Abstract
Proton therapy of prostate cancer (PCPT) was linked with increased levels of gastrointestinal toxicity in its early use compared to intensity-modulated radiation therapy (IMRT). The higher radiation dose to the rectum by proton beams is mainly due to anatomical variations. Here, we demonstrate an approach to monitor rectal radiation exposure in PCPT based on prompt gamma spectroscopy (PGS). Endorectal balloons (ERBs) are used to stabilize prostate movement during radiotherapy. These ERBs are usually filled with water. However, other water solutions containing elements with higher atomic numbers, such as silicon, may enable the use of PGS to monitor the radiation exposure of the rectum. Protons hitting silicon atoms emit prompt gamma rays with a specific energy of 1.78 MeV, which can be used to monitor whether the ERB is being hit. In a binary approach, we search the silicon energy peaks for every irradiated prostate region. We demonstrate this technique for both single-spot irradiation and real treatment plans. Real-time feedback based on the ERB being hit column-wise is feasible and would allow clinicians to decide whether to adapt or continue treatment. This technique may be extended to other cancer types and organs at risk, such as the oesophagus.
Subject terms: Prostate, Prostate cancer, Applied physics, Techniques and instrumentation, Biomedical engineering, Radiotherapy
Introduction
Range verification is one of the most important problems to be solved in particle therapy1,2. Offline positron emission tomography (offline PET) has verified range uncertainties of approximately 6 mm3. Offline PET scans are performed after irradiation, and the activated tissue is imaged. However, this technique suffers from low signal and biological washout over time. More recent results with in-beam PET have demonstrated the online capabilities of this technique4. Prompt gamma imaging (PGI) has emerged as an alternative that relies on the prompt nature of the gamma radiation emitted during particle therapy. Range verification can be accomplished in real time during treatment, thus providing a means to avoid unwanted irradiation of healthy tissues. Since 2006, several concepts based on imaging and non-imaging systems have been developed5–11. Eventually, two of them—the knife-edge slit camera and prompt gamma spectroscopy—reached the clinical phase12,13 and are currently being used at proton facilities.
Proton therapy for prostate cancer (PCPT) has been a reality since the 1990s14,15. Several clinical studies have estimated the toxicity of prostate cancer therapy with photons16,17, protons14,18–23, and carbon ions24,25. At the outset, PCPT was considered to deliver less dose than photon radiation to normal tissues surrounding the prostate, such as the rectum and bladder26–28. PCPT had, however, a major setback, with two clinical studies reporting higher toxicity than conventional photon therapy19,21. Sheets et al. showed that although intensity-modulated radiation therapy (IMRT) delivered three times more radiation to the body, it presented 50% less gastrointestinal morbidity. Proton therapy-treated patients were more likely to receive a diagnosis of gastrointestinal morbidity and undergo gastrointestinal procedures. There were, however, no significant differences in urinary nonincontinence or incontinence diagnoses or procedures, erectile dysfunction, or hip fractures21. Kim et al. also showed that proton therapy had the highest rate of grade 3/4 toxicity among radiotherapy modalities (20.1 per 1000 person-years)19. However, the authors pointed out that the sample size for the proton cohort was quite small because the study included patients diagnosed from 1992 to 2005, a period when proton therapy was in its relative infancy and only passively scattered proton therapy (PSPT) was available. In the meantime, intensity-modulated proton therapy (IMPT) was developed both for protons29,30 and carbon ions31. More recent studies have demonstrated more favourable toxicity outcomes with proton therapy20,22,23.
Another drawback of PCPT is the irradiation of the femoral heads, which lay in the path of the beam to the prostate. These structures influence the stopping power due to their high density. This effect is even more pronounced in the presence of prosthetic hips, where proton therapy is not recommended at all. Anterior-oblique (AO) fields have been proposed to circumvent this problem with proton32,33 and carbon beams34. However, the target coverage is substantially reduced after considering interfractional variations in AO plans, thus increasing the susceptibility to underdosing33. Adaptive range verification could play a decisive role in recovering the target coverage. In these cases, superior sparing of the rectum was demonstrated. For this purpose, an in vivo diode-based range verification system was developed and commissioned at the Massachusetts General Hospital (MGH)35,36. This system was designed for passive-scattering proton delivery and relies on a matrix of 1 mm diodes mounted in a water balloon. An accuracy on the order of 1 mm for the water equivalent path length (WEPL) measurements was demonstrated. The rectal wall was shown to receive doses of 1.6% for anterior fields and 0.4% for AO fields36. Another technique to improve gastrointestinal toxicity and mitigate the uncertainties in proton relative biological effectiveness (RBE) is the use of rectal hydrogel spacers located between the prostate and the rectum37–39. This is, however, an invasive technique that demands surgery, as the spacers are later absorbed.
Endorectal balloons (ERBs) have long been used to stabilize the prostate location40,41, especially during radiotherapy with photons, such as three-dimensional conformal radiotherapy (3D CRT)40,42,43 and intensity-modulated radiotherapy (IMRT)44–47. Clinical studies using ERBs in IMRT evaluated acute toxicity48,49 and rectal wall sparing50. Reduction in the dose to the rectal wall by means of an ERB has been observed by several authors40,51,52.
Among the range verification techniques, prompt gamma spectroscopy (PGS) has demonstrated the ability to measure absolute range deviations during the course of treatment. This technique relies on the analysis of the prompt gamma energy spectrum, which is characterized by specific energy lines corresponding to the reaction channels of the irradiated protons with the elements of the human body. The most common reactions are those with oxygen and carbon atoms, which become excited and eventually emit prompt gamma rays up to 10 MeV6,53. However, other reactions emitting low-energy prompt gammas following the irradiation of metals by protons54 and helium ions55 have been shown.
