A Quantitative ELISA Protocol for Detection of Specific Human IgG against the SARS-CoV-2 Spike Protein

Rémi Vernet 1,2,*, Emily Charrier 1,2,3, Julien Grogg 3, Nicolas Mach 1,2
Editor: Ralph A Tripp
PMCID: PMC8309958  PMID: 34358186

Abstract

Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has caused a worldwide pandemic with at least 3.8 million deaths to date. For that reason, finding an efficient vaccine for this virus quickly became a global priority. The majority of vaccines now marketed are based on the SARS-CoV-2 spike protein that has been described as the keystone for optimal immunization. In order to monitor SARS-CoV-2 spike-specific humoral responses generated by immunization or infection, we have developed a robust and reproducible enzyme-linked immunosorbent assay (ELISA) protocol. This protocol describes a method for quantitative detection of IgG antibodies against the SARS-CoV-2 spike protein using antigen-coated microtiter plates. Results showed that antibodies could be quantified between the range of 1.953 ng/mL to 500 ng/mL with limited inter- and intra-assay variability.

Keywords: SARS-CoV-2, COVID-19, ELISA, antibodies

1. Introduction

COVID-19, the disease caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), has triggered a worldwide pandemic with over 176 million confirmed cases globally and more than 3.8 million deaths as of June 2021 [1]. In response to this major public health crisis, global efforts have been dedicated to developing different SARS-CoV-2 vaccines using a variety of technologies [2,3,4,5,6] and the SARS-CoV-2 spike protein (S protein). Indeed, the SARS-CoV-2 spike protein has often been described as the keystone for obtaining an optimal and efficient immunization by eliciting neutralizing antibodies and targeting the structural domains of the S protein: S1 RBD, S1 N-terminal domain, or the S2 region [7,8].

Following vaccination or infection, cellular and humoral responses are generated by the host immune system [9,10]. Immediately, innate immunity defenses are triggered to slow down or inhibit initial infection by protecting cells from infection or by eliminating virus-infected cells. This initial reactivity allows time for the adaptive immune response to begin. The adaptive immune response is based on T cell- and B cell-mediated responses. Firstly, T cell-mediated response will take place, principally by recognition and destruction of virus-infected cells [11]; secondly, specific antiviral antibodies released by mature secreting plasma B cells and commonly named immunoglobulins (Ig) create an immune-protective barrier against infection [12].

Consequently, to obtain a durable protective immunity, memory B cells and/or memory T cells must be obtained [13]. Memory cells are lasting immune cells capable of recognizing foreign proteins to which they have previously been exposed. These immune cells will facilitate a faster secondary response when the antigen is encountered on a subsequent occasion. Memory plasma cells go on to continuously secrete antibodies which allow the immune system to maintain a stable humoral immunological memory over long periods. In this way, multiple studies are investigating seroprevalence within different populations, for example, in patients with cancer [14] and within different environmental contexts [15,16]. These studies are mainly focused on IgG [17] and IgM patterns for each cohort or in relation to COVID-19 symptoms and severity [18,19]. Given the development of many different vaccines, they are also evaluating their effectiveness to generate short- and long-term immune responses.

For this research, the scientific community requires sound operational diagnostic tools [20]. Nucleic acid amplification, such as polymerase chain reaction (PCR), can be used for the detection of the viral genome, but several time-consuming steps need to be performed. This test is optimal and is generally recommended for the diagnosis of an acute SARS-CoV-2 infection. However, specific antibody-based technology will allow detection of specific antibodies directed against the SARS-CoV-2 virus, and this serological testing may be considered more reliable for the diagnosis of suspected patients presenting with negative viral genomic results and for the analysis of asymptomatic infections [21]. Specific antibody quantification technology will therefore prove essential in studying the effectiveness of the various vaccinations commercialized or under development. To date, there is a lack of commercially available products in different countries; most commercialized kits were only available outside the European area. Moreover, no detailed protocols have been yet shared by the scientific community.

In the present study, we have developed a robust and reproducible enzyme-linked immunosorbent assay (ELISA) protocol. This protocol describes a method for quantitative detection of IgG antibodies against the SARS-CoV-2 spike protein with limited inter- and intra-assay variability. This limited variability has been validated with batches of independently prepared standard curves.

