Reporting of Death in US Food and Drug Administration Medical Device Adverse Event Reports in Categories Other Than Death

Christina Lalani 1,, Elysha M Kunwar 2, Madris Kinard 3, Sanket S Dhruva 4,5,6,7, Rita F Redberg 4,6,8
PMCID: PMC8314174  PMID: 34309624

Key Points

Question

When a medical device adverse event report in the US Food and Drug Administration Manufacturer and User Facility Device Experience database describes a patient death, how often is that report classified as injury, malfunction, other, or missing instead of death?

Findings

In this study and manual review of a random sample of 1000 adverse event reports that were selected using a natural language processing algorithm to identify patient deaths, the percentage of reports classified as injury, malfunction, other, or missing (instead of death) was 23% (95% CI, 20%-25%).

Meaning

Many adverse event reports for medical devices that involve a patient death are classified in categories other than death, and as the US Food and Drug Administration only routinely reviews all adverse events that are reported as patient deaths, improving the accuracy of adverse event reporting may enhance patient safety.

Abstract

Importance

In the US, most postmarket medical device safety data are obtained through adverse event reports that are submitted to the US Food and Drug Administration (FDA)’s Manufacturer and User Facility Device Experience (MAUDE) database. Adverse event reports are classified by the reporter as injury, malfunction, death, or other. If the device may have caused or contributed to a death, or if the cause of death is unknown, the FDA requires that the adverse event be reported as a death.

Objective

To determine the percentage of medical device adverse event reports submitted to the MAUDE database that were not classified as death even though the patient died.

Design, Setting, and Participants

In this study, a natural language processing algorithm was applied to the MAUDE database, followed by manual text review, to identify reports in the injury, malfunction, other or missing categories that included at least 1 term that suggested a patient death, such as patient died or patient expired, from December 31, 1991, to April 30, 2020, for any medical device.

Exposures

Manual review of a random sample of 1000 adverse event reports not classified as death and of selected reports for 62 terms that are associated with deaths but were not classified as death.

Main Outcomes and Measures

Percentage of adverse event reports in which the patient was said to have died in the narrative section of the report but the reporter classified the report in a category other than death.

Results

The terms in the natural language processing algorithm identified 290 141 reports in which a serious injury or death was reported. Of these, 151 145 (52.1%) were classified by the reporter as death and 47.9% were classified as malfunction, injury, other, or missing. For the overall sample, the percentage of reports with deaths that were not classified as deaths was 23% (95% CI, 20%-25%), suggesting that approximately 31 552 reports in our sample had deaths that were classified in other categories. The overall percentage of missed deaths, defined as the percentage of deaths that were classified in other categories, was 17% (95% CI, 16%-19%).

Conclusions and Relevance

Many of the findings of this study suggest that many medical device adverse event reports in the FDA’s MAUDE database that involved a patient death are classified in categories other than death. As the FDA only routinely reviews all adverse events that are reported as patient deaths, improving the accuracy of adverse event reporting may enhance patient safety.


This study examines the percentage of medical device adverse event reports submitted to the Us Food and Drug Administration’s Manufacturer and User Facility Device Experience database that were not classified as death even though the patient died.

Introduction

Before marketing, the clinical trial evidence supporting medical device approval is often limited; postmarket surveillance provides additional evidence on safety and effectiveness.1,2,3 For the highest-risk (class III) medical devices, the US Food and Drug Administration (FDA) requires premarket approval and the submission of clinical data for the agency to review. However, premarket approval is often based on a single, small, nonrandomized study with a short period of follow-up.1 Examples of class III medical devices are ventricular assist bypass devices, deep brain stimulators, and coronary drug-eluting stents. Class II (usually moderate-risk) devices typically receive clearance based on substantial equivalence to an existing device; the submission of clinical data are not required.4 Thus, postmarket surveillance allows for longer-term follow-up of a larger and more diverse group of patients than those enrolled in premarket trials.2,5

Passive surveillance is the primary source of safety signals for medical device safety communications.6 Manufacturers, distributers, and health care facilities are required to submit adverse events to the FDA’s Manufacturer and User Facility Device Experience (MAUDE) database.6,7,8,9 Others (eg, physicians, hospitals, and patients) can submit reports, but such voluntary reports are submitted infrequently.10

