by up to the duration of a cardiac frame, and the shift leading to the smallest k-space gap is picked for image reconstruction. Second, a triggered GA scheme is introduced that applies a new rotation angle to consecutive arms and the standard GA at the beginning of each heartbeat using prospective ECG-triggering. Compared to the segmented golden ratio approach (36), this triggered GA scheme is suitable for retrospective rejection of data corrupted by motion or other reasons. Monte-Carlo numerical simulations are performed to determine the optimal rotation angles in the triggered GA scheme, and to quantitatively evaluate the performance of the proposed methods to reduce k-space gaps. The proposed methods are tested at 3 T in healthy human subjects, validated against published methods (18,19), and intra-scan repeatability, inter-scan reproducibility, and intra-observer reproducibility are determined.

Methods

Coronary PC MRI with Standard GA Scheme

Coronary PC cine with both high temporal and high spatial resolutions was implemented by combining GASSP with interleaved two-sided VE (39), referred to here as standard GA scheme. Spiral arms were acquired in pairs of positive and negative VE with the same azimuthal angle, and then rotated by GA for the next pair (top row of Figure 1a). The GA was 137.508° (26,29). Spiral arm pairs were continuously acquired while the ECG signal was recorded for retrospective data binning.

Figure 1:

Figure 1:

Rotation angle and velocity encoding directions of the triggered GA scheme compared to the standard GA scheme (a) and shifted binning process (b). The velocity encoding direction in the standard GA scheme is toggled each TR while it is toggled every heartbeat in the triggered GA scheme. The rotation angle in the standard GA scheme is GA = 137.508° throughout, whereas in the triggered GA scheme it is θ = 83.8° except at the beginning of each heartbeat, when the rotation is set to GA/2 = 68.754° relative to the first spiral arm of the previous heartbeat (a). Both schemes are subject to shifted binning process (b). Retrospective binning with a fixed temporal resolution τ = TR was used and non-linear stretching (44) was applied due to RR-interval variations. Bins were shifted N times by temporal resolution τ divided by N (N=20) and for each shift the mean-max-gap was determined. The shift resulting in the smallest mean-max-gap was selected for image reconstruction.

To achieve high temporal resolution and to mitigate blurring effects due to B0 inhomogeneity the spiral readout window was set to ~14 ms leading to a repetition time (TR), and hence minimal τ, of <21 ms. The spatial resolution was set to 0.8 mm for a 350×350 mm2 field-of-view (FOV). This would require 34 spiral arms to obtain a fully sampled k-space, i.e. 68 heartbeats to acquire both VE datasets, which is too long for a single breath-hold scan. Therefore, the spiral trajectory was undersampled with a variable density pattern (41,42). A numerical algorithm (43) was used to optimally design the spiral trajectory so that the central 15% of k-space were fully sampled with 10 arms similar to (32). The outer 70% of k-space were uniformly undersampled by a factor of 4.6, and the undersampling factor was linearly increased in between. 500 spiral arm pairs (1000 arms) were acquired in total, leading to a scan time of ~21 s. Maximum encoded velocity (Venc) was set to ±35cm/s. Other parameters of the two-dimensional (2D) gradient echo sequence were echo time (TE) = 3.3 – 3.5ms, 1-2-1 binomial spectral spatial water excitation to suppress fat, flip angle (FA) = 20°, no RF spoiling, and slice thickness (ST) = 8 mm.

Retrospective Gating with Shifted Binning

The continuously acquired spiral data were binned into cardiac frames based on the ECG signal using non-linear stretching (44). Positive and negative VE data were binned separately. The number of cardiac frames was determined by dividing the acquisition’s average RR-interval, the time between two consecutive R-waves, by the required temporal resolution τ (= TR <21 ms), and was, therefore, subject specific. As a result, subjects with lower HR had more cardiac frames and less data assigned to each cardiac frame (i.e. higher undersampling factor). Assuming uniformly sampled k-space, the undersampling factor was around 3.0 for HR = 80, but increased to 3.9 and 4.7 for HR = 60 and 50, respectively, but was higher in practice because binning leads to non-uniform undersampling, and hence, gaps in k-space (3336).

In this work, we tested two methods to reduce the k-space gaps. The first method, referred to as shifted binning, shifted the binning windows by up to the duration of τ, trying N = 20 equally spaced shifts and selecting the shift that minimized the largest gaps (Figure 1b). For this, the k-space gaps after binning were quantitatively evaluated after binning the azimuthal angle of each spiral readout to determine first the maximal angular spacing between the sorted spiral arms (max-gap) in each bin (for each cardiac frame and VE direction), and, second the average of the max-gaps (mean-max-gap) over all cardiac frames and both VE directions. This was repeated N times, each time shifting the binning windows by τ/N, leading to N different mean-max-gaps (Figure 1b). The shift resulting in the lowest mean-max-gap was eventually chosen to bin the data, and resulting images were labeled with “best shift”. For comparison, the shift associated with the largest mean-max-gap was applied as well, and labeled with “worst shift”.

Coronary PC MRI with triggered GA scheme

In addition to shifted binning, the triggered GA scheme was the second method to improve k-space uniformity. The triggered GA scheme uses two different rotation angles and prospective ECG-triggering to reset the rotation angle and to toggle the VE direction at the beginning of each heartbeat. As shown in the bottom row of Figure 1a, spiral arms with the same VE direction, either positive or negative, were continuously acquired within one heartbeat. The rotation angle between consecutive arms within a heartbeat was modified to θ = 83.8°, which was selected based on the numerical simulations below. When an R-wave was detected, the rotation angle was reset so that the first spiral arm in a new heartbeat was rotated by half the GA (68.754°) compared to the first spiral arm of the previous heartbeat. The following arms were rotated again by θ. Concurrently with the angle reset, the VE direction was inverted to toggle its direction every heartbeat. Hence, the first spiral arms of heartbeats with the same VE direction were rotated by the GA.

1000 spiral arms were acquired in total for the triggered GA scheme, but due to varying RR-intervals, the total number of spiral arms in each VE direction was usually slightly different than 500. All other acquisition parameters of the triggered GA scheme were identical to standard GA scheme including the shifted binning gating process.

Simulation of k-space gap after binning

A Monte-Carlo numerical simulation was performed to determine the optimal rotation angle θ in the triggered GA scheme, and to quantitatively evaluate the performance of the shifted binning method and the triggered GA scheme. To quantify k-space uniformity, both mean-max-gap and the largest max-gap (max-max-gap) were determined and called ‘one-frame mean-max-gap’ and ‘one-frame max-max-gap’, respectively. Additionally, because the reconstruction applies temporal sparsity constraints, k-space gaps in one cardiac frame can be compensated for during the reconstruction by spiral arms from neighboring frames. For this, spiral arms from consecutive frames should be located at the center of the largest gap, which can be achieved if θ ≈ max-max-gap / 2. Therefore, θ was chosen to minimize the absolute of (max-max-gap / 2) - θ. To test if the triggered GA scheme indeed reduces gaps that overlap in consecutive cardiac frames, the mean-max-gap and max-max-gap were also calculated after combining the arms of three consecutive frames, and called ‘three-frame mean-max-gap’ and ‘three-frame max-max-gap’, respectively.