We describe the implementation of an ERB inflated with a mixture of water and silicon dioxide (SiO) to wirelessly monitor the proton range in PCPT via prompt gamma rays. The presence of silicon in the mixture allows the differentiation of the characteristic gamma emission lines resulting from the bombardment of the Si atoms by protons. The cross-sections associated with such reactions have been investigated56–59. A gamma line emitted at 1.78 MeV provides a unique signature for differentiation. Instantaneous feedback from the ERB being hit by protons allows for prompt binary output regarding the irradiation of the rectum.
We irradiated single spots with beams, hitting only water or other water mixtures and solutions inside the ERB. We evaluated a worst-case scenario of an ERB surrounded by a phantom filled with water that emits prompt gamma rays under irradiation, with energy lines strongly competing with those of interest in the ERB. The water flasks in the beam path played a stronger role in signal deterioration. The prompt gamma attenuation within the water flasks placed in the path from the interaction point to the detectors was less pronounced. Finally, we evaluated two real treatment plans with a single 2 Gy anterior field and a 1 Gy AO field. A realistic phantom with an inserted ERB filled with the SiO water mixture was irradiated with active pencil beams. A scenario with the plan overlapping the ERB and another one with the real plan were considered. The iso-energy layers (IELs) crossing the balloon were identified, as well as the columns parallel to the ERB within each IEL.
Results
We started by irradiating different water solutions and mixtures with single-spot proton beams. Figure 1 shows the detectors, the targets, and the beam nozzle. Figure 1a shows an ERB filled with a water mixture to be irradiated with the lowest energy available (48 MeV). Afterwards, we increased the energy of our proton beam to an energy applicable in PCPT. Figure 1b shows two flasks of water in front of our target. To evaluate the prompt gamma attenuation in the patient, we placed two water flasks on each side of the target, i.e., in the path from the target to the detectors, as shown in Fig. 1c. Figure 1d shows a prostate phantom with a custom-made insert filled with a commercial silicone sealant. Two tungsten collimators were placed in a semi-collimation configuration in front of each detector in the beam direction to prevent scattered particles in the nozzle from hitting the detectors and to collimate the prompt gammas only from the most downstream region. These collimators had a strong impact in reducing the detector count rate, thus allowing higher beam intensities.
Figure 1.
Experimental setup. (a) Photo of the experimental setup, consisting of an irradiating nozzle, a target, and two CeBr detectors. (b) Photo of a similar setup with the target placed behind two flasks of water in the beam path. (c) That with two flasks of water placed on each side of the target. Two tungsten collimators were placed in front of each detector in the beam direction. (d) Photo of the same setup with the target of a prostate phantom with a commercial silicone insert.
In Fig. 2a, we show the energy spectra of several water solutions and mixtures irradiated by single-spot proton beams at the lowest energy. The mixture with silicon dioxide (SiO) exhibits several differences from the other solutions. The solution of heptahydrate magnesium sulphate (MgSO7HO), also known as Epsom salt, responds to higher temperatures with higher solubility. This is not observed in the SiO mixture. The addition of sodium hydroxide (NaOH) to the SiO mixture creates a solution of sodium metasilicate (NaSiO), but the quantity in grams of dissolved solute remains the same as that in the mixture. The limit for SiO, either mixed or dissolved in 60 mL of water, is 40 g. Above that quantity, the viscosity increases, and the mixture or solution cannot flow inside the small diameter tube between the syringe and the balloon. A commercial silicone sealant was also irradiated for the sake of comparison with the expected silicon gamma lines. Figure 2b shows the spectra of these targets with two water flasks placed in front of them. Due to the increased lateral spread and range straggling, the 1.78 MeV silicon gamma line is smeared out, and the nearby 1.635 MeV energy line resulting from oxygen irradiation becomes more prominent. The addition of NaOH creates a sodium line at 1.278 MeV, increases the oxygen and sodium lines at 1.635 MeV, and decreases the silicon line at 1.78 MeV. In view of these results and due to the simplicity of operation and its harmlessness (lack of toxic effects), we decided to continue our studies with a mixture of water and SiO. Figure 2c shows the spectra obtained with two water flasks placed on each side of the target. The prompt gamma attenuation effect is hardly visible. All sequential effects were combined, thus mimicking a worst-case scenario of a target mostly made of water. In this case, the prompt gamma water lines compete strongly with the prompt gamma silicon lines. Figure 2d compares the energy spectra from the proton irradiation of a prostate phantom with either a silicone insert or an ERB filled with a water mixture of SiO. The differences in the prominences of the peaks of interest are negligible.
Figure 2.
Energy spectra obtained from the irradiation of several targets by protons. (a) The targets irradiated with the minimum proton energy exhibit very prominent silicon energy peaks. A sulphur energy peak at 1.28 MeV is also visible. (b) Targets located after two flasks of water were irradiated with an energy of 113.6 MeV. The silicon energy peak at 1.78 MeV exhibits relatively low prominence for all materials and is absent for water. A sodium energy peak at 1.278 MeV is also visible. (c) The effect of the prompt gamma attenuation by the two flasks of water on each side of the target in the path to the detectors is shown for the three targets. (d) The commercial silicone insert and the balloon with the mixture of SiO inserted in the prostate phantom present similar spectra under irradiation by proton beams of 121.95 MeV.
We then aimed to evaluate the cumulative effects of range straggling and prompt gamma attenuation in a prostate phantom with an inserted ERB filled with a mixture of water and SiO. Therefore, we irradiated the prostate and the ERB with single-spot proton beams at different phantom positions. To reproduce a real treatment scenario within a rotating gantry, the phantom was rotated by 90 in the transaxial direction and irradiated by a horizontal beam. Figure 3a–c shows the phantom at three gantry angles: 0, 90, and 270. Figure 3d–f shows the spectra resulting from the irradiation of the prostate and the ERB with single spot beams in the three positions. A 1.78 MeV silicon line is present in the ERB irradiation and absent in the prostate irradiation. For the lateral beams, the closer the ERB is to the detector, the better the signal from the 1.78 MeV prompt gammas. Detector 1 collects a higher signal for the 90 angle, while detector 2 collects a higher signal for the 270 angle. To increase the signal at a 0 angle, we used the timing information of the arrival time of the protons provided by the scintillating fibres placed between the nozzle and the target60. The trigger was not further used in the treatment plans due to the strong impact in the statistics and due to the intensity constraints (increasing pile-up above p/s).