The individuals or organizations reporting adverse events to MAUDE choose whether to report them as malfunction, injury, death, or other. The FDA requires that reports be classified as death if the device “may have caused or contributed to a death,” or if the cause of death is unknown.11 If a report in which a patient dies is not classified as death (eg, is reported instead as a device malfunction), the identification of safety signals may be delayed. Although the FDA must review all reports classified as death, it does not routinely review all reports that are classified as injury or malfunction.10

A prior report from our research group12 showed high rates of miscategorization of deaths as malfunctions or injuries for 2 high-risk cardiac devices, the Sapiens 3 aortic valve (Edwards Lifesciences) and the MitraClip (Abbott Vascular), a device used for transcatheter mitral valve repair. In this study, we expand on our earlier article by examining adverse event reports for all medical devices in MAUDE to determine how often a death was reported in a category other than death.

Methods

Data Source

We reviewed adverse event reports in MAUDE from its inception on December 31, 1991, to April 30, 2020. Given that MAUDE is a publicly accessible database that stores only deidentified information, this study was exempt from institutional review board approval. We used a proprietary software (Device Events) that exports MAUDE data into an interface that can be used to navigate the FDA database.13 The software uses a natural language processing algorithm to help identify reports in which there was a death. The natural language processing algorithm searches the free text in reports for terms such as patient died, patient expired, could not be resuscitated, and time of death, which are likely to be associated with death. The algorithm allows for plural and past tense versions of words (eg, death includes deaths and die includes died). The algorithm also excludes reports with selected phrases that are unlikely to be associated with death (eg, patient passed out should be excluded, although the term patient passed should not be excluded).

Contents of MAUDE Adverse Event Reports

The terms included in a MAUDE report include report ID, date of event, date of manufacturer receipt, date of report, date FDA received report, report type, report source, reporter occupation, company name, product code, device name, device brand name, and device generic name. Reporters must also complete an unstructured free text box to provide a more detailed event description. The report type refers to whether the adverse event reporter classified the patient event outcome as a death, malfunction, injury, other, or did not classify the report (missing). The report source describes whether a report was submitted by a user facility, manufacturer, voluntary reporter, or distributor. The Device Events software added a term for the device class (class I-III) associated with the report by matching the device product code with the appropriate class of the device.

Search Query

We included MAUDE reports with at least 1 of 70 terms that were associated with death. As our interest was in the reporting of deaths in categories other than death, we then limited the data set to reports in which the outcome was classified as injury, malfunction, other, or missing and manually reviewed selected reports.

Primary Outcome

The primary outcome was the percentage of reports in which the patient’s ultimate outcome was death, but the report had not been classified as death by the reporter. One of 2 authors (C.L. and E.M.K.) read the report descriptions to determine if a death occurred.

Statistical Analysis

We reviewed the event descriptions for a simple random sample of up to 50 injury, malfunction, other, and missing reports for each of the 70 terms for descriptions of a patient death (Figure). We selected all random samples by assigning a random number to each report in Microsoft Excel and selecting the reports with the 50 lowest assigned numbers. If there were fewer than 50 total reports for a given term, we reviewed all reports.

Figure. Report Review Algorithm.

Figure.

NLP indicates natural language processing.

We calculated the percentage of reports in which a death was identified in the narrative among all reviewed reports (ie, all reports among those classified as injury, malfunction, other, or missing). If there were no reports in which the patient’s ultimate outcome was death, the search term was removed from analysis; 8 terms were removed for this reason, leaving 62 terms.

For each term, we determined the confidence interval for the percentage of these reports that were classified not as death but as injury/malfunction/other/missing despite a patient death in the full data set based on the percentage of death reports calculated from our manual review. We used the epitools package in the R statistical software (R Foundation) to calculate an exact binomial proportion confidence interval. We then determined the number of reports (in multiples of 50) to be reviewed to ensure a confidence interval width of less than 5% for the estimate for the true percentage of these death reports. We calculated 95% confidence intervals, which resulted in statistical significance for P < .05. This calculation was performed by assuming that the estimated percentage of reports in which a patient death occurred but the death was classified as injury/malfunction/other/missing instead of death in a random sample of 50 reports for a given term would generalize to the entire population of reports. For example, if all 50 of 50 manually reviewed reports (100%) for a given term were determined to have a patient outcome of death, the estimated percentage of such death reports was considered to be 100%, and approximately 100 reports would have to be reviewed to ensure a confidence interval at least as narrow as 95% to 100% for the true percentage. If the total number of reports that needed to be reviewed was greater than the total number of reports for a given term, all reports were reviewed, and an exact percentage was calculated. We reviewed all reports for 44 of 62 terms (71.0%). We estimated the total number of death reports that were classified in other categories for each of the 18 terms, for which we calculated confidence intervals by multiplying the estimated percentage of reports with deaths that were classified in other categories for the manually reviewed random sample of reports by the total number of reports that were classified as injury, malfunction, other, or missing.

