Abstract
This study presents a comparison of quantitative MRI methods based on an efficiency metric that quantifies their intrinsic ability to extract information about tissue parameters. Under a regime of unbiased parameter estimates, an intrinsic efficiency metric was derived for fully-sampled experiments which can be used to both optimize and compare sequences. Here we optimize and compare several steady-state and transient gradient-echo based qMRI methods, such as magnetic resonance fingerprinting (MRF), for joint and mapping. The impact of undersampling was also evaluated, assuming incoherent aliasing that is treated as noise by parameter estimation. In vivo validation of the efficiency metric was also performed. Transient methods such as MRF can be up to 3.5 times more efficient than steady-state methods, when spatial undersampling is ignored. If incoherent aliasing is treated as noise during least-squares parameter estimation, the efficiency is reduced in proportion to the SNR of the data, with reduction factors of 5 often seen for practical SNR levels. In vivo validation showed a very good agreement between the theoretical and experimentally predicted efficiency. This work presents and validates an efficiency metric to optimize and compare the performance of qMRI methods. Transient methods were found to be intrinsically more efficient than steady-state methods, however the effect of spatial undersampling can significantly erode this advantage.
Keywords: efficiency, quantitative MRI (qMRI), T1 mapping, T2 mapping, steady-state, fingerprinting (MRF)
Introduction
Over time many methods have been developed that aim to estimate and as effectively as possible, from classical inversion recovery and spin-echo sequences, to steady-state sequences (Deoni et al 2003, Welsch et al 2009, Heule et al 2014, Teixeira et al 2017, Shcherbakova et al 2018) and more recently MR fingerprinting (MRF) (Ma et al 2013). Selecting which method to favor for any given scenario can be challenging since the achieved precision (statistical uncertainty) and accuracy (proximity to true value), and the way these change with acquisition time are generally complex functions of the pulse sequence settings, tissue properties, specific details of the hardware used and the type of image reconstruction.
Nevertheless, comparisons have been made using metrics that strive to evaluate the intrinsic merits of each method (Crawley and Henkelman 1988, Jones et al 1996, Deoni Rutt and Peters 2003, Ma et al 2013, Assländer et al 2019a, Assländer 2020), accounting for the differences external to the methods themselves. Fundamentally, these differences are the amount of data and the SNR of the experiment. The amount of data has been normalized using either the total number of measurements (Jones et al 1996) or the total acquisition time (Edelstein et al 1983, O’Donnell et al 1986, Crawley and Henkelman 1988, Deoni et al 2003, Ma et al 2013, van Valenberg et al 2017, Assländer 2020). The SNR of the experiment has been normalized using the voxel volume (Deoni et al 2003), a combination of the thermal noise level with the proton density (Edelstein et al 1983, O’Donnell et al 1986, Jones et al 1996, van Valenberg et al 2017, Assländer et al 2019a, Assländer 2020) or the signal dynamic range of each method (Crawley and Henkelman 1988). Some of these studies (Ma et al 2013, Assländer et al 2019a, Assländer 2020) showed that balanced MRF (using a balanced readout) outperforms driven equilibrium (DE) single pulse observation of / (DESPOT) for both and estimation, but it is as yet unclear how other commonly used methods, like spoiled MRF (Jiang et al 2015) (using a gradient spoiled readout) or double echo steady-state (DESS) (Welsch et al 2009), compare. Furthermore, methods based entirely on gradient spoiled readouts are popular for their insensitivity to off-resonance at a cost of an SNR penalty—it is unclear whether this trade-off benefits and estimation.
In this work we focus on precision and propose a general efficiency metric that integrates the concepts from figures of merit used so far, quantifying the encoded information from a spin dynamics perspective, but not how it is decoded by the reconstruction. The resulting metric is then used to optimize and comprehensively compare a range of well-established methods that simultaneously estimate and Finally, we discuss the utility of the efficiency metric and other important considerations when comparing qMRI methods.