The simulation covers a wide range of HR from 40 to 100 beats per minute with a step size of 1 beat per minute. TR = 20.65 ms was used, leading to a duration of 20.65 s for 1000 spiral arms. For each HR, a series of RR-intervals were simulated to cover the 20.65 s duration using a normal distribution with a mean value of the given RR-interval and a standard deviation (SD) of 50 ms (45). A series of spiral rotation angles were generated using standard or triggered GA scheme. These angles were then binned with shifted binning, and mean-max-gap and max-max-gap determined. The simulation was repeated 200 times for each HR and the corresponding values averaged. For the triggered GA scheme, these simulations were repeated for θ ranging from 1° – 160° in steps of 3.6°. For the chosen θ (83.8°), the best shift was compared to the worst shift, and the triggered GA scheme to the standard GA scheme.

In vivo experiments and reference acquisitions

All human studies were approved by the Johns Hopkins School of Medicine Institutional Review Board and informed, written consent was obtained from all study subjects. In vivo studies were conducted on 8 healthy human subjects with no history of heart disease (37±7 years old, 1 female). MR studies were carried out on a 3T scanner (Achieva, Philips Healthcare, Best, The Netherlands) using a 32-channel cardiac receive coil and subject-specific shimming of the excitation radiofrequency field (46).

All 2D coronary PC scans were acquired within a single end-expiratory breath-hold and planned orthogonally to either proximal or mid segments of the coronary arteries that was straight for at least 20 mm. To facilitate careful planning of the image orientations, targeted 3D DIXON scans were performed to provide water and fat images of both RCA and LAD in 3D double oblique views (47,48) with TR/TE1/TE2 = 5.6–6.1/1.9–2.2/3.6–3.9 ms, FOV = 200×200×32 mm3, acquired voxel size = 1×0.85×3 mm3, and scan duration of ~1.5–6 min depending on the efficiency of the respiratory navigator. These targeted views were defined using the three-point planning tool on the target vessel of a whole heart coronary angiography scan with intermediate resolution acquired in ~1 minute. All scans for image planning were gated by respiratory navigator-echo and ECG-triggered to mid-diastole.

The standard and triggered GA scheme were compared to existing methods with either high spatial but low temporal resolution (8,19) (the ‘high-spatial’ sequence), or high temporal but low spatial resolution (18) (the ‘high-temporal’ sequence).

The high-spatial sequence was ECG-triggered and acquired a fixed (depending on the patient’s heartrate) number of arms with the same spiral rotation angle in each heartbeat. VE direction was inverted every heartbeat, and 10 spiral arms were acquired for each VE, leading to a total scan time of 20 heartbeats. High in plane spatial resolution of 0.8mm led to long spiral readouts of ~33 ms and long TR of ~40 ms. Other acquisition parameters were identical to standard GA sequence except for the FOV = 250×250 mm2. Because of ECG-triggering, the last end diastolic cardiac frames were missed as the sequence was interrupted to wait for and detect the next R-wave. Images and velocity maps were generated on the vendor-provided platform.

In the high-temporal sequence, VE direction was also inverted every heartbeat, with positive and negative arms acquired in 13 heartbeats each. With a high temporal resolution of TR = ~20 ms (spiral readout = ~13 ms), the spatial resolution was sacrificed to 1.4 mm to fully cover the k-space with 13 arms. Retrospective ECG-gating was used to bin the data into cardiac frames. Other acquisition parameters were identical to the standard GA sequence. Images were reconstructed off-line.

To determine inter-scan reproducibility of the triggered GA sequence, the subjects were removed from the scanner and given a 10-minutes break, before repeating the examination. For 4 of the subjects, the triggered GA sequences were also repeated immediately without repositioning to evaluate the intra-scan repeatability.

Image reconstruction and Flow analysis

Image reconstruction and off-resonance deblurring (except for the high-spatial sequence) were implemented in the graphical programming interface (49). Retrospective binning of k-space was performed with shifted binning and data were reconstructed by GASSP (37), minimizing parallel imaging data consistency and temporal total variation (TV) sparsity constraints (30). In this work, the original GASSP reconstruction for structural CINE images was adapted for PC by joint reconstruction of both VE steps with an additional TV sparsity constraint along the flow dimension. The flow sparsity constraint was only applied to the magnitude of the signal to preserve the phase information of each VE direction. The weight was set to 0.03 for the temporal sparsity constraint, and 0.015 for the additional flow sparsity constraint.

Flow-compensated images were generated by geometric mean, by which positive and negative VE phases were canceled by complex multiplication. Flow-compensated images were utilized to determine the deblurring frequency at the location of the coronary artery (50), which was subsequently used to deblur the original two VE images I+ and I. The deblurred VE images were used to compute deblurred flow-compensated images for vessel segmentation and deblurred velocity maps for velocity quantification. The velocity maps were computed by phase-difference with pixel-wise conjugated multiplication (20)

V=Vencπarg(I+I*). (1)

Here V is the velocity and arg is the argument of complex values.

A semi-automatic flow analysis software was implemented in MATLAB (Mathworks, Natick, MA, USA) that included vessel segmentation and correction for background phase and through-plane motion. All images were Fourier-interpolated to 0.1 mm spatial resolution for more precise segmentation (13). First, the vessel area Av was semi-automatically computed from three consecutive end-diastolic frames by full-width-half-maximum thresholding. Second, the vessel in each cardiac frame was segmented with a circular region of interest (ROI) with the same area Av, by clicking the center of the interpolated vessel as recommended by Brandts et al (19). Another circular ROI, the myocardial ROI with the same area Av was manually selected on surrounding myocardial tissue to correct for background phase and through-plane motion (7,18,51), and the mean velocity in myocardial ROI was subtracted from the velocity in the vessel ROI.

Eleven metrics were determined. The vessel area Av was determined in the first step of the segmentation. Eight velocity metrics were measured: peak systolic and diastolic velocity (PSV, PDV) together with the trigger delays to those peaks (TSV, TDV); and those velocity metrics were determined in two ways, mean (mean) and maximum (max) velocity spatially within the vessel ROI. And two flow metrics were measured: product of mean velocity and Av integrated over all cardiac frames as flow in [ml/cycle] or flow rate in [ml/min].

All datasets were tested for normal distribution using the Kolmogorov-Smirnov test. To limit the number of statistical tests, only four of the metrics (mean PSV, mean PDV, flow rate, and Av) acquired with the proposed triggered GA scheme were compared to both high-spatial and high-temporal sequences using paired Student’s t-test with a modified Bonferroni correction to account for the eight tests. Additionally, Bland-Altman analysis was performed for the same four metrics.

The intra-scan repeatability, inter-scan reproducibility, and intra-observer reproducibility were determined by the intraclass correlation coefficient (ICC) for absolute agreement. For intra-scan repeatability, Bland-Altman analysis was performed as well. To test the intra-observer agreement, the same observer repeated the flow analysis at least one month after the first observation, whereas the two analyses of the intra-scan repeatability and inter-scan reproducibility were performed on the same day. A P-value smaller than 0.05 was considered significant for all tests.