To estimate the total number of reports of deaths that were classified in categories other than death, we also reviewed a random sample of 1000 reports from the full data set (138 996 reports in categories other than death). We counted the number of reports with death classified in other categories from this manually reviewed random sample of 1000 reports and applied that percentage to the overall sample (138 996 reports). We defined the percentage of missed deaths as the number of reports in which a patient’s outcome was classified as injury/malfunction/other/missing (ie, not as death), but in which the patient had died, divided by the total number of reports with the outcome of patient death.

Results

Sample Characteristics

The terms in the natural language processing algorithm identified 290 141 reports in which a serious injury or death was reported. Of these, 151 145 (52.1%) were classified by the reporter as death (Figure and Table 1). Most reports (162 147 [55.9%]) were submitted for class III devices; 116 358 (40.1%) were for class II devices. Of the reports, 50.1% were received by the FDA between January 2015 and April 2020. Nearly all reports were submitted by manufacturers (95.9%). There was more than 1 term associated with death, such as expired, could not be resuscitated, and pt passed in 52.8% of reports. There was no specified reporter occupation for 139 436 reports (48.1%); when specified, the most frequent reporter occupation was physician (58 787 [20.3%]), which likely refers to either physicians reporting adverse events to manufacturers or physicians employed by manufacturers. The most common product codes among all adverse event reports were ventricular (assist) bypass, with 38 708 reports; dialysate concentrate for hemodialysis (liquid or powder), with 25 261 reports; and transcervical contraceptive tubal occlusion device, with 14 387 reports.

Table 1. Characteristics of the 290 141 Reports in Which the Natural Language Processing Algorithm Identified a Serious Injury or Death.

Characteristic No. (%)
Report type
Death 151 145 (52.1)
Injury 77 516 (26.7)
Malfunction 58 025 (20.0)
Other/missing 3455 (1.2)
Device class
I 7399 (2.6)
II 116 358 (40.1)
III 162 147 (55.9)
Undetermined 4237 (1.5)
Year of report receipt by FDA
Before 1995 1161 (0.4)
1995-1999 11 467 (4.0)
2000-2004 13 198 (4.5)
2005-2009 24 675 (8.5)
2010-2014 94 170 (32.5)
2015-April 2020 145 377 (50.1)
Missing 93 (0.0)
Report source
Manufacturer 278 251 (95.9)
User facility 6172 (2.1)
Voluntary 4778 (1.6)
Distributor 940 (0.3)
Reporter occupation
Unspecifieda 139 436 (48.1)
Physician 58 787 (20.3)
Health care personnelb 41 329 (14.2)
Attorney 26 314 (9.1)
Nurse 10 965 (3.8)
Patient, patient family, or caregiver 6757 (2.3)
Risk manager 4792 (1.7)
Medical equipment company technician/representative 1538 (0.5)
Service/testing personnel 223 (0.1)
Device product codec
DSQ: ventricular (assist) bypass 38 708 (13.3)
KPO: dialysate concentrate for hemodialysis (liquid or powder) 25 261 (8.7)
HHS: transcervical contraceptive tubal occlusion device 14 387 (5.0)
NIQ: coronary drug-eluting stent 11 296 (3.9)
LWS: implantable cardioverter-defibrillator (non-CRT) 10 006 (3.4)
MVK: wearable automated external defibrillator 9273 (3.2)
DTB: permanent pacemaker electrode 8332 (2.9)
MIH: system, endovascular graft, aortic aneurysm treatment 6562 (2.3)
DTK: filter, intravascular, cardiovascular 5777 (2.0)
LZG: pump, infusion, insulin 5005 (1.7)

Abbreviations: CRT, cardiac resynchronization therapy; FDA, US Food and Drug Administration.

a

Includes no information, not applicable, other, and unknown.

b

Includes audiologist, biomedical engineer, dental assistant, dental hygienist, dentist, dietician, emergency medical technician, health professional, home health aide, hospital service technician, medical assistant, medical technologist, nuclear medicine technologist, nursing assistant, occupational therapist, other health care professional, paramedic, personal care assistant, pharmacist, phlebotomist, physical therapist, physician assistant, physicist, radiology technologist, and respiratory therapist.

c

Includes only the top 10 most frequently represented product codes.