Theory
We consider that each voxel contains a single pool of spins characterized by unique values of and that the signal models for the different qMRI methods are accurate, subject to additive Gaussian noise; and that parameter estimation results in an unbiased estimate of the parameters of interest (i.e. and but potentially other parameters). Therefore, the error in the estimates is defined by the precision that is characterized by the standard deviation of the estimated parameter Analogous to the definition of SNR, the precision can also be represented by the parameter-to-noise-ratio ():
Although the directly relates to the SNR, it also depends on how much information about the parameter being measured is encoded in the data. Further, the SNR can be broken down to consist of an intrinsic SNR (relating to the receiver system, field strength, resolution etc) and the amount of data acquired. These dependencies have been highlighted in previous works (Edelstein et al 1983, O’Donnell et al 1986, Crawley and Henkelman 1988, Jones et al 1996) and can be expressed in the following equation that serves to define efficiency with which a parameter is estimated,
Here, represents the maximum SNR of any one measurement. is defined as the maximum signal from a voxel that would be measured by applying a pulse with the magnetization in thermal equilibrium; this is a characteristic of the system (field strength and coil) and acquisition geometrical parameters (resolution and field of view); is the receiver noise standard deviation (i.e. what would be measured during one k-space data readout scaled to account for differences in scaling between k-space and image domain) and is characteristic of the k-space readout and its bandwidth; is the total acquisition duration for all data required to estimate the parameter Rewriting equation (2) with as its subject gives:
An upper bound on the efficiency can be obtained without need for physical measurement by calculating the Cramér–Rao lower bound (CRLB) for (Sengupta and Kay 1995), resulting in The ability to achieve this bound depends on the full parameter reconstruction pipeline, which we assume to extract all encoded information so that equality holds.
To illustrate the utility of the efficiency concept, figure 1 shows an example simplified ‘fingerprinting’ experiment with only 5 radiofrequency (RF) pulses, each followed by a measurement. When optimized for maximum efficiency in estimating and two pulses get set to zero amplitude leaving a 3 pulse, 60°–180°–90°, sequence that basically combines two familiar sequences: a spin-echo and an inversion recovery with optimized inversion and echo times for maximum information. Interestingly, it is more efficient to measure fewer signals but allow magnetization to recover, in this case.
Figure 1.

Optimized fingerprint with just 5 pulses that are applied cyclically, with a spoiling gradient preceding each pulse. Schematic representation of (a) the optimized flip angles and intervals between the pulses and (b) the signal and its derivatives w.r.t. and at echo time (2 ms). Note that the regions in gray are already repetitions of the main block of 5 pulses. Because the efficiency measure incorporates the acquisition time it allows the best structure of flip angles and their timings to be found such its averaging extracts the most information about the parameters of interest.
Although the results from equations (2) to (3) are general, when computing efficiency with the CRLB we have assumed that each measurement is fully-sampled and acquired using a single channel coil with uniform sensitivity. This allows a quick evaluation of efficiency, crucial for use in optimization, but is not realistic. In practice, data is acquired using multi-channel receiver coils, often with some degree of undersampling, either for parallel imaging or as an inherent part of the method as in MRF. For multi-channel coils, is still defined as the standard deviation in one readout in one channel, as after pre-whitening all channels should have the same noise level (Pruessmann et al 1999). In this case is the result obtained from optimally combining all channels. Use of multi-channel coils with optimal combination would not affect the efficiency; better intrinsic SNR would improve the which is captured by in equation (3). On the other hand, undersampling will affect both and impacting the efficiency of the sequence. According to Hu and Peters (2019), the standard deviation of in an experiment with an undersampling factor can be related to the fully-sampled case as:
where is the so-called ‘dynamics-factor’ that expresses the parameter error amplification due to the ill-conditioning of the parameter estimation (Hu and Peters 2019). The subscript represents the n th voxel, as may be spatially varying since it will include effects from coil encoding such as the g-factor (Pruessmann et al 1999) as well as sampling effects that may arise from use of time-varying non-cartesian k-space trajectories. Hence the efficiency of an undersampled experiment can be related back to the fully-sampled experiment efficiency by:
For a least-squares estimator so can only reduce the efficiency compared to the fully-sampled experiment. In general, must be estimated by analysis of the full parameter reconstruction pipeline, which will be problem dependant and could become very large for a non-cartesian image reconstruction. In the following subsection we approximate for the special case of a zero-filled reconstruction.
Zero-filled reconstruction
The original MRF paper (Ma et al 2013) proposed a zero-filled reconstruction that treated undersampling artifacts as noise. In this case for a least-squares estimator the final parameter standard deviations are proportional to the signal standard deviations; thermal noise and ‘aliasing noise’ both contribute noise that have a similar impact on parameter estimation but may differ in relative strength. Assuming ‘aliasing noise’ follows an independent Gaussian distribution the image domain signal standard deviation in an undersampled experiment may be written:
where is the image domain noise standard deviation in a fully-sampled experiment. Hence, we may write
This may further be written in terms of the signal-to-noise-ratio in a ‘fully-encoded’ image () and the signal-to-aliasing ratio () from an undersampled image without thermal noise as:
It is therefore expected that the dynamics-factor will become larger if the signal-to-noise-ratio of the data improves; although at first this seems non-intuitive it highlights that aliasing effects are proportional to the signal, so a stronger signal leads to a larger contribution. Note that in equations (6)–(8) we have considered a constant signal amplitude and an image domain SNR since aliasing is fundamentally treated in the image domain.