2. Materials and Methods

2.1. SARS-CoV-2 Spike Protein-Specific IgG Quantification by ELISA

The following steps were carried out in compliance with good laboratory practices (GLP):

  1. Clear Flat-Bottom Immuno Nonsterile MEDISORP 96-Well Plates (Thermo Fisher Scientific, Waltham, MA, USA, 467320) were coated with 100 nanograms (ng) of SARS-CoV-2 spike protein S1/S2 (S-ECD) (aa11-1208) (Thermo Fisher Scientific, RP87680) in 100 microliters (μL) phosphate-buffered saline (PBS 1X) (Sigma Aldrich, St. Louis, MO, USA, D8537–500 ML) per well for 16 to 18 h at 4 °C.

  2. 1X wash buffer was prepared from 25X concentrated solution (Thermo Fisher Scientific, 3501004) by allowing it to reach room temperature (RT), then mixing to ensure that any precipitated salts had been dissolved before diluting with deionized water (Reactolab, Chroma, Servion, Switzerland, 99389.9010).

  3. Plates were washed 4 times for 5 min with 250 μL of the 1X wash buffer under 200 rpm agitation using a plate orbital shaker.

  4. Plates were blocked with 250 μL of PBS 1X + 0,05% Tween 20 (AppliChem, Darmstadt, Germany, A1389,0500) + 1% Bovine Serum Albumin (BSA) (Sigma Aldrich, A4503-50G) for 1 h at RT under 200 rpm agitation using a plate orbital shaker.

  5. Plates were incubated either with 100 μL of samples or, for the standard curve, 100 μL of primary antibody S-RBD Chimeric Recombinant Mouse Monoclonal Antibody dilution (D005) (Thermo Fisher Scientific, MA5-35958). The standard curve was prepared using serial dilution from 500 ng/mL until 1.953 ng/mL. The samples and antibody were diluted in blocking buffer to obtain 100 μL of each analyte. These were then incubated for 90 min at RT under 200 rpm agitation using a plate orbital shaker.

  6. Plates were washed 4 times for 5 min with 300 μL of 1X wash buffer under 200 rpm agitation using a plate orbital shaker.

  7. Plates were incubated with 100 μL of a 1 μg/mL secondary antibody, Mouse anti-Human IgG1 Fc secondary antibody, HRP solution (Thermo Fisher Scientific, A-10648) for the detection of bound antibodies for 1 h at RT under 200 rpm agitation using a plate orbital shaker.

  8. Plates were washed 4 times for 5 min with 300 μL of 1X wash buffer under 200 rpm agitation using a plate orbital shaker.

  9. Plates were revealed by chromogenic revelation using 100 μL of 1X TMB substrate (Thermo Fisher Scientific, SB01) for 30 min at RT in the dark and stopped with 100 μL of 1X stop solution (Thermo Fisher Scientific, SS01).

  10. Plates were read immediately with a SpectraMax reader for optical density/absorbance of the samples at 450 nm. The concentration of the samples was then calculated using the equation of the standard curve and taking into account the dilution of each sample.

2.2. Data Analysis

Data analyses were performed using SoftMax Pro 7.1 (Molecular Devices Software) and Microsoft® Excel® 2013 (Microsoft Office Professional Plus 2013).

3. Results

3.1. Reproducibility of SARS-CoV-2 Spike Protein-Specific IgG Quantification by ELISA

This protocol has been designed on the hypothesis that a quantitative, sensitive, and reproducible homemade ELISA is feasible with individual commercially available consumables.

This standardized protocol allows quantitative detection of anti-SARS-CoV-2 IgG antibodies in serum samples. Data presented in Table 1 shows optical density (O.D.) values for nine standard concentration curves across the range of 1.953 ng/mL to 500 ng/mL human-specific IgG against the SARS-CoV-2 spike protein. Four independent standard curves were developed and run over two independent experiments on two different days and on different plates. They are all represented by a 5 Parameter Logistic curve in Figure 1.

Table 1.

Results for the various standard curves over the range of 1.953 ng/mL to 500 ng/mL.