Results

The simulation results demonstrate the dependence between rotation angle θ of the triggered GA scheme, RR-interval, and mean-max-gap and max-max-gap, respectively (Figure 2). The one-frame mean-max-gap increases as functions of both RR-interval and θ (Figure 2a). A similar behavior but with larger gaps is seen for one-frame max-max-gap (Figure 2b). The |one-frame max-max-gap / 2 – θ| averaged over the range of simulated HR in Figure 2c is minimized for θ = 83.8°. Therefore, the rotation angle θ in the triggered GA scheme was chosen to be 83.8°. The three-frame mean-max-gap and three-frame max-max-gap at θ = 83.8° are smaller than even the best one-frame gaps, and are also smaller than the three-frame gaps for smaller θ (Figures 2d and 2e).

Figure 2:

Figure 2:

Surface plots of mean-max-gap (a, d) and max-max-gap (b, e) of the proposed triggered GA scheme with best binning shift were determined for a range of RR-intervals and rotation angles θ using Monte Carlo simulations. The gaps were determined for a single cardiac frame (a, b) or for three consecutive frames (d, e). In (c), the minimum of |max-max-gaps/2-θ| averaged over all heartrates (HR) is used to determine the rotation angle θ = 83.8° that ensures that neighboring cardiac frames have k-space data in the center of the largest gap which is beneficial for reconstruction with temporal sparsity constraints. The purple line in the surface plots indicates the chosen θ.

Figures 3a and 3b illustrate the simulated one-frame and three-frame mean-max-gaps, respectively, as a function of the mean RR-interval of both standard and triggered GA schemes with both best and worst shift. As the RR-intervals increase, larger gaps are observed. By using shifted binning, the one-frame mean-max-gaps of the best shift dropped about 14°–22° for triggered GA and 14°–27° for standard GA scheme (Figure 3a). The one-frame and three-frame mean-max-gaps of the triggered GA scheme increase approximately linearly with the mean RR-interval (608–1503ms) whereas the ones of the standard GA scheme exhibit an additional modulation pattern. The shifted binning has little effect on the three-frame mean-max-gaps (Figure 3b). The three-frame mean-max-gaps of the standard GA scheme is larger than those of the triggered GA scheme for most RR-intervals and the modulation pattern is even stronger than the one-frame mean-max-gaps.

Figure 3:

Figure 3:

One-frame mean-max-gap (a, c) and three-frame mean-max-gap (b, d) as a function of RR-interval of simulated datasets (a, b) and in-vivo datasets (c, d) for both standard GA and triggered GA schemes with both best and worst binning shift. Each marker in (a) and (b) is an average of 200 repeated simulations. In (c) and (d), each data point of the standard GA scan (purple rectangles and blue circles) is from a single scan, while the points of the triggered GA data (yellow crosses and red triangles) are averages of either 2 (inter-scan) or 3 (intra- and inter-scan) repetitions. In the simulations, the one-frame and three-frame mean-max-gaps of the triggered GA scheme are approximately linear whereas the ones of the standard GA scheme exhibit an additional modulation which is magnified in the three-frame case. Mean and standard deviation (error bars) of in-vivo experiments (c, d) are demonstrated to the right of their respective graphs. T-tests show significant reduction of the gaps for triggered GA scheme with best binning shift compared to standard GA scheme with best binning shift.

Complete datasets of both RCA and LAD were obtained from all 8 subjects. However, the RCA of one subject could not be visualized throughout the cardiac cycle (Supporting Information Figure S1), leading to a total of n = 15 analyzed vessels. Off-line image reconstruction took ~24 min per dataset. All metrics among the 15 vessels were normally distributed.

Figures 3c and 3d illustrate the one-frame and three-frame mean-max-gaps, respectively, measured in human subjects. In Figure 3c, the one-frame mean-max-gaps with best shift are smaller compared to those with worst shift. The mean-max-gaps averaged among all cases using the triggered GA scheme (73.6°±13.3°, best shift) are significantly smaller (t-test P=0.03) than those of the standard GA scheme (80.6°±18.0°, best shift). When combining 3 consecutive frames (Figure 3d), some large gaps were observed in the scatter plot of the standard GA scheme at RR-interval of around 872 ms and 916 ms, which is consistent with the simulation results in Figure 3b. As a result, the three-frame average mean-max-gaps of the standard GA scheme (42.8°±13.7°, best shift) is significantly larger (t-test P=0.04) than the one of the triggered GA scheme (35.3°±3.6°, best shift). The difference is especially large for the SD.

Figure 4 shows an example dataset acquired with both GA schemes and worst and best shifts. By using best shift, the mean-max-gaps were reduced by 21° and 19° for standard and triggered GA scheme, respectively. With best shift, the triggered GA scheme has lower mean-max-gaps (−5.5°) compared to the standard GA scheme. Swirling artifacts were observed in the right ventricle and the left ventricular wall (yellow arrows) in the standard GA images, for both VE and flow-compensated magnitudes, but not in the triggered GA scheme. Additional examples comparing standard and triggered GA schemes are provided as videos for both RCA (Supporting Information Video S1) and LAD (Supporting Information Video S2) demonstrating reduced artifacts with the triggered GA scheme. Supporting Information Video S3 shows magnitude and velocity cine videos acquired with the proposed triggered GA scheme of all 15 coronary arteries.

Figure 4:

Figure 4:

Representative LAD PC images of a 39-year-old male showing gap angle and images artifact reduction by applying triggered GA scheme and shifted binning. The max-gaps of the spiral arms are visualized by two black arms form the origin to the end of the corresponding spiral arms, with the angle labeled in degrees at the left top corner of each k-space. The LAD (red arrow) is visualized in cross-sectional view above the left ventricle (LV). Zoomed insets of the LAD are shown at the right-bottom corner in each image. Artifacts in the right ventricle (RV) and LV wall (yellow arrows) in the images acquired with the standard GA scheme are substantially reduced in the proposed triggered GA scheme.

Figure 5 shows an example dataset of a mid-diastolic frame acquired and reconstructed with all techniques. The flow compensated image of the high-spatial sequence demonstrated in Figure 5a suffers from aliasing artifacts (yellow arrow). The high-temporal sequence, on the other hand, shows pixelated pattern on the RCA due to low spatial resolution. The velocity map of the triggered GA method with best binning shift (red frame) has excellent quality across the heart and appears to have less noise / artifacts compared to other methods though the differences are only subtle. Flow rate curves in Figure 5b and 5c agree well across sequences and shifts. Of note, in this case though not on average (Table 1), the high-spatial sequence underestimates PSV by about half because of the lower temporal resolution.