Figure 3.
Prostate phantom irradiated at different gantry angles. (a–c) Photos of the prostate phantom with an ERB filled with SiO and lying on a rotating platform to reproduce the three gantry angles of 0, 90, and 270. (d–f), Energy spectra from the irradiation of the prostate and the rectal balloon at the three angles. The silicon energy peak of 1.78 MeV is clearly distinguishable in the ERB spectra.
In the last setup, we considered real treatment-like plans. Figure 4 shows the computed tomography (CT) and the plans of an anterior beam irradiating the prostate either conformally or overlapping with the ERB. In Fig. 4a–c, the sagittal views through the prostate and the ERB clearly show their structure and the spacing between the ERB and the prostate. The CT also shows the seminal vesicles, the bladder, and the small tube inside a larger tube that transports the solution or mixture from the syringe to the ERB. Figure 4d shows a coronal plan where the IELs as well as the spots overlapping the prostate and the ERB are visible. While IEL 17 has all spots overlapping within the ERB, IEL 12 only has six central overlapping spots.
Figure 4.
Phantom computed tomography and anterior proton plans. (a) Sagittal view of a computed tomography of the prostate phantom with the ERB. The spacing between the prostate and the ERB is visible as well as the seminal vesicles and the bladder. (b–c) Sagittal view of two anterior plans either overlapping with the ERB or conforming to the prostate. (d) Coronal view of the overlapping plan where all 19 IELs are present. All spots in the distal IELs overlap with the ERB, while the proximal IELs (12–14) just overlap with its central part.
Our goal was to determine at which IEL the protons hit the ERB with the overlapping anterior beam. However, since not every spot within each IEL overlapped with the ERB, we sorted the irradiation within each IEL by columns parallel to the ERB and attributed time stamps to each column. Figure 5 shows the prompt gamma spectra from the irradiation of the phantom at IELs 12, 13, and 14. IEL 12 is at the interface between the prostate and the ERB. Columns were detected from the first to the last starting in beam-eye view (BEV) on the left for detector 1 and on the right for detector 2. While detector 1 detects the columns to the left in BEV with higher sensitivity, detector 2 has a higher count rate for columns to the right in BEV. For detector 1, we observe a 1.78 MeV silicon line emerging from column 8 to column 11 at IEL 12 (103.22 MeV). At IEL 14 (107.51 MeV), a 1.78 MeV silicon energy line emerges from column 4. For detector 2, this energy line emerges from column 4 for IEL 13 (105.42 MeV) and column 3 for IEL 14. The number of protons per column is on the order of 1–3.7 10, and the detected prompt gammas per column are on the order of 25-85 kcts.
Figure 5.
Energy spectra from the irradiation of the prostate phantom according to an anterior proton plan overlapping with the ERB. (a,b) Detector 1 detects the prompt gammas emitted from the irradiation of IELs 12 and 14. (c,d) Detector 2 detects the prompt gamma emitted from the irradiation of IELs 13 and 14. The plots are reordered into columns such that the central columns present a silicon energy line of 1.78 MeV. Detectors 1 and 2 detect the columns to the left and the right in the BEV, respectively, with higher sensitivity. For both detectors, the central columns, despite the lower counts due to attenuation, present silicon energy lines of 1.78 MeV.
In the AO plan, we also reordered each IEL of the plan in such way that they were irradiated in columns parallel to the ERB from left to right in the BEV. Figure 6a and b shows a photo of the prostate phantom at an angle of 279 and schematics of the irradiation of IEL 12 from column 1 to column 15. The plans with and without overlap with the ERB are shown in Fig. 6c and d.
Figure 6.
Anterior-oblique plan for a 279 gantry angle. (a) Photo of the prostate phantom lying on a rotating platform commanded remotely from the control room. The ERB was closer to detector 2. (b) Schematics of the irradiation of each IEL in a column-wise fashion. The Y-coordinates represent the horizontal direction and increase from left to right in the BEV, and the X-coordinates represent the vertical direction and increase from top to bottom. (c) Plans of the AO beam conforming to the prostate (d), and that overlapping with the ERB.
In Fig. 7a, we observe that the columns to the right overlapping with the ERB produce a 1.78 MeV prompt gamma line, while those to the left irradiate the prostate and therefore present no such line. Such tracking is possible with columns comprising less than 108 protons. For IEL 12, the protons start hitting the ERB at column 6 with particles. In Fig. 7b, we confirm that the real plan without overlap with the ERB does not yield a 1.78 MeV energy line for the last columns to the right. For the sake of irradiation speed, the first depicted column aggregates several columns to the left in the prostate region. An independent measurement undertaken after one month with the same gantry angle demonstrates the existence of 1.78 MeV energy lines for the columns overlapping with the ERB (Fig. 7c). An additional measurement at a symmetric position of 81 shows 1.78 MeV energy lines for the columns to the left closer to detector 1 (Fig. 7d).
Figure 7.
Energy spectra from the irradiation of the prostate phantom by an AO plan. (a) Detector 2 detects a 1.78 MeV silicon line starting at column 6 of the IEL 12 (102.6 MeV) during irradiation with 8.4 10 protons, thus indicating the entrance of protons into the ERB. (b) The real plan without overlap with the ERB presents no silicon energy peaks for the last columns of IEL 14 (108.2 MeV). (c) Another measurement separated in time by one month also shows a 1.78 MeV silicon peak for the AO beam overlapping with the ERB. (d) A symmetric configuration with an AO plan for an 81 gantry angle results in the detection of prompt gamma radiation by the first detector with the same energy lines of interest.