Concentration
(ng/mL)
Run 1.1 Run 1.2 Intra Run 1 Value
O.D. 1–450 nm O.D. 2–450 nm O.D. 3–450 nm Mean O.D. Value SD CV O.D. 1–450 nm O.D. 2–450 nm O.D. 3–450 nm Mean O.D. Value SD CV Mean O.D. Value SD CV
500 3.933 3.953 3.953 3.946 0.012 0.3 3.924 3.935 3.924 3.927 0.006 0.2 3.937 0.013 0.3
250 3.748 3.795 3.775 3.773 0.024 0.6 3.671 3.648 3.680 3.666 0.016 0.4 3.719 0.061 1.6
125 2.916 2.925 2.947 2.929 0.016 0.6 2.764 2.732 2.734 2.743 0.018 0.7 2.836 0.103 3.6
62.5 1.772 1.823 1.828 1.808 0.031 1.7 1.675 1.673 1.673 1.674 0.001 0.1 1.741 0.076 4.4
31.25 0.942 0.975 0.984 0.967 0.022 2.2 0.931 0.918 0.923 0.924 0.007 0.7 0.945 0.028 2.9
15.625 0.478 0.488 0.497 0.488 0.009 1.9 0.488 0.483 0.490 0.487 0.004 0.7 0.487 0.006 1.3
7.813 0.234 0.242 0.244 0.240 0.005 2.3 0.260 0.265 0.255 0.260 0.005 1.8 0.250 0.012 4.8
3.906 0.134 0.138 0.131 0.135 0.004 2.7 0.126 0.125 0.132 0.128 0.004 3.0 0.131 0.005 3.9
1.953 0.066 0.067 0.063 0.065 0.002 3.0 0.060 0.060 0.061 0.060 0.001 1.4 0.063 0.003 5.1
Mean CV = 1.7 Mean CV = 1.0 Mean CV = 2.8
Concentration
(ng/mL)
Run 2.1 Run 2.2 Intra Run 2 Value
O.D. 1–450 nm O.D. 2–450 nm O.D. 3–450 nm Mean O.D. Value SD CV O.D. 1–450 nm O.D. 2–450 nm O.D. 3–450 nm Mean O.D. Value SD CV Mean O.D. Value SD CV
500 3.945 3.922 3.941 3.936 0.012 0.3 3.952 3.949 3.952 3.951 0.002 0.0 3.944 0.011 0.3
250 3.847 3.932 3.916 3.898 0.045 1.1 3.918 3.905 3.927 3.917 0.011 0.3 3.908 0.031 0.8
125 3.474 3.538 3.573 3.528 0.050 1.4 3.592 3.632 3.672 3.632 0.040 1.1 3.580 0.070 1.9
62.5 2.530 2.619 2.657 2.602 0.065 2.5 2.701 2.670 2.679 2.683 0.016 0.6 2.643 0.061 2.3
31.25 1.626 1.644 1.658 1.643 0.016 1.0 1.728 1.672 1.677 1.692 0.031 1.9 1.667 0.035 2.1
15.625 0.908 0.909 0.932 0.917 0.013 1.5 0.980 0.948 0.979 0.969 0.018 1.9 0.943 0.032 3.4
7.813 0.482 0.506 0.497 0.495 0.012 2.5 0.524 0.519 0.539 0.527 0.010 2.0 0.511 0.020 4.0
3.906 0.244 0.244 0.246 0.244 0.001 0.4 0.285 0.274 0.278 0.279 0.006 2.1 0.262 0.019 7.3
1.953 0.141 0.140 0.128 0.136 0.007 5.2 0.145 0.139 0.133 0.139 0.006 4.2 0.138 0.006 4.4
Mean CV = 1.8 Mean CV = 1.6 Mean CV = 2.9
Concentration
(ng/mL)
Inter-Assay to Assay Value
Mean O.D. Value SD CV
500 3.940 0.012 0.3
250 3.813 0.108 2.8
125 3.208 0.397 12.4
62.5 2.192 0.476 21.7
31.25 1.306 0.378 29.0
15.625 0.715 0.239 33.4
7.813 0.380 0.137 36.1
3.906 0.196 0.069 35.4
1.953 0.100 0.039 39.3
Mean CV = 23.4

Figure 1.

Figure 1

Graphical representation by a 5 Parameter Logistic (5PL) Curve of the Results for the various standard curves across the range of 1.953 ng/mL to 500 ng/mL human-specific IgG against the SARS-CoV-2 spike protein.