Method
Optimized sequence design
In order to make a fair comparison between all analyzed methods, sequence acquisition settings (repeat time, flip angles etc) were optimized similarly to other works (Gras et al 2017, Nataraj et al 2017, Teixeira et al 2017, Assländer et al 2019b, Zhao et al 2019), improving and efficiencies for a range of tissue parameters represented by the set of parameters For clarity, is the set of parameters of interest, while is all parameters that may affect the signal; in general but this dependence is kept implicit for notation simplicity. In this work while depends on the types of sequence used—if spoiled sequences are used then but if balanced sequences are used, then off-resonance frequency and phase of the measurement are also relevant, such that
The optimization solved to find the acquisition parameters that maximizes efficiency for each method follows:
where and are method-dependent constraint functions and are detailed in supporting information table S1 (available online at stacks.iop.org/PMB/66/15NT02/mmedia). The set of parameters consists of (corresponding to white matter at 3 T Bojorquez et al 2017); and in steps of 5 Hz. Inclusion of a range of forces methods based on balanced sequences to achieve good efficiencies over a range of frequencies.
The optimization problem was solved using the sequential quadratic program algorithm from Matlab function fmincon. Whenever the number of design variables was a multi-start strategy was employed consisting of random initializations, otherwise a single initialization was used consisting of the originally published acquisition settings for the respective method.
Efficiency comparison
We have studied five steady-state methods: DESPOT (Deoni et al 2003) and a variant called joint system relaxometry (JSR) (Teixeira et al 2017) (analyzed together due to their similarity), PLANET (Shcherbakova et al 2018), DESS (Welsch et al 2009) and triple echo steady-state (TESS) (Heule et al 2014); and MRF sequences with gradient spoiled (Jiang et al 2015) or balanced (Ma et al 2013, Assländer et al 2016) readouts. MRF sequences started from thermal equilibrium or were in a DE mode (Ma et al 2018; Assländer et al 2019b), in which a pulse train of fixed length is cycled such the final magnetization is equal to the initial magnetization. For each method optimized acquisition settings were determined using equation (9) with efficiency calculated using the CRLB in equation (3) assuming fully-sampled measurements. This was repeated for several acquisitions with different numbers of measurements (table 1). The signal model, optimization constraints and acquisition settings for each method are in supporting information table S1.
Table 1.
Number of measurements of the several acquisitions for which the and efficiencies of every method were optimized. is the number of measurements in the transient method (length of the fingerprint). For the transient methods (in orange), fingerprints with less than 400 measurements (in gray) were not considered for further analysis as these could be incompatible with spatial encoding.
|
For each method the acquisition with the lowest cost function (highest efficiency) was evaluated for a wider range of in steps of 40 ms and in steps of 4 ms. We found that MRF sequences with very small numbers of pulses can achieve very high efficiencies, particularly if starting from thermal equilibrium. However, in practice such short pulse trains make only a limited number of measurements, so cannot support spatial encoding. Therefore, MRF acquisitions with less than 400 excitations were not evaluated over the extended parameter range (values shown in gray in table 1).
For the main cross comparisons between methods, all efficiencies were calculated assuming fully-sampled measurements. This makes a concise and transparent presentation and is a reasonable approach for steady-state methods. However, MRF methods are most often acquired with a considerable degree of undersampling, and in these cases the obtained efficiency would also depend on the dynamics-factor according to equation (5). To address this we estimated for the sub-optimal zero-filled reconstruction using equation (8). Both random and spiral (Glover 1999, Pipe and Zwart 2013, Wundrak et al 2015) undersampling were explored for the Shepp–Logan phantom and Monte-Carlo simulations (100 000 trials each with independent Gaussian additive noise) were performed to estimate the standard deviation of the undersampled data Several undersampling factors and different SNR levels were used to estimate
Validation experiments
To experimentally validate the efficiency metric the standard deviation of needs to be determined, requiring multiple estimates obtained from different datasets. To stay within an acceptable acquisition time, we focused on validation of efficiency prediction for estimation of a single parameter— using DESPOT1 (Christensen et al 1974). One healthy volunteer (male, age 25) was scanned on 3 T Achieva MRI systems (Best, Netherland) using a 32-channel head coil. Brain images at a resolution were obtained using 3D Cartesian encoding of a transverse slab with 7 slices, such that the middle slice could be analyzed free of slice profile effects; no parallel imaging acceleration was used. The acquisition consisted of repeats of SPGR sequences with and yielding a maximum root mean squared RF field of 0.46 μT for to minimize bias from magnetization transfer (MT) effects (Ou and Gochberg 2008, Teixeira et al 2019a). To compare multiple examples with different efficiencies, all combinations of at least 3 SPGR were considered. Additionally, a transmit field () map was acquired using actual flip angle imaging method (Yarnykh 2007) with an isotropic resolution of
All numerical simulations and analyses were performed in a workstation with 64 GB of RAM and with an Intel Xeon E5-2687 W 0 @ 3.10 GHz, using MATLAB R2017b (The MathWorks, Natick, MA, USA) with some functions implemented in C++/MEX using the Eigen linear algebra library (Guënnebaud and Jacob 2010).