A peak analysis within the region of interest for the spectra presented in Fig. 7 is depicted in Fig. 8. The prominence and the width at half prominence are shown for the peaks of interest. The top four peaks that result from the irradiation of the ERB are indicative of the prompt gamma lines associated with the reaction between the protons and the silicon atoms.
Figure 8.
Analysis of the region of interest for the energy spectra obtained during irradiation of the prostate phantom by the AO plans. (a) The 1.635 MeV and 1.78 MeV peaks are more prominent in the four top spectra. These peaks correspond to columns 6 to 9 of IEL 12 (Fig. 7a). (b) The four top spectra presenting the peaks of interest correspond to columns 8 to 11 of IEL 13 (Fig. 7c). (c) The gantry angle of 81 also presents the peaks of interest for columns 7 to 9 of IEL 12 (Fig. 7d). The vertical scale has been adapted for visualization purposes.
Discussion
Prompt gamma spectroscopy (PGS) is currently one of the most promising techniques for particle range monitoring and measurements of the elemental composition of irradiated targets in particle therapy6,13,55,61. This technique facilitates absolute range measurements with millimetre precision due to accurate knowledge of the nuclear reaction cross-sections between the irradiated particles and the types of atoms in the patient. Two PGI modalities, PGS and the knife-edge slit camera, have now reached the level of clinical prototypes12,13. The combination of in vivo range monitoring and adaptation methods has been proposed for the treatment of prostate cancer with either anterior beams35 or anterior oblique (AO) beams33. An in vivo range verification system has already been commissioned36. This system is composed of a 4 by 3 array matrix of silicon diodes attached by a self-adhesive surface to an ERB and presents a WEPL measurement accuracy on the order of 1 mm.
In this paper, we propose a wireless solution that uses prompt gamma rays to monitor the interaction of protons within an ERB filled with a silicon dioxide water mixture and inserted in a prostate phantom. This concept aims to monitor the proton range in PCPT in real time. The irradiation of atomic nuclei within the human body by protons emits prompt gamma rays with characteristic energy lines6,56. The irradiation of carbon and oxygen atoms is followed by the emission of prompt gamma radiation with low and high energies (0.511 MeV, 0.718 MeV, 1.022 MeV, 1.635 MeV, 2.31 MeV, 2.8 MeV, 4.4 MeV, 5.2 MeV, and 6.1 MeV)6,54. Conversely, during the irradiation of metals, prompt gamma rays are emitted with lower energy (below 3 MeV)54,55. This radiation exits the patient under proton bombardment and may be detected by scintillating crystals, e.g., CeBr. The signals are digitally converted and processed to extract energy and time information.
Metals usually not present within the human body are good candidates for ranging probes. Although not a metal, silicon dioxide has been shown to be a good choice due to the unique signature provided by the emission of a prompt gamma energy line at 1.78 MeV. This line is distinguishable from the remaining spectrum and can therefore provide binary information about the elemental composition of the material being hit. However, even with good dose confinement to the target, the patient is still exposed to a dose in the organ at risk (OAR) and very likely prompt gammas emitted from the ERB. Therefore, a possible solution would be to set a threshold on the 1.78 MeV prompt gammas detected at a certain IEL and neighbouring IELs. This binary output might trigger a decision on whether to continue or stop/adapt the treatment since an organ at risk may be endangered.
Proton beam delivery with spot- or raster-pencil-beam scanning (PBS) is particularly suitable for such an approach. A synchronization between beam delivery and prompt gamma detection may allow real-time monitoring of the voxels being hit and simultaneous comparison to the prediction. A standard 2 Gy prostate treatment provides sufficient statistics for such monitoring. Due to the round shape of the rectum, an anterior beam requires column-wise delivery parallel to the rectum so that which IEL column the nuclear reactions with the silicon take place in can be inferred. The range monitoring also requires detectors closer to the irradiated column. Therefore, the right columns in the beam-eye view require detectors on the right side, and the left columns are better detected by detectors on the left side. The AO beams present an even more preferable solution, as the geometry allows the detectors to be placed closer to the ranging probe. All columns within IELs overlapping with the range probe are prone to be detected with higher sensitivity. In the case of a range probe located in the rectum or the oesophagus, the AO beams are especially suitable, as the detector may be positioned at right angles with the patient and close to the probe.
Range monitoring by means of PGS is feasible in PCPT. Once the proton range is under control, one may use fields other than the commonly used bilateral opposing fields that are more robust to range uncertainties. The two AO beams may assume variable angles due to the flexibility provided by the method presented in this paper. Therefore, SiO-filled ERB combined with PBS delivery and PGS monitoring may allow AO beams to be sensitive against rectal changes within and between treatment fractions.
Methods
Prostate phantom
The phantom was a prostate training phantom, CIRS model 070L (CIRS Inc., Norfolk, USA). This phantom is commonly used for ultrasound images and biopsy through the ZSkin rectal wall or perineal membrane. The main inner composition is Zerdine. It includes a urethra with a diameter of 0.7 cm, seminal vesicles with a diameter of 0.7 cm and length of 10 cm, and two lesions. The container has a volume of 9 cm 10 cm 10 cm and a probe opening of 1.2 cm.
Endorectal balloon
The endorectal balloon (ERB) was a QLRAD Rectal Pro75 (QLRAD International, Larnaca, Cyprus), which is commonly used to stabilize prostate movement in radiotherapy. It is coupled to a syringe via a smaller tube, and a latch closes the liquid flow. Each ERB was filled with 50 mL. Two ERBs operated over six shifts of six hours with different water solutions.