Figure 5:

Figure 5:

Example PC data of the RCA (red arrow) of a 20-year old male acquired and reconstructed with all the different techniques applied in this work. Magnitude of flow compensated images and velocity maps of a mid-diastolic cardiac frame are shown in (a) and flow rate curves in (b-d). Aliasing artifacts (yellow arrow) are seen in the magnitude images of the high-spatial sequence. Images acquired with the proposed triggered GA method with best binning shift and combined GASSP reconstruction are highlighted with a red frame. The flow rate curves from the four different acquisition sequences (b) and using best and worst binning shift (c), generally agree well with each other. RA, right atrium and RV, right ventricle.

Table 1:

Coronary flow velocity metrics acquired with reference and proposed methods.

RCA (n=7) High-spatial High-temporal Standard GA Triggered GA
PSV max (cm/s) 21.2 ± 5.4 18.8 ± 4.7 18.9 ± 4.2 19.3 ± 3.8
TSV max (ms) 87.9 ± 40.0 65.9 ± 9.7 51.2 ± 20.6 69.8 ± 9.3
PDV max (cm/s) 23.7 ± 5.1 20.5 ± 4.5 19.8 ± 5.8 19.0 ± 5.9
TDV max (ms) 400.1 ± 35.5 398.1 ± 42.0 386.8 ± 29.0 394.5 ± 33.4
PSV mean (cm/s) 15.7 ± 5.4 15.8 ± 3.8 14.5 ± 4.0 14.4 ± 3.4
TSV mean (ms) 76.4 ± 19.4 74.4 ± 29.9 51.2 ± 20.6 69.8 ± 9.3
PDV mean (cm/s) 15.8 ± 4.8 17.8 ± 4.1 14.6 ± 5.4 14.3 ± 5.2
TDV mean (ms) 416.9 ± 57.0 395.2 ± 35.3 381.0 ± 41.7 391.5 ± 34.0
flow (ml/cycle) 0.69 ± 0.18* 0.76 ± 0.13 0.75 ± 0.26 0.82 ± 0.18
flow rate (ml/min) 48.8 ± 16.4* 52.8 ± 12.7 52.5 ± 19.3 55.7 ± 15.4
area (mm2) 9.1 ± 3.0 9.1 ± 1.9 11.6 ± 4.3 12.4 ± 3.7
LAD (n=8) High-spatial High-temporal Standard GA Triggered GA
PSV max (cm/s) 6.3 ± 4.1 5.0 ± 5.4 9.1 ± 3.6 7.8 ± 6.8
TSV max (ms) 137.8 ± 72.9 130.1 ± 65.9 118.5 ± 59.5 106.6 ± 84.4
PDV max (cm/s) 27.1 ± 8.1 22.5 ± 5.4 23.2 ± 6.4 21.4 ± 5.5
TDV max (ms) 451.4 ± 57.9 435.2 ± 56.3 432.5 ± 55.4 428.4 ± 51.5
PSV mean (cm/s) 2.5 ± 2.2 3.9 ± 2.7 3.8 ± 1.6 5.0 ± 4.1
TSV mean (ms) 142.8 ± 42.0 105.1 ± 49.9 115.8 ± 66.9 96.2 ± 69.9
PDV mean (cm/s) 16.7 ± 6.1 17.4 ± 4.4 15.4 ± 4.6 14.0 ± 3.7
TDV mean (ms) 436.4 ± 41.0 437.7 ± 47.2 440.4 ± 28.7 438.7 ± 45.2
flow (ml/cycle) 0.41 ± 0.16* 0.44 ± 0.15 0.44 ± 0.12 0.48 ± 0.16
flow rate (ml/min) 28.3 ± 12.0* 31.1 ± 14.4 29.9 ± 9.7 33.8 ± 14.3
area (mm2) 8.4 ± 2.3 9.1 ± 2.5 10.1 ± 2.2 11.1 ± 3.5

Flow velocity metrics are in mean ± standard deviation.

Abbreviations: PSV = peak systolic velocity, TSV = time to PSV, PDV = peak diastolic velocity, TDV=time to PDV, max = maximum velocity in vessel ROI, mean = average velocity in vessel ROI.

*

Due to prospective triggering used in high-spatial sequences, flow data was calculated based on incomplete cardiac cycle.

Both standard GA and triggered GA use the best binning shift.

Figure 6 demonstrates the agreement of flow rate curves of the repeated scans using the triggered GA scheme with best binning shift of all 15 coronary arteries. The 8 intra-scan repeats all show excellent agreement with the initial scan. The inter-scan repeats still agreed well in most cases.

Figure 6:

Figure 6:

Flow rate curves of the repeated scans acquired with the triggered GA scheme and reconstructed with the best binning shift and combined reconstruction of 7 RCA and 8 LAD cases. All 15 cases were repeated after repositioning while 4 RCA and 4 LAD cases were repeated without repositioning.

The velocity and flow metrics for all 4 techniques are shown in Table 1, separately averaged among 7 RCA and 8 LAD scans. After combining all 15 vessels, Supporting Information Figure S2 shows Bland-Altman analysis of mean PSV, mean PDV, flow rate, and Av comparing the proposed triggered GA scheme with the high-spatial and high-temporal reference methods. There is a high agreement for velocity and flow metrics, validating the new technique. Bonferroni corrected t-tests show that the area measured with triggered GA scheme is significantly larger than with both high-spatial and high-temporal sequences, and PDV is significantly lower compared to high-temporal sequence.

Table 2 demonstrates the results of intra-scan repeatability and inter-scan reproducibility. Spatial max PDV, spatial mean PSV and PDV, flow metrics and area are highly correlated (ICC≥0.97, P<0.01) for intra-scan repeats. Spatial max PSV is also well correlated (ICC=0.92, P<0.01). TSV and TDV are not correlated (−0.12≤ICC≤0.35, 0.16<P<0.62). Figure 7 demonstrates the high intra-scan repeatability of mean PSV, mean PDV, flow rate, and Av metrics using Bland-Altman analysis. For inter-scan reproducibility, the correlation of flow rate, area, PSV and PDV were still high and statistically significant (0.78≤ICC≤0.91, P<0.01).

Table 2:

Intra-scan repeatability and inter-scan reproducibility of coronary flow velocity metrics using proposed triggered GA scheme with shifted binning.