Results
Efficiency comparison
Figure 2 shows a comparison of optimized efficiencies for steady-state and transient methods assuming fully-sampled measurements in all cases, whilst the optimized acquisition settings are in supporting information table S2 and figure 3. Figures 2(a) and (b) show the distribution of efficiency values over different and values and averaged over off-resonance, while figures 2(c) and (d) show spread over different off-resonance values averaged over and values. Consistently we see that the steady-state methods are less efficient than their transient counterparts; DESPOT/JSR is the most efficient steady-state method while balanced MRF starting from thermal equilibrium is the most efficient transient method. In general the best transient method is approximately 3–3.5 times more efficient than the best steady-state method. The transient methods have an apparently greater spread in efficiency as a function of and Only the methods that include balanced readouts are sensitive to off-resonance, and of these the transient methods seem more sensitive than the steady-state ones. Nevertheless, methods using balanced readouts are more efficient than those using spoiled readouts despite having to estimate two additional nuisance parameters. The optimized acquisition settings for the best transient sequences in figure 3 also reveal recovery periods of zero flip angle in lieu of performing more measurements. Supporting information figure S1 compares efficiency of different optimized MRF trains with different numbers of measurements.
Figure 2.

Efficiency comparison for steady-state (blue) and transient (orange) methods as described in subsection ‘efficiency comparison’; in each case the results shown are for the most efficient acquisition of each method (table 1). (a), (b) efficiency averaged over all off-resonance values; spread corresponds to variability over (c), (d) efficiency averaged over all { }; spread corresponds to variability over off-resonance frequencies. As may be expected the balanced sequences show greater sensitivity to off-resonance.
Figure 3.

Optimized settings for each of the transient methods used for cross comparisons. In spoiled MRF (a) and (b) both flip angle and TR were optimized, whilst for balanced MRF (c) and (d) only flip angle was optimized (TR fixed at 5 ms). For those starting from thermal equilibrium (a) and (c) the shortest feasible sequence was the most efficient (N = 400; see table 1), whereas the opposite was verified for driven equilibrium (b) and (d) sequences (N = 1000; see table 1).
Figure 4 shows the results of the Monte-Carlo investigation of the dynamics-factor for zero-filled reconstructions computed with random and spiral sampling. For this special case, increases quickly for lower undersampling factors R, but then plateaus at higher R; the level it reaches is directly proportional to (figure 4(b)), with spiral sampling achieving lower values than random sampling. Figure 4(c) plots the ‘aliasing-to-signal ratio’ as a function of R for a scenario with zero thermal noise; empirical fits to data show that to a good approximation this is proportional to for both sampling schemes used. The full maps are in supporting information figure S2.
Figure 4.

Analysis of the effect of undersampling for random and spiral sampling. (a) Average in the non-zero locations of the Shepp–Logan phantom as a function of the undersampling factor R and (b) as a function of the SNR in the image domain. (c) Aliasing-to-signal ratio as a function of the undersampling factor R and its empirical fit to the expression
In vivo validation
Figure 5 depicts comparisons of efficiency for DESPOT1 estimation in vivo for several subsets of the total dataset. Figure 5(a) shows efficiency maps for a selection of combinations (the full set of maps is in supporting information figure S3). Figure 5(c) shows the average of the experimental efficiency inside the white and gray matter regions (figure 5(b)) plotted against the average of the theoretical efficiency in the same regions for each combination tested. The goodness-of-fit for this comparison was calculated to be
Figure 5.

In vivo validation results for the efficiency of DESPOT1. (a) Theoretical and experimental efficiency maps obtained for some combinations of the acquired SPGRs. (b) White and gray matter masks obtained using FSL FAST (Zhang et al 2001). (c) Average experimental efficiency inside the gray and white matter masks plotted against the respective theoretical efficiency for each combination of SPGRs. A table with all SPGR combinations is provided in supporting table S3 and all maps are in supporting information figure S3. The correction for incomplete spoiling in SPGR proposed by Baudrexel et al (2018) was implemented. Brain extraction was performed using FSL BET (Smith 2002) and all images were registered using MIRTK (Schuh et al 2013). Gold standard values for and were estimated from fitting to the average of all repetitions and the noise standard deviation was estimated from the image domain (supporting information figure S4).
Discussion
This work presented a comparison of qMRI methods for simultaneous and estimation based on an efficiency metric, that quantifies the information encoded about the parameters per square root of acquisition time. The efficiency metric was validated in vivo using the DESPOT1 method (figure 5). The results showed an excellent agreement between experimental and theoretical efficiency () when averaged over white/gray matter masks.