Water solutions and mixtures
The mixture of water and silicon dioxide (SiO) consisted of 90 mL of deionized water and 60 g of diatomaceous earth (Kieselgur) from Health Leeds (Health Leeds UK Ltd, Horeb, UK). The solution of sodium metasilicate (NaSiO) consisted of 60 mL of water, 27 g of sodium hydroxide, and 40 g of diatomaceous earth. The magnesium sulphate solution consisted of 90 mL of water and 100 g of MgSO7HO (Epsom salts) from Health Leeds previously heated to 40 C.
The HIT facility
The Heidelberg Ion-Beam Therapy Center (HIT)62 accelerates proton, helium, carbon, and oxygen ions from 48 MeV/u to 430 MeV/u. While protons and carbon ions are routinely implemented in the clinical setting, helium ions are currently being commissioned63,64, and oxygen ions remain a research beam species.
The intensities in clinical practice range from 2 10 p/s for carbon ions to 3.2 10 p/s for protons. There are two horizontal beam rooms and a 360 gantry for therapy. There is a horizontal experimental room where all the experiments referred to in this paper were performed.
Computed tomography
The computed tomography (CT) followed the routine CT protocol for ion beam therapy planning at HIT with the Siemens SOMATOM Confidence RT Pro (Siemens Healthineers, Erlangen, Germany). The phantom and the inserted ERB were scanned with a tube voltage of 120 kV, and the image was reconstructed for a field of view (FOV) of 50 cm with a convolution kernel B40s and a spacing between slices of 3 mm.
Plans
The treatment-like plans were optimized using a RayStation 10A (RaySearch Laboratories, Stockholm, Sweden) and calculated with a Monte Carlo algorithm. The plans conforming to the prostate (either with 2 oblique fields mimicking AO 279 and 81 degree beams or a single AP beam at 0 degrees) were designed to prevent any proton Bragg peak localization within the ERB and for a maximum rectum dose below 0.3 Gy (RBE) per fraction. The plans overlapping with the ERB for the same beam configurations covered an extended target including the prostate and 1.5 cm extension in the posterior direction (towards the ERB). The dose prescriptions to the targets were a median dose of 2 Gy (RBE) per fraction.
Experimental setup
The main components of the experimental setup were the nozzle, the target, and the CeBr detectors. These detectors are scintillation detectors with very good time and energy resolution. They feature a measured energy resolution of 3.49%55 and a measured time resolution of 0.85 ns60. They are mainly used for range verification of the proton and ion beams in a patient. The CeBr detectors were aligned with the isocentre and positioned at a distance of 15 cm from the beam axis. The CeBr crystals are identical in size (diameter d = 3.81 cm and length l = 7.62 cm). One crystal was coupled to a Hamamatsu R13089 photomultiplier tube (PMT), and the other crystal was coupled to a Hamamatsu R9420-100 PMT. Both detectors were plugged to a voltage divider. The anode output fed the data acquisition system (DAQ)65. This is a module of a FlashCam FADC system, originally designed for the Cherenkov Telescope Array (CTA)66.
Intensities, acquisition times, and counts
The results shown in Figs. 2 and 3 were obtained with an intensity of 2 10 p/s and the spill time lasted 1:07 min (14 spills). A total of 1.36 10 protons were delivered and the counts were variable. The results shown in Fig. 5 were obtained with an intensity of 2 10 p/s and the spill time lasted 2:59 min. A total of 3.43 10 protons were delivered and a total of 7.8 Mcts were recorded. The results shown in Fig. 7a were obtained with an intensity of 2 10 p/s and the spill time lasted 1:33 min. A total of 1.79 10 protons were delivered and a total of 5.1 Mcts were recorded. The results shown in Fig. 7b were obtained with an intensity of 2 10 p/s and the spill time lasted 1:11 min. A total of 1.36 10 protons were delivered and a total of 3.8 Mcts were recorded. The results shown in Fig. 7d were obtained with an intensity of 2 10 p/s and the spill time lasted 1:35 min. A total of 1.83 10 protons were delivered and a total of 5.7 Mcts were recorded.
Peak analysis
The presence or absence of the silicon line could not be visually verified. Therefore, a simple method was developed to identify the presence of 1.635 MeV and 1.78 MeV peaks within a region of interest. We subtracted the background from the peaks by fitting a straight line through their high and low energy values. The MATLAB function findpeaks was adapted to identify the peaks within a certain energy interval and to meet certain criteria. The parameters, such as the minimum peak height or prominence, the minimum peak width at half prominence, and the maximum and minimum distances between energy peaks, were adjusted after the spectra were properly calibrated. Other methods, such as that presented by Dal Bello et al.55, could also have been used.
Acknowledgements
P.M.M. is supported by a research fellowship for postdoctoral researchers from the Alexander von Humboldt Foundation, Bonn, Germany. H.F. is supported by Erasmus+ Programme of the European Union. P.M.M. would like to thank M.C.F. Magalhães for fruitful discussions and support with the preparation of the water solutions. The authors would like to thank German Hermann and Thomas Kihm for providing the data acquisition system. The authors would like to thank Semi Harrabi for suggesting the use of the ERB and to Nami Saito for proposing the use of silicon as a ranging probe. The authors thank Arjen Winkel for providing the ERBs. The authors thank the staff of the DKFZ, in particular Gernot G. Echner, Armin Runz, and Ruediger Schmidt for the technical support. The authors also thank the radiation protection department of the DKFZ, in particular Mechthild Kaemmer, for the support with calibration sources.
Author contributions
P.M.M. and J.S. conceptualized the project. P.M.M. and H.F. performed experiments and acquired data. B.A. performed the computed tomography and supported the experiments. T.T. prepared the treatment plans. S.B. provided continuous support throughout all experiments. P.M.M., H.F., T.T., B.A., S.B. and J.S. discussed content and researched data. P.M.M processed and analysed the data. P.M.M. and T.T. prepared figures. J.S. acquired funding and supervised the project.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Competing interests
The international patent application number PCT/EP2020/070128 for a PGS system including the features presented in this work has been filed with J.S. and P.M.M. as inventors. H.F., T.T., B.A. and S.B. declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Paulo Magalhaes Martins, Email: pjmartins@fc.ul.pt.