Intra-Scan Repeatability (n=8) Inter-Scan Reproducibility (n=15)
First Scan Repeat ICC P First Scan Repeat ICC P
PSV max (cm/s) 11.8 ± 9.8 13.6 ± 7.7 0.92 <0.01 13.1 ± 8.0 12.6 ± 8.0 0.88 <0.01
TSV max (ms) 79.7 ± 76.9 53.0 ± 10.8 −0.11 0.609 89.4 ± 62.9 89.6 ± 65.8 0.52 0.018
PDV max (cm/s) 22.3 ± 5.7 21.6 ± 5.4 0.98 <0.01 20.3 ± 5.6 20.9 ± 5.8 0.81 <0.01
TDV max (ms) 394.8 ± 35.7 398.6 ± 29.5 0.35 0.164 412.6 ± 45.9 437.7 ± 98.5 0.41 0.054
PSV mean (cm/s) 8.9 ± 7.2 9.5 ± 7.2 0.99 <0.01 9.4 ± 6.1 8.5 ± 6.8 0.84 <0.01
TSV mean (ms) 84.7 ± 66.5 53.0 ± 10.8 −0.12 0.619 83.9 ± 51.6 77.5 ± 43.8 0.66 <0.01
PDV mean (cm/s) 15.3 ± 4.8 15.2 ± 4.7 0.99 <0.01 14.2 ± 4.3 14.4 ± 4.5 0.78 <0.01
TDV mean (ms) 400.0 ± 35.0 393.5 ± 25.0 0.33 0.176 416.7 ± 45.9 440.4 ± 95.2 0.36 0.079
flow (ml/cycle) 0.65 ± 0.29 0.65 ± 0.29 0.98 <0.01 0.64 ± 0.24 0.58 ± 0.24 0.86 <0.01
flow rate (ml/min) 48.0 ± 22.6 48.2 ± 21.5 0.99 <0.01 44.0 ± 18.2 40.6 ± 18.8 0.91 <0.01
area (mm2) 12.2 ± 3.7 12.2 ± 4.1 0.97 <0.01 11.7 ± 3.5 11.1 ± 3.1 0.85 <0.01

Flow velocity metrics are in mean ± standard deviation.

Abbreviations: PSV = peak systolic velocity, TSV = time to PSV, PDV = peak diastolic velocity, TDV=time to PDV, max = maximum velocity in vessel ROI, mean = average velocity in vessel ROI, ICC = intraclass correlation coefficients for absolute agreement

Figure 7:

Figure 7:

Bland Altman analysis of the intra-scan repeatability for PSV (a), PDV (b), coronary blood volume (c), and cross-sectional vessel area (d).

The intra-observer reproducibility is demonstrated in Table 3. Area analysis correlation (ICC=0.98, P<0.01) is very high. Also the flow (ICC=0.91, P<0.01), flow rate (ICC=0.94, P<0.01), and PSV and PDV (ICC≥0.94, P<0.01) correlate very well. The correlation of TDV and TSV are lower, especially for TSV, but are still significant P≤0.01.

Table 3:

Intra-observer reproducibility of coronary flow velocity metrics using proposed triggered GA scheme with shifted binning.

Intra-Observer Reproducibility (*n=38,18 RCA and 20 LAD)
First Observation Second Observation ICC P Value
PSV max (cm/s) 13.0 ± 7.7 13.2 ± 9.0 0.94 <0.01
TSV max (ms) 81.8 ± 58.2 95.8 ± 79.0 0.38 <0.01
PDV max (cm/s) 20.8 ± 5.5 21.1 ± 5.4 0.98 <0.01
TDV max (ms) 419.6 ± 69.8 414.2 ± 71.9 0.92 <0.01
PSV mean (cm/s) 9.1 ± 6.4 9.4 ± 7.1 0.96 <0.01
TSV mean (ms) 74.9 ± 43.5 90.8 ± 72.9 0.37 =0.01
PDV mean (cm/s) 14.5 ± 4.4 14.7 ± 4.1 0.96 <0.01
TDV mean (ms) 421.2 ± 68.3 414.2 ± 72.4 0.88 <0.01
flow (ml/cycle) 0.62 ± 0.25 0.64 ± 0.31 0.91 <0.01
flow rate (ml/min) 43.5 ± 18.9 44.6 ± 21.4 0.94 <0.01
area (mm2) 11.6 ± 3.4 11.3 ± 3.3 0.98 <0.01

Flow velocity metrics are in mean ± standard deviation.

Abbreviations: PSV = peak systolic velocity, TSV = time to PSV, PDV = peak diastolic velocity, TDV = time to PDV, max = maximum velocity in vessel ROI, mean = average velocity in vessel ROI, ICC = intra-class correlation coefficients for absolute agreement

*

18 RCA scans include 7 first scans, 4 intra-scan repeats without repositioning, and 7 inter-scan repeats after repositioning. 20 LAD scans include 8 first scans, 4 intra-scan repeats without repositioning, and 8 inter-scan repeats after repositioning

Discussion

The proposed triggered GA method combines GASSP and shifted binning to enable single breath-hold coronary PC MRI with both high spatial (0.8mm) and high temporal (<21ms) resolutions. Feasibility of the technique has been demonstrated in 8 healthy subjects leading to high image quality and highly reproducible quantitative coronary flow measures. The method was validated against two previously published methods with either high spatial or high temporal resolution, and the flow rate and PSV agreed well.

However, the vessel areas were measured larger with the proposed method than either of the other two, and PDV was measured lower compared to the high-temporal sequence. With regards to the area, it has been shown that a larger voxel size, such as in the high-temporal sequence, in combination with Fourier interpolation (used to improve precision) leads to area measurement underestimation (13). Therefore, the larger area compared to the high-temporal sequence was expected. The similarly small area measured with the high-spatial sequence was not expected, and this may be due to remaining image blurring as it was not possible to deblur the images reconstructed and processed on the vendor platform. Lower PDV compared to high-temporal sequence may be due to the aforementioned area underestimation of the high-temporal sequence. This is supported by the agreement in the flow metric where both vessel area and measured velocities are combined. Additionally, the presented values agree well with those in the original publications that presented the reference methods (18,19).

Intra-scan repeatability, inter-scan reproducibility, and intra-observer reproducibility were determined (Tables 13). Intra-scan repeatability of coronary flow rate, peak systolic velocity (PSV) and peak diastolic velocity (PDV, both using spatial mean) were highest with ICC = 0.99, comparing favorably to the high-temporal method that reported ICCs of 0.975, 0.985, and 0.97, respectively (18). Intra-scan repeatability is important when coronary flow/velocity are measured before and during an intervention such as when assessing coronary flow reserve or coronary endothelial function. Inter-scan reproducibility was determined after the participant was given a 10-minute break outside of the scanner, and coronary flow rate had the highest ICC (ICC = 0.91), while PSV and PDV correlated well with ICC from 0.78 to 0.88; similar to the high-spatial method that reported RCA ICCs of 0.89, 0.88, and 0.84, respectively. Intra-observer reproducibility was also very high with ICC ≥ 0.94 for cross-sectional area, PSV and PDV, and coronary flow rate. Interestingly, intra-scan repeatability was higher than the intra-observer reproducibility. This is likely because the intra-scan analysis was conducted on the same day allowing the operator to remember what cardiac frames were selected to determine the area, while the intra-observer analysis was conducted after at least 30 days.

Binning of GA rotated data into cardiac frames may cause large gaps in k-space, and two methods to reduce those have been tested here. First, shifted binning reduced the one-frame mean-max-gap by about 20° in agreement with numerical simulations. Second, the triggered GA scheme reduced the three-frame mean-max-gap compared to the standard GA scheme from 42.8°±13.7° to 35.3°±3.6°, and, hence, reduced the overlap of k-space gaps across adjacent cardiac frames. For this, the rotation angle θ = 83.8° was chosen based on numerical simulations, in order to compensate for missing data in k-space with data from neighboring cardiac frames acquired at this k-space position. Simulations show that the three-frame mean-max-gap and max-max-gap (Figure 2d and 2e) increase if θ is chosen smaller even though the one-frame mean-max-gap and max-max-gap could be reduced with smaller θ, especially for θ < 40°.