Reproducibility within the assay was assessed over these four independent nine-standard concentration curves. Each standard concentration curve was performed in triplicate. Results obtained show an overall calculated intra-assay coefficient of variation (CV) of 1.53% (range of 1.0% to 1.8%).

Reproducibility of the standard curves between assay to assay, performed on the same day and on two distinct plates, gave rise to a CV of 2.8% for the first two plates (range of 0.3% to 5.1%) and 2.9% for the second two plates (range of 0.3% to 7.3%).

All of these intra-assay O.D. values, whether inside the same plate or during the same run, show robust reproducibility for all of the nine-standard concentration curves.

Inter-assay precision within our laboratory was calculated between all of the values obtained during these four independent experiments and gave rise to an overall interassay CV of 23.4% (range of 0.3% to 39.3%). This overall inter-assay CV shows a proportionally inverse increase compared to the nine-standard concentration curves, meaning that inter-assay precision may be around 30% when concentrations lower than 15 ng/mL are analyzed on different days.

3.2. Analytical Sensitivity of SARS-CoV-2 Spike Protein-Specific IgG Quantification by ELISA

Regarding the sensitivity of this assay, twelve values were obtained using the dilution buffer (Table 2). These values were obtained at the same time and in the same independent manner as the previously generated results used for the validation of this assay reproducibility.

Table 2.

Evaluation of the Lower Limit of Detection.

Concentration 0 ng/mL
O.D. 1–450 nm O.D. 2–450 nm O.D. 3–450 nm Mean O.D. Value SD LLOD
RUN 1.1 −0.000199 0.000600 0.000400
RUN 1.2 0.000600 0.000200 0.000400 0.00016 0.00039 0.0040
RUN 2.1 −0.000433 0.000007 0.000367
RUN 2.2 0.000433 −0.000567 0.000133

The lower limit of detection (LLOD) was defined as the overall mean O.D. value plus ten standard deviations (SD) as obtained with the dilution buffer. The LLOD for this assay was validated with an O.D. of approximately 0.004 at 450 nm and is therefore considerably lower than the lowest O.D. tested concentration of this assay (1.953 ng/mL). Effectively, the mean overall O.D. for the 1.953 ng/mL is 0.1 (SD 0.039). By extrapolating, we can say that the LLOD is, therefore, below the 1.953 ng/mL concentration.

In parallel, negative human serum control was used in order to check whether the background is generated with this analyte. The result obtained was similar to the LLOD, with an O.D. of 0.004 at 450 nm.

4. Discussion

This standardized quantitative enzyme-linked immunosorbent assay protocol for detection of specific human IgG against the SARS-CoV-2 spike protein is an assay designed to detect and quantify the level of Human IgG anti-SARS-CoV-2 spike protein.

All of the components (plastic plates, coating buffer, concentration of the coated protein/secondary antibody) and parameters within each step of this protocol have been optimized to allow researchers to perform a quantitative, sensitive, and reproducible homemade ELISA. Parameters have been selected based on the broad range of obtained O.D.s for the standard curve and on the lowest background generated by plastic and chemical consumables. Moreover, different volumes of samples have been tested in order to minimize the use of the rare and precious samples that can be obtained from studied patients and/or laboratory animals. Indeed, a maximum of 5 μL is needed for the less concentrated samples.

The results obtained leave us confident about the good reproducibility of this quantitative ELISA protocol. However, inter-assay precision within our laboratory shows overall inter-assay CV higher than 30% when analyzed concentrations are smaller than 15 ng/mL. While notably, these data fall within the range of commercialized products but with a much lower cost and with worldwide availability of products.

The limitation of the ELISA assay technology is that the scientific community should keep in mind that when comparing same-subject samples, it is preferable to analyze them at the same time. Furthermore, for long-term studies, it may be advisable to include samples with previously determined concentrations as internal positive quality controls of assay performance. These controls should be used to determine if a run is acceptable or needs to be redone.

Finally, this protocol presents the potential to adapt the revelation antibodies to the species studied, allowing it to be applied to many different sample species. This was demonstrated by our previous study [5], where we analyzed mice samples. We obtained qualitative data from each mice sample and were able to compare baseline samples with sacrifice samples. It can also be useful for seroconversion studies for which IgG and IgM levels need to be quantified separately. In this way, identically prepared and diluted samples can be analyzed and compared at the same time and on the same coated plate, using several secondary revelation antibodies.