Reports in Which the Patient Died but Which Were Not Classified as Death

For the overall sample, the percentage of reports with deaths that were not classified as deaths was 23% (95% CI, 20%-25%), suggesting that approximately 31 552 reports in our sample had deaths that were classified in other categories. Of the 70 natural language processing terms, 62 (88.6%) were associated with deaths that were categorized in other categories. We reviewed a total of 7951 reports for all terms. The reports for 44 (71.0%) of the terms were reviewed in full because of small sample sizes. For the 18 terms with sample sizes that were large enough to calculate confidence intervals, the term death(s) was associated with the highest total number of reports in which the patient died but had been classified as injury/malfunction/other/missing (instead of death). For the term death(s), the percentage of such reports not classified as deaths was 12% (95% CI, 9%-14%) based on our review of a random sample of 600 reports containing the term death(s), suggesting that approximately 13 156 reports with this term had deaths that were classified in other categories. Of the 62 terms, 17 (27.4%) had an estimated percentage of 100%, which means that every time that term was used, the patient had died, even though the reporter had not classified the report as death (Table 2).

Table 2. Reports in the MAUDE Database in Categories Other Than Death in Which the Patient Had Died in All Reports Reviewed.

Term Total No. No. % of Reports that were classified as death (with 95% CI when calculated)a
Reports Nondeath reports Reports reviewed Reports in which the patient had died
Patient
Expired 28 239 3416 100 100 100 (96-100)
Died 23 482 3155 100 100 100 (96-100)
Subsequently expired 26 936 2688 100 100 100 (96-100)
Decedent 20 903 2657 100 100 100 (96-100)
Patients expired 8779 1331 100 100 100 (96-100)
Patient died 1462 1019 100 100 100 (96-100)
Time of death 8329 718 100 100 100 (96-100)
Patient died 4363 598 100 100 100 (96-100)
Patient later expired 431 153 100 100 100 (96-100)
Found dead 769 76 76 76 100
Patient who died 148 64 64 64 100
Could not be resuscitated 281 38 38 38 100
Baby died 54 20 20 20 100
Patients who expired 27 9 9 9 100
Organ failure and expired 47 9 9 9 100
Unreported cause of death 5 3 3 3 100
Babies died 6 3 3 3 100
Total No. of reviewed reports for all 62 terms NA NA 7951 NA NA

Abbreviations: MAUDE, manufacturer and user facility device experience; NA, not applicable.

a

Percentage of reports that were classified as death equal to the number of deaths that were identified on manual review/total number of nondeath reports ×100. A 95% confidence interval was calculated for this percentage in the cases in which only a subset of reports were reviewed.

Missed Deaths

In the random sample of 1000 reports that were not classified as death, the overall percentage of missed deaths, defined as the percentage of all deaths that were classified in other categories, was 17% (95% CI, 16%-19%). The term with the highest percentage of missed deaths among the 44 terms reviewed in full was unreported cause of death, with a total of 5 reports and a 60% missed death rate (Table 3). The term with the highest percentage of missed deaths among the 18 terms reviewed partially was patients died, which had a missed death percentage of 70% (95% CI, 69%-70%).

Table 3. Terms in the MAUDE Database Associated With the Highest Percentages of Missed Deaths.

Term Total No. No. % of Missed death reports (with 95% CI when calculated)a
Reports Nondeath reports Reports reviewed Deaths that were not classified as deaths
Patients died 1462 1019 100 100 70 (69-70)
Unreported cause of death 5 3 3 3 60
Babies died 6 3 3 3 50
Patient who died 148 64 64 64 43
Baby died 54 20 20 20 37
Patient later expired 431 153 100 100 35 (35-35)
Patients who expired 27 9 9 9 33
Patients died 74 25 25 24 33
Fatal outcome 140 49 49 40 31
Patients expired 125 38 38 37 30

Abbreviation: MAUDE, manufacturer and user facility device experience.

a

Percentage of missed deaths was equal to the number of deaths identified on manual review /(number of deaths identified on manual review + number of known deaths) × 100. A 95% confidence interval was calculated for this percentage in the cases in which only a subset of reports were reviewed.