The presented simulation results are only valid for acquisitions with the same duration, temporal resolution, and TR but could easily be adjusted to evaluate k-space gaps of other acquisitions that bin GA-rotated readouts into cardiac frames ranging from 2D radial (31) to 5D flow (52) sequences.

Despite significantly reducing k-space gaps using the proposed shifted binning and triggered GA scheme, only minor artifacts were removed compared to the standard GA scheme without shifted binning (Figure 4). This indicates that k-t sparse reconstruction performs well even with larger gaps, potentially allowing to shorten the breath-hold duration in combination with the gap reduction in the future.

Because spectral-spatial pulses on our scanner use fly-back gradients (53), the slice-thickness is at least ~8-mm thick leading to a rather large slice-thickness vs. in-plane-resolution ratio of 10, requiring careful planning of the slice location in straight segments. Lotz et al have shown as long as motion during the cardiac cycle keeps the vessel within 15° of being orthogonal to the imaging plane the error in flow measures remains within 10% (54).

Conclusion

We developed and validated a novel single breath-hold coronary PC MRI technique with both high temporal (<21 ms) and spatial resolution (0.8 mm). To achieve the high resolutions, a GA rotated spiral acquisition was combined with k-t sparse SENSE reconstruction. Gaps in k-space, that occur after binning GA rotated data into cardiac frames, were reduced by using shifted binning and a triggered GA scheme, leading to a reduction of undersampling aliasing artifacts and improved image quality. PSV, PDV, coronary flow rate and cross-sectional area acquired with the proposed triggered GA scheme exhibited excellent intra-scan repeatability and high inter-scan and intra-observer reproducibility.

Supplementary Material

figS1

Supporting Information Figure S1: The RCA of one subject could not be visualized throughout the cardiac cycle and was excluded from analysis. A screenshot demonstrates the planning on the 3D double oblique DIXON image (left). The small caliber RCA was badly visualized and only clearly visible in the mid-section of the RCA where the PC MRI imaging plane was planned. On the right, velocity compensated PC MRI images of a systolic frame (top row) and a diastolic frame (bottom row) are shown for all methods (high spatial, high temporal, standard GA and triggered GA). The RCA cannot be observed in the systolic frames (purple insets) and its signal is low in the diastolic frames (purple arrows). All other image acquisitions were successful with both RCA and LAD in 7 out of 8 subjects, leading to a success rate of 94% (15 out of 16 vessels).

Supporting Information Figure S2: Bland Altman analysis of the validation of the proposed triggered GA scheme with best binning shift against the high-spatial (a-d) and the high-temporal (e-h) sequences for PSV (a, e), PDV (b, f), coronary blood volume (c, g), and cross-sectional vessel area (d, h).

videoS1

Supporting Information Video S1: RCA flow-compensated image and velocity map of a 29-year-old male, comparing standard GA to triggered GA.

Download video file (8.4MB, avi)
videoS2

Supporting Information Video S2: LAD flow-compensated image and velocity map of a 45-year-old male, comparing standard GA to triggered GA.

Download video file (8.8MB, avi)
videoS3

Supporting Information Video S3: Flow-compensated images and velocity maps of all 15 RCA and LAD arteries shown in the same order as the flow rate curves in Figure 6.

Download video file (15.9MB, avi)

Funding:

AHA 17SDG33671007 and the District of Columbia Women’s Board (DCWB), and NIH HL61912, HL120905, HL125059.

Footnotes

Data Availability Statement

To promote reproducible science, graphical programming interface (49) reconstruction code is available at https://github.com/jhu-cardiac-mri/triggeredGASSP/ (SHA-1: 86fee6aa3d7baf885c33dae6ad27ffcde0700750) including a subject-approved, anonymized dataset.