Within the pandemic scenario, we believe that a balance must be achieved between rapid development and obtaining robust and reproducible analytical methods. Indeed, it is essential for the scientific community to have resilient operational tools in order to generate comparable data and minimize variability. In the context of the COVID-19 pandemic, this quantitative ELISA protocol can be utilized for research used only to meet this demand without compromising good laboratory practices.

5. Conclusions

In conclusion, this protocol describes a method for quantitative detection of IgG antibodies against the SARS CoV 2 spike protein using antigen-coated microtiter plates. Results showed that antibodies could be quantified between the range of 1.953 ng/mL to 500 ng/mL with limited inter- and intra-assay variability.

Acknowledgments

We would like to thanks Filippo Passardi from Thermo Fisher Scientific for support regarding technical tools and commercial products.

Author Contributions

Conceptualization, R.V.; formal analysis, R.V.; writing—original draft preparation, R.V.; writing—review and editing, R.V., E.C., J.G., and N.M. All authors have actively contributed to the manuscript’s preparation and agree with its scientific content. All authors have read and agreed to the published version of the manuscript.

Funding

This work was funded by MaxiVAX. This financial support has not played any role in the study design, collection, analysis, and interpretation of data, in the writing of the report, or in the decision to submit the paper for publication.

Informed Consent Statement

Informed consent was obtained from the subject involved in the study.

Data Availability Statement

Data can be accessed upon request.

