Diagnostic applications of simultaneously acquired dual-isotope single-photon emission CT scans.

D Mathews 1, B S Walker 1, B C Allen 1, H Batjer 1, P D Purdy 1
PMCID: PMC8332071  PMID: 8141067

Abstract

PURPOSE

To report the development and validation of a technique of dual tracer single-photon emission CT brain imaging using technetium-99m hexamethylpropyleneamine oxime and iodine-123 iodoamphetamine agents and the application of this technique in patients with a variety of diagnoses.

METHODS

Contamination between the two isotopes' energy windows was calculated by opening both energy windows while scanning a group of patients using a single isotope. To compare uniformity of I-123 down-scatter, Tc-99m studies were performed both before and after the administration of I-123 in five of 24 dual studies. The 24 patients studied with the dual-isotope technique were evaluated during acetazolamide testing, trial balloon occlusion, or embolization of an arteriovenous malformation.

RESULTS

In a dual acquisition, average count contamination of an I-123 study by Tc-99m was less than 1% of the total I-123 counts, and contamination of a Tc-99m study by I-123 was approximately 12% of the total Tc-99m counts. Tc-99m studies performed both before and after the administration of I-123 demonstrated that contaminating counts do not adversely affect scan interpretation. Dual-tracer scans were completed in all 24 patients, 10 of whom showed changes after intervention.

CONCLUSIONS

Dual-tracer single-photon emission CT brain scans of adequate diagnostic quality are possible using Tc-99m and I-123.