References

  • 1.Moran PR, Moran RA, Karstaedt N. Verification and evaluation of internal flow and motion. True magnetic resonance imaging by the phase gradient modulation method. Radiology 1985;154:433–441 doi: 10.1148/radiology.154.2.3966130. [DOI] [PubMed] [Google Scholar]
  • 2.Edelman RR, Manning WJ, Gervino E, Li W. Flow velocity quantification in human coronary arteries with fast, breath-hold MR angiography. J. Magn. Reson. Imaging 1993;3:699–703 doi: 10.1002/jmri.1880030503. [DOI] [PubMed] [Google Scholar]
  • 3.Keegan J, Firmin D, Gatehouse P, Longmore D. The application of breath hold phase velocity mapping techniques to the measurement of coronary artery blood flow velocity: Phantom data and initialin vivo results. Magn. Reson. Med 1994;31:526–536 doi: 10.1002/mrm.1910310509. [DOI] [PubMed] [Google Scholar]
  • 4.Gould KL, Lipscomb K, Hamilton GW. Physiologic basis for assessing critical coronary stenosis. The American Journal of Cardiology 1974;33:87–94 doi: 10.1016/0002-9149(74)90743-7. [DOI] [PubMed] [Google Scholar]
  • 5.Clarke GD, Eckels R, Chaney C, et al. Measurement of Absolute Epicardial Coronary Artery Flow and Flow Reserve With Breath-Hold Cine Phase-Contrast Magnetic Resonance Imaging. Circulation 1995;91:2627–2634 doi: 10.1161/01.CIR.91.10.2627. [DOI] [PubMed] [Google Scholar]
  • 6.Hundley WG, Lange RA, Clarke GD, et al. Assessment of Coronary Arterial Flow and Flow Reserve in Humans With Magnetic Resonance Imaging. Circulation 1996;93:1502–1508 doi: 10.1161/01.CIR.93.8.1502. [DOI] [PubMed] [Google Scholar]
  • 7.Sakuma H, Blake LM, Amidon TM, et al. Coronary flow reserve: noninvasive measurement in humans with breath-hold velocity-encoded cine MR imaging. Radiology 1996;198:745–750 doi: 10.1148/radiology.198.3.8628864. [DOI] [PubMed] [Google Scholar]
  • 8.Hays AG, Hirsch GA, Kelle S, Gerstenblith G, Weiss RG, Stuber M. Noninvasive visualization of coronary artery endothelial function in healthy subjects and in patients with coronary artery disease. J. Am. Coll. Cardiol 2010;56:1657–1665. [DOI] [PubMed] [Google Scholar]
  • 9.Hays AG, Iantorno M, Soleimanifard S, et al. Coronary vasomotor responses to isometric handgrip exercise are primarily mediated by nitric oxide: a noninvasive MRI test of coronary endothelial function. Am. J. Physiol. Heart Circ. Physiol 2015;308:H1343–1350. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Maroules CD, Chang AY, Kotys-Traughber MS, Peshock RM. Coronary Endothelial Dysfunction in Women With Type 2 Diabetes Measured by Coronary Phase Contrast Flow Velocity Magnetic Resonance Imaging: Journal of Investigative Medicine 2015;63:856–861 doi: 10.1097/JIM.0000000000000226. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Deng Z, Fan Z, Lee S-E, et al. Noninvasive measurement of pressure gradient across a coronary stenosis using phase contrast (PC)-MRI: A feasibility study: Pressure Gradient Measurement Across a Coronary Artery. Magnetic Resonance in Medicine 2017;77:529–537 doi: 10.1002/mrm.26579. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Hofman MBM, Visser FC, Van Rossum AC, Vink GQM, Sprenger M, Westerhof N. In Vivo Validation of Magnetic Resonance Blood Volume Flow Measurements with Limited Spatial Resolution in Small Vessels. Magn. Reson. Med 1995;33:778–784 doi: 10.1002/mrm.1910330606. [DOI] [PubMed] [Google Scholar]
  • 13.Schär M, Soleimanifard S, Bonanno G, Yerly J, Hays AG, Weiss RG. Precision and accuracy of cross-sectional area measurements used to measure coronary endothelial function with spiral MRI. Magnetic Resonance in Medicine 2019;81:291–302 doi: 10.1002/mrm.27384. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Hofman MBM, Wickline SA, Lorenz CH. Quantification of in-plane motion of the coronary arteries during the cardiac cycle: Implications for acquisition window duration for MR flow quantification. J. Magn. Reson. Imaging 1998;8:568–576 doi: 10.1002/jmri.1880080309. [DOI] [PubMed] [Google Scholar]
  • 15.Marcus JT, Smeenk HG, Kuijer JP, Van der Geest RJ, Heethaar RM, Van Rossum AC. Flow profiles in the left anterior descending and the right coronary artery assessed by MR velocity quantification: effects of through-plane and in-plane motion of the heart. J Comput Assist Tomogr 1999;23:567–576. [DOI] [PubMed] [Google Scholar]
  • 16.Keegan J, Gatehouse P, Yang GZ, Firmin D. Interleaved spiral cine coronary artery velocity mapping. Magn Reson Med 2000;43:787–792 doi: . [DOI] [PubMed] [Google Scholar]
  • 17.Keegan J, Gatehouse PD, Mohiaddin RH, Yang G-Z, Firmin DN. Comparison of spiral and FLASH phase velocity mapping, with and without breath-holding, for the assessment of left and right coronary artery blood flow velocity. J Magn Reson Imaging 2004;19:40–49 doi: 10.1002/jmri.10434. [DOI] [PubMed] [Google Scholar]
  • 18.Keegan J, Raphael CE, Parker K, et al. Validation of high temporal resolution spiral phase velocity mapping of temporal patterns of left and right coronary artery blood flow against Doppler guidewire. Journal of Cardiovascular Magnetic Resonance 2015;17 doi: 10.1186/s12968-015-0189-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Brandts A, Roes SD, Doornbos J, et al. Right coronary artery flow velocity and volume assessment with spiral K-space sampled breathhold velocity-encoded MRI at 3 tesla: Accuracy and reproducibility. Journal of Magnetic Resonance Imaging 2010;31:1215–1223 doi: 10.1002/jmri.22144. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Bernstein MA, King KF, Zhou ZJ. Handbook of MRI pulse sequences. Amsterdam ; Boston: Academic Press; 2004. [Google Scholar]
  • 21.Pruessmann KP, Weiger M, Börnert P, Boesiger P. Advances in sensitivity encoding with arbitrary k-space trajectories. Magn Reson Med 2001;46:638–651. [DOI] [PubMed] [Google Scholar]
  • 22.Heidemann RM, Griswold MA, Seiberlich N, et al. Direct parallel image reconstructions for spiral trajectories using GRAPPA. Magn. Reson. Med 2006;56:317–326 doi: 10.1002/mrm.20951. [DOI] [PubMed] [Google Scholar]
  • 23.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med 2007;58:1182–1195 doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
  • 24.Adluru G, Awate SP, Tasdizen T, Whitaker RT, DiBella EVR. Temporally constrained reconstruction of dynamic cardiac perfusion MRI. Magn. Reson. Med 2007;57:1027–1036 doi: 10.1002/mrm.21248. [DOI] [PubMed] [Google Scholar]
  • 25.Gamper U, Boesiger P, Kozerke S. Compressed sensing in dynamic MRI. Magn Reson Med 2008;59:365–373 doi: 10.1002/mrm.21477. [DOI] [PubMed] [Google Scholar]
  • 26.Winkelmann S, Schaeffter T, Koehler T, Eggers H, Doessel O. An optimal radial profile order based on the Golden Ratio for time-resolved MRI. IEEE Trans Med Imaging 2007;26:68–76 doi: 10.1109/TMI.2006.885337. [DOI] [PubMed] [Google Scholar]
  • 27.Chan RW, Ramsay EA, Cheung EY, Plewes DB. The influence of radial undersampling schemes on compressed sensing reconstruction in breast MRI. Magn. Reson. Med 2012;67:363–377 doi: 10.1002/mrm.23008. [DOI] [PubMed] [Google Scholar]
  • 28.Otazo R, Kim D, Axel L, Sodickson DK. Combination of compressed sensing and parallel imaging for highly accelerated first-pass cardiac perfusion MRI. Magn Reson Med 2010;64:767–776 doi: 10.1002/mrm.22463. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kim D, Dyvorne HA, Otazo R, Feng L, Sodickson DK, Lee VS. Accelerated phase-contrast cine MRI using k-t SPARSE-SENSE. Magn Reson Med 2012;67:1054–1064 doi: 10.1002/mrm.23088. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Feng L, Grimm R, Block KT, et al. Golden-angle radial sparse parallel MRI: combination of compressed sensing, parallel imaging, and golden-angle radial sampling for fast and flexible dynamic volumetric MRI. Magn Reson Med 2014;72:707–717 doi: 10.1002/mrm.24980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Yerly J, Ginami G, Nordio G, et al. Coronary endothelial function assessment using self-gated cardiac cine MRI and k - t sparse SENSE: Coronary Endothelial Function Assessment Using Self-Gated Cardiac Cine MRI. Magn. Reson. Med 2016;76:1443–1454 doi: 10.1002/mrm.26050. [DOI] [PubMed] [Google Scholar]
  • 32.Bonanno G, Hays AG, Weiss RG, Schär M. Self-gated golden angle spiral cine MRI for coronary endothelial function assessment: Self-Gated Spiral Cine MRI. Magnetic Resonance in Medicine 2018;80:560–570 doi: 10.1002/mrm.27060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Krämer M, Herrmann K-H, Biermann J, Reichenbach JR. Retrospective reconstruction of cardiac cine images from golden-ratio radial MRI using one-dimensional navigators: 1D Navigator Golden-Ratio Radial Cine Imaging. Journal of Magnetic Resonance Imaging 2014;40:413–422 doi: 10.1002/jmri.24364. [DOI] [PubMed] [Google Scholar]
  • 34.Edelman RR, Giri S, Murphy IG, Flanagan O, Speier P, Koktzoglou I. Ungated radial quiescent-inflow single-shot (UnQISS) magnetic resonance angiography using optimized azimuthal equidistant projections: Ungated Quiescent-Inflow Single-Shot MRA. Magnetic Resonance in Medicine 2014;72:1522–1529 doi: 10.1002/mrm.25477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Kawaji K, Foppa M, Roujol S, Akçakaya M, Nezafat R. Whole Heart Coronary Imaging with Flexible Acquisition Window and Trigger Delay Guo Y, editor. PLOS ONE 2015;10:e0112020 doi: 10.1371/journal.pone.0112020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Han F, Zhou Z, Rapacchi S, Nguyen K-L, Finn JP, Hu P. Segmented golden ratio radial reordering with variable temporal resolution for dynamic cardiac MRI: Segmented Golden Ratio Radial Reordering for Cardiac MRI. Magnetic Resonance in Medicine 2016;76:94–103 doi: 10.1002/mrm.25861. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Schär M, Bonanno G, Hays AG, et al. Golden-Angle Spiral Sparse Parallel-Imaging for Coronary Lumen Area Measurements in Short Breath-Holds. In: Washington, D.C.; 2017. p. P321. [Google Scholar]
  • 38.Dumoulin CL, Hart HR. Magnetic resonance angiography. Radiology 1986;161:717–720 doi: 10.1148/radiology.161.3.3786721. [DOI] [PubMed] [Google Scholar]
  • 39.Zhu D, Bonanno G, Weiss RG, Schär M. Phase contrast coronary blood velocity mapping with both high temporal and spatial resolution using Golden Angle rotated Spiral k-t Sparse Parallel imaging (GASSP). In: Proc of 27th Meeting, ISMRM. Montreal, Canada; 2019. p. 1951. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Zhu D, Schär M. Reducing k-space gaps of golden angle rotated spiral MRI binned into cardiac frames applied to flow measurements of the coronary arteries. In: Proceedings of the 28th Annual Meeting of ISMRM. virtual; 2020. p. 2259. [Google Scholar]
  • 41.Spielman DM, Pauly JM, Meyer CH. Magnetic resonance fluoroscopy using spirals with variable sampling densities. Magn Reson Med 1995;34:388–394. [DOI] [PubMed] [Google Scholar]
  • 42.Tsai CM, Nishimura DG. Reduced aliasing artifacts using variable-density k-space sampling trajectories. Magn Reson Med 2000;43:452–458. [DOI] [PubMed] [Google Scholar]
  • 43.Pipe JG, Zwart NR. Spiral trajectory design: A flexible numerical algorithm and base analytical equations: Spiral Trajectory Design. Magnetic Resonance in Medicine 2014;71:278–285 doi: 10.1002/mrm.24675. [DOI] [PubMed] [Google Scholar]
  • 44.Feinstein JA, Epstein FH, Arai AE, et al. Using cardiac phase to order reconstruction (CAPTOR): A method to improve diastolic images. J. Magn. Reson. Imaging 1997;7:794–798 doi: 10.1002/jmri.1880070505. [DOI] [PubMed] [Google Scholar]
  • 45.Nunan D, Sandercock GRH, Brodie DA. A quantitative systematic review of normal values for short-term heart rate variability in healthy adults. Pacing Clin Electrophysiol 2010;33:1407–1417 doi: 10.1111/j.1540-8159.2010.02841.x. [DOI] [PubMed] [Google Scholar]
  • 46.Schär M, Ding H, Herzka DA. Improvement in B1+ homogeneity and average flip angle using dual-source parallel RF excitation for cardiac MRI in swine hearts. PLoS ONE 2015;10:e0139859. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Stuber M, Botnar RM, Danias PG, et al. Double-oblique free-breathing high resolution three-dimensional coronary magnetic resonance angiography. J. Am. Coll. Cardiol 1999;34:524–531. [DOI] [PubMed] [Google Scholar]
  • 48.Börnert P, Koken P, Nehrke K, Eggers H, Ostendorf P. Water/fat-resolved whole-heart Dixon coronary MRA: an initial comparison. Magn Reson Med 2014;71:156–163 doi: 10.1002/mrm.24648. [DOI] [PubMed] [Google Scholar]
  • 49.Zwart NR, Pipe JG. Graphical programming interface: A development environment for MRI methods. Magn. Reson. Med 2015;74:1449–1460 doi: 10.1002/mrm.25528. [DOI] [PubMed] [Google Scholar]
  • 50.Noll DC, Pauly JM, Meyer CH, Nishimura DG, Macovski A. Deblurring for non-2D Fourier transform magnetic resonance imaging. Magn Reson Med 1992;25:319–333. [DOI] [PubMed] [Google Scholar]
  • 51.Nagel E, Bornstedt A, Hug J, Schnackenburg B, Wellnhofer E, Fleck E. Noninvasive determination of coronary blood flow velocity with magnetic resonance imaging: Comparison of breath-hold and navigator techniques with intravascular ultrasound. Magnetic Resonance in Medicine 1999;41:544–549. [DOI] [PubMed] [Google Scholar]
  • 52.Ma LE, Yerly J, Piccini D, et al. 5D Flow MRI: A Fully Self-gated, Free-running Framework for Cardiac and Respiratory Motion-resolved 3D Hemodynamics. Radiol Cardiothorac Imaging 2020;2:e200219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Zur Y Design of improved spectral-spatial pulses for routine clinical use. Magnetic Resonance in Medicine 2000;43:410–420. [DOI] [PubMed] [Google Scholar]
  • 54.Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation. RadioGraphics 2002;22:651–671. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