The metric considers intrinsic efficiency related to the dynamics of the spin system only. Therefore, it considers what mainly distinguishes one method from another: the intrinsic way magnetization is manipulated to achieve greater or lesser sensitivity to the tissue parameters of interest. Nevertheless, according to equation (2) there are two other factors that affect the final parameter-to-noise-ratio, and these do need to be considered in the bigger picture. One factor is the amount of data acquired, as efficiency is normalized by Thus, the SNR advantage yielded by 3D encoding compared to 2D encoding should be accounted for when comparing the expected parameter-to-noise-ratio. The other is the intrinsic SNR of the experiment, specified by While increasing the repetition time might appear to reduce efficiency, one could for example make use of that time to reduce the receiver bandwidth, thus increasing Moreover, different k-space sampling strategies could be employed to make more efficient use of that time, like spiral sampling (Jiang et al 2015) or even EPI (Rieger et al 2017, 2018). However, there are pragmatic limits to which such trade-off can be explored due to decay and increased sensitivity to water-fat shift, off-resonance, among others. In the end, intrinsic SNR, amount of data acquired, and efficiency have to weighed together to determine the best method for a specific application.
A Cramér–Rao approach was adopted for determining the uncertainty of parameter estimation, providing a secure mathematical basis for interpretation of results with a clear domain of applicability based on unbiased estimators (Sengupta and Kay 1995). This connects to a key assumption in relaxometry—the biophysical model. In this work we assumed that the magnetization dynamics are well represented by the Bloch equations and that each voxel contains only one tissue type. In practice this ignores factors such as MT (Ou and Gochberg 2008, Teixeira et al 2019a, 2019b) and diffusion (Kobayashi and Terada 2018) that can produce systematic differences between the model and data, biasing parameter estimation, and to which different methods have different sensitivities.
Figure 2 suggests that after optimization, the transient methods are generally more efficient than the steady-state methods for fully-sampled experiments. From the latter, DESPOT/JSR has the highest optimized efficiency. This approach combines SPGR and balanced SSFP sequences, giving it a large degree of flexibility. We found that the most efficient acquisitions consisted of mainly bSSFP sequences but having at least one SPGR enhances efficiency by decorrelating and information. The PLANET method uses only bSSFP sequences, which results in slightly lower efficiency as it constitutes a constrained case of JSR that excludes SPGR measurements and forces a single flip angle and TR. Optimizing bSSFP sequences to maximize leads to use of multiple different flip angles. Because TESS and DESS obtain multiple echoes per TR period, they are often thought of as efficient, however our results indicate this may not be the case. Although they measure multiple echo pathways, the information in these echoes is correlated since they share the same flip angle and TR, and the higher order echoes often have lower signal amplitudes; it could be more efficient to obtain more diverse data instead.
The transient (MRF) sequences are divided into those in DE and those that start from thermal equilibrium, and between either gradient spoiled or balanced readouts. The results in figure 2 compared only the most efficient transient acquisition with at least 400 measurements (figure 3). Supporting information figure S1 expands this with efficiencies for different numbers of measurements. It was seen that very high efficiencies can be achieved for short MRF pulse trains that start from thermal equilibrium; these may not be viable for performing the spatial encoding required for 3D but could potentially be used for 2D encoding. Methods that start from thermal equilibrium (i.e. the ‘non-DE’ sequences) are ‘privileged’ by assuming sufficient time has elapsed following previous acquisitions to allow full recovery. If this recovery time is also included in the efficiency calculation these methods lose their high efficiency. In the case of longer acquisitions, both MRF sequences in DE and starting from thermal equilibrium converged to a similar efficiency, which is still higher than steady-state methods.
Overall the best transient method was found to be 3–3.5 times more efficient than the best steady-state method considering fully-sampled measurements. It is important to note that this apparently superior efficiency comes from considering spin dynamics alone, i.e. information encoded into the data, but excludes spatial sampling and reconstruction. Our analysis assumed that reconstruction faithfully represents the state of the system at each timepoint. In practice undersampling of the data, which is particularly a feature of transient methods (Ma et al 2013), reduces efficiency according to the dynamics-factor This factor depends on the reconstruction method. Many such methods are available—an exhaustive review of these is beyond the scope of this study. The quoted efficiencies excluding could thus be considered best-case scenarios. At the other extreme, a zero-filled reconstruction (initially proposed for MRF Ma et al 2013) is a straightforward entry-level approach that represents a reasonable worst case for Figure 4 shows that for higher undersampling factors R, is independent of R but scales linearly with the SNR of the data. To understand these relationships, figure 4(c) shows empirically that the ‘alias-to-signal’ ratio scales as Substituting this into equation (8) we find:
where is a scaling constant, which yields for large R and large SNR. Though at first unintuitive this should be expected since high SNR implies that the dominant source of ‘noise’ in the reconstruction comes from aliasing. This implies that actually MRF using a simple zero-filled reconstruction will be less efficient than steady-state imaging in many circumstances.