Discussion

We found that a large percentage of adverse event reports in the FDA’s MAUDE database that involved a patient death were classified in categories other than death by the reporter. The classification chosen by the reporter is vital, as the FDA must review all adverse events reported as deaths, which is not the case for the other reporting categories.10 Per FDA instructions, reports of a device malfunction or injury that include a patient death should be reported as death if the death may be associated with the device malfunction or if the cause of death is unknown.11 However, it is possible that manufacturers or other reporters do not classify the patient outcome as death if they do not believe that the death was definitively caused by the device.

The MAUDE database allows for the identification of safety signals for medical devices that are repeatedly associated with serious adverse events. Accurate reporting of patient deaths allows the FDA to pursue investigations to determine if there is cause for concern, notify clinicians, and, when appropriate, take regulatory actions to protect patient safety.14 Thus, it is essential that all deaths be classified and reported appropriately as deaths. There is a disclaimer on the MAUDE database that states, “Submission of a medical device report and the FDA's release of that information is not necessarily an admission that a product, user facility, importer, distributor, manufacturer, or medical personnel caused or contributed to the event.”7

Our findings suggest that the FDA could improve the identification of fatalities among patients with medical devices by using search terms, such as decedent and pt expired, within reports to help identify patterns and clusters of reports that are associated with higher rates of patient death. Using a natural language processing algorithm, as in our study, could help streamline the full review of reports that are more likely to be associated with patient death at a time when the numbers of medical devices and adverse event reports are substantially increasing.14,15 Future studies could evaluate other terms or consider using a machine learning algorithm (with the terms in our study or others) to allow for more rapid evaluation of reports to identify patient deaths, such as has been explored for drug adverse event reporting.16

Currently, manufacturers report, and thus choose, the reporting category for more than 95% of adverse event reports. There is an inherent conflict with manufacturers being the primary reporters, as it may not be in their interest to facilitate identification of serious problems with their own devices in a timely manner. There have been multiple instances of delays by manufacturers in reporting serious malfunctions and deaths that were associated with medical devices, as well as complete failures to report.17,18,19 As a result, it seems likely that more patients have been unknowingly treated with devices that turned out to be dangerous than would have been the case if reporting had not been delayed. Some examples of delayed reporting of adverse events associated with implantable cardioverter-defibrillator leads include delays by St Jude Medical17 and Medtronic.18 In another reporting failure, between 2002 and 2013, 32 000 women reported adverse events associated with Essure (Conceptus), a medical device used for permanent birth control, to the manufacturer, yet the FDA only received 1023 events (3.2%) during that period from the manufacturer.13,19 Bard received multiple warning letters following inspections in 2014 and 2015 that cited multiple failures, including a failure to comply with the medical device reporting regulations by misreporting patient deaths that were associated with vena caval filters as malfunctions.20 For these reasons, physicians, hospitals, and patients should submit reports directly to the FDA instead of or in addition to reporting through the manufacturer. The FDA could facilitate such reporting by making a freely available user-friendly mobile application that patients and physicians could use to report adverse events directly to the agency.

Limitations

Our study has limitations. We performed a manual review of a subset of adverse event reports. Specifically, we reviewed all of the reports for 44 of the 62 terms that were associated with death reports in categories other than death and used random samples with narrow confidence intervals (<5% confidence interval width) to create estimates of misclassified reports and missed deaths that were associated with the other 18 terms. As we only included reports that contained at least 1 term that was likely to be associated with a death and not all reports in the MAUDE database, our findings likely underestimate the actual number of deaths that were not classified as death, even when the patient died.

Conclusions

Many deaths in persons with medical devices are reported to the FDA as adverse events in categories other than death. Our findings suggest that the FDA’s adverse event reporting form should be revised to make it even clearer that if the cause of death in a patient with a medical device is unknown, as it often is,21 it still should be reported in the category of death. As the FDA only routinely reviews all adverse events that are reported as patient deaths, improving the accuracy of adverse event reporting may enhance patient safety.