Joao Seco, Email: j.seco@dkfz.de.
References
- 1.Paganetti H. Range uncertainties in proton therapy and the role of Monte Carlo simulations. Phys. Med. Biol. 2012;57:R99–R117. doi: 10.1088/0031-9155/57/11/R99. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Parodi K, Polf J. In vivo range verification in particle therapy. Med. Phys. 2018;45:e1036–e1050. doi: 10.1002/mp.12764. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Moteabbed M, España S, Paganetti H. Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy. Phys. Med. Biol. 2011;56:1063–1082. doi: 10.1088/0031-9155/56/4/012. [DOI] [PubMed] [Google Scholar]
- 4.Ferrero V, et al. Online proton therapy monitoring: Clinical test of a Silicon-photodetector-based in-beam PET. Sci. Rep. 2018;8:4100. doi: 10.1038/s41598-018-22325-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Smeets J, et al. Prompt gamma imaging with a slit camera for real-time range control in proton therapy. Phys. Med. Biol. 2012;57:3371–3405. doi: 10.1088/0031-9155/57/11/3371. [DOI] [PubMed] [Google Scholar]
- 6.Verburg J, Seco J. Proton range verification through prompt gamma-ray spectroscopy. Phys. Med. Biol. 2014;59:7089–7106. doi: 10.1088/0031-9155/59/23/7089. [DOI] [PubMed] [Google Scholar]
- 7.Golnik C, et al. Range assessment in particle therapy based on prompt -ray timing measurements. Phys. Med. Biol. 2014;59:5399–5422. doi: 10.1088/0031-9155/59/18/5399. [DOI] [PubMed] [Google Scholar]
- 8.Krimmer, J., Dauvergne, D., Létang, J. & Testa, É. Prompt-gamma monitoring in hadrontherapy: A review. Nucl. Instrum. Methods A878, 58–73 (2018).
- 9.Draeger E, et al. 3D prompt gamma imaging for proton beam range verification. Phys. Med. Biol. 2018;63:035019. doi: 10.1088/1361-6560/aaa203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Wronska, A. & Dauvergne, D. Range verification by means of prompt-gamma detection in particle therapy. Radiation Detection Systems, In press. hal-03085504 (2021).
- 11.Magalhaes Martins, P., Dal Bello, R. & Seco, J. in Radiation Therapy Dosimetry: A Practical Handbook (1st ed.) (ed. Darafsheh, A.) Ch. 27 (CRC Press, 2021).
- 12.Richter C, et al. First clinical application of a prompt gamma based in vivo proton range verification system. Radiother. Oncol. 2016;118:232–237. doi: 10.1016/j.radonc.2016.01.004. [DOI] [PubMed] [Google Scholar]
- 13.Hueso-González F, Rabe M, Ruggieri TA, Bortfeld T, Verburg J. A full-scale clinical prototype for proton range verification using prompt gamma-ray spectroscopy. Phys. Med. Biol. 2018;63:185019. doi: 10.1088/1361-6560/aad513. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Slater JD, et al. Proton therapy for prostate cancer: The initial Loma Linda University experience. Int. J. Radiat. Oncol. Biol. Phys. 2004;59:348–352. doi: 10.1016/j.ijrobp.2003.10.011. [DOI] [PubMed] [Google Scholar]
- 15.Ishikawa H, et al. Particle therapy for prostate cancer: The past, present and future. Int. J. Urol. 2019;26:971–979. doi: 10.1111/iju.14041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Storey MR, et al. Complications from radiotherapy dose escalation in prostate cancer: Preliminary results of a randomized trial. Int. J. Radiat. Oncol. Biol. Phys. 2000;48:635–642. doi: 10.1016/S0360-3016(00)00700-8. [DOI] [PubMed] [Google Scholar]
- 17.Pollack A, et al. Prostate cancer radiation dose response: Results of the M.D. Anderson phase III randomized trial. Int. J. Radiat. Oncol. Biol. Phys. 2002;53:1097–1105. doi: 10.1016/S0360-3016(02)02829-8. [DOI] [PubMed] [Google Scholar]
- 18.Mock, U., Bogner, J., Georg, D., Auberger, T. & Pötter, R. Comparative treatment planning on localized prostate carcinoma. Strahlenther. Onkol.181, 448–455 (2005). [DOI] [PubMed]
- 19.Kim S, et al. Late gastrointestinal toxicities following radiation therapy for prostate cancer. Eur. Urol. 2011;60:908–916. doi: 10.1016/j.eururo.2011.05.052. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Nihei K, et al. Multi-institutional phase II study of proton beam therapy for organ-confined prostate cancer focusing on the incidence of late rectal toxicities. Int. J. Radiat. Oncol. Biol. Phys. 2011;81:390–396. doi: 10.1016/j.ijrobp.2010.05.027. [DOI] [PubMed] [Google Scholar]
- 21.Sheets NC, et al. Intensity-modulated radiation therapy, proton therapy, or conformal radiation therapy and morbidity and disease control in localized prostate cancer. JAMA. 2012;307:1611–1620. doi: 10.1001/jama.2012.460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Mendenhall NP, et al. Five-year outcomes from 3 prospective trials of image-guided proton therapy for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2014;88:596–602. doi: 10.1016/j.ijrobp.2013.11.007. [DOI] [PubMed] [Google Scholar]
- 23.Bryant C, et al. Five-year biochemical results, toxicity, and patient-reported quality of life after delivery of dose-escalated image guided proton therapy for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2016;95:422–434. doi: 10.1016/j.ijrobp.2016.02.038. [DOI] [PubMed] [Google Scholar]