Of course, it is important to reiterate that zero filling is a simple but sub-optimal approach to MRF reconstruction, and many more sophisticated and superior methods do exist. Current reconstructions using machine learning (Cohen et al 2018, Hamilton and Seiberlich 2020, Hermann et al 2021) and/or some prior knowledge or form of regularization (Pierre et al 2016, Assländer et al 2018a, Zhao et al 2018) should be expected to reduce since they attempt to resolve the aliased signal rather than treat it as noise (Stolk and Sbrizzi 2019). Hence the values computed here might be considered as upper bounds, yet it is still to be shown how far can be reduced by more modern methods while still maintaining accuracy. A further consideration is that more complex reconstructions employing regularization or prior knowledge may introduce bias, which is beyond the scope of the proposed efficiency metric that considers only precision assuming unbiased estimation.
Our comparison analyzed error amplification from thermal noise considering gradient-echo based methods for brain imaging that simultaneously estimate and For other applications (e.g. cardiac or abdominal imaging) different considerations (e.g. physiological noise, SAR, or motion) need to be taken into account. These can be incorporated by changing the model and/or constraining the design space, which might change the comparison landscape obtained here, making particular methods unviable and could favor methods that would otherwise seem inefficient.
Conclusion
This work presents a comparison of several qMRI methods based on an efficiency metric that is the ratio of the best-case parameter-to-noise-ratio to the maximum achievable SNR, normalized to the square root of the acquisition time. This allows the performance of different qMRI methods to be quantified and optimized, and has been used here to compare a range of well-established methods that simultaneously estimate and . We found that methods based on balanced readouts outperformed methods based on spoiled readouts, and that transient qMRI sequences can be 3–3.5 times more efficient than steady-state alternatives for both and mapping in fully-sampled experiments. However, transient sequences which need to capture rapidly fluctuating signals often deploy highly undersampled acquisition strategies. The manner with which this undersampling is resolved can reduce the overall efficiency. For example, MRF using zero-filled reconstruction that treats aliasing artifacts as noise (Ma et al 2013), could easily drop efficiency by a factor of thereby negating the gains offered by a transient acquisition. More advanced reconstructions will be important to realize the potential gains offered by MRF.
The concept of an efficiency metric, such as the one proposed in this paper, can capture all aspects of both acquisition and reconstruction providing a means to assess relative performance and so can help in selection of optimal qMRI methods.
Acknowledgments
This research was funded in whole, or in part, by the Wellcome Trust [WT 203148/Z/16/Z]. For the purpose of open access, the author has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising from this submission. This work was supported by the Wellcome/EPSRC Centre for Medical Engineering [WT 203148/Z/16/Z] and by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London and/or the NIHR Clinical Research Facility, and funded by the King’s College London & Imperial College London EPSRC Centre for Doctoral Training in Medical Imaging [EP/L015226/1]. The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care.
Ethical statement.
Healthy volunteer gave written informed consent according to local ethics requirements (Guy’s Research Ethics Committee, ID 01/11/12).