figS1

Supporting Information Figure S1: The RCA of one subject could not be visualized throughout the cardiac cycle and was excluded from analysis. A screenshot demonstrates the planning on the 3D double oblique DIXON image (left). The small caliber RCA was badly visualized and only clearly visible in the mid-section of the RCA where the PC MRI imaging plane was planned. On the right, velocity compensated PC MRI images of a systolic frame (top row) and a diastolic frame (bottom row) are shown for all methods (high spatial, high temporal, standard GA and triggered GA). The RCA cannot be observed in the systolic frames (purple insets) and its signal is low in the diastolic frames (purple arrows). All other image acquisitions were successful with both RCA and LAD in 7 out of 8 subjects, leading to a success rate of 94% (15 out of 16 vessels).

Supporting Information Figure S2: Bland Altman analysis of the validation of the proposed triggered GA scheme with best binning shift against the high-spatial (a-d) and the high-temporal (e-h) sequences for PSV (a, e), PDV (b, f), coronary blood volume (c, g), and cross-sectional vessel area (d, h).

videoS1

Supporting Information Video S1: RCA flow-compensated image and velocity map of a 29-year-old male, comparing standard GA to triggered GA.

Download video file (8.4MB, avi)
videoS2

Supporting Information Video S2: LAD flow-compensated image and velocity map of a 45-year-old male, comparing standard GA to triggered GA.

Download video file (8.8MB, avi)
videoS3

Supporting Information Video S3: Flow-compensated images and velocity maps of all 15 RCA and LAD arteries shown in the same order as the flow rate curves in Figure 6.

Download video file (15.9MB, avi)

RESOURCES