Blood flow in major cerebral arteries measured by phase-contrast cine MR.

D R Enzmann 1, M R Ross 1, M P Marks 1, N J Pelc 1
PMCID: PMC8332086  PMID: 8141043

Abstract

PURPOSE

To measure mean blood flow in individual cerebral arteries (carotid, basilar, anterior cerebral, middle cerebral, and posterior cerebral) using a cine phase contrast MR pulse sequence.

METHODS

Ten healthy volunteers (22 to 38 years of age) were studied. The cine phase-contrast section was positioned perpendicular to the vessel of interest using oblique scanning planes. This pulse sequence used a velocity encoding range of 60 to 250 cm/sec. From the velocity and area measurements on the cine images, mean blood flow was calculated in milliliters per minute and milliliters per cardiac cycle. In the same subjects, transcranial Doppler measurements of blood velocity in these same vessels were also obtained.

RESULTS

There was no difference in blood flow in the paired cerebral arteries. Carotid arteries had mean blood flow in the range of 4.8 +/- 0.4 ml/cycle, the basilar artery 2.4 +/- 0.2 ml/cycle, the middle cerebral artery 1.8 +/- 0.2 ml/cycle, the distal anterior cerebral artery 0.6 +/- 0.1 ml/cycle, and the posterior cerebral artery 0.8 +/- 0.1 ml/cycle. Overall, there was poor correlation between MR-measured and transcranial Doppler-measured peak velocity.

CONCLUSION

Although careful attention to technical detail is required, mean blood flow measurements in individual cerebral vessels is feasible using a cine phase-contrast MR pulse sequence.