References
- Assländer J, Glaser Steffen J, Hennig Jürgen. Pseudo Steady-State Free Precession for MR-Fingerprinting. Magn. Reson. Med. 2016;77:1151-61. doi: 10.1002/mrm.26202. [DOI] [PubMed] [Google Scholar]
- Assländer J, et al. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn. Reson. Med. 2018a;79:83–96. doi: 10.1002/mrm.26639. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Assländer J, Novikov D S, Lattanzi R, et al. Hybrid-state free precession in nuclear magnetic resonance. Commun Phys. 2019a;2:73. doi: 10.1038/s42005-019-0174-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Assländer J, et al. Optimized quantification of spin relaxation times in the hybrid state. Magn. Reson. Med. 2019b;82:1385–97. doi: 10.1002/mrm.27819. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Assländer J. A perspective on MR fingerprinting. J. Magn. Reson. Imaging. 2020;53:676-685. doi: 10.1002/jmri.27134. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Baudrexel S, et al. T1 mapping with the variable flip angle technique: a simple correction for insufficient spoiling of transverse magnetization. Magn. Reson. Med. 2018;79:3082–92. doi: 10.1002/mrm.26979. [DOI] [PubMed] [Google Scholar]
- Bojorquez J Z, et al. What are normal relaxation times of tissues at 3 T? Magn. Reson. Imaging. 2017;35:69–80. doi: 10.1016/j.mri.2016.08.021. [DOI] [PubMed] [Google Scholar]
- Christensen K A, et al. Optimal determination of relaxation times of Fourier transform nuclear magnetic resonance. determination of spin-lattice relaxation times in chemically polarized species. J. Phys. Chem. 1974;78:1971–6. doi: 10.1021/j100612a022. [DOI] [Google Scholar]
- Cohen O, Zhu B, Rosen M S. MR fingerprinting deep reconstruction network (DRONE) Magn. Reson. Med. 2018;80:885–94. doi: 10.1002/mrm.27198. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Crawley A P, Henkelman R M. A comparison of one‐shot and recovery methods in T1 imaging. Magn. Reson. Med. 1988;7:23–34. doi: 10.1002/mrm.1910070104. [DOI] [PubMed] [Google Scholar]
- Deoni S C L, Rutt B K, Peters T M. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magn. Reson. Med. 2003;49:515–26. doi: 10.1002/mrm.10407. [DOI] [PubMed] [Google Scholar]
- Edelstein W A, et al. Signal, noise, and contrast in nuclear magnetic resonance (NMR) imaging. J. Comput. Assist. Tomogr. 1983;7:391–401. doi: 10.1097/00004728-198306000-00001. [DOI] [PubMed] [Google Scholar]
- Glover G H. Simple analytic spiral K-space algorithm. Magn. Reson. Med. 1999;42:412–5. doi: 10.1002/(SICI)1522-2594(199908)42:2<412::AID-MRM25>3.0.CO;2-U. [DOI] [PubMed] [Google Scholar]
- Gras V, et al. Diffusion-weighted DESS protocol optimization for simultaneous mapping of the mean diffusivity, proton density and relaxation times at 3 Tesla. Magn. Reson. Med. 2017;78:130–41. doi: 10.1002/mrm.26353. [DOI] [PubMed] [Google Scholar]
- Guënnebaud G, Jacob B. 2010. Eigen v3 .
- Hamilton J I, Seiberlich N. Machine learning for rapid magnetic resonance fingerprinting tissue property quantification. Proc. IEEE. 2020;108:69–85. doi: 10.1109/JPROC.2019.2936998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hermann I, et al. Accelerated white matter lesion analysis based on simultaneous T1 and T2* quantification using magnetic resonance fingerprinting and deep learning. Magn. Reson. Med. 2021;86:471–86. doi: 10.1002/mrm.28688. [DOI] [PubMed] [Google Scholar]
- Heule R, Ganter C, Bieri O. Triple echo steady-state (TESS) relaxometry. Magn. Reson. Med. 2014;71:230–7. doi: 10.1002/mrm.24659. [DOI] [PubMed] [Google Scholar]
- Hu C, Peters D C. SUPER: a blockwise curve-fitting method for accelerating MR parametric mapping with fast reconstruction. Magn. Reson. Med. 2019;81:3515–29. doi: 10.1002/mrm.27662. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jiang Y, et al. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn. Reson. Med. 2015;74:1621–31. doi: 10.1002/mrm.25559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jones J A, et al. Optimal sampling strategies for the measurement of spin–spin relaxation times. J. Magn. Reson. B. 1996;113:25–34. doi: 10.1006/jmrb.1996.0151. [DOI] [Google Scholar]
- Kobayashi Y, Terada Y. Diffusion-weighting caused by spoiler gradients in the fast imaging with steady-state precession sequence may lead to inaccurate T2 measurements in MR fingerprinting. Magn. Reson. Med. Sci. 2018;18:1–9. doi: 10.2463/mrms.tn.2018-0027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ma D, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187–192. doi: 10.1038/nature11971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ma D, et al. Fast 3D magnetic resonance fingerprinting for a whole-brain coverage. Magn. Reson. Med. 2018;79:2190–7. doi: 10.1002/mrm.26886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nataraj G, Nielsen J, Fessler J A. Optimizing MR scan design for model-based T1 , T2 Estimation From Steady-State Sequences. IEEE Transactions on Medical Imaging. 2017;36:467–77. doi: 10.1109/TMI.2016.2614967. [DOI] [PMC free article] [PubMed] [Google Scholar]