- 24.Akakura K, et al. Phase I/II clinical trials of carbon ion therapy for prostate cancer. The Prostate. 2004;58:252–258. doi: 10.1002/pros.10328. [DOI] [PubMed] [Google Scholar]
- 25.Fukahori M, et al. Estimation of late rectal normal tissue complication probability parameters in carbon ion therapy for prostate cancer. Radiother. Oncol. 2016;118:136–140. doi: 10.1016/j.radonc.2015.11.023. [DOI] [PubMed] [Google Scholar]
- 26.Vargas C, et al. Dose-volume comparison of proton therapy and intensity-modulated radiotherapy for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2008;70:744–751. doi: 10.1016/j.ijrobp.2007.07.2335. [DOI] [PubMed] [Google Scholar]
- 27.Trofimov A, et al. Radiotherapy treatment of early-stage prostate cancer with IMRT and protons: A treatment planning comparison. Int. J. Radiat. Oncol. Biol. Phys. 2007;69:444–453. doi: 10.1016/j.ijrobp.2007.03.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Chera BS, et al. Dosimetric study of pelvic proton radiotherapy for high-risk prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2009;75:994–1002. doi: 10.1016/j.ijrobp.2009.01.044. [DOI] [PubMed] [Google Scholar]
- 29.Lomax A. Intensity modulation methods for proton radiotherapy. Phys. Med. Biol. 1999;44:185–205. doi: 10.1088/0031-9155/44/1/014. [DOI] [PubMed] [Google Scholar]
- 30.Mohan R, Grosshans D. Proton therapy - Present and future. Adv. Drug Deliv. Rev. 2017;109:26–44. doi: 10.1016/j.addr.2016.11.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Debus J, et al. Bestrahlung von Schädelbasistumoren mit Kohlenstoffionen bei der GSI Erste klinische Ergebnisse und zukünftige Perspektiven. Strahlenther. Onkol. 2000;176:211–216. doi: 10.1007/s000660050002. [DOI] [PubMed] [Google Scholar]
- 32.Cuaron JJ, et al. Anterior-oriented proton beams for prostate cancer: A multi-institutional experience. Acta Oncol. 2015;54:868–874. doi: 10.3109/0284186X.2014.986288. [DOI] [PubMed] [Google Scholar]
- 33.Moteabbed M, et al. Proton therapy of prostate cancer by anterior-oblique beams: Implications of setup and anatomy variations. Phys. Med. Biol. 2017;62:1644–1660. doi: 10.1088/1361-6560/62/5/1644. [DOI] [PubMed] [Google Scholar]
- 34.Kubota Y, et al. Changes in rectal dose due to alterations in beam angles for setup uncertainty and range uncertainty in carbon-ion radiotherapy for prostate cancer. PLOS ONE. 2016;11:1–11. doi: 10.1371/journal.pone.0153894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bentefour EH, et al. Validation of an in-vivo proton beam range check method in an anthropomorphic pelvic phantom using dose measurements. Med. Phys. 2015;42:1936–1947. doi: 10.1118/1.4915923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Hoesl M, et al. Clinical commissioning of an in vivo range verification system for prostate cancer treatment with anterior and anterior oblique proton beams. Phys. Med. Biol. 2016;61:3049–3062. doi: 10.1088/0031-9155/61/8/3049. [DOI] [PubMed] [Google Scholar]
- 37.Rucinski A, et al. Ion therapy of prostate cancer: Daily rectal dose reduction by application of spacer gel. Radiat. Oncol. 2015;10:56. doi: 10.1186/s13014-015-0348-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Underwood TSA, et al. Hydrogel rectum-prostate spacers mitigate the uncertainties in proton relative biological effectiveness associated with anterior-oblique beams. Acta Oncol. 2017;56:575–581. doi: 10.1080/0284186X.2016.1275781. [DOI] [PubMed] [Google Scholar]
- 39.Dinh T-KT, et al. Rectal hydrogel spacer improves late gastrointestinal toxicity compared to rectal balloon immobilization after proton beam radiation therapy for localized prostate cancer: A retrospective observational study. Int. J. Radiat. Oncol. Biol. Phys. 2020;108:635–643. doi: 10.1016/j.ijrobp.2020.01.026. [DOI] [PubMed] [Google Scholar]
- 40.D’Amico AV, et al. A practical method to achieve prostate gland immobilization and target verification for daily treatment. Int. J. Radiat. Oncol. Biol. Phys. 2001;51:1431–1436. doi: 10.1016/S0360-3016(01)02663-3. [DOI] [PubMed] [Google Scholar]
- 41.McGary JE, Teh BS, Butler EB, Grant W. Prostate immobilization using a rectal balloon. J. Appl. Clin. Med. Phys. 2002;3:6–11. doi: 10.1120/jacmp.v3i1.2590. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.van Lin ENT, Hoffmann AL, van Kollenburg P, Leer JW, Visser AG. Rectal wall sparing effect of three different endorectal balloons in 3D conformal and IMRT prostate radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2005;63:565–576. doi: 10.1016/j.ijrobp.2005.05.010. [DOI] [PubMed] [Google Scholar]
- 43.van Lin EN, et al. Reduced late rectal mucosal changes after prostate three-dimensional conformal radiotherapy with endorectal balloon as observed in repeated endoscopy. Int. J. Radiat. Oncol. Biol. Phys. 2007;67:799–811. doi: 10.1016/j.ijrobp.2006.09.034. [DOI] [PubMed] [Google Scholar]
- 44.Bastasch M, et al. Tolerance of endorectal immobilization balloon in 396 patients treated with intensity-modulated radiation therapy (IMRT) for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2002;54:270. doi: 10.1016/S0360-3016(02)03523-X. [DOI] [PubMed] [Google Scholar]
