Skip to main content
AJNR: American Journal of Neuroradiology logoLink to AJNR: American Journal of Neuroradiology
. 1982 Sep-Oct;3(5):561-5.

Dynamic CT Scanning of Spinal Column Trauma

Ben Maurice Brown 1,2, Michael Brant-Zawadzki 1,2, Christopher E Cann 2
PMCID: PMC8333133

Abstract

Dynamic sequential computed tomographic scanning with automatic table incrementation uses low milliampere-second technique to eliminate tube cooling delays between scanning slices and, thus, markedly shortens examination times. A total of 25 patients with spinal column trauma involving 28 levels were studied with dynamic scans and retrospectively reviewed. Dynamic studies were considerably faster than conventional spine examinations and yielded reliable diagnoses. Bone disruption and subluxation was accurately evaluated, and the use of intrathecal metrizamide in low doses allowed direct visualization of spinal cord or radicular compromise. Multiplanar image reformation was aided by the dynamic incrementation technique, since motion between slices (and the resulting misregistration artifact on image reformation) was minimized. A phantom was devised to test spatial resolution of computed tomography for objects 1–3 mm in size and disclosed minimal differences for dynamic and conventional computed tomographic techniques in resolving medium-to-high–contrast objects.

Full Text

The Full Text of this article is available as a PDF (2.4 MB).


Articles from AJNR: American Journal of Neuroradiology are provided here courtesy of American Society of Neuroradiology

RESOURCES