Conflicts of Interest

The authors declare that they have no known competing financial interests or personal relationships that could appear to influence the work reported in this paper. N. Mach is a founder and minority stockholder of MaxiVAX SA; E. Charrier and J. Grogg are employees of MaxiVAX SA. MaxiVAX SA is a Geneva-based biotech company involved in the development of cell encapsulation technologies.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.World Health Organization World Health Organization Coronavirus Disease (COVID-19) Dashboard. [(accessed on 20 October 2020)]; Available online: https://covid19.who.int/
  • 2.Polack F.P., Thomas S.J., Kitchin N., Absalon J., Gurtman A., Lockhart S., Perez J.L., Marc G.P., Moreira E.D., Zerbini C., et al. Safety and Efficacy of the BNT162b2 mRNA COVID-19 Vaccine. N. Engl. J. Med. 2020;383:2603–2615. doi: 10.1056/NEJMoa2034577. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Voysey M., Clemens S.A.C., Madhi A.S., Weckx L.Y., Folegatti P.M., Aley P.K., Angus B., Baillie V.L., Barnabas S.L., Bhorat E.Q., et al. Safety and efficacy of the ChAdOx1 nCoV-19 vaccine (AZD1222) against SARS-CoV-2: An interim analysis of four randomised controlled trials in Brazil, South Africa, and the UK. Lancet. 2021;397:99–111. doi: 10.1016/S0140-6736(20)32661-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Baden L.R., El Sahly H.M., Essink B., Kotloff K., Frey S., Novak R., Diemert D., Spector S.A., Rouphael N., Creech C.B., et al. Efficacy and Safety of the mRNA-1273 SARS-CoV-2 Vaccine. N. Engl. J. Med. 2021;384:403–416. doi: 10.1056/NEJMoa2035389. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Vernet R., Charrier E., Cosset E., Fièvre S., Tomasello U., Grogg J., Mach N. Local Sustained GM-CSF Delivery by Genetically Engineered Encapsulated Cells Enhanced Both Cellular and Humoral SARS-CoV-2 Spike-Specific Immune Response in an Experimental Murine Spike DNA Vaccination Model. Vaccines. 2021;9:484. doi: 10.3390/vaccines9050484. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Sadoff J., Gray G., Vandebosch A., Cárdenas V., Shukarev G., Grinsztejn B., Goepfert P.A., Truyers C., Fennema H., Spiessens B., et al. Safety and Efficacy of Single-Dose Ad26.COV2.S Vaccine against COVID-19. New Engl. J. Med. 2021;384:2187–2201. doi: 10.1056/NEJMoa2101544. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Jiang S., Hillyer C., Du L. Neutralizing Antibodies against SARS-CoV-2 and Other Human Coronaviruses. Trends Immunol. 2020;41:355–359. doi: 10.1016/j.it.2020.03.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wang D., Mai J., Zhou W., Yu W., Zhan Y., Wang N., Epstein N., Yang Y. Immunoinformatic Analysis of T- and B-Cell Epitopes for SARS-CoV-2 Vaccine Design. Vaccines. 2020;8:355. doi: 10.3390/vaccines8030355. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Chowdhury M.A., Hossain N., Kashem M.A., Shahid A.M., Alam A. Immune response in COVID-19: A review. J. Infect. Public Health. 2020;13:1619–1629. doi: 10.1016/j.jiph.2020.07.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Paces J., Strizova Z., Smrz D., Cerny J. COVID-19 and the Immune System. Physiol. Res. 2020:379–388. doi: 10.33549/physiolres.934492. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Toor S.M., Saleh R., Nair V.S., Taha R.Z., Elkord E. T-cell responses and therapies against SARS-CoV-2 infection. Immunology. 2021;162:30–43. doi: 10.1111/imm.13262. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Laing K. Immune Responses to Viruses. 2. [(accessed on 20 October 2020)]; Available online: https://www.immunology.org/public-information/bitesized-immunology/pathogens-and-disease/immune-responses-viruses.
  • 13.Dan J.M., Mateus J., Kato Y., Hastie K.M., Yu E.D., Faliti C.E., Grifoni A., Ramirez S.I., Haupt S., Frazier A., et al. Immunological memory to SARS-CoV-2 assessed for up to 8 months after infection. Science. 2021;371:eabf4063. doi: 10.1126/science.abf4063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Addeo A., Shah P., Bordry N., Hudson R., Albracht B., Marco M.D., Kaklamani V., Dietrich P.-Y., Taylor B., Simand P.-F., et al. Immunogenicity of SARS-CoV-2 MRNA Vaccines in Patients with Cancer: Findings from an International Collaborative Cohort Study. Cancer Cell. 2021;39:1–8. doi: 10.21203/rs.3.rs-554993/v1. In Review. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Liu X., Wang J., Xu X., Liao G., Chen Y., Hu C.-H. Patterns of IgG and IgM antibody response in COVID-19 patients. Emerg. Microbes Infect. 2020;9:1269–1274. doi: 10.1080/22221751.2020.1773324. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Hormati A., Poustchi H., Ghadir M.R., Jafari S., Jafari N., Ashtari A., Ahmadpour S. Production of SARS-CoV-2 Antibodies and Emergence of the Clinical Symptoms of COVID-19. Iran. J. Allergy Asthma Immunol. 2021 doi: 10.18502/ijaai.v20i2.6057. [DOI] [PubMed] [Google Scholar]
  • 17.Prakash O., Solanki B., Sheth J., Makwana G., Kadam M., Vyas S., Shukla A., Pethani J., Tiwari H. SARS-CoV2 IgG antibody: Seroprevalence among health care workers. Clin. Epidemiol. Glob. Heal. 2021;11:100766. doi: 10.1016/j.cegh.2021.100766. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Wang Y., Zhang L., Sang L., Ye F., Ruan S., Zhong B., Song T., Alshukairi A.N., Chen R., Zhang Z., et al. Kinetics of viral load and antibody response in relation to COVID-19 severity. J. Clin. Investig. 2020;130:5235–5244. doi: 10.1172/JCI138759. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Marklund E., Leach S., Axelsson H., Nyström K., Norder H., Bemark M., Angeletti D., Lundgren A., Nilsson S., Andersson L.-M., et al. Serum-IgG responses to SARS-CoV-2 after mild and severe COVID-19 infection and analysis of IgG non-responders. PLoS ONE. 2020;15:e0241104. doi: 10.1371/journal.pone.0241104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Sreepadmanabh M., Sahu A.K., Chande A. COVID-19: Advances in diagnostic tools, treatment strategies, and vaccine development. J. Biosci. 2020;45:1–20. doi: 10.1007/s12038-020-00114-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Long Q.-X., Liu B.-Z., Deng H.-J., Wu G.-C., Deng K., Chen Y.-K., Liao P., Qiu J.-F., Lin Y., Cai X.-F., et al. Antibody responses to SARS-CoV-2 in patients with COVID-19. Nat. Med. 2020;26:845–848. doi: 10.1038/s41591-020-0897-1. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data can be accessed upon request.