- 45.Hardcastle N, Metcalfe PE, Rosenfeld AB, Tomé WA. Endo-rectal balloon cavity dosimetry in a phantom: Performance under IMRT and helical tomotherapy. Radiother. Oncol. 2009;92:48–56. doi: 10.1016/j.radonc.2009.03.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Smeenk RJ, Teh BS, Butler EB, van Lin EN, Kaanders JH. Is there a role for endorectal balloons in prostate radiotherapy? A systematic review. Radiother. Oncol. 2010;95:277–282. doi: 10.1016/j.radonc.2010.04.016. [DOI] [PubMed] [Google Scholar]
- 47.Smeenk RJ, Hopman WP, Hoffmann AL, van Lin EN, Kaanders JH. Differences in radiation dosimetry and anorectal function testing imply that anorectal symptoms may arise from different anatomic substrates. Int. J. Radiat. Oncol. Biol. Phys. 2012;82:145–152. doi: 10.1016/j.ijrobp.2010.08.023. [DOI] [PubMed] [Google Scholar]
- 48.Teh BS, et al. Intensity-modulated radiation therapy (IMRT) for prostate cancer with the use of a rectal balloon for prostate immobilization: acute toxicity and dose-volume analysis. Int. J. Radiat. Oncol. Biol. Phys. 2001;49:705–712. doi: 10.1016/S0360-3016(00)01428-0. [DOI] [PubMed] [Google Scholar]
- 49.Teh, B. S. et al. Clinical experience with intensity-modulated radiation therapy (IMRT) for prostate cancer with the use of rectal balloon for prostate immobilization. Med. Dosim.27, 105–113 (2002). [DOI] [PubMed]
- 50.Teh BS, et al. Rectal wall sparing by dosimetric effect of rectal balloon used during Intensity-Modulated Radiation Therapy (IMRT) for prostate cancer. Med. Dosim. 2005;30:25–30. doi: 10.1016/j.meddos.2004.10.005. [DOI] [PubMed] [Google Scholar]
- 51.Patel RR, Orton N, Tomé WA, Chappell R, Ritter MA. Rectal dose sparing with a balloon catheter and ultrasound localization in conformal radiation therapy for prostate cancer. Radiother. Oncol. 2003;67:285–294. doi: 10.1016/S0167-8140(03)00056-2. [DOI] [PubMed] [Google Scholar]
- 52.Dubouloz A, et al. Urethra-sparing stereotactic body radiotherapy for prostate cancer: how much can the rectal wall dose be reduced with or without an endorectal balloon? Radiat. Oncol. 2018;13:114. doi: 10.1186/s13014-018-1059-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Kelleter L, et al. Spectroscopic study of prompt-gamma emission for range verification in proton therapy. Phys. Med. 2017;34:7–17. doi: 10.1016/j.ejmp.2017.01.003. [DOI] [PubMed] [Google Scholar]
- 54.Martins PM, et al. Prompt gamma spectroscopy for range control with CeBr. CDBME. 2017;3:113–117. [Google Scholar]
- 55.Dal Bello R, et al. Results from the experimental evaluation of CeBr scintillators for He prompt gamma spectroscopy. Med. Phys. 2019;46:3615–3626. doi: 10.1002/mp.13594. [DOI] [PubMed] [Google Scholar]
- 56.Foley K, Clegg A, Salmon G. Gamma-radiation from the medium energy proton borbardment of sodium, magnesium, aluminium, silicon phosphorus and sulphur. Nucl. Phys. 1962;37:23–44. doi: 10.1016/0029-5582(62)90243-2. [DOI] [Google Scholar]
- 57.Kozlovsky B, Murphy RJ, Ramaty R. Nuclear deexcitation gamma-ray lines from accelerated particle interactions. ApJS. 2002;141:523–541. doi: 10.1086/340545. [DOI] [Google Scholar]
- 58.Boromiza M, et al. Nucleon inelastic scattering cross sections on O and Si. Phys. Rev. C. 2020;101:024604. doi: 10.1103/PhysRevC.101.024604. [DOI] [Google Scholar]
- 59.Kiener J, et al. Gamma-ray emission in alpha-particle reactions with C, Mg, Si, Fe. J. Phys. Conf. Ser. 2020;1555:012011. doi: 10.1088/1742-6596/1555/1/012011. [DOI] [Google Scholar]
- 60.Magalhaes Martins P, et al. A single-particle trigger for time-of-flight measurements in prompt-gamma imaging. Front. Phys. 2020;8:169. doi: 10.3389/fphy.2020.00169. [DOI] [Google Scholar]
- 61.Magalhaes Martins P, et al. PIBS: Proton and ion beam spectroscopy for in vivo measurements of oxygen, carbon, and calcium concentrations in the human body. Sci. Rep. 2020;10:7007. doi: 10.1038/s41598-020-63215-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Haberer T, et al. The Heidelberg ion therapy center. Radiother. Oncol. 2004;73:S186–S190. doi: 10.1016/S0167-8140(04)80046-X. [DOI] [PubMed] [Google Scholar]
- 63.Tessonnier T, et al. Proton and helium ion radiotherapy for meningioma tumors: A Monte Carlo-based treatment planning comparison. Radiat. Oncol. 2018;13:2. doi: 10.1186/s13014-017-0944-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Mein S, et al. Biophysical modeling and experimental validation of relative biological effectiveness (RBE) for He ion beam therapy. Radiat. Oncol. 2019;14:123. doi: 10.1186/s13014-019-1295-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Werner F, et al. Performance verification of the FlashCam prototype camera for the Cherenkov Telescope Array. Nucl. Instrum. Methods A. 2017;876:31–34. doi: 10.1016/j.nima.2016.12.056. [DOI] [Google Scholar]
- 66.The CTA Consortium et al. Design concepts for the Cherenkov Telescope Array CTA: An advanced facility for ground-based high-energy gamma-ray astronomy. Exp. Astron.32, 193–316 (2011).