Abstract
Objectives
Precise small lung nodule resection is challenging in minimally invasive thoracoscopic surgery. Various methods that help surgeons to locate the target nodule have been devised; however, the ideal way that satisfies the demand has not yet been realized. We have developed and applied a novel marking system to localize small lung nodules for the first time in humans.
Methods
A radiofrequency identification tag (1.8 mm in diameter and 7 mm in length) that can communicate with a wand-shaped antenna (10 mm in diameter) from the distance of 3 cm was prepared. The tag was delivered adjacent to a 7-mm subsolid nodule in the right lower lobe of a patient under cone beam computed tomography guidance and video-assisted thoracoscopic surgery wedge resection was subsequently performed.
Results
The delivery of the tag was smooth, and the tag was almost immediately detected by the antenna. Wedge resection was successfully performed with the guiding signal from the tag.
Conclusions
We have so far demonstrated that this technology could be applicable for small lung nodule detection in preclinical studies. In this first clinical experience, this system proved to provide accurate positional information of small lung nodules with depth.
Key Words: lung cancer, small nodule, minimally invasive surgery
Abbreviations and Acronyms: CT, computed tomography; RFID, radiofrequency identification
Central Message.
We report the first clinical case using a novel marking system that allows precise resection of small pulmonary nodules during thoracoscopic surgery.
Perspective.
Tis or T1mi peripheral lung adenocarcinoma has been defined and described in the 8th edition of the Union for International Cancer Committee TNM classification and these small and low-grade lung cancers are increasing. Under these circumstances, new techniques are required to enable precise resection with adequate margins in minimally invasive treatment.
See Commentaries on pages 305 and 307.
We have developed a new system using near-field communication technology that enables precise localization of small lung nodules. The system consists of 3 components: a small tag designed to be delivered adjacent to the target lesion through flexible bronchoscope, a detector that receives the signal from the tag, and a signal processing unit that identifies the signal from the tag. During resection, the surgeon can determine the distance between the tip of the detector and the tag by an audiovisual interface from the signal processing unit (Figure 1), and system monitor informs the surgeon by flicking the targeted tag that was previously registered in the system. The system was approved as class 3 medical device by Japanese Pharmaceuticals and Medical Devices Agency in July 2018.
Clinical Summary
A 75-year-old female patient with a 7-mm subsolid nodule in the basal segment of the right lower lobe was referred to our department (Figure 2). Periodical thin-slice computed tomography (CT) inspection confirmed the nodule had been gradually enlarging; thus, wedge resection was planned for her. She once had developed severe pneumonia in the past, and the flattened diaphragm shadow in her chest radiograph suggested some adhesion in the thoracic cavity. To localize the nodule in the thoracoscopic procedure, a novel radiofrequency identification (RFID) marking system was employed. In the hybrid operating theater, the patient was intubated with a single-lumen endotracheal tube under general anesthesia, and an RFID tag (1.8 mm in diameter and 7 mm in length) was delivered via a bronchoscope to the adjacent of the target nodule under fluoroscopic and virtual bronchoscopic guidance. Cone-beam CT confirmed that the tag was placed 10 mm caudal, 5 mm lateral to the target. After the endotracheal tube was changed to a double-lumen tube, the patient was placed in the left decubitus position, and three 10-mm ports were set in the 5th and 8th intercostal space. Whole lung adhesion to the thoracic wall was found. After we detached the adhesion, the wand-shaped locating detector (10 mm in diameter) was introduced into the thoracic cavity to receive the signal from the RFID tag. Detection of the signal was almost immediately achieved; then, successful wedge resection was performed with secure margin of 15 mm (Video 1). Afterward, pathologic examination confirmed that the target nodule was papillary adenocarcinoma (7 mm). A signed informed consent in publishing the result and the clinical course was obtained from the patient.
