Commentary: Early failure of the Trifecta GT bioprosthesis: Innovation is not always progress

Jean Porterie 1, Dimitri Kalavrouziotis 1, Siamak Mohammadi 1,
PMCID: PMC8307754  PMID: 34317980

graphic file with name fx1.jpg

Jean Porterie, MD, Dimitri Kalavrouziotis, MD, FRCSC, and Siamak Mohammadi, MD, FRCSC

Central Message.

There is an urgent need to elucidate all possible mechanisms of early and late prosthetic valve failure, especially as treatment options for severe aortic stenosis continue to rapidly evolve.

See Article page 106.

Structural valve degeneration (SVD) is a major disadvantage of surgical aortic valve replacement (AVR) with a bioprosthesis. A recent single-center series demonstrated that the Trifecta valve (Abbott Vascular, Santa Clara, Calif) can be safely implanted in the aortic position with excellent hemodynamics and midterm durability,1 whereas others have shown unusually high early SVD rates with this prosthesis.2 The newer-generation Trifecta with Glide Technology (GT) was designed to improve these results, using novel antileaflet calcification and anti-support strut distortion technology.

In this issue of the Journal, Tchouta and colleagues3 reported 3 cases of early SVD after AVR with the Trifecta GT, for a cumulative incidence of 3.3% at 3.5 years among a population of 106 patients. Mean age at implantation was 56.7 years and all 3 patients had bicuspid aortic valve. The SVD occurred between 1.9 and 3.2 years after implantation, and all patients presented with aortic insufficiency and heart failure. There was no patient–prosthesis mismatch and endocarditis. All patients underwent redo AVR, and intraoperative findings were similar in the 3 cases, namely a detached prosthetic leaflet at the stent post between the noncoronary and right coronary cusps, without leaflet calcification. The authors postulated that the externally mounted leaflet design of the Trifecta GT and mechanical abrasion from the aortic wall may have contributed to the early failure of the prosthesis.

The report by Tchouta and colleagues3 suggests that biologic/immunologic factors may be less relevant in early SVD, whereas the mechanical/fluid dynamic properties directly associated with particular elements of the prosthetic valve design may be more important contributing factors.2,4 No prosthetic leaflet calcification was observed in any of the 3 cases. Conversely, the leaflet disinsertion found at the same location in all patients was also observed in a majority of the cases of SVD presenting with pure aortic insufficiency occurring on the first-generation Trifecta prosthesis.2 In vitro fluid dynamic studies have shown that the externally mounted design of the Trifecta valve is associated with favorable hemodynamic properties compared with other stented bioprostheses, in which leaflets are mounted internally, including a lower resistance to flow and a larger effective orifice area. However, a greater degree of prosthetic leaflet flutter was observed with the Trifecta prosthesis.5 The Mitroflow valve (LivaNova Group Inc, London, England), another bioprosthesis in which the leaflets are externally-mounted, is well-known to have high rates of accelerated SVD.6 Another hypothesis is that the asymmetrical aortic annulus of bicuspid aortic valve may have increased the likelihood of abnormal shear-stress patterns and may potentially explain the monomorphic nature of the observed prosthetic failures. The implant technique itself may have a major impact in the durability of prostheses with externally mounted leaflets, including oversizing, bending of the strut posts during hand-tying of sutures, or the use of suture-tying devices with rigid knot stabilizers. Although oversizing was unlikely to play a significant role in the SVD reported by Tchouta and colleagues,3 2 of the 3 patients required an aortic annulus enlargement, and the third was a female with small body surface area and aortic root diameter, suggesting anatomical conditions likely to favor strut distortion during implantation.

In summary, the findings by Tchouta and colleagues3 are concerning, especially given the current trend of extending bioprosthetic AVR to younger patients. Well-designed, multicenter studies are urgently needed to elucidate all possible mechanisms of early SVD, to allow for a fine-tuning of the surgical implant, and optimal selection of the appropriate prosthesis for each individual patient. The report by Tchouta and colleagues3 further highlights the fact that close longitudinal echocardiographic surveillance becomes paramount in an era in which the treatment of severe aortic stenosis is rapidly evolving.

Footnotes

Disclosures: Dr Porterie received a grant from Edwards Lifesciences for his fellowship training at the Quebec Heart and Lung Institute. All other authors reported no conflicts of interest.

The Journal policy requires editors and reviewers to disclose conflicts of interest and to decline handling or reviewing manuscripts for which they may have a conflict of interest. The editors and reviewers of this article have no conflicts of interest.

References

  • 1.Kilic A., Sultan I., Navid F., Aranda-Michel E., Chu D., Thoma F., et al. Trifecta aortic bioprosthesis: midterm results in 1,953 patients from a single center. Ann Thorac Surg. 2019;107:1356–1362. doi: 10.1016/j.athoracsur.2018.10.063. [DOI] [PubMed] [Google Scholar]
  • 2.Fukuhara S., Shiomi S., Yang B., Kim K., Bolling S.F., Haft J., et al. Early structural valve degeneration of trifecta bioprosthesis. Ann Thorac Surg. 2020;109:720–727. doi: 10.1016/j.athoracsur.2019.06.032. [DOI] [PubMed] [Google Scholar]
  • 3.Tchouta L., Liesman D., Kim K., Fukuhara S. Early failure of the Trifecta GT bioprostheses. J Thorac Cardiovasc Surg Tech. 2020;4:106–108. doi: 10.1016/j.xjtc.2020.08.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Gallo M., Demertzis S. Xenoantigens and structural valve degeneration of Trifecta bioprosthesis: connecting the dots. Ann Thorac Surg. 2020;110:343. doi: 10.1016/j.athoracsur.2019.10.054. [DOI] [PubMed] [Google Scholar]
  • 5.Marx P., Kowalczyk W., Demircioglu A., Shehada S.-E., Wendta H., Mourad F., et al. An in vitro comparison of flow dynamics of the Magna Ease and the Trifecta prostheses. Minim Invasive Ther Allied Technol. 2020;29:78–85. doi: 10.1080/13645706.2019.1586732. [DOI] [PubMed] [Google Scholar]
  • 6.Ius F., Schulz J., Roumieh M., Fleissner F., Ismail I., Tudorache I., et al. Long-term results of the Mitroflow aortic pericardial bioprosthesis in over 800 patients: limited durability and mechanisms of dysfunction. Eur J Cardiothorac Surg. 2017;52:264–271. doi: 10.1093/ejcts/ezx161. [DOI] [PubMed] [Google Scholar]

Articles from JTCVS Techniques are provided here courtesy of Elsevier

RESOURCES