- O’Donnell M, Gore J C, Adams W J. Toward an automated analysis system for nuclear magnetic resonance imaging: I. Efficient pulse sequences for simultaneous T1–T2imaging. Med. Phys. 1986;13:182–90. doi: 10.1118/1.595943. [DOI] [PubMed] [Google Scholar]
- Ou X, Gochberg D F. MT effects and T1 quantification in single-slice spoiled gradient echo imaging. Magn. Reson. Med. 2008;59:835–45. doi: 10.1002/mrm.21550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pierre E Y, et al. Multiscale reconstruction for MR fingerprinting. Magn. Reson. Med. 2016;75:2481–92. doi: 10.1002/mrm.25776. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pipe J G, Zwart N R. Spiral trajectory design: A flexible numerical algorithm and base analytical equations. Magnetic Resonance in Medicine. 2013;71 doi: 10.1002/mrm.24675. [DOI] [PubMed] [Google Scholar]
- Pruessmann K P, et al. SENSE: sensitivity encoding for fast MRI. Magn. Reson. Med. 1999;42:952–62. doi: 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S. [DOI] [PubMed] [Google Scholar]
- Rieger B, et al. Magnetic resonance fingerprinting using echo-planar imaging: joint quantification of T1 and T2* relaxation times. Magn. Reson. Med. 2017;78:1724–33. doi: 10.1002/mrm.26561. [DOI] [PubMed] [Google Scholar]
- Rieger B, et al. Time efficient whole-brain coverage with MR Fingerprinting using slice-interleaved echo-planar-imaging. Sci. Rep. 2018;8:1–12. doi: 10.1038/s41598-018-24920-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schuh A, Rueckert D, Schnabel J. 2013. Medical Image Registration ToolKit (MIRTK)
- Sengupta S K, Kay S M. Fundamentals of statistical signal processing: estimation theory. Technometrics. 1995;37:27–81. doi: 10.2307/1269750. [DOI] [Google Scholar]
- Shcherbakova Y, et al. PLANET: an ellipse fitting approach for simultaneous T1 and T2 mapping using phase-cycled balanced steady-state free precession. Magn. Reson. Med. 2018;79:711–22. doi: 10.1002/mrm.26717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith S M. Fast robust automated brain extraction. Hum. Brain Mapp. 2002;17:143–55. doi: 10.1002/hbm.10062. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stolk C C, Sbrizzi A. Understanding the combined effect of k-space undersampling and transient states excitation in MR fingerprinting reconstructions. IEEE Trans. Med. Imaging. 2019;38:2445–55. doi: 10.1109/TMI.2019.2900585. [DOI] [PubMed] [Google Scholar]
- Teixeira R P A G, Malik S J, Hajnal J V. Joint system relaxometry (JSR) and Cramer–Rao lower bound optimization of sequence parameters: a framework for enhanced precision of DESPOT T1 and T2 estimation. Magn. Reson. Med. 2017;79:234–45. doi: 10.1002/mrm.26670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Teixeira R P A G, Malik S J, Hajnal J V. Fast quantitative MRI using controlled saturation magnetization transfer. Magn. Reson. Med. 2019a;81:907–20. doi: 10.1002/mrm.27442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Teixeira R P A G, et al. Controlled saturation magnetization transfer for reproducible multivendor variable flip angle T1 and T2 mapping. Magn. Reson. Med. 2019b;84:1–16. doi: 10.1002/mrm.28109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- van Valenberg W, Vos F M, Klein S, van Vlie L J, Poot D H J. Determining the time efficiency of quantitative MRI methods using bloch simulations.. 25th Proc. Intl. Soc. Mag. Reson. Med. ; Honululu, HI. 22–27 April, 2017; 2017. https://archive.ismrm.org/2017/1470.html. [Google Scholar]
- Welsch G H, et al. Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3 Tesla. Magn. Reson. Med. 2009;62:544–9. doi: 10.1002/mrm.22036. [DOI] [PubMed] [Google Scholar]
- Wundrak S, Paul J, Ulrici J, Hell E, Geibel M-A, Bernhardt P, Rottbauer W, Rasche V. Golden ratio sparse MRI using tiny golden angles. Magn. Reson. in Med. 2015;75 doi: 10.1002/mrm.25831. [DOI] [PubMed] [Google Scholar]
- Wundrak S, et al. Golden ratio sparse MRI using tiny golden angles. Magn. Reson. Med. 2016;2378:2372–8. doi: 10.1002/mrm.25831. [DOI] [PubMed] [Google Scholar]
- Yarnykh V L. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn. Reson. Med. 2007;57:192–200. doi: 10.1002/mrm.21120. [DOI] [PubMed] [Google Scholar]
- Zhang Y, Brady M, Smith S. Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm. IEEE Trans. Med. Imaging. 2001;20:45–57. doi: 10.1109/42.906424. [DOI] [PubMed] [Google Scholar]
- Zhao B, et al. Improved magnetic resonance fingerprinting reconstruction with low-rank and subspace modeling. Magn. Reson. Med. 2018;79:933–42. doi: 10.1002/mrm.26701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhao B, et al. Optimal experiment design for magnetic resonance fingerprinting: Cramér–Rao bound meets spin dynamics. IEEE Trans. Med. Imaging. 2019;38:844–61. doi: 10.1109/TMI.2018.2873704. [DOI] [PMC free article] [PubMed] [Google Scholar]