Discussion
Current guidelines for the incidentally detected pulmonary nodule recommend resection for suspicious pulmonary nodules ≥6 mm after a certain observation period.1 However, precise localization and resection of such small and subsolid or pure ground-glass lesion are extremely difficult without any means of preoperative marking. So far, various procedures have been attempted and applied, including placement of a hook wire and delivery of fiducials, coils, dyes, indocyanine green, contrast medium, barium, intraoperative ultrasound, and injection of radioactive tracers and their combinations through percutaneous and bronchoscopic approaches.2 The multitude of reports indicate that this is one of the issues to be addressed by the modern thoracic surgeon. Lesions just beneath the pleura, even pure-ground-glass ones, might be located and removed without preoperative marking, but still some measures would be useful in determining the precise resection area (or volume). Deeper lesions (2 cm) are always challenging for the thoracic surgeon, and some measures should be taken to localize the target; however, they are not always practical and may cause severe adverse event including air embolism. Among many marking measures, dye marking through a bronchoscope is well investigated and documented by Sato and colleagues. They conducted a multi-institutional prospective study investigating the efficacy of the dye marking method, which they reported in the Journal.3 They reported that insufficient dye marking was noted in 14% of patients, and the most common reason was central location, failure to show the deep margin, followed by anthracotic lung and emphysema. So far, we have demonstrated that our methodology using near-field communication could be applicable in the detection of small lung nodules and showed the deep margin precisely in preclinical studies.4,5 In these previous studies using a canine model, the tag delivered to the designated place was retained for months without dislocation, the use of a stapler caused no problems at the time of excision, and no remarkable adverse events, including pneumothorax, hemoptysis, and pneumonia, were observed.
In this first clinical experience, we were able to accurately localize the small lung nodule with this novel system despite the whole lung adhesion, which might have hindered visual localization with dye from the lung surface, and successfully performed wedge resection. This RFID lung marking system provides accurate positional information with depth because the tag was designed to be delivered and stay both at the subsegmental bronchus and into the lung parenchyma namely just beneath the pleura. Furthermore, since each tag has its own ID, real navigational surgery by placing multiple tags for sublobar resection is expected to be performed.5 Regarding tag placement before resection, we deliver the tag under cone-beam CT guidance, but a more accurate and simple delivery will be realized through multiple platforms, such as electromagnetic navigational bronchoscope and a robotic system, in the near future. We are applying this technique not only for wedge resection but also subsegmentectomy or segmentectomy and are planning to conduct a multicenter trial to prove efficacy and benefit for the patients with small lung nodules.
Conflict of Interest Statement
Drs Sato, Yutaka, and Date are receiving research support from HOGY Medical Inc, Tokyo, Japan. Dr Yutaka has an issued patent for RFID reader of positional localization. Dr Nakamura has reported no conflicts of interest.
The Journal policy requires editors and reviewers to disclose conflicts of interest and to decline handling or reviewing manuscripts for which they may have a conflict of interest. The editors and reviewers of this article have no conflicts of interest.
Footnotes
Accepted for the 100th Annual Meeting of The American Association for Thoracic Surgery.
Supplementary Data
References
- 1.MacMahon H., Naidich D.P., Goo J.M., Lee K.S., Leung A.N.C., Mayo J.R., et al. Guidelines for management of incidental pulmonary nodules detected on CT images: from the Fleischner Society 2017. Radiology. 2017;284:228–243. doi: 10.1148/radiol.2017161659. [DOI] [PubMed] [Google Scholar]
- 2.Velasquez R., Martin A., Abu Hishmeh M., DeLorenzo L., Dhillon S.S., Harris K. Placement of markers to assist minimally invasive resection of peripheral lung lesions. Ann Transl Med. 2019;7:360. doi: 10.21037/atm.2019.03.50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sato M., Kobayashi M., Kojima F., Tanaka F., Yanagiya M., Kosaka S., et al. Effect of virtual-assisted lung mapping in acquisition of surgical margins in sublobar lung resection. J Thorac Cardiovasc Surg. 2018;156:1691–1701.e5. doi: 10.1016/j.jtcvs.2018.05.122. [DOI] [PubMed] [Google Scholar]
- 4.Kojima F., Sato T., Takahata H., Okada M., Sugiura T., Oshiro O., et al. A novel surgical marking system for small peripheral lung nodules based on radio frequency identification technology: feasibility study in a canine model. J Thorac Cardiovasc Surg. 2014;147:1384–1389. doi: 10.1016/j.jtcvs.2013.05.048. [DOI] [PubMed] [Google Scholar]
- 5.Yutaka Y., Sato T., Matsushita K., Aiba H., Muranishi Y., Sakaguchi Y., et al. Three-dimensional navigation for thoracoscopic sublobar resection using a novel wireless marking system. Semin Thorac Cardiovasc Surg. 2018;30:230–237. doi: 10.1053/j.semtcvs.2018